11 | P a g e
No suction in chemical
intake tubes
As water flows through the unit, the
chemical intake tubes should begin to fill. The
product intake tube
should fill within
seconds. The activator intake tube will take longer
to fill, approximately 15-30 seconds
.
Verify suction by lifting each intake tube one at a
time from its supply drum, it should
begin to suction in
air, creating small air bubbles in
the intake tube. If this does not happen, follow
the steps below until the
problem is resolved.
1. Check the product levels in the
chemical supply drums.
2. Make sure the orifice tip at each intake
point is not clogged or crystallized.
3. Make sure each foot valve, inline check valve
, and intake tube is not clogged or damaged.
4. Check to see that water supply line meets p
si and flow requirements.
If the above steps do not facilitate
suction through the intake tubes, one or both sides of
the venturi intake
assembly
may need to be repaired or replaced.
No yellow color present
If the directions for setting up the
unit have been followed, but the product shows no sign
of activation within
5 minutes
, follow the steps below until the problem is resolved.
1. Verify suction in the product in
take tube and the activator intake tube.
Read "No suction in chemical
intake
tubes" for help verifying suction and the
steps to be taken for correcting.
2. Check the pH of the sample
. Make sure the pH of the solution is within the range of
2.0 to 3.0 or as
specified by application
. To do this, simply remove the 4" access port
plug on the top of the lid. Slid
e a
plastic cup through the access port and under the
solution outlet inside the solution tan
k. Collect
approximately 6 ounces of sample and
turn the control valve to the "off" position, then
check the pH. Set
the sample aside for
about 5 minutes. Observe the sample. The
solution should have a slight yellow color
present.
If the color is present, activation
is occurring and the factory product parameters are adequate
.
If the pH of
the solution is
out of range, read "Changing pH of activated solution" to correct.
Changing the pH of the
activated solution
Before making any changes to the
pH or concentration, check the QC sheets that accompany
this manual
for the factory preset
pH and concentration levels of this unit. Note the color o
f the
orifice to be changed.
Make no more
than one orifice size changes up or down between
adjustments (see
color orifice
chart
in
orifice pack).
The Alpha is a venturi based system
that draws chemicals through orifices to control pH
and
concentration
(ppm). Therefore,
when a change of orifice is made on one side of the vent
uri
assembly,
the
other side is
slightly
affected as well. For example, if a change is made to a
larger
size orifice in the activator side of
the