User-Replaceable Parts and Accessories ............................................. 51
ii
Introduction
The Fluke 2638A HYDRA Series III Data Acquisition Unit (the Product or
Instrument) is a 67 analog channel bench-top data logger that measures and
records dc volts, ac volts, dc current, ac current, resistance, frequency, and
temperature. See the Specifications section for information on the types and
ranges of the measurement inputs the Product can accept.
1
2638A
Calibration Manual
Safety Information
Symbol Description Symbol Description
A Warning identifies conditions and procedures that are dangerous to the user.
A Caution identifies conditions and procedures that can cause damage to the
Product or the equipment under test.
See Table 1 for a list of symbols used in this manual and on the Product.
Table 1. Symbols
W
X
or
Ã
~
Risk of Danger. Important
information. See Manual.
Hazardous voltage. Voltage
>30 V dc or ac peak might be
present.
Earth ground. Digital signal
AC or DC (Alternating or Direct
Current)
Recycle.
Conforms to relevant South
Korean EMC Standards.
Measurement Category II is applicable to test and measuring circuits connected directly to
utilization points (socket outlets and similar points) of the low-voltage MAINS installation.
Measurement Category III is applicable to test and measuring circuits connected to the
distribution part of the building’s low-voltage MAINS installation.
Measurement Category IV is applicable to test and measuring circuits connected at the
source of the building’s low-voltage MAINS installation.
This product complies with the WEEE Directive marking requirements. The affixed label
indicates that you must not discard this electrical/electronic product in domestic household
waste. Product Category: With reference to the equipment types in the WEEE Directive
Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation”
product. Do not dispose of this product as unsorted municipal waste.
DC (Direct Current)
AC or DC (Alternating or Direct Current)
T
+
Double insulated
Power ON / OFF
PConforms to European Union directives.
2
HYDRA Series III Data Acquisition Unit
Safety Information
WXWarning
To prevent possible electrical shock, fire, or personal injury:
• Read all safety information before you use the Product.
• Carefully read all instructions.
• Use the Product only as specified, or the protection supplied
by the Product can be compromised.
• Examine the case before you use the Product. Look for cracks
or missing plastic. Carefully look at the insulation around the
terminals.
• Do not use the Product if it operates incorrectly.
• Do not use the Product if it is damaged.
• Disable the Product if it is damaged.
• Use only the mains power cord and connector approved for
the voltage and plug configuration in your country and rated
for the Product.
• Replace the mains power cord if the insulation is damaged or
if the insulation shows signs of wear.
• Make sure the ground conductor in the mains power cord is
connected to a protective earth ground. Disruption of the
protective earth could put voltage on the chassis that could
cause death.
• Do not put the Product where access to the mains power cord
is blocked.
• Use only correct measurement category (CAT), voltage, and
amperage rated probes, test leads, and adapters for the
measurement.
• Use only cables with correct voltage ratings.
• Do not use test leads if they are damaged. Examine the test
leads for damaged insulation and measure a known voltage.
• Do not exceed the Measurement Category (CAT) rating of the
lowest rated individual component of a Product, probe, or
accessory.
• Keep fingers behind the finger guards on the probes.
• Do not apply more than the rated voltage, between the
terminals or between each terminal and earth ground.
• Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.
• Limit operation to the specified measurement category,
voltage, or amperage ratings.
• Measure a known voltage first to make sure that the Product
operates correctly.
3
2638A
Calibration Manual
• Consider all accessible channels to be hazardous live and an
electric shock hazard if any channel is connected to a
hazardous voltage source.
• Do not remove, touch, or change the internal wiring of
hazardous inputs until the input source is turned off.
• Remove inputs from hazardous voltage sources before an
input module is opened.
• Use the correct terminals, function, and range for
measurements.
• Use this Product indoors only.
• Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
4
HYDRA Series III Data Acquisition Unit
General Specifications
General Specifications
Mains Voltage
100 V Setting ...................................................... 90 V to 110 V
120 V Setting ...................................................... 108 V to 132 V
220 V Setting ...................................................... 198 V to 242 V
240 V Setting ...................................................... 216 V to 264 V
Frequency .............................................................. 47 Hz to 440 Hz
Power Consumption ............................................. 36 VA peak (24 W average)
Environment Temperature
Operating ............................................................ 0 °C to 50 °C
Full accuracy ....................................................... 18 °C to 28 °C
Storage ............................................................... -20 °C to 70 °C
Warm-up ............................................................. 1 hour to full accuracy specifications
Relative Humidity (non-condensing)
Operating ............................................................ 0 °C to 28 °C <90 %
28 °C to 40 °C <75 %
40 °C to 50 °C <45 %
Storage ............................................................... -20 °C to 70 °C <95 %
Altitude
Operating ............................................................ 2,000 m
Storage ............................................................... 12,000 m
Channel Capacity
Total analog channels ........................................ 67
Measurement Front Panel .................................. IEC 61010-2-030: CAT II 300 V
Measurement Rear Panel ................................... IEC 61010-2-030: CAT II 150 V CAT II, 250 V rms with maximum
Electromagnetic Compatibility (EMC)
International ........................................................ IEC 61326-1: Basic Electromagnetic Environment
transient voltage of 1000 V peak. These terminals are not intended for
connection to mains voltage above 150 V without external transient
suppression. The maximum input that can be applied between rearmodule terminals or between any rear-module terminal and earth
ground is 250 V dc or ac rms.
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for
the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low-voltage power
supply network that supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated
disturbances.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.
5
2638A
Calibration Manual
Korea (KCC) ....................................................... Class A Equipment (Industrial Broadcasting & Communication
USA (FCC) .......................................................... 47 CFR 15 subpart B. This product is considered an exempt device per
Math Channels
Number of channels............................................ 20
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user should take
notice of it. This equipment is intended for use in business
environments and not to be used in homes.
Accuracy specifications generally are valid for 6 ½ digit resolution mode (unless otherwise noted) for front panel input
(Channel 001), after a minimum of 1-hour warm-up, and within an environment temperature range of 18 °C to 28 °C.
24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per EN
61326. The confidence level for accuracy specifications is 99 % within 1 year of calibration (unless otherwise noted).
Scan rate (typical, depending on function and range)
Fast ..................................................................... 46 channels per second max (0.02 seconds per channel)
Medium ............................................................... 10 channels per second (0.1 seconds per channel)
Slow .................................................................... 2 channels per second (0.5 seconds per channel)
Display Resolution ................................................ 4 ½ to 6 ½ digits, depending on Sample Rate or NPLC
DC Voltage
Maximum Input ...................................................... 300 V on any range
Common Mode Rejection ..................................... 140 dB at 50 Hz or 60 Hz (1 kΩ unbalance for NPLC of 1 or greater,
±500 V peak maximum in the low lead)
Normal Mode Rejection ........................................ 55 dB for NPLC of 1 or greater and power-line frequency ±0.1 %,
• For conducted disturbances on mains input >1 V from 10 MHz to 20 MHz, add 0.02 % of range. For disturbances
>3 V, accuracy is unspecified.
• For radiated disturbances >1V/m from 450 MHz to 550 MHz, add 0.02 % of range. For disturbances > 3 V/m,
accuracy is unspecified.
