Welcome to the newest innovation in hearing aid testing! The 7000 Test System
is the long-awaited successor to the 6500-CX, the analyzer that arguably has
become the standard by which all other hearing aid test equipment is measured.
The 7000 Test System has many design improvements, including a high-resolution color display, a flexible and easy-to-use interface, and built-in help screens.
Throughout the design process of the 7000 Test System, the engineers at Frye
Electronics have followed two important mantras: Flexibility and Usability. We
wanted all the features of the existing 6500-CX system in a new package that was
easier to use and easier to upgrade. We are confident that the 7000 Test System is a
success on both counts
1.1 History
The 6500-CX has been one of the fastest and most accurate hearing aid analyzers in the industry. Since its introduction in 1985 at the International Audiology
Congress in Prague, Czechoslovakia, it has evolved through three model num
bers: the original 6500, the 6500-C with color, and the 6500-CX with a faster
CPU. These hardware upgrades, and the software upgrades that went with them,
kept the 6500 analyzer current as hearing aid technology evolved.
-
The original 6500 operator interface was originally designed around a set of
tests that were needed at the time of its introduction. Linear and AGC hearing
aids were tested for ANSI coupler performance, and real ear tests were done to
check on insertion gain performance. Its Composite signal gave the industry the
first real time, interactive frequency response measurement for both coupler
and real ear tests. This enabled an order of magnitude increase in the collec
tion speed of response data. This feature has been particularly useful in real ear
tests, where artifacts caused by ambient noise in the acoustical environment are
a factor.
The 6500 evolved to keep pace with the technology of the hearing aid industry.
The 6500-C with color video came in 1992. The CX version with its faster CPU
was introduced in 1994. Functions such as Multi-curve, Target 2-cc, ANSI 92,
CIC, Digital-Speech-in-Noise, the real-ear SPL screen, and Enhanced DSP were
gradually introduced, making the 6500-CX analyzer more powerful with each
release.
The number of tests on the 6500-CX grew, but the operator interface remained
much the same. To accommodate new functions, 6500 front panel buttons were
programmed with functions different from their original purpose and labeling.
While the help messages on the screen allowed the first time user to operate the
instrument, it was obvious that improvements could be made. Some of these
-
2FONIX 7000 Hearing Aid Analyzer
were implemented and tested on other Frye instruments. The FP40 analyzer
introduced “function buttons” that adapt to handle the specific needs of each
test screen and sequence. The FP35 analyzer added help screens and local
menus to make the analyzer more intuitive and easier to use.
The lessons learned in the FP40 and FP35 analyzers have been incorporated into the
7000 Test System. It is hoped that these improvements will make the 7000 flexible
enough to adapt to any new test procedures as they become needed with the changing testing requirements of the hearing aid industry
1.2 Basic Test Functions
The basic 7000 can perform the following coupler tests:
• Pure-tone signal measurements
• Composite and Digital Speech signal measurements
• Phase measures for binaural hearing aid pairs
• Group delay (processing delay) measurement of digital hearing aids
• Battery Current as a function of amplitude and frequency
For both composite and pure tone measurements, the following are possible:
An additional function that comes standard with the 7000 is RS232 communication
with an external personal computer. This allows remote operation of the analyzer
from the computer, and lets you save test results onto your computer’s hard drive.
Firmware upgrades to the 7000 Test System can also be performed though the computer.
1.3 Options
The 7000 test system has the following software (and hardware) options available:
Real-Ear Option
• I/O Curves
• Attack and Release
• Special CIC hearing aid coupler measurements
• Behind the ear (BTE), In the ear (ITE), in the canal (ITC), Open ear, and
body hearing aid measurements
• Choice of either IEC 60118-7 or ANSI S3.22-2003/96/87/92 standard tests
(All tests can be installed as options)
The 7000 test system can be ordered with the Real-Ear Option so that tests can
be done on the hearing aid while it is in the client’s ear. These measurements
are also commonly called “probe measurements” because they are performed
Introduction and Setup 3
with a probe microphone. Using real-ear measurements makes it possible to
individualize the fitting of a hearing aid since a coupler measurement can sel
dom tell the operator exactly how what sound is received by the client. Many
factors affect the sound on its way to the ear drum. When measuring with a
probe microphone, you will know what is happening in the “real-ear.”
ANSI Option
The ANSI Option gives the user access to the ANSI 03, ANSI 96, ANSI 87, and
ANSI 92 test sequences. These perform measurements according to the ANSI
S3.22 2003, ANSI S3.22 1996, ANSI S3.22-1987, and ANSI S3.42-1992 stan
dards, respectively.
