Fluke Calibration 9118A Instruction Manual

9118A
Calibration Furnace

Users Manual

September 2012 Rev. 1, 2/2014
© 2012-2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
11/99
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands

Table of Contents

Title Page
Introduction .................................................................................................................... 1
How to Contact Fluke .................................................................................................... 1
Safety Information .......................................................................................................... 1
Installation ..................................................................................................................... 6
Unpack and Inspection ............................................................................................. 6
Placement ................................................................................................................. 6
Dry-Out Period .......................................................................................................... 6
How to Move the Product .......................................................................................... 7
Product Components ..................................................................................................... 7
Front-Panel Controls ................................................................................................. 10
Rear-Panel Connections ........................................................................................... 12
Probe Installation ...................................................................................................... 17
The Isothermal Block ................................................................................................ 20
Connect to Mains Power ........................................................................................... 21
How to Turn On the Product .......................................................................................... 21
How to Turn Off the Product .......................................................................................... 22
Display ........................................................................................................................... 23
Setup ............................................................................................................................. 23
Set the Language ...................................................................................................... 23
Set Display Contrast ................................................................................................. 23
i
9118A
Users Manual
Operation ....................................................................................................................... 24
How to Set Parameters in the Product ...................................................................... 25
Value List ............................................................................................................. 25
Set Number Value ................................................................................................ 25
Softkey Menus .......................................................................................................... 26
Temperature (TEMP) Setup Menu ............................................................................ 29
Scan Rate ............................................................................................................ 29
Cutout ................................................................................................................... 29
Program (PROG) Menu ............................................................................................ 29
RAMP/SOAK Menu .............................................................................................. 29
RUN PROG Function ........................................................................................... 30
Calibration (CALIB) Menu ......................................................................................... 30
Calibration (CAL) Points Menu ............................................................................. 30
PID Setup Menu ................................................................................................... 31
Furnace Setup Menu ................................................................................................. 31
System Menu ............................................................................................................ 31
Display (DISP) Menu ............................................................................................ 31
Communications (COMM) Setup Menu ................................................................ 31
Password Setup Menu ......................................................................................... 32
System Information (INFO) Menu ......................................................................... 32
Adjust Furnace Temperature .................................................................................... 32
Set the Furnace Temperature with a Preset ......................................................... 32
Set the Furnace Temperature Manually ............................................................... 33
How to Stop the Furnace Heat Cycle ........................................................................ 33
Change a Preset Temperature .................................................................................. 33
Probe Insertion .............................................................................................................. 34
Cutouts .......................................................................................................................... 34
Soft Cutout ................................................................................................................ 34
Hard Cutout ............................................................................................................... 35
Chassis Thermostat Cutout ....................................................................................... 35
Fan Fault Cutout ....................................................................................................... 35
TC or Heater Fault Cutout ......................................................................................... 35
Remote Operation ......................................................................................................... 35
Digital Communication Interface ............................................................................... 36
ii
Contents (continued)
How to Configure the Digital Interface ...................................................................... 37
Input Terminators ...................................................................................................... 37
SCPI Commands ...................................................................................................... 37
Commands by Function or Group ........................................................................ 37
Alphabetic List of Commands ........................................................................................ 42
Maintenance .................................................................................................................. 59
Cutout Test ............................................................................................................... 60
How to Clean the Product ......................................................................................... 60
Furnace Calibration ................................................................................................... 60
Calibration Terminology ....................................................................................... 61
Environment Conditions ....................................................................................... 61
Calibration Equipment .......................................................................................... 61
Calibration ............................................................................................................ 61
Calibration Procedure for Tube Furnace ........................................................................ 64
Calibration procedure for 9118A-ITB (with Isothermal Block) ................................... 70
Calibration Equipment Setup .................................................................................... 70
Stability Test ......................................................................................................... 73
Accuracy Test ...................................................................................................... 73
Accuracy Adjustment ........................................................................................... 74
Axial Uniformity .................................................................................................... 74
Specifications for 9118A and 9118A-ITB ....................................................................... 78
General Specifications .............................................................................................. 78
Accuracy Specifications ............................................................................................ 79
Radial Uniformity ....................................................................................................... 79
Axial Uniformity ......................................................................................................... 79
Temperature Stability ................................................................................................ 79
Heating Time (23 °C to 1200 °C) .............................................................................. 80
Cooling Time (1200 °C to 300 °C) ............................................................................ 80
iii
9118A
Users Manual
iv

