FSN FS-L190 D, FS-L190 DT, FS-L240 D, FS-L240 DT, FS-L260 D User Manual

...
MEDICAL LCD MONIT
MEDICAL LCD MONIT
OR
OR
USER'S GUIDE
Before connecting, operating or adjusting this product, please read this instruction booklet carefully and completely .
Rev . 4.0, August 201 13
Model definition
F S- L X X X* D T
1 2 3 4 5 6
1. FS : Monitor manufacturer.
2. L : Panel manufacturer.
3. XXX : Display size.
4. * : Signal input option.
5. D : Medical grade.
6. T : Touch screen installed.
MEDICAL LCD MONITOR
FS-L190*D (19”) FS-L190*DT(19”) FS-L240*D (24”) FS-L240*DT(24”) FS-L260*D (26”) FS-P260*D (26”) FS-L320*D (32”) FS-L420*D (42”) FS-L550*D (55”)
English - 2
Symbol Definitions ………………………………………………………………3 Safety Instructions ………………………………………………………………5 Caution………………………………………………………………………………8 FCC Information …………………………………………………………………11 Parts
FS-L190*D / FS-L190*DT……………………………………………………16
FS-L240*D / FS-L240*DT……………………………………………………17
FS-L260*D / FS-P260*DT ……………………………………………………18
FS-L320*D ……………………………………………………………………19
FS-L420*D / FS-L550*D ……………………………………………………20
Connector
FS-L190*D / FS-L240*D / FS-L260*D / FS-P260*D / FS-L320*D ……21
FS-L190*DT / FS-L240*DT / FS-P260*DT ……………………………… 22
FS-L420*D / FS-L550*D …………………………………………………… 23
Mechanical Product Drawing
FS-L190*D / FS-L190*DT ………………………………………………… 24
FS-L240*D / FS-L240*DT ………………………………………………… 25
FS-L260*D …………………………………………………………………… 26
FS-L260*D(LED) / FS-P260*D(LED)……………………………………… 27
FS-L320*D …………………………………………………………………… 28
FS-L420*D …………………………………………………………………… 29
FS-L550*D …………………………………………………………………… 30 Control ……………………………………………………………………………31 Power management ……………………………………………………………35 OSD ………………………………………………………………………………36 Standard Signal table …………………………………………………………51 Signam connector Pin Assignments …………………………………………52 Specification ……………………………………………………………………55 Cleaning instructions ……………………………………………………………63 DVI ②(Fiber) install guide (Rx) ………………………………………………66
Table of Contents
3- English
Symbol Definitions
The following symbols appear on the product, its labels, or the product package.
Each symbol carries a special definition, as defined below.
Dangerous : High Voltage.
Consult accompanying documents.
Direct Current.
lndicates protective earth ground.
DC Power control switch.
lndicates equipotential earth ground.
Serial Number.
Consult the operating instructions.
Top-Bottom.
Fragile.
Maximum Stacking.(19”/24”/26”)
Do not get wet.
Maximum Stacking.(32”/42”/55”)
Representative in the USA.
Authorised representative in the European community.
Manufacturer.
Indicates the manufacturing date.
Power adaptor.
English - 4
Indicates proof of conformity to applicable European Economic Community Council directives and to harmonized standards published in the official journal of the European Communities.
Medical LCD monitor is in accordance with UL 60601-1 and CAN/CSA C22.2 No.601.1 in regards to electric shock, fire hazards, and mechanical hazard.
Tested to comply with FCC Class B standard.
This symbol indicates that the waste of medical LCD monitor must not be disposed as unsorted municipal waste and must be collected separately. Please contact the manufacturer or other authorized disposal company to decommission your medical LCD monitor.
Language: English Note: A printed copy of the manual in English is provided with the product. Users within EU member state, please contact local distributor or Foreseeson Custom Displays, Inc. for other languages or refer to the CD manual enclosed with the product. This applies to EU member state where the product has been purchased through authorized channels.
Indicates the display is approved according to the CCC regulations.
China RoHS labels.
Humidity limitation.
Temperature limitation.
Atmospheric pressure limitation.
5- English
Safety Instructions
On Safety
1. Before connecting the AC power cord to the DC adapter outlet make sure the voltage designation of the DC adapter corresponds to the local electrical supply.
2. Never insert anything metallic into the cabinet openings of the medical LCD monitor. Doing so may create the danger of electric shock.
3. To reduce the risk of electric shock, do not remove cover. No user-serviceable parts inside. Only a qualified technician should open the case of the medical LCD monitor.
4. Never use your medical LCD monitor if the power cord has been damaged. Do not allow anything to rest on the power cord, and keep the cord away from areas where people can trip over it.
5. Be sure to hold the plug, not the cord,when disconnecting the medical LCD monitor power cord from an electric socket.
6. Unplug your medical LCD monitor power cord when it is going to be left unused for an extended period of time.
7. Unplug your medical LCD monitor power cord from the AC outlet before any service.
8. If your medical LCD monitor does not operate normally, in particular, if there are any unusual sounds or smells coming from it, unplug it immediately and contact an authorized dealer or service center.
9. Please contact the manufacturer if the set should be installed in an inaccessible
area.
Warning
Do not touch input or output connectors and the patient simultaneously.
English - 6
Warning
This medical LCD monitor is intended for connection to input/output signals and other connectors that comply with relevant IEC standard (e.g.,IEC60950 for IT equipment and IEC60601 series for medical electrical equipment). In addition, all such combination-system shall comply with the standard IEC 60601-1-1, safety requirements for medical electrical systems. Any person who has formed a combination-system is responsible for the system to comply with the requirements of IEC 60601-1-1. If in doubt, contact qualified technician or your local representative.
On installation
1. Openings in the medical LCD monitor cabinet are provided for ventilation. To prevent overheating, these openings should not be blocked or covered. If you put the medical LCD monitor in a bookcase or some other enclosed space, be sure to provide adequate ventilation.
2. Put your medical LCD monitor in a location with low humidity and a minimum of dust.
3. Do not expose the medical LCD monitor to rain or use it near water (in kitchens, near swimming pools, etc.). If the medical LCD monitor accidentally gets wet, unplug it and contact an authorized dealer immediately.
You can clean the medical LCD monitor with a damp cloth if necessary,
but be sure to unplug the medical LCD monitor first.
4. Place your medical LCD monitor near an easily accessible AC outlet.
5. High temperature can cause problems. Don't use your medical LCD monitor in direct sunlight and keep it away from heaters, stoves, fireplaces,and sources of heat.
6. Don’t place your medical LCD Monitor on an unstable stand, Medical LCD monitor may malfunction or fall.
7. This medical LCD monitor should not topple over when tilted at a 5
° angle, in any
position, during
NORMAL USE, excluding transport.
8. In the position specified for transport, medical LCD monitor shall not overbalance when tilted at a 10
degree
angle.
9.
When you carry this product, please use both handles on left and right side of
the product which carrying more two persons. If you want the product to be installed on another place, please call A/S center.
7- English
10. Do not use other cable or accessary that are not provided.
11. Do not lay this monitor on the other equipment.
Environmental Conditions for operation and Storage
- Temperature range within 0°C to 40°C(operation), -20°C to 60°C(storage)
- Relative humidity range 10% to 85% Atmospheric pressure range within 500 to 1060hPa.
Intended Use
- This Medical LCD Monitor is an accessory intended for use with Medical Equipment to display alphabetical, numerical and graphical data.
English - 8
This symbol alerts the user that important literature concerning the operation of this unit has been included. Therefore, it should be read carefully in order to avoid potential problems.
This symbol warns user that un-insulated voltage within the unit may have sufficient magnitude to cause electrical shock. Therefore, it is dangerous to make contact with any part inside the unit. To reduce the risk of electric shock, DO NOT remove cover (or back). There are no user serviceable parts inside. Refer servicing to qualified service personnel.
To prevent fire or shock hazards, do not expose this unit to rain or moisture. Also, do not use this unit's polarized plug with an extension cord receptacle or other outlets unless the prongs can be fully inserted. The display is designed to meet the medical safety requirements for a patient vicinity device. This device may not be used in connection with life support equipment.
CAUTION
Underwriters Laboratories (UL) Classification:
UL safety Compliance: This medical LCD monitor is U.L. Classified WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1/CAN/CSA C22.2 NO. 601.1
9- English
Recycling
Follow local governing ordinances and recycling plans regarding the recycling or disposal of this equipment.
EEC Safety Compliance
This medical LCD monitor unit meets the requirements of EN-60601-1 so as to conform to the Medical Device Directive 93/42/EEC (general safety information). Use 120V rating 5-15P type plug only in the U.S
This medical LCD monitor complies to the above standards only when used with the supplied medical grade power supply.
19”(FS-L190*D / FS-L190*DT) - JMW190KA1200F02(BRIDGE POWER CORP.) 24”/26”(FS-L240*D / FS-L240*DT / FS-L260*D)
JMW1150KA2400F04(BRIDGE POWER CORP.) 26”LED (FS-P260*D / FS-L260*D) - BPM150S24F10(BRIDGE POWER CORP.) 32”(FS-L320*D) -JMW1180KA2400F01(BRIDGE POWER CORP.)
