Fuji Film 01000005 User Manual

Operation Manual
2nd Edition : July 2013
For Safe Operation
System Conguration (Product Overview)
Basic Operation
Daily Inspection and Maintenance
Appendix
897N102280A
Maintenance and Inspection
This Operation Manual describes details on how to operate the DR-ID 900PU and cautions to be observed when operating it. Please read the Operation Manual thoroughly before actually operating the DR-ID 900PU along with “DR-ID 300CL Reference Guide” and other manuals for the related products. After reading this manual, store it nearby the DR-ID 900PU so that you can see it whenever necessary.
DR-ID 900PU Operation Manual
ii
897N102280A
Introduction
DR-ID 900PU is an X-ray equipment which acquires a general radiograph from the indirect­conversion at panel sensor.
DR-ID 900PU consists of DR-ID 900MP and DR-ID 911SE/DR-ID 601SE/DR-ID 602SE/DR-ID 611SE/ DR-ID 613SE. DR-ID 911SE, communication mode or wired communication mode is available. When used in wireless communication mode, an access point* required. *1 In the countries other than the U.S., an access point is not included as a component of the
system. For details including installation, consult our ofcial dealer. Each at panel sensor complies with IEC 62220-1 (MEDICAL ELECTRICAL EQUIPMENT ­CHARACTERISTICS OF DIGITAL X-RAY IMAGING DEVICES - ) as a general X-ray radiography equipment.
The detector of at panel sensors features 150 micron pixel pitch, a wide 16-bit dynamic range and exposure times up to 3.8 seconds.
This Operation Manual includes descriptions of matters necessary when using the DR-ID 900PU such as the equipment overview, operation procedures and precautions to observe, as well as daily inspections and maintenance. Accompanying documents were originally drafted in the English language.
DR-ID 601SE, DR-ID 602SE, DR-ID 611SE and
1
, battery pack (optional) and battery charger (optional) are
Installation may only be conducted by authorized service personal.
CAUTIONS
DR-ID 613SE
:Wireless
1. No part or all of this manual may be reproduced in any form without prior permission.
2. The information contained in this manual may be subject to change without prior notice.
3. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from installation, relocation, remodeling, maintenance, and repair performed by other than dealers specied by FUJIFILM Corporation.
4.
FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM Corporation products due to products of other manufacturers not supplied by FUJIFILM Corporation.
5.
FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from remodeling, maintenance, and repair using repair parts other than those specied by FUJIFILM Corporation.
6. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from negligence of precautions and operating methods contained in this manual.
7. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use under environment conditions outside the range of using conditions for this product such as power supply, installation environment, etc. contained in this manual.
8. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from natural disasters such as res, earthquakes, oods, lightning, etc.
This system is classied as a medical device under EC Directive 93/42/EEC.
Process waste correctly, as stipulated by local law or any regulations that apply.
Caution : Rx Only in the United States (Federal law restricts this device to sale by or on the order
of a physician.)
Open-Source Software Used in This Product
This product uses third party’s software that is made available as open source software or free software. For information on open source software used in this product, please see the attached CD. Source codes for certain type of open source software used in this product are available at delivery cost. If you would like to receive such source codes, please contact FUJIFILM dealer or the service representatives at the agency from which you purchased this product. (Please be noted that any inquiries concerning the contents of source codes should be directed to original licensers of open source software.)
Note : FUJIFILM has successfully performed verication and validation testing on all third party
software and has conrmed its suitability to be used in this system.
Trademarks
All company names and product names described in this manual are the trademarks or registered trademarks of FUJIFILM Corporation or their respective holders. Windows is the registered trademark of US Microsoft Corporation in the U.S.A. and other countries.
Copyright © 2013 FUJIFILM Corporation. All rights reserved.
