Fluke Calibration Fluke Calibration 1523-P4-256 Manual

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
1523, 1524
Thermometer Readout
PN 4427817 March 2014 © 2014 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.
Table of Contents
1 Before You Start .......................................................................1
1.1 Introduction ............................................................................................... 1
1.2 Standard Equipment ................................................................................. 1
1.3 Safety Information ..................................................................................... 2
1.3.1 Warning .....................................................................................................2
1.3.2 Cautions ..........................................................................................................3
1.4 CE Comments ........................................................................................... 5
1.4.1 EMC Directive .................................................................................................5
1.4.2 Immunity Testing .............................................................................................5
1.5 Using Clamp-On Ferrites .......................................................................... 5
1.6 Emissions Testing ..................................................................................... 6
1.7 Low Voltage Directive (Safety) .................................................................. 6
1.8 Authorized Service Centers ...................................................................... 7
2 Quick Start ................................................................................9
2.1 Setup ......................................................................................................... 9
2.2 Specifications ......................................................................................... 24
i
1523, 1524 Thermometer Readout
Figures
Figure 1 Clamp-On Ferrite ................................................................................. 6
Figure 2 Input/Output Connections - 1523 ........................................................ 9
Figure 3 Input/Output Connections - 1524 ...................................................... 10
Figure 4 Keys ................................................................................................... 11
Figure 5 1523 Menu ......................................................................................... 14
Figure 6 1523 Menu (cont) ............................................................................. 15
Figure 7 1523 Menu (cont) ............................................................................. 16
Figure 8 1524 Menu ........................................................................................ 19
Figure 9 1524 Menu (cont) ............................................................................. 20
Figure 10 1524 Menu (cont) ........................................................................... 21
Figure 11 1524 Menu (cont) ........................................................................... 22
Figure 12 1524 Menu (cont) ........................................................................... 23
Figure 13 1524 Menu (cont) ........................................................................... 24
ii
Tables
Table 1 International Symbols ............................................................................ 4
Table 2 1523 Input/Output Connections ............................................................ 9
Table 3 1524 Input/Output Connections .......................................................... 10
Table 4 1523 Key Functions ............................................................................. 12
Table 5 1524 Key Functions ............................................................................. 17
Table 6 General Specifications ........................................................................ 24
Table 7 Millivolt Measurement .......................................................................... 25
Table 8 Ohms Measurement, RTDs ................................................................. 25
Table 9 Ohms Measurement, Thermistor ......................................................... 26
Table 10 Temperature, Thermocouples ........................................................... 26
Table 11 Temperature, RTD Ranges, and Accuracies (ITS-90) ....................... 28
Table 12 Temperature, Thermistor ................................................................... 28
iii
Before You Start
Standard Equipment
Before You Start1
Introduction1.1
The Reference Thermometer Readouts (1523, 1524) are designed to be reliable, stable, temperature measur­ing instruments that can be used in the eld or laboratory. They offer accuracy, portability, and speed for near­ly every eld calibration application. The instruments have been designed with the eld user in mind and are easy to use while maintaining stability, uniformity, and accuracy comparable to some laboratory instruments. Your Fluke 1523 and 1524 thermometer readout is a handheld, battery operated instrument that measures tem­perature using Platinum resistance Thermometers (PRT), Thermistors, and Thermocouples (TC).
Standard Equipment1.2
Unpack the instrument carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:

1523/1524 Reference Thermometer Readout with 3 AA batteries

AC Adapter, with power cord

RS-232 Cable

User’s Guide

Documentation CD

Report of Calibration and calibration label

Clamp-on ferrite(s)
If all items are not present, contact an Authorized Service Center. (See Section 1.8, Authorized Service Cen­ters on page 7.)
1
1523, 1524 Thermometer Readout
Safety Information
Safety Information1.3
The Reference Thermometer is designed in accordance with EN 61010-1 {2nd Edition}, and CAN/CSA 22.2 No 61010.1-04. Use this instrument only as specied in this manual, otherwise the protection provided by the instrument may be impaired.
A Warning identies conditions and actions that pose hazard(s) to the user; a Caution identies conditions and actions that may damage the instrument being used.
International symbols used on the reference thermometer and in this manual are explained in Table 1 on page
4.
1.3.1 Warning
To avoid possible electric shock or personal injury:

Do not use the reference thermometer in environments other than those listed in the user’s guide.