Caution: Make sure the power cord is the correct type that is required in your area. This medical LCD monitor has a universal power supply that allows operation in either 100-120V AC or 200-240V AC voltage areas (no user adjustment is required).
Use the proper power cord with correct attachment plug type. If the power source is 120 V AC, use a power cord which is a Hospital Grade Power Cord with NEMA 5-15 style plug, labeled for 125 volts AC with UL and C-UL approvals. If the power source is a 240 V AC supply, use the tandem (T blade) type attachment plug with ground conductor power cord that meets the respective European country's safety regulations.
The hospital-grade plug for medical products intended for use in Denmark has DEMKO approval and is rated 13 amps at 250Vac. Plug is recommended for use in medical applications and specifications are being added to the standard SB 107-2-D1. Plug mates with maker's Danish hospital-grade socket. Hospital sockets have slightly different shaped openings allowing only the hospital plug, not the standard Danish plug, to be inserted, to protect the ac circuit in specific medical settings. A ground post, located on the back of the display, may be used for the purpose of grounding the display’s chassis. Any such ground must be installed in accordance with applicable electrical codes. The groundpost is shown on the mechanical drawing found on page 20~22.
English - 10
Servicing
Do not attempt to service the medical LCD monitor yourself, as opening or removing covers may expose you to dangerous voltages or other hazards, and will void the warranty. Refer all servicing to qualified service personnel. Unplug the medical LCD monitor from its power source and refer servicing to qualified personnel under the following conditions:
- If the power cord or plug is damaged or frayed.
- If liquid has been spilled into the medical LCD monitor.
- If objects have fallen into the medical LCD monitor.
- If the medical LCD monitor has been exposed to rain or moisture.
- If the medical LCD monitor has been subjected to excessive shock by being dropped.
- If the cabinet has been damaged.
- If the medical LCD monitor seems to be overheated.
- If the medical LCD monitor emits smoke or abnormal odor.
- If the medical LCD monitor fails to operate in accordance with the operating instructions.
Accessories
Use only accessories specified by the manufacturer, or sold with the medical LCD monitor.
Classification
- Protection against electric shock : Class I including AC/DC Adapter
- Applied Parts : No Applied Parts
- Degree of safety in the presence of flammable anesthetics mixture with air or with oxygen or with nitrous oxide. Not suitable for use in the presence of a flammable anesthetics mixture with oxygen or with nitrous oxide.
- Mode of operation : Continuous.
Cleaning Instructions
Follow your hospital protocol for the handling of blood and body fluids. Clean the display with a diluted mixture of mild detergent and water. Use a soft towel or swab. Use of certain detergents may cause degradation to the labels and plastic components of the product. Consult cleanser manufacturer to see if agent is compatible. Do not allow liquid to enter the display.
11 - English
FCC Information
This medical LCD monitor unit has been tested and found to comply with the limits of a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against interference. This monitor can radiate radio frequency energy and, if not installed and used in accordance with the instructions, it may interfere with other radio communications equipment. There is no guarantee that interference will not occur in a particular installation. If this equipment is found to cause harmful interference to radio or television reception, the user is encouraged to try to correct the interference by carrying out one or more of the following measures:
1. Reorient or relocate the receiving antenna.
2. Increase the distance between the medical LCD monitor and the subject of interference.
3. Plug the monitor into an outlet on a different electrical circuit than that to which the subject of interference is connected.
4. Consult the dealer or an experienced radio/TV technician for help.
NOTICES TO USER
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
FCC WARNING
This medical LCD monitor generates or uses radio frequency energy. Changes or modifications to this medical LCD monitor may cause harmful interference unless the modifications are expressly approved in the instruction manual. The user could lose authority to operate this equipment if an unauthorized change or modification is made.
English - 12
1. Guidance and manufacturer's declaration - electromagnetic emissions
The medical LCD monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the medical LCD monitor should assure that it is used in such an environment.
Electromagnetic environment -guidance
The medical LCD monitor uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment
The medical LCD monitor is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes
2. Guidance and manufacturer's declaration - electromagnetic immunity
This medical LCD monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the medical LCD monitor should assure that it is used in such an environment.
Compliance level
6kV Contact 8kV air
2kV for power supply lines
1kV for input/output lines
Emission test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations IEC 61000-3-3
Compliance
Group 1
Class B
D
Complies
Immunity test
Electrostatic discharge(ESD) IEC61000-4-2
Electrical fast transient/burst IEC 61000-4-4
IEC 60601 Test level
6kV Contact 8kV air
2kV for power supply lines 1kV for input/output
lines
Electromagnetic environment -guidance
Floors should be wood,concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%
Mains power quality should be that of a typical commercial or hospital environment.
13 - English
Surge IEC 61000-4-5
Conducted RF IEC 61000-4-6
1kV differential mode 2kV common mode
3 Vrms
150 kHz to 80MHz
1kV differential mode 2kV common mode
3 Vrms
150 kHz to 80MHz
Mains power quality should be that of a typical commercial or hospital environment.
Portable and mobile RF communications equipment should be used no closer to any part of the medical LCD monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance : d
where P is the maximum output power rating of the transmitter in watts (W)
3. Guidance and manufacturer's declaration - electromagnetic immunity
This medical LCD monitor is intended for use in the electromagnetic environment specified below. The customer or the user of monitor should assure that it is used in such an environment.
Immunity test
Power frequency (50/60Hz) magnetic field IEC 61000-4-8
Voltage dips, short interruptions and voltage variations on power supply
input lines IEC 61000-4-11
IEC 60601 Test level
3.0A/m
<5 % U T (>95 % dip in U T) for 0.5 cycle 40 % U T (60 % dip in U T ) for 5 cycle 70 % U T (30 % dip in U T) for 25 cycle <5 % U T (<95 % dip in U T) for 5 sec.
Compliance level
3.0A/m
<5 % U
T
(>95 % dip in U T) for 0.5 cycle 40 % U T (60 % dip in U T ) for 5 cycle 70 % U T
(30 % dip in U T) for 25 cycle <5 % U T
(<95 % dip in U T) for 5 sec.
Electromagnetic environment -guidance
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Main power quality should be that of a typical commercial or hospital environment. If the user of monitor requires continued operation during power mains interruptions, it is recommended that monitor be powered from an uninterruptible power supply or a battery.
NOTE :
T the A.C. mains voltage prior to
application of the test level.
English - 14
Immunity test
Radiated RF IEC 61000-4-3
IEC 60601 Test level
3 V/m
80.0 MHz to 2.5 GHz
Compliance level
3 V/m
80.0 MHz to 2.5 GHz
Electromagnetic environment -guidance
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, It should be less than the compliance level in each frequency range.
80MHz to 800MHz
80MHz to 2.5GHz
4. Recommended separation distances between portable and mobile RF communications equipment and this medical LCD monitor.
- The medical LCD monitor is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the medical LCD monitor as recommended below, according to the maximum output power of the communications equipment.
