Fukuda Denshi Co LX8300M User Manual

Thank you for purchasing our product. Before using this product, read this operation manual thoroughly for correct handling and operation.
Safety Precautions
The safety precautions shown in this manual contain important details on the safe use of this product, and must be obeyed. Symbols and their meanings are shown below. Make sure to understand the following before reading the rest of the manual.
Indicates a potentially hazardous situation which, if
DANGER
WARNING
CAUTION
NOTE
Precaution from Fukuda Denshi
Fukuda Denshi is liable for the safety, reliability, and performance of its equipment only if; Maintenance, modifications, and repairs are carried out by authorized personnel. Components are used in accordance with Fukuda Denshi operating instructions. If the equipment is used incorrectly and become unusable, Fukuda Denshi is not liable for the malfunction. Use the equipment only for the purpose specified in this manual.
not avoided, will result in death, serious injury, or fire.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury, or property damage.
A note is not related to product safety, but provides information about the correct use and operating procedures to prevent incorrect operation and malfunction of the equipment.
i
Intended Use of this Equipment
This equipment is designed for the following <Intended Use>.
<Intended Use> This equipment is intended for patient monitoring in surgery room, ICU, ward, emergency room in the medical facility by measuring ECG, respiration and SpO and transmitting the measured data by wireless network to the central monitor continuously.
This equipment is intended to be used by healthcare professionals. Users should have a thorough knowledge of the function and operation before using this equipment. The maintenance of this equipment should be performed by skilled personnel who received a training of possible hazards and measures to avoid those hazards. Also, your local regulation must be followed. If this equipment is used for the purpose other than intended, or if the user does not follow the safety instructions, the following hazard may result.
Hazard to the Life and Health of the Patient or the User A Problem Related to Medical Practice Damage to the Equipment
2
ii
Graphic Symbols
The following symbols are used for this equipment.
Symbols indicated on the main unit
Symbol Description
Warning (indicated in yellow)
Follow operating instructions (Warning); (indicated in blue) Indicates that the failure to follow operating instructions could place the patient or operator at risk.
Follow operating instructions (Information); Indicates the need to refer to the related accompanying documents before operation.
Type CF Applied Part with Defibrillation-Proof Indicates that the degree of protection against electric shock is Type CF Applied Part with defibrillation-proof.
Indicates the power ON/OFF status.
Indicates the point to close the battery compartment lid.
Indicates the battery type and direction.
Indicates that the alarm function is not provided.
Date of Manufacture Indicates the date of manufacture.
Non-ionizing electromagnetic radiation Indicates the radio transmitting device.
iii
Symbols displayed on the screen
Symbol Description
The following icons are displayed only on the all data display.
HR Synchronized Mark This mark flashes synchronizing to the heartbeat.
Indicates the remaining battery level.
Indicates that the expiration date of the SpO2 sensor is approaching.
Indicates that probe is disconnected or damaged.
Indicates that sensor check, etc. is required.
Indicates that the amplitude of the pulse waveform is too low, or the sensor is not positioned correctly.
Indicates that the probe is damaged, or the usable life of the sensor has expired.
iv
Precautions for Safe Operation of Medical Electrical Equipment
This section contains general information on how to handle this equipment safely for the patient and users. The precautions specific to this equipment are described afterwards.
CAUTION
1. User should have a thorough knowledge of the operation before using the equipment.
2. For installation and storage of the equipment, pay attention to the following.
Install or store in a place where the equipment will not be exposed to
splashing water.
Install or store in an area where environmental conditions such as
atmospheric pressure, temperature, humidity, ventilation, sunlight, dust, sodium, and sulfur will not adversely affect the system.
Place the equipment on a stable surface where there is no inclination,
vibration, or shock (including during transportation).
Do not install or store in an area where chemicals are stored or gases are
evolved.
3. Before operating the equipment, verify the following items.
Check the cable connection, polarity, etc. to ensure safe and proper
operation of the equipment.
Ensure that all cables are firmly and safely connected. Especially, make
sure to check the attachment and connection condition of electrodes and transducers.
Pay special attention when the equipment is used in conjunction with other
equipments as it may cause erroneous judgment and dangerous situation.
