FW Fisher Wallace Stimulator Instruction Manual

!
1
THE FISHER WALLACE STIMULATOR FOR THE TREATMENT OF PAIN ON THE BODY
INSTRUCTION MANUAL. Call us with any questions: (800) 692 - 4380
The Fisher Wallace Stimulator is a portable, battery-powered device which has been cleared by the FDA for symptomatic relief of chronic pain, post-traumatic acute pain and post-surgical pain when used on the body. The device is also cleared by the FDA for the treatment of depression, anxiety and insomnia. Although the device poses no significant risks to health, patients should be monitored by their physician at least once every three months while using the device; more frequent monitoring is recommended. Do not attempt to reduce or cease use of other medication without first consulting your doctor.
The device delivers a micro-electrical stimulus (1-4 mA) that is carried by electrical cables to wet sponges (supplied with the device) that are applied to the skin. When treating site-specific pain on the body, the electrodes are placed on the body. As clearly illustrated in this manual on pages 6-12, the red wire electrode is typically placed at the rear base of the neck and the black wire electrode is placed on the pain site (with the exception of gastro pain treatment). The electrodes are secured to the body with Velcro body straps, which are included in the kit.
Read section 10.0 APPLICATION OF THE DEVICE ON THE BODY for instructions for use, as well as the Treatment Diagrams on pages 6-12 for guidance regarding the treatment of Back Pain, Shoulder Pain, Gastro Pain, Elbow Pain, Phantom Limb Pain, Arthritic Pain and Neck Pain.
Please email us at info@fisherwallace.com or call us at (800) 692 - 4380 with any questions.
2.0 CONTRAINDICATIONS
The device is contraindicated for use in patients who have demand or sensing type cardiac pacemakers.
3.0 WARNINGS
Keep the device out of the reach of children. Electronic monitoring equipment (such as ECG monitors, ECG alarms) may not operate properly when CES stimulation is in use. High Frequency surgical equipment may not be used when CES stimulation is in use. Do not operate the CES stimulator in close proximity to short-wave or microwave therapy equipment. This device is not suitable for use with oxygen or in the presence of a flammable anesthetic mixture with air or oxygen, or with nitrous oxide.
4.0 PRECAUTIONS
This device should not be used on the throat or neck, except the rear base of the neck as indicated in the diagrams for treating pain (see the Chronic Pain Instruction Manual). This device should not be used on the eyes. This device should not be used on or near areas of the body, including the head, that contain implanted devices, such as stimulators, stents, and active or inactive implants such as deep brain stimulators and vagus nerve stimulators. Patients should avoid using the device near areas of the body where there is embedded shrapnel or metal plates. There is no danger in using the device if you have dental fillings. Do not allow water to enter this device. The Fisher Wallace Stimulator should not be exposed to environmental conditions where the system may get wet.
5.0 ADVERSE REACTIONS (VERY RARE)
Skin irritation may occur at the site of sponge electrode placement, especially if the sponges deteriorate (sponges should be replaced EVERY TWO WEEKS) or if contact is made between the skin and the metal part of the sponge receptacle, or if the sponges are not thoroughly wet before each use. A mild headache or dizziness may occur, and should cease after you stop using the device. Such reactions are very rare.
!
2
6.0 FEDERAL LABELING
Federal Law (USA) states that this device must only be purchased by or under the order of a health care practitioner, licensed in the state in which he/she practices.
7.0 INDICATIONS FOR USE
The Fisher Wallace Stimulator is a portable battery powered pulse generator that may be used for symptomatic relief of chronic pain, post-traumatic acute pain and post-surgical pain. Your physician will prescribe the appropriate duration to utilize the Fisher Wallace Stimulator. Please refer to the enclosed diagrams regarding electrode placement and standard treatment durations. Typically, patients should use the device on the body
two to three times a day for twenty to forty minutes per treatment session, at Level #3 or #4.
8.0 DEVICE CONTROLS
The Fisher Wallace Stimulator has only one knob that encompasses both the ON/OFF switch for turning the device ON or OFF as well as adjusting the level of current.
There is a green LED indicator light that flashes when the device is turned ON.
There are four (4) yellow LED indicator lights that flash according to the intensity of stimulation. Level #2 is typically used when the sponge electrodes are placed on the head, and level #3 and #4 are typically used when the sponge electrodes are placed on the body.
Both the green and yellow lights flicker when lit – this is normal.
9.0 SKIN PREPARATION
Good skin care is important in minimizing any skin irritation, which may be encountered with the active use of the electrodes. Thoroughly wash the skin sites where the electrodes will be placed.
10.0 APPLICATION OF THE DEVICE ON THE BODY
To treat pain on a particular area of the body, such as the lower back or elbow, apply the red-lead sponge electrodes to the rear base of the neck (except when treating gastro pain) and the black-lead electrode to the pain area. Use the Velcro straps to hold the electrodes in place. Refer to the body diagrams below on pages 6-12 for electrode placement and application. If you have any questions, please call or email us.
10.1 Electrode Preparation
New headsets come with new sponges (dehydrated yellow disks) already installed in the white sponge receptacles. The sponges will expand when submerged in water. You may wet the sponges separately or while they are in the plastic receptacles. The entire plastic sponge receptacle, including the wire, may be dunked into water. The wet sponges should fit into the white receptacles in such a way that the edge of the receptacles overlaps the sides of the sponges.
