Table 1 lists the International Electrical Symbols. Some or all of these symbols
may be used on the instrument or in this manual.
Table 1. International Electrical Symbols
Symbol Description
Chapter 1
Before You Start
Ο
І
CAT II
AC (Alternating Current)
AC-DC
Battery
CE
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Guide (Important Information)
Off
On
Canadian Standards Association
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree
2 per IEC1010-1 refers to the level of Impulse Withstand Voltage
protection provided. Equipment of OVERVOLTAGE
CATEGORY II is energy-consuming equipment to be supplied
from the fixed installation. Examples include household, office,
and laboratory appliances.
1-1
9150
User's Guide
1.2 Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
“Warning” identifies conditions and actions that may pose hazards to th
user.
“Caution” identifies conditions and actions that may damage the
instrument being used.
To avoid personal injury, follow these guidelines.
•BURN HAZARD - DO NOT touch the well access surface of
the unit.
•The temperature of the well access is the same as the
actual temperature shown on the display. If the unit is set
at 600°C and the display
1.2.1
•
•
Warnings
C-TICK Australian EMC mark
The European Waste Electrical and Electronic Equipment
(WEEE) Directive (2002/96/ EC) mark.
Warnings
reads 600°C, the well
is at 600°C.
e
1-2
•The top sheet metal of the furnace may exhibit extreme
temperatures for areas close to the
•The air over the well can reach temperatures greater than
200°C. Pro
the unit
200°C.
•
DO NOT turn off the unit at temperatures higher than
100°C. This could create a hazardous situation. Select a
set-point less than 100°C and allow
turning it off.
•DO NOT remove inserts and insulators at high
temperatures. Inserts and insulators are the same
temperature as the display temperature.
•DO NOT operate this unit without a properly
properly polarized power cord.
•DO NOT co
polarized outlet.
•HIGH
SEVERE IN- JURY OR DEATH may result if personnel fail to
observe safety precautions. Before working inside the
equipment, turn the power off and disconnect the power
cord.
bes should only be inse
when the unit is set at temperatures
nnect this unit to a non-grounded, non-
VOLTAGE is use
d in the operation of this equipment.
well acc
rted and removed from
the unit to
ess.
less than
cool before
grounded,
•Always replace the fuse with one of the same rating,
voltage, and ty
pe.
Before You Start
1.2 Safety Information1
•Overhead cl
instrument under a cabinet or other structure
•DO NOT use this unit for any appl
calibration w
•DO NOT us
earance is required. DO NOT place this
.
ication other than
ork.
e this unit in environments other than those
listed in the user’s guide.
•DO NOT turn the unit upside down with the inserts in place;
the inserts w
ill fall out
of the unit.
•DO NOT operate near flammable materials.
•Use of this instrument at HIGH TEMPERATURES for
extended periods of time requires caution.
•Completely unattended high temperature operation is not
recommended for safety
reasons
.
•Before initial use, after transport, and anytime the furnace
has not been energized for more than 10 da
calibrator must be energized for a dry
-out pe
ys,
the
riod of 1 to 2
hours before it can be assumed to meet all of the safety
requirements of the IEC1010-1.
•Materials u
sed in this furnace may
be irritating to skin,
eyes, and respiratory tract. Consult the material
manufacturer’s MSDS (Material Safety Data Sheet). Proper
ventilation and safety precautions must be observed.
1.2.2
•Follo
•Calibration Equipment should only
w all safety guidelines listed in
the user’s guide.
be used by
Trained
Personnel.
Cautions
Cautions
•To avoid possible damage to the instrument, follow t
guidelines.
•DO NOT plug the unit into 230 V if the heater switches and
fuse holder read 115 V. This action will cause the fuses to
blow and may damage the instrument.
•Components and heater lifetime can be shortened by
continuous high temperature operation.
•Most probes have handle temperature limits. Be sure that
the probe hand temperature limit is not exceeded in the air
above the unit.
•Allow for test probe expansion inside the well as the
furnace heats.
•DO NOT use fluids to clean out the well.
hese
1-3
9150
User's Guide
•Never introduce foreign material into the probe hole of the
insert. Fluids, etc. can leak into the calibrator causing
damage.
•DO NOT change the values of the calibration constants
from the factory set values. The correct setting of these
parameters is important to the safety and proper operation
of the calibrator.
•DO NOT drop or force the probe stems into the well. This
type of action can cause a shock to the sensor and affect
the calibration.
•DO use a ground fault interrupt device.
1-4
Before You Start
1.3 Authorized Service Centers1
1-5
9150
User's Guide
1-6
The Hart Scientific 9150 thermocouple furnace can be used for calibrating thermocouple and RTD temperature probes. Calibrations may be done over a range
of 150°C to 1200°C (302°F to 2192°F). Temperature display of the 9150 is 0.1
degrees below 1000°and 1 degrees above 1000°.
The furnace features:
• Rapid heating and cooling
• Interchangeable multiple hole probe sleeves
Chapter 2
Introduction
• Convenient handle
• RS-232 interface
Built in programmable features include:
• Temperature scan rate control
• Eight set-point memory
• Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s hybrid analog/digital controller. The controller uses a Type S thermocouple as a sensor and controls the
well temperature with a solid state relay (triac) driven heater. The furnace should
be calibrated yearly by trained personnel at Hart Scientific.
The LED front panel continuously shows the current well temperature. The
temperature may be easily set with the control buttons to any desired temperature within the specified range. The calibrator’s multiple fault protection de- vices
insure user and instrument safety and protection.
The 9150 calibrator furnace was designed for portability, low cost, and ease of
operation. Through proper use, the instrument will provide continued accurate
calibration of temperature sensors and devices. The user should be familiar with
the safety guidelines and operating procedures of the furnace as described in
this manual.
2-1
9150
User's Guide
2-2
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.