Check the remaining battery level. When replacing the batteries, make sure that the batteries polarity are
correct. Do not charge the batteries.
4. During operation of the system, verify the following items:
Do not operate the equipment beyond the time period required for
diagnosis and medical care.
Do not hold the probe or cable part to pick up the equipment. It may
damage the equipment and lead to measurement error.
Always observe the equipment and patient to ensure safe operation. If any abnormality is found on the equipment or patient, take appropriate
measures such as ceasing operation of the equipment and/or detaching the probe (sensor) and/or electrode, in the safest way for the patient.
Do not allow the patient to come in contact with other equipments.
v
CAUTION
5. After using the equipment, verify following items.
Return all operating switches, knobs etc to the position before using the
equipment, and then switch off the power.
When unplugging the cables, make sure to pull from the connector part of
the cable to avoid excessive force on the cable.
Clean the accessories and cables, and keep them together in one place. Keep the equipment clean to ensure proper operation for the next use. Make sure to remove the batteries if the equipment is not used for a long
time. The leakage from the batteries may damage the equipment, or an explosion from the batteries may occur.
6. If the equipment is damaged, do not attempt service. Ensure patient safety by immediately turning off the power and removing the electrodes and cables from the patient. Label the unit “OUT OF ORDER” and contact your nearest service representative.
7. Do not disassemble or remodel the equipment.
8. Maintenance and Inspection
Make sure to periodically check the equipment and parts. (It is
recommended to conclude a maintenance contract.)
Before reusing the equipment that has been left unused for a while, make
sure that the equipment operates properly and safely.
9. When using electrosurgical knives or defibrillator with this equipment, follow the precautions below.
To prevent patient from burn injury, verify proper attachment of patient
ground plate, ECG electrode type when using the electrosurgical knife, and verify paste volume, output energy when using the defibrillator.
Some types of equipment other than the above may cause accidental
hazards to the patient and operator due to the conditions of the equipment. Read the operation manual attached to each equipment and fully understand the precautions prior to use.
Non-Explosion Proof
DANGER
Never operate the equipment in the presence of flammable anesthetics,
high concentration of oxygen. It may cause an explosion or fire.
Never use the equipment in the hyperbaric oxygen therapy chamber.
It may cause an explosion or fire.
Never operate the equipment where flammable gas or fluid such as
anesthetic, oxygen, and hydrogen are used. It may cause an explosion or fire.
vi
Precautions about Magnetic Resonance Imaging (MRI)
WARNING
Do not use this equipment in magnetic resonance imaging (MRI)
environments.
When conducting MRI test, remove the electrodes and sensors
connected to the patient. The local heating caused by the induced electromotive force may cause burn injury to the patient. For details, refer to the operation manual for the MRI testing device.
Electrosurgery Safety
WARNING
The monitoring system contains protection against interference generated by electrosurgical instruments. However, depending on the operating conditions, surgery site with respect to the location of ECG electrodes, ground plate attachment condition, or the type of instrument used, it may cause burn injury at the electrode site or noise on the ECG. The noise is generated at the tip of the electrosurgical knife and is difficult to completely eliminate because of the frequency components of the ECG. To reduce electrosurgical interference, take the following precautions:
Location:
Locate the electrosurgical unit as far as possible from this equipment and the patient cable. This will help reduce interference on the ECG through the monitor or cables.
Electrode Placement:
The amount of noise interference is considerably different depending on the electrode position and surgery site. Place the ECG electrodes as far away as possible from the surgery site and the ground plate. Do not place electrodes in the path between the surgery site and the ground plate. If the electrodes are placed in this path, the amount of interference will be quite large. Position (+) and (–) electrodes as close as possible to each other.
Ground Plate:
When using electrosurgical instruments, make sure the contact between the patient and the ground plate is secure. If the connection is incomplete, the patient may suffer from burn at the electrode site.
vii
Precautions about Using with the Defibrillator
WARNING
When using this equipment with a defibrillator, use only the specified
lead cable. If unspecified lead cable is used, it may damage the equipment and safety cannot be ensured.
When defibrillating, keep away from the electrodes or medicament
applied to the patient chest. If this is not possible, remove the electrodes or medicament before defibrillating. If the defibrillator paddles directly touch the electrodes or medicament, an electrical shock may result by the discharged energy.