DR-ID 900PU Operation Manual 897N102280A
iii
Contents
Introduction ...........................................................................................................................iii
Chapter 1 For Safe Operation
1.1 Safety ......................................................................................................................1-1
1.2 Electromagnetic Compatibility (EMC) ......................................................................1-9
1.2.1 DR-ID 900PU ...........................................................................................................1-9
Chapter 2 System Configuration (Product Overview)
2.1 DR-ID 900PU .......................................................................................................... 2-1
2.1.1 System Conguration ..............................................................................................2-1
2.2 Unit Names and the Functions ................................................................................ 2-2
2.3 Locations of Labels and Signs ................................................................................2-3
2.3.1 Locations of Labels ..................................................................................................2-3
2.3.2 Safety and Other Symbols .......................................................................................2-6
Chapter 3 Basic Operation
3.1 Preparing the Flat Panel Sensor ............................................................................. 3-1
3.1.1 Type of Flat Panel Sensor .......................................................................................3-1
3.1.2 Number of the Connectable Flat Panel Sensors .....................................................3-1
3.1.3 Connecting/Disconnecting the Flat Panel Sensor Connector ..................................3-2
3.1.4 Inserting/Removing the Flat Panel Sensor into/
from the Radiographic Examination Stand ..............................................................3-3
[1] In the case of the at panel sensor DR-ID 911SE ..............................................3-3
[2] In the case of the at panel sensors DR-ID 601SE/DR-ID 602SE/
DR-ID 611SE/DR-ID 613SE ...............................................................................3-3
3.1.5 Changing the Direction of the Flat Panel Sensor Connector ...................................3-5
3.1.6 Charging the Battery Pack (optional) for the Flat Panel Sensor ..............................3-5
3.1.7 Installing/Removing the Battery Pack (optional) for the Flat Panel Sensor .............3-6
3.2 Starting Up and Shutting Down the DR-ID 900PU .................................................. 3-7
3.2.1 Starting Up the DR-ID 900PU ..................................................................................3-7
3.2.2 Shutting Down the DR-ID 900PU ............................................................................3-7
Chapter 4 Daily Inspection and Maintenance
4.1 Daily User Inspection and Maintenance .................................................................. 4-1
4.1.1 Periodical Inspection ................................................................................................4-1
Appendix A Specifications
A.1 Specications ..........................................................................................................A-1
A.1.1 Reduced Equivalent (DR-ID 900) ........................................................................... A-1
A.1.2 Power Supply Conditions ........................................................................................ A-1
A.1.3 Environmental Conditions ....................................................................................... A-1
A.1.4 Image performance ................................................................................................. A-1
Appendix O Use of Optional Items
O.1 Optional Items ....................................................................................................... O-1
O.2 Using the Retaining Bracket for MP ....................................................................... O-2
Maintenance and Inspection
DR-ID 900PU Operation Manual
iv
897N102280A
Chapter 1
For Safe Operation
1.1 Safety
Before using the DR-ID 900PU, read this section thoroughly to ensure that you use the product properly.
Electric Shock Warnings and Cautions
WARNING
The power supply to the DR-ID 900PU is AC100 to 240V. To avoid electric shocks, users should always take the following precautions:
Never open any covers of the equipment.
Install the equipment in a location where it will not be exposed to water.
Check that the equipment is securely earthed.
Check that all of the cords and cables are completely and securely connected.
Keep the image processing unit and the control cabinet out of reach of patients.
WARNING
Do not touch the patient’s body while touching the control cabinet. Otherwise, the patient may receive an electric shock.
WARNING
Do not use a multiple tap connector or extension cable for powering the devices constituting the system. Otherwise, re or electric shock may occur due to the electrical load exceeding the allowable limit.
1
For Safe Operation
WARNING
Observe the following precautions when using the cables.
Turn off each unit before connecting/disconnecting the cable. Do not touch the plug and connector with wet hands. Otherwise, electric shock may result, causing death or severe injury.
Hold the plug or connector when removing the cable. Pulling the cable or carrying by holding it may damage the cable, causing re or electric shock.
Do not damage or remodel the cable. Do not place a heavy object on the cable or lay it under the at panel sensor. Do not step on, pull, forcibly bend, or bundle the cable. Otherwise, fire or electric shock may result.
Do not use the flat panel sensor for the radiographic examination stand if its cable becomes overloaded. Otherwise, the cable may be damaged, causing re or electric shock.
WARNING
Do not turn on the system with dew condensation on the flat panel sensor. Otherwise, fire or electric shock may result.
\
Do not use the equipment in a location where metal particles could come into the equipment. This may cause an electric shock.
WARNING
DR-ID 900PU Operation Manual 897N102280A
1-1
WARNING
Do not disassemble or remodel the equipment. Otherwise, re or electric shock may result. Keep away from the parts inside the product, which may cause electric shock. If you touch them accidentally, death or severe injury may result.
WARNING
1
For Safe Operation
Do not hit or drop the equipment or subject it to severe shock. Otherwise, the equipment may be damaged. If the damaged equipment is used, re or electric shock may result.
Do not use the at panel sensor when the battery cover and the battery pack are not installed. Doing so may cause electric shock
Make sure to use the optional parts and accessories recommended by us. Failure to use the optional parts and accessories recommended by us may result in electric shock or injury.