150kHz to 80MHz
V1=3Vrms
Rated maximum output power of transmitter [W]
Separation distance according to frequency of transmitter[m]
80MHz to 800MHz
E1=3V/m
800MHz to 2.5GHz
E1=3V/m
15 - English
0.116
0.368
1.166
3.687
11.660
0.116
0.3687
1.1660
3.6872
11.6600
0.2333
0.7378
0.2333
0.7375
23.333
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
English - 16
19" FS-L190*D / FS-L190*DT Monitor
Accessories
Parts
User Manual D-SUB Cable
(6ft/1.8m)
AC-DC Adaptor
(JMW190KA1200F02)
S-Video (Y/C) Cable
(Option)
DVI Cable
(6ft/1.8m)
BNC Cable
(6ft/1.8m)
AC Power cord
(6ft/1.8/m US,UK,EU)
(Hospital Grade)
SCREW FH M3X6
SCREW BH M4X10
(Option)
17 - English
24" FS-L240*D / FS-L240*DT Monitor
Accessories
User Manual D-SUB Cable
(6ft/1.8m)
AC-DC Adaptor
(JMW1150KA2400F04)
S-Video (Y/C) Cable
(Option)
DVI Cable
(6ft/1.8m)
BNC Cable
(6ft/1.8m)
AC Power cord
(6ft/1.8/m US,UK,EU)
(Hospital Grade)
DC Cable Terminal
Male / female (Option)
SCREW FH M3X6
SCREW BH M4X10
(Option)
English - 18
26" FS-L260*D / FS-P260*D(LED) Monitor
Accessories
User Manual D-SUB Cable
(6ft/1.8m)
AC-DC Adaptor
(JMW1150KA2400F04)
(BPM150S24F10)
S-Video (Y/C) Cable
(Option)
DVI Cable
(6ft/1.8m)
BNC Cable
(6ft/1.8m)
AC Power cord
(6ft/1.8/m US,UK,EU)
(Hospital Grade)
DC Cable Terminal
Male / female (Option)
SCREW FH M3X6
SCREW BH M4X10 SCREW BHM4X6
(Option)
19 - English
32" FS-L320*D Monitor
Accessories
User Manual D-SUB Cable
(6ft/1.8m)
AC-DC Adaptor
(JMW1180KA2400F01)
S-Video (Y/C) Cable
(Option)
DVI Cable
(6ft/1.8m)
BNC Cable
(6ft/1.8m)
AC Power cord
(6ft/1.8/m US,UK,EU)
(Hospital Grade)
DC Cable Terminal
Male / female (Option)
SCREW FH M3X6
Remote controller
Battery(AAA)
SCREW BH M4X10
English - 20
42" FS-L420*D / 55” FS-L550*D Monitor
User Manual
Accessories
D-SUB Cable
(6ft/1.8m)
S-Video (Y/C) Cable
(Option)
DVI Cable(6ft)
BNC Cable
(6ft/1.8m)
AC Power cord
(6ft/1.8/m US,UK,EU)
(Hospital Grade)
Remote controller
Battery(AAA)
21 - English
DC INLET DC ON/OFF SWITCH OPTICAL IN DVI IN DVI OUT VGA IN GPIO RS232C SDI IN SDI OUT
S-VIDEO /Y S-VIDEO /C C-VIDEO RED /Pr GREEN /Y BLUE /Pb C-SYNC/H-SYNC V-SYNC EARTH GROUND
FS-L190*D / FS-L240*D / FS-L260*D / FS-P260*D / FS-L320*D Input connector
Connector
DC INLET DC ON/OFF SWITCH DVI 2 IN DVI 1 IN DVI 1 OUT VGA IN GPIO RS232C SDI IN SDI OUT
S-VIDEO /Y S-VIDEO /C C-VIDEO RED /Pr GREEN /Y BLUE /Pb C-SYNC/H-SYNC V-SYNC EARTH GROUND
DVI fiber optic(4port LC / 1port SC) input
Dual DVI input
NO. 3 Optical IN do not support on FS-L190*D
NO. 3 DVI 2 IN do not support on FS-L190*D
English - 22
DC INLET DC ON/OFF SWITCH OPTICAL IN DVI IN DVI OUT VGA IN GPIO USB TOUCH SERIAL TOUCH RS232C SDI IN
SDI OUT S-VIDEO /Y S-VIDEO /C C-VIDEO RED /Pr GREEN /Y BLUE /Pb C-SYNC/H-SYNC V-SYNC EARTH GROUND
FS-L190*DT / FS-L240*DT / FS-P260*DT input connector
DC INLET DC ON/OFF SWITCH DVI 2 IN DVI 1 IN DVI 1 OUT VGA IN GPIO USB TOUCH SERIAL TOUCH RS232C SDI IN
SDI OUT S-VIDEO /Y S-VIDEO /C C-VIDEO RED /Pr GREEN /Y BLUE /Pb C-SYNC/H-SYNC V-SYNC EARTH GROUND
DVI fiber optic(4port LC / 1port SC) input
Dual DVI input
NO. 3 Optical IN do not support on FS-L190*D
NO. 3 DVI 2 IN do not support on FS-L190*D
23 - English
AC ON/OFF SWITCH AC INLET OPTICAL IN DVI IN DVI OUT VGA IN GPIO RS232C SDI IN SDI OUT
S-VIDEO /Y S-VIDEO /C C-VIDEO RED /Pr GREEN /Y BLUE /Pb C-SYNC/H-SYNC V-SYNC EARTH GROUND
FS-L420*D / FS-L550*D Input connector
DVI fiber optic input
AC ON/OFF SWITCH AC INLET DVI 2 IN DVI 1 IN DVI 1 OUT VGA IN GPIO RS232C SDI IN SDI OUT
S-VIDEO /Y S-VIDEO /C C-VIDEO RED /Pr GREEN /Y BLUE /Pb C-SYNC/H-SYNC V-SYNC EARTH GROUND
Dual DVI input
English - 24
395
76.5
31
423
351.5
88.5 89
100100
100
100
4-M4X12(VESA MOUNT HOLE)4-M4X12(VESA MOUNT HOLE)
114 60
OSD MEMBRANE SWITCH
Unit : mm
Mechanical Product Drawing
19”
FS-L190*D / FS-L190*DT Dimension
Front view
Rear view (Without I/O Cover)
Rear view (Installed I/O Cover)
25 - English
580
386
124.3 68.7
95
100
100
4451
OSD MEMBRANE SWITCH
4-M4X12(VESA MOUNT HOLE)
100
100
4-M4X12(VESA MOUNT HOLE)
Unit : mm
24” FS-L240*D / FS-L240*DT Dimension
Front view
Rear view (Without I/O Cover)
Rear view (Installed I/O Cover)
English - 26
618
99.5
55.5 44
412
107 99
200
100
OSD MEMBRANE SWITCH
4-M4X12(VESA MOUNT HOLE)
200
100
4-M4X12(VESA MOUNT HOLE)
100
100
Unit : mm
26” FS-L260*D Dimension
Front view
Rear view (Without I/O Cover)
Rear view (Installed I/O Cover)
27 - English
26” FS-L260*D(LED) / FS-P260*D(LED) Dimension
Unit : mm
Front view
Rear view (Without I/O Cover)
Rear view (Installed I/O Cover)
English - 28
100
795
100
100
200
100
200
300
400
56 44
485
242.5
4-M4X12(VESA MOUNT HOLE)
4-M6(VESA MOUNT HOLE)
100
300
400
4-M4X12(VESA MOUNT HOLE)
4-M6(VESA MOUNT HOLE)
OSD MEMBRANE SWITCH
Unit : mm
32” FS-L320*D Dimension
Front view
Rear view (Without I/O Cover)
Rear view (Installed I/O Cover)
29 - English
1024.6
600
200
300
617.4
1221.1 495.4 70.3
111.1 55 50
OSD MEMBRANE SWITCH
10-M8(USER MOUNT HOLE)
Unit : mm
42” FS-L420*D Dimension
Front view
Rear view
English - 30
IR
6-M8(VESA MOUNT HOLE)
OSD MEMBRANE SWITCH
684.4
1213.6
1293
600
400
86.3
36.3 50
777
Unit : mm
55” FS-L550*D Dimension
Front view
Rear view
31 - English
Control
An 8 button keypad, located in bottom right corner on the front of the display, allows the user to make adjustments to various display parameters using the On Screen Display (OSD) system.
Power Indicating LED Normal mode (ON): Green Standby mode : Blinking Green
Off mode : Monitor Off
Note 1 : LED normal and off mode sign can be changed according to the
requirement of the customer whether normal mode green is on or off
Note 2 : The main AC power switch, on the back panel, should be in the ON position.
The DC power button is used to turn on the monitor
On-Screen Display (OSD) Function Button
1. POWER : Soft power turns the monitor ON or OFF
2. MENU : Used to activate to the OSD menu and exit from main menu or sub menu.
3. PIP : Enables PIP(picture in Picture) function. Selects PIP,PBP1,PBP2.
4. UP () : With the OSD deactivated, it is a Hot Key for increasing brightness.
With the OSD activated, moves the cursor upward.
5. DOWN (): With theOSD deactivated, it is a Hot Key for decreasing brightness.
With the OSD activated, moves the cursor downward.
OSD Button
English - 32
6. PLUS (+) : With the OSD deactivated, it is a Hot key for increasing contrast. With the OSD activated, enter sub menu and increases the adjustment of the selected function.
7. MINUS (-) : With the OSD deactivated, it is a Hot key for increasing contrast. With theOSD activated, it decreases the adjustment of the selected function.
8. INPUT : With the OSD deactivated, if pressed down for over 1 sec., it is a Hot Key for auto-adjustment control under DSUB ANALOG / RGBs signals. With theOSD activated, it changes the displayed signal source.
33 - English
1. SOURCE : Changes the display signal source.
2. POWER : Soft power turns the monitor ON or OFF.
3. UP () : With the OSD deactivated,it is a Hot key for increasing brightness. With the OSD activated, moves the cursor upward.
4. DOWN () : With the OSD deactivated, it is a Hot key for decreasing
brightness. With the OSD activated, moves the cursor downward.
5. MINUS (-) : With the OSD deactivated, it is a Hot key for decreasing contrast.
With the OSD activated, it decreases the adjustment of the selected function.
6. PLUS (+) : With the OSD deactivated, it is a Hot key for increasing contrast.
With the OSD activated, it increases the adjustment of the Selected function.
7. MENU : With the OSD deactivated, activates the OSD menu.
With the OSD activated, exits from main menu or sub menu.
8. AUTO : Fits to the most appropriate screen on the D-SUB Analog signal.
9. PIP : Enables PIP(picture in Picture) function.
Selects PIP,PBP1,PBP2
10. MUTE : Sound muted.(Not used)
11. SWAP : Swaps the position of the Primary and Secondary images.
Remote button function
<N
otice
> Remote control is available in FS-L320*D / FS-L420*D / FS-L55*D models only.
English - 34
Pin 1. Primary and Secondary Swap.
Grounding this pin will swap the primary and secondary image.
Pin 2. PIP,PBP1,PBP2 Single
Continuously grounding this pin causes the position and size choices to cycle.
Pin 3. Record Indicator
The record indicator is displayed in the
top left corner when the pin is ground to pin 4. The indicator will vanish when the contact is opened.
Pin 4. Connector Ground
This is the common ground location.
Monitor
RJ9
GPIO
There are four pins on the RJ9 GPIO connector. Each pin has a preprogrammed function assigned to it. The function is initiated when the pin is grounded.