Do not use the reference thermometer for any application other than that which is specied. The instrument was designed for temperature measurement and calibration. Any other use of the instrument may cause unknown hazards to the user.

If the reference thermometer is used in a manner not in accordance with the equipment design, the operation and the protection provided by the instrument may be impaired. In addition, safety hazards may arise.

Do not apply more than the rated voltage, as marked on the reference thermometer, between the inputs, or between any input and earth ground (30 V, 24 mA max all terminals).

Follow all equipment safety procedures.

Calibration equipment should only be used by trained personnel.

The reference thermometer is intended for indoor use only.

Before you use the instrument, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors. Do not use the reference thermometer if it
2
Before You Start
Safety Information
appears damaged or operates abnormally. Protection may be impaired. When in doubt, have the instrument serviced.

Always use an isolated RTD or PRT (metal sheath isolated from lead wires).

Make sure the battery door is closed and latched before you operate the reference thermometer.

Do not operate the reference thermometer around explosive gas, vapor, or dust.

For battery operation use only 3 AA batteries, properly installed in the reference thermometer case.

1524 model, thermocouples can only be used on channel 1.
Cautions1.3.2
To avoid possible damage to the reference thermometer or to equipment under test:

Do not apply more than the rated voltage, as marked on the reference thermometer, between the inputs, or between any input and earth ground (30 V 24 mA max all terminals).

Unless recalibrating the instrument DO NOT change the values of the calibration constants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the instrument.

The instrument and any thermometer probes used with it are sensitive instruments that can be easily damaged. Always handle these devices with care. DO NOT allow them to be dropped, struck, stressed, or overheated.

DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty environment.

Use the proper probes, function and range for your measurement.