When defibrillating, do not touch the patient and the metal part of the
equipment or cables. Electric shock may result from the discharged energy.
viii
Precautions about the Pacemaker
WARNING
Minute ventilation rate-adaptive implantable pacemakers can
occasionally interact with certain cardiac monitoring and diagnostic equipment, causing the pacemakers to pace at their maximum programmed rate. The cardiac monitoring and diagnostic equipment may possibly send wrong information. If such event occurs, disconnect the cardiac monitoring and diagnostic equipment, or follow the procedures described in the operation manual of the pacemaker. (For more details, contact FUKUDA DENSHI personnel, your institution’s professionals, or your pacemaker distributors.)
Rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms. Keep pacemaker patients under close surveillance.
Reference “Minute Ventilation Rate-Adaptive Pacemakers” FDA alerts health professionals that minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac monitoring and diagnostic equipment, causing pacemakers to pace at their maximum programmed rate. [October 14, 1998 – FDA]
Precautions for Using This Equipment
WARNING
Do not connect cables not authorized by Fukuda Denshi to any I/O
connector. If unspecified cable is connected, not only that the equipment cannot deliver its maximum performance, the equipment may be damaged and safety cannot be ensured.
Do not use this equipment with multiple patients simultaneously.
CAUTION
Do not hold the cable part and hang or swing the transmitter. It may cause wire break, injury, or damage to the surrounding equipment.
ix
Precautions about Waterproof
CAUTION
To maintain the waterproof performance, replace the battery
compartment lid periodically. Otherwise, the quality of the lid will deteriorate and cannot keep the waterproof performance. For periodic replacement, contact your local service representative.
When the equipment is subjected to high impact, the damage to the
enclosure or lid may degrade the waterproof performance. In such case, contact your local service representative to check the waterproof performance.
The SpO
Do not use the transmitter when it is wet. Wipe the transmitter with a soft
probes are not waterproof. Do not take a bath with the probes
2
attached, and keep them away from liquids.
cloth and dry it thoroughly before use.
Precautions about ECG Monitoring
CAUTION
When removing electrodes from the patient, remove them carefully and
slowly. Do not apply excessive force to remove them. Otherwise, it may damage the skin.
If any electrodes get detached from the patient after being connected to
the lead cable and patient monitor, pay attention that the metal part of the electrode does not get in touch with any metal parts of the bed or any conductive parts. Also, the operator should not touch any conductive parts with bare hands. Otherwise, it may cause electric shock to the patient and/or operator due to excessive leakage current.
The indication for continuous use of the electrode is about one day. Replace the electrode if the skin contact gets loosen due to perspiration,
etc.
When an electrode is attached to the same location for a long period,
some patients may develop skin irritation. Check the patient's skin condition periodically and change the electrode site as required.
For stable ECG monitoring, verify proper electrode placement, lead, and
waveform size. If not properly selected, it may cause erroneous detection.
There are some cases when the pacemaker pulse cannot be detected
depending on the pacemaker type, pulse voltage, pulse width, electrode lead type (unipolar, bipolar), or electrode placement which causes the pacemaker pulse amplitude to decrease, and disables the pacemaker pulse detection.
If signals similar to a pacemaker pulse are present, such as electric
blanket noise or excessive AC frequency noise, these may be erroneously detected and displayed as a pacemaker pulse. In such case, check the condition of the electrodes and lead cables to resolve the cause or turn off the pacemaker detection setting on the receiving monitor.
If a pacemaker pulse is continuously detected due to AC frequency
interference, QRS detection will not be properly displayed.
x
Precautions about SpO2 Monitoring
WARNING
For SpO2 monitoring, use only the specified probe. Check the probe
before usage to make sure that it is the specified probe. If unspecified probe is used, not only that the equipment cannot deliver its maximum performance, the equipment may be damaged and safety cannot be ensured.
If the nail is rough, dirty, or manicured, accurate measurement will not
be possible. Change the finger or clean the nail before attaching the sensor.
If irritation such as skin reddening appears with the sensor use, change
the attachment site or stop using the sensor.
Do not use a tape to attach the sensor. When the SpO
measurement/waveform will not be displayed on the receiving monitor. Also, the alarms will not be generated. Make sure that the SpO2 probe is securely connected to this equipment.