35 - English
This monitor does not adhere to the VESA DPMS standard when no signal is present on the video input.
Model
Status LED sign Power Consumption
FS-L190*D Normal mode Green on <60W
Standby mode Green blinking <20W
FS-L190*DT Normal mode Green on <60W
Standby mode Green blinking <20W
FS-L240*D Normal mode Green on <100W
Standby mode Green blinking <20W
FS-L240*DT Normal mode Green on <100W
Standby mode Green blinking <20W
FS-L260*D Normal mode Green on <130W
Standby mode Green blinking <20W
FS-L260*D(LED) Normal mode Green on <70W
Standby mode Green blinking <20W
FS-P260*D(LED) Normal mode Green on <70W
Standby mode Green blinking <20W
FS-L320*D Normal mode Green on <150W
Standby mode Green blinking <20W
FS-L420*D Normal mode Green on <260W
Standby mode Green blinking <20W
FS-L550*D Normal mode Green on <200W
Standby mode Green blinking <20W
Power management
English - 36
ADJUST
1. BRIGHTNESS I
ncreases or decreases the brightness.(Range
: 0~100)
2. CONTRAST I
ncreases or decreases the contrast.(Range
: 0~100)
3. CLOCK I
ncreases or decreases the sampling frequency. (Range
: 0~100)
4.
PHASE I
ncreases or decreases the Phase level. (Range
: 0~100)
5. BACKLIGHT Adjusts backlight dimming level
. (Range
: 0~100)
6.
AUTO ADJUST F
its to the most appropriate screen on the D-SUB Analog / RGBs signal
.
OSD
50 50
50 50
ADJUST BRIGHTNESS
CONTRAST CLOCK PHASE AUTO ADJUST
EXIT DSUB ANALOG
SELECT
1280X1024/80K x75Hz
MOVE
COLOR TEMP
COLOR TEMP MODE C1
EXIT DSUB ANALOG 1280X1024/80K x75Hz
MOVE
SELECT
1. MODE C
hanges the color mode (C1(Reddish,6500K), C2(Bluish,9300K),
USER(7200K)
2. RED R
ed balance
.(O
nly works with
USER Mode) (R
ange
: 0~100)
3. GREEN G
reen balance
.(O
nly works with
USER Mode) (R
ange
: 0~100)
4. BLUE B
lue balance
.(O
nly works with
USER Mode) (R
ange
: 0~100)
50 50
IMAGE IMAGE SIZE FILL
NORMAL
MODE 1
H POSITION V POSITION
FILTER
2.0
GRMMA
0 1 2 3 4 5 6 7 8
OVER SCAN IMAGE SETTING
DSUB ANALOG
1280X1024/80K x75Hz
EXIT
MOVE
SELECT
IMAGE
1. IMAGE SIZE C
hanges the image size. (Full,Fill aspect,1:1,Normal)
2. H POSITION A
djusts the horizontal position of the displayed source image.(Range
: 0~100)
3. V POSITION A
djusts the vertical position of the displayed source image. (Range
: 0~100)
4. GAMMA A
djusts GAMMA value (VIDEO,BYPASS,1.8,2.0,2.2,2.4,2.6,PACS)
5. FILTER S
ets the sharpness of image ( Softest, Soft, Normal, Sharp, Sharpest)
6. OVER SCAN A
djusts the displayed size. (0~8)
7. IMAGE SETTING Changes the image setting.(Preset 1,2 / User 1,2,3)
8.
ZOOM / PAN
Enlarges the image, moves images left and right.
9. FREEZE FRAME K
eeps the image still.
DSUB ANALOG / RGBS
input source
37 - English
SETUP
1. LANGUAGE C
hanges the OSD language (8 language)
2. OSD COLOR A
djusts the OSD background from white opaque to translucent.
3 .OSD POSITION
C
hanges the osd position. (9 Positions)
4. DURATION
Adjusts the length of time the OSD menu is present on the screen
.
(5, 10, 20, 30, 60, 90, 120, 180, 240 seconds)
5. RESET SETTING Changes all the OSD values make to factory default.
6. AUTO SOURCE SELECT D
isables or enable auto source select.
ON: Searches through all possible input source untill an active video
source is found.
OFF: Video input is manually selected.
7. INACTIVE INPUT Changes the input source between RGBs and YPbPr
.
8. PICTURE DELAY
Changes the interace mode setting. (0 : OFF, 1 : ON)
PIP
50
SETUP LANGUAGE ENGLISH
120 SEC
OSD COLOR OSD POSITION DURATION RESET SETTINGS AUTO SOURCE SELECT ON COLOR SPACE RGB
DSUB ANALOG
1280X1024/80K x75Hz
EXIT
MOVE
SELECT
PIP LAYOUT SINGLE
DSUB ANALOG
1280X1024/80K x75Hz
EXIT
MOVE
SELECT
1. LAYOUT C
hanges the OSD layout. (Single, PIP, PBP1, PBP2)
2. SOURCE C
hanges the secondary source.
3. SIZE Changes the PIP size(Small, Large)
.
4. POSITION
Changes the PIP Position
.
5. SWAP S
waps the position and size of the Primary and Secondary image.
English - 38
ADJUST
1. BRIGHTNESS I
ncreases or decrease the brightness.
(R
ange
: 0~100)
2. CONTRAST I
ncreases or decreases the Contrast.
(R
ange
: 0~100)
3. BACKLIGHT A
djusts backlight dimming level. (Range : 0~100)
50 50
100
ADJUST BRIGHTNESS
CONTRAST BACK LIGHT
EXIT DVI DIGITAL
MOVE
1280X1024/80K x75Hz
SELECT
COLOR TEMP
COLOR TEMP MODE C1
EXIT DVI DIGITAL 1280X1024/80K x75Hz
MOVE
SELECT
1. MODE C
hanges the color temperature mode. (C1(Reddish,6500K), C2(Bluish,9300K),
USER(7200K)
2. RED R
ed balance.(Only works with
USER Mode) (R
ange
: 0~100)
3. GREEN G
reen balance.
(O
nly works with
USER Mode) (R
ange
: 0~100)
4. BLUE B
lue balance.(Only works with
USER Mode) (R
ange
: 0~100)
IMAGE IMAGE SIZE FILL
NORMAL
PRESET1
FILTER
2.0
GRMMA
0 1 2 3 4 5 6 7 8
OVER SCAN IMAGE SETTING ZOOM/PAN FREEZE FRAME
DVI DIGITAL
1280X1024/80K x75Hz
EXIT
MOVE
SELECT
ON
IMAGE
1. IMAGE SIZE C
hanges the image size
. (F
ull, Fill aspect**, 1:1**, Norma)
**Only in DVI Optical
2. GAMMA A
djusts GAMMA value. (VIDEO,BYPASS,1.8,2.0,2.2,2.4,2.6,PACS)
3. FILTER S
ets the sharpness of image. ( Softest, Soft, Normal, Sharp, Sharpest)
4.
OVER SCAN A
djusts the displayed size. (0~8)
5.
IMAGE SETTING Changes the image setting. (Preset 1,2 / User 1,2,3)
6.
ZOOM/PAN Enlarges the image, moves images left and right.
7.
FREEZE FRAME Keeps the image still.
DVI OPTICAL / DVI DIGITAL
input source
39 - English
SETUP
1. LANGUAGE C
hanges the OSD language (8 language)
2. OSD COLOR A
djusts the OSD background from white opaque to translucent.
3 .OSD POSITION
C
hanges the OSD position. (9 Positions)
4. DURATION Adjusts the length of time the OSD menu is present on the screen
.
(5, 10, 20, 30, 60, 90, 120, 180, 240 seconds)
5. RESET SETTING Changes all the OSD values to factory outgoing state
.
6. AUTO SOURCE SELECT D
isables or enables auto source select.
ON: Searches through all possible input source until an active video
source is found.
OFF: Video input is manually selected.
7. INACTIVE INPUT Changes the input source between RGBs and YPbPr
.
8. PICTURE DELAY
Adjust the picture delay.
PIP
50
SETUP LANGUAGE ENGLISH
120 SEC
OSD COLOR OSD POSITION DURATION RESET SETTINGS AUTO SOURCE SELECT ON
DVI DIGITAL
1280X1024/80K x75Hz
EXIT
MOVE
SELECT
PIP LAYOUT SINGLE
DSUB ANALOG
1280X1024/80K x75Hz
EXIT
MOVE
SELECT
1. LAYOUT C
hanges the OSD layout. (Single, PIP, PBP1, PBP2)
2. SOURCE C
hanges the secondary source.
3. SIZE Changes the PIP size(Small, Large)
.
4. POSITION Changes the PIP position
.
5. SWAP S
waps the position and size of the Primary and Secondary image.