This option also includes the Telewand, which is used for performing ANSI 96
telecoil measurements.
IEC Option
The IEC Option gives the user access to the IEC test sequence, which performs
measurements in accordance with the IEC 118-7, the performance part of the
IEC standard.
OES Option
The Occluded Ear Simulator (OES) Option includes the MZ-1, MZ-2, and MZ-3
couplers and corresponding correction factors meant to simulate the results of a
real-ear simulator (Zwislocki) coupler when simulating occluded ear measure
ments in the sound chamber. The results produced with the OES Option will
not be an exact duplication of the results produced with a Zwislocki coupler.
-
-
-
VA CORFIG
The VA-CORFIG Option supplies a frequency response in the ANSI test
sequence that includes correction factors that make the measurement closer to a
real-ear response. These correction factors were developed by the United States
Veterans Administration.
4FONIX 7000 Hearing Aid Analyzer
1.4 Accessories
The following are accessories that are standard and optional on your FONIX
7000 hearing aid analyzer.
1.4.1 Standard Accessories
The following accessories come standard with each 7000 test system.
Coupler Microphone—M1950E
14 mm pressure-type electret instru
mentation microphone.
Note: Older 7000 analyzers may use
the M1750E microphone.
HA-1 2-cc Coupler
Dimensions per ANSI S3.7 for testing
in-the-ear aids, canal aids, and aids fit
ted with earmolds.
-
-
HA-2, 2-cc Coupler
Dimensions per ANSI S3.7 for testing
behind-the-ear aids, eyeglass aids, and
body aids.
Ear-Level Adapter (BTE)
Snaps into the 1/4" (6.35 mm) diameter
cavity of the HA-2 and MZ-2 couplers.
Equipped with a 0.6" (15 mm) length
of 0.076" (1.93 mm) ID tubing, the
adapter allows ANSI S3.22 specified
connection of an ear-level aid to the
coupler.
Introduction and Setup 5
Microphone Adapter
14 mm to 1" (25.4) mm adapter for
coupling the M1950E microphone
and reference microphone to a 1-inch
device such as a sound level calibra
tor.
Test Chamber Cable
Connects the main module to the test
chamber.
-
Other Standard Accessories:
Thermal Paper
Fun-Tak
RS232 Cable
Dummy Microphone (for testing to ANSI specifications)
Sound Chamber Replacement Feet
Operator’s Manual
Maintenance Manual (upon request at time of purchase)
Battery Substitution Pills
Used to power hearing aids, measure
battery drain, and estimate battery life.
Standard sizes:
675/76
13
312
10A/230
6FONIX 7000 Hearing Aid Analyzer
1.4.2 Real-Ear Accessories
When the Real-Ear Option is ordered, the following accessories are
included.
Remote Moduleand Microphones
The remote module allows the user to
control the analyzer while remaining
close to the patient. The probe and ref
erence microphones, used for perform
ing real-ear measurements, are plugged
into this module.
Sound Field Speaker
Produces the output for most real-ear
measurements. The Real-Ear Option
can be ordered with a floor stand,
remote module shelf, or with a swing
arm.
-
-
Loudspeaker Floor Stand
The lightweight, sturdy floor stand is
adjustable and has a low-profile tripod
base.
Remote Module Shelf
The shelf fits directly on the floor
stand, providing a stable, convenient
platform for operating and storing the
remote module. This is the default
configuration.
Loudspeaker Swing Arm
This swing arm may be substituted for
the speaker floor stand & shelf. It is a
convenient device that can be mounted
to a wall or desktop. The four-piece
arm swivels and extends along five
different dimensions to allow precise
placement and aiming of the loud
speaker. Spring tension and friction
keep the loudspeaker exactly where
you put it.
-
Introduction and Setup 7
Monitor Headset
Plugs into the remote module, and
allows the user to listen to the signal
being measured by the probe micro
phone. Model of headset may differ
from one shown.
-
Wedge-style Ear Hook
Standard Size. Holds probe and refer
ence microphone during real-ear test
ing.
Probe Tubes
Used for performing real-ear measure
ments. Probe tubes are not reusable.
-
-
-
8FONIX 7000 Hearing Aid Analyzer
1.4.3 Optional Accessories
Infant/Child Headband
Includes infant, child, and adult head
bands, six bendable earhooks, and two
sets of “animal ears.”