When not measuring, unplug the SpO2 probe from the connector.
Otherwise, the outside light may affect to falsely display the measurements.
Check the sensor attachment site constantly every 4 hours when probes
or reusable sensor are used, and at least every 8 hours when single patient use sensors are used. Be especially careful of a patient with bad perfusion. If the sensor attachment position is not changed constantly, skin irritation or skin necrosis due to compression may be developed. For the patient with bad perfusion, check the sensor attachment position at least every 2 hours.
As skin for neonate, premature infant is immature, change the sensor
attachment site more frequently depending on the condition.
When measuring the SpO2 of patient with high fever or peripheral
circulatory insufficiency, check the sensor attachment periodically and change the attachment site. The temperature of the attachment site will rise due to the sensor heat which may result in burn injury.
Even attachment for a short duration may inhibit the blood flow and
generate compression necrosis. Also, blood flow inhibition may prevent correct measurements.
Direct sunlight to the sensor area can cause a measurement error.
Place a black or dark cloth over the sensor if using in direct sunlight.
The pulse wave is normalized for SpO
indicate perfused blood volume. Check proper sensor attachment by observing the pulse wave.
Precautions for Reusable Sensors
The light-emitting part of the sensor should be over the root of the fingernail or as instructed per the related sensor instruction manual. Do not insert the finger too far into the sensor as it may hurt the patient. For details, refer to the SpO
Precautions for Single-Patient-Use Type Sensors
The sensor can be reused on the same patient as long as the adhesive tape attaches without slippage. Do not reuse on other patients to avoid
probe is disconnected from the equipment, the SpO2
2
measurement, and does not
2
sensor instruction manual.
2
xi
cross contamination. It is intended for single patient use only. For details, refer to the operation manual of the SpO
Measuring on a limb with NIBP cuff, arterial catheter, or intracatheter
may result in incorrect measurement. Venous congestion may cause under reading of actual oxygen saturation. Therefore, assure proper venous outflow from monitored site. Sensor should not be below heart level (e.g. sensor on hand of a patient in a bed with arm dangling to the floor).
sensor.
2
CAUTION
WARNING
For the following case, accurate measurement may not be possible.
Patient with excessive abnormal hemoglobin (COHb, MetHb) Patient with the pigment injected to the blood Patient receiving CPR treatment When a probe is applied to a limb with NIBP cuff, arterial catheter, or
intracatheter
When measuring at position with venous pulse Patient with body motion Patient with small pulse Excessive body motion (patient’s motion) Excessive light (direct sunlight, fluorescent, light therapy equipment,
surgical light, infrared heat ramp, etc.)
External colorant such as nail polish Abnormally low or high hemoglobin concentration
Precautions about Output Signal
WARNING
Do not use the output signal of the monitor that receives radio wave signal from this equipment as the trigger signal for MRI, echocardiographic, or defibrillator. It may lead to a delay of operating timing due to the delay time of waveform transmission. A trigger signal unrelated to the heart rate may be generated due to the interfusion of spike noise at weak electric field.
Precautions about Accessories and Optional Accessories
WARNING
Use only the specified disposable electrodes, lead cable, SpO2 probes, etc. Otherwise, this equipment cannot deliver its maximum performance and may be damaged, resulting in a safety hazard.
xii
CAUTION
Do not reuse disposable products. Store the disposable products properly as mentioned in their user
manuals.
Precautions about the Alkaline Batteries
WARNING
Use new "AA" size (“LR06” size) alkaline batteries which is within the
expiration date.
Install the batteries with the correct polarity. Do not charge the batteries. Any attempt to charge the batteries may
cause them to leak or break.
Do not short the (+) and (-) terminals. It may result in exothermic heat
and fire.
Do not use different types of batteries at the same time. The leakage
from the batteries may damage the equipment, or an explosion from the batteries may occur.
Precautions about Disposing of Equipment, Accessories, or Components
CAUTION
When disposing of the equipment, accessories, or components, use an
industrial waste distributor. Do not dispose of as ordinary waste.
Used disposal items (ECG electrodes, etc.) shall be discarded as
medical waste.
