Fluke Networks OF500 Data Sheet

OptiFiber® Specifications and Ordering Information
General Specifications
Weight Main with module and battery: 4.5 lbs (1.9 kg)
Dimensions Main with module and battery:
Battery Li-Ion
Battery life 8 hours typical
Connections USB, RS-232, FiberInspectorTM camera port, PS-2 keyboard
Memory card SD MMC
Environmental Specifications
Operating temperature 0°C to 40°C
Non-operating temperature -20°C to 60°C
Relative humidity (%RH operating without condensation)
Vibration Random, 2 g, 5-500 Hz
Shock 1 meter drop test while module is both
11.4 x 7.5 x 2.5 in (29.0 x 19.1 x 6.4 cm)
95% (10 to 35°C); 75% (35 to 40°C) Uncontrolled < 10°C
inserted inside and outside the instrument enclosure
Key OTDR Specs (23°C)
Output/input connector SC/UPC removeable/cleanable SC/UPC removable/cleanable
Wavelengths 850 ± 20 nm and 1300 ± 25 nm 1310 ± 25 nm and 1550 ± 30 nm
Fiber types tested 50/125 µm or 62.5/125 µm multimode 9/125 µm singlemode
Event deadzone 850 nm: 0.5 m typical
Attenuation deadzone 850 nm: 4.5 m typical
Pulse width 850 nm: 4 ns, 20 ns
Max distance range 850 nm: 3 km, 1300 nm: 7 km 1310/1550 nm: 60 km
Dynamic range 850 nm: 15 dB typical
Testing speed <10s for two wavelengths at 2 km with 25 cm resolution
Output power 850 nm: >110 mW-pk
Distance accuracy ± 1 m ± 0.005% of distance ± 50% of resolution ± IOR error ± event location error
Loss threshold .2 dB .01 dB to 1.5 dB inclusive
Linearity ± .07 dB/dB ± .05 dB/dB
Sample resolution 3 cm to 50 cm 3 cm to 400 cm
Reflectance accuracy ± 4 dB ± 4 dB
ORL accuracy ± 4 dB ± 4 dB
Laser classification Class 1 CDRH, complies to EN 60825-2
Multimode: OFTM-5610B, OFTM-5611B, OFTM-5612B
1300 nm: 1.3 m typical
1300 nm: 10.5 m typical
1300 nm: 8 ns, 40 ns, 100 ns, 200 ns, 400 ns, 650 ns
1300 nm: 14 dB typical
<30s for two wavelengths at 400 m with 3 cm resolution
1300 nm: >22 mW-pk
Singlemode: OFTM-5730, OFTM-5731, OFTM-5732
1310/1550 nm: 1 m typical
1310/1550 nm: 8 m typical
1310 nm/1550 nm: 5 ns, 20 ns, 40 ns, 100 ns, 300 ns, 1 µs, 3 µs, 10 µs
1310 nm: 26 dB typical 1550 nm: 24 dB typical
Auto OTDR 15 seconds typical Manual OTDR 15 seconds to 3 min
1310 nm: >28 mW-pk 1550 nm: >24 mW-pk
Key Power Meter and Loss/Length
OFTM-5612B, OFTM-5732
Specifications (23°C)
Input/output connectors Removable/interchangeable input connector; SC/PC output connector
Nominal output wavelengths OFTM-5612B: LED source: 850 nm and 1300 nm
Measurement range OFTM-5612B: 5 km of 62.5 µm or 50 µm multimode fiber; 20 km in far-end source mode
Power measurment detector type InGaAs
Calibrated wavelengths for power measurment 850 nm, 1310 nm, 1550 nm
Power measurement range 0 to -60 dBm (1300/1310 nm and 1550 nm); 0 to -52 dBm (850 nm)
Output power (nominal) OFTM-5612B: -19.5 dBm; OFTM-5732: -7 dBm
Note:
Above specifications are subject to change without notice
OFTM-5732: Laser source: 1310 nm and 1550 nm
OFTM-5732: 10 km of 9 µm singlemode fiber; 60 km in far-end source mode
OptiFiber® Certifying OTDR. Bringing Network SuperVision to premise fiber networks with the most complete fiber optic test solution.