6-cc Coupler
Enables checking the response of stan
dard audiometer headphones. It is
designed around the NBS 9A coupler
specifications.
-
-
MZ Couplers
MZ-3, MZ-2, and MZ-1 couplers are
supplied with the OES Option.
CIC Coupler
Non-standard (0.4 cc) coupler used for
realistic testing of CIC hearing aids.
This coupler is always used in con
junction with the CIC software correc
tion factors
-
-
Introduction and Setup 9
Open Ear Coupler
Non-standard coupler used for realistic
testing of open ear hearing aids.
#5 battery pill
Provides battery current measurements
for many CIC hearing aids.
Also available:
#41 battery pill
AA
Sound Level Calibrator A portable,
self-contained, field-type calibrator for
calibrating the microphone amplifier.
It operates on one nine-volt transistor
battery. Use with 14 mm-to-1" adapter.
Calibration is traceable to the U.S.
National Institute of Standards and
Technology. Conforms to ANSI @1.401984 and IEC 942: 1988.
10FONIX 7000 Hearing Aid Analyzer
Sound Chamber Stand
A secure, steel-tube stand that
improves sound isolation and brings
the testing area of the test chamber to
convenient table height. Color matches
the electronics module.
Sound Chamber Spring Damper
Factory-installed.or Assembly Kit.
This spring assembly fits onto the back
of the sound chamber, making it easier
to open the sound chamber lid. Once
the sound chamber is open past a cer
tain point, the user can release the lid
and let the spring open it the rest of the
way. Great when performing repetitive
coupler measurements.
FM Kit
Facilitates coupler and real-ear tests
of FM systems. The kit includes a
telescoping floor stand with a test plat
form and plenty of extra Fun-Tak and
a 6-inch (15 cm) square foam pad for
using the 2-cc coupler outside the test
chamber.
-
-
Introduction and Setup 11
POWER
STANDBY
1.5 Layout, Controls, and Safety
This section describes the basic layout of the front and rear panel of the 7000
test system, and special information pertaining to safety regulations.
1.5.1 Front Panel Layout
The front panel of the 7000 test system consists of 8 function keys, 4 arrow
keys, and 9 other keys with varying operations. Together, all of these keys are
used to control all the operations of the 7000 test system. A wide internal ther
mal printer is located to the right of the front panel. See Figure 1.5.1.
RECD Earphone Package
Consists of one ER3A earphone with
a 1⁄4 inch plug, a 72 inch cable, an
assortment of ear tips, a calibration
certificate, and a lapel clip. This pack
age is suitable for performing an RECD
measurement with the 7000 Test
System.
-
Figure 1.5.1—Front panel
FUNCTION KEYS:
The top row of buttons on the 7000 front panel consists of “function keys”
labeled [F1] through [F8]. The function of these keys change from screen to
screen, according to the need of the screen. There are three basic uses of the
function keys:
12FONIX 7000 Hearing Aid Analyzer
• To navigate from screen to screen
• To toggle a selection
• To pop-up up a selection menu
For a more complete description of the operation of the function keys, see
Section 2.1.1.
[RESET]: Resets the analyzer and returns you to the opening screen.
Many settings, measurements, and test conditions are erased
or returned to their default conditions with the push of this
button. Use [EXIT] instead of [RESET] to avoid the loss of
data.
[HELP]: Brings up a pop-up help window, which tells you what func-
tions can be performed in the current screen.
[MENU]: Opens a local menu, or the Default Settings menu.
[EXIT]: Exits from the current screen, leaving all settings, measure-
ments, and test conditions as is.
[LEVEL]: Levels the sound chamber or sound field speaker.
[FEED]: Advances the internal thermal paper feed.
[PRINT]: Prints current screen.
[STOP]: Stops testing.
[START]: Starts testing.
ARROW KEYS: The arrow keys perform different functions depending upon
the screen and previous buttons pushed. In a test situation,
they adjust the source amplitude and frequency. If you press
one of the function buttons to bring up a pop-up menu, they
are used to scroll through the available options and make
selections.
LEDs: Power: When green, this LED indicates that the 7000 Test
System is powered on.
Standby: When green, this LED indicates that the screensaver
mode is engaged (screensaver not available at this time)
Introduction and Setup 13
FONIX 7000 Serial No. 1234
Made in Tigard, Oregon, USA
by
Frye Electronics, Inc.