24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
T.C./ °C
Outside
18 °C to 28 °C
DC Voltage Additional Errors
Digits NPLC Ch. x01 – x20
6 ½ 200
6 ½
6 ½
5 ½ 1 (Medium)
4 ½ 0.2 (Fast) 4 ½ 0.02 -
100
10 (Slow)
add 2 μV
add 2 μV
add 2 μV
add 2 μV
Additional NPLC Noise
Error
-
-
-
add 0.0008 % of range
add 0.002 % of range + 12 μV
add 0.014 % of range + 17 μV
AC Voltage
AC voltage specifications are for ac sine wave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range. For 50 kHz to 100 kHz, add 0.13 % of range.
Maximum Input ...................................................... 300 V rms or 425 V peak or 3 × 10
less) for any range.
Measurement Method ........................................... AC-coupled true-rms. Measures the ac component of input with up to
6 ½ 200 - 6 ½ 100 - 6 ½ 10 (Slow) - 5 ½ 1 (Medium) 0.0008 % of range 0.0008 % of range
4 ½ 0.2 (Fast) 0.002 % of range + 1.2 µA 0.002 % of range + 12 nA
4 ½ 0.02 0.014 % of range + 1.7 µA 0.014 % of range + 17 nA
Additional NPLC Noise Error
for 10 mA, 100 mA
Additional NPLC Noise
Error for 100 µA, 1 mA
Burden
Voltage
<1 mV
<1 mV
<1 mV
<1 mV
T.C./ °C
Outside
AC Current
Input Protection .................................................... 0.15 A / 600 V resettable PTC
Measurement Method ........................................... AC-coupled true-rms, dc-coupled to the shunt (no blocking capacitor).
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for a sinusoidal signal with
amplitude greater than 5 % of range. For input signals between 1 % to 5 % of range, add 0.1 % of range.
Gate Times ............................................................. 100 milliseconds to 1 second.
Measurement Method ........................................... Flexible counting technique. AC-coupled input using the ac voltage
Settling Considerations........................................ When measuring frequency after a dc offset voltage change, errors
Measurement Considerations.............................. To minimize measurement errors, shield inputs from external noise
measurement function.
may occur. For the most accurate measurement, wait up to 1 second
for the input blocking capacitor to settle.
when measuring low-voltage, low-frequency signals.
Frequency Accuracy
Accuracy is given as ± % of measurement.
Range Frequency
100 mV to
[1] [2]
300 V
[1] Input >100 mV. For 10 mV to 100 mV, multiply percent measurement error by 10.
[2] Limited to 3 x 10
20 Hz to 40 Hz 0.03 % 0.03 % 0.03 % 0.001 %
40 Hz to 1 MHz
7
volt-Hertz
24 Hour
(23 ±1 °C)
0.006 % 0.01 % 0.01 % 0.001 %
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
T.C./ °C
Outside 18
°C to 28 °C
Resistance
Measurement Method ........................................... Current source referenced to LO input.
Max. Lead Resistance (4-wire ohms) .................. 10 Ω per lead for 100 Ω, 1 kΩ ranges. 1 kΩ per lead on all other
Input Protection .................................................... 300 V on all ranges.
ranges.
10
HYDRA Series III Data Acquisition Unit
Measurement Specifications
Resistance Input Characteristics
Resolution
Range Resolution
100 Ω 100.0000 Ω 10 mΩ 1 mΩ 0.1 mΩ
1 kΩ 1.000000 kΩ 100 mΩ 10 mΩ 1 mΩ
10 kΩ 10.00000 kΩ 1 Ω 100 mΩ 10 mΩ
100 kΩ 100.0000 kΩ 10 Ω 1 Ω 100 mΩ
1 MΩ 1.000000 MΩ 100 Ω 10 Ω 1 Ω
10 MΩ 10.00000 MΩ 1 kΩ 100 Ω 10 Ω
100 MΩ 100.0000 MΩ 10 kΩ 1 kΩ 100 Ω
Fast
4½ Digits
Medium
5 ½ Digits
Slow
6 ½ Digits
Source Current
1 mA / 4 V
1 mA / 4 V
100 µA / 6 V
100 µA / 12 V
10 µA / 12 V
1 µA / 12 V
0.1 µA / 12 V
Resistance Accuracy
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for 4-wire resistance. For 2-wire
resistance add 0.02 Ω internal resistance if using Channel 1 or 1.5 Ω if using channels x01 through x20, and add external
lead wire resistance.
Range
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
Note:
For conducted disturbances on mains input >1 V from 10 MHz to 40 MHz, add 0.6 % of range. For disturbances >3 V,
6 ½ 200 0 % of range
6 ½ 100 0 % of range
6 ½ 10 (Slow) 0 % of range
5 ½ 1 (Medium) 0.0008 % of range
4 ½ 0.2 (Fast)(only for 2-wire)
4 ½ 0.02 (only for 2-wire)
0.002 % of range + 12 mΩ
0.01 % of range + 17 mΩ
RTD
Temperature Range .............................................. -200 °C to 1200 °C (depending on the sensor)
Resistance Range ................................................. 0 Ω to 4 kΩ
Maximum Lead Resistance (4-wire Ω) ................ 2.5 % of range per lead for 400 Ω and 4 kΩ ranges
Medium ............................................................... 2 PLC
Fast ..................................................................... 1 PLC
11
2638A
Calibration Manual
RTD Temperature Accuracy
Accuracy is for 4-wire 100 Ω nominal RTD, using the slow sample rate. With 3-wire PRT/RTD add 0.015 °C if using
Channel 1, or add 0.15 °C internal resistance mismatch to the accuracy specification if using channels x01 through x20,
and add external lead wire resistance mismatch. When using medium or fast sample rate (NPLC <10), add the number
given in the table to the accuracy specification. If the environment temperature is outside the specified range, multiply the
Temperature Coefficient number by the temperature deviation and add to the accuracy specification. Linear interpolation
may be used between points in the table. Specifications do not include sensor accuracy. The practical range of
temperature measurement depends on the sensor and characterization.
Temperature Accuracy
-200 °C 0.016 °C add 0.02 °C 0.0026 °C
0 °C 0.038 °C add 0.02 °C 0.0041 °C
300 °C 0.073 °C add 0.03 °C 0.0063 °C
600 °C 0.113 °C add 0.03 °C 0.0089 °C
Note:
For conducted disturbances on mains input >1 V from 10 MHz to 40 MHz, add 0.2 Celsius. For disturbances >3 V,
accuracy is unspecified.
Medium/Fast Sample
Rate (NPLC <10)
T.C./ °C Outside
18 °C to 28 °C
RTD Measurement Characteristics
Temperature Display Resolution
Range
0 Ω to 400 Ω
400 Ω to 4 kΩ
Slow
Sample Rate
0.001 °C 0.01 °C
0.001 °C 0.01 °C
Medium/Fast
Sample Rate
Source Current
1 mA
0.1 mA
Thermistor
Temperature Range .............................................. -200 °C to 400 °C (depending on the sensor)
Resistance Range ................................................. 0 Ω to 1 MΩ
Medium ............................................................... 2 PLC
Fast ..................................................................... 1 PLC
12
Thermistor Temperature Accuracy
Accuracy specifications are for a 4-wire thermistor using medium or slow sample rate. With 2-wire thermistor add the
number given in the table to the accuracy specification for internal resistance. When using fast sample rate (NPLC <10),
multiply the accuracy specification by 3. If the environment temperature is outside the specified range, increase the
accuracy specification by 25 % for every 1 °C outside the specified environment temperature range. Specifications do not
include sensor accuracy. The practical range of temperature measurement depends on the sensor.
