This Service Manual contains descriptions, installation instructions, maintenance
instructions, troubleshooting, specifications and schematics for the Infratec™ 1241
ISW 5.xx. Service of Infratec™ 1241 should only be done by authorized personnel.
NOTE: This manual covers the Infratec 1241 Generation 2 and 3
only. For reference to the Infratec 1241 Generation 1 and Infratec
1256 use Service Manual p/n 10012177.
1.2Service Documents
This Service Manual is a part of the Service USB (if available) which contains the
complete support documentation for this product. The Service USB may contain
additional information in the form of animations for the purpose of illustrating and
explaining e.g. working principles, process flows and service and adjustment
procedures. See also section 8 Document References for an overview of User and
Service Documentation.
1.3Self-Service Support
The Customer Support Toolbox (CST) on FOSS EXTRAnet (www.foss.dk) contains
all available support information. Please visit the CST at regular intervals for latest
updates of documents and software.
1.4Contacting Customer Support
For support on this product, please contact the responsible Technical Support
engineer at FOSS or file a support request in the Global HelpDesk on FOSS
EXTRAnet.
1.5ESD Information
Parts of this instrument, e.g. PCBs, are sensitive to Electro Static Discharge (ESD).
All sensitive parts should be handled using ESD protection.
Follow these rules for effective ESD protection:
•Handle all ESD sensitive parts with an ESD wrist band connected to earth.
•Transport all ESD sensitive parts in ESD protected bags or boxes.
•Check your ESD protection at regular intervals to secure its function and quality.
Service Manual 1001 5015 / Rev. 41:1
Infratec™ 1241 Grain Analyzer
8000189a
The following ESD protection material is available from FOSS:
•436220 Field-ESD service kit including wrist band
•463238 Blue table mat with wrist strap and ground wire
•436246 Coiled wire, 360 cm with stretch-wrist band
•436261 Wrist-strap tester
Fig. 1:1 ESD equipment
1.6Precautions
The layout of safety symbols used in this manual is described below. Symbols are
used whenever a safety related message is written to warn the reader of potential
hazards.
Safety Symbols
Explanation of safety symbols used in this manual:
SymbolDescription
General hazard.
Electrical shock hazard.
Hot surface.
Heavy object.
Safety Terminology
Explanation of safety terms used in this manual.
TermDescription
WarningDanger to human safety.
CautionDanger to product performance/operation.
NoteImportant supplementary information.
1:2Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
2Technical Description
2.1History
The Infratec 1241 was first launched in the year 2000. It was then equipped with a
black & white display and a diskette drive as means of importing and exporting data
to the instrument. These instruments are now referred to as Generation 1.
In 2007 the Infratec 1241 was hardware and software wise redesigned. It was then
equipped with a color display and USB ports for importing and exporting data. The
"Basic concept" was also introduced which made it possible to limit the amount of
calibrations on the instrument and thus reduce the selling price. These instruments
are now referred to as Generation 2.
In 2010 the exterior was redesigned to match the EyeFoss image analyzer design.
Few internal parts where altered. These instruments are now referred to as
Generation 3.
ModelDisplayDataISW version
Generation 1Black & whiteDiskette3.xx
Generation 2ColorUSB5.xx
Generation 3ColorUSB5.xx
The Infratec 1256 is electronically and technology wise almost identical to the 1241.
Over the years different hardware and software features have been continuously
introduced, e.g. TCP/IP, calculated constituents. Many of these can be added to older
instruments. Upgrade of ISW is always free of charge.
The Basic Concept
Model# of allowed AppM's# of allowed parameters
FullManyMany
Basic 1OneThree
Basic 2TwoThree per AppM
Basic 3OneSix
Basic 4TwoFour per AppM
Basic 5FiveFour per AppM
Basic 6SixFour per AppM
Service Manual 1001 5015 / Rev. 42:1
Infratec™ 1241 Grain Analyzer
Related Information
Service personnel should be familiar with the following documentation:
•User Manual; for function descriptions, maintenance and operating instructions
•Spare Parts Manual; for spare part illustrations and spare part numbers
•Quick Guide; for a quick start instruction
Also see chapter 8 Document References.
Abbreviations
Explanation of abbreviations used in this manual.
ADCAnalog to Digital Converter
DIBDisplay Interconnection Board
DSPDigital Signal Processing
EIDElectronic ID
GAGrain Analyzer
I2CStandardised serial interface (Inter Integrated Communication)
I/OInput/Output
ISWInstrument Software
LIMSLaboratory Information and Management System
MUMeasuring Unit
OSOperating System
PCBPrinted Circuit Board
PSMPower Supply Module
RTOSReal Time Operating System
STMSample Transport Module, optional
TWMTest Weight Module, optional
2:2Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
2.2Component Descriptions
Compatibility with previous version
ModuleCompatibleComments
PSUYesAll versions
DetectorYesAll versions
MonochromatorYesAll versions
MUYesAll versions
TWMYesAll versions (TN 1234)
STMYes
LCD DisplayNoSee Spare Parts Manual
PC module
DSP
PC- board
Compact Flash
Floppy disk
No
No
No
No/Yes
No
Requires OS vers 5.01 or higher
Requires OS vers 5.01 or higher
Dependency. Requires ISW 5.xx
Disabled in BIOS, USB Floppy disk support
Cable StalkNo
InterfaceNoUSB
ChassisYes
System Overview
Fig. 2:1
Service Manual 1001 5015 / Rev. 42:3
Infratec™ 1241 Grain Analyzer
2.2.1PC-Module
The PC-module is a PC104 embedded hardware platform with QNX real-time
operating system (RTOS). PC104 is a popular standardized form-factor for small
computing modules typically used in embedded system. A system composed of PC/
104 modules is often referred to as a "stack".
The PC module "stack" assembly consists of:
•Mounting plate
•Processor board, that acts as a controller for the peripheral components
•DSP board, that acts as an extension Processor board and is also used as an I/O
system
Communication between the PC-module and the instrument modules is routed via
the DSP board I
2
C bus, which is a simple standard 2-wire connection where the DSP
acts as the master device.
NOTE: Due to the QNX software platform, USB plug & play is not
supported. An USB-floppy drive solution may however work.
Fig. 2:2 PC-Module
2:4Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
2.2.2Processor Board
The processor board is a single board processor (SBC) with an onboard CPU
processor. Processor board acts as a controller for the peripheral components. The
processor board communicates with the DSP board via the PC/104 bus.
The processor board has three main interfaces:
•DSP board PC/104 bus
•TFT LCD display
•Connectors to chassis Interface panel, providing access to external keyboard,
USB, comm. ports, LAN etc
The Processor Board has replaceable external CMOS battery.
Fig. 2:4 Processor Board
2.2.3DSP Board
The DSP module is accessed from the ISW via the PC/104 bus interface (P3) and acts
as slave under the ISW.
The DSP software acts as an extension of th e ISW and is also used as an I/O system.
The PC-module makes it possible for the ISW to have control over the different HW
modules.
There is no ROM memory on the DSP Board. For this reason, the DSP software is
uploaded from the processor board by ISW during start-up. The ISW installation
package contains and installs the latest and matching DSP software version.
The DSP has an I2C bus interface, implemented as a SW driver in the DSP software
and is used for communication between devices inside the instrument. I
wire serial bus, as shown in
Fig. 2:5.
Fig. 2:5
2
C is a two-
There are a number of on board LEDs on the DSP Board. These LEDs provide
information as to the status of the DSP Board.
LEDFunctionColor
D3Power OnGreen, constant
D4DSP-ISW communicationRed, flashing when active
D1DSP runningRed, constant after boot-up
D2
2:6Service Manual 1001 5015 / Rev. 4
Monochromator motor
direction
Red, blinking
Infratec™ 1241 Grain Analyzer
Fig. 2:6 DSP Board
P1KeypadP13Detector
P3PC/104 BusP14I2C Bus
P6Power SupplyP15I2C Bus
P7Inverter Interface BoardD1DSP running (red LED)
P8MonochromatorD2Monochromator (red LED)
P9MonochromatorD3Power On (green LED)
P12MonochromatorD4I2C communication (red LED)
DSP Board
FromDesignationTo
P1Keypad overlay
P3PC/104 Bus InterfaceProcessor Board
P6Power SupplyPower Supply Board (P7)
P7Backlight PowerDisplay Interconnection Board
P8Encoder ConnectorMonochromator
P9Grating Motor ConnectorMonochromator
P12Shutter Board ConnectorMonochromator
P13Detector SignalDetector
P14I2C BusMeasuring Unit
Service Manual 1001 5015 / Rev. 42:7
Infratec™ 1241 Grain Analyzer
P15I2C BusPower Supply Board
2.2.4CompactFlash Disk
Type II CompactFlash™ Memory
The CompactFlash memory is preinstalled with latest OS version and ISW software.
As spare part it is available as Basic 1 configuration only.
The Flash Disk is an electronic memory mounted in a slot on the Processor Board.
Due to market conditions you may find different sizes in the instruments over the
years. There is no problem upgrading the Flash Disk to a larger size. The disk is
partitioned in three parts, 5 Mb for the operative system, 5 Mb for ISW. The
remaining space is for data.