Snap the red and black wires onto the receptacles and then connect the other ends of the wires into the stimulator. Black and red colors should be matched.
After use, sponges will dry more quickly if removed from the receptacles and allowed to dry by themselves.
SPONGES SHOULD BE REPLACED EVERY 2-4 WEEKS FOR OPTIMAL PERFORMANCE.
Replacement sponges may be purchased on www.FisherWallace.com.
10.2 Electrode Placement
Please refer to the diagrams on pages 6-12 regarding electrode placement and standard treatment durations.
!
3
Make sure the sponges are wet. Use the enclosed Velcro body straps to secure the red wire electrode to the rear base of the neck (see below) and the black wire electrode to the area of the body that is experiencing pain as indicated in the diagrams.
Below is an example of how to use the Velcro body straps to secure the sponge electrodes. You may use other kinds of straps if you prefer, as long as the sponge electrodes are secure:
Velcro Strap on base of neck Red Wire sponge electrode tucked
beneath strap, wet sponge against skin
Velcro Strap on elbow Black Wire sponge electrode tucked beneath strap, wet sponge against skin
10.3 Treatment Duration
When used on the body, the device may be used 2-3 times a day, for 20-40 minutes per treatment session. The device will turn off automatically after 20 minutes. To reactivate the device for another 20 minutes, turn the dial clockwise to the OFF position until you hear an audible click, and then turn the device back on.
10.5 Device cleaning
The device and headset should be thoroughly cleaned every two weeks. To clean the device, remove the black and red wires from the base unit, remove the old sponges from the sponge electrodes, dampen a cloth or paper towel with alcohol and clean the inside and outside of the plastic electrode receptacles in a circular motion. Insert a new pair of sponges and dispose of used sponges. To clean the base unit, dampen a cloth or a paper towel
!
4
with alcohol and wipe down the base unit.
Once clean, reinsert the wires into the base unit, matching the wire colors to the circles on the base unit. Now you can continue to use the device.
11.0 BATTERY REPLACEMENT
In order to replace the battery, remove the battery compartment cover on the back of the device by sliding the cover into the open position. Use only high quality AA batteries.
12.0 TROUBLE SHOOTING
If your device turns off before the 20 minutes are up, or when the device is set down on a table or lightly jarred or tapped, re-insert the batteries. Sometimes the batteries will misalign with the metal receptors in the
battery compartment and as a result, the device will turn off when lightly jarred or tapped.
If the green “ON” light does not illuminate when the device is turned ON, first turn the device completely OFF by rotating the dial clockwise until you hear a click, then try turning the device ON again. Replace the batteries with fresh ones if you still have trouble.
If the yellow indicator lights (1-4) do not illuminate, be sure the sponges are new, clean and thoroughly wet. The hair / skin beneath the sponge electrodes must also be sufficiently wet. People with thick hair may have more difficulty wetting the hair enough to cause the electrodes to conduct. Some people with small heads will require a tighter headband to make the electrodes work properly – adequate pressure is needed against the electrodes by the headband.
To test if the electrodes are working, press the two sponge electrodes together (like an “Oreo cookie”), turn the device ON (green light flashing), and, while continuing to squeeze the sponge electrodes together, continue dialing up (counterclockwise) until you see the yellow lights illuminate.
If the yellow lights illuminate when you test them but not when you have them in use on your head or body, then the wetness of the sponges / hair / skin is insufficient or the tightness of the headband or strap holding the electrodes in place may be insufficient. Make sure the headband or strap provides a snug fit. If the headband or strap is not pressing the electrodes firmly enough, adequate conductivity may not occur. Press the wet electrodes firmly while the device is on and see if this causes the yellow lights to illuminate. If they illuminate, you will need a smaller headband or to tighten the strap.
If your device beeps throughout the treatment session, or fails to beep at the end of the session, replace the batteries.
If, after trying the above suggestions, you are still having trouble with your device, call us at (800) 692-4380 or email us at info@fisherwallace.com. If we cannot fix your problem over the phone, we will repair or replace your device.
If you are required to return your device for repair or replacement, you will be responsible for shipping it to us. We will repair or replace the device for free if it is still under warranty, and ground ship it back to you at no cost. If you would like the repaired or replaced device shipped overnight to you, you will be charged $58.95 for expedited shipping. In-person drop offs by customers are not accepted at our administrative offices in New York City.
13.0 THEORY OF OPERATION
The electronic waveform contains a 15,000Hz square wave carrier which is rectified, varying from zero to a maximum of 4 milliamperes. The first modulating signal of 15Hz provides an "ON" time of 50 milliseconds and an "OFF " time of 16.7 milliseconds. The second modulating signal of 500Hz changes the "ON" time series of 15,000Hz carrier pulses (750 pulses in 50 milliseconds) into 25 smaller bursts of 15 pulses each of the 15,000Hz carrier signal 375 pulses in the same 50 milliseconds. The subject device is a bipolar version of a TENS device, wherein the first major burst of energy (50 milliseconds is positive [above the zero axis], followed by a 16.7 millisecond "OFF " time, is then followed by a second major burst of energy (50 milliseconds is negative [below the axis]) followed by a 16.7 millisecond "OFF " time. Thus, the consecutive positive burst and OFF time is followed by an equal and opposite negative burst and OFF time, balancing the direct current component to zero.
Loading...
+ 9 hidden pages