Accuracy
Temperature
-40 °C 0.002 °C 0.007 °C 0.007 °C add 0.002 °C
0 °C 0.005 °C 0.004 °C 0.003 °C add 0.004 °C
25 °C 0.013 °C 0.007 °C 0.005 °C add 0.016 °C
50 °C 0.019 °C 0.01 °C 0.011 °C add 0.05 °C
100 °C 0.116 °C 0.054 °C 0.026 °C add 0.34 °C
150 °C 0.527 °C 0.239 °C 0.1 °C add 1.7 °C
Note:
For conducted disturbances on mains input >1 V from 10 MHz to 40 MHz, add 0.2 Celsius. For disturbances >3 V,
accuracy is unspecified.
2.2 kΩ
thermistor
5 kΩ
thermistor
10 kΩ
thermistor
2-wire
HYDRA Series III Data Acquisition Unit
Measurement Specifications
Thermistor Measurement Characteristics
Temperature Display Resolution
Range
0 Ω to 98 kΩ
95 kΩ to 1 MΩ
Slow
Sample Rate
0.001 °C 0.01 °C
0.001 °C 0.01 °C
Medium/Fast
Sample Rate
Source Current
10 μA
1 μA
Thermocouple
Temperature Range .............................................. -270 °C to 2315 °C (depending on the sensor)
Voltage Range ....................................................... -15 mV to 100 mV
Medium ............................................................... 2 PLC
Fast ..................................................................... 1 PLC
Thermocouple Temperature Accuracy
Accuracy specifications apply using a slow sample rate. When using a medium or fast sample rate (NPLC < 10), increase
the accuracy specification by 25 %. If the environment temperature is outside the specified range, increase the accuracy
specification by 12 % for every 1 °C outside the specified environment temperature range. Accuracy with fixed/external
CJC does not include the accuracy of the reference junction temperature. Linear interpolation may be used between
points in the table. Specifications do not include sensor accuracy. The practical range of temperature measurement
depends on the sensor.
Type
(Range)
K
-270 °C to 1372 °C
T
-270 °C to 400 °C
R
-50 °C to 1768 °C
S
-50 °C to 1768 °C
J
-210 °C to 1200 °C
N
-270 °C to 1300 °C
E
-270 °C to 1000 °C
Temperature
-200 °C
0 °C
1000 °C
-200 °C
0 °C
200 °C
400 °C
0 °C
300 °C
1200 °C
1600 °C
0 °C
300 °C
1200 °C
1600 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
500 °C
1000 °C
-200 °C
0 °C
300 °C
700 °C
Accuracy
Fixed / External CJC Internal CJC
Channel 1 Ch. x01 – x20 Ch. x01 – x20
0.28 °C
0.10 °C
0.14 °C
0.27 °C
0.10 °C
0.08 °C
0.08 °C
0.76 °C
0.42 °C
0.33 °C
0.34 °C
0.74 °C
0.45 °C
0.37 °C
0.39 °C
0.20 °C
0.08 °C
0.11 °C
0.42 °C
0.15 °C
0.12 °C
0.14 °C
0.17 °C
0.07 °C
0.06 °C
0.08 °C
0.41 °C
0.15 °C
0.20 °C
0.40 °C
0.15 °C
0.12 °C
0.11 °C
1.13 °C
0.63 °C
0.47 °C
0.49 °C
1.11 °C
0.67 °C
0.54 °C
0.56 °C
0.29 °C
0.12 °C
0.14 °C
0.62 °C
0.23 °C
0.17 °C
0.19 °C
0.25 °C
0.10 °C
0.09 °C
0.10 °C
1.60 °C
0.62 °C
0.64 °C
1.60 °C
0.65 °C
0.47 °C
0.41 °C
1.28 °C
0.71 °C
0.52 °C
0.54 °C
1.26 °C
0.76 °C
0.60 °C
0.63 °C
1.41 °C
0.61 °C
0.53 °C
1.69 °C
0.64 °C
0.44 °C
0.45 °C
1.42 °C
0.61 °C
0.46 °C
0.45 °C
13
2638A
Calibration Manual
B
100 °C to 1820 °C
C
0 °C to 2315 °C
D
0 °C to 2315 °C
G
0 °C to 2315 °C
L
-200 °C to 900 °C
M
-50 °C to 1410 °C
U
-200 °C to 600 °C
W
0 °C to 2315 °C
300 °C
600 °C
1200 °C
1600 °C
600 °C
1200 °C
2000 °C
600 °C
1200 °C
2000 °C
600 °C
1200 °C
2000 °C
-200 °C
0 °C
800 °C
0 °C
500 °C
1000 °C
-200 °C
0 °C
400 °C
600 °C
1200 °C
2000 °C
1.32 °C
0.68 °C
0.41 °C
0.38 °C
0.23 °C
0.28 °C
0.44 °C
0.22 °C
0.26 °C
0.39 °C
0.24 °C
0.22 °C
0.33 °C
0.13 °C
0.08 °C
0.09 °C
0.11 °C
0.10 °C
0.10 °C
0.25 °C
0.10 °C
0.08 °C
0.24 °C
0.22 °C
0.33 °C
1.97 °C
1.02 °C
0.60 °C
0.55 °C
0.33 °C
0.40 °C
0.60 °C
0.32 °C
0.36 °C
0.53 °C
0.36 °C
0.32 °C
0.46 °C
0.19 °C
0.12 °C
0.12 °C
0.16 °C
0.15 °C
0.14 °C
0.37 °C
0.15 °C
0.11 °C
0.36 °C
0.32 °C
0.46 °C
1.97 °C
1.02 °C
0.60 °C
0.55 °C
0.54 °C
0.63 °C
0.91 °C
0.44 °C
0.49 °C
0.69 °C
0.36 °C
0.33 °C
0.46 °C
0.99 °C
0.62 °C
0.48 °C
0.64 °C
0.51 °C
0.41 °C
1.48 °C
0.63 °C
0.40 °C
0.36 °C
0.33 °C
0.46 °C
Thermocouple Measurement Characteristics
Temperature Display Resolution
Range
Slow
Sample Rate
-270 °C to 2315 °C 0.01 °C 0.1 °C
Digital I/O
Absolute Voltage Range ......................................... -4 V to 30 V
Input Minimum Logic High ...................................... 2.0 V
Input Maximum Logic Low ...................................... 0.7 V
Output Type ............................................................ open drain active low
Output Logic Low (<1 mA) ...................................... 0 V to 0.7 V
Maximum Sink Current ........................................... 50 mA
Basic CJC Accuracy ............................................... 0.6 °C
notice in this manual under input protection)
15
2638A
Calibration Manual
General Maintenance
If the Product Does Not Turn On
Clean the Product
This section supplies information on how to clean the Product, reset the memory,
and replace the fuse in the rear of the Product.
To help solve problems encountered when turning on the Product:
1. Verify that the power switch is in the “On” position.
2. Make sure that the mains power cord is firmly plugged into the power module
on the rear of the Product.