The Basic Concept
Model# of allowed AppM's# of allowed parameters
FullManyMany
Basic 1OneThree
Basic 2TwoThree per AppM
Basic 3OneSix
Basic 4TwoFour per AppM
Basic 5FiveFour per AppM
Basic 6SixFour per AppM
2.2.5Monochromator
The Monochromator is developed and manufactured by FOSS. It is to be considered
a closed, factory calibrated box. There is under no circumstance any reason to open
it for maintenance, inspection or other purpose. On the contrary an opened
monochromator will not be covered by warranty, neither will it be accepted in the
exchange part system. The monochromator is the only spare part in the Infratec that
is enrolled on the spare part exchange system.
In the Infratec the monochromation of the "white light" from the lamp is done before
the sample. This technology is called pre-dispersive monochromation. This is
optimal when using transmission of light through the sample, e.g. for grain analysis.
A disadvantage of this technology is that the system becomes very sensitive to stray
light from the environment, e.g. sunlight and fluorescent light from the ceiling.
In the Infratec the light from the light source (the instrument lamp) is transported in
an optic fiber bundle into the heart of the monochromator. This fiber bundle consists
of thousands of randomized fibers. A lamp change will therefore not affect the output
of the fiber bundle and thus makes it un-necessary to linearize after the lamp is
change. The fibre optic cables are sensitive to sharp bends and should be treated with
care.
Inside the monochromator the lamp light is split up (monochromated) in a diffraction
grating. The grating is turned by a motor so that only one wavelength is exiting the
monochromator at any given time. The grating is producing a spectrum covering the
wavelength range from 570 to 1100 nm. The bandwidth is 7 nm.
At the out put there is a second fiber bundle to transport the monochrome light over
to the right side of the instrument where the sample cell is located.
2:8Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
Before the monochromator output there is a shutter with 3 positions, Open, Closed
and Filter. The closed position is used to block the light during the offset
measurement which is done prior to scanning the sample. Open position is used
during the sample scanning and Filter position is used during performance tests to
determine the wavelength accuracy compared to the factory status.
The monochromator is supplied with an EEPROM which stores the O- and Pconstants. These constants are obtained from the factory calibration and cannot be
altered or re-calibrated in the field. The EEPROM resides on a PCB underneath the
monochromator base plate. This PCB is therefore a unique part dedicated for each
monochromator and cannot be changed.
NOTE: The Monochromator is sealed and it is absolutely
forbidden to open it by other than FOSS personnel.
Fig. 2:7
1Lamp fiber4Output fiber
2Grating5Motor/Encoder
3Shutter6PCB with EEPROM
Service Manual 1001 5015 / Rev. 42:9
Infratec™ 1241 Grain Analyzer
2.2.6Lamp Module
The Infratec 1241 light source solution is accommodated in a heat sink aluminium
housing. It has no internal electronics other than the power supply cable and light
bulb. The module includes:
•Aluminium housing as heat sink
•Lamp Holder
•Lamp
•Aperture (if applicable)
•Optic Fibre connection
Lamp holder
Three versions of lamp holders have been manufactured for Infratec 1241. The latest
version, introduced 2008-10-17, has socket screw terminals to secure firm contact.
See picture below. Previous versions are exposed to oxidation coating causing bad
contact between lamp and terminals. It is recommended to replace the previous
versions with P/N 10013351 which is the latest version (see picture below).
Infratec 1241, InfraXact and FoodScan analysers all have the same lamp holder.
Fig. 2:8 Lamp Holder
Lamp source
The halogen lamp is FOSS custom manufactured and can not be substituted with
regular lamps. Use P/N 60023564 to order.
Aperture
Exchangeable apertures are used to adjust the amount of light admitted to the
monochromator inlet fibre. Monochromator replacement requires that light intensity
is checked due to dispersion variation in the Monochromator Optical Fiber.
Exchangeable apertures are available in fixed openings of 5, 5.5, 6, 6.5, 7 and 7.5 mm
diameter. An Aperture Kit (P/N 60023539) with a full range of apertures is supplied
with replacement monochromators. Please refer to TN1355 for replacement
instructions.
2:10Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
2.2.7Detector
The detector converts light energy input to digital data output. It is constructed in a
sealed and shielded box with a window for light input and a connector for digital
input and output. The analog electronics inside the box are galvanic isolated (opto
coupled) from all I/O signals.
The detector and the DSP communicate with each other in two ways. The I2C bus is
used to transfer commands from the DSP to the detector, while a high speed serial
channel is used to transfer data from the detector's A/D converter.
The detector module has an I2C expander on board that is accessed from the DSP and
is used to set the detector gain to values specified by the ISW. It also has a
temperature sensor from which ISW reads current temperature inside the detector
and it's accessed via the I
An E2PROM onboard the detector module contains the detector constants (e.g. serial
number).
2
C interface as well.
Fig. 2:9 Detector Block Diagram
1Front end optics6I/O expander for control signals
2Large area silicon detector diode7Temperature sensor
Pre-amplifier stage 1 with selecta-
3
ble gain
Pre-amplifier stage 2 with selecta-
4
ble gain
516 bit 100 ksamp/sec A/D-converter
Service Manual 1001 5015 / Rev. 42:11
EEPROM for calibration data stor-
8
age
9Isolated DC/DC converter
Infratec™ 1241 Grain Analyzer
2.2.8Power Supply
The Power Supply takes the 24V AC power it receives from the transformer and
converts it to a variety of DC voltages (+ 5.5V, +5V, +12V, -12V, and +10.5V).
These voltages are used by a variety of instrument components, including the PC
module, measuring unit, lamp, display, and optional equipment (TWM, STM etc.).
The Power Supply receives command signals from the DSP Board via the I2C bus.
2
C decoder processes these signals and controls the lamp and display
The I
accordingly. For example, the lamp and display backlight can be placed in power
saving stand-by modes, the display itself can be extinguished during certain
sequences such as the initial instrument boot-up procedure.
Fig. 2:10 Power Supply Block Diagram
2:12Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
Fig. 2:11 Power Supply Board
J1LCD BacklightP8Measuring Unit Power
J3LCDF1Fuse 6.3 AT M
J6I2C BusF2Fuse 1.25 AT T
J7Remote I/O Power (optional)F3Fuse 4.0 AT
P1From TransformerF4Fuse 500 mAT
P5LampF6Fuse 4.0 AT
P7DSP Board Power
Service Manual 1001 5015 / Rev. 42:13
Infratec™ 1241 Grain Analyzer
2.2.9LCD Display
The display is a 6.4 inch color active matrix LCD module, VGA 640 x 480 pixel
resolution, incorporating amorphous silicon TFT (Thin Film Transistor). It is
composed of a color TFT-LCD panel, driver ICs, control circuits, power circuits and
a backlight unit.
Other features include a wide viewing angle with the best angle at 6 o'clock, high
contrast, anti glare polarizing filter, normally white display mode that produces
natural colors.
The display needs voltage from three sources to be operational:
•5VDC from the Processor Board (the display on/off signal)
•Display backlight from the Display Interconnection Board
•Contrast voltage
The LCD Display is not repairable, it has to be replaced if a failure occurs.
The keypad is a passive matrix (7´4) with an LED power indicator. It is sometimes
here referred to as the keypad overlay.
2:14Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
2.2.10 Display Interconnection Board
The Display Interconnection Board is connecte d between the Power Supply Board,
DSP Board and the Display. It functions as a switch whereby ISW controls LCD
Backlight and LCD Display by on/off function.
Fig. 2:12 Inverter Interface Board
JP1I2C BusJP2LCD Display
The Display Interconnection Board is powered by DSP Board P7 sock et.
NOTE: Backlight. Actually 600 V AC, supplied by a DC/AC
converter.
The Electronic ID is installed as factory standard and enables the extended
wavelength range needed for predictions in the area 570 nm - 850 nm.
The effective wavelength range of the instrument will be 570 nm-1098 nm. This
range will be activated when using an Application Model with extended range.
Service Manual 1001 5015 / Rev. 42:15
Infratec™ 1241 Grain Analyzer
2.2.12 TWM (Optional)
Test Weight Unit (TWU)
The device which performs the actual weighing of the sample. It consists of the
following basic items;
•A load cell (balance).
•Control electronics PCB. Referred to as the TWM PCB.
•A wiper arm to level of the cuvette.
•Level sensor to detect when the cuvette is full.
•A shutter to allow emptying of the cuvette after measurement.
Fig. 2:14 Test Weight Module
See the Spare Part manual for more detailed information on available spare parts.
Test Weight Module (TWM)
TWM is the terminology used for the entire assembly, including the TWU and
chassis. This module is an optional device that can be fitted to any existing Infratec
1241. To measure the test weight of a sample the used Application model must have
support for the TWM.
The Infratec software (ISW) will during boot-up detect if a TWM is electrically
connected to the instrument.
The TWM is connected to the Infratec through the TWM power/ communication
cable and the drawer sensor cable. The TWM power/ communication cable can be
connected in either of the two optional contacts on the Measuring Unit. The TWM
then receives commands from the ISW. The weight is mathematically converted to a
density and the completed result is presented on the display as a bulk density.
The instrument handles both metric units and imperial units making it possible to
present test weight in both kg/hl and lbs/bu. More information can be found in the
User Manual for Infratec 1241 Modules, p/n 60043623.
2:16Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
The TWU is designed to receive sample from the Infratec in small sub-samples into
a measurement cuvette. The surface of the measurement cuvette is then skimmed flat
with a wiper arm and the measurement taken.
The TWU then discharges the sample into the sample drawer. The commands to
operate the TWU originate from the DSP board. All commands come over the I2C
serial bus. The I2C is a 2-wire serial bus consisting of data and clock. There are two
buses of this type on the TWU PCB, one to the DSP and the other to the EEPROM.
The EEPROM fitted on the TWM PCB is provided to hold the mass calibration
constants. The control of the wiper arm is done with a step motor with an encoder.
At start-up the wiper arm moves between the physical end positions. Then a couple
of pulses are taken away in each end. The calculated positions are HOME and FAR
positions.
Below is a description of the weighing process for Infratec 1241 Test Weight
Module:
1.Activate the Test Weight Unit.
Check in the User menu under Application models/Test Weight/Setup/General
settings, that "Enabled" is set to ON.
2.Choose an Application model that include test weighing. It's important that the
twm.ini file is included. It is also possible to create a new Application model. Go
into the User menu and choose Application models/Test Weight/Setup/Test
Weight Models.
3.Pour the sample into the hopper.
4.Press the analyse button (or use F´9 on the external keyboard)
The empty weight of the test cell is first measured. Then the conveyor in the
Infratec Measuring Unit deposits sub-samples in the sample cell, where NIT
analysis takes place. After analysis of each sub-sample, the bottom shutter of the
sample cell opens, depositing the grain in the cuvette of the Test Weight Unit. A
level sensor detects when the TWU cuvette has been filled and analysis of the
remaining NIT sub-samples are temporarily halted.
A wiper arm levels off the test cell to ensure proper cell packing and consistent
sample volume. The weight of the grain in the test cell is then determined as the
difference between the full measurement and the empty measurement after
which the Infratec makes all necessary calculations and adjustments. The shutter
at the bottom of the test cell is then opened, allowing the grain to fall into the
sample collection drawer, and the analysis of the remaining NIT subsamples is
resumed. The grain from these sub-samples simply passes directly through the
TWU cuvette).
5.The NIT results, along with test weight, are then presented on the LCD display
of the Infratec.
Service Manual 1001 5015 / Rev. 42:17
Infratec™ 1241 Grain Analyzer
2.2.13 STM (Optional)
The Sample Transport Module is an optional device used for samples that are not fit
to be taken through the Measuring Unit sample cell, e.g. moist samples or even
liquids. Such samples may be green-malt, beer, boiled rice etc. Several different
cuvette models and sizes are available for different kinds of sample types.
The STM consist hardware wise of an Elevator Unit which includes a motor that
drives the sample cuvette down in front of the detector and electronics to
communicate commands to the DSP. The position of the cuvette is detected by two
magnet sensors in the elevator unit.
Electrically the STM is connected to either of the two ports on the measuring Unit.
For installation and usage see the User Manual for Infratec 1241 Modules, p/n
60043623
2:18Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
2.2.14 Remote I/O (Optional)
Introduction
This document describes the I/O Interface and specifies the physical interfaces used
for connecting it to an Infratec 1241 or Infratec 1256. The Infratec is assumed to be
prepared for this hardware such that the back plate has got place for a High Density
DSUB15 connector for this purpose.
Connections
The I/O Board has three different interfaces:
•The 5-pin I2C/Power interface. The cable connecting to this interface has a 4way , 2-wire I²C connector that sh ould be connected to an available I²C connector
inside the Infratec 1241/1256. The 4-way, 3-wire power connector should be
connected to an available power connector inside the Infratec 1241/1256.
•The 3-pin sensor interface could be used for connecting a "trigger"-sensor inside
the instrument.
•The 15-pin High Density DSUB15 is the interface to the outside world and is
mounted on the I/O Board and should be fastened at the back of the instrument.
3IO7 - TriggerTrigger input used to start analysis
4IO6 - ReservedNot used
5GND
6IO2 - AM Selection 0
7IO3 - AM Selection 1
8IO4 - AM Selection 2
9IO5 - AM Selection 3
10GNDGND to external sensor
1112V Output12V supply to external sensor
12NCNC
Indicates that the Infratec 1241/1256 is busy
analysing
Indicates that the Infratec 1241/1256 is connected
Bit 0 of the 4 bit nibble used for selecting the
application model
Bit 1 of the 4 bit nibble used for selecting the
application model
Bit 2 of the 4 bit nibble used for selecting the
application model
Bit 3 of the 4 bit nibble used for selecting the
application model
13NCNC
14NCNC
1510-24V InputExternal power 10-24V for opto-couplers
2:20Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
Hardware
The hardware of the I/O Board is basically an I²C controlled 8 bit I/O expander where
2 pins are allocated as outputs (IO0, IO1) and 6 pins as inputs (IO2-IO7). All I/O:s
are protected with opto-couplers and the external part of the opto-couplers are
powered with 12/24V from the outside.
One pin on the connector is connected to the Infratec 1241/1256 12V supply to
enable connection of an external sensor without extra hardware.
Software
The functionalities for the I/O Board require at least ISW 3.40. In this version there
are however some bugs regarding the statuses 'Busy' (not reported correct at all
times) and 'Connected' (not enabled). This will be fixed in ISW 3.41 that will be
released in 2006.
Fig. 2:16
Service Manual 1001 5015 / Rev. 42:21
Infratec™ 1241 Grain Analyzer
2.3Software Description
The operative system used in the instrument is QNX which is a commercial Unixlike real-time operating system (RTOS), aimed primarily at the embedded systems
market.
The DSP software acts as an extension of the instrument software (ISW) and is
located as an IO module to the PC board. The DSP software is accessed via the
standard PC/104 bus on the PC board. The DSP module makes it possible for the
ISW to have control over the different components of the instrument. This document
describes the interface (IF 1) between the ISW (SW 1) and the DSP software (SW 2).
Fig. 2:17
SW 1: Instrument software system (ISW) (10008304)
PC-software located in the instrument. It contains functions for analysing samples,
predicting results, creating scan files for mathematical treatment.
SW 2: DSP software system (10008494 & 10014320)
Software located in the DSP. Containing functions for Detector, Monochromator,
MU, TWM and PSM communication.
SW 3: Monochromator microcontroller software
PIC microcontroller software to control the monochromator shutter functions and the
E2PROM access on the monochromator board
The O and P constants are stored together with the monochromator module serial
number in an E2PROM onboard the monochromator PCB.
SW 4: Measuring Unit (MU) microcontroller software
PIC microcontroller software to control the MU functions.
SW 5: Test Weight Module (TWM) microcontroller software
PIC microcontroller software to control the TWM functions.
SW 6: Electronic ID (EID) microcontroller software
PIC microcontroller software to control the EID functions.
2:22Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
SW 7: Sample Transport Module (STM) microcontroller software
PIC microcontroller software to control the STM functions.
IF 1: Instrument Software/DSP gen II (10014320)
Interface between the Instrument software and the DSP software.
IF 2: DSP/Monochromator
I2C interface between the DSP and the monochromator microcontroller.
IF 3: DSP/Detector Module
I2C interface between the DSP and the detector module E2PROM, I2C expander and
temperature sensor.
IF 4: DSP/MU
I2C interface between the DSP and the MU microcontroller and temperature sensor.
IF 5: DSP/PSM
I2C interface between the DSP and the PSM I2C expander.
The signals used to adjust the contrast of the display, the display on/off function, to
set the display backlight on/off and the monochromator lamp in standby mode is sent
to the power supply module via the i2C bus interface. This connector is identical to
the connector interfacing the STM.
IF 6: DSP/TWM
I2C interface between the DSP and the TWM microcontroller via the MU board.
IF 7: DSP/Hopper Temp.
I2C interface between the DSP and the hopper temperature sensor via the MU board.
IF 8: DSP/EID
I2C interface between the DSP and the EID microcontroller via the MU board.
IF 9: DSP/STM
I2C interface between the DSP and the STM microcontroller via the MU board.
IF 10: DSP/Remote IO
I2C interface between the DSP and the IO expander circuit on the remote IO board
via the MU board.
Service Manual 1001 5015 / Rev. 42:23
Infratec™ 1241 Grain Analyzer
2:24Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
FOSS TECATOR
3Installation
3.1General
See User Manual - Infratec™ 1241 Grain Analyzer.
Information about installation of Sample Transport Module and Flour Module can be
found in User Manual - Infratec™ 1241 Modules, P/N 60043623.
Site Preparation
The site preparation is intended to be carried out prior to the installation of the
instrument at the customer site. The site preparation documents can be found at
FOSS Intranet – Customer Support Toolbox.
Filling Out Installation Documents
Each instrument is delivered with an Owners Guide binder where you will find all
installation documents to be filled out.
3.2Installation of Test Weight Module on
Infratec 1241 Generation 2
The installation of the Test Weight Module occurs in the three steps:
1.Preparation of the Infratec for assembly
2.Preparation of the Test Weight Module for assembly
3.Final ass embly of Infratec and Test Weight Module into a complete assembly
3.2.1Preparation of the Infratec for assembly
To prepare the Infratec for assembly to the Test Weight Module, perform the
following steps:
1.Remove the screw on the back of the instrument's lower right hand side using a
3 mm (2,5 mm for later models) allen wrench, see
Fig. 3:1 below.
Fig. 3:1 Screw on backside of instrument
Service Manual 1001 5015 / Rev. 43:1
Infratec™ 1241 Grain Analyzer
3
2
1
1241298a
2.Disconnect the collection drawer sensor cable connector (1) from the Measuring
Unit, see
Fig. 3:2 below.
Fig. 3:2 Collection drawer sensor cable,
Measuring Unit Gen. I (left) and Measuring Unit Gen. II (right)
The next step requires that the instrument be tilted on its left side. Before this can be
done, there must be adequate workbench space and a soft padding (or equivalent)
must be laid where the instrument will be tilted on its side.
Warning
The tilting of the instrument must be performed by two persons.
3.Gently tilt the instrument and lay it to rest on its left side.
4.Remove the sample collec tion drawer from the instrument.
Caution
Open the right-hand door carefully when the instrument is in
tilted position. The door may brake if incautios.
5.Lift the hopper section to its fully upright position and then open the right-hand
door, see
Fig. 3:3 below.
Fig. 3:3 Right hand door opened
3:2Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
6.Remove the bottom pla t e by loosening the four screws with
a 3 mm allen wrench, see Fig. 3:4 below.
Fig. 3:4 Bottom plate
7.Remove the collection drawer guides (two on each side ). This is done by
unscrewing the screws with a 4 mm allen wrench, see
Fig. 3:5 below. Remount
four short screws in the holes on the left side (the instrument side). This is to
prevent dirt from getting into the Infratec.
Fig. 3:5 Collection drawer guides
8.Close the right-hand door and lower the hopper section to its fully closed
position.
Service Manual 1001 5015 / Rev. 43:3
Infratec™ 1241 Grain Analyzer
3.2.2Preparation of the Test Weight Module for assembly
To prepare the Test Weight Module for assembly, perform the following:
1.Place the Test Weight Module on a workbench. The workbench should be as
level as possible. The workbench should also be as free from vibrations as
possible.
2.Adjust the four feet under the plinth, see Fig. 3:6 below. Use a level guide to
level the Test Weight Module.
Fig. 3:6 Adjustment of feet
3.Loosen the three letter screws on the Front Transport Securing Device, see
Fig. 3:7 below. Store them in the extra holes on the Front Transport Securing
Device.
Fig. 3:7 Letter screws on Front Transport Securing Device
3:4Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
4.Loosen the two letter screws on the Back Transport Securing Device and grip
the Test Weight Module at the same time so that it will not fall, see
below. Use the two letter screws to fasten the Front Transport Securing Device
on the block situated behind the Test Weight Module.
Fig. 3:8 Screws on Back Transport Securing Device
Fig. 3:8
Warning
Never perform step 5 with the Back Transport Securing Device
tightened.
5.Fasten the two allen screws hard with a 4 mm allen wrench, see Fig. 3:9 below.
Grip the Test Weight Module at the same time so that you can feel that it is
getting positioned.
Fig. 3:9 Allen screws
6.Use the enclosed labels to cover the two holes from the letter screws.
Service Manual 1001 5015 / Rev. 43:5
Infratec™ 1241 Grain Analyzer
3.2.3Final assembly of Infratec and Test Weight Module
into a complete assembly
The Infratec and Test Weight Module should be assembled into a complete assembly
as follows:
1.Tilt the Test Weight Module so that it can be introduced in the Infratec.
Caution
Extreme caution should be used when mounting the Test Weight
Module to the Infratec. The test cell of the Test Weight Module is
a very delicate and sensitive item. There should be no contact
whatsoever with this item during the mounting process.
Caution
Open the right-hand door carefully when the instrument is in
tilted position. The door may brake if incautios.
2.Lift the hopper section to its fully upright position and then open the right-hand
door.
3.Mount the screw on the back side of the Infratec using a 3 mm (2,5 mm for later
models) allen wrench, see
Fig. 3:10 Screw on back side of instrument
4.Mount the two screws located on either side of the collection drawer opening
using a 3 mm allen wrench, see
Fig. 3:10 below.
Fig. 3:11 below.
Fig. 3:11 Screws at collection drawer opening
3:6Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
5.Fasten the guiding bracket on the right side of the Test Weight Module. Use the
holes market with arrows in
bracket with two short screws using a 3 mm allen wrench.
Fig. 3:12 Guiding bracket hole
Fig. 3:12 below and Fig. 3:13 on page 3:7. Attach
Fig. 3:13 Guiding bracket hole
Caution
Steps 6-8 are extremly important for the instrument to be EMC
approved.
6.To earth the TWM at least one of the holes marked with arrows in Fig. 3:14
below, has to be scraped. Use an 8 mm plane countersink.
7.Plane countersink the corresponding hole on the TWM.
Service Manual 1001 5015 / Rev. 43:7
Infratec™ 1241 Grain Analyzer
8.Position the cover plate assembly in the collection drawer opening of the
Infratec, see
Fig. 3:14 below. Fasten the cover plate assembly with two long
screws using a 3 mm allen wrench.
Note: Fasten a toothed plate connector in the plane countersinked hole.
Fig. 3:14 Scrape at least one of these holes
9.Close the right-hand door and lower the hopper section to its fully closed
position. Tilt the complete instrument back to upright position.
10. Mount the block ment to hold the Front Transport Securing device behind the
TWM. Dry off the space behind the TWM with alcohol to make the selfadhering tape adhere better, see
Fig. 3:15 below.
Fig. 3:15 Self-adhering tape
3:8Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
3
2
1
4
1241298a
1. Drawer sensor cable
2. Optional
3. Optional
4. Hood temp sensor cable
11. Connect the drawer sensor cable and the Test Weight Module power cable to the
Measuring Unit, see
Fig. 3:16 below. The Test Weight Module power cable can
be connected in either of the two optional contacts 2 or 3.
Fig. 3:16 Connections on the Measuring Unit,
Measuring Unit Gen. I (left) and Measuring Unit Gen. II (right)
12. Install the Test Weight ini-file. First of all put the diskette into the diskette holder
of the instrument. From the User Menu, choose Read disk. The Test Weight inifile will be imported to the instrument, see User Manual.
13. Replace the sample collection drawer.
Service Manual 1001 5015 / Rev. 43:9
Infratec™ 1241 Grain Analyzer
3.3Installation of Test Weight Module on
Infratec 1241 Generation 3
1.Remove the two plastic slides for the drawer. They should be moved to the
TWM later.
Fig. 3:17
2.Remove the metal plate covering the drawer sensor by loosen the four sc rews.
Fig. 3:18
3:10Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
Fig. 3:19
3.Turn the Infratec on the left side. Place some soft material on the table to prevent
scratches.
Fig. 3:20
Service Manual 1001 5015 / Rev. 43:11
4.Remove three of the four feet.
Infratec™ 1241 Grain Analyzer
Fig. 3:21
5.Remove the two beams 1 and 2 by loosen the screws.
Fig. 3:22
3:12Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
6.Remove the bottom plate by removing the three screws 1.
Keep the screws because they are needed later.
Fig. 3:23
7.Disconnect the drawer sensor and remove the sensor plate by removing the two
screws 1.
Fig. 3:24
Service Manual 1001 5015 / Rev. 43:13
Infratec™ 1241 Grain Analyzer
8.Prepare the grain deflector (p/n 60039237) by putting the adhesive (p/n
10014399) on the outside of the right part of the deflector (see arrow below). It
is used to hold it in place before it is fasten by the screws later in the assembly.
9.Mount the grain deflector at the front by using four screws (p/n 55320017).
Just fix it don't fasten it yet.
Fig. 3:25
10. Put together the front cover (p/n 60034842) and cover bracket (p/n 60039398)
by using the four screws (p/n 15320098).
Fig. 3:26
3:14Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
11. Fasten the cover plate to the deflector and adjust. Then fasten the screws holding
the deflector.
Fig. 3:27
12. Use the three screws that were holding the bottom plate to cover the screw holes
(one screw is not shown below).
Fig. 3:28
Service Manual 1001 5015 / Rev. 43:15
Infratec™ 1241 Grain Analyzer
13. Prepare the Test Weight Unit by mounting the two plastic slides from the
Infratec.
Fig. 3:29
14. Transfer the drawer sensor from the holder that you have taken out from the
Infratec (removed in step 7) to the holder for TWM (p/n 60039395). Put the
sensor as much to the left as possible. Also put the rubber sealing (p/n 15420126)
onto the left wing (see arrow below).
Fig. 3:30
3:16Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
15. Put in the holder into the TWM. Note that the cable from the drawer sensor
should be placed between the rubber sealing and the chassis. Fasten with three
screws (p/n 55320017).
Fig. 3:31
16. Loosen the three letter screws 1 on the Front Transport Securing Device. Store
them in the extra holes on the Front Transport Securing Device.
Fig. 3:32
Service Manual 1001 5015 / Rev. 43:17
Infratec™ 1241 Grain Analyzer
17. Loosen the two letter screws 1 on the Back Transport Securing Device and grip
the Test Weight Module at the same time so that it will not fall. Store them as
well in the holes in the Front Transport Securing Device.