Precautions about Disposing of Battery
CAUTION
Obey the local municipal rule to dispose the used dry cell battery.
xiii
Precautions for Use of Medical Telemeter
WARNING
The LX-8300M transmitter must not be co-located or operated in
conjunction with any other antenna or transmitter.
For the receiving monitor of the LX-8300M transmitter, make sure to
use the Fukuda Denshi products with the receiving range of 608 MHz­614 MHz.
This equipment complies with FCC/IC radiation exposure limits set forth
for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines and RSS-102 of the IC radio frequency (RF) Exposure rules. This equipment has very low levels of RF energy that are deemed to comply without testing of specific absorption rate (SAR).
Operation of LX-8300M requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry Service.
This radio frequency device is susceptible to interference from outside
sources. Interference may prevent the monitoring of patients connected to this equipment. If a problem exists, contact your local service representative.
The LX-8300M transmits vital signs to the receiving monitor using radio
wave signal. Under unstable radio wave signals, the receiving monitor will not generate any alarms. This situation may miss sudden change in the patient's condition and may cause a serious accident. Under unstable radio wave signals, check the patient status consistently under this situation. To get stable radio wave signals, make sure to have a proper telemetry installation.
CAUTION
For installation, make sure to follo w the precautions below.
The medical institution (hereinafter referred to as the “Institution”) must
decide the telemetry installation plan for the medical department in order to prevent interference and interference between transmitters (telemetry based on destination country’s radio law). When telemetry has already been installed and been used, radio format, frequency, and antenna power are required to be examined to prevent interference.
When laying receiver antenna for each transmitter, the Institution has to
examine the installation so that electronic interference does not occur.
Based on the above examination result, the Institution should install each
receiver antenna as required.
xiv
CAUTION
For management, make sure to follow the precautions below.
The Institution should appoint a person (hereinafter referred to as the
"Coordinator") to manage the wireless channels for the whole Institution.
The Coordinator must be selected from people who understand the
characteristics and functionality of telemetry systems, and are skilled in operating telemetry.
When installing telemetry, the Coordinator has to understand the
precautions for use of the telemetry in advance.
The Coordinator is responsible for maintenance of wireless channels and
storage and maintenance of telemeter in the overall medical facilities to give proper instructions to the telemetry users.
The Coordinator should create a management log (hereinafter referred to
as the "log"), which contains a list of the management status of the wireless channels for the whole Institution. When changing a wireless channel, register it in the log and give proper instructions to the user.
The telemetry user verifies operation of the transmitter/receiver before use. When interference or breakdown occurs in telemetry communication, the
user is required to inform the Coordinator of the problems. The Coordinator is to deal with the problem properly and/or contact their nearest Fukuda
Electromagnetic Compatibility
Denshi representative for service.
This equipment complies with IEC 60601-1-2 (2014), safety standard regarding the electromagnetic disturbances of medical electrical equipment. To ensure maximum performance against the electromagnetic disturbances, make sure to follow the precautions for installation and usage described in this manual.
This equipment is intended for use in the medical facility (except inside the shield room of MRI device), and satisfies the immunity level for professional healthcare facility environment stipulated in IEC 60601-1-2.
An excessive magnetic disturbance may degrade the HR and SpO measurement accuracy (refer to “15. Specification”), which is the essential performance of this equipment, and may cause delay in treatment or inaccurate diagnosis.
When using this equipment, interference with other medical electrical equipments or non-medical electrical equipments may occur. Make sure that no interference is present before usage.
To ensure basic safety and essential performance related to electromagnetic distrubances during the expected service life of this equipment, “Daily Check” and “Periodic Check” must be performed. (refer to “13. Maintenance and Inspection”)
2
xv
Precautions for Safe Operation under Electromagnetic Influence
CAUTION
If any sorts of electromagnetic wave, magnetic field, or static electricity exist around the equipment, noise interference or malfunction of the equipment may occur. If any unintended malfunction or noise occurs during monitoring, check the magnetic influence and take appropriate countermeasures.
The followings are examples of the common cause and countermeasures.
Mobile Phone
The radio wave may cause malfunction to the equipment. Mobile phones and radio sets should be turned off in the room (building) where medical device is located.
Static Electricity
In a dry environment (room), static electricity is likely to occur. Take the following countermeasures.
