F&P Optiflow junior OPT312, Optiflow junior OPT314, Optiflow junior OPT316, Optiflow junior OPT318 Application Manual

Nasal CaNNula FittiNg guiDE
SELECT CANNULA SIZE
SIZe rANGe eNSUre GAP ArOUND PrONGS
Premature Size
Neonatal Size
Infant Size
Pediatric Size
THE FOLLOWING IS A SUGGESTED PROCEDURE FOR APPLYING THE OPTIFLOW JUNIOR NASAL CANNULA (OPT312, OPT314, OPT316, OPT318):
FITTING
Setup
1
Set up the gas supply, humidifier and breathing circuit.
Select Cannula Size
2
• The sizing chart on the following page can be used as a guide.
• If the infant can fit two sizes, select the smaller size.
Prepare Skin
3
Ensure infant’s face is clean and dry. Follow your hospital’s protocol for skin preparation.
Connect to Gas Source
4
Connect the cannula to the gas source.
Tip: Place hand close to nasal prongs to ensure that there is gas flow exiting the prongs.
Remove First Wigglepad Tabs
5
Without touching the adhesive, remove the first backing tabs from Wigglepads, leaving the second backing tabs in place.
2
5
6
Second backing tabs
First backing tabs
Place Cannula
6
• Once applied, the cannula bridge may move away from the septum as the cheeks relax. Refer to Placement Checks for correct positioning during use.
• Hold the ends of the pads and apply slight tension to the cannula.
• Position the prongs as far into the nares as possible such that the cannula bridge rests just underneath the septum.
• Position horizontally across the face and stick Wigglepads onto the cheeks.
Secure Cannula
7
Remove the second backing tabs and stick the Wigglepads onto the infant’s cheeks. Ensure the Wigglepads are well adhered to the face.
185048033 rev B © 2013 Fisher & Paykel Healthcare Limited
www.fphcare.com
7
Nasal CaNNula FittiNg guiDE
F&P OPTIFLOW JUNIOR
OPTIFLOW JUNIOR NASAL CANNULA
ITEM CODE APPROX WEIGHT (KG) SPARE
WIGGLEPADS
2 4 6 8 10 12 14 16 18 20 22
Premature Size
OPT312 OPT010
Neonatal Size
OPT314
OPT012
Infant Size
OPT316
Pediatric Size
OPT318
PLACEMENT CHECKS
Sensitive Septum
1
Check that the nasal cannula does not apply any pressure to the infant’s septum. Ensure a slight gap is visible between the cannula and the septum. Adjust cannula position if required.
Cheeky Check
2
• Gently squish the infant’s cheeks to check the prong placement in the nares.
• If prongs come out of the nares, adjust cannula further into the nares according to instructions. Repeat the cheeky check.
Tidy Tubing
3
Ensure that the baby is not lying on the tubing and that the flow path is maintained.
CANNULA REMOVAL/ADJUSTMENT
Always readjust cannula while patient is supine.
To temporarily remove or adjust the cannula on the Wigglepads:
• Place fingertip on the outside edge of Wigglepad.
• Gently peel the cannula pad away from the Wigglepad starting from the outside and working in towards the infant’s nose.
• To adjust, reposition prongs into nares and place cannula back onto Wigglepads.
2
WIGGLEPAD REMOVAL/ REPLACEMENT
Lift edge of the Wigglepad and using a wet cloth wipe the skin and underside of the Wigglepad while gently peeling Wigglepad from face. Refer to sizing chart for Wigglepad spares.
Note: Attach spare Wigglepad to cannula prior to applying to face.
SIZING CHART
IMPORTANT: Always refer to the user instructions supplied with the product for full set up instructions, warnings, contraindications and explanations.
Max. flow
8 L/min
Max. flow
8 L/min
Max. flow 20 L/min
Max. flow 25 L/min
www.fphcare.com
Loading...