Fig. 3:33
18. Fasten the two allen screws 1 hard with a 4 mm allen wrench. Grip the Test
Weight Module at the same time so that you can feel that it is getting centered.
Fig. 3:34
3:18Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
19. Use the enclosed labels 1 (p/n 10009083) to cover the two holes from the letter
screws.
Fig. 3:35
20. Take the Test Weight Unit and connect it to the Infratec. Place the balance
behind the deflector so that the level sensors don't touch the deflector.
Fig. 3:36
Service Manual 1001 5015 / Rev. 43:19
Infratec™ 1241 Grain Analyzer
21. Secure the TWM to the Infratec by using eight screws 1 (p/n 55320017) (six
screws are not shown below).
Fig. 3:37
22. Refit the two beams 1 and 2 and the three feet. Place the Infratec into standing
position.
Fig. 3:38
3:20Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
23. Connect the TWM to the measuring unit. Make sure that no cables are touching
the balance.
Fig. 3:39
24. Swich on the Infratec.
25. Enable the TWM function of the Infratec 1241 according to User Manual
Infratec 1241 Modules (p/n 60043623).
3.4Installation of Sample Transport Module
Software Options
In the Service Menu there is an option in “3 Manual Tests” called “7 STM” which
offers three options.
1.Position settings
2.Service
3.Exercise
Position settings can be used to run the cuvette holder up and down manually. It can
also be used to step between the sub sample positions. It is easy to use and stops when
it comes to the end position.
Service, displays the position sensors value and gives an option of running the motor
Up/ Down/ Off.
Note: It will not stop in the end position so be very careful.
Exercise, will only continuously run the cuvette holder up and down, no real use for
this option.
Service Manual 1001 5015 / Rev. 43:21
Infratec™ 1241 Grain Analyzer
3.5Mosaic Connection
Assigning Instrument type to Network
1.The instrument type must be assigned to the Network. Right click on Network
Name. Select “Assign instrument types” from the context menu.
Fig. 3:40
2.Highlight the instrument type that should be assigned and click “OK”.
Fig. 3:41
3:22Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
Add New Group to Instrument type
3.Add an instrument group to the instrument type by right clicking on instrument
type, select New Group from the context menu. In the example a New Group is
added to Infratec 1241.
Fig. 3:42
4.New instrument group will show up:
Fig. 3:43
Service Manual 1001 5015 / Rev. 43:23
Infratec™ 1241 Grain Analyzer
5.Right click on the New instrument group and select Properties from the context
menu.
Fig. 3:44
6.Give the Instrument group a new Name, for example customer name, site,
factory or similar. Click OK.
Fig. 3:45
3:24Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
7.The new Instrument Group Name will appear under the instrument type.
Fig. 3:46
Connect Infratec 1241 to the Mosaic Server
Fig. 3:47
Service Manual 1001 5015 / Rev. 43:25
8.In analyze mode click on the Menu button.
Select 1 Main Menu.
Fig. 3:48
Infratec™ 1241 Grain Analyzer
9.Enter password. The default password is “123456”.
Fig. 3:49
3:26Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
10. Select “5 Instrument Settings”.
11. Select “2 Setup”.
Fig. 3:50
12. Select “1 Communication”.
Fig. 3:51
Fig. 3:52
Service Manual 1001 5015 / Rev. 43:27
13. Select “2 LAN Settings”.
Infratec™ 1241 Grain Analyzer
Fig. 3:53
The customer IT department will assign if Static or Dynamic IP should be used. The
IT department should also supply all information needed.
14. If the instrument will have a Static IP select “1 Static IP”.
If using Dynamic IP please select “2 Dynamic IP”.
Fig. 3:54
3:28Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
15. Static IP: Enter at least the Instrument IP No., Subnet Mask and Gateway which
the customer IT department has assigned for the instrument. Press Activate.
Fig. 3:55
16. Dynamic IP: Enter Instrument Name assign by Customer IT department. Press
Activate.
Fig. 3:56
Service Manual 1001 5015 / Rev. 43:29
Infratec™ 1241 Grain Analyzer
17. Select “3 Network Information” to check the LAN settings is correct in Infratec.
Fig. 3:57
18. Select “2 LAN Settings”.
Fig. 3:58
19. Select “4 Mosaic Settings” to enter the Mosaic server settings.
Fig. 3:59
3:30Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
20. Type in the IP number to the Mosaic Server in Mosaic Host field. Remember
that Infratec software cannot resolve DNS name, therefore use IP number to
Mosaic server. Enter the right port for the Infratec to connect. Select Update type
On Command or On Timer. Select OK. In selecting On Command the customer
needs to go into menu to be able to synchronize, see next picture. Easiest way
for the customer is to have Update Type On Timer.
Fig. 3:60
21. Select “5 Enable Mosaic Communication” to start synchronizing with the
Mosaic System.
Fig. 3:61
Service Manual 1001 5015 / Rev. 43:31
Infratec™ 1241 Grain Analyzer
22. When the first synchronization is done with Mosaic server, the message
“Instrument is waiting registration by Mosaic Manager” will appear if the
synchronization was successful. Continue to register the instrument in Mosaic.
Fig. 3:62
Connect Infratec to Network in Mosaic
23. In Unassigned Instrument the newly synchronized Infratec 1241 will appear.
Fig. 3:63
3:32Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
24. Right click on the Instrument and select “Register…” from the context menu.
Fig. 3:64
25. Select the Network and Instrument group that the instrument should be assigned
to and click OK.
Fig. 3:65
Service Manual 1001 5015 / Rev. 43:33
Fig. 3:66
Infratec™ 1241 Grain Analyzer
3:34Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
4Diagnostics and Troubleshooting
4.1General
As a Service technician you should try to gather as much information as possible before proceeding with a course of action when a customer calls to you with an instrument problem. The majority of problems that the customer experiences can be
divided into two main categories, "poor" results or instrument faults. Instrument
faults express themselves in error codes or that the instrument does not work at all.
Further, a problem reported may be a repeatable problem or an intermittent problem.
The later requires more information to be pinpointed. You should then ask the user
questions like;
•When in time did the problem occur first?
•How frequent is it?
•Did it occur in conjunction with any other event, e.g. relocating the instrument,
adding new hardware, adding new software etc?
•Has it escalated?
•Does the problem happen when certain events (keystrokes, selecting things etc)
happen?
•Try to think "outside the box" when you ask questions. Consider the
environment, e.g. vibrations, temperature, straylight, electricity, user log-in, time
in the day/week or anything else relevant that may cross your mind.
•Has the user himself observed anything that is out of the ordinary?
Consider the first contact with the customer as a golden opportunity to get more information by asking questions like the ones above. A telephone call is a duplex communication and many times you will get closer to the root cause even if you may not
be able to pinpoint the exact cause of the problem.
A very good way of getting closer to a solution is to learn how a working system behaves. Learn how "normal" logfiles and scans look like. Look at a normal startup and
learn the procedure for the hardware like LED´s, beeps, hardware movements, soft
ware messages etc. A deviation in this kind of information may also give clues to a
solution.
Discussing "poor results" may be the most difficult situation due to the fact that the
customer´s expectations can sometimes exceed the typical specified performance of
the instrument. For NIR instruments, which always is an in-direct method of meas
urement, you must consider the robustnes of the application, the sample itself and
perhaps environmental influence. This without saying that the problem might be re
lated to hardware or software in the system.
If you wan't to escalate a support incident to the 2nd line support please always add
all the information you have gained during your first attempts to troubleshoot. Also
describe the action you may have taken and the effect of those actions, if any. For the
Infratec 1241 ALWAYS supply the
This chapter helps you to quickly find the explanation and likely cause to most of the
error messages and faults. The beginning of the chapter includes questions that help
you to sort out what kind of problem that has occurred.
complete set of log files from the instrument.
-
-
-
Service Manual 1001 5015 / Rev. 44:1
Infratec™ 1241 Grain Analyzer
4.1.1Poor Results
The points listed below can be considered as a check list for trouble shooting “poor”
results. The reasons for “poor” results can be divided into:
•User problems
•Poor calibrations
•Outliers
•Poor reference data
•Instrument problems (Time for cleaning? see User Manual)
4.1.2User Problems
•Are they using the right calibration?
•Has there been an adjustment to the slope or intercept in the calibration?
•Are they sure that the moisture basis of the Infratec results is the same as that of
the reference results?
•Do they use adjustment to dry content or constant moisture content?
•How many sub-samples are used?
4.1.3Poor Calibrations
•Was the calibration developed for this type of sample?
•Was there sufficient data to develop the calibration?
•Was the calibration stabilised for instrument and temperature differences?
4.1.4Outliers
Results with outliers should always be confirmed by a second analysis.
A sample may be classified as an outlier for a number of reasons:
•The constituent concentration is outside of the calibration range.
•The sample is of a variety for which the NIT spectrum differs greatly from the
samples in the calibration set.
•The product is analysed with wrong AM.
•There is too much foreign material in the sample.
•Movements of the sample during the scan.
•Too high absorbance.
•Packing of sample not good
For each constituent, an outlier will be displayed as a letter (A-E) and a number
(1-5). The letter indicates what kind of outlier it is, as described below, and the
number describes the severity (the higher the number, the stronger the outlier).