3. Make sure the power source that the Product is plugged into is energized.
4. Verify that the line power fuse is good. See “Replace the Line Power Fuse”.
If these steps do not solve the problem, then contact Fluke. See the “Contact
Fluke”.
To clean the Product, wipe the instrument with a cloth that is lightly dampened
with water or mild detergent. Do not use aromatic hydrocarbons, chlorinated
solvents, or methanol based fluids.
To prevent possible damage to the Product, do not use
solvents or abrasive cleansers.
For safe operation and maintenance of the product:
• Have an approved technician repair the Product.
• Do not allow water to get inside the Product.
Replace the Line Power Fuse
The Product has a fuse that protects from overcurrent. Each voltage selection
requires a specific fuse. See Table 2 for the correct fuse for each of the four linevoltage selections. This fuse is located on the rear panel.
To prevent possible electrical shock, fire, or personal injury,
use only specified replacement parts.
Voltage Selector Fuse Fluke Part Number
100 V 0.25 A, 250 V (slow blow) 166306
120 V 0.25 A, 250 V (slow blow) 166306
WCaution
WCaution
XWWarning
Table 2. Fuses
16
220 V 0.160 A, 250 V (slow blow) 4394437
240 V 0.160 A, 250 V (slow blow) 4394437
HYDRA Series III Data Acquisition Unit
General Maintenance
To replace the fuses, see Figure 1:
1. Remove any high-capacity modules or test leads from the Product where
hazardous voltage may be present.
2. Disconnect the mains-power cord from the power-entry module.
3. Insert a small, flat screwdriver blade into the narrow recess to the left of the
fuse holder and pry to the right until the holder pops out. The Product is
shipped with a replacement fuse of the same rating as the fuse installed in
the fuse block.
4. Replace the fuse with the replacements as listed in Table 2.
5. Slide the fuse holder back into the Product until it locks into place.
12
Battery Replacement
The battery in the Product is used for Product memory (menu settings, readings,
or other functions).
To replace the battery:
1. Disassemble the Product as described in the Disassembly section.
2. Remove the battery (see Figure 9) and replace with one rated appropriately
for the selected voltage button battery.
3. Reassemble the Product.
Figure 1. Fuse Replacement
hcn027.eps
17
2638A
Calibration Manual
Memory and Factory Reset
The Product has two memory reset functions to remove data from the memory
and reset the Product:
• Clear all Files
• Factory Reset
See Table 3 for a comparison of these functions.
Note
All memory reset functions require the Admin password.
Table 3. Memory Clear Functions
Task Clear All Files Factory Reset
Deletes Test Setup files, DMM Data files, and Scan
Data files from the internal memory.
Clears the configuration of the Channel Setup, Test
Setup, and Instrument Setup
[1] Does not remove data from the USB drive.
[2] Does not reset the MAC address, the serial number, calibration, clock time, or the Admin or User passwords.
[1]
[2]
•
•
To clear all files:
1. Push .
2. Push twice.
3. Enter the Admin password.
4. Push .
5. Push to confirm.
To reset the Product to factory settings:
1. Push .
2. Push .
3. Enter the Admin password, then push .
18
HYDRA Series III Data Acquisition Unit
Firmware Update
Firmware Update
For Products with a firmware version 1.03 or higher, firmware may be updated by
the product administrator with the front-panel USB port. New firmware versions
are available from the Fluke Calibration web page for the Product under the
Knowledge and Information tab.
For Products that have firmware versions below 1.03, contact Fluke Calibration
Service to schedule a firmware update as these versions do not support the USB
firmware update process.
Update instructions for Products that use version 1.03 and above:
WCaution
• Do not cycle the power or remove the USB device while the
firmware update is in process. These actions will cause an
instrument failure during the update process and require a
return to a Fluke Service center to correct.
• The Product automatically re-boots when the update
process is complete. Updates can take several minutes to
complete and the display screen will not illuminate until the
re-boot.
Note
Fluke Calibration recommends that you move all internal data files
and setup files from the Product to a USB device before the update
of the firmware.
Before the update process starts, verify that the current firmware is version 1.03
or above by viewing Firmware under . If your firmware version is below
1.03, do not continue the update process. You must return the Product to Fluke
Calibration Service for the firmware update.
Firmware update steps:
1. Download and extract the firmware update files (AuxInfo.txt and mtv.flt) to
your computer from the Fluke Calibration website.
2. Create a new folder firmware on a USB flash drive (1 GB or greater), and
then copy these files into this folder. To prevent file corruption, use the
Windows utility to safely eject your USB flash drive from your computer.
3. Plug the USB flash drive into the Product and wait for the red light on the
USB port to illuminate and show that the drive is recognized.
Note
Some USB flash drives may not be recognized by the Product. Test
the USB flash drive prior to starting this update process to ensure it
is recognized.
4. Push to enter the Setup menu.
5. Push the Up/Down arrow button to select Firmware then push to enter
the firmware menu.
6. Push and enter the administrator password and then push to start
the update process.
7. WAIT until the Product automatically re-boots. There will be no screen display
for several minutes during this time. Do not cycle the power or remove the
USB device while the firmware update is in process.
19
2638A
Calibration Manual
Required Equipment
Function Instrument Type Model Comments
Volts DC
8. After the automatic re-boot, push to verify that the update was
successful. Verify the firmware version that you have installed is displayed on
the screen.
Table 4 lists the equipment required for performance tests and calibration
adjustment of the Product.
Table 4. Required Test Equipment
Standard Fluke 5720A
8½ digit meter Fluke 8508A
4-wire short
Alternate standard
Standard Fluke 5720A
[1]
Fluke low thermal 4-wire
short or equivalent
Fluke 5522A
Used to characterized the
5522A
Fluke PN 2653346
Must be characterized with
8508A
Volts ac
Frequency
Ohms
8½ digit meter Fluke 8508A
4-wire short
Alternate standard
Standard Fluke 5522A
Alternate standard Function generator
Standard Fluke 5720A
4-wire short
Alternate standard
[1]
Fluke low thermal 4-wire
short or equivalent
Fluke 5522A
Fluke low thermal 4-wire
short or equivalent
The standard resistors
which have equivalent
spec with Fluke 5720A.
Used to characterized the
5522A
Fluke PN 2653346
Must be characterized with
8508A
Specifications include
0.075 % frequency
accuracy from 20 Hz to
40 Hz and 0.0025 %
accuracy for frequencies
up to 1 MHz
Fluke PN 2653346
20
HYDRA Series III Data Acquisition Unit
Required Equipment
Table 4. Required Test Equipment (cont.)
Function Instrument Type Model Comments
PRT
Thermistor
Current DC
Current ac
Fluke 5522A
(for 1 kΩ value)
Standard
Standard resistors of
25 Ω, 100 Ω, 200 Ω, and
400 Ω
4-wire short
Fluke low thermal 4-wire
short or equivalent
The standard resistors
Alternate standard
which have equivalent
specification.
