● Before using this equip ment, rea d this opera tion manual thoroughly.
● Keep this manual near the device for future reference.
CAUTION
Federal Law restricts this device to sale by or on the order of a physician.
CAUTION
Users are advised to periodically contact the FCC or specified frequency
coordinator and determine if other or your transmitter frequencies that
may cause interference.
CAUTION
The manufacturers, installers and users of Wireless Medical Telemetry
System equipment are cautioned that the operation of this equipment
could result in harmful interference to other nearby medical devices.
CAUTION:
• This equipment for sale by or on the order of a physician.
• The company and product names used in this manual are trademarks or
registered trademarks.
• If this manual has pages missing or out of order, contact Fukuda Denshi for
replacement.
• Only physician or persons instructed by physicians are allowed to use the
equipment.
• The information contained in this document is subject to change without
Thank you for purchasing this product.
Before using this product, read this operation manual thoroughly for
correct handling and operation.
Safety Precautions
Read the “Safety Precautions” thoroughly before use to ensure correct and
safe use of the product.
Make sure to follow the precautions indicated below, as these are important
messages related to safety.
Failure to follow this message may cause immediate
DANGER
threat of death or serious injury.
Failure to follow this message may result in death or
WARNING
serious injury.
Failure to follow this message may cause injury or
CAUTION
failure to the equipment.
A note is not related to product safety, but provides
information about the correct use and operating
NOTE
procedures to prevent incorrect operation and
malfunction of the equipment.
Precaution from Fukuda Denshi
Fukuda Denshi is liable for the safety, reliability, and performance of its
equipment only if;
Maintenance, modifications, and repairs are carried out by authorized
personnel.
Components are used in accordance with Fukuda Denshi operating
instructions.
If the equipment is used incorrectly and become unusable, Fukuda Denshi is
not liable for the malfunction. Use the equipment only for the purpose specified
in this manual.
i
Graphic Symbols
Refer to the following symbols indicated on the LX-7120(G) for their meanings.
Symbols indicated on the main unit
Symbol Description
Caution: Refer To Accompanying Documents
Indicates the need to refer to the related accompanying
documents before operation.
Type CF Applied Part with Defibrillation-Proof
Indicates that the degree of protection against electric
shock is Type CF Applied Part with defibrillation-proof.
No Alarm Function
Symbols indicated on the LCD screen
Symbol Description
Indicates no alarm function.
Battery Type and Direction
Indicates the battery type and direction.
Heart Rate Synchronization Mark
This mark flashes synchronizing to the heartbeat.
ii
Battery Mark
Indicates the remaining battery level.
Precautions for Safe Operation o f Medical Electrical Equipment
Cautions described here are regarding the general instructions for safety use
to the patient and users. As for cautions about the LX-7120(G), please refer to
the following pages.
CAUTION
1. Users should have a thorough knowledge of the operation before
using this equipment.
2. Pay attention to the following when installing or storing the
equipment.
Do not install or store in an area where the equipment will be subject
to splashing water.
Do not install or store in an area where the environmental
conditions, such as atmospheric pressure, temperature, humidity,
ventilation, sunlight, dust, sodium, sulfur, will adversely affect the
system.
Place the equipment on a stable surface where there is no
inclination, vibration, or shock (including during transportation).
Do not install or store in an area where chemicals are stored or
gases are evolved.
3. Before operating the equipment, verify the following items.
Check the cable connection and polarity to ensure proper operation
of the equipment.
Ensure that all cables are firmly and safely connected. Especially,
recheck the attachment and connection condition of electrode.
Pay special attention when the equipment is used in conjunction with
other equipment as it may cause erroneous judgment and danger.
Check the remaining battery level.
When replacing the battery, make sure that the battery polarity is
correct. Do not charge the battery.
4. During operation of the equipment, verify the following items.
Do not operate the equipment beyond the time period required for
diagnosis and medical care.
Do not pick up and/or swing the equipment pulling/grabbing the
cable part. It may damage the equipment and lead to measurement
error.
Always observe the equipment and patient to ensure safe operation
of the equipment.
If any abnormality is found on the equipment or patient, take
appropriate measures such as ceasing operation of the equipment
and/or detaching the electrode, in the safest way for the patient.
Do not allow the patient to come in contact with other equipments.
iii
CAUTION
5. After using the equipment, verify the following items.
Make sure to turn off the power of the equipment.
When unplugging the cables, do not apply excessive force on the
cable and pull from its connector.
Clean the accessories and cables, and keep them together in one
place.
Keep the equipment clean to ensure proper operation for the next
use.
Make sure to remove the battery if the equipment is not used for a
long time. The leakage from the battery may damage the equipment
or an explosion from the battery may occur.
6. If the equipment is damaged and in need of repair, ensure patient
safety by immediately turning the eq uipment off and remove the
electrodes and/or cable from the patient. User should not attempt
service. Label the unit “OUT OF ORDER” and contact Fukuda
Denshi represen tative.
7. Do not remodel the equipment.
8. Maintenance Check
Make sure to periodically check the equipment, and accessories.
Before reusing the equipment that has been left unused for a while,
make sure that the equipment works normally and safely.
9. When using electrosurgical knives or defibrillator with this
equipment, take care of the following.
To prevent burn injury to the patient, verify proper attachment of
patient ground plate, ECG electrode type for the electrosurgical
knives, and the quantity of gel, output energy for the defibrillator.
Also, verify that a proper ground is selected.
Some types of equipment other than the above may cause
accidental hazards to the patient and operator due to the conditions
of the equipment
equipment and understand the precautionary instructions prior to
use.
Non-Explosion Proof
Never operate the equipment in the presence of flammable anesthetics,
high concentration of oxygen. It may cause an explosion or fire.
Never operate the equipment inside a hyperbaric chamber. It may
. Read the operation manual attached to each
DANGER
iv
cause an explosion or fire.
Never operate the equipment where flammable gas or fluid such as
anesthetic, oxygen, and hydrogen are used. It may cause an explosion
or fire.
Precautions about Magnetic Resona nce Imaging (MRI)
WARNING
Do not operate this equipment in magnetic resonance imaging (MRI)
environments.
When conducting MRI test, remove the electrodes and sensors
connected to the patient (test subject).
The local heating caused by the induced electromotive force may cause
burn injury to the patient (subject). For details, refer to the operation
manual for the MRI testing device.
Electrosurgery Safety
When using electrosurgical instrument, make sure the contact between
the patient and the ground plate is secured. If the connection is
incomplete, the patient may suffer a burn at the electrode site.
When using an electrosurgical instrument, it may misidentify noise from
the electrosurgical instrument as a heartbeat or arrhythmia.
Defibrillation Safety
Use only the lead cable specified by Fukuda Denshi when defibrillating.
If used by unspecified lead cable, the equipment may be damaged,
resulting in a safety hazard.
WARNING
WARNING
When using the defibrillator, keep away from the electrodes or
medicament applied to the patient chest. If this is not possible, remove
the electrodes or medicament before using it.
If the defibrillator paddles are directly in contact with the electrodes or
medicament, an electrical shock may result from the discharged energy.
When using the defibrillator, do not touch the patient and the metal part
of the equipment or cables. Electric shock may result from the
discharged energy.
v
Precautions about the Pacemaker
WARNING
Minute ventilation rate-adaptive implantable pacemakers can
occasionally interact with certain cardiac monitoring and diagnostic
equipment, causing the pacemakers to pace at their maximum
programmed rate. The cardiac monitoring and diagnostic equipment
may possibly send wrong information.
If such event occurs, disconnect the cardiac monitoring and diagnostic
equipment, or follow the procedures described in the operation manual
of the pacemaker.
(For more details, contact FUKUDA DENSHI personnel, your
institution’s professionals, or your pacemaker distributors.)
Reference
“Minute Ventilation Rate-Adaptive Pacemakers”
FDA alerts health professionals that minute ventilation rate-adaptive
implantable pacemakers can occasionally interact with certain cardiac
monitoring and diagnostic equipment, causing pacemakers to pace at their
maximum programmed rate.
[October 14, 1998 (Letter: www .fda.gov/cdrh/safety.html) – FDA]
ECG meter may continue to count the pacemaker rate during
occurrences of cardiac arrest or arrhythmias. Do not rely entirely upon
the ECG meter alarms. Keep pacemaker patients under close
surveillance. Check this manual for disclosure of the pacemaker pulse
rejection capability of this instrument.
Precautions about the LX-7120(G)
WARNING
Do not connect cables not authorized by Fukuda Denshi to any I/O
connector. If done so by mistake, the LX-7120(G) cannot deliver its
maximum performance and may be damaged, resulting in a safety
hazard.
Do not use this equipment with multiple patients simultaneously.
This equipment itself has no alarm function. Do not use it if an alarm
function is necessary. The alarm function with the receiving monitor,
refer to the operation manual of the receiving monitor.
CAUTION
Do not pick up and/or swing the LX-7120(G) pulling/grabbing the cord part.
The cable could break or get disconnected from the LX-7120(G). And it
vi
may hit people or damage other equipment around.
Precautions about Waterproof
Replace the “Battery Compartment Lid” of the LX-7120(G) regularly to
keep the performance of waterproof. If not regularly replaced, the quality
of the lid will deteriorate and cannot keep the waterproof performance.
For details about the regular replacement, contact your local Fukuda
Denshi service representative.
The lid may be damaged from high impact. If the LX-7120(G) is dropped
or is subjected to a high impact, make sure that the lid is not damaged.
Do not use the LX-7120(G) wet. Always wipe the LX-7120(G) with a soft
cloth and dry it thoroughly before use.
CAUTION
Precautions about ECG
When removing electrodes from the patient, remove them carefully and
slowly. Do not apply excessive force to remove them. Otherwise, it may
damage the skin.
There are some cases when the pacemaker pulse cannot be detected
depending on the pacemaker type, pulse voltage, pulse width, electrode
lead type (unipolar, bipolar), electrode placement, or lead method which
causes the pacemaker pulse amplitude to decrease and disables
pacemaker pulse detection.
If signals similar to a pacemaker pulse are present, such as electric
blanket noise or excessive AC frequency noise, these may be
erroneously detected and displayed as a pacemaker pulse. In this case,
check the condition of the electrodes and ECG lead cable to resolve the
CAUTION
cause or turn off the pacemaker detection setting on the receiving
monitor.
Precautions about Outpu t Signal
WARNING
Do not use the output signal of the monitor that receives radio wave signal
from the LX-7120(G) as the trigger signal for IABP, MRI
echocardiographic, or defibrillator for the following reasons.
It may lead to a delay of operating timing due to the delay time of
waveform transmission.
A trigger signal unrelated to the heart rate may be generated due to
the interfusion of spike noise at weak electric field.
vii
Precautions about Accessories and Optional Accessories
WARNING
Use only the accessories, such as ECG Lead cable, specified by Fukuda
Denshi for the LX-7120(G). Otherwise, the LX-7120(G) cannot deliver its
maximum performance and may be damaged, resulting in a safety hazard.
CAUTION
Do not reuse disposable products.
Store the disposable products properly as mentioned in their user
manuals.
Precautions about Battery
WARNING
Use new "AA" size (“LR06” size) alkaline cell.
Install the battery with the correct polarity.
Do not charge the battery. Any attempt to charge the battery may cause
it to leak or break.
Do not short the (+) and (-) terminals. It may result in exothermic heat
and fire.
Do not throw the battery into fire. It may explode.
Precautions about Disposing of Equi pment, Accessories, or
Components
CAUTION
When disposing of the equipment, accessories, or components, use an
industrial waste distributor. Do not dispose of as ordinary waste.
Used disposal items (ECG electrodes, etc.) shall be discarded as
medical waste.
Precautions about Disposing of Batte ry
CAUTION
Obey the local municipal rule to dispose the used dry cell battery.
viii
Precautions for Use of Medical Telem eter
WARNING
The LX-7120(G) transmitter must not be co-located or operated in
conjunction with any other antenna or transmitter.
This equipment complies with FCC/IC radiation exposure limits set forth
for an uncontrolled environment and meets the FCC radio frequency
(RF) Exposure Guidelines and RSS-102 of the IC radio frequency (RF)
Exposure rules. This equipment has very low levels of RF energy that
are deemed to comply without testing of specific absorption rate(SAR).
Operation of LX-7120(G) requires the prior coordination with a
frequency coordinator designated by the FCC for the Wireless Medical
Telemetry Service.
This radio frequency device is susceptible to interference from outside
sources. Interference may prevent the monitoring of patients connected
to this equipment. If a problem exists, contact your local service
representative.
The LX-7120(G) transmits vital signs to the receiving monitor using
radio wave signal. Under unstable radio wave signals, the receiving
monitor will not generate any alarms. This situation may miss sudden
change in the patient's condition and may cause a serious accident.
Under unstable radio wave signals, check the patient status consistently
under this situation. To get stable radio wave signals, make sure to
have a proper telemetry installation.
CAUTION
For installation, make sure the following.
The medical institution (hereinafter referred to as the “Institution”) must
decide the telemetry installation plan for the medical department in order
to prevent interference and interference between transmitters (telemetry
based on destination country’s radio law). When telemetry has already
been installed and been used, radio format, frequency, and antenna power
are required to be examined to prevent interference.
When using telemetry, which requires zone location, the Institution is to
set up the zones as an operation unit for each transmitter to prevent
electronic interference between telemetry throughout the Institution.
When using telemetry, which requires zone location, display and identify
each prepared zone in the equipment.
When laying receiver antenna for each transmitter, the Institution has to
examine the installation so that electronic interference does not occur.
Based on the above examination result, the Institution should install each
receiver antenna as required.
For management, make sure to follow the precautions below.
The Institution should appoint a person (hereinafter referred to as the
“Overall Manager”) to manage the wireless channels for the whole
Institution.
ix
CAUTION
And when using telemetry, which requires zone location, the Institution
should nominate a person (hereinafter referred to as the “Zone Manager”)
to manage the wireless channels in each zone. However, when using such
telemetry in a local Institution, one person can perform both functions.
The Overall Manager and Zone Manager must be selected from people
who understand the characteristics and functionality of telemetry systems,
and are skilled in operating telemetry.
When installing telemetry, the Overall Manager and the Zone Manager
have to understand the precautions for use of telemetry in advance.
The Overall Manager is responsible for maintenance of wireless channel
and storage and maintenance of telemeter in the overall medical facilities
to give proper instructions to the Zone Manager when using telemetry
needing zone alignment, and to the telemetry user when using telemetry
not-needing zone alignment.
The Overall Manager should create a management log (hereinafter
referred to as the “log”), which contains a list of the management status of
the wireless channels for the whole Institution. When changing a wireless
channel, register it in the log and give proper instructions to the Zone
Manager or to the user.
The Zone Manager assumes responsibility for managing the wireless
channels, storing, and managing telemetry.
The Zone Manager assigns the transmitter to the user, and provides
enough education for use inside the zone.
The telemetry user verifies operation of the transmitter/receiver before
use.
The telemetry user, if using the telemetry in a zone location, follows the
instructions of the Zone Manager for the zone and gives instructions to the
patient if required.
When interference or breakdown occurs in telemetry communication, the
user is required to inform the Zone Manager and the Overall Manager of
the problems. The Zone Manager and Overall Manager are to deal with
the problem properly and/or contact their nearest Fukuda Denshi
representative for service.
x
Electromagnetic Compatibility
The performance of this equipment under electromagnetic environment
complies with IEC 60601-1-2 (2007).
Precautions for Safe Operation u nder Electromagnetic Influence
CAUTION
If any sorts of electromagnetic wave, magnetic field, or static electricity
exist around the equipment, noise interference or malfunction of the
equipment may occur. If any unintended malfunction or noise occurs
during monitoring, check the magnetic influence and take appropriate
countermeasures.
The following are examples of the common cause and countermeasures.
●Mobile Phone
The radio wave may cause malfunction to the equipment.
Mobile phones and radio sets should be turned off in the room
(building) where medical device is located.
●Static Electricity
In a dry environment (room), static electricity is likely to occur. Take
the following countermeasures.
Both operator and patient should remove any static electricity before
entering the room
Humidify the room
xi
EMC Guidance
This equipment complies with IEC 60601-1-2 (2007). However, if portable
transmitter or wireless LAN equipment is used extremely nearby, the
electromagnetic influence may largely exceed the compliance level and may
cause unexpected phenomenon such as noise interference on the waveform,
etc.
Therefore, this equipment should be used in a location specified by each
medical institution. If any unexpected noise interference on the waveform or
failure to the peripheral device occurs, stop using the equipment and follow the
instruction of the technician.
The following is the information relating to EMC (Electromagnetic
Compatibility).
(When using this equipment, verify that it is used within the environment
specified below.)
●Compliance to the Electromagnetic Emissions
The LX-7120(G) is intended for use in the electromagnetic environment
specified below. The customer or the user of the LX-7120(G) should assure
that it is used in such an environment.
Emission Test Compliance
RF Emission
Group 1
CISPR 11
RF Emission
Class A
CISPR 11
The LX-7120(G) uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
Electromagnetic Environment –
Guidance
Harmonic Emission
IEC 61000-3-2
Voltage Limit /
Flicker Emission
IEC 61000-3-3
N/A
This LX-7120(G) is suitable for use in all
establishments other than domestic
establishments.
N/A
xii
e
●Compliance to the Electromagnetic Immunity (1)
The LX-7120(G) is intended for use in the electromagnetic environment
specified below. The customer or the user of the LX-7120(G) should assure
that it is used in such an environment.
Immunity Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
IEC 60601-1-2
Test Level
±6kV contact
±8kV air
±2kV:
Power supply lines
±1kV:
Input/output lines
Compliance
Level
±6kV contact
±8kV air
N/A
Electromagnetic
Environment
Guidanc
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30%.
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines.
IEC 61000-4-11
±1kV:
differential mode
±2kV:
common mode
<5% U
(>95% dip in U
T
)
T
for 0.5 cycle
40% U
(60% dip in U
T
)
T
for 5 cycles
70% U
(30% dip in U
T
)
T
for 25 cycles
N/A
N/A
Power
Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
<5% U
(>95% dip in U
T
)
T
for 5sec.
Power frequency
magnetic fields should
be at levels
3A/m 3A/m
characteristic of a
typical location in a
typical commercial or
hospital environment.
xiii
)
●Compliance to the Electromagnetic Immunity (2)
The LX-7120(G) is intended for use in the electromagnetic environment
specified below. The customer or the user of the LX-7120(G) should assure
that it is used in such an environment.
Immunity
Test
Portable and mobile RF
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC60601-1-2
Test Level
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
Complianc
e Level
3Vrms
3V/m
Electromagnetic En vi ro nment
Guidance
communications equipment should
be used no closer to any part of the
LX-7120(G), including cables, than
the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended Separation Distance
d = 1.2
d = 1.2
d = 2.3
P
80MHz to 800MHz
P
800MHz to 2.5GHz
P
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range
Interference may occur in the vicinity
of equipment marked with the
following symbol:
b)
.
a)
, should
xiv
Note 1: At 80MHz and 800MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the LX-7120(G) is used exceeds the applicable RF
compliance level above, the LX-7120(G) should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the LX-7120(G).
b)
Over the frequency range 150kHz to 80MHz, field strength should be less than
3V/m.
●Recommended Separation Distances between Portable and
Mobile RF Communications Equipment and the LX-7120(G)
The LX-7120(G) is intended for use in an environment in which radiated RF
disturbances are controlled. The customer or the user of the LX-7120(G) can
help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters)
and the LX-7120(G) as recommended below, according to the maximum
output power of the communications equipment.
Rated Maximum
Output Power of
Transmitter
(W)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Separation Distance according to Frequency of Transmitter (m)
150kHz to 80MHz
d = 1.2
P
80MHz to 800MHz
d = 1.2
P
800MHz to 2.5GHz
d = 2.3
P
Note 1 : At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
Note 2 : These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
xv
Blank Page
xvi
CONTENTS
Safety Precautions ..................................................................................... i
Precaution from Fukuda Denshi ....................................................... i
Graphic Symbols .............................................................................. ii
Precautions for Safe Operation of Medical Electrical Equipment .... iii
Non-Explosion Proof ....................................................................... iv
Precautions about Magnetic Resonance Imaging (MRI) ................. v
Electrosurgery Safety ...................................................................... v
Defibrillation Safety ......................................................................... v
Precautions about the Pacemaker .................................................. vi
Precautions about the LX-7120(G) ................................................. vi
Precautions about Waterproof ....................................................... vii
Precautions about ECG ................................................................. vii
Precautions about Output Signal ................................................... vii
Precautions about Accessories and Optional Accessories ........... viii
Precautions about Battery ............................................................. viii
Precautions about Disposing of Equipment, Accessories, or Components
................................................................................................. viii
Precautions about Disposing of Battery ........................................ viii
Precautions for Use of Medical Telemeter ...................................... ix
Electromagnetic Compatibility ............................................................. xi
Precautions for Safe Operation under Electromagnetic Influence ... xi
EMC Guidance ............................................................................... xii
1. General Description ............................................................................. 1
2. Names of Parts and Their Functions .................................................... 3
■ Details of the “ELECTRODE?” Message ................................. 53
■ List of Setup Items ................................................................... 54
xviii
1. General Description
The LX-7120(G) is a radio telemetry transmitter designed to measure the ECG
and respiration waveform with one (1) “AA” size (“LR06” size) alkaline battery.
Information such as ECG measurements, respiration waveform, battery level,
and the conditions of the ECG electrodes are displayed on the LCD of the front
panel.
ECG lead selection is available using the two buttons (Enter and
front panel. (In case of using a 3-electrode lead cable or a 5-electrode chest
lead cable)
Before using the LX-7120(G), read also the operation manual of the patient
monitor at the receiving side thoroughly.
▽) on the
External Appearance
1
1. General Description
Blank Page
2
2. Names of Parts and Their Functions
Channel Number Label
Indicates transmitter
channel number.
LCD
Displays measurement
waveform and transmitter
information.
Enter button
Uses for setup.
Neck Strap Hole
Attaches the neck strap.
ECG/RESP Input Connector
Connects the ECG lead cable
to measure ECG and
respiration waveform.
▽ button
Uses for setup.
EVENT Button
Activates the function
assigned on the receiving
monitor.
Battery Compartment Lid
This is the lid for the battery
compartment.
Make sure to close it when the
transmitter is operated.
Refer to “Safety
Precautions” in this
manual’s preface.
Battery Lid Lock Lever
Uses to open/lock the
battery compartment lid.
Power Swit c h
Turn ON/OFF the power.
Electrode Position Label
Indicates standard ECG
electrode position.
3
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.