Frye Electronics products are protected by
U.S.
and/or foreign patents and/or patents pending
T630 mAL
1.5.2 Rear Panel Layout
The rear panel of the 7000 test system contains most of the external connections for the analyzer. See Figure 1.5.2 for a diagram of the back panel.
Figure 1.5.2—Back panel
1. PRINTER:Connects an external printer to the 7000 test system.
2. SPKR: Connects the sound field speaker for real-ear measurements.
3. EARPHONE: Connect a insert earphone to the analyzer for RECD measurements. A 50 ohm earphone should always be used, to
avoid harm to the earphone.
4. SOUND
CHAMBER: Connects the main module to the sound chamber.
5. RS232: Connects the 7000 test system to a personal computer.
6. RS232 LEDs: The red LED indicates the analyzer is sending a poll. This
will flash whether or not a computer is connected. The
green LED indicates the analyzer is receiving a command
from a connected computer.
7. MONITOR: Connects the main module to a video monitor.
8. GAIN: Adjusts the calibration of the M1950E microphone.
9. MICROPHONE: Connects the M1950E microphone.
10. SCOPE: Connects to an external scope for external measuring purposes.
14FONIX 7000 Hearing Aid Analyzer
11. PROBE: Connects the main module to the remote module (used for
real-ear measurements).
KEYBD: Not currently active.
AUX: Not currently active.
The following safety symbols can be found on the back of the 7000 hearing aid
analyzer:
1.5.3 Safety
Safety Classification for IEC 60601-1
Type of protection against electric shock: Class I
Degree of protection against electric shock: Type B
Read the accompanying documents. Please read this manual
This symbol indicates that Frye Electronics, Inc. conforms to
Replace fuses only with the same type and rating.
Type B Equipment. The probe microphone and insert earphone are type B applied parts, according to IEC 60601-1.
before operating the 7000 test system. A separate maintenance
manual exists for the 7000 test system. If you wish to obtain a
maintenance manual, please contact Frye Electronics, Inc. or
your Frye representative.
the Medical Device Directive 93/42/EEC. Any attached video
monitor, external printer, or external computer should also
have a CE mark in order for the 7000 test system to remain
compliant.
Protection against harmful ingress of water: Ordinary
Mode of operation: Continuous
The 7000 does not require sterilization or disinfection.
Warning: This equipment is not suitable for use in the presence of flammable
anaesthetic mixture with air or with oxygen or nitrous oxide.
Introduction and Setup 15
Connection of peripheral equipment to the 7000
Compliance with IEC 60601-1-1 Safety requirements for medical electrical sys
tems must be determined on a case-by-case basis.
All electrical equipment attached to the 7000, such as video monitors, computer
equipment, etc. must, at a minimum, meet one of the following conditions:
a. The equipment complies with IEC 60601-1
b. The equipment complies with relevant IEC and ISO safety standards and is
supplied from a medical grade isolation transformer.
c. The equipment complies with relevant IEC and ISO safety standards and is
kept at least 1.5 meters from the patient.
The allowable leakage currents of IEC 60601-1-1 must not be exceeded. IEC
60601-1-1 should be consulted when assembling such a system.
Electromagnetic compatibility
The 7000 complies with IEC 60601-1-2.
The 7000 generates and uses radio frequency energy. In some cases the 7000
could cause interference to radio or television reception. You can determine if
the 7000 is the source of such interference by turning the unit off and on.
-
If you are experiencing interference caused by the 7000, you may be able to cor
rect it by one or more of the following measures:
1. Relocate or reorient the receiving antenna.
2. Increase the distance between the 7000 and the receiver.
3. Connect the 7000 to a different outlet than the receiver.
In some cases radio transmitting devices, such as cellular telephones, may cause
interference to the 7000. In this case try increasing the distance between the
transmitter and the 7000.
Disposal of the 7000 and accessories
The 7000 and some of its accessories contain lead. At the end of its useful life,
please recycle or dispose of the 7000 according to local regulations.
If you are located in the European Union, please report all safety-related con
cerns to our authorized representative:
Siemens Hearing Instruments Ltd.
Alexandra House
Newton Road
Manor Royal
Crawley
West Sussex RH109TT
ENGLAND
-
-
16FONIX 7000 Hearing Aid Analyzer
Otherwise, please report all safety related concerns to the Frye factory:
Frye Electronics, Inc.
9826 SW Tigard St.
Tigard, OR 97223
Ph: (503) 620-2722 or (800) 547-8209
Fax: (503) 639-0128
email: support @frye.com
sales@frye.com
Frye Electronics, Inc. is a Registered Firm of British Standards
Institution, and we conform to the ISO 13485 standard.
