Before using this device, read this operation manual thoroughly.
● Keep this manual near the device for future reference.
CAUTION
Federal Law restricts this device to sale by or on the order of a physician.
CAUTION
Users are advised to periodically contact the FCC or specified frequency
coordinator and determine if other or your transmitter frequencies that
may cause interference.
CAUTION
The manufacturers, installers and users of Wireless Medical Telemetry
System equipment are cautioned that the operation of this equipment
could result in harmful interference to other nearby medical devices.
CAUTION:
• This device for sale by or on the order of a physician.
• The company and product names used in this manu al are trademarks or
registered trademarks.
• If this manual has pages missing or out of order, contact Fukuda Denshi
for replacement.
• Only physician or persons instructed by physicians are allowed to use the
equipment.
• The information contained in this document is subject to change without
Thank you for purchasing this product from Fukuda Denshi.
Before use, read this operation manual thoroughly for correct handling
and operation.
Safety Precautions
Read the “Safety Precautions” thoroughly before use to ensure correct and
safe use of the product.
Make sure to follow the precautions indicated below, as these are important
messages related to safety.
DANGER
WARNING
CAUTION
NOTE
Failure to follow this message may cause immediate
threat of death or serious injury.
Failure to follow this message may result in death or
serious injury.
Failure to follow this message may cause injury or
failure to the equipment.
A note is not related to product safety, but provides
information about the correct use and operating
procedures to prevent incorrect operation and
malfunction of the equipment.
Precaution from Fukuda Denshi
Fukuda Denshi is liable for the safety, reliability, and performance of its device
only if;
Maintenance, modifications, and repairs are carried out by authorized
personnel.
Components are used in accordance with Fukuda Denshi operating
instructions.
If the equipment is used incorrectly and become unusable, Fukuda Denshi is
not liable for the malfunction. Use the equipment only for the purpose specified
in this manual.
i
Equipment Symbols
Refer to the following symbols indicated on the LX-7230 for their meanings.
Symbols indicated on the main unit
ol Description
Symb
Caution; read the operation manual.
Indicates the need to refer to the operation manual before
operation.
Type CF Applied Part with Defibrillation-Proof
Indicates that the degree of protection against electric
shock is Type CF Applied Part with defibrillation-proof.
Indicates the battery type and direction.
Symbols indicated on the LCD screen
Symbol Description
Heart Rate Synchronization Mark:
This mark flashes synchronizing with the heartbeat.
Battery Mark:
Indicates the remaining battery level.
Intermittent SpO
Indicates that the SpO
activated.
Mode:
2
Intermittent measurement is
2
Displayed only on the LX-7230KM.
ii
Precautions for Safe Operation of Medical E lectrical Equipment
Cautions described here are regarding the general instructions for safety use
to the patient and the operator. As for cautions about the LX-7230, please refer
to the following pages.
CAUTION
1. Users should have a thorough knowledge of the operation before
using this equipment.
2. Pay attention to the following when installing or storing the
equipment.
Do not install or store in an area where the equipment will be subject
to splashing water.
Do not install or store in an area where the environmental
conditions, such as atmospheric pressure, temperature, humidity,
ventilation, sunlight, dust, sodium, sulfur, will adversely affect the
system.
Place the equipment on a stable surface where there is no
inclination, vibration, or shock (including during transportation).
Do not install or store in an area where chemicals are stored or
gases are generated.
3. Before operating the equipment, verify the following items.
Check the cable connection and polarity to ensure proper operation
of the equipment.
Ensure that all cables are firmly and safely connected. Especially,
recheck the attachment and connection condition of electrode and
the probe (sensor).
Pay special attention when the equipment is used in conjunction with
other equipment as it may cause erroneous judgment and danger.
Check the remaining battery level.
When replacing the battery, make sure that the battery polarity is
correct. Do not charge the battery.
4. During operation of the equipment, verify the following items.
Do not operate the equipment beyond the time period required for
diagnosis and medical care.
Do not pick up and/or swing the equipment pulling/grabbing the
probe (sensor) or cable part. It may damage the equipment and lead
to measurement error.
Always observe the equipment and patient to ensure safe operation
of the equipment.
If any abnormality is found on the equipment or patient, take
appropriate measures such as ceasing operation of the equipment
and/or detaching the probe (sensor) and/or electrode, in the safest
way for the patient.