Standard Fluke 5522A
8½ digit meter Fluke 8508A
4-wire short
Alternate standard
[1]
Fluke low thermal 4-wire
short or equivalent
Fluke 5720A
Standard Fluke 5522A
8½ digit meter Fluke 8508A
Alternate standard
[1]
Fluke 5720A
Fluke 5522A
Standard
Fluke 5720A (for 10 μA
adjustment)
8½ digit meter Fluke 8508A
Resistance accuracy
specification (k = 2) should
be or better than:
40 ppm (for 25 Ω)
20 ppm (for 100 Ω, 200 Ω
and 400 Ω)
Fluke PN 2653346
Must be characterized with
8508A
Used to characterized the
5522A
Fluke PN 2653346
Must be characterized with
8508A
Used to characterized the
5522A
Fluke 5522A must be
characterized with 8508A
Used to characterized the
5522A
Alternate standard Fluke 5720A
Metrology Drywell
Reference thermistor
probe
Fluke 9171 with insert
block
Fluke 5610-9
Accuracy 0.013 °C or
better at 25 °C
2638A-100
Module
Reference
thermometer
E-type
thermocouples
Fluke 1586A Only for CJC verification
Must be calibrated.
Omega TT-E-24-SLE
Accuracy 0.026 °C or
better at 25 °C
To reduce the possibility of inducing errors with ac signals picked up by the test
Cables
leads, use short, shielded twisted-pair PTFE-insulated test cables between the test
equipment and the Meter. Fluke makes a 2 foot (PN 738716) and 4 foot (PN
738724) PTFE insulated test cable for this purpose
[1] Other alternate standards beside those listed can be used as long as they provide sufficient traceable [Test
uncertaintity Ratios (TURs)] at each calibration and verification point.
21
2638A
Calibration Manual
Test Considerations
For optimum performance, all test procedures should comply with these
recommendations:
• Assure the calibration ambient temperature (T
and 28 °C. Ideally, the calibration should be done at 23 °C ±2 °C.
• Assure ambient relative humidity is <80%.
• Allow a 60-minute warm-up period.
• Use shielded twisted-pair PTFE-insulated cables to reduce settling and noise
errors.
• Keep all input cables as short as possible.
• Ensure that the calibration standards and test procedures used do not
introduce additional errors.
Ideally, the standards used to verify and adjust the Product should
be four times more accurate than each full-scale error specification
of the Product.
• User the Fluke low-thermal 4-Wire short for all voltages and Ohm shorts. See
Table 4 for Fluke part number.
Performance Tests
This section provides performance tests to verify that the Product operates within
published specifications as well as a complete calibration procedure. The
performance test and, if necessary, the calibration adjustment procedure can be
done both periodically and after service or repair.
Note
) is stable and between 18 °C
cal
The performance tests can be used as an acceptance test upon receipt of the
Product. Use the 90-day specifications when an acceptance test is done after the
Product is calibrated.
Fluke provides these performance tests to ensure that the Product is in proper
operating condition. If the Product fails any of the performance tests, calibration
adjustment and/or repair are needed. The performance test works best if done in
the sequence shown in Table 5.
Each of the measurements listed in the following tests assumes the Product is
being tested after a one-hour warm-up in an environment with an ambient
temperature of 18 °C to 28 °C and a relative humidity of <80 %.
Note
All instrument settings for verification use power up conditions
except as noted by the verification step.
22
HYDRA Series III Data Acquisition Unit
Performance Tests
Front Panel Test
To test the keypad and LEDs on the front panel:
1. Push Instrument Setup.
2. Push and hold for 3 seconds to enter keypad diagnostic.
3. Push every key. The key name should be shown correctly on the screen.
4. Push , , , and again, to check the LED.
5. Push to exit the keypad diagnostic program.
DC Volts Verification
Connect the Product to the test equipment as shown in Figure 2 and apply the
voltages listed in Table 5.
Note
For the 0 V tests, use the 4-wire short to short the Hi/Lo and Sense
inputs.
1 M Ω1 M 1.00009 MΩ 0.99991 MΩ 1.00011 MΩ 0.99989 MΩ
0 Ω10 M 0.0001 MΩ -0.0001 MΩ 0.0001 MΩ -0.0001 MΩ
10 M Ω10 M 10.0021 MΩ 9.9979 MΩ 10.0041 kΩ 9.9959 MΩ
[1]
0 Ω
100 M 0.01 MΩ -0.01 MΩ 0.01 MΩ -0.01 MΩ
100 M Ω
[1] 2-Wire
[1]
100 M 100.81 MΩ 99.19 MΩ 100.81 MΩ 99.19 MΩ
27
2638A
Calibration Manual
PRT Verification Steps
Connect the Product as in Figure 5 for the 1 kΩ value listed in Table 9. For all
other values in the table, use standard resistors. Configure the Ch001 as 4-wire
PRT function with resistance display.
UUT
NC
NC
Figure 5. PRT Verification Connection
Configure Resistance Display
After configuring the channel Ch001 to 4-wire PRT function, send the following
command through the remote interface to configure the resistance display:
TEMP:FPRT:CALC:RES ON,(@1)
Table 9. PRT Verification Steps
Nominal
Input
Range
1 ke
5522A Calibrator
hcn312.eps
1-Year Test Limits
High Low
0 Ω 100 0.005 Ω -0.005 Ω
25 Ω 100 25.0075 Ω 24.9925 Ω
100 Ω 100 100.015 Ω 99.985 Ω
1k Ω1 k 1.0002 kΩ 0.9998 kΩ
28
HYDRA Series III Data Acquisition Unit
Performance Tests
Thermistor Verification Steps
Connect the Product to the test equipment as shown in Figure 6 and apply the
resistance listed in Table 10. Configure Ch001 as a 4-wire thermistor function
with resistance display.
UUT
NC
NC
Figure 6. Thermistor Test Equipment Setup
Configure Resistance Display
After configuring the channel Ch001 to 4-wire thermistor function, send this
command through the remote interface to configure the resistance display. Apply
the values in Table 10:
TEMP:FTH:CALC:RES ON, (@1)
Table 10. Thermistor Verification Steps
Nominal
Input
Range
5522A Calibrator
hcn304.eps
1-Year Test Limits
High Low
0 Ω 2.2 k 0.5 Ω -0.5 Ω
2 kΩ
90 kΩ 98 k 90010 Ω 89990 Ω
900 kΩ 1 M 900370 Ω 899630 Ω
2.2 k 2000.7 Ω 1999.3 Ω
29
2638A
Calibration Manual
DC Current Verification
Connect the Product to the test equipment as shown in Figure 7 and apply the
resistance listed in Table 11.
Note
When the 5720A/5522A is used as source, connect Hi (source) to Lo
(Product) and Lo (source) to mA (Product). For the 0 A tests, open
Lo inputs.
1 -0.999839 mA -1.000161 mA -0.999839 mA -1.000161 mA
0 10 0.35 μA -0.35 μA 0.35 μA -0.35 μA
[1]
10
10 10.00185 mA 9.99815 mA 10.00185 mA 9.99815 mA
[1]
-10
10 -9.99815 mA -10.00185 mA -9.99815 mA -10.00185 mA
0 100 3.5 μA -3.5 μA 3.5 μA -3.5 μA
[1]
100
100 100.0185 mA 99.9815 mA 100.0185 mA 99.9815 mA
[1]
-100
100 -99.9815 mA -100.0185 mA -99.9815 mA -100.0185 mA
[1] 5720A or 5522A must be used with 8508A to obtain suitable test uncertainty ratio.