As the cables of the equipment are long, be careful not to entangle the cables during use. Also, be careful not to trip over the cables. Falls could result in injury.
Be sure to follow the specied procedure when turning off the power of the equipment. Otherwise, the at panel sensor could be damaged.
WARNING
WARNING
CAUTIONS
CAUTIONS
CAUTIONS
Do not store magnetic media near the DR system and control cabinet. Otherwise, magnetism generated by the equipment may cause the data to be lost.
CAUTIONS
Keep the equipment away from patient’s body uids, chemicals, water, etc. Otherwise, it may become damaged, causing re or electric shock. If necessary, protect the at panel sensor by covering it with a disposable bag.
Explosion Warnings
WARNING
Because this equipment is not explosion-proof, do not use combustible and explosive gases near the equipment.
WARNING
Flammable gasses may stay in the room after disinfection. Ventilate the room well before powering on the equipment following disinfection.
DR-ID 900PU Operation Manual 897N102280A
1-2
Warnings for Abnormalities
WARNING
If any of the following occurs, immediately turn off the power of each unit, unplug the power cable from the outlet, and then contact a FUJIFILM dealer.
When smoke, strange odor, or abnormal sound is present.
When a foreign object (such as a metal object) or liquid enters the product.
When the equipment is dropped or hit and is damaged.
1
Installation Precautions
CAUTIONS
Do not install the equipment in a location with the following conditions.
Where the temperature changes sharply.
Close to heat sources such as a heater.
Where the equipment may be exposed to water due to water leakage or ingress.
Where corrosive gas may be generated.
Where there is excessive dust.
Where the equipment is subject to frequent or excessive vibration/shock.
Where the equipment is exposed to direct sunlight.
CAUTIONS
For veterinary or mobile applications, please contact a certied FUJIFILM service representative.
CAUTIONS
Use the system indoor in wireless communication mode. For details, contact a FUJIFILM dealer.
CAUTIONS
For Safe Operation
Do not place any object in a place where removal of the power cable is prevented.
Connection Instructions
WARNING
Make sure that the devices to be connected to the equipment are authorized for connection.
WARNING
Connect the DR-ID 900PU only to the designated control cabinet (DR-ID 900MC). When the DR-ID 900PU is connected to X-ray equipment, make sure that the equipment complies with IEC 60601-1.
Warning regarding radiographic examination stands
WARNING
In the case of using this product in combination with an upright-type or bed-type radiographic examination stand, select an upright-type or bed-type radiographic examination stand whose aluminum equivalent, of the front panel of the upright-type radiographic examination and the top of the bed-type radiographic examination or the like, complies with IEC 60601-1-3: 2008/EN 60601-1-3: 2008.
DR-ID 900PU Operation Manual 897N102280A
1-3
Precautions on External Network Connection
When a setting of the network to which the equipment is connected has been changed, check that the change does not affect the system operation and take measures if necessary. The setting change may include the following:
1
For Safe Operation
Change of connection destination
Addition of devices
Removal of devices
Update of devices
Upgrade of devices
Software Precautions
Connect to the Ethernet Network of 1000BASE-T, 100BASE-TX, or 10BASE-T prescribed in the IEEE standard 802.3. Do not connect telephone lines to LAN connector. Only UTP-type straight LAN cables of 4-pair Category 5 cable (CAT 5E) or higher are appropriate for connection to this connector.
CAUTIONS
CAUTIONS
CAUTIONS
After connecting this system to the network with other systems,conrm that the other systems are not affected. If they are affected, take countermeasures such as network separation.
System Isolation Instructions
WARNING
To ensure complete system isolation, never install any unauthorized accessories or other such items. When it is necessary to install authorized accessories or optional items, contact a FUJIFILM dealer.
WARNING
Keep equipment other than those used for patients out of their reach to ensure appropriate system isolation.
WARNING
Do not move the Console from where it is installed.
In normal use, have a patient take a proper positioning for exposure. The operator should operate the system in a place where safety from radiation is ensured. The operator should also make sure before exposure that no one but the patient is in the exposure area and the operating area of the system.
DR-ID 900PU Operation Manual 897N102280A
1-4
WARNING
Software Precautions
CAUTIONS
Do not install additional software to the system. Do not uninstall any of the software preinstalled in the system. The system is preinstalled with the appropriate software. If other software is installed or if the existing software is uninstalled, various operational errors may result. Various operational errors may result.