Ensure probe coefcients are downloaded.
3
1523, 1524 Thermometer Readout
Safety Information
Table 1 International Symbols
Symbol Description Symbol Description
AC (Alternating Current) PE Ground
AC-DC Hot Surface (Burn Hazard)
Battery Read the User’s Guide (Important Information)
Complies with European Union directives Off
DC On
Double Insulated Canadian Standards Association
Electric Shock C-TICK Australian EMC mark
Fuse
4
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) mark.
Before You Start
Using Clamp-On Ferrites
CE Comments1.4
EMC Directive1.4.1
Fluke’s equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 2004/108/EC ). The Declaration of Conformity for your instrument lists the specific stan-dards to which the unit was tested.
The instrument was designed specically as a test and measuring device. Compliance to the EMC direc­tive is through EN 61326-1:2006 Electrical equipment for measurement, control and laboratory use – EMC requirements
As noted in the EN 61326-1, the instrument can have varying congurations. The instrument was tested in a typical conguration with shielded RS-232 cables.
Immunity Testing1.4.2
The instrument was tested to the requirements for laboratory locations.
Using Clamp-On Ferrites1.5
Clamp-on ferrites are provided for use in improving the instrument’s electromagnetic (EM) immunity in envi­ronments of excessive EM interference, like areas of heavy industrial equipment. We recommend placing the ferrites on the cables of probes attached to the instrument.
To attach a ferrite to a probe cable, make a loop in the cable near the connector and clamp the ferrite around half of the loop as shown in the diagram. The ferrite can be easily detached and moved to a new probe when needed. (See Figure 1 on next page.)
5
1523, 1524 Thermometer Readout
clamp-on ferrite
probe cable
Emissions Testing
Figure 1
Clamp-On Ferrite
Emissions Testing1.6
The instrument fullls the limit requirements for Class B.
Low Voltage Directive (Safety)1.7
In order to comply with the European Low Voltage Directive (2006/95/EC), Fluke equipment has been de­signed to meet the EN 61010-1.
6
Before You Start
Authorized Service Centers
7
Quick Start
RS
232
12
V DC
30
V MAX
Setup
Quick Start2
Setup2.1
1523 Input/Output ConnectionsTable 2
No. Name Description
1 Serial Serial interface connector
2 Connector, T1 Sensor Connector, Channel 1
4 Power External Power adapter connection
1
Figure 2 Input/Output Connections - 1523
2
4
9
1523, 1524 Thermometer Readout
RS232
12
V DC
30 V MAX
T1 T2
Setup
1
Figure 3 Input/Output Connections - 1524
2
3
1524 Input/Output ConnectionsTable 3
No Name Description
1 Serial Serial interface connector
2 Connector, T1 Sensor Connector, Channel 1
3 Connector, T2 Sensor Connector, Channel 2
4 Power External Power adapter connection
4
10
Quick Start
Setup
V DC
12
30 V MAX
T1 T2
RS232
Figure 4 Keys
CALIBRATION
THERMOMETER
READOUT
1524
3
2
1
6
TRENDmV
5
RESET
4
HOLDSTATS
F
°
C
°
SETUP
7
HOMELOG
10
ENTER
SAVE
8
12
NEXT
9
RECALL
11
11
1523, 1524 Thermometer Readout
STATS
°C °
F
HOLD
SETUP
SAVE
ENTER
Setup
1523 Key FunctionsTable 4
No Key Description
1
Power on or off
2
3
4
5
6
7
8
9
10
Yellow Second or Special Function Key
Turns the backlight on or off
1st Press: MAX, 2nd Press: MIn, 3rd Press: AVE, 4th Press: STD DEV
Units, °C/°F
1st press - Holds value on screen "-- HOLD --" across bottom of screen. 2nd press - Releases Screen hold.
Enters setup menu, see menu structure
Saves measurement as a logged data point
Arrows increment or decrement selections in an active field. In Graph Mode, Arrows change the scale of the graph.
Selects highlighted selection, Saves a new selection.
12
No Key Description
RECALL
NEXT
STATS
°C °
F
HOLD
ENTER
Quick Start
Setup
11
12
13
+
14
+
15
+
16
+
1st Press - Enters Recall menu, 2nd Press - Exits Recall Menu
Moves down to next option on screen.
“RESET” - Resets Stats Data
mV” - Toggles from °C to or to °C (PRT, thermistor), °C to mV or mV to °C (TC)
“TREND” - Starts Graphing data
“HOME” Returns user to main screen
13
1523, 1524 Thermometer Readout
Setup
Figure 5 1523 Menu
14
Figure 6 1523 Menu (cont)
Quick Start
Setup
15
1523, 1524 Thermometer Readout
Setup
Figure 7 1523 Menu (cont)
16
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
1524 Key FunctionsTable 5
STATS
°C °
F
HOLD
SETUP
SAVE
No Key Description
1
Power on or off
Quick Start
Setup
2
3
4
5
6
7
8
9