31
2638A
Calibration Manual
AC Current Verification Steps
Connect the Meter to the test equipment as shown in Figure 8 and, depending on
which meter you are calibrating, apply the resistance listed in Table 12.
Note
When use 5720A/5522A as source, connect Hi (source) to Lo
(meter) and Lo (source) to mA (meter). For the zero (0) A tests, open
Lo inputs.
5522A Calibrator
CALIBRATOR
5520A
2638A HYDRA Series
UUT
X
EX SNS
: OFF
A.
NC
EX GRD
: OFF
Figure 8. mA AC Current Equipment Setup
HI
LO
HI
CURRENT
OUTPUT
V A
AUX
NORMALAU X
V, , ,
20V PK MAX
RTD
GUAR D
A, -SENSE,
HI
LO
AUX V
FUSED
20A
20V PK MAX
TC
5720A Calibrator
SENSE
V
WIDEBAND
HI
LO
GUARD GROUND
SCOPE
OUT
TRIG
hcn311.eps
32
HYDRA Series III Data Acquisition Unit
Performance Tests
Table 12. AC Current Verification Steps
Nominal Input
Ampl. Frequency High Low High Low
100.0 μA
100.0 μA
1.0 mA
1.0 mA
10.0 mA
10.0 mA
100.0 mA
100.0 mA
[1] 5522A must be used with 8508A to obtain suitable test uncertainty ratio.
[1]
20 Hz 0.1 100.31 µA 99.69 µA 100.36 µA 99.64 µA
[1]
2 kHz 0.1 100.31 µA 99.69 µA 100.36 µA 99.64 µA
[1]
20 Hz 1 1.0031 mA 0.9969 mA 1.0036 mA 0.9964 mA
[1]
2 kHz 1 1.0031 mA 0.9969 mA 1.0036 mA 0.9964 mA
[1]
20 Hz 10 10.031 mA 9.969 mA 10.036 mA 9.964 mA
[1]
2 kHz 10 10.031 mA 9.969 mA 10.036 mA 9.964 mA
[1]
20 Hz 100 100.31 mA 99.69 mA 100.36 mA 99.64 mA
[1]
2 kHz 100 100.31 mA 99.69 mA 100.36 mA 99.64 mA
Range
90-day Test Limits 1-year Test Limits
2638A-100 Module Verification
1. Connect the calibrated E-type thermocouples to channel 10 on the Product.
2. Insert the thermocouples into a drywell calibrator which is set and stabilized
at 25 °C.
3. Configure the channels of the Product for an E-thermocouple.
4. Insert the reference thermistor probe into the drywell to measure the actual
temperature.
5. Use the reference thermometer with the reference thermistor probe and the
thermocouples on the Product to measure the drywell temperature until all
readings are stable.
6. Compare the thermocouple readings and reference thermometer reading.
The difference should be <0.6 °C.
33
2638A
Calibration Manual
Calibration Adjustment
Calibration adjustment should be done at the specified interval, or whenever a
verification test indicates that a Product function is out of tolerance. Product
accuracy stays within specifications only if the adjustment procedure is done at
regular intervals. A one-year interval is adequate for most applications. Product
accuracy specifications are not valid beyond the one-year interval.
Adjustments are accessed through both the remote interface and front panel with
a series of adjustment steps. The remote program directs the test equipment to
apply a series of shorts, opens, voltages, currents, and resistances to the
Product. At each step, the Product internally makes the necessary adjustment to
bring the Product into specification. No internal mechanical adjustments are
necessary.
With an automated, computer-controlled procedure, the calibration and
verification procedures can be done in under an hour. A simple adjustment
program is listed in the Sample Adjustment Program section.
Adjustments are password protected to prevent accidental or unauthorized
adjustments. The admin password must be entered through the front panel or
remote interface.
Unlock the Product
To unlock the Product for adjustments from the front panel:
1. Push Instrument Setup.
Push Calibrate.
2.
3. Use the numeric keypad to enter the 4-digit admin password.
4.
Push OK to
The Product is shipped from the factory with the password set to 2638.
Unlock the Product with a Remote Interface
To unlock the Product with a remote interface, send the command:
CAL:SEC:STAT OFF,“2638”
To relock the Product, send the command:
CAL:SEC:STAT ON
enter the password and continue the adjustments procedure.
34
HYDRA Series III Data Acquisition Unit
Calibration Adjustment
Reset the Admin Password
If the admin password is lost or forgotten, the password can be reset to 2638 with
these steps:
Note
Before doing these steps, try to use the factory default password:
2638.
1. Do the general disassembly steps in the “Disassembly” section.
2. Connect a jumper across J8, as shown in Figure 9.
3. Reconnect the mains power cable between the Product and a power outlet.
4. Turn the Product on.
5. When the Product is started, the password will automatically be reset to
2638.
6. Turn the Product off and disconnect the mains power cable.
7. Remove the jumper connected in step 2.
8. Reassemble the Product by doing the reverse procedure in “Disassembly”.
Display
Cable
InGuard
Shield
Short J8, then turn on the power.
After the instrument is successfully
started, the admin password is
reset to 2638.
Battery
J8
RST PSW
A6
PCB
Assembly
Fan
Figure 9. Password Reset Jumper
hcn103.eps
35
2638A
Calibration Manual
Disassembly
XWWarning
To prevent possible electrical shock or personal injury,
disconnect the mains power cord before you remove the
Product covers.
Only disassemble the Product to reset the password or replace the battery. See
Figure 10.
A 2# Phillips screwdriver and small crescent wrench are required to disassemble
the Product.
1. Remove all high-capacity modules and test leads from the Product.
2. Turn off the mains power at the rear of the Product and remove the mains
power cord. The front panel power key only puts the Product into a powersave mode and does not remove mains power.
3. Remove the boots by pulling from a corner and stretching the bumpers off
the Product.
4. Remove the bail by rotating the handle upright to a 90 ° angle from the top
of the Product and pulling the bail out from the sides.
5. Remove the top cover by removing the two screws on the sides of the
chassis, and slide the cover towards the back of the Product .
36
HYDRA Series III Data Acquisition Unit
Calibration Adjustment
1
Boot Removal
3
To Remove, Pull End From Unit
2
4
6
5
Battery
hcn309.eps
Figure 10. Disassembly
37
2638A
Calibration Manual
Change the Calibration Date Remotely
“Mainframe” Adjustment Process
The calibration date is automatically updated when the “CALibrate:STORe”
command is sent.
Table 13 lists:
• Steps
• Description of the adjustment
• Measurement adjustment type (open, zero, or gain adjustment)
• Product value or range that is to be adjusted
• Amplitude of the adjustment signal
• The frequency of the adjustment signal (if necessary)
Table 13. Adjustment Steps
Step
MAIN
A2D open ADC self-calibration Y
ACV
AC100MV 0.1
AC_100MV_1 100.0 mV, 1 kHz N
AC_50MV_1 50.0 mV, 1 kHz N
AC_25MV_1 25.0 mV, 1 kHz N
AC_5MV_1 5.0 mV, 1 kHz N
AC_0V_1 0.0 V N
AC1V 1.0
AC_1V_2 1.0 V, 1 kHz N
AC_0_5V_2 0.5 V, 1 kHz N
AC_0_25V_2 0.25 V, 1 kHz N
Value
Range
Input Signal Description
Start calibration of
ac 100 mV range
Start calibration of
ac 1 V range
Y
Y
Entry
Point
38
AC_0_05V_2 0.05 V, 1 kHz N
AC_0V_2 0.0 V N
AC_1V_LF 1 V, 20 Hz
AC10V 10.0
AC_10V_3 1.0 V, 1 kHz N
AC_5V_3 5.0 V, 1 kHz N
AC LF
compensation
Start calibration of
ac 10 V range
N
Y
HYDRA Series III Data Acquisition Unit
Calibration Adjustment
Table 13. Adjustment Steps (cont.)