1
Disinfection Instructions
WARNING
Conrm that the respiratory density of disinfectant including solvent is under legal regulation. Certain disinfectants may damage health. When using a disinfectant, follow instructions supplied by the manufacturers.
WARNING
Ethanol for disinfection is recommended as the disinfectant. Carefully read the instructions and cautions supplied with the disinfectant before use. Do not use the following when disinfecting. Otherwise, the equipment may be damaged, and its quality, performance, and safety cannot be guaranteed.
Chloric disinfectant which is strongly corrosive to metals and rubber. parts.
Disinfectant whose use on metals, plastics, and coating is forbidden according to the
instructions supplied with the disinfectant.
Formalin gas and disinfectant sprays that may get inside the equipment.
Ultraviolet sterilizers
CAUTIONS
Clean the sensor unit of the at panel sensor with ethanol for disinfection, etc. for each patient to prevent infection.
For Safe Operation
Precautions for Charging the Battery
CAUTIONS
Use the battery charger recommended by FUJIFILM Corporation.
For details on operations, refer to the instruction manual for the battery charger.
Do not charge the battery pack near fire or under strong sunshine. If the built-in protection
mechanisms are activated by a high temperature, the battery pack cannot be charged. Also, if the built-in protection mechanisms are damaged, the battery pack may be charged with extremely high current and voltage, and abnormal chemical reactions may occur inside the battery pack, causing it to overheat, emit smoke, explode or ignite.
Immediately stop charging the battery pack, if charging is not completed within the specified
time. Otherwise, the battery pack may overheat, emit smoke, explode or ignite.
Do not use the at panel sensor near the AC adapter.
Do not use the broken battery charger.
Battery Pack (Optional) Precautions
WARNING
If this equipment is not in use for while, store it with the battery pack removed. Not removing the battery pack may cause malfunction.
DR-ID 900PU Operation Manual 897N102280A
1-5
Be sure to observe the following cautions when using the battery pack (optional).
Do not use the battery pack with any device other than those designated by FUJIFILM
Charge the battery pack only with the designated battery charger. If the battery pack is
1
For Safe Operation
When storing the battery pack (125N100050) for a long period, charge the battery fully,
Do not leave the removed battery pack in a car or other place exposed to high temperature.
Use and store the battery pack only in the environmental conditions specied by FUJIFILM
When disposing of the battery pack, consult our ofcial dealer.
Do not disassemble or remodel the battery pack. The battery pack is equipped with built-in
Do not connect the positive (+) and negative (-) terminals of the battery pack using any
Do not use or leave the battery pack in a place where it is exposed to high temperature (80°C
Do not immerse the battery pack in water or seawater, and do not allow it to become wet.
Do not pierce the battery pack with a nail, hit it with a hammer, or step on it. Otherwise, the
Do not subject the battery pack to strong impact or throw it. If the built-in protection
Do not use an apparently damaged or deformed battery pack. Otherwise, the battery pack
Do not solder the battery pack directly. Otherwise, its insulator may melt, or its gas release
Do not reverse the positive (+) and negative (-) terminals of the battery pack. Otherwise,
The battery pack has a predetermined polarity. If you cannot connect the battery pack to the
Do not connect the battery pack to an electrical outlet or cigarette lighter socket in a car.
Do not use the battery pack for equipment other than those specied. Otherwise, the
CAUTIONS
Corporation.
charged under the charging conditions (voltage, current and charging method) different from those specified by FUJIFILM Corporation, the battery pack may emit smoke, ignite, explode or leak fluid.
remove it from the at panel sensor and then store it in a cool and dark place. Recharge the stored battery every six months or every year. Otherwise a decrease in battery capacity or other problems may result.
Corporation. If the battery pack is used or stored in a place where it is exposed to high temperature, the battery pack may emit smoke, ignite, explode or leak fluid.
safety and protection mechanisms. If they are damaged, the battery pack may overheat, emit smoke, explode or ignite.
metal object, such as a wire. Do not carry or store the battery pack together with metal objects such as necklaces or hairpins. Otherwise, the battery pack may short-circuit and overcurrent may flow, causing the battery pack to overheat, emit smoke, explode or ignite. Metal objects such as necklaces or hairpins may also become hot. Do not throw the battery pack into re or expose it to excessive heat. Otherwise, its insulator may melt, its gas release vent or safety mechanisms may be damaged, and/or its electrolyte may catch fire, causing the battery pack to overheat, emit smoke, explode or ignite.
or higher), such as re or a heater. If the resin separator is damaged due to heat, the battery pack may short-circuit, causing it to overheat, emit smoke, explode or ignite.