1.6 Setup
This section describes the unpacking and setup of the 7000 test system. A
description of the front and back panels is also included.
1.6.1 Unpacking the FONIX 7000
Remove the FONIX 7000 test system from the shipping cartons. Store the shipping cartons in a dry place so that they can be used again in the event that the
unit must be returned to the factory for repair or upgrade.
See Figure 1.6.1 for a picture of the 7000 test system and a labeling of its basic
components.
Figure 1.6.1. Caption: The 7000 Test system.
Introduction and Setup 17
1.6.2 Locating and arranging the 7000 test system
Locate the 7000 test system in a moderately quiet area, such as a private office
or laboratory. In order for test results to reliably conform to specifications, the
ambient noise, mechanical vibrations, electrical or magnetic fields must not
affect test results by more than 0.5 dB (ANSI S3.22). Low-noise acoustic condi
tions, as found in sound treated rooms and booths, are recommended but not
necessary.
Locate the main module near the sound chamber. The video monitor may be
placed on top of the main module.
If your 7000 has the Real-ear Option, further consideration needs to be given
to the placement of the sound field speaker. The speaker should be placed in
a manner in which the sound will not reflect off walls, chairs, tables, or other
objects. Such sound reflections could negatively affect test results.
1.6.3 Connecting the Main Module & Components
Connect the main module to the system components as follows (see Figure
1.6.3):
-
1. Connect the video monitor to the jack marked “Monitor” on the back of the
main module. Tighten the screws on the connector to secure the connection.
2. Plug the M1950E microphone into the connector marked “Microphone”
located on the back of the main module.
3. Connect the sound chamber cable into the plug marked “sound chamber” on
the back of the main module. Plug the other end of the cable into the sound
chamber. This cable provides all the connections necessary for operating the
internal loudspeaker, telecoil, and battery voltage supply of the 7020 test
chamber. Tighten the screws of the plug at each end of the cable to secure
the connection.
4. Plug the power cord into the three-pronged jack on the rear panel of the
main module. Plug the power cords of both the main module and the video
monitor into an AC outlet. We recommend using a switchable multiple out
let power strip. Any voltage between 100 and 240 VAC is acceptable, 50 or
60 Hz line frequency.
-
18FONIX 7000 Hearing Aid Analyzer
PRINTER
SOUND CHAMBER
EARPHONE
SPKR
MONITOR
RS232
AU
X
KEYBD
MICROPHONE
GAIN
SCOPEPROBE
FONIX 7000 Serial No. 1234
Manufactured in 2003
CALIBRATED
Frye Electronics, Inc.
800-547-8209
www
.frye.com
DATE CAL
DATE DUE
ATTENTION: Ground reliability can be achieved
ONLY when equipment is connected to “HOSPITAL
GRADE” receptacles.
For service refer to qualified
personnel only.
Consult accompanying documents.
ELECTRICAL RATING:
100–240
V
~
50-60 Hz .6A
Made in Tigard, Oregon, USA
by Frye Electronics, Inc.
Frye Electronics products are protected by
U.S.
and/or foreign patents and/or patents pending
T630 mAL
Earphone
Speaker
External printer
(not provided)
Monitor
Headphones
Computer
(not provided)
Telecoil board
(optional
)
Te
lewand (optional)
7020 Sound Chamber
Quik-Probe
remote modul
e
M1950E
Microphone
Probe/Reference mics
Figure 1.6.3 —7000 test system setup
1.6.4 Connecting the Real-Ear Equipment
If you ordered the Real-Ear Option, connect the real-ear equipment as follows.
See Figure 1.6.3.
1. Plug the remote module cable into the jack labeled “Probe” on the back of
the main module. Tighten the connector screw to secure the connection.
2. Plug the round, 8-pin connector of the dual microphone cable into the top
surface of the remote module where it is marked “PROBE MIC.”
3. Plug the monitor headset into the remote module by attaching the ¼” adapt
er and plugging it into the jack marked “earphone,” if desired.
-
Introduction and Setup 19
4. Depending on whether you have a floor stand or a swing arm, either (1)
Unfold and extend the floor stand to the desired height, about the level of a
seated client’s ear, or (2) Mount the wall/desk bracket of the swing arm in a
convenient location. Attach the loudspeaker to the stand or arm, using the
threaded connection.
5. Plug the dual banana plug into the rear jacks of the loudspeaker. Then plug
the other end of the cable into the main module where it is marked “SPKR.”
1.6.5 Powering up the 7000 test system
1. Turn on the electronics module by reaching back on the right side of the
main module (while you are facing the front of the unit), and pressing
the POWER switch. The green LED on the front panel, above the words
“POWER” will turn on after a couple seconds.
2. Turn on the video monitor. See the video operator’s manual for instructions,
if necessary.
The Opening screen of the 7000 test system will be displayed. It contains the
FONIX logo, along with the software version, the serial number, and the avail
able options. This information is important in identifying your equipment when
you communicate with the company about your FONIX 7000 test system.
-
1.7 Maintenance
In order to ensure accurate testing, we recommend yearly calibrations of your
coupler microphone. For those units with the Real-ear Option, we also recom
mend yearly calibration of the probe & reference microphones and the sound
field speaker. These calibrations can be performed by any Frye distributor or by
yourself with the aid of a sound level calibrator such as a Quest QC-10. Please
see Appendix B for calibration instructions.
1.7.1 Cleaning
For your safety, disconnect the 7000 from mains power while cleaning. Wipe the
7000 with a slightly moist but not dripping cloth. Use plain water or water with
mild dish washing detergent. Wipe away any detergent with a slightly moist
cloth. Finally, use a soft dry cloth to eliminate any lingering moisture.
Never allow fluid to enter the 7000:
• enclosure
• power switch
• power connector
• electrical connectors
• front panel buttons
-
20FONIX 7000 Hearing Aid Analyzer
The 7000 microphones should be wiped with a dry cloth. Excess moisture may
damage the microphones.
Solvents and abrasives will cause permanent damage to the 7000.
1.7.2 Care & Maintenance of your M1950E Coupler Microphone
Microphones are perhaps the most delicate of FONIX products. They require
care and concern when handled. At the risk of sounding negative, the following
is a list of the five microphone DON’Ts.
1. Don’t bend the microphone cord at the connection of the cord and the body!
This is perhaps the most common cause of microphone damage. The place
where the cord and the body of the microphone meet is the most fragile part
of their connection. The less strain you put on this area, the longer your
microphone will last.
2. Don’t hit the microphone against the sound chamber or any other object!
The microphone head is delicate and easily damaged. Hitting the micro
phone against anything will do you no good and the microphone much
harm. Similarly, be careful not the drop the microphone on the floor.
-
3. Don’t pull the microphone from the coupler by the cord!
When removing the microphone from the coupler, grip the microphone
body, not the cord. Pulling on the cord damages the connection.
4. Don’t get Fun-Tak on the grill of the microphone!
Fun-tak that accumulates on the grill will fall into the microphone body and
onto the diaphragm which will damage the microphone.
5. Don’t touch the diaphragm of the microphone!
The diaphragm is located inside the grill of the microphone and is extremely
sensitive. If you need to take the grill off the microphone, please do so with
extreme care, and be especially careful not to touch the diaphragm, as this
will cause damage!
Overall, just be aware that your coupler microphone is a delicate instrument
that needs to be treated with care and respect. If you do so, the two of you will
be rewarded with a long, happy relationship.
1.7.3 Locating the serial number and software version
When contacting the factory, it is very useful if you have your serial number
and software version handy. This information will help us help you regarding
repair, upgrade, and technical questions.
The serial number of your 7000 test system can be found on the back of the
main module, just to the right of the power switch. See Figure 1.7.3A.
Introduction and Setup 21
Figure 1.7.3A—Serial number location
The software version of your 7000 test system can be found on the Opening
screen. See Figure 1.7.3B.
Press [HELP] to open a help window with more detailed service information.
This data may be requested if you call in with a repair-related question. Press
[HELP] again to close the help window.
Figure 1.7.3B—Software version location
22FONIX 7000 Hearing Aid Analyzer
1.7.4 Contacting the Factory
If you have any questions regarding your instrument, have your serial number
and software version ready (see Section 1.5.3), and contact us at:
Mailing address: Shipping address:
Frye Electronics, Inc. Frye Electronics
PO Box 23391 9826 SW Tigard St.
Tigard OR 97281-3391 Tigard, OR 97223
USA USA
Phone: (800) 547-8209 or (503) 620-2722
Fax: (503) 639-0128
The 7000 test system and its accessories are guaranteed to be free of manufacturing defects that would prevent the product from meeting its specifications
(given in Appendix A of this manual) for a period of one year from the date of
purchase.
Loading...
+ 186 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.