Both operator and patient should remove any static electricity before
entering the room.
Humidify the room.
CAUTION
If this equipment is installed close to, or stacked with other equipment,
malfunction may occur. Make sure to verify that the equipments operate properly in a used location.
Use of accessories, probes, or cables other than specified may cause
increase in electromagnetic emission or decrease in electromagnetic immunity resulting in malfunction of the equipment.
The portable RF communications equipment (including antenna cable
and peripheral equipment such as external antenna) with the specified cable should be used in a location at least 30 cm apart from any part of this equipment. Otherwise, it may result in performance degradation of this equipment.
xvi
EMC Guidance
This equipment complies with IEC 60601-1-2 (2014). However, if portable transmitter or wireless LAN equipment is used extremely nearby, the electromagnetic influence may largely exceed the compliance level and may cause unexpected phenomenon such as noise interference on the waveform, etc. Therefore, this equipment should be used in a location specified by each medical institution. If any unexpected noise interference on the waveform or failure to the peripheral device occurs, stop using the equipment and follow the instruction of the technician.
The following is the information relating to EMC (Electromagnetic Compatibility). (When using this equipment, verify that it is used within the environment specified below.)
Compliance to the Electromagnetic Emissions
This equipment complies with the following emission standard.
Emission test Compliance
RF Emission CISPR 11 Group 1 Class A
CAUTION
The emission performance of this equipment is suitable for use in industrial environment and hospital environment (CISPR 11 Class A). To use in home environment (generally, CISPR 11 Class B is required), this equipment may not be properly protected from wireless frequency communication service. It may be necessary to take measures such as changing the installation location or equipment orientation.
xvii
Compliance to the Electromagnetic Immunity
The LX-8300M is intended for use in the electromagnetic environment specified below. The customer or the user of the LX-8300M should assure that it is used in such an environment.
Basic EMC standard or
test method
Electrostatic discharge IEC 61000-4-2
Radiated RF EM fields IEC 61000-4-3
Proximity fields from RF wireless communications equipment IEC 61000-4-3 Conducted disturbances induced by RF fields IEC 61000-4-6
Rated power frequency magnetic fields IEC 61000-4-8
Immunity test specifications for RF wireless communications equipment
Test frequency
(MHz)
710, 745, 780 PM, 217 Hz 0.2 0.3 9
810, 870, 930 PM, 18 Hz 2 0.3 28
1720, 1845, 1970 PM, 217 Hz 2 0.3 28
2450 PM, 217 Hz 2 0.3 28
5240, 5500, 5785 PM, 217 Hz 0.2 0.3 9
IEC 61000-4-3: Proximity fields from RF wireless communications equipment
Since TETRA 400 is a service in Europe and this product for the US does not emit close proximity, the test frequency of 385 MHz is not implemented.
GMRS 460, FRS 460 are general and leisure radios and have a test frequency of 450 MHz because they are not radiated in close proximity with this product, which is intended for use on a patient in a professional healthcare environment. Not implemented.
xviii
8 kV contact 2 kV, 4 kV, 8 kV, 15 kV air
3 V/m 80 MHz to 2.7 GHz 1 kHz 80%AM Refer to the following table.
3 V
0.15 MHz to 80 MHz 1 kHz 80%AM 6 V
0.15 MHz to 80 MHz (in ISM bands between 0.15 MHz and 80 MHz) 1 kHz 80%AM 30 A/m 60 Hz
Modulation
Maximum
power
(W)
Immunity test levels
Distance
Immunity test level
(m)
(V/m)
Contact
If you need more information, please contact the following.
(1) Fukuda Denshi Co., Ltd., Head Office
3-39-4 Hongo, Bunkyo-ku, Tokyo, Japan Tel: +81-3-5684-1455 Fax: +81-3-3814-1222 E-mail: info@fukuda.co.jp Home Page: http://www.fukuda.com
(2) Sales Representative
Write the name, address, phone, fax number of your local sales representative.