If the built-in protection mechanisms are damaged, the battery pack may overheat, emit smoke, explode or ignite.
battery pack may be damaged or deformed and short-circuit, causing it to overheat, emit smoke, explode or ignite.
mechanisms are damaged, the battery pack may be charged with extremely high current and voltage, and abnormal chemical reactions may occur inside the battery pack, causing it to overheat, emit smoke, explode or ignite.
may overheat, emit smoke, explode or ignite.
vent or safety mechanisms may be damaged, causing the battery pack to overheat, emit smoke, explode or ignite.
the battery pack may be reverse-charged during charging. As a result, abnormal chemical reactions may occur inside the battery pack, or extremely high current may ow during discharging, causing it to overheat, emit smoke, explode or ignite.
battery charger or other equipment, do not connect the battery pack forcefully. Make sure that the terminals are correctly oriented. If the battery pack is connected in reverse, it will be reverse-charged, and abnormal chemical reactions may occur inside the battery pack, causing it to overheat, emit smoke, explode or ignite.
Overcurrent may flow to the battery pack due to high voltage applied, causing the battery pack to overheat, emit smoke, explode or ignite.
guaranteed performance will be reduced and/or the service life will be shortened. Depending on the equipment to which the battery pack is connected, extremely high current may flow,
DR-ID 900PU Operation Manual 897N102280A
1-6
causing the battery pack to be damaged, overheat, emit smoke, explode or ignite.
If the electrolyte leaked from the battery pack enters the eyes, do not rub them. Wash the
eyes immediately with clean water such as tap water, and consult a doctor. Otherwise, eye injury may result.
Do not use the battery pack in combination with a primary battery such as a dry battery or
other battery of a different capacity, type and/or brand. Otherwise, the battery pack may be overcharged during charging, and abnormal chemical reactions may occur inside the battery pack, causing it to overheat, emit smoke, explode or ignite.
Do not put the battery pack in a microwave oven or high-pressure container. Otherwise, the
battery pack may be rapidly heated or damaged, causing it to overheat, emit smoke, explode or ignite.
If the battery pack leaks or emits an unusual odor, remove it from re immediately.
Otherwise, the leaked electrolyte may catch re, causing the battery pack to overheat, emit smoke, explode or ignite.
If you notice an unusual odor, heat, discoloration, deformation or any other abnormality during use, charging or storage, remove the battery pack from the equipment or battery charger, and stop using it. Otherwise, the battery pack may overheat, emit smoke, explode or ignite.
Do not use the battery pack exposed to a strong magnetic eld of an MRI system, etc.
Do not use the battery pack immersed in liquid.
.
Other Warnings and Cautions
WARNING
1
For Safe Operation
No modication of this equipment is allowed.
CAUTIONS
Install the system in accordance with what is provided by IEC60601-1-1. Contact a FUJIFILM dealer for installation and relocation (except the at panel sensor) of the system.
CAUTIONS
Do not hit or drop the equipment. Otherwise, injury or damage to images, etc. may result.
CAUTIONS
Be sure to inspect the system periodically. To assure optimum performance of the equipment, it is necessary to systematically perform maintenance and inspection. For information on maintenance and inspection, contact a FUJIFILM dealer.
CAUTIONS
The institution where the equipment is installed is responsible for its use and maintenance. In addition, this equipment should not be used by persons other than doctors or suitably trained staff.
CAUTIONS
Do not place the cable terminal on the oor.Doing so may cause infection. Also, periodically clean the cable and the terminal.
CAUTIONS
The institution where the equipment is installed is responsible for its use and maintenance. In addition, this equipment should not be used by persons other than doctors or suitably trained staff.
DR-ID 900PU Operation Manual 897N102280A
1-7
1
Contraindications and Prohibitions
No contraindications present.
Classification
According to the type of protection against electrical shock Wired communication mode: Class 1 equipment (stationary) Wireless communication mode: Internal power supply
For Safe Operation
According to the degree of protection against electrical shock Type B applied part
According to the degree of protection against harmful ingress of water IPX0
According to the degree of safety of application in the presence of a ammable anesthetics mixture with air or with oxygen or nitrous oxide.
Equipment not suitable for use in the presence of a ammable anesthetics mixture with air or with
oxygen or nitrous oxide.
According to the mode of operation
CONTINUOUS OPERATION
DR-ID 900PU Operation Manual 897N102280A
1-8