Do not allow the patient to come in contact with other equipments.
iii
CAUTION
5. After using the equipment, verify the following items.
Make sure to turn off the power of the equipment.
When unplugging the cables, do not apply excessive force on the
cable and pull from its connector.
Clean the accessories and cables, and keep them together in one
place.
Keep the equipment clean to ensure proper operation for the next
use.
Make sure to remove the batteries if the equipment is not used for a
long time. The leakage from the batteries may damage the device or
an explosion from the batteries may occur.
6. If the equipment is damaged and in need of repair, ensure patient
safety by immediately turning the equipment off and remo ve the
electrodes and/or probe from the patient. User should not
attempt service. Label the unit “OUT OF ORDER” and contact
Fukuda Denshi representative.
7. Do not remodel the equipment.
8. Maintenance Check
Make sure to periodically check the equipment, and accessories.
Before reusing the equipment that has been left unused for a while,
make sure that the equipment works normally and safely.
9. When using electrosurgical knives or defibrillator with this
equipment, take care of the following.
To prevent burn injury to the patient, verify proper attachment of
patient ground plate, ECG electrode type for the electrosurgical
knives, and the quantity of gel, output energy for the defibrillator.
Also, verify that a proper ground is selected.
Some types of equipment other than the above may cause
accidental hazards to the patient and operator due to the conditions
of the equipment. Read the operation manual attached to each
device and understand the precautionary instructions prior to use.
Non-Explosion Proof
DANGER
Never operate the equipment in the presence of flammable anesthetics,
high concentration of oxygen. It may cause an explosion or fire.
Never operate the equipment inside a hyperbaric chamber. It may
cause an explosion or fire.
Never operate the equipment where flammable gas or fluid such as
anesthetic, oxygen, and hydrogen are used. It may cause an explosion
or fire.
iv
Precautions about Magnetic Resonance Im aging (MRI)
WARNING
Do not operate this equipment in magnetic resonance imaging (MRI)
environments.
When conducting MRI test, remove the electrodes and sensors
connected to the patient (test subject).
The local heating caused by the induced electromotive force may cause
burn injury to the patient (subject). For details, refer to the operation
manual for the MRI testing device.
Electrosurgery Safety
WARNING
When using electrosurgical instrument, make sure the contact between
the patient and the ground plate is secure. If the connection is
incomplete, the patient may suffer a burn at the electrode site.
When using an electrosurgical instrument, it may misidentify noise from
the electrosurgical instrument as a heartbeat or arrhythmia.
Defibrillation Safety
WARNING
Use only the lead cable specified by Fukuda Denshi when defibrillating.
If used by unspecified lead cable, the device may be damaged, resulting
in a safety hazard.
When using the defibrillator, keep away from the electrodes or
medicament applied to the patient chest. If this is not possible, remove
the electrodes or medicament before using it.
If the defibrillator paddles are directly in contact with the electrodes or
medicament, an electrical shock may result from the discharged energy.
When using the defibrillator, do not touch the patient and the metal part
of the device or cables. Electric shock may result from the discharged
energy.
v
Precautions about the Pacemaker
WARNING
Minute ventilation rate-adaptive implantable pacemakers can
occasionally interact with certain cardiac monitoring and diagnostic
equipment, causing the pacemakers to pace at their maximum
programmed rate. The cardiac monitoring and diagnostic equipment
may possibly send wrong information.
If such event occurs, please disconnect the cardiac monitoring and
diagnostic equipment, or follow the procedures described in the
operation manual of the pacemaker.
(For more details, contact FUKUDA DENSHI personnel, your
institution’s professionals, or your pacemaker distributors.)
Reference
“Minute Ventilation Rate-Adaptive Pacemakers”
FDA alerts health professionals that minute ventilation rate-adaptive
implantable pacemakers can occasionally interact with certain cardiac
monitoring and diagnostic equipment, causing pacemakers to pace at their
maximum programmed rate.
[October 14, 1998 (Letter: www .fda.gov/cdrh/safety.html) – FDA]
ECG meter may continue to count the pacemaker rate during
occurrences of cardiac arrest or arrhythmias. Do not rely entirely upon
the ECG meter alarms. Keep pacemaker patients under close
surveillance. Check this manual for disclosure of the pacemaker pulse
rejection capability of this instrument.