Power Meter Specifications OFTM-5611B, OFTM-5731
Input connector Removable/interchangeable SC input connector
Detector type InGaAs
Calibrated wavelengths 850 nm, 1310 nm, 1550 nm
Power measurement range 0 to -60 dBm (1300/1310 nm and 1550 nm), 0 to -52 dBm (850 nm)
FiberInspector™ Pro Specifications OFTM-5352
Magnification Switchable – 250x / 400x
Dimensions (without tip) 1.8 x 1.7 x 5.5 in (4.6 x 4.3 x 14.0 cm)
Weight 0.4 lb (180 g)
(Length depends on adapter tip)
VFL Specifications OFTM-5730, OFTM-5731, OFTM-5732
On/off control Controlled by OptiFiber software (no hardware switch)
Output power (into singlemode fiber)
Operating wavelength 650 nm nominal
Spectral width ± 3 nm
Output modes Continuous and pulse mode (2Hz-3Hz blink)
Connector adapter 2.5 mm universal
Laser safety Class II CDRH complies to EN60825-2
316 um (-5 dBm) < or equal to peak power > or equal to 1 mw (0 dBm)
OptiFiber Packages Spec Table
Model Description Includes
OF-500-01 OptiFiber Multimode OTDR Mainframe, battery, MM OTDR module, 62.5 µm launch fiber cable, CDs with LinkWare
OF-500-01/50M OptiFiber Multimode OTDR (50 µm) 50µm launch fiber replaces 62.5 µm launch fiber
OF-500-02 OptiFiber Multimode OTDR with
Power Meter
OF-500-02/50M OptiFiber Multimode OTDR with Power
Meter (50 µm)
OF-500-03 OptiFiber Singlemode OTDR Mainframe, battery SM OTDR module, launch fiber, CDs with LinkWare and manuals,
OF-500-10 OptiFiber Multimode Certifying OTDR Mainframe, battery, MM OTDR/OPM/LL module, 62.5 µm launch fiber, CDs with LinkWare
OF-500-10/50M OptiFiber Multimode Certifying
OTDR (50 µm)
OF-500-13 Singlemode Certifying Smart Remote
OTDR
OF-500-15 OptiFiber SR Certifying OTDR Mainframe, battery, MM OTDR/OPM/LL module, 62.5 µm launch fiber, CDs with
OF-500-15/50M OptiFiber SR Certifying OTDR (50 µm) 50 µm launch fiber and test reference cords replace 62.5 um launch fiber and test refer-
OF-500-35 OptiFiber Multimode/Singlemode OTDR Mainframe, battery, MM OTDR/OPM/LL module, SM OTDR/OPM/LL module, protective car-
OF-500-35/50M OptiFiber Multimode/Singlemode OTDR
(50 µm)
OF-500-45 OptiFiber Multimode/Singlemode
Certifying OTDR
and manuals, user guide, 32 MB MMC, USB cable, AC adapter, protective carrying case
Mainframe, battery, MM OTDR/OPM module, 62.5 µm launch fiber, CDs with LinkWare and manuals, user guide, 32 MB MMC, USB cable, AC adapter, carrying strap, protective carrying case
50 µm launch fiber and test reference cords replace 62.5 µm launch fiber
user guide, 32 MB MMC, USB cable, AC adapter, carrying strap, protective carrying case
and manuals, user guide, 32 MB MMC, USB cable, AC adapter, carrying strap, protective carrying case, 62.5 µm test reference cords, 250X/400X inspection camera, MMC reader
50 µm launch fiber and test reference cords replace 62.5 um launch fiber and test refer­ence cords replace 62.5 µm
Mainframe, battery, SM OTDR/OPM/LL module, launch fiber, CDs with LinkWare and manuals, user guide, 32MB MMC, USB cable, AC adapter (x2), carrying strap, test reference cords, 250X/400X inspection camera, hard carrying case, MMC reader, Smart Remote unit, Smart Remote SM Module, Smart Remote user guide, Smart Remote carrying strap, mini-B USB cable
LinkWare and manuals, user guide, 32 MB MMC, USB cable, AC adapter (x2), carrying strap, 62.5 µm test reference cords, 250X/400X inspection camera, hard carrying case, MMC reader, Smart Remote unit, Smart Remote MM Module, Smart Remote user guide, Smart Remote carrying strap, mini-B USB cable
ence cords
rying case, 62.5 µm, 8.3 µm launch fibers, CDs with LinkWare and manuals, user guide, 32 MB MMC, USB cable, AC adapter, carrying strap, 62.5 µm,
8.3 µm test reference cords, 250X/400X inspection camera, hard carrying case
50 µm launch fiber and test reference cords replaces 62.5 µm, launch fiber and test reference cords
