Figure 8 Serial Cable Wiring ........................................................................... 43
viii
Before You Start
Symbols Used
Before You Start1
Symbols Used1.1
Table 1 lists the International Electrical Symbols. Some or all of these symbols may be
used on the instrument or in this manual.
Table 1 International Electrical Symbols
SymbolDescription
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Guide (Important Information)
Off
On
Canadian Standards Association
1
7103 Micro-Bath User’s Guide
Safety Information
SymbolDescription
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re fers to
the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE
CATEGORY II is energy-consuming equipment to be supplied from the xed installation.
Examples include household, office, and laboratory appliances.
C-TICK Australian EMC mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/
EC) mark.
Safety Information 1.2
Use this instrument only as specif ed in this manual. Otherwise, the protection
provided by the instrument may be impaired.
The following def nitions apply to the terms “Warning” and “Caution”.
L
“Warning” identif es conditions and actions that may pose hazards to the user.
L
“Caution” identif es conditions and actions that may damage the instru ment
being used.
1.2.1 Warnings
Disclaimer: Hart Scientific manufactures instruments for the purpose of
temperature calibration. Instruments used for applications other than
calibration are used at the discretion and sole responsibility of the customer.
Hart Scientific cannot accept any responsibility for the use of instruments for
any application other than temperature calibration.
GENERAL
DO NOT use the instrument for any application other than calibration work.
The instrument was designed for temperature calibration. Any other use of the unit
may cause unknown hazards to the user.
DO NOT use the unit in environments other than those listed in the user’s guide.
Completely unattended operation in not recommended.
DO NOT install access cover without holes onto a bath that is energized. Dangerous
pressures may result from f uids vaporizing.
Follow all safety guidelines listed in the user’s manual.
2
Before You Start
Safety Information
Calibration Equipment should only be used by Trained Personnel.
If this equipment is used in a manner not specif ed by the manufacturer, the protection
provided by the equipment may be impaired or safety hazards may arise.
Inspect the instrument for damage before each use. DO NOT use the instru ment if it
appears damaged or operates abnormally.
Before initial use, or after transport, or after storage in humid or semi-humid
environments, or anytime the instrument has not been energized for more than 10
days, the instrument needs to be energized for a “dry-out” period of 2 hours before
it can be assumed to meet all of the safety requirements of the IEC 61010-1. If the
product is wet or has been in a wet environment, take necessary measures to remove
moisture prior to applying power such as storage in a low humidity temperature
chamber operating at 50°C for 4 hours or more.
The instrument is intended for indoor use only.
Lift the instrument by the handle provided to move the instrument. DO NOT move
the instrument until the display reads less than 25°C (77°F) and the unit has been
drained or the Transport Lid installed.
BURN HAZARD
ALWAYS ensure the instrument is COOL before closing the instrument for storage.
DO NOT mix water and oil when exceeding temperatures of 90°C (194°F).
DO NOT touch the well access surface of the unit.
The temperature of the well access is the same as the actual temperature shown on
the display, e.g. if the unit is set at 125°C and the display reads 125°C, the well is at
125°C.
Ensure the power cord is positioned in such a way as it cannot contact hot sur faces
or temperature probes. Always inspect power cord before use for any damage to the
insulation due to contact with hot surfaces, cuts or abrasions.
The top sheet metal of the instrument may exhibit extreme temperatures for ar eas
close to the well access.
DO NOT turn off the unit at temperatures higher than 100°C. This could create a
hazardous situation. Select a set-point less than 100°C and allow the unit to cool
before turning it off.
DO NOT remove the f uid at high temperatures. The f uid will be the same
temperature as the display temperature.
DO NOT operate near f ammable materials. Extreme temperatures could ignite the
f ammable material.
Use of this instrument at HIGH TEMPERATURES for extended periods of time
requires caution.
3
7103 Micro-Bath User’s Guide
Safety Information
ELECTRICAL HAZARD
These guidelines must be followed to ensure that the safety mechanisms in this
instrument will operate properly. This instrument must be plugged into a 115 VAC
(± 10%) or 230 VAC (± 10%) 50/60 Hz only electric outlet as indicated on the back
panel. The power cord of the instrument is equipped with a three-pronged grounding
plug for your protection against electrical shock haz ards. It must be plugged directly
into a properly grounded three-prong recepta cle. The receptacle must be installed
in accordance with local codes and ordinances or adapter plug. DO NOT use an
extension cord. Consult a qualif ed electrician. Always inspect the power cord
before use for any damage to the in sulation due to contact with hot surfaces, cuts or
abrasions.