Precautions about the LX-7230
WARNING
Do not connect cables not authorized by Fukuda Denshi to any I/O
connector. If done so by mistake, the LX-7230 cannot deliver its
maximum performance and may be damaged, resulting in a safety
hazard.
Do not use this device with multiple patients simultaneously.
CAUTION
Do not pick up and/or swing the LX-7230 pulling/grabbing the probe
(sensor) or cord part. The cable could break or get disconnected from the
LX-7230. And it may hit people or damage other equipment around.
vi
Precautions about Waterproof
CAUTION
Replace the “Battery Compartment Lid” of the LX-7230 regularly to keep
the main unit performance of waterproof. If not regularly replaced, the
quality of the lid will deteriorate and cannot keep the waterproof
performance. For details about the regular replacement, contact your
local Fukuda Denshi service representative.
The lid may be damaged from high impact. If the LX-7230 is dropped or
is subjected to a high impact, make sure that the lid is not damaged.
However, the SpO
Do not take a bath with them, and ensure to be away from liquid.
Do not use the LX-7230 wet. Always wipe the LX-7230 with a soft cloth
and dry it thoroughly before use.
probes (sensors/relay cables) are not waterproof.
2
Precautions about ECG
CAUTION
When removing electrodes from the patient, remove them carefully and
slowly. Do not apply excessive force to remove them. Otherwise, it may
damage the skin.
There are some cases when the pacemaker pulse cannot be detected
depending on the pacemaker type, pulse voltage, pulse width, electrode
lead type (unipolar, bipolar), electrode placement, or lead method which
causes the pacemaker pulse amplitude to decrease and disables
pacemaker pulse detection.
If signals similar to a pacemaker pulse are present, such as electric
blanket noise or excessive AC frequency noise, these may be
erroneously detected and displayed as a pacemaker pulse. In this case,
check the condition of the electrodes and ECG lead cable to resolve the
cause or turn off the pacemaker detection setting on the receiving
monitor.
vii
Precautions about SpO
2
WARNING
During SpO2 monitoring, always use the probe (sensor)/relay cable
specified by Fukuda Denshi. If any other probe (sensor)/relay cable is
used, a high temperature rise of the probe (sensor) may place the
patient in danger of burns in the worst case.
When the SpO2 probe (sensor) is in a connector-off condition, the SpO
alarm will not be generated on the receiving monitor. Make sure that the
probe (sensor)/ relay cable is securely connected. If the SpO
SpO
2
waveform/numeric data is not displayed, check the patient’s condition
and pay attention not to miss the connector-off condition.
When measuring the SpO2 of a patient with high fever or peripheral
circulatory insufficiency, check the probe (sensor) attachment
periodically and change the attachment site. The temperature of the
attachment site will rise 2 to 3C due to the sensor heat which may
result in compression necrosis and burn injury.
Even a short duration of attachment may inhibit the blood flow and
generate compression necrosis and burn injury.
When securing the probe (sensor) with tape, do not apply the tape too
tight. At the same time, check the blood flow constantly so that
congestion is not generated at the peripheral. When removing the tape,
remove it slowly with care not to damage the patient’s skin.
CAUTION
For the following case, accurate measurement may not be possible.
Patient with excessive abnormal hemoglobin (COHb, MetHb)
Patient with the pigment injected to the blood
Patient receiving CPR treatment
Placement of SpO
arterial catheter, or intravascular line
When measuring at placement position with venous pulse
Patient with body motion
Patient with small pulse
Excessive body motion (patient’s motion)
Excessive light (direct sunlight, fluorescent, light therapy equipment,
surgical light, infrared heat ramp, etc.)
External colorant such as nail polish
Abnormally low or high hemoglobin concentration
Electrosurgery
Influence of electromagnetic waves from other electronics devices
High-intensity radio waves from cell phone
probe (sensor) on limb with a blood-pressure cuff,
2
2
2
viii
Precautions about Output Signal
WARNING
Do not use the output signal of the monitor that receives radio wave signal
from the LX-7230 as the trigger signal for IABP, MRI echocardiographic, or
defibrillator for the following reasons.
It may lead to a delay of operating timing due to the delay time of
waveform transmission.
A trigger signal unrelated to the heart rate may be generated due to
the interfusion of spike noise at weak electric field.
Precautions about Accessories and Optiona l Accessories
WARNING
Use only the accessories, such as ECG Lead cable and SpO2 probe
(sensor)/ relay cable, specified by Fukuda Denshi for the LX-7230.
Otherwise, the LX-7230 cannot deliver its maximum performance and may
be damaged, resulting in a safety hazard.
CAUTION
Do not reuse disposable products.
Store the disposable products properly as mentioned in their user
manuals.
Precautions about Battery
WARNING
Use new "AA" size (“LR06” size) alkaline cell.
Install the battery with the correct polarity.
Do not charge the battery. Any attempt to charge the battery may cause
it to leak or break.
Do not short the (+) and (-) terminals. It may result in exothermic heat
and fire.
Do not throw the battery into fire. It may explode.
ix
Precautions about Disposing of Equipmen t, Accessories, or
Compo
nents
CAUTION
When disposing of the equipment, accessories, or components, use an
industrial waste distributor. Do not dispose of as ordinary waste.
Used disposal items (ECG electrodes, etc.) shall be discarded as
medical waste.
Precautions about Disposing of Batte ry
CAUTION
Obey the local municipal rule to dispose the used dry cell battery.
Precautions for Use of Medical Telem eter
WARNING
The LX-7230 transmitter must not be co-located or operated in
conjunction with any other antenna or transmitter.
The LX-7230 complies with FCC radiation exposure limits set forth for a
controlled environment and meets the FCC radio frequency (RF)
Exposure Guidelines in Supplement C to OET65. The LX-7230 has very
low levels of RF energy that are deemed to comply without testing of
specific absorption ratio (SAR).
Operation of LX-7230 requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry
Service.
This radio frequency device is susceptible to interference from outside
sources. Interference may prevent the monitoring of patients connected
to this device. If a problem exists, contact your local service
representative.
The LX-7230 transmits vital signs to the receiving monitor using radio
wave signal. Under unstable radio wave signals, the receiving monitor
will not generate any alarms. This situation may miss sudden change in
the patient's condition and may cause a serious accident. Under
unstable radio wave signals, check the patient status consistently under
this situation. To get stable radio wave signals, make sure to have a
proper telemetry installation.
CAUTION
For installation, make sure the following.
The medical institution (hereinafter referred to as the “Institution”) must
decide the telemetry installation plan for the medical department in order
to prevent interference and interference between transmitters (telemetry
x
based on destination country’s radio law). When telemetry has already
been installed and been used, radio format, frequency, and antenna
power are required to be examined to prevent interference.
When using telemetry, which requires zone location, the Institution is to
set up the zones as an operation unit for each transmitter to prevent
electronic interference between telemetry throughout the Institution.
When using telemetry, which requires zone location, display and identify
each prepared zone in the equipment.When laying receiver antenna for
each transmitter, the Institution has to examine the installation so that
electronic interference does not occur.
Based on the above examination result, the Institution should install each
receiver antenna as required.
For management, make sure to follow the precautions below.
The Institution should appoint a person (hereinafter referred to as the
“Overall Manager”) to manage the wireless channels for the whole
Institution.
And when using telemetry, which requires zone location, the Institution
should nominate a person (hereinafter referred to as the “Zone Manager”)
to manage the wireless channels in each zone. However, when using such
telemetry in a local Institution, one person can perform both functions.
The Overall Manager and Zone Manager must be selected from people
who understand the characteristics and functionality of telemetry systems,
and are skilled in operating telemetry.
When installing telemetry, the Overall Manager and the Zone Manager
have to understand the precautions for use of telemetry in advance.
The Overall Manager is responsible for maintenance of wireless channel
and storage and maintenance of telemeter in the overall medical facilities
to give proper instructions to the Zone Manager when using telemetry
needing zone alignment, and to the telemetry user when using telemetry
not-needing zone alignment.
The Overall Manager should create a management log (hereinafter
referred to as the “log”), which contains a list of the management status of
the wireless channels for the whole Institution. When changing a wireless
channel, register it in the log and give proper instructions to the Zone
Manager or to the user.
The Zone Manager assumes responsibility for managing the wireless
channels, storing, and managing telemetry.
The Zone Manager assigns the transmitter to the user, and provides
enough education for use inside the zone.
The telemetry user verifies operation of the transmitter/receiver before
use.
The telemetry user, if using the telemetry in a zone location, follows the
instructions of the Zone Manager for the zone and gives instructions to the
patient if required.
When interference or breakdown occurs in telemetry communication, the
user is required to inform the Zone Manager and the Overall Manager of
the problems. The Zone Manager and Overall Manager are to deal with
the problem properly and/or contact their nearest Fukuda Denshi
representative for service.
xi
Electromagnetic Compatibility
The performance of this device under electromagnetic environment complies
with IEC 60601-1-2 (2007).
Precautions for Safe Operation under Electr omagnetic Influence
CAUTION
If any sorts of electromagnetic wave, magnetic field, or static electricity
exist around the device, noise interference or malfunction of the device
may occur. If any unintended malfunction or noise occurs during
monitoring, check the magnetic influence and take appropriate
countermeasures.
The following are examples of the common cause and countermeasures.
●Cellular Phone
The radio wave may cause malfunction to the device.
Cellular phones and radio sets should be turned off in the room
(building) where medical device is located.
●Static Electricity
In a dry environment (room), static electricity is likely to occur. Take
the following countermeasures.
Both operator and patient should remove any static electricity before
entering the room.
Humidify the room.
xii
EMC Guidance
This equipment complies with IEC 60601-1-2 (2007). However, if portable
transmitter or wireless LAN equipment is used extremely nearby, the
electromagnetic influence may largely exceed the compliance level and may
cause unexpected phenomenon such as noise interference on the waveform,
etc.
Therefore, this equipment should be used in a location specified by each
medical institution. If any unexpected noise interference on the waveform or
failure to the peripheral device occurs, stop using the equipment and follow the
instruction of the technician.
The following is the information relating to EMC (Electromagnetic
Compatibility).
(When using this equipment, verify that it is used within the environment
specified below.)
●Compliance to the Electromagnetic Emissions
The LX-7230 is intended for use in the electromagnetic environment specified
below. The customer or the user of the LX-7230 should assure that it is used in
such an environment.
Emission Test Compliance
RF Emission
RF Emission
Harmonic Emission
IEC 61000-3-2
Voltage Limit /
Flicker Emission
IEC 61000-3-3
CISPR 11
CISPR 11
Group 1
Class A
N/A
N/A
Electromagnetic Environment –
Guidance
The LX-7230 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
This LX-7230 is suitable for use in all
establishments other than domestic and
those directly connected to the public lowvoltage power supply network that
supplies buildings used for domestic
purposes.
xiii
●Compliance to the Electromagnetic Immunity (1)
The LX-7230 is intended for use in the electromagnetic environment specified
below. The customer or the user of the LX-7230 should assure that it is used in
such an environment.
Electromagnetic
Environment
Guidance
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Immunity Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines.
IEC 61000-4-11
Power
Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
IEC 60601-1-2
Test Level
±6kV contact
±8kV air
Compliance
Level
±6kV contact
±8kV air
±2kV:
Power supply lines
±1kV:
N/A
Input/output lines
±1kV:
differential mode
±2kV:
N/A
common mode
<5% UT
(>95% dip in U
)
T
for 0.5 cycle
40% U
(60% dip in U
T
)
T
for 5 cycles
70% U
T
(30% dip in U
N/A
)
T
for 25 cycles
<5% U
(>95% dip in U
T
)
T
for 5sec.
3A/m 3A/m
xiv
●Complian
The LX-7230 is intended for use in the electromagnetic environment specified
below. The customer or the user of the LX-7230 should assure that it is used in
such an environment.
Immunity
Portable and mobile RF
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Where P is the maximum output
Note 1: At 80MHz and 800MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
a)
Field strengths from fixed transmitters, such as base stations for radio
b)
Over the frequency range 150kHz to 80MHz, field strength should be less than
ce to the Electromagnetic Immunity (2)
IEC60601-1-2
Test
affected by absorption and reflection from structures, objects and people.
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the LX-7230 is used exceeds the applicable RF compliance
level above, the LX-7230 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the LX-7230.
3V/m.
Test Level
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
Complianc
e Level
3Vrms
3V/m
Electromagnetic Environment
Guidance
communications equipment should
be used no closer to any part of the
LX-7230, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
d = 1.2
d = 1.2
d = 2.3
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range
Interference may occur in the vicinity
of equipment marked with the
following symbol:
P
80MHz to 800MHz
P
800MHz to 2.5GHz
P
a)
b)
.