List of Tables

Table Title Page
1. Symbols ................................................................................................................................ 5
2. Front-Panel Controls ............................................................................................................. 11
3. Rear-Panel Connections ....................................................................................................... 13
4. Display Elements .................................................................................................................. 23
5. Commands by Function or Group ......................................................................................... 38
6. Calibration Equipment........................................................................................................... 61
7. Furnace Calibration Tests and Adjustments ......................................................................... 63
8. Stability Test Data ................................................................................................................. 66
9. Axial Uniformity Test ............................................................................................................. 67
10. Skew Correction Values ........................................................................................................ 67
11. SKEW Parameter ................................................................................................................. 67
12. Zone Correction Values ........................................................................................................ 68
13. ZONE Parameter .................................................................................................................. 68
14. Accuracy Parameter ............................................................................................................. 69
15. Stability Test Data with Isothermal Block .............................................................................. 73
16. Accuracy Parameter for Isothermal Block ............................................................................. 74
17. Axial Uniformity Test with Isothermal Block .......................................................................... 75
18. Skew Correction Values for Isothermal Block ....................................................................... 76
19. SKEW Parameter for Isothermal Block ................................................................................. 76
20. Zone Correction Values For Isothermal Block ...................................................................... 77
21. Zone Parameter for Isothermal Block ................................................................................... 77
v
9118A
Users Manual
vi

List of Figures

Figure Title Page
1. Calibration Tube Furnace ..................................................................................................... 8
2. Front View with Isothermal Block .......................................................................................... 9
3. Rear View with Isothermal Block .......................................................................................... 9
4. Front-Panel Controls ............................................................................................................. 10
5. Rear-Panel Connections ....................................................................................................... 12
6. Rear View of 9118A Thermocouple Calibration Furnace with Isothermal Block ................... 14
7. Short Tube and Long Tube Insertion .................................................................................... 15
8. Rear Plug Insertion ............................................................................................................... 15
9. Attach the Rear Shield .......................................................................................................... 16
10. Attach the Support Bracket ................................................................................................... 16
11. Front Plug Insertion .............................................................................................................. 17
12. Probe Installation A ............................................................................................................... 18
13. Probe Installation B ............................................................................................................... 19
14. Insert Alignment .................................................................................................................... 20
15. Power-Up Screen ................................................................................................................. 21
16. Main Screen .......................................................................................................................... 22
17. Preset Selection .................................................................................................................... 25
18. SETPOINT Temperature ...................................................................................................... 25
19. Softkey Menu Tree ............................................................................................................... 27
20. Menu Softkey Displays ......................................................................................................... 28
vii
9118A
Users Manual
21. Preset Selection .................................................................................................................... 32
22. Change Preset Temperature ................................................................................................. 33
23. RS-232 Cable Connections................................................................................................... 36
24. Tube Furnace Calibration Flow Chart ................................................................................... 62
25. Test Equipment Connections ................................................................................................ 65
26. Calibration Equipment Setup ................................................................................................ 71
27. Calibration Procedure for Furnace with Thermo Block .......................................................... 72
viii

Introduction

 Warning
To prevent possible electrical shock, fire, or personal injury, read all safety information before you use the Product.
The Product is used to calibrate Platinum Resistance Thermometers (PRTs), fiber optic sensors, and thermocouples to a maximum temperature of 1200 °C.

How to Contact Fluke

To contact Fluke Calibration, call one of these telephone numbers:
Technical Support USA: 1-877-355-3225
Calibration/Repair USA: 1-877-355-3225
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31-40-2675-200
Japan: +81-3-6714-3114
Singapore: +65-6799-5566
China: +86-400-810-3435
Brazil: +55-11-3759-7600
Anywhere in the world: +1-425-446-6110
To see product information and download the latest manual supplements, visit Fluke Calibration’s website at
www.flukecal.com.
To register your product, visit http://flukecal.com/register-
product.