Step
AC_2_5V_3 2.5 V, 1 kHz N
AC_0_5V_3 0.5 V, 1 kHz N
AC_0V_3 0.0 V N
AC100V 100.0
AC_100V_4 100.0 V, 1 kHz N
AC_50V_4 50.0 V , 1 kHz N
AC_25V_4 25.0 V, 1 kHz N
AC_5V_4 5.0 V, 1 kHz N
AC_0V_4 0.0 V N
AC300V 300.0
AC_300V_5 300.0 V, 1 kHz N
AC_200V_5 200.0 V, 1 kHz N
AC_100V_5 100.0 V, 1 kHz N
Value
Range
Input Signal Description
Start calibration of ac
100 V range
Start calibration of ac
300 V range
Y
Y
Entry
Point
AC_50V_5 50.0 V, 1 kHz N
AC_0V_5 0.0 V N
DCV
DC100MV 0.1
DC_P100MV_1 100.0 mV N
DC_0V_1 0.0 V N
DC_N100MV_1 -100.0 mV N
DC1V 1.0
DC_P1V_2 1.0 V N
DC_0V_2 0.0 V N
DC_N1V_2 -1.0 V N
DC10V 10.0
DC_P10V_3 10.0 V N
DC_0V_3 0.0 V N
Start calibration of dc
100 mV range
Start calibration of dc 1 V
range
Start calibration of dc
10 V range
Y
Y
Y
39
2638A
Calibration Manual
Table 13. Adjustment Steps (cont.)
Step Value Range
DC_N10V_3 -10.0 V N
DC100V 100.0
DC_P100V_4 100.0 V N
DC_0V_4 0.0 V N
DC_N100V_4 -100.0 V N
DC300V 300.0
DC_P300V_5 300.0 V N
DC_0V_5 0.0 V N
DC_N300V_5 -300.0 V N
OHM
R100 100.0
R_100_1 100.0 Ω N
Input
Signal
Start calibration of dc 100 V
range
Start calibration of dc 300 V
range
Start calibration of FRES
100 Ω range
Description
Entry
Point
Y
Y
Y
R_19_1 19.0 Ω N
R_0_1 0.0 Ω N
R1K 1.0E+3
R_1K_2 1.0 kΩN
R_190_2 190.0 Ω N
R_0_2 0.0 Ω N
R10K 10.0E+3
R_10K_3 10.0 kΩN
R_1_9K_3 1.9 kΩN
R_0_3 0.0 Ω N
R100K 100.0E+3
R_100K_4 100.0 kΩN
R_19K_4 19.0 kΩN
Start calibration of FRES
1 kΩ range
Start calibration of FRES
10 kΩ range
Start calibration of FRES
100 kΩ range
Y
Y
Y
40
HYDRA Series III Data Acquisition Unit
Calibration Adjustment
Table 13. Adjustment Steps (cont.)
Step Value Range
R_0_4 0.0 Ω N
R1M 1.0E+6
R_1M_5 1.0 MΩN
R_190K_5 190.0 kΩN
R_100K_5 100.0 kΩN
R_0_5 0.0 Ω N
R10M 10.0E+6
R_10M_6 10.0 MΩN
R_1_9M_6 1.9 MΩN
R_1_0M_6 1.0 MΩN
R_0_6 0.0 Ω N
R100M 100.0E+6
Input
Signal
Start calibration of FRES
1 MΩ range
Start calibration of RES
10 MΩ range
Start calibration of RES
100 MΩ range
Description
Entry
Point
Y
Y
Y
R_100M_7 100.0 MΩN
R_19M_7 19.0 MΩN
R_0_7 0.0 Ω N
THERM
T2K 2.2E+3
T_1_9K_1 1.9 kΩN
T_1_0K_1 1.0 kΩN
T_190_1 190.0 Ω N
T_0_1 0.0 Ω N
T90K 90.0E+3
T_100K_2 100.0 kΩN
T_19K_2 19.0 kΩN
T_10K_2 10.0 kΩN
T_0_2 0.0 Ω N
Start calibration of FTH
2 kΩ range
Start calibration of FTH
90 kΩ range
Y
Y
T1M 90.0E+3
Start calibration of FTH
1 MΩ range
Y
41
2638A
Calibration Manual
Table 13. Adjustment Steps (cont.)
Step Value Range
T_1M_3 1.0 MΩN
T_190K_3 190.0 kΩN
T_100K_3 100.0 kΩN
T_0_3 0.0 Ω N
HZ
HZ1K
HERTZ
ACI
AC100UA 100.0E-6
AC_100UA_1
AC_50UA_1
Input
Signal
1000.0 Hz,
1 V
100.0 μA,
500 Hz
50.0 μA,
500 Hz
Description
Start calibration of
frequency
N
Start calibration of ac
100 μA range
N
N
Entry
Point
Y
Y
AC_25UA_1
AC_10UA_1
AC_0A_1 Open N
AC1MA 1.0E-3
AC_1MA_2
AC_0_5MA_2
AC_0_25MA_2
AC_0_05MA_2
AC_0A_2 Open N
AC10MA 10.0E-3
AC_10MA_3
25.0 μA,
500 Hz
10.0 μA,
500 Hz
1.0 mA,
500 Hz
0.5 mA,
500 Hz
0.25 mA,
500 Hz
0.05 mA,
500 Hz
10.0 mA,
500 Hz
N
N
Start calibration of ac 1 mA
range
N
N
N
N
Start calibration of ac
10 mA range
N
Y
Y
42
HYDRA Series III Data Acquisition Unit
Calibration Adjustment
Table 13. Adjustment Steps (cont.)
Step Value Range Input Signal Description
AC_5MA_3 5.0 mA, 500 Hz N
AC_2_5MA_3 2.5 mA, 500 Hz N
AC_0_5MA_3 0.5 mA, 500 Hz N
AC_0A_3 Open N
AC100MA 100.0E-3
AC_100MA_1
AC_50MA_1 50.0 mA, 500 Hz N
AC_25MA_1 25.0 mA, 500 Hz N
AC_5MA_1 5.0 mA, 500 Hz N
AC_0A_1 Open N
DCI
DC100UA 100.0E-6
100.0 mA,
500 Hz
Start calibration of ac
100 mA range
N
Start calibration of dc
100 μA range
Entry
Point
Y
Y
DC_P100UA_1 100.0 μA N
DC_0A_1 Open N
DC_N100UA_1 -100.0 μA N
DC1MA 1.0E-3
DC_P1_0MA_2 1.0 mA N
DC_0A_2 Open N
DC_N1_0MA_2 -1.0 mA N
DC10MA 10.0E-3
DC_P10MA_3 10.0 mA N
DC_0A_3 Open N
DC_N10MA_3 -10.0 mA N
DC100MA 100.0E-3
DC_P100MA_4 100.0 mA N
DC_0A_4 Open N
Start calibration of dc
1 mA range
Start calibration of dc
10 mA range
Start calibration of dc
100 mA range
Y
Y
Y
DC_N100MA_4 -100.0 mA N
43
2638A
Calibration Manual
Remote Commands for Calibration
Table 14 alphabetically lists the command set for calibration.