, should
xv
●Recomme
nded Separation Distances between Portable and
Mobile RF Communications Equipment and the LX-7230
The LX-7230 is intended for use in an environment in which radiated RF
disturbances are controlled. The customer or the user of the LX-7230 can help
prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters)
and the LX-7230 as recommended below, according to the maximum output
power of the communications equipment.
Rated Maximum
Output Power of
Transmitter
(W)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1 : At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
Note 2 : These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Separation Distance according to Frequency of Transmitter (m)
150kHz to 80MHz
d = 1.2
P
80MHz to 800MHz
d = 1.2
800MHz to 2.5GHz
P
d = 2.3
P
xvi
CONTENTS
Safety Precautions..................................................................................... i
Precaution from Fukuda Denshi....................................................... i
■ Details of the “ELECTRODE?” Message................................. 89
■ List of Setup Items................................................................... 90
xviii
1. General Description
The LX-7230 is a radio telemetry transmitter designed to measure the ECG,
respiration waveform, SpO
two “AA” size (“LR06” size) alkaline batteries.
Information such as ECG measurements, respiration waveform, SpO2
measurements
electrodes and SpO
pulse waveform, battery level, and the conditions of the ECG
,
2
panel.
ECG lead selection is available using the two buttons (Enter and
front panel. (In case of using a 3-electrode lead cable or a 5-electrode chest
lead cable)
The LX-7230 can also function as a transmitter to measure only the
ECG/Respiration without SpO2 or to measure only the SpO2 without
ECG/Respiration.
Before using the LX-7230, read also the operation manual of the patient
monitor at the receiving side thoroughly.
There are two types of transmitters depending on the SpO2 Module, which is
built-in as below.
probe (sensor) are displayed on the LCD of the front
▽) on the
1
1. General Description
Blank Page
2
A
A
2. Names of Parts and Their Functions
Refer to “Safety
Precautions” in this
manual’s preface.
Enter button
Uses for setup.
SpO2 Input Connector
Connects the SpO
cable.
▽ button
Uses for setup.
Neck Strap Hole
ttaches the neck strap.
ECG/RESP Input Connector
Connects the ECG lead cable to
measure ECG and respiration
waveform.
Refer to “Safety Precautions” in this
manual’s preface.
probe (sensor)/relay
2
Chan
nel Number Label
Indicates transmitter channel number.
LCD
Displays measurement waveform and
transmitter information.
Power Switch
Turn ON/OFF the power.
EVENT Button
ctivates the function assigned on the
receiving monitor.
Electrode Position Label
Indicates standard ECG electrode
position.
3
2. Names of Parts and Their Functions
p
Battery Lid Lock Lever
Uses to open/lock the battery
compartment lid.
Battery Compartment Lid
This is the lid for the battery
compartment.
Make sure to close it when the
transmitter is o
erated.
4
3. Preparation
1) Installing the Batteries
The LX-7230 functions with two "AA" size (“LR06” size) alkaline batteries.
With new batteries, the LX-7230 is capable of the following operation.
LX-7230N: approximately 2.5 days
LX-7230KM: approximately 5 days
(However, continuous operating time may be shorter than the above
mentioned time depending on the application of the SpO
WARNING
Unplug the ECG lead cable when the battery compartment lid is
opened. Otherwise, patient leakage current beyond the allowable value
may occur.
Use new "AA" size (“LR06” size) alkaline batteries.
Do not short out the (+) and (-) terminals. It may result in exothermic
heat and fire, the leakage from the batteries may damage the device, or
an explosion from the batteries may occur.
Install the batteries with the correct polarity.
Do not use a disassembled or a damaged battery due to drop or shock.
The leakage from the batteries may damage the device, or an explosion
from the batteries may occur.
Do not use different types of batteries at the same time. The leakage
from the batteries may damage the device, or an explosion from the
batteries may occur.
Remove the exhausted batteries immediately. The leakage from the
batteries may damage the device, or an explosion from the batteries
may occur.
If the transmitter is not in use for a long period of time, remove the
batteries and store the device in an appropriate place. If the batteries
are left in the transmitter for a long period of time, the leakage from the
batteries may damage the device or an explosion from the batteries
may occur.
Make sure to replace the two batteries simultaneously. If a new and
used battery are mixed, a leakage from the batteries may damage the
device or an explosion from the batteries may occur.
CAUTION
Use only alkaline batteries. Other batteries will shorten the continuous
operating time.
Once the power switch is on the OFF position, then open the battery
compartment lid.
Do not replace the batteries with wet hands.
In case of storing the used or unused batteries, make sure that the
terminals are not touching other batteries or metal parts.
probe (sensor).)
2
5
3. Preparation
Lift the lock lever to open the battery
compartment lid as shown in the left picture.
Install new batteries according to the polarity
indication inside the battery compartment.
Make sure to first Insert the battery into the
battery compartment from the minus (-)
terminal as shown in the left picture.
Hook the lock lever on the projection from
the body and press it down until it is
horizontal (flat position).
CAUTION
Make sure that any foreign particles, such as hairs, are not held on
the battery compartment lid and dust is not adhered to the edge of the
lid to prevent water entering into the battery compartment area.
Make sure to only turn ON the LX-7230 after closing the battery
compartment lid.
6
2) Operating Power Switch
Turning the power switch to “ON”
Rotate the power switch to the left until it clicks.
LCD screen turns ON and measurement starts.
Regarding the LCD screen, refer to page 45 (7. Measurement).
The screen automatically turns itself OFF after a minute.
After the power is turned ON, make sure to check the remaining battery level
on the LCD screen.
Refer to the following symbol about the remaining battery level.
Battery Symbol Remaining Battery Level
Full
Getting low but still available
Nearly empty;
Replace the battery
The battery level estimation is in case of using alkaline batteries.
Turning the power switch to “OFF”
Rotate the power switch to the right until it clicks.
3. Preparation
7
3. Preparation
Blank Page
8
4. ECG Monitoring
When the transmitter is used without the SpO2 probe (sensor), it will measure
only ECG and respiration.
CAUTION
When using the transmitter with only the ECG lead cable, SpO2
measurements on the receiving monitor shall be turned off to prevent an
erroneous alarm.
■ Connecting the ECG Lead Cable and Electrodes
The optional ECG lead cables for LX-7230 are as follows.
ECG Lead Cables
AHA color code:
Item No. Applicable Lead Remark
CMT-01HTH-0.8DA
CMT-02HTH-0.8DA
CMT-03HTH-0.8DA
CMT-01FTH-0.8DA
CMT-02FTH-0.8DA
CMT-03FTH-0.8DA
Limb Lead (1CH)
Limb Lead (2CH)
Limb Lead (1CH)
+Chest (1CH)
Limb Lead (1CH)
Limb Lead (2CH)
Limb Lead (1CH)
+Chest (1CH)
WARNING
Use only the specified lead cable from Fukuda Denshi. Otherwise, proper
monitoring may not be performed, and also it may fail defibrillation or cause
a malfunction of the device when the device is used with a defibrillator.
3-electrode Hook Type
(White, Black, Red)
4-electrode Hook Type
(White, Black, Red, Green)
5-electrode Hook Type
(White, Black, Red, Green, Brown)
3-electrode Clip Type
(White, Black, Red)
4-electrode Clip Type
(White, Black, Red, Green)
5-electrode Clip Type
(White, Black, Red, Green, Brown)
9
4. ECG Monitoring
(LA)
(RA)
(RL)
The relations between the attached electrode positions and lead method are
as follows. Attach the electrodes to monitor proper waveform.
For 3-electrode lead cable
For AHA color code electrode position (No. CMT-01HTH-0.8DA, CMT-01FTH-0.8DA)
te
Whi
Red
Standard Limb leads
(LL)
Black
Standard Limb leads can be selected from lead I, lead II, or lead III under
the setting of the device.
Refer to “8. Operation ■Changing Setup ●ECG Display Screen (1)
<<Switching Lead>>” in page 57.
F
or 4-electrode lead cable
For AHA color code electrode position (No. CMT-02HTH-0.8DA, CMT-02FTH-0.8DA)
White
(RA)
Black
(LA
)
Green
Standard Limb leads
Red
(LL)
Two leads measurements, lead I and II are fixed. Lead III, aVR, aVL, and
aVF can be also displayed from the setting on the receiving monitor. For
details, refer to the operation manual of the receiving monitor.
10
Loading...
+ 82 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.