Safety Information

A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
 Warning
To prevent possible electrical shock, fire, or personal injury:
Read all safety information before you use the
Product.
Carefully read all instructions.
1
9118A
Users Manual
Use the Product only as specified, or the
protection supplied by the Product can be compromised.
Do not use the Product around explosive gas,
vapor, or in damp or wet environments.
Use this Product indoors only.
Use only the mains power cord and connector
approved for the voltage and plug configuration in your country and rated for the Product.
Replace the mains power cord if the insulation
is damaged or if the insulation shows signs of wear.
Connect one end of the supplied ground wire
to the protective earth ground terminal on the rear panel and the other end to an earth ground.
Make sure the ground conductor in the mains
power cord is connected to a protective earth ground. Disruption of the protective earth could put voltage on the chassis that could cause death.
Do not put the Product where access to the
mains power cord is blocked.
Make sure that the Product is grounded before
use.
Do not use an extension cord or adapter plug.
Do not use the Product if it operates
incorrectly.
Do not use the Product if it is damaged.
Disable the Product if it is damaged.
Make sure the power cord does not touch with
hot parts of the Product.
Make sure that the space around the Product
meets minimum requirements.
Do not keep the Product in operation and
unattended at high temperatures.
Use caution when you install and remove
probes and inserts from the Product. They can be hot.
Do not put the Product on a flammable surface
and keep a fire extinguisher nearby.
Do not set anything on top of the Product. Very
high temperatures can cause the ignition of flammable materials.
Do not remove the isothermal block and rear
plug at high temperatures. The isothermal block and rear plug are the same temperature as shown on the display.
Do not lift or move the Product with the
isothermal block in place. The isothermal block and rear plug may fall out of the instrument.
2
Calibration Furnace
Safety Information
Applicable personal protective equipment
must be worn when you move hot temperature probes or do work near the furnace apertures.
Do not use the Product for applications other
than calibration work.
The Product must only be used by approved
personnel.
Do not touch surfaces near the well access
port. They can be very hot.
Do not touch the insulation material in the
Product. The insulation material can cause mild mechanical irritation with skin, eyes, and upper respiratory system. If it gets on your skin, flush area and wash gently. Do not rub or scratch exposed skin. If it gets into your eyes, flush abundantly with water. Do not rub eyes. If the nose and throat become irritated, move to a dust-free area and blow nose. If symptoms continue, get medical advice.
Caution
To prevent damage to the Product:
Operate the Product at room temperature.
See specifications.
Keep a minimum of 45 cm clearance from
the front, left, and right sides. Keep 150 cm from the top of the Product.
Do not put the Product in a corner or cause
a blockage behind the Product.
Do not use fluids to clean out the well.
Fluids in the Product can cause damage.
Do not put unwanted material into the
probe protection tube or hole of the insert. Fluids in the Product can cause damage.
Only change calibration constant values
from the factory set values when you calibrate the Product. These values are important in the safe operation of the Product.
Read and understand the operation before
you turn on the Product.
Do not operate the Product in wet, oily,
dusty, or dirty environments.
Make sure the probe handle temperature
does not go above its limit.
3
9118A
Users Manual
Be careful with thermometer probes. Do
not drop, stress, or let them become too hot.
Do not operate the Product without the
control probes correctly connected and put into the furnace. The furnace does not operate correctly without the control probes.
Make sure the main probe and zone probe
are connected to the correct connector.
Turn off the furnace when a mains supply
power fluctuation occurs. Turn on the furnace only when the mains power is stable.
Make sure that all reference sensors used
with the furnace are specified for the temperature range of the furnace.
Do not disconnect the Product from the
mains while the furnace temperature is above 80 °C.
Do not attempt to push the block into the
furnace. The isothermal block expands as the furnace heats due to thermal expansion. It will extend beyond the front of the furnace at high temperatures. This is normal.
Remove the isothermal block and rear plug
prior to transporting the furnace. The ceramic furnace tube is fragile and can be damaged by the block.
To prevent the probe under test from
getting stuck in the isothermal block due to different rates of thermal expansion, allow a clearance of at least 0.4 mm between the outer diameter of the probe and the hole of the isothermal block.
Never introduce any foreign material into
the probe hole of the isothermal block. Fluids, etc. can leak into the Product and cause damage.
Minimize vibration in the calibration
environment.
The furnace can be used with or without an
isothermal block installed. Unique calibration parameters are necessary for each mode of operation.
Before operating the furnace, select the
appropriate mode of operation in the calibration menu (see Furnace Setup Menu). If necessary, a calibration for the new mode of operation may be required.
4
Calibration Furnace
Safety Information
The symbols on the Product and in this manual are shown in Table 1.
Table 1. Symbols
Symbol Description Symbol Description
Risk of Danger. Important information. See Manual.
Hazardous voltage. Risk of electric shock.
Caution. Hot surface. Protective Earth Ground
USB connection Conforms to European Union directives.
Conforms to relevant North American Safety Standards.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
Conforms to relevant Australian EMC standards.
5
9118A
Users Manual
Installation
The subsequent sections give you information on best practices for Product installation.
Unpack and Inspection
Carefully unpack the Product and check for all parts. If there is damage, tell the shipper immediately and make a claim. The parts included are:
9118A Calibration Furnace
Type-S Thermocouple, Long
Type-S Thermocouple, Short
Protection Tube, Long
Protection Tube, Short
TC Support Bracket
L-Key, Hex 3 mm
Button Head Screw, M5X12 (2 each)
Getting Started Manual
Users Manual CD-ROM
Ground Wire
Serial Cable
Isothermal Block
Rear Plug
Rear Plug for backup
The last 3 items are only for the thermocouple calibration furnace with isothermal block.