The instrument is equipped with operator accessible fuses. If a fuse blows, it may
be due to a power surge or failure of a component. Replace the fuse once. If the fuse
blows a second time, it is likely caused by failure of a component part. If this occurs,
contact a Hart Scientif c Authorized Service Center. Always replace the fuse with
one of the same rating, voltage, and type. Never replace the fuse with one of a higher
current rating.
Always replace the power cord with an approved cord of the correct rating and type. If
you have questions, contact a Hart Scientif c Authorized Service Center (see Section
1.3).
High voltage is used in the operation of this equipment. Severe injury or death may
result if personnel fail to observe the safety precautions.
1.2.2 Cautions
Always operate this instrument at room temperatures listed in Section 3.2,
Environmental Conditions. Allow suff cient air circulation by leaving at least 6 inches
(15 cm) of clearance around the instrument.
Overhead clearance is required. DO NOT place this instrument under a cabinet or
other structure.
Never introduce any foreign material into the well.
DO NOT change the values of the calibration constants from the factory set values.
The correct setting of these parameters is important to the safety and proper operation
of the unit.
DO NOT slam the probe stems into the well. This type of action can cause a shock to
the sensor and affect the calibration.
DO use a ground fault interrupt device.
DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty
environment.
4
Before You Start
Authorized Service Centers
The unit is a precision instrument. Although it has been designed for optimum
durability and trouble free operation, it must be handled with care.
Most probes have handle temperature limits. Be sure that the probe handle tem perature
limit is not exceeded in the air above the instrument.
The instrument and any thermometer probes used with it are sensitive instru ments that
can be easily damaged. Always handle these devices with care. Do not allow them to
be dropped, struck, stressed, or overheated.
When calibrating PRTs always follow correct calibration procedure and cali brate
from high temperatures to low temperatures with the appropriate triple point of water
checks.
Components and heater lifetimes can be shortened by continuous high tempera ture
operation.
If a mains supply power f uctuation occurs, immediately turn off the bath. Power
bumps from brown-outs and black-outs can damage the instrument. Wait until the
power has stabilized before re-energizing the bath.
Fluids may expand at different rates. Allow for f uid expansion inside the well as
the instrument heats. Otherwise, the f uid may overf ow the well and leak into the
instrument.
5
7103 Micro-Bath User’s Guide
Authorized Service Centers
6
Introduction
Introduction 2
The Hart Scientif c 7103 Micro-Bath may be used as a portable instrument or bench
top temperature calibrator for calibrating thermocouple and RTD temper ature probes.
The 7103 is small enough to use in the f eld, and accurate enough to use in the lab.
With an ambient temperature of 23°C (74°F), calibrations may be done over a range
of -30°C to 125°C (-22°F to 257°F). The resolution of the 7103 temperature display is
0.01 degrees.
The Micro-Bath calibrator features:
L
Convenient handle
L
RS-232 interface
L
Switchable AC Input (115 VAC or 230 VAC)
Built in programmable features include:
L
Temperature scan rate control
L
Temperature switch hold
L
Eight Set-point memory
L
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s hybrid analog/digital con troller. The
controller uses a precision platinum RTD as a sensor and controls the well temperature
with thermal electric devices (TED).
The LED front panel continuously shows the current well temperature. The
temperature may be easily set with the control buttons to any desired tempera ture
within the specif ed range. The calibrator’s multiple fault protection de vices insure
user and instrument safety and protection.
The 7103 Micro-Bath was designed for portability, low cost, and ease of opera tion.
Through proper use, the instrument will continuously provide accurate calibration
of temperature sensors and devices. The user should be familiar with the safety
guidelines and operating procedures of the calibrator as de scribed in the instruction
manual.