Yellow Second or Special Function Key
Turns the backlight on or off
1st Press: Max, 2nd Press: Min, 3rd Press: Ave, 4th Press: STD DEV
Units, °C/°F
1st press - Holds value on screen "-- HOLD --" across bottom of screen. 2nd press - Releases Screen hold.
Enters setup menu, see menu structure
Saves measurement as a logged data point
Arrows increment or decrement selections in an active field. In Graph Mode, Arrows change the scale of the graph.
17
1523, 1524 Thermometer Readout
ENTER
RECALL
NEXT
STATS
°C °
F
HOLD
SAVE
ENTER
Setup
No Key Description
10
11
12
13
+
14
+
15
+
16
+
17
+
Selects highlighted selection, Saves a new selection.
1st press - Enters Recall Menu, 2nd press - Exits Recall Menu
Moves down to next option on screen.
“RESET” - Resets Stats Data
mV” - Toggles from °C to or to °C (PRT, thermistor), °C to mV or mV to °C (TC)
“TREND” - Starts Graphing data
“LOG” - Log a series of measurements, see Auto Log in menu structure
“HOME” Returns user to main screen
18
Figure 8 1524 Menu
Quick Start
Setup
19
1523, 1524 Thermometer Readout
Setup
Figure 9 1524 Menu (cont)
20
Quick Start
Setup
Figure 10 1524 Menu (cont)
21
1523, 1524 Thermometer Readout
Setup
22
Figure 11 1524 Menu (cont)
Quick Start
Setup
Figure 12 1524 Menu (cont)
23
1523, 1524 Thermometer Readout
Specications
Figure 13 1524 Menu (cont)
Specications2.2
Specications are based on a one year calibration cycle and apply from 13 °C to 33 °C unless stated otherwise. All specications assume a ve minute warm up period.
General SpecificationsTable 6
Operating Temperature
Storage Temperature –20 °C to 70 °C
Operating altitude 10,000 meters above mean sea level
Relative Humidity (% RH operating without condensation)
Vibration Random, 2g, 5–500 Hz
–10 °C to 60 °C
0 % to 90 % (non condensing)
24
Quick Start
Specications
Power requirements 3 AA alkaline batteries
Size 96 x 200 x 47 mm
Weight 0.65 kg (1.4 lb)
Safety EN 61010-1:2001, CAN/CSA C22.2 No.
Environmental conditions for all specifications: 13 °C to 33 °C
Millivolt MeasurementTable 7
Range Resolution Accuracy
–10 mV to 75 mV 0.001 mV
Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):
± (0.001 %/°C + 1 µV/°C)
Ohms Measurement, RTDsTable 8
Ohms Range
0 to 400 ± (0.004 % + 0.002 )
Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):
0.0008 %/°C + 0.0004
Excitation Current: 1 mA
12 V dc universal power supply
(3.75 x 7.9 x 1.86 inches)
61010.1-04
± (0.005 % + 5 µV)
Accuracy ± 4 Wire
25
1523, 1524 Thermometer Readout
Specications
Ohms Measurement, ThermistorTable 9
Ohms Range
200 to 50 k ± (0.01 % + 0.5 )
50 k to 500 k
Temperature Coefficient ( –10 °C to 13 °C , +33 °C to 60 °C):
0.002 %/°C + 0.1(0 to 50 k)
0.06 %/°C + 0.1(50 k to 500 k)
Excitation Current:
Equivalent temperature accuracies derived from primary specications (, mV)
Temperature, ThermocouplesTable 10
Type Range
B 600 °C to 800 °C
800 °C to 1000 °C
1000 °C to 1800 °C
C 100 °C to 550 °C
550 °C to 2300 °C
E –200 °C to 0 °C
0 °C to 950 °C
J –200 °C to 0 °C
0 °C to 1200 °C
K –200 °C to 0 °C
0 °C to 1370 °C
Accuracy ± Ω, 4 Wire
± (0.03 %)
10 µA (0 to 50 k) 2 µA (50 k to 500 k)
Measure
Accuracies (ITS-90)
0.85 °C
0.68 °C
0.57 °C
0.32 °C
0.71 °C
0.52 °C
0.22 °C
0.52 °C
0.23 °C
0.61 °C
0.24 °C
26
Quick Start
Specications
Type Range
L –200 °C to 0 °C
0 °C to 900 °C
M –20 °C to 0 °C
0 °C to 400 °C
400 °C to 1400 °C
N –200 °C to 0 °C
0 °C to 1300 °C
R –20 °C to 0 °C
0 °C to 500 °C
500 °C to 1750 °C
S –20 °C to 0 °C
0 °C to 500 °C
500 °C to 1750 °C
T –200 °C to 0 °C
0 °C to 400 °C
U –200 °C to 0 °C
0 °C to 400 °C
Resolution: 0.01 °
Note 1: Accuracies are based on internal Reference Junction Compensation. Refer to Technical manual for equivalent external reference accuracies.
Measure
Accuracies (ITS-90)
0.36 °C
0.23 °C
0.26 °C
0.25 °C
0.22 °C
0.72 °C
0.28 °C
1.09 °C
0.97 °C
0.49 °C
1.05 °C
0.95 °C
0.56 °C
0.60 °C
0.25 °C
0.54 °C
0.24 °C
27
1523, 1524 Thermometer Readout
Specications
Temperature, RTD Ranges, Table 11
and Accuracies (ITS-90)
Accuracy ± °C 4 Wire Probe
± 0.011 at –100 °C
± 0.015 at 0 °C
± 0.019 at 100 °C
± 0.023 at 200 °C
± 0.031 at 400 °C
± 0.039 at 600 °C
Resolution: 0.001°C (0.001°F)
Temperature, ThermistorTable 12
Accuracy ± °C
± 0.002 at 0 °C
± 0.003 at 25 °C
± 0.006 at 50 °C
± 0.014 at 75 °C
± 0.030 at 100 °C
Resolution: 0.001 °C (0.001 °F)
Based on a 10 k (at 25 °C) thermister with a beta value of 4000 . See technical manual for details.
28