Placement

For best results, choose a location to setup the Product where room temperature changes are minimum. To safely remove heat, put the Product in a position with a minimum of 45 cm of free air space around it.
Fluke Calibration recommends that you do a test for axial uniformity and adjust if necessary after the Product is installed. Temperature uniformity of the Product could be affected by the environment where the Product operates or by equipment or material placed inside or at the ends of the central tube. Refer to the Calibration section of this manual.

Dry-Out Period

Before initial use, after transport, or the Product has not been energized for more than 10 days, turn on the Product for a minimum of 2 hours.
If the Product is wet or was in a wet environment, remove moisture before you turn on the Product.
USB Cable
Super Wool Insulator
Calibration Certificate
6
Calibration Furnace
Product Components

How to Move the Product

 Warning
To prevent personal injury, use proper techniques to lift the Product.
When you lift the Product, bend your knees and lift with your legs. Before you move the Product, let it cool to ambient temperatures and remove the control and cutout thermocouples and all objects in the furnace well. To safely move the Product, two people are necessary. Put one person in the front and one person at the rear of the Product. Carefully put your hands under the Product and lift in unison.
Caution
To prevent damage to the Product, remove the isothermal block and rear plug prior to transporting the furnace. The ceramic furnace tube is fragile and can be damaged by the block.

Product Components

As shown in Figure 1 the two primary components of the Product are the furnace and the controller. Figures 2 and 3 show the thermocouple calibration furnace with the isothermal block. The isothermal block shown in Figure 2 is made of aluminum-oxide and provides a constant and accurate temperature environment for the sensor under test. Four 6.7 mm diameter wells are provided.
7
9118A
Users Manual
Furnace
Controller
8
Figure 1. Calibration Tube Furnace
gxs009.eps
Calibration Furnace
Product Components
Rear Plug
Thermal Block
Figure 2. Front View with Isothermal Block
gxs121.eps
gxs133.eps
Figure 3. Rear View with Isothermal Block
9
9118A
Users Manual
Front-Panel Controls
Figure 4 shows the front-panel controls of the Product and Table 2 is a list of control descriptions.
1
10
2
9
3
8
5
4
7
6
gxs007.eps
Figure 4. Front-Panel Controls
10
Calibration Furnace
Product Components
Table 2. Front-Panel Controls
Item Description Item Description
LCD Display  Mains power switch
Set a temperature setpoint into the Product.
Sets the Product to show temperatures in Celsius or
Fahrenheit.
Moves through menus and adjusts numbers.
Furnace fuses
Used to set a value into the Product and choose a menu.
Menu button shows the main menu in the display.
Exits the menu shown in the display and moves to the previous menu.
Softkeys (F1 through F4) are used to select the function that shows in the display above the softkey.
11
9118A
Users Manual