7
Specications and Environmental Conditions3
3.1 Specications
Table 2 Specifications
Range-30 to 125°C (-22 to 257°F)
Accuracy±0.25°C
Stability±0.03°C at -25°C (oil, 5010)
Uniformity±0.02°C
Resolution0.01°C/F
Operating Temperature5 to 45°C (41 to 113°F)
Heating Time25°C to 100°C: 35 minutes
Cooling Time25°C to –25°C: 45 minutes
Well Size2.5" dia. x 5.5" deep (64 x 139 mm)
Size9" W x 13.2" H x 10.5" D (23 x 34 x 26 cm)
Weight22 lb. (10 kg) with uid
DisplayLED, Switchable °C or °F
CoolingFan and Thermal Electric Devices (TED)
Fault ProtectionSensor burnout and short protection
Power94 – 234 VAC (±10%), 50/60 Hz, 400 W
SafetyOvervoltage (Installation) Category II, Pollution Degree 2 per IEC 61010-01
Fuse Rating250 V 5 A F(fast acting) NO USER SERVICEABLE PARTS
Specications and Environmental Conditions
Specications
±0.05°C at 125°C (oil, 5010)
(access opening is 1.9" [48 mm] in diameter)
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and clean ing recommendations
can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
L
ambient temperature range: 5 – 45°C (41 – 113°F)
L
ambient relative humidity: maximum 80% for temperature <31°C, de creasing
linearly to 50% at 40°C
L
mains voltage within ± 10% of nominal
L
vibrations in the calibration environment should be minimized
L
altitudes less than 2,000 meters
9
Quick Start4
Caution: READ SECTION 6 ENTITLED BATH USE before placing the bath in
service. Incorrect handling can damage the bath and void the war ranty.
4.1 Unpacking
Unpack the Micro-Bath carefully and inspect it for any damage that may have
occurred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
L
7103 Micro-Bath
L
Transport/Pour Access Lid
L
Probe Basket
L
Stir Bar
L
Power Cord
L
Manual
Access Cover, if purchasedL
Quick Start
Unpacking
Setup 4.2
Caution: DO NOT operate the instrument without fluid.
Place the calibrator on a f at surface with at least 6 inches of free space around the
instrument. Plug the power cord into a grounded mains outlet. Observe that the
nominal voltage corresponds to that indicated on the back of the calibrator.
Carefully insert the probe basket into the well. Fill the well with the appropriate f uid.
The set-point temperature and the number of and size of probes deter mine the f uid
level. Be sure to keep the f uid level an adequate distance below the top of the well
to prevent overf owing the f uid when the probes are in serted. For example, placing
200.05 oil at room temperature (25°C) into the bath and heating the unit to 126°C,
causes a 1-inch (2.54 cm) expansion of the f uid inside the well.
Keep the f uid level at least 1.9 cm (0.75 inches) below the top of the well at all
times. With the probe (probes) in the well f ll the tank 3/4 full. Heat to the maximum
temperature of the f uid. Slowly f ll the well to 2.54 cm (1 inch) be low the top of the
basket at the maximum temperature of the f uid, or to the second ring on the basket.
11
7103 Micro-Bath User’s Guide
Power
Turn on the power to the calibrator by toggling the switch on the power entry module.
The fan should begin blowing air through the instrument and the con troller display
should illuminate after 3 seconds. After a brief self test the con troller should begin
normal operation. If the unit fails to operate please check the power connection.
The display shows the well temperature and the well TEDs start operating to bring
the temperature of the well to the set-point temperature. Insure that the f uid is being
stirred.
Power 4.3
Plug the Micro-Bath power cord into a mains outlet of the proper voltage, fre quency,
and current capability. Refer to Section 3.1, Specif cations, for power details.
Turn the bath on using the rear panel “POWER” switch. The Micro-Bath will turn on
and begin to heat to the previously programmed temperature set-point. The front panel
LED display will indicate the actual bath temperature.
Setting the Temperature 4.4
Section 9.2 explains in detail how to set the temperature set-point on the cali brator
using the front panel keys. The procedure is summarized here.
Press “SET” twice to access the set-point value. 1.
Press “UP” or “DOWN” to change the set-point value. 2.
Press “SET” to store the new set-point. 3.
Press “EXIT” to return to the temperature display. 4.
When the set-point temperature is changed the controller switchs the well heater on
or off to raise or lower the temperature. The displayed well tempera ture gradually
changes until it reaches the set-point temperature. The well may require 90 minutes
to reach the set-point depending on the span. Another 10 to 15 minutes is required to
stabilize within ±0.04°C of the set-point. Ultimate sta bility may take 20 to 30 minutes
more of stabilization time.
12
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.