A-outlier (residual) and B-outlier (leverage)
An outlier denoted by the letters A and B indicate how closely the NIT spectra of the
sample correlate with the spectra in the calibration set.
C-outlier (sub sample deviation)
An outlier denoted by the letter C indicates that the standard deviation between the
sub predicted values from the individual sub samples is above the set limit.
D-outlier (out of range)
An outlier denoted by the letter D indicates that the low or high limit for the constituent has been exceeded by the predicted value of at least one of the sub samples. This
means that it is possible to get a D-outlier even if the average result on the display is
within the calibration range.
4:2Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
E-outlier (difference between sample and ambient temperature)
An outlier denoted by the letter E indicates that the difference between the sample
temperature and the ambient temperature is outside the set limits.
Sensitivity of outlier detection
The sensitivity for detection of outliers can be set in the instrument software, see User
Menu, Application Models®Setup®Outlier limits. For the outlier constants, it is im
portant that a relevant value is set, so only real outliers will be indicated. If the value
selected is too low, samples that are not real outliers will be indicated as such. If the
value is set too high, some real outliers may be missed.
The sensitivity setting for C is the maximum acceptable standard deviation for each
set of sub samples. The appropriate value for this parameter, as well as the appropri
ate values for A and B parameters, are determined when the calibration is developed.
Values for the A,B, and C constants will be recommended for AMs supplied by Foss
Analytical.
The low and high limits for D should be set according to the range for each constituent in the actual AM.
The following are some questions that should be asked concerning the outlier:
•How many of the samples give outliers?
•In which positions are the outliers (A, B, C, D or E)?
•What is the value of the outlier (1-5)?
•Have they had outlier indications on that type of sample before?
•Are the outlier indications reproducible?
•Is the used calibration the correct one?
-
-
Conditions that will cause a sample to be flagged as an outlier
•Scan data does not fit with the calibration
•The sample differs from the typical samples in the calibration
•Standard deviation between sub-sample predictions outside preset limit
•Sample outside calibration range
Possible causes for outliers
•Wrong sample analysed
•Variety not included in the calibration
•Extreme chemical composition
•Impure sample (has too much foreign material)
•Instrument error
•Sample too hot or too cold (i.e. sample is frozen)
Corrective actions in the event of an outlier
•Re-run the analysis and verify it to be a true outlier
•Check to see if the sample meets any of the criteria mentioned in the section
“Possible causes for outliers” above. If so, take corrective measures.
•Use reference method (Wet chemistry) instead
4.1.5Poor Reference Data
•Which laboratory was used?
•Was more than one laboratory used, and if so, were there level differences?
•Are there systematic errors lab to lab bias?
•What reference method was used?
Service Manual 1001 5015 / Rev. 44:3
Infratec™ 1241 Grain Analyzer
•Were determinations made induplicated?
•How was sample selection performed?
•Was grinding performed?
•What is the reference methods reproducibility?
•Is the moisture basis the same in the laboratory results as in the Infratec results?
4.1.6Instrument Problems
•The Measuring Unit compartment of the instrument should be kept as clean as
possible. Do not forget to sometimes also clean the Detector Window on the Cell
House and the Cell Window. By cleaning the instrument regularly you avoid
excessive dust build-up and thereby you lower the risk of stoppage. For cleaning
instructions, see User Manual.
•Do they have a reference set of samples that they measure frequently on the
instrument? This should be done to see if the instrument has changed! What
changes?
•How is the sub-sample standard deviation for each constituent?
•Does it change during the day?
•Do the results drift during the day? If so, how much and does this occur every
day?
4.1.7I2C Trouble Shooting
Infratec 1241 is basically a PC with some extra hardware to perform analysis. The
hardware is connected to the PC via a two-wire bus called the I2C. This is a Phillips
developed bus.
All communication (commands, analytical data, error messages) is performed
through the bus.
equipment is separate but connected to the bus. When the extra modules are
connected the bus will detect the extra nodes (addresses) and open up the software to
control them.
The instrument software can be divided into three levels, not counting the BIOS and
the operating system. The top level is the ISW (instrument Software), which is handling all information to for example the display. The DSP software is controlling the
I2C communication as well as being a co processor for mathematical treatment of the
data signal. In certain modules there are software (firm ware) controlling low level
tasks like opening and closing the shutter in the Measuring Unit.
When the power is turned on the BIOS at the processor starts working. After some
seconds there is a beep as in a normal PC and then the operative system QNX (UNIX
type) starts working. After some initiation the ISW starts and the DSP software is
down loaded from the flash disc to the DSP processor and starts to run. About this
time the signal to turn on the display is sent to the power supply via the I2C bus and
information starts to flow showing the start up procedure.
Fig. 4:1 shows the bus and all connected equipment. Note that some
4:4Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
This table gives the approximate times for these events.
Service Manual 1001 5015 / Rev. 44:5
Infratec™ 1241 Grain Analyzer
Processor
Module
PC/104
External PC
Modem
Printer
Com1
Com 2
LPT1
DSP
Module
Fan
Flash
disk
+5V
+12V
GND
GND
Display
Keypad
PROCESSOR MODULE BLOCK DIAGRAM
IDE
1241026a
+5V
DECODE
CONTROL
LED
CONTROL
Monochromator
Detector
MU
TWM
PSM
Onboard LED's
Detector
ADC
Mono-
chromator
grating
DSP
1241021a
I2C
PC
PC/104
FILTER
STM
1241010b
1
2
3
4
5
6
7
8
11
12
10
9
If the bus is corrupted in any way, the communication will not work. It is therefore
possible to have a black display, when the error is actually in the I2C commun ication.
The following troubleshooting scheme is suggested.
Situation: A black display or an error log with information pointing in different direc-
tions.
1.Disconnect all external equipment (STM, TWM) if present.
2.Disconnect the i2c cable from the PSU, monochromator, detector and MU to the
DSP board. Attach a jumper on the open connector on the Display
interconnection board or use an external monitor. Start the instrument and if the
display starts up you should get 4 beeps and the following error message:
Any other error message indicates an error in the DSP board. If the display stays
black go to step 8.
3.Reattach the PSU i2c cable, remove the jumper on the display interconnection
board and restart the instrument. The following error message should appear:
Any other error message indicates an error in the detector.
6.Reattach the MU i2c cable and restart the instrument. If you get an error message
at this stage disconnect the ambient senor and the E-id from the MU and restart
the instrument.
7.Reconnec t STM and TWM one at a time if prese nt. Restart the instrument and
check for errors.
8.Measure the voltage on the connection JP3 blue wire on the power supply. It
should be more then 4V. If the voltage is over 4V and the display still is black
there is a problem in the display or the display interconnection board.
Service Manual 1001 5015 / Rev. 44:9
Infratec™ 1241 Grain Analyzer
4.2 Service Menu
In order to troubleshoot an instrument, Service technicians will have to access the
Service Menu. The Service Menu assume a more detailed knowledge about the In
fratec 1241 Grain Analyzer. The Service Menu is therefore password protected.
Password is only given to personnel approved by FOSS.
Most of the headings in the Service Menu have sub menus, see menu structure on
next page.
Service menu for Infratec 1241
Level 1Level 2Level 3Level 4Comments
1 Quality Control
1 Self Test
1 Run Self-TestRun a Self-Test
2 ExportCopy Start-Up test result or
3 PrintPrint Start-Up test result or
2 Clean Measuring
Unit
3 Audit LogView latest log by user
4 Software Error
Log
5 Hardware Error
Log
6 Export Logs and
Configuration
7 Wavelength Stability
8 Lamp CheckChecks the condition of the
Self-Test result to a USB stick
Self-Test result
Prepare the cell for cleaning
View latest Software Error
Log
View latest Hardware Error
Log
View error logs or copy
them to a USB stick
Testing the Monochromator
Wavelength Stability
Monochromator lamp
-
2 Automatic Tests
3 Manual Tests
1 Detector
2 Monochromator
3 I/O
1 Detector
2 Monochromator
1 ADC No Missing
Codes
2 Offset with Empty
Cell
1 Grating System
2 Air Stability
3 Shutter
4 Internal BG20
5 External BG20
6 Reference Scan
7 Encoder Information
1 COM 1
2 COM 2
3 Lpt 1
4 Keyboard
5 External Keyboard
4:10Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
Level 1Level 2Level 3Level 4Comments
3 Measuring Unit
1 Position Setting
2 Sensors
3 Set Ambient Temperature
4 Display
5 Scan
6 Test WeightGreyed out if not installed
1 Position Settings
2 Calibrate
1 Balance Control
2 Mass Calibration
3 Adjust Balance
3 Test Weight Cycle
4 AD Values
5 Flush
7 Sample Transport
Module
1 Position SettingsPossibility to manually posi-
2 ServiceTest of sensors and motor
3 Exercise
8 Bottle UnitGreyed out on 1241
1 Position settingsPossibility to manually posi-
2 SensorsPossibility to check the Bot-
Greyed out if not installed
tion the Cuvette Holder
tion the Bottle Unit
tle Unit position sensors
4 Manual AnalysisLengths, Speeds, Shutter
1 Settings
2 Fill Sample Cell
3 Scan
4 Offset
5 Empty Sample
Cell
6 Flush Sample Cell
5 Instrument Info
1 HardwareInstrument version of DSP,
2 SoftwareSoftware version
3 BiosBios version
4 Monochromator
Constants
5 Operating Temperature
6 Volume Information
7 Voltage Information
6 Clone InstrumentExport of configuration for
Positions etc. can be set
manually from here
Monochromator, Detector
etc.
The O- and P-constants
Min. and Max. operating
temperature
View available space on the
database disk
modem, menu and lan
guage
-
Service Manual 1001 5015 / Rev. 44:11
Infratec™ 1241 Grain Analyzer
Level 1Level 2Level 3Level 4Comments
7 Settings
1 Default SettingsResets the instrument to its
2 Scan Mode
(Standard)
default settings
8 Read DiskRead contents of USB stick
and install possible options
e.g. AM-s or programs
4.3Quality Control
4.3.1Self-test
Fig. 4:4 Self-test sub-menus
When the instrument is powered-up or when manually chosen, a Self-test is executed. The different tests take place in the order as they are specified in this section.
Screen Display
The PC-board is not capable to send screen display to both LCD and CRT. The default screen mode is LCD.
The LCD backlight is intentionally disabled at boot-up. It is enabled at the end of th e
DSP initialization where the PSU initialization activates the IIB..
For troubleshooting purpose you can bypass the disabled backlight buy shortcircuiting the jumper JP3 on the Display Interconnection Board.
PC Board System Test and Initialization
These routines test and initialize PC board hardware. If the routines encounter an error during the tests, you will either hear a few short beeps or see an error message on
the screen. There are two kind s of errors: fatal and non-fatal. The system can usually
continue the boot up sequence with non-fatal errors. Non-fatal error messages usual
ly appear on the screen along with the following instructions:
Press <F1> to RESUME
Write down the message and press the F1 key to continue the boot up sequence.
-
4:12Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
PC Module System configuration verification
These routines check the current system configuration against the values stored in the
CMOS memory. If they do not match, the program outputs an error message. You
will then need to run the BIOS setup program to set the configuration information in
memory.
There are three situations in which you will need to change the CMOS settings:
1.You are starting your sys tem for the first time
2.You have changed the hardware attached to your system
3.The CMOS memory has lost power and the configuration information has been
erased.
The PC-board memory has an external lithium battery backup for data retention.
Power-up beeper and DSP LEDs
When the instrument is powered-up, there is a beeper function to notify if a failure in
the instrument software start-up routine, or DSP software loading, has occurred:
•1 beep OK
•2 beeps Could not open DSP device
•3 beeps DSP communication failure
•4 beeps LCD display enable failure (see Trouble Shooting scheme)
LEDs on the DSP
First the power (green) is lit. After some time the LED 1 is lit. The DSP software is
now running. When the DSP software is communicating the LED 4 is flashing rap
idly. At this state the display should be lit. For more information regarding the DSP
LEDs see Technical Description.
Communication
Communications between the instrument software and the DSP board are first
checked when the instrument is powered-up. Failure in the instrument software to
2
DSP communication is notified by a beeper signal, as described above. Then the I
C
communications from the DSP to each hardware module (Monochromator comm.,
Detector comm., Measuring Unit comm. and PSM comm.) are tested. The detector
temperature sensor and the ambient temperature sensor (placed on the hood) are also
tested and reported as “True” or “False”.
NOTE: The "1241xxxx" written on each line symbolise the serial
number of the instrument.
This check is using the same accuracy as the voltage display in the service menu 5.7.
Monochromator
Grating system
The grating system functionality test checks the functionality of the grating shaft encoder. The test results report the maximum (positive) and the minimum (negative)
encoder unlinearity when the grating makes a complete scan, and the total difference
(i.e. max-min). The number of encoder pulses to complete a single revolution (for
ward and reverse) of the grating is also recorded, along with the time to complete the
scan. The following table lists the names, typical values and limits for the above pa
rameters.
This test validates O- and P-constants, serial number of the Monochromator and peak
value for internal BG20 filter. If the checksum does not match what is stored in the
EEPROM, an error occur and the instrument will be unable to analyse.
Measures the offset voltage at the ADC input with no light falling on the detector,
e.g. monochromator shutter closed. The detector gain is stepped up from 1*1 to
1000*16 with the minimum, maximum, mean ADC signal and standard deviation re
corded for each gain setting, as shown in the table below. The detector temperature
is recorded together with the ADC data.
This test validates the serial number and gain settings. If the checksum does not
match what is stored in the EEPROM, an error occur and the instrument disables
analysis
This test checks the stability of the air reference scans after power-up.
The test checks the level of the reference scan and reports test “Failed” if it is outside
limits (15000<Max level<65280). If the test fails due to the reference scan limits, the
result will only show one row displaying the ADC level.
The difference between two succeeding groups of 6 scans are calculated and should
not exceed the limit 100. The first rows indicate the level of the reference scan at the
start of the test, because the first differences are calculated against a zero value. If the
difference is >100 another group of 6 scans are calculated against the previous group
and again the difference should be below 100. There is always a 10 second delay be
tween the groups. This continues until the difference is less than 100 or a time limit
of 20 minutes is reached. The minimum number of steps is 13.
Line description including typical valueUnitsMin.Max.
* Level of reference scan, calculated against zero vector.
Monochromator Wavelength Stability
Wavelength stability is measured by calculating the 2
nd
absorbency peak from the internal BG20 filter. Measurements are made on ten subsamples.
The mean value of the peaks is then compared to the one stored in the monochromator EEPROM. The mean, maximum, minimum, standard deviation and difference are
displayed on the screen. The difference between the measured wavelength and the
stored wavelength is presented as Peak
2
Diff. The maximum limit is +/- 1nm (one),
if the value is outside the limits, an error message is created.
Line description including typical valueUnitsMin.Max.
Once the Self-test is completed a report line is given the overall "Passed/Failed" status of each of the communication, monochromator, detector, air stability and monochromator wave length tests.
The date and time of the Self-test is also shown. There will be a report on the screen
with the test result as below.
The details from each test could be viewed by choosing one of the individual tests.
When the Self-test is done at start-up, the test report will only be shown if one or
more of the tests have "Failed". For detailed information of the Self-test, the test must
be exported.
Line description including typical valueUnitsMin.Max.
Makes the instrument run a sequence for cleaning of the cell. This should be done on
daily basis. This function can also be reached from the User Menu.
4.3.3Audit Log
Displays the audit trail log. When choosing the Export Logs and configuration
option an “1241xxxx.al“ file is created, see Appendix “Logs Description” in User
Manual.
4.3.4Software Error Log
Displays the software Error log. The information in the software log file is basically
the messages that the operator sees on the display. The time stamps in this log can
many times be matched to the timestamps in the hardware log and thus give more in
formation when you are troubleshooting. Note that the software log information is
always in English despite the selected language.
When choosing the Disk option an “xxxxxxxx.sl” file is created.
-
Service Manual 1001 5015 / Rev. 44:19
Infratec™ 1241 Grain Analyzer
4.3.5Hardware Error Log
Fig. 4:6
Displays the hardware Error log. When choosing the Disk option an “xxxxxxxx.h l”
file is created. For decoding of the Hardware error log, see
4.12.
4.3.6Export Logs and Configuration
Exports the record of errors that occurred during analysis, and the instrument configuration, to a USB memory stick. For decoding of the Export logs, see 4.13.
4.3.7Wavelength Stability
Tests and displays the monochoromator wavelength stability, see 4.3.1.
4.3.8Lamp Check
Displays the intensity of the lamp. Measured value is presented for gain 1x2.
4:20Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
4.4Automatic Tests
From here all of the tests carried out in the Self-test routine can be executed individually. The test results are presented in the same way as described in section Self-test,
with an option to save to disk.
Fig. 4:7 Automatic Tests
4.4.1Detector
Fig. 4:8 Detector Tests
ADC No Missing Codes
Identical to the "ADC No Missing Codes" test carried out in the Self-test routine.
Offset with Empty Cell
Identical to the "Detector Offset" test carried out in the Self-test routine.
Service Manual 1001 5015 / Rev. 44:21
Infratec™ 1241 Grain Analyzer
4.4.2Monochromator
Fig. 4:9 Monochromator Tests
Grating System
Identical to the "Grating System" test carried out in the Self-test routine.
Air Stability
Almost identical to the "Air stability test" test carried out in the Self test routine, with
one exception: This test has an option to break the test by using the Stop function that
makes it possible to go on measuring stability after the test has passed. On the screen
the maximum difference values are shown together with the incremental values and
the scan peak value.
Shutter
The Shutter test consists of two parts: The first is checking the repeatability of the
shutter positioning and the second is measuring the light leakage.
For the shutter repeatability positioning, a reference scan is taken with the lamp on
and shutter open. A new scan is taken after the shutter is closed. Then a third scan
after the shutter is reopened again. The difference between the first and third scan has
to be below limit. Failure can be viewed on screen and logged to Error log.
For the Light Leakage test the shutter is now closed again and a fourth scan is taken.
The second and fourth scans must also be within limits.
The max. and min. levels of the second and fourth scans must also be within limits.