Rear-Panel Connections

Figure 5 shows the rear-panel connections of the Product and Table 3 is a list of connector descriptions.
N10140
MAIN TCZONE TC
BEFORE OPENING DISCONNECT MAINS.
230V 50/60Hz 20A
12
EARTH CONNECTION ESSENTIAL BEFORE CONNECTING SUPPLY.
1
2
3 4 5 6
RS-232
Serial no.
FLUKE CORPORATION
www.flukecal.com
gxs008.eps
Figure 5. Rear-Panel Connections
Calibration Furnace
Product Components
Table 3. Rear-Panel Connections
Item Description Item Description
Mains power cord Main thermocouple input jack
Protective Earth Terminal USB connector
Zone thermocouple input jack RS-232 connector
13
9118A
Users Manual

Protection Tube and Support Bracket Installation (and Optional Isothermal Block)

Warning
To prevent personal injury, disconnect the Product from mains power and when applicable, allow to cool to 80 °C prior to performing any maintenance or installation work on the Product.
To install two protection tubes and a thermocouple (TC) support bracket and optional isothermal block (9118A­ITB):
1. Remove the five socket-head screws and take down the rear shield from unit. See Figure 6.
Rear Shield
Socket
Head Screw
Figure 6. Rear View of 9118A Thermocouple
Calibration Furnace with Isothermal Block
gxs124.eps
14
Calibration Furnace
Product Components
2. Insert the long and short protection tubes into flange
nuts and screw them into position.
Long tube in the right side.
Short tube in the left side. See Figure 7.
Short Tube
Long Tube
gxs125.eps
Figure 7. Short Tube and Long Tube Insertion
3. This step is only for the optional isothermal block (9118A-ITB). Insert the rear plug carefully into position. See Figure 8.
Rear Plug
gxs126.eps
Figure 8. Rear Plug Insertion
15
9118A
Users Manual
4. Attach the rear shield with the five socket-head screws. See Figure 9.
Rear Shield
Socket
Head Screw
Figure 9. Attach the Rear Shield
gxs127.eps
5. Install the thermocouple support bracket on the rear shield. See Figure 10.
Rear Shield
TC Support
Bracket
gxs128.eps
Figure 10. Attach the Support Bracket
16
Calibration Furnace
Product Components
6. This step is only for the optional isothermal block (9118A-ITB). Insert the isothermal block into the ceramic tube from the front side of the Product. See Figure 6.
a. Make sure that the front plug is in contact with the
side of the ceramic tube.
b. The position of block must be as shown in Figure 11.
Right Position Wrong Position
Figure 11. Front Plug Insertion
gxs130.eps
Probe Installation
Note
When you put the temperature control probes into their applicable protection tube (see steps 2 and 6 below), make sure the end surface of the probe handle is aligned with the edge of the TC support bracket. Failure to do so may cause the Product to be out of calibration.
Two thermocouples are used as temperature control probes. These probes are put into the rear of the furnace and held in position with the TC support bracket.
To install the control probes:
1. Put the long probe into the MAIN TC tube on the rear panel of the furnace.
2. Align the end surface of the thermocouple handle with the front edge of the bracket. See Figures 12 and 13 .
3. Tighten the bracket screw to hold the thermocouple in position.
4. Connect the cable of the long probe to the MAIN TC connector on the rear panel.
17
9118A
Users Manual
Zone TC
Main TC
Support Bracket
N10140
BEFORE OPENING DISCONNECT MAINS.
230V 50/60Hz 20A
MAIN TCZONE TC
EARTH CONNECTION ESSENTIAL BEFORE CONNECTING SUPPLY.
RS-232
Serial no.
FLUKE CORPORATION
www.flukecal.com
18
gxs131.eps
Figure 12. Probe Installation A
Calibration Furnace
Product Components
Zone TC Handle
Locking Screw
Main TC
Handle
Align these two surfaces
Support Bracket
gxs132.eps
Figure 13. Probe Installation B
19
9118A
Users Manual
Caution
To prevent damage to the thermocouple, do not over tighten the screw.
To prevent contamination of the thermocouple, always wear clean gloves when you handle the thermocouples.
5. Put the short probe into the ZONE TC tube on the rear panel of the furnace.
6. Align the end surface of the thermocouple handle with the front edge of the bracket. See Figures 13 and 12.
7. Tighten the bracket screw to hold the thermocouple in position.
8. Connect the cable of the short probe to the ZONE TC connector on the rear panel.