Table 14. List of Commands
Remote Command Meaning
CALibrate:ABORt
CALibrate:BACKup
CALibrate:CONSt?
CALibrate:DATE?
CALibrate:INFO?
CALibrate:MOD:DATE
<slot>,<year>,<month>,<day>
CALibrate:NEXT [<reference>] Continue a calibration procedure if it is waiting.
CALibrate:REF? Return nominal value expected for reference entry.
CALibrate:SECure:STATe
<boolean>,<admin_password>
CALibrate:SECTion Skip to next section of calibration procedure.
CALibrate:SKIP Skip to next entry point in calibration procedure.
CALibrate:STARt <procedure>[,<step>] Start a calibration procedure.
Instruct unit to abort calibration procedure after
present step.
Backup to previous entry point in calibration
procedure.
Retrieve the value in use of the given calibration
constant.
Return a CAL date associated with stored
calibration constants.
Return message or instructions associated with
running step.
Set CAL date of module in specified slot.
Instruct unit to enable calibration.
44
CALibrate:STATe? Return state of calibration.
CALibrate:STEP? Return name of step currently running.
CALibrate:STORe Store new calibration constants.
CALibrate:MODule:DATE
<slot>,<year>,<month>,<day>
Set module CAL date.
HYDRA Series III Data Acquisition Unit
Remote Commands for Calibration
Remote Programming Examples
This section gives examples of command sequences for several likely scenarios.
Start a Full Calibration
Table 15 shows an example to run a full calibration.
Table 15. Full Calibration Example
Command Action
CAL:SEC:STAT OFF,”2638” Disable the security for calibration.
CAL:STAR MAIN
CAL:NEXT Continue to show instruction of ACV calibration.
CAL:NEXT
CAL:NEXT
CAL:NEXT 0.1
CAL:NEXT 0.05
CAL:NEXT 0.025
CAL:NEXT 0.005
Start MAIN calibration procedure, and show instruction of MAIN
calibration procedure.
Continue to perform ACV calibration.
After it is completed, show instruction to connect the calibrator to
the instrument for ACV calibration.
Show instruction to ask for 100 mV/1 kHz signal input, and the
reference value.
Continue to perform measurements for this point, after it is
completed, show instruction to ask for 50 mV/1 kHz signal input
and the reference value.
Continue to perform measurements for this point, after it is
completed, show instruction to ask for 25 mV/1 kHz signal input
and the reference value.
Continue to perform measurements for this point, after it is
completed, show instruction to ask for 5 mV/1 kHz signal input
and the reference value.
Continue to perform measurements for this point, after it is
completed, show instruction to ask for 0V signal input and the
reference value.
Continue to perform measurements for this point, after it is
completed, calculate the calibration constants for ac 100 mV
CAL:NEXT 0.0
……
CAL:STOR Store the calibration constants.
CAL:ABOR Abort the calibration procedure.
range, then show instruction to ask for 1 V/1 kHz signal input and
the reference value (the first calibration point of next ac 1 V
range.
45
2638A
Calibration Manual
Calibrate 1 V DC Range only
CAL:SEC:STAT OFF,”2638” Disable the security for calibration.
CAL:STAR MAIN,DC1V
CAL:NEXT 1
CAL:NEXT 0
CAL:NEXT -1
Table 16 shows an example to run a calibration for 1 V dc.
Table 16. 1 V DC Calibration Example
Command Action
Start MAIN calibration procedure, and jump to 1 V dc range
directly, it will show ask for 1 V signal input, and the reference
value.
Continue to perform measurements for this point, after it is
completed, show instruction to ask for 0 V signal input and the
reference value.
Continue to perform measurements for this point, after it is
completed, show instruction to ask for -1 V signal input and the
reference value.
Continue to perform measurements for this point, after it is
completed, calculate the calibration constants for DC 1V range,
then show instruction to ask for 10 V signal input and the
reference value (the first calibration point of next 10 V dc range.
CAL:STOR Store the calibration constants.
CAL:ABOR Abort the calibration procedure.
Write Calibration Date to a Module
Table 17 shows an example to write the calibration date into a module after the
CJC accuracy is validated.
Table 17. Write Calibration Date to a Module Example
Command Action
CAL:SEC:STAT OFF,”2638” Disable the security for calibration.
CAL:MOD:DATE 1,2014,11,1 Write “2014/11/1” as new calibration date into module in slot 1.
46
HYDRA Series III Data Acquisition Unit
Remote Commands for Calibration
Command References
CALibrate:ABORt
Description: Instruct unit to abort calibration procedure after present
step.
Example: CAL:ABOR
Related Commands:
CALibrate:STATe?
CALibrate:BACKup
Description: Backup to previous entry point in calibration procedure.
Example: CAL:BACK
Related Commands:
CALibrate:SKIP
CALibrate:SECTion
CALibrate:CONSt? “<cco_name>”
Description: Retrieves the value in use of the given calibration constant.
Example: CAL:CONS? “DC100MV_A1”
Response: -1.401686110719e-04
Related Commands:
CALibrate:SECure:STATe CALibrate:STORe
CALibrate:DATE?
Description: Return a CAL date associated with stored calibration
constants.
Example: CAL:DATE? Response: 2013,11,1
Related Commands:
CALibrate:MODule:DATE?
47
2638A
Calibration Manual
CALibrate:INFOrmation?
Description: Return message or instructions associated with
running step.
Example: CAL:INFO? Response: “Connect calibrator to Volt terminals”
Related Commands:
CALibrate:STATe?
Description: Skip to next section of calibration procedure.
Example: CAL:SECT
Related Commands:
CALibrate:BACKup
CALibrate:SKIP
CALibrate:SKIP
Description: Skip to next entry point in calibration procedure.
Example: CAL:SKIP
Related Commands:
CALibrate:BACKup
CALibrate:SECTion
49
2638A
Calibration Manual
CALibrate:STARt <procedure>[,<step>]
Description: Start a calibration procedure. As parameter, the name
of procedure should be provided (MAIN is the
procedure for full instrument calibration), an optional
parameter <step> can be provided to start from, if it is
omitted, it starts at the beginning. Before any
calibration procedure can be started, the calibration
secure state should be disabled.
Example: CAL_START MAIN
CAL_START MAIN,DC1V
Related Commands:
CALibrate:SECure:STATe
CALibrate:STATe?
Description: Return state of calibration.
Example: CAL:STAT?
Response: RUN - running a calibration step
REF - waiting for a CAL_NEXT with reference
(measurement) value
INS - instruction available, waiting for a CAL_NEXT NOT - not in a calibration procedure (or at end of one)
Related Commands:
CALibrate:STEP? CALibrate:REFerence?
CALibrate:INFOrmation?
50
HYDRA Series III Data Acquisition Unit
Remote Commands for Calibration
CALibrate:STEP?
Description: Return name of step currently running.
Example: CAL:STEP?
Response: DC1V
Related Commands: