Frederiksen 4410.30 User Manual

Note:
The purpose of this apparatus is to
measure very small electrical
currents and charges. It is there ­fore very sensitive to disturbances that you normally would neglect completely. Please read the section “Application hints”
The electrometer has on its left side input connect­ors for current (1) and charge (2). The reading is shown on a zero centered analogue instrument (5). The scale goes from “-50” to “50” and is read in connection with the setting of the switch (4). If for in­stance the switch is set to 500 nC, the meter will read from -500 nC to 500 nC.
During charge measurements the apparatus can be zeroed with the switch (8).
An output is provided between the sockets (9) and (10) for data logging or a demonstration meter. The voltage varies between -500 mV and 500 mV, corres ponding to meter readings between “-50” and “50”.
The yellow/green socket (7) to the left as well as the black socket (10) to the right are connected to chas­sis ground. The apparatus is driven by a 9 V battery, placed in the holder (3).
Current measurements
The input I is used with the switch settings 5 pA, 50 pA and 500 pA for measuring small currents.
The current goes through a 1 GΩ resistor. This results in a voltage drop in the three measuring ranges of respectively 5 mV, 50 mV and 500 mV. The current input is not protected against an over­voltage larger than 1 kV (corresponding to a current of 1 µA or 1,000,000 pA).
Charge measurements
The input Q is used with the switch settings 5 nC, 50 nC and 500 nC for measuring charge. The charge measurement can be zeroed by pressing
Reset Q.
Charge is transferred to an internal 1 µF capacitor. This results in a voltage in the three measuring ranges of respectively 5 mV, 50 mV and 500 mV. In connection with typical electrostatic experiments
Manual for electrometer no. 4410.30
09.12.08 Ae 4410.30
A/S Søren Frederiksen, Ølgod Tel. +45 7524 4966 info@frederiksen.eu Viaduktvej 35 · DK-6870 Ølgod Fax +45 7524 6282 www.frederiksen.eu
®
1
2
3 4 5 6
7 8 9
10
(with voltages in the kilovolts range), this voltage can be considered to be 0 V.
In other words, when measuring for instance the charge of a metal sphere, the measuring process can be viewed as a discharge of the sphere while concurrently keeping track of the amount of charge that leaves the sphere.
The charge input withstands a lasting voltage up to 70 V (corresponding to a charge of 70 µC or 70,000 nC).
During electrostatic experiments where charge is transferred to the input by a ball or the like, there is no risk of damaging the apparatus – only if the input is connected directly to a high voltage supply, bad things will happen.
APPLICATION HINTS Grounding
In all experiment with static electricity or tiny cur­rents it is an advantage to have a well-defined zero for all voltages. Normally you chose “earth” as this zero point.
A correctly installed, grounded mains socket pro ­vides this zero potential. When you are using for in­stance our 6kV supply 3660.50 or the 500V supply
3655.60/65 with the appropriate power cord, the ground connection is available as the yellow/green socket on the back. This should be connected to the electrometer chassis – either the yellow/green socket to the left or the black socket to the right.
If you work on a table with a metal frame, you may with advantage ground connect this also (eventually by means of an alligator clip).
If you also ground a 4410.02 insulated rod or a ter­minal like 4350.10 you have an un-insulated ground to discharge diverse metal balls as well as the expe­rimenter.
Removing static charges
If you want to experiment with electric charges by means of for instance a 4415.00 metal plated ball on rod, the experiment may be disturbed by small char­ges on the rod.
If a radioactive alpha source with a Perspex handle is used with the 4410.35 ionization chamber you may again observe that the apparatus reacts to charges on the handle.
Likewise charges may be created on the plastic in­sulation of the sockets of the apparatus when a cord is pulled out. Such charges may seep in to the I in­put and show itself as a current that only very slowly falls towards zero – and that may be strong enough to throw the instrument off scale.
In the last case the problem is usually remedied by moistening a cotton swab with 96 % ethanol and in­serting it into the socket. After removing it again and the liquid has evaporated, the dial should go to rest at zero.
To remove charges on insulated rods, you may also use 96 % ethanol. You should be aware that there is a risk that Perspex may get frosted or even crackle by contact with ethanol. It should be mentioned that our 4415.00 metal plated ball on rod uses a Perspex rod – but we have in fact never observed problems with the use of ethanol in connection with our work with the apparatus.
EXPERIMENTS
The accessories mentioned are not part of 4410.13 but may be ordered separately.
The ionization chamber
The ionization chamber (4410.35) consists of a cen­ter electrode and a surrounding grid electrode. The center electrode is put into the I input such that the connector to the grid electrode is positioned to the left of the cabinet. Connect the grid electrode to a normal low voltage supply (for instance 3630.00) which has its other pole connected to the electro­meter’s ground (black socket to the right – or even­tually yellow/green socket to the right). The measu­ring range should be 5 or eventually 50 nA. Check that the socked has not been charged. While turning up the voltage to something like 10 V, you will ob­serve a momentary current as the grid is charged – but this should vanish fast, making the current ap­proximately zero.
Prepare an alpha source (like our 37 kBq Am-241 source) so the handle is not charged, and move it close to the ionization chamber. When it gets close enough, you will observe a current running between the electrodes – the air has been partially ionized.
Likewise, you can place a burning match near the io­nization chamber and gently blow the flame in the direction of the chamber. The free ions from the flame are enough to carry a measurable current.
A small signal diode as a photo detector
A solar cell is in fact just a diode with a special ap­pearance. An ordinary small signal diode in a glass housing (as for instance 1N4148) has in fact the same photoelectrical properties – the light sensitive area is just substantially smaller, giving rise to photocurrents in the picoampere range.
Set the range switch at 500 pA and connect the diode between the I input and zero ( ) with a couple of leads and alligator clips.
®
Loading...
+ 2 hidden pages