The Isothermal Block

The Isothermal Block consists of:
Front ceramic plug
Four ceramic guide tubes
Ceramic block
Rear ceramic plug
accurate temperature environment for the sensor under test. Four 6.7 mm diameter wells are provided.
To install the insert:
1. Carefully remove the isothermal block from the packaging.
The well must be clear of any foreign objects such as dirt or grit before the isothermal block is installed.
2. Carefully insert the isothermal block into the well from the front of the furnace until the front plug contacts the well.
3. Rotate the insert to the correct position, as shown in Figure 14.
Right Position Wrong Position
Items 1-3 are fixed together by high temperature cement. The isothermal block is intended to stabilize temperature fluctuations and improve heat conduction between the furnace and the sensor under test. The “block” is made of aluminum-oxide and provides a relatively constant and
20
gxs130.eps
Figure 14. Insert Alignment
4. Insert the rear plug into the well from the rear of the furnace until it contacts the tube.
Calibration Furnace
How to Turn On the Product
The fit between the block and the tube is typically loose in order to accommodate thermal expansion.

Connect to Mains Power

 Warning
For safe operation and maintenance of the product, remove moisture from the Product before you connect to mains power. Fluke Calibration recommends the Product be put in a low humidity temperature chamber at 50 °C for 4 hours or more.
To prevent possible electrical shock or personal injury, connect mains power cord correctly and make sure the product is grounded.
Be sure the power cord to the mains supply is wired correctly. The power cord wires are color-coded as follows:
Black: Line
White: Neutral
Green: Ground
Attach a separate safety ground wire between the Protective Earth Terminal on the rear panel of the Product and an earth ground conductor.

How to Turn On the Product

Push the “I” side of the power switch, on the front panel of the Product. The screen shown in Figure 15 shows in the display while the Product does a self test.
qrs001.bmp
Figure 15. Power-Up Screen
Connect the supplied ground wire between the protective earth terminal on the rear panel and earth ground. Use the
2.5 meter power cord to connect the Product to a 230 V ac outlet rated for 20 amps.
21
9118A
Users Manual
When the self test is complete and no errors are sensed, the screen shown in Figure 16 shows in the display. The Product is in a standby mode. You must set a setpoint temperature to start control of the furnace. See the “Adjust Furnace Temperature” section to learn how to start the temperature control of the furnace.
qxs002.bmp
Figure 16. Main Screen

How to Turn Off the Product

Warning
To avoid personal injury, do not disconnect the product from mains power if the furnace temperature is greater than 80 °C.
Caution
To prevent possible damage to the Product, disable the temperature control first and then turn off power.
To turn off the Product:
1. Push .
2. Push the softkey under SAVE/DISABLE to disable the heaters.
3. Push the “O” side of the power switch, on the front panel of the Product.
22
Calibration Furnace
Display
Display
Table 4 is a list of display elements with descriptions.
Table 4. Display Elements
1
2
3
5
Item Description
Furnace temperature
Setpoint temperature
4
gxs006.eps

Setup

Set the Language

Text in the LCD display can be set to show in a different language. To set the language:
1. Push .
2. Push the softkey below System Menu.
3. Push the softkey below Display Menu.
4. Push  or  to scroll through the languages.
5. When your language is shown in the display, push
.
To toggle between the set language and English, push and at the same time.