(Name of Sales Representative, Address, Phone/Fax)
xix
xx
Contents
Safety Precautions .................................................................................... i
Precaution from Fukuda Denshi ....................................................... i
Intended Use of this Equipment ....................................................... ii
Graphic Symbols ............................................................................. iii
Precautions for Safe Operation of Medical Electrical Equipment .... v
Non-Explosion Proof ....................................................................... vi
Precautions about Magnetic Resonance Imaging (MRI) ................ vii
Electrosurgery Safety ..................................................................... vii
Precautions about Using with the Defibrillator .............................. viii
Precautions about the Pacemaker .................................................. ix
Precautions for Using This Equipment ............................................ ix
Precautions about Waterproof ........................................................ x
Precautions about ECG Monitoring ................................................. x
Precautions about SpO2 Monitoring ................................................ xi
Precautions about Output Signal ................................................... xii
Precautions about Accessories and Optional Accessories ............ xii
Precautions about the Alkaline Batteries ...................................... xiii
Precautions about Disposing of Equipment, Accessories, or Components
................................................................................................. xiii
Precautions about Disposing of Battery ........................................ xiii
Precautions for Use of Medical Telemeter .................................... xiv
Electromagnetic Compatibility ............................................................ xv
Precautions for Safe Operation under Electromagnetic Influence . xvi
EMC Guidance ............................................................................. xvii
Contact .................................................................................................. xix
1. General Description ............................................................................. 1
2. Names of Parts and Their Functions .................................................... 3
3. Preparation .......................................................................................... 5
1) Installing the Batteries ................................................................ 5
2) Operating the Power Switch ....................................................... 7
4. ECG Monitoring ................................................................................. 11
Connecting the ECG Lead Cable and Electrodes ....................... 11
Attaching the Electrodes ............................................................. 12
Connecting the ECG Lead Cable to the LX-8300M .................... 13
5. Respiration Monitoring ....................................................................... 15
6. SpO2 Monitoring ................................................................................. 17
Preparation for Monitoring .......................................................... 17
Connecting the SpO2 sensor to the LX-8300M ........................... 29
7. Measurement ..................................................................................... 31
Monitoring Screen ....................................................................... 31
All Data Display .......................................................................... 33
8. Operation ........................................................................................... 39
Setup Mode ................................................................................ 39
ECG Setup .................................................................................. 39
Respiration Setup ....................................................................... 41
SpO2 Setup ................................................................................. 41
Restarting the Display ................................................................. 43
Pressing the [EVENT] Button ...................................................... 44
9. Other Setting Items ............................................................................ 45
Time Constant ............................................................................ 46
Detection Sensitivity of the Pacemaker Pulse ............................. 46
QRS Detection ............................................................................ 47
Respiration Detection Signal ON/OFF ........................................ 48
Display Brightness ...................................................................... 48
Display Timeout Duration ............................................................ 48
Sound ON/OFF ........................................................................... 49
Displayed Color of SpO2 ............................................................. 49
Transmit PI Information .............................................................. 49
10. Changing the Transmitter Channel and Group ID ............................ 51
Transmitter Channel ................................................................... 51
Group ID ..................................................................................... 51
11. Troubleshooting ............................................................................... 53
In Case of Dropping the LX-8300M into Water ........................... 59
12. Cleaning and Disinfection ................................................................ 61
Cleaning the Housing ................................................................. 61
Cleaning the Connector .............................................................. 62
Disinfection ................................................................................. 62
Cleaning the ECG lead cable ...................................................... 62
13. Maintenance and Inspection ............................................................ 63
Daily Check ................................................................................. 63
Periodic Check ............................................................................ 63
Periodic Replacement Parts ....................................................... 63
Periodic Check ............................................................................ 65
Repairing the Equipment ............................................................ 71
Replacing the Battery Compartment Lid Unit ............................. 71
14. Standard and Optional Accessories ................................................. 75
Standard Accessories ................................................................. 75
Optional Accessories .................................................................. 75
15. Specification ..................................................................................... 79
Specification ............................................................................... 79
1. General Description
The LX-8300M is a radio telemetry transmitter designed to measure the ECG, respiration waveform, SpO hemoglobin), pulse waveform with two “AA” size (“LR06” size) alkaline batteries. Information such as ECG measurements, respiration waveform, SpO measurements electrodes and SpO ECG lead selection is available using the two buttons ([ENTER] and [ the front panel (In case of using a 3-electrode lead cable or a 5-electrode chest lead cable). The LX-8300M can also function as a transmitter to measure only the ECG/Respiration without SpO2 or to measure only the SpO2 without ECG/Respiration. Before using the LX-8300M, read also the operation manual of the patient monitor at the receiving side thoroughly.
pulse waveform, battery level, and the conditions of the ECG
,
External Appearance
(functional oxygen saturation of arterial
2
probe (sensor) are displayed on the front panel.
2
2
]) on
1
1. General Description
2
2. Names of Parts and Their Functions
Refer to “Safety Precautions” in the preface of this manual.
SpO2 Input Connector Connects the SpO
ECG/RESP Input Connector
Connects the ECG lead cable to measure ECG and respiration waveform
Refer to “Safety Precautions” in the preface of this manual.
probe (sensor).
2
Electrode Position Label
Indicates standard ECG electrode position.
Power Switch
Power Supply LED
Turns ON/OFF the power
Indicates the power supply status
Light ON: In normal operation Light OFF: Power Off
[ENTER] Button
For Setup
[ ] Button
For Setup
Display
Displays measurement waveform and transmitter information.
Channel Number
Paste the label with the number.
3
2. Names of Parts and Their Functions
[EVENT] Button
Transmits EVENT Information
Refer to “Safety Precautions” in the preface of this manual.
Battery Compartment Lid Open/Close Lever
To open/close the battery compartment lid, slide this lever up/down.
Battery Lid Lock Button
Unlocks the battery compartment lid.
4
Strap Hole
Attaches the strap.
3. Preparation
1) Installing the Batteries
The LX-8300M functions with two “AA” size (“LR6” size) alkaline batteries. The battery operation time of LX-8300M is as follows.
When SpO
When SpO
However, continuous operating time may be shorter than the above mentioned time depending on the application of the SpO2 probe (sensor).
Unplug the ECG lead cable when the battery compartment lid is opened.
Otherwise, patient leakage current beyond the allowable value may occur.
Use new "AA" size (“LR06” size) alkaline batteries. Do not short out the (+) and (-) terminals. It may result in exothermic
heat and fire, the leakage from the batteries may damage the equipment, or an explosion from the batteries may occur.
Install the batteries with the correct polarity. Do not charge alkaline batteries. Any attempt to charge the batteries
may cause them to leak or break.
Do not use a disassembled or a damaged battery due to drop or shock.
The leakage from the batteries may damage the equipment, or an explosion from the batteries may occur.
Do not use different types of batteries at the same time. The leakage
from the batteries may damage the equipment, or an explosion from the batteries may occur.
Remove the exhausted batteries immediately. The leakage from the
batteries may damage the equipment, or an explosion from the batteries may occur.
If the equipment is not in use for a long period of time, remove the
batteries and store the equipment in an appropriate place. If the batteries are left in the equipment for a long period of time, the leakage from the batteries may damage the equipment or an explosion from the batteries may occur.
Make sure to replace the two batteries simultaneously. If a new and
used battery are mixed, a leakage from the batteries may damage the equipment or an explosion from the batteries may occur.
measurement is ON: Approximately 2.5 days (new batteries)
2
Conditions: When measuring ECG, RESP, SpO operating temperature 23C
measurement is OFF: Approximately 6.5 days (new batteries)
2
Conditions: When measuring ECG, RESP with default settings, SpO measurement OFF, operating temperature 23C *Disconnecting the SpO It is necessary to set the SpO OFF the SpO2 measurement“.
probe does not satisfy the above condition.
2
measurement to OFF. Refer to “Turning
2
with default settings,
2
WARNING
2
5
3. Preparation
CAUTION
Use only alkaline battery (AA). Other battery will shorten the continuous
operating time.
Once the power switch is on the OFF position, then open the battery
compartment lid.
Do not replace the batteries with wet hands. In case of storing the used or unused batteries, make sure that the
terminals are not touching other batteries or metal parts.
(1)
(2)
(2)
(1)
Unlock and open the battery compartment lid by sliding the open/close lever towards OPEN while pressing the lock button.
Install new batteries according to the polarity indication inside the battery compartment.
After installing the batteries, lock the battery compartment lid by sliding the open/close lever towards CLOSE while pressing over “PUSH” on the lid.
6
Loading...
+ 80 hidden pages