Includes contents of OF-500-35 plus: 50 µm launch fiber, 50 µm test reference cords protective carrying case, Smart Remote unit, Smart Remote MM module, Smart Remote carry strap, Smart Remote user guide, Mini-B USB cable, a second AC adapter, MMC reader
OptiFiber Mainframe and Modules Ordering Information
Model Description Use
OF-500 OptiFiber Mainframe and Battery Mainframe, battery, CD with LinkWare and manuals, user manual, AC adapter,
OFTM-5610B OptiFiber Multimode OTDR Module Pinpoint the location of problems within a multimode fiber link, even if spaced
OFTM-5611B OptiFiber Multimode OTDR + PM Module Measure the attenuation of multimode fiber link using a separate far-end
OFTM-5612B OptiFiber Multimode OTDR + PM
+ Loss/Length Module
OFTM-5730 OptiFiber Extended Range Singlemode
OTDR Module
OFTM-5731 OptiFiber Extended Range Singlemode
OTDR + PM Module
OFTM-5732 OptiFiber Extended Range Singlemode OTDR +
PM + Loss/Length Module
32 MB MMC, USB cable, carrying strap
as closely as .5 m apart
850/1300 nm source
Certify a multimode link using the proven standards-based two fiber, dual wavelength methodology
Pinpoint the location of problems within a singlemode fiber links, even if spaced as closely as 1 m apart
Measure the loss of singlemode fiber links using a separate far-end 1310/1550 nm source
Certify a singlemode link using the proven standards-based two fiber, dual wavelength methodology
OptiFiber Options and Accessories Ordering Information
Model Description Use
OFTM-5352 FiberInspector Pro Video Probe, 250x/400x High resolution 250x/400x inspection of fiber connectors on patch panels and
OFSR-MMREM Smart Remote with multimode option Use with OptiFiber mainframe for Smart Remote loss/length certification.
OFSR-SFM Singlemode Smart Remote module option Use with the OptiFiber Smart Remote Option (OFSR-MMREM) to allow Smart
OFSR-MFM Multimode Smart Remote module option Use with a DTX CableAnalyzer™ to allow its remote unit to work with an OptiFiber
NF360 FiberInspector MT-RJ Probe Tip For inspecting MT-RJ small form factor connectors
NF362 FiberInspector LC Probe Tip For inspecting LC small form factor connectors
NF364 FiberInspector MU Probe Tip For inspecting MU small form factor connectors
NF366 FiberInspector E2000 Probe Tip For inspecting E2000 connectors
NFK1-LAUNCH SC-SC 62.5/125 µm launch cable, 100 m OptiFiber launch and receive fibers
NFK2-LAUNCH SC-SC 50/125 µm launch cable, 100 m
NFK3-LAUNCH SC-SC Singlemode launch cable, 130 m
OFBP-LI OptiFiber Li-Ion battery Test a full 8 hours between charges with integrated charger
OFCC-SOFTCASE OptiFiber protective carry case Protective carrying case for mainframe, module, launch fiber and accessories
OFCC-HCASE OptiFiber hard carrying case Ultimate protection for the mainframe, modules, batteries and accessories
OPV-KB External mini-keyboard Enter information quickly with the plug and play keyboard
32MB Secure digital memory card Record storage
DTX-SMC32 32MB secure digital memory card
MMC CASE MMC softcase Easily keep track of your MMC cards in this handy case that stores up to 8 MMC
DSP-MCR-U MMC Reader, USB cable Upload saved results to the PC for results management, analysis and reporting
8251-13 SimpliFiber 850/1300 LED Source, SC Optional sources for use with OptiFiber OTDR + power meter modules
8251-11 SimpliFiber 1310 Laser Source, SC
8251-12 SimpliFiber 1550 Laser Source, SC
NF400 Soft Case for Patch Cords & Accessories Easily keep track of patch cords and small accessories
NF430 Cleaning kit Properly clean fibers with optical grade cleaning pads, swabs and alcohol
800695 Adapter cable for cigarette lighter jacks Charge the OptiFiber while route to the jobsite
OPV-PS AC adapter with line cord Keep your battery charged up with AC adapter