English - 40
ADJUST
1. BRIGHTNESS I
ncreases or decreases the brightness.(Range
: 0~100)
2. CONTRAST I
ncreases or decreases the Contrast.(Range
: 0~100)
3. SHARPNESS Adjusts the sharpness of video image
.(Range
: 0~100)
4. SATURATION Changes the tone of color
. (Range : 0~100)
5. COLOR Changes the richness of color. (Range : Greenish 0~50, Redish 0~50)
6. BACKLIGHT A
djusts backlight dimming level. (Range : 0~100)
7. CLOCK Increases or decreases the sampling
. (Range : 0~100)
8. PHASE Increases or decreases the Phase level
. (Range : 0~100)
50 50
ADJUST BRIGHTNESS
CONTRAST
EXIT YPbPr
1920X1080/80K x60Hz
50
SHARPNESS
65
SATURATION
0
COLOR
100
BACKLIGHT
65
PHASE
SELECT
MOVE
COLOR TEMP
COLOR TEMP MODE C1
YPbPr 1920X1080/80K x60Hz
SELECT
EXIT MOVE
1. MODE C
hanges the color temperature mode. (C1(Reddish,6500K), C2(Bluish,9300K),
USER(7200K)
2. RED R
ed balance.(Only works with
USER Mode) (R
ange
: 0~100)
3. GREEN G
reen balance.
(O
nly works with
USER Mode) (R
ange
: 0~100)
4. BLUE B
lue balance.(Only works with
USER Mode) (R
ange
: 0~100)
IMAGE IMAGE SIZE FILL
NORMAL
PRESET1
FILTER
2.0
GRMMA
0 1 2 3 4 5 6 7 8
OVER SCAN IMAGE SETTING
YPbPr
1920X1080/80K x60Hz
50
V POSITON
50
H POSITON
EXIT
MOVE
SELECT
IMAGE
1. IMAGE SIZE C
hanges the image size.(Full, Fill aspect, 1:1, Normal, Anamorphic)
2. H POSITION A
djusts the horizontal position of the displayed source image.(Range
: 0~100)
3. V POSITION A
djusts the vertical position of the displayed source image. (Range
: 0~100)
4. GAMMA A
djusts GAMMA value.(VIDEO,BYPASS,1.8,2.0,2.2,2.4,2.6,PACS))
5. FILTER S
ets the sharpness of image.( Softest, Soft, Normal, Sharp, Sharpest)
6.
OVER SCAN A
djusts the displayed size. (0~8)
7. IMAGE SETTING Changes the image setting.(Preset 1,2 / User 1,2,3)
8.
ZOOM / PAN Enlarges the image, moves images left and right.
9.
FREEZE FRAME
K
eeps the image still.
YP
bPr input source
41 - English
SETUP
1. LANGUAGE C
hanges the OSD language (8 language)
2. OSD COLOR A
djusts the OSD background from white opaque to translucent.
3 .OSD POSITION
C
hanges the osd position. (9 Positions)
4. DURATION
Adjusts the length of time the OSD menu is present on the screen
.
(5, 10, 20, 30, 60, 90, 120, 180, 240 seconds)
5. RESET SETTING Changes all the OSD values to factory outgoing state
.
6. AUTO SOURCE SELECT D
isables or enables auto source select.
ON: Searches through all possible input source until an active video
source is found.
OFF: Video input is manually selected.
7. INACTIVE INPUT Changes the input source between RGBs and YPbPr
.
8. PICTURE DELAY
Changes the interace mode setting. (0 : OFF, 1 : ON)
PIP
50
SETUP LANGUAGE ENGLISH
120 SEC
OSD COLOR OSD POSITION DURATION RESET SETTINGS AUTO SOURCE SELECT ON COLOR SPACE YPbPr
YPbPr
1920X1080/80K x60Hz
EXIT
MOVE
SELECT
PIP LAYOUT SINGLE
YPbPr
1920X1080/80K x60Hz
EXIT
MOVE
SELECT
1. LAYOUT C
hanges the OSD layout. (Single, PIP, PBP1, PBP2)
2. SOURCE C
hanges the secondary source.
3. SIZE Changes the PIP size(Small, Large)
.
4. POSITION Changes the PIP Position
.
5. SWAP S
waps the position and size of the Primary and Secondary image.
English - 42
ADJUST
1. BRIGHTNESS
I
ncreases or decreases the brightness.(Range
: 0~100)
2. CONTRAST
I
ncreases or decreases the Contrast.(Range
: 0~100)
3. SHARPNESS
A
djusts the sharpness of video image.(Range
: 0~100)
4. SATURATION
C
hanges the tone of color. (Range : 0~100)
5. COLOR
C
hanges the richness of color. (Range : Greenish 0~50, Redish 0~50)
6. BACKLIGHT
A
djusts backlight dimming level. (Range : 0~100)
50 50
ADJUST BRIGHTNESS
CONTRAST
EXIT CVIDEO
MOVE
NTSC 60Hz / 4.43MHz
50
SHARPNESS
65
SATURATION
0
COLOR
100
BACKLIGHT
SELECT
COLOR TEMP
COLOR TEMP MODE C1
EXIT CVIDEO NTSC 60Hz / 4.43MHz
MOVE
SELECT
1. MODE C
hanges the color temperature mode. (C1(Reddish,6500K), C2(Bluish,9300K),
USER(7200K)
2. RED R
ed balance.(Only works with
USER Mode) (R
ange
: 0~100)
3. GREEN G
reen balance
.(O
nly works with
USER Mode) (R
ange
: 0~100)
4. BLUE B
lue balance.(Only works with
USER Mode) (R
ange
: 0~100)
IMAGE IMAGE SIZE FILL
NORMAL
PRESET1
FILTER
2.0
GRMMA
0 1 2 3 4 5 6 7 8
OVER SCAN IMAGE SETTING
CVIDEO
NTSC 60Hz / 4.43MHz
50
V POSITON
50
H POSITON
EXIT
MOVE
SELECT
IMAGE
1.IMAGE SIZE C
hanges the image size.
(F
ull, Fill aspect, 1:1, Normal, Anamorphic)
2. H POSITION A
djusts the horizontal position of the displayed source image.(Range
: 0~100)
3. V POSITION A
djusts the vertical position of the displayed source image. (Range
: 0~100)
4. GAMMA A
djusts GAMMA value (VIDEO,BYPASS,1.8,2.0,2.2,2.4,2.6,PACS)
5. FILTER S
ets the sharpness of image ( Softest, Soft, Normal, Sharp, Sharpest)
6.
OVER SCAN A
djusts the displayed size. (0~8)
7. IMAGE SETTING Changes the image setting.(Preset 1,2 / User 1,2,3)
8.
ZOOM / PAN Enlarges the image, moves images left and right.
9.
FREEZE FRAME
K
eeps the image still.
SVIDEO / CVIDEO
input source
43 - English
SETUP
1. LANGUAGE C
hanges the OSD language (8 language)
2. OSD COLOR A
djusts the OSD background from white opaque to translucent.
3 .OSD POSITION
C
hanges the osd position. (9 Positions)
4. DURATION
Adjusts the length of time the OSD menu is present on the screen
.
(5, 10, 20, 30, 60, 90, 120, 180, 240 seconds)
5. RESET SETTING Changes all the OSD values to factory outgoing state
.
6. AUTO SOURCE SELECT D
isables or enables auto source select.
ON: Searches through all possible input source until an active video
source is found.
OFF: Video input is manually selected.
7. INACTIVE INPUT Changes the input source between RGBs and YPbPr
.
8. PICTURE DELAY
Adjust the picture delay.
PIP
50
SETUP LANGUAGE ENGLISH
120 SEC
OSD COLOR OSD POSITION DURATION RESET SETTINGS AUTO SOURCE SELECT ON
CVIDEO
NTSC 60Hz / 4.43MHz
EXIT
MOVE
SELECT
PIP LAYOUT SINGLE
CVIDEO
NTSC 60Hz / 4.43MHz
EXIT
MOVE
SELECT
1. LAYOUT C
hanges the OSD layout. (Single, PIP, PBP1, PBP2)
2. SOURCE C
hanges the secondary source.
3. SIZE Changes the PIP size(Small, Large)
.
4. POSITION Changes the PIP Position
.
5. SWAP S
waps the position and size of the Primary and Secondary image.
English - 44
ADJUST
1. BRIGHTNESS I
ncreases or decreases the brightness.(Range
: 0~100)
2. CONTRAST I
ncreases or decreases the Contrast.(Range
: 0~100)
3. SHARPNESS A
djusts the sharpness of video image.(Range
: 0~100)
4. SATURATION C
hanges the tone of color. (Range : 0~100)
5. COLOR C
hanges the richness of color. (Range : Greenish 0~50,Redish 0~50)
6. BACKLIGHT A
djusts backlight dimming level. (Range : 0~100)
50 50
ADJUST BRIGHTNESS
CONTRAST
EXIT SDI
MOVE
1920X1080i / 30Hz
50
SHARPNESS
65
SATURATION
0
COLOR
100
BACKLIGHT
SELECT
COLOR TEMP
COLOR TEMP MODE C1
EXIT SDI 1920X1080i / 30Hz
MOVE
SELECT
1. MODE C
hanges the color temperature mode. (C1(Reddish,6500K), C2(Bluish,9300K),
USER(7200K)
2. RED R
ed balance.(Only works with
USER Mode) (R
ange
: 0~100)
3. GREEN G
reen balance
.(O
nly works with
USER Mode) (R
ange
: 0~100)
4. BLUE B
lue balance
.(O
nly works with
USER Mode) (R
ange
: 0~100)
IMAGE IMAGE SIZE FILL
IMAGE SETTING PRESET1 ZOOM/PAN
ON
FREEZE FRAME
SDI
1920X1080i / 30Hz
EXIT
MOVE
SELECT
0 1 2 3 4 5 6 7 8
OVER SCREN
NORMAL
FILTER
2.0
GRMMA
IMAGE
1. IMAGE SIZE C
hanges the image size.(Full, Fill aspect, 1:1, Normal, Anamorphic)
2. GAMMA A
djusts GAMMA value (VIDEO,BYPASS,1.8,2.0,2.2,2.4,2.6,PACS)
3. FILTER S
ets the sharpness of image ( Softest, Soft, Normal, Sharp, Sharpest)
4.
OVER SCAN A
djusts the displayed size. (0~8)
5. IMAGE SETTING Changes the image setting.(Preset 1,2 / User 1,2,3)
6.
ZOOM / PAN Enlarges the image, moves images left and right.
7.
FREEZE FRAME
K
eeps the image still.
SDI
input source
45 - English
SETUP
1. LANGUAGE C
hanges the OSD language (8 language)
2. OSD COLOR A
djusts the OSD background from white opaque to translucent.
3 .OSD POSITION
C
hanges the osd position. (9 Positions)
4. DURATION Adjusts the length of time the OSD menu is present on the screen
.
(5, 10, 20, 30, 60, 90, 120, 180, 240 seconds)
5. RESET SETTING Changes all the OSD values to factory outgoing state
.
6. AUTO SOURCE SELECT D
isables or enables auto source select.
ON: Searches through all possible input source until an active video
source is found.
OFF: Video input is manually selected.
7. PICTURE DELAY
Changes the interace mode setting. (0 : OFF, 1 : ON)
PIP
50
SETUP LANGUAGE ENGLISH
120 SEC
OSD COLOR OSD POSITION DURATION RESET SETTINGS AUTO SOURCE SELECT ON
SDI
1920X1080i / 30Hz
EXIT
MOVE
SELECT
PIP LAYOUT SINGLE
SDI
1920X1080i / 30Hz
EXIT
MOVE
SELECT
1. LAYOUT C
hanges the OSD layout. (Single, PIP, PBP1, PBP2)
2. SOURCE C
hanges the secondary source.
3. SIZE Changes the PIP size(Small, Large)
.
4. POSITION
Changes the PIP Position
..
5. SWAP S
waps the position and size of the Primary and Secondary image.
English - 46
OSD System overview
BRIGHTNESS Press the BRIGHTNESS botton to display the ADJUST menu or UP,DOWN hot key.
Setting the brightness too high or too low will decrease the amount of visible grayscales.
CONTRAST Press the CONTRAST button to display the ADJUST menu or +,- hot key.
Setting the Contrast too high or too low will cause loss of some grascales.
CLOCK Do not adjust. It will adjust automatically after auto adjustment.
When frequency value is wrong, the horizontal image will have a wrong size or noise.
PHASE Do not adjust . It will adjust automatically after auto adjustment.
When frequency value is wrong, the image will have a noise.
BACKLIGHT Adjusts backlight dimming level. Setting the backlight too low will cause dark image and too high
will decrease the backlight life time.
AUTO ADJUST Fits to the most appropriate screen on the D-SUB Analog/RGBS signal SHARPNESS Adjusts the sharpness of video image
SATURATION Changes the tone of color COLOR Changes the richness of color(Range Greenish 0~50,Redish 0~50) COLOR TEMP C1 Default 6500K color setting COLOR TEMP C2 Default 9300K color setting
COLOR TEMP Default 7200K color setting,but it's changeable value by user USER
IMAGE SIZE
DSUB/ DVI OPTICAL / DVI DIGITAL input source
FULL FILL ASPECT
1:1 NORMAL
YPbPr / RGBS / SDI / CVIDEO / SVIDEO input source
FULL FILL ASPECT 1:1
NORMAL ANAMORPHIC
CAUTION : FILL ASPECT,NORMAL Size depends on input size ratio
Menus Function Descriptions
47 - English
1.8 2.0 2.2 2.4 2.6 PACS VIDEO
Note: BYPASS depends on panel gamma value, please refer to the panel specification.
A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W
A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W A B C D E F G H I J K L M N O P Q R S T U V W
SOFTEST
NORMAL
SHARP
SHARPEST
SOFT
0: 1: 2:
A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U
A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U A B C D E F G H I J K L M N O P Q R S T U
B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T
B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T B C D E F G H I J K L M N O P Q R S T
C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R
C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R
C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R
C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R C D E F G H I J K L M N O P Q R
B C D E F G H I J K L M N O P Q R S T
D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q
D E F G H I J K L M N O P Q
D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q
D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q D E F G H I J K L M N O P Q
3: 4: 5:
6: 7: 8:
Menus Function Descriptions
H POSITION
Adjusts the Horizontal position of the image. It will return to the default state when executing
AUTO ADJUST or RESET SETTINGS.
V POSITION Adjusts the Verticall position of the image. It will return to the default state when executing
AUTO ADJUST or RESET SETTINGS.
GAMMA Adjusts the gamma curve of video image.
FILTER Adjusts the gamma curve of video image.
OVER SCAN Enables a 10% over scan of orignal input image.
English - 48
IMAGE SETTING Saves 5 user’s settings : BRIGHTNESS,CONTRAST,BACKLIGHT,COLOR,TEMP and FILTER
separately.(PRESET1,2 / USER 1,2,3)
ZOOM / PAN Controls the zoom in/out of the image.
FREEZE FRAME Freezes the main image. Does not freeze secondary image in PIP mode. LANGUAGE Changes the OSD to one of 8 languages.
ENGLISH / GERMAN / FRENCH / SPANISH / ITALIAN / JAPANESE / CHINESE / KOREAN
OSD COLOR Adjusts the color of OSD.
OSD POSITION Adjusts the position of the OSD.
0 Step : 4 Step : 9 Step :
13 Step : 18 Step : 22 Step :
26 Step : 30 Step :
Controls the PAN in/out of the image.
50 50 50
100
65
ADJUST BRIGHTNESS
CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS
CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
TRANSPARENT
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
50 50 50
100
65
ADJUST BRIGHTNESS CONTRAST CLOCK PHASE BACK LIGHT AUTO ADJUST
EXIT DSUB ANALOG
MOVE ADJUST
1280X1024/80K x75Hz
Menus Function Descriptions
49 - English
OSD DURATION Adjusts the time the OSD menu remains on the screen. RESET SETTING Resets the monitor back to factory settings.. AUTO SOURCE When AUTO SOURCE SELECT is on, the monitor will automatically search for an active MAIN
SELECT input Source. COLOR SPACE
Changes the input source between RGBs and YPbPr
.
PIP LAY OUT Selets on of four possible layouts(SINGLE,PIP,PBP1,PBP2)
Note : The aspect ratio does not change in PIP or PBP1.
PIP SOURCE Selects a PIP source input. You can change other sub windows through OSD pip menu.
Below chart is PIP
O : Support up to UXGA,60Hz (162MHz) : Option to have DVI 2
PIP SIZE
Changes PIP mode sub window size
PIP POSITION Changes PIP mode sub window Position
Sub window
DVI OPTICAL DVI DIGITAL1 DSUB SDI YPbPr RGBS CVIDEO SVIDEO
Input source DVI DIGITAL 2 ANALOG
DVI OPTICAL DVI DIGITAL 2
DVI DIGITAL 1
DSUB ANALOG
SDI
YPbPr
RGBS
CVIDEO
SVIDEO
XO
OOOO
OO
OX
XOXX
OO
OX
XOXX
O
O
OO
OX OO
XX
OX
XOXX
OO
O
X
X
O
X
X
O
O
OO
OX OO
XX
OO
OX OO
XX
SMALL LARGE
25% / Panel size 33% / Panel size
Menus Function Descriptions
1
1
1
1
1
1
1
English - 50
SWAP Changes main window and sub window position in PIP,PBP1,PBP2.
Menus Function Descriptions
PIP
PBP1
PBP2
51 - English
640 X 350 @70Hz 720 X 400 @70Hz 640 X 480 @60Hz 640 X 480 @75Hz 800 X 600 @60Hz
800 X 600 @75Hz 1024 X 768 @60Hz 1024 X 768 @75Hz 1152 X 864 @60Hz 1152 X 864 @75Hz 1280 X 720 @60Hz 1280 X 1024@60Hz 1280 X 1024@75Hz 1360 X 768@75Hz 1600 X 1200@60Hz 1920 X 1080@60Hz 1920 X 1200@60Hz
31.50
31.50
31.50
37.50
37.90
46.90
48.40
60.00
54.35
67.50
44.70
64.00
80.00
47.70
75.00
67.50
74.00
70.00
70.00
60.00
75.00
60.00
75.00
60.00
75.00
60.00
75.00
60.00
60.00
75.00
60.00
60.00
60.00
60.00
25.17
28.32
25.18
31.50
40.00
49.50
65.00
78.75
80.000
108.00
74.40
108.00
135.00
108.75
162.00
148.50
154.12
Resolution
Horizontal
Frequency
(KHz)
Vertical
Frequency
(Hz)
Clock
Frequency
(MHz)
PC Supported Mode
Standard Signal table
FS-L190*D(T)
FS-L260*D / FS-P260*D / FS-L320*D / FS-L420*D / FS-L550*D
FS-L240*D(T) / FS-L260*D
SMPTE-424M 1080p (3G) SMPTE-292M 720p , 1080i (HD) SMPTE-259M 480i , 576i (SD)
SDI Video format
Output Signal
Description
English - 52
15
11 15
6
10
Pin No. Assignment Pin No. Assignment
1 Red 9 No Connection 2 Green 10 Ground-Sync 3 Blue 11 Ground 4 Ground 12 DDC Data 5 DDC 5V Standby 13 H.Sync
Cable Connection check 14 V.Sync
6 Ground-Red 15 DDC Clock 7 Ground-Green 8 Ground-Blue
18
17 24
9
16
2
1
Pin No. Assignment Pin No. Assignment
1 T.M.D.S. Data2- 13 No Connection 2 T.M.D.S. Data2+ 14 +5V Power 3 T.M.D.S. Data2 Shield 15 Ground
4 No Connection 16 Hot Plug Detect
5 No Connection 17 T.M.D.S. Data0­6 DDC Clock 18 T.M.D.S. Data0+ 7 DDC Data 19 T.M.D.S. Data0 Shield 8 No Connection 20 No Connection
9 T.M.D.S. Data1- 21 No Connection 10 T.M.D.S. Data1+ 22 T.M.D.S. Clock Shield 11 T.M.D.S. Data1 Shield 23 T.M.D.S. Clock+ 12 No Connection 24 T.M.D.S. Clock-
Pin No. Assignment
1 Composite 2 Ground
VGA (
15Pin D-SUB)
DVI In,Out (
24Pin DVI-D)
C-Video (BNC)
Signal connector Pin Assignments
53 - English
1 2 3 4
Pin No. Assignment
1 P,S Swap 2 PIP,PBP1,PBP2 Select 3 Record lndicator 4 Ground
GPIO
15
6
9
Pin No. Assignment
1 No Connection 2 TXD 3 RXD 4 No Connection 5 Ground 6 No Connection
7 No Connection 8 No Connection 9 No Connection
RS232C (D-SUB 9Pin)
2
1
Pin No. Assignment
1 S-VIDEO/Y (Luma)
2 S-VIDEO/C (Chroma)
3 Ground
S-Video (BNC)
3
Pin No. Assignment
1 SDI IN
2 SDI OUT
3 Ground
SDI (BNC)
2
1
3
English - 54
R/Pr
B/Pb
Assignment
Pin No.
RGBHV/RGBS/YPbPr (BNC)
G/Y
CS/HS
VS
1
3
Pin No. Assignment
1 OPTICAL Clock 2 OPTICAL Blue 3 OPTICAL Green 4 OPTICAL Red
OPTICAL , LC Type (Matched with DDL receivers)
2 4
1
23
45
1 Red Pr 2 Green Y 3 Blue Pb 4 H-Sync / C-Sync No Connection 5 V-Sync No Connection 6 Ground
RGBS Y Pb Pr
6
Pin No. Assignment
1 OPTICAL IN
OPTICAL , SC Type (Matched with DDL receivers)
Pin No. Assignment
1 OPTICAL IN
OPTICAL , LC Type (Matched with DOL receivers)
55 - English
Type 19” TFT-LCD Screen Size 19 inch(5:4) Maximum Resolution 1280 X 1024 @ 60Hz Pixel Pitch
0.294(H) mm X 0.294(V) mm Display Colors 16.7M Contrast Ratio(Typical) 800:1 Viewing Angle 85° / 85° / 85° / 85° Response Time 14 msec(Gray to Gray)
Luminance(Typical) 270cd/
Touch Panel ELO 5wire resistive touch screen Interface USB & SERIAL
Acrylic Double side Anti-Reflection coating
Signal input VGA~SXGA
Maximum Max 60W Standby Mode Max 20W
Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1x Optical DVI 1(DVI 2 option), 1x D-SUB, 1xBNC (CVBS)Video, 2 x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SDI),
5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS)
Video 1xDVI, 1xBNC (SD/HD/3G-SDI) DC 12V, 7A Max
423(W)X351.5(H)X76.5(D) (
mm
)
7.3Kg -Without stand
16.653(W)X13.838(H)X3.011(D) (inch)
16.09 lbs - Without stand
Optical
Characteristics
Touch Specification
(F-L190*DT only)
Model
FS-L190*D / FS-L190*DT
Front Filter
(FS-L190*D only)
Resolution
Specification
FS-L190*D / FS-L190*DT
Power
Control Key
Input Signal
Output Signal
Input Power
Dimension
Size and Weight
Video
English - 56
Type 24” TFT-LCD Screen Size 24 inch(16:10) Maximum Resolution 1920 X 1200 @ 60Hz Pixel Pitch
0.270(H) mm X 0.270(V) mm Display Colors 1.06Billion Contrast Ratio(Typical) 1000:1 Viewing Angle 89° / 89° / 89° / 89° Response Time 6 msec(Rising+Falling) Luminance(Typical) 400cd/
Touch Panel ELO 5wire resistive touch screen Interface USB & SERIAL
Acrylic Double side Anti-Reflection coating
Signal input VGA~WUXGA
Maximum Max 100W Standby Mode Max 20W
Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1x Optical DVI 1(DVI 2 option), 1x D-SUB, 1xBNC (CVBS)Video, 2 x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SDI), 5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS)
Video 1xDVI, 1xBNC (SD/HD/3G-SDI) DC 24V, 6.25A Max
580(W)X386(H)X95(D) (
mm
)
7.5Kg -Without stand
22.834(W)X15.196(H)X3.740(D) (inch)
16.53 lbs - Without stand
Optical
Characteristics
Touch Specification
(F-L240*DT only)
MODEL
FS-L240*D / FS-L240*DT
Front Filter
(FS-L240*D only)
Resolution
FS-L240*D / FS-L240*DT
Power
Control Key
Input Signal
Output Signal
Input Power
Dimension
Size and Weight
Video
57 - English
Type 26” TFT-LCD Screen Size 26 inch(16:10) Maximum Resolution 1920X 1200 @ 60Hz Pixel Pitch
0.2685(H) mm X 0.2685(V) mm Display Colors 16.7M Contrast Ratio(Typical) 1000:1 Viewing Angle 89° / 89° / 89° / 89° Response Time 6 msec(Rising+Falling) Luminance(Typical) 400cd/
Video input VGA~WUXGA Acrylic Double side Anti-Reflection coating
Maximum Max 130W Standby Mode Max 20W
Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1x Optical DVI 1(DVI 2 option), 1x D-SUB, 1xBNC (CVBS)Video, 2 x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SDI),
5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS) Video 1xDVI, 1xBNC (SD/HD/3G-SDI) DC 24V, 6.25A Max
618(W)X412(H)X99.5(D) (
mm
)
8.8Kg -Without stand
24.330(W)X16.220(H)X3.917(D) (inch)
19.4 lbs - Without stand
Optical
Characteristics
MODEL
FS-L260*D
FS-L260*D
Power
Input Signal
Input Power
Dimension
Size and Weight
Resolution
Front Filter
Control Key
Output Signal
Video
English - 58
Type 26” TFT-LCD Screen Size 26 inch(16:9) Maximum Resolution 1920X 1080 @ 60Hz Pixel Pitch
0.3(H) mm X 0.3(V) mm Display Colors 16.7M Contrast Ratio(Typical) 1000:1 Viewing Angle 89° / 89° / 89° / 89° Response Time 6 msec(Rising+Falling) Luminance(Typical) 350cd/
Video input VGA~1080p Acrylic Double side Anti-Reflection coating
Maximum Max 70W Standby Mode Max 20W
Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1x Optical DVI 1(DVI 2 option), 1x D-SUB, 1xBNC (CVBS)Video, 2 x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SDI),
5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS) Video 1xDVI, 1xBNC (SD/HD/3G-SDI) DC 24V, 6.25A Max
638(W)X389(H)X74.7(D) (
mm
)
7.3Kg -Without stand
25.118(W)X15.315(H)X2.941(D) (inch)
16.1 lbs - Without stand
Optical
Characteristics
MODEL
FS-L260*D(LED)
FS-L260*D(LED)
Power
Input Signal
Input Power
Dimension
Size and Weight
Resolution
Front Filter
Control Key
Output Signal
Video
59 - English
Type 26” TFT-LCD Screen Size 26 inch(16:9) Maximum Resolution 1920X 1080 @ 60Hz Pixel Pitch
0.3(H) mm X 0.3(V) mm Display Colors 1.06B Contrast Ratio(Typical) 1400:1 Viewing Angle 89° / 89° / 89° / 89° Response Time 8 msec(Rising+Falling)
Luminance(Typical) 450cd/
Video input VGA~1080p Acrylic Double side Anti-Reflection coating
Touch(Option) 26” IR Multi Touch(USB) Glass(Option) Optical bonding Maximum Max 70W
Standby Mode Max 20W Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1x Optical DVI 1(DVI 2 option), 1x D-SUB, 1xBNC (CVBS)Video, 2 x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SDI), 5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS)
Video 1xDVI, 1xBNC (SD/HD/3G-SDI)
DC 24V, 6.25A Max
638(W)X389(H)X74.7(D) (
mm
)
7.6Kg -Without stand
25.118(W)X15.315(H)X2.941(D) (inch)
16.76 lbs - Without stand
Optical
Characteristics
MODEL
FS-P260*D(LED)
FS-P260*D(LED)
Power
Input Signal
Input Power
Dimension
Size and Weight
Resolution
Front Filter
Control Key
Output Signal
Video
English - 60
Type 32” TFT-LCD Screen Size 32 inch(16:9) Maximum Resolution 1920X 1080 @ 60Hz Pixel Pitch
0.363(H) mm X 0.363(V) mm Display Colors 1.06 billion Contrast Ratio(Typical) 1300:1 Viewing Angle 89° / 89° / 89° / 89° Response Time 6msec(Rising+Falling) Luminance(Typical) 500cd/
Video input VGA~1080P Glass Double side Anti-Reflection coating
Maximum Max 150W Standby Mode Max 20W
Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1xOptical DVI 1(DVI 2 option), 1x D-SUB, 1xBNC (CVBS)Video, 2 x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SD),
5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS) Video 1xDVI, 1xBNC (SD/HD/3G-SDI) DC 24V, 7.5A Max
795(W)X485(H)X100(D) (
mm
)
15.8Kg -Without stand
31.299(W)X19.094(H)X3.937(D) (inch)
34.8 lbs - Without stand
Optical
Characteristics
MODEL
FS-L320*D
FS-L320*D
Power
Input Signal
Input Power
Dimension
Size and Weight
Resolution
Front Filter
Control Key
Output Signal
Video
61 - English
Type 42” TFT-LCD Screen Size 42 inch(16:9) Maximum Resolution 1920X 1080 @ 60Hz Pixel Pitch
0.4845(H) mm X 0.4845(V) mm Display Colors 1.06 billion Contrast Ratio(Typical) 1300:1 Viewing Angle 89° / 89° / 89° / 89° Response Time 6 msec(Rising+Falling) Luminance(Typical) 500cd/
Video input VGA~1080P Glass Double side Anti-Reflection coating
Maximum Max 260W Standby Mode Max 20W
Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1xOptical DVI 1(DVI 2 option), 1xD-SUB, 1xBNC (CVBS)Video, 2x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SDI), 5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS)
Video 1xDVI, 1xBNC (SD/HD/3G-SDI) AC 100~230V, 50~60Hz,3A Max
1024.6(W)X617.4(H)X111.1(D) (
mm
)
28Kg -Without stand
40.338(W)X24.307(H)X4.374(D) (inch)
61.7 lbs - Without stand
Optical
Characteristics
MODEL
FS-L420*D
FS-L420*D
Power
Input Signal
Output Signal
Dimension
Size and Weight
Resolution Front Filter
Control Key
Input Power
Video
English - 62
Type 55” TFT-LCD Screen Size 55 inch(16:9) Maximum Resolution 1920X 1080 @ 60Hz Pixel Pitch
0.630(H) mm X 0.630(V) mm
Display Colors 1.06 billion Contrast Ratio(Typical) 1300:1 Viewing Angle 89° / 89° / 89° / 89° Response Time 6 msec(Rising+Falling) Luminance(Typical) 450cd/
Video input VGA~1080P Glass Double side Anti-Reflection coating
Maximum Max 200W Standby Mode Max 20W
Front Side INPUT, -, +,
, , PIP, MENU, POWER
1xDVI, 1xOptical DVI 1(DVI 2 option), 1xD-SUB, 1xBNC (CVBS)Video, 2x BNC (SVHS Y/C), 1 x BNC(SD/HD/3G-SDI),
5 x BNC (Component Y/G, Pb/B, Pr/R, H/CS, VS)
Video 1xDVI, 1xBNC (SD/HD/3G-SDI) AC 100~230V, 50~60Hz,3A Max
1293.0(W)X777.0(H)X86.3(D) (
mm
)
37Kg -Without stand
50.91(W)X30.59(H)X3.4(D) (inch)
81.57 lbs - Without stand
Optical
Characteristics
MODEL
FS-L550*D
FS-L550*D
Power
Input Signal
Output Signal
Dimension
Size and Weight
Resolution Front Filter
Control Key
Input Power
Video
63 - English
Cleaning Instructions
Precautions
Before cleaning, switch the display in stand-by position to prevent the control touch panel from being activated inadvertently by sweeping over the front filter. In stand-by position the touch panel cannot be activated by just sweeping over them. To switch the display on again, you must press the stand-by key again.
Take care not to damage or scratch the front filter or LCD panel.
Be careful with rings or other jewellery that can touch the front filter.
Do not apply pressure on the front filter or LCD panel.
Do not apply or spray liquid directly to the front filter, panel or
cabinet as excess liquid may cause damage to internal electronics. Instead, apply the liquid to the cleaning cloth.
Follow your hospital protocol for the handling of blood and body fluids.
The display is not disinfected or packed in sterile environment.
Follow your hospital protocol in case the display needs to be
disinfected prior to installation.
Front Filter
Proceed as follows:
1. Remove dust with a dry, lint-free, non-abrasive soft cotton cloth.
2. Remove fingerprints or grease using a lint-free, non-abrasive soft cotton cloth that is lightly moistened with plain water or a mild commercial glass cleaning product suited for coated glass surfaces.
3. Gently wipe dry with a dry cloth.
The following products are tested and approved:
Misty Clear Lemon 10 Disinfectant
Bohle glass cleaner
Zep Heavy-duty glass & all surface cleaner
English - 64
Klear Screen
Screen TFT (Kontakt Chemie)
Incidin Foam (Ecolab)
Microzid
Mild detergent
Isopropyl alcohol with concentration < 5%
Household bleach (generic sodium hypochlorite, solutions of 5.25%
sodium hypochlorite diluted with water between 1:10 and 1:100)
Do NOT use:
Alcohol/solvents at higher concentration > 5%
Strong alkalis, strong solvents
Acid
Detergents with fluoride
Detergents with ammonia
Detergents with abrasives
Steel wool
Sponge with abrasives
Steel blades
Cloth with steel thread
Cabinet
Proceed as follows:
Clean the cabinet using a soft cotton cloth, lightly moistened with a recognized cleaning product for medical equipment.
Repeat with water only.
Wipe dry with a dry cloth.
The cabinet has been tested for resistance to the following products:
Virex Ready-to-use Disinfectant Cleaner
Misty Clear Lemon 10 Disinfectant
Misty Multi-Purpose Disinfectant Cleaner
Misty Multi-Purpose Disinfectant Cleaner II
65 - English
Zep Heavy-duty glass & all surface cleaner
Klear Screen
Screen TFT (Kontakt Chemie)
Incidin Foam (Ecolab)
Microzid
Mild detergent
Isopropyl alchohol with concentration < 5%
Household bleach (generic sodium hypochlorite, solutions of 5.25%
sodium hypochlorite diluted with water between1:10 and 1:100)
Precise Hospital Foam Cleaner Disinfectant
English - 66
DVI ②(Fiber) install guide (Rx)
Pull out the DVI ②slot gender by loosening 2 screws on the bracket
DDL(4ch LC) Rx installation
Insert DDL Rx into blank DVI2 slot and fasten it by 2 screws with DDL bracket.
DDL(1ch LC) Rx installation
Insert DOL Rx into blank DVI2 slot and fasten it by 2 screws with DOL bracket.
67 - English
DSP/DSK (1ch SC) Rx installation
Insert DSP Rx into blank DVI2 slot and fasten it by 2 screws with DSP/DSK bracket.
DSL (1ch SC) Rx installation
Insert DSL Rx into blank DVI2 slot and fasten it by 2 screws with DSL bracket.
Fiber extender Order code Drawing Application models
DDL 639494070400
DOL 639494083000
DSP/DSK 639494070300
DSL 639494083100
Please place an order to provider to get each bracket for fiber extenders
FS-L240*D FS-L260*D FS-P260*D FS-L320*D FS-L420*D FS-L550*D
(*means numeric depend on customer)
Thank you for choosing our product.
Service
Please contact our customer service if you need any
information or help with our products.
Warranty
One year, parts and labor
ASIA FORESEESON KOREA
Room404-b, Pangyolnnovalley B, Sampyeong-Dong621, Bundang-gu, Seongnam-si, Gyeonggi-do,korea Tel.+82(31)8018-0780 Fax. +82(31)8018-0786 www.fsnmed.com
USA FORESEESON CUSTOM DISPLAY INC
2210 E. Winston Road, Anaheim, CA 92806 USA Tel.1-714-300-0540 Fax. 1-714-300-0546
EC Representative FORESEESON GmbH
Industriestrasse 18a, 63150 Heusenstamm, Germany Tel.+49(0)6104-643980
FORESEESON UK
Unit 2, Kingsmill Business Park Chapel Mill Road,Kingston upon Thames,Surrey KT1 3GZ,UK Tel.+44-(0)208-546-1047
D&T Inc. Jang-dong, Daedeok Valley, 26-121, Gajeongbuk-ro, Yuseong-gu, Daejeon, Korea Tel : 82-2-360-8000 Fax. +82(2)360-8005
949494061302-English
Loading...