The lamp is then turned off (shutter still closed) and a fifth scan is taken. The differ
ence between the fourth and fifth scan must be within limits. Failure can be viewed
on screen and logged to Error log.
Line description including typical valueUnitsMin.Max.
Identical to the "Monochromator Wavelength Stability" test carried out in the Selftest routine.
The same limits are used.
External BG20
With this test it is possible to compare one master BG20 filter with the results from
the internal BG20 filter test. Insert a permanent sample with a BG20 glass filter in
4:22Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
the measuring unit when prompt for. Ten sub samples are taken and the 2nd peak is
detected and displayed on the screen.
Line description including typical valueUnitsMin.Max.
A reference scan is taken and the maximum level of the scan is displayed.
No software error indication on a scan level failure. The raw data values can be exported to disk by choosing the Disk option.
Line description including typical valueUnitsMin.Max.
It is possible to test communication ports and parallel ports by using loop-back connectors.
1001 0301Loop-back connector Serial
1001 0302Loop-back connector Parallel (printer)
For COM 1 and COM2 always make sure that you use the standard (Default)
settings for modem and serial line before starting this test.
This is changed in User Menu: Instrument settings → View → Communication →
Communication settings (1 Set Modem) (2 Set Serial). Choose “Default Modem
RS232“ in Set Modem. Restart the instrument before starting the loop-back test.
Choose “Default Direct RS232“ in Set Modem. Transmitted data values are
0x00-0xFF.
LPT1 One byte written and read back.
Service Manual 1001 5015 / Rev. 44:23
Infratec™ 1241 Grain Analyzer
The test runs the ports individually as chosen in the "I/0" menu. If any error occur
this is reported to screen and Error log.
Line description including typical valueUnitsMin.Max.
The test is going through every key from upper left to down right by stating which
key to press. When the correct key is pressed the next key is stated.
If no key is detected within 10 seconds from last pressed key, the test aborts. If aborted, this will be logged to Error log with the message on which key that should be
pressed when the sequence was aborted.
External Keyboard
The test shows which key on the external keyboard that is pressed at the moment. If
no key is detected within 10 seconds after the last pressed key,
the test aborts without any Error log message.
4:24Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
4.5Manual Tests
To exercise all of the instrument low-level hardware commands. All the tests below
are addressed one by one and performed in manual order.
Fig. 4:11 Manual Tests
4.5.1Detector
Available options:
•Set Gain value
•Read Gain value
•Read detector temperature
GainPre GainMain GainTotal Gain
1111x1
2121x2
3141x4
4181x8
51161x16
631131x1
731231x2
831431x4
931831x8
10311631x16
11100011000x1
12100021000x2
13100041000x4
14100081000x8
151000161000x16
4.5.2Monochromator
Available options:
•Set shutter position
•Read shutter position
•Set lamp On/Off
•Set grating motor On/Off
Service Manual 1001 5015 / Rev. 44:25
4.5.3Measuring Unit
Fig. 4:12 Measuring Unit Tests
Position Setting
Available options:
•Set path length
•Set bottom shutter Open/Closed
•Set top shutter Open/Closed
•Set conveyor belt On/Off
Infratec™ 1241 Grain Analyzer
Sensors
Available options:
•Read ambient temperature (Calculated using the ambient temp. sensor in the
hood together with the detector temp. sensor)
•Read speckle detectors
•Read drawer sensor
•Read internal temperature (Using the temp. sensor on the Measuring Unit board)
Set Ambient Temperature
Possibility to force the ambient temperature input to an arbitrary value.
The ambient temperature value will return to real values if the "Set Ambient temperature" is set to "0.0" or if the "Default Settings" command is executed or if the instrument is restarted. Used to temporarily bypass the calculations mentioned under
Position Settings and Sensors.
4.5.4Display
Available options:
•Set display On/Off
•Set back-light On/Off
•Adjust display contrast value
4.5.5Scan
To collect a scan with the present settings and return the maximum value together
with the option to save scan data values on disk or cancel. The raw data values can
be exported to disk by choosing the Disk option.
4:26Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
4.5.6Test Weight
Fig. 4:13 Test Weight
Position Settings
This function is for testing of the Wiper, Shutter and Sensor.
Calibrate
– Balance Control; This function can also be reached from the User Menu. Instructions given on the display.
– Mass Calibration; Instructions given on the display. Use two weights with good accuracy and two decimals. The result is stored in the EEPROM of the TWM.
– Adjust Balance; Should only be used after discussion with Product specialist at
FOSS.
Test Weight Cycle
This function runs a complete Test weight cycle (sequence) but with possibilities to
check part sequences and continue. The raw data will be presented on the display to
gether with the calculated value. It is possible to catch the sample in the cuvette and
measure on another scale of sufficient accuracy.
AD Values
Continues reading of AD values with an interval of 10 seconds.
Flush
This function empties the hopper content into the drawer.
4.5.7Sample Transport Module
Position Settings
Possibility to manually position the Cuvette Holder.
Service
Test of sensors and motor. Can be used to run the motor up and down.
Exercise
No use for this option on the field.
-
Service Manual 1001 5015 / Rev. 44:27
Infratec™ 1241 Grain Analyzer
4.6Manual Analysis
Fig. 4:14 Manual Analysis
4.6.1Settings
The parameters used in the manual analysis sequence:
•Set path length
•Set fill speed
•Set flush speed
4.6.2Fill Sample Cell
Fills the measuring unit sample cell at the speed pre-set in settings.
4.6.3Scan
Collects a scan at a gain and a shutter position chosen in the window popped up when
choosing Scan. When scanning is finished the maximum value of the scan will be
displayed together with options to save the scan data values on Disk or Cancel.
4.6.4Offset
Calculates the offset at the present gain. The gain can be set manually in Manual
Tests→Detector. The offset value is displayed with the present settings, together
with the maximum, minimum, standard deviation and detector tempera ture. There is
an option to save the offset value to Disk or Cancel.
4.6.5Empty Sample Cell
Empties the measuring unit sample cell.
4.6.6Flush Sample Cell
Flushes the measuring unit sample cell and the hopper at the speed pre-set in settings.
4:28Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
4.7Instrument Info
Fig. 4:15 Instrument Info
4.7.1Hardware
Displays the following:
•DSP software version
•Monochromator software version and serial number
•Detector serial number
•Measuring unit software version
•Optional module and serial number
4.7.2Software
Displays the different instrument software module versions.
4.7.3BIOS
Displays the BIOS information.
4.7.4Monochromator Constants
Displays the O- and P-constant and internal BG20 peak values on the screen.
4.7.5Operating Temperature
Displays the min. and max. operating temperatures which the instrument is intended
to work within. The digits can not be changed.
4.7.6Volume Information
The Flash disk is partitioned in three parts, 5 Mb for the operative system, 5 Mb for
the ISW. The remaining space is for data.
4.7.7Voltage Information
See Voltage Check in section 4.3.1.
4.8Clone Instrument
For copying of an instrument´s settings and transfer them to another instrument. Put
a USB stick in the disk drive of the instrument which settings you wish to copy. After
finished copying, put the USB stick in the other instrument and run function Read Disk. A script file will automatically adjust its settings to the ones on the USB stick.
IP number and TWM calibration will only be cloned to disk (not copied to the next
instrument). These settings has to be handled in other ways.
Service Manual 1001 5015 / Rev. 44:29
Infratec™ 1241 Grain Analyzer
4.9Settings
4.9.1Default Settings
To be able to return to a normal instrument status the "Default Settings" command
sets the instrument to the following status:
Monochromator lampOn
Grating motorOn
ShutterOpen
Detector gain1x1
Initialise Measuring Unit
Set Ambient temperature0.0
4.9.2Scan Mode (Standard)
Switches to the extended wavelength mode. Allows measurement in the wavelength
range 570 - 1100nm.
4.10Read Disk
Imports updates and application models from a USB memory stick.
4:30Service Manual 1001 5015 / Rev. 4
Infratec™ 1241 Grain Analyzer
4.11Error Codes in Software Log (SL) and Hardware Log (HL)
When reviewing the SL and HL log files it's a good idea to look at the two files
simultaneously. Try to compare the entries using the time stamps. This way you will
many times get a better understanding because you can see both what the customer
reported (SL) and what command was sent at the same time. In the HL file you will
also see the status reported back by the module that was addressed.
SL
Error
HL
Error
Message to the User (What
possibly can be done before
contacting Service)
CommentsNote
327680x8000 Hardware error.See
4.11.7
330250×8101 Grating motor error.Monochromator Motor cannot
move.
330260×8102 Monochromator EEPROM er-
ror.
The Monochromator was not
able to save constants in its
EEPROM.
Message to the User (What
possibly can be done before
contacting Service)
CommentsNote
335380x8302TWM operation mode switch.
335390x8303 TWM EEPROM errorProblems to read/write the data
in the PROM
335400x8304 TWM EEPROM errorSlope data outside limits. try to
re calibrate the balance.
340500x8502 Lamp On/Off error.
340510x8503 Set display On/Off error.
340520x8504 Set back light On/Off error.
340530x8505 Set contrast On/Off error.
10x0001 Instrument communication er-
ror.
20x0002 Detector ADC errorThere is no signal from the De-