OPV-KB External keyboard Use optional keyboard for fast set up
NFA-SC SC connector adapters for OFTM Power Meter and
DTX Module inputs, set of two
cable assemblies. Includes Probe Adapter Tips (ST, SC, FC, and universal 2.5 mm patch cord tip)
Includes smart remote unit with a smart remote multimode module and accessories
Remote loss/length certification of singlemode fibers
mainframe as an OptiFiber smart remote unit for loss/length testing of multimode fibers
cards
Replacement SC adapters for Power Meter Ports
OptiFiber Options and Accessories Ordering Information
Model Description Use
NFA-LC LC connector adapters for OFTM Power Meter and
DTX Module inputs, set of two
NFA-ST ST connector adapters for OFTM Power Meter
and DTX Module inputs, set of two
NFA-FC FC connector adapters for OFTM Power Meter
and DTX Module inputs, set of two
NFK1-DPLX-SC SC/SC to SC/SC Duplex 62.5 um test reference
cord, 2 m, set of two
NFK1-DPLX-LC SC/LC to LC/LC Duplex 62.5 um test reference
cord, 2 m, set of two
NFK1-DPLX-ST SC/ST to ST/ST Duplex 62.5 um test reference
cord, 2 m, set of two
NFK1-DPLX-FC SC/FC to FC/FC Duplex 62.5 um Test Reference
cord, 2 m, set of two
NFK2-DPLX-SC SC/SC to SC/SC Duplex 50 um test reference cord,
2 m, set of two)
NFK2-DPLX-LC SC/LC to LC/LC Duplex 50 um test reference cord,
2 m, set of two
NFK2-DPLX-ST SC/ST to ST/ST Duplex 50 um test reference cord,
2 m, set of two
NFK2-DPLX-FC SC/FC to FC/FC Duplex 50 um test reference cord,
2 m, set of two
NFK3-DPLX-SC SC/SC to SC/SC Duplex Singlemode test
reference cord, 2 m, set of two
NFK3-DPLX-LC SC/LC to LC/LC Duplex Singlemode test
reference cord, 2 m, set of two
NFK3-DPLX-ST SC/ST to ST/ST Duplex Singlemode test
reference cord, 2 m, set of two
NFK3-DPLX-FC SC/FC to FC/FC Duplex Singlemode test
reference cord, 2m, set of two
NF-Mandrel-Kit Mandrel Kit 2 each for testing 50 um and
62.5 um with 3 mm jacket
NF-Mandrel -50 Mandrel for 50 um fiber with 3 mm jacket Use mandrels for accurate, standard compliant multimode measurements
NF-Mandrel-625 Mandrel for 62.5 um fiber with 3 mm jacket Use mandrels for accurate, standard compliant multimode measurements
LinkWare LinkWare cable test management software CD The best results management tool to document your results
NFK1-LAUNCH-ST Launch Fiber, Multimode 62.5 um SC/ST 100 m Hybrid launch fiber for testing cabling with ST style connectors
NFK2-LAUNCH-ST Launch Fiber, Multimode 50 um SC/ST 100 m
NFK3-LAUNCH-ST Launch Fiber, Singlemode SC/ST 130 m
NFK1-LAUNCH-LC Launch Fiber, Multimode 62.5 um SC/LC 100 m Hybrid launch fiber for testing cabling with LC style connectors
NFK2-LAUNCH-LC Launch Fiber, Multimode 50 um SC/LC 100 m Hybrid launch fiber for testing cabling with LC style connectors
NFK3-LAUNCH-LC Launch Fiber, Singlemode SC/LC 130 m Hybrid launch fiber for testing cabling with LC style connectors
NFK2-LAUNCH-E2K Launch Fiber, Multimode 50 um SC/E2000 100 m Hybrid launch fiber for testing cabling with E2000 style connectors
NFK3-LAUNCH-E2K Launch Fiber, Singlemode SC/E2000 130 m
NF-OPRT-SC SC OTDR Port Adapter Replacement adapter for OptiFiber OTDR port
Replacement LC adapters for Power Meter Ports
Replacement ST adapters for Power Meter Ports
Replacement FC adapters for Power Meter Ports
Use high quality, durable, test reference cords and mandrels for the best power and loss length measurements
Use mandrels for accurate, standard compliant multimode measurements
N E T W O R K S U P E R V I S I O N
Fluke Networks
P.O. Box 777, Everett, WA USA 98206-0777
Fluke Networks operates in more than 50 countries worldwide. To find your local office contact details, go to www.flukenetworks.com/contact.
©2004 Fluke Corporation. All rights reserved. Printed in U.S.A. 7/2006 2032360 D-ENG-N Rev F
Loading...