Set Display Contrast

With the main screen shown in the display, push to decrease display contrast. Push to increase display contrast.
Heat/Cool Status
Stability Indicator (
= stable, = unstable)
Softkey functions
23
9118A
Users Manual
Operation
 Warning
For safe operation and maintenance of the product:
Energize the Product for a 2 hour dry-out
period before use, if the Product was:
o In transport
o In a humid or semi-humid storage
environment
o Not energized for more than 10 days
Remove moisture from the Product before
you connect to mains power. Fluke Calibration recommends the Product be put in a low humidity temperature chamber at 50 °C for 4 hours or more. While the Product dries, do not assume the Product complies with all safety standards.
Put temperature probes you remove from
the Product on a heat resistant rack or surface until they are safe to touch.
Keep a minimum of 45 cm clearance from
the front, left, and right sides. Keep 150 cm from the top of the Product.
Notes
In order for the Product to perform within specification, the open ends of the furnace tube (both front and rear) must be filled with a temperature-resistant fiber ceramic insulating material such as a Superwool Plus Blanket.
Fiber ceramic insulation contains a binding material that will off-gas when first exposed to high temperatures producing a harmless odor. Once the binding material has off-gassed completely, the odor will no longer persist with subsequent uses of the furnace.
Warning
To avoid personal injury or skin irritation, wear disposable latex gloves (or equivalent) when handling fiber ceramic material.
The front-panel controls let you set the parameters of the Product and monitor its performance. You can also set the parameters of the communication interface. Four softkeys are used to navigate through the menus of the Product. The menus let you setup the Product and change system parameters as necessary.
24
Calibration Furnace
Operation

How to Set Parameters in the Product

There are two methods to set a parameter in the Product. You can set a parameter from a list of values or in the case of numerical values, set a parameter with the cursor keys.

Value List

An example where you select from a list of values is the preset selection.
1. From the main screen, push .
2. Push the softkey below SELECT PRESET in the display.
As shown in Figure 17, the number 1 is highlighted to the right of PRESET:.
gxs111.eps
Figure 17. Preset Selection
4. Push  when the correct preset number is shown
in the display.
This is the selection method for value lists.

Set Number Value

An example where you set a number is where you set the setpoint temperature.
1. From the main screen, push .
2. Push .
As shown in Figure 18 the left most digit in the SETPOINT temperature is highlighted.
SELECT PRESET
Figure 18. SETPOINT Temperature
SAVE/
DISABLE
gxs110.eps
3. To change the preset number, push and to scroll through the preset numbers.
25
9118A
Users Manual
3. Push  or  to increase or decrease the highlighted
digit.
4. Push  to move the highlight to the next digit.
5. Do steps 3 and 4 again for each digit in the number.
6. After all digits are set, push .
Softkey Menus
The softkeys are identified in Table 2. The softkeys are used with the front-panel key. The functions of the four softkeys are identified by label information shown directly above each key.
A group of softkey labels is a menu. A group of connected menus makes a menu tree. Figure 19 shows the menu­tree structure for the Product. Figure 20 shows each menu in the menu tree.
26
Calibration Furnace
Operation
MENU
A
B
F
E
C
D
gxs003.eps
Figure 19. Softkey Menu Tree
27
9118A
Users Manual
TEMP
A
SETUP
PROG MENU
CAL MENU
F1 F2 F3 F4
To B To C To E To F
B

SETUP CUTOUT

F1 F2 F3 F4
Set
Set soft cutout
View hard cutout
RAMP/ SOAK
C

Scan Rate

RUN PROG
F1 F2 F3 F4
Self Test
Set # of setpoints,
Status
set soak time
set # of cycles
set direction
To D
SYSTEM MENU
SET
D
POINTS
F1 F2 F3 F4
Set up to
8 setpoints
CAL
E
POINTS
F1 F2 F3 F4
Set calibration
parameter
DISPLAY
F
MENU
F1 F2 F3 F4
Set language,
set decimal,
set audio
Figure 20. Menu Softkey Displays
PID SETUP
Set control parameter
COMM SETUP
Set baud
rate, set linefeed
FURNACE
SETUP
Set work
mode
PASSWORD
Set password,
set protection
SYSTEM INFO
View model,
serial number,
firmware
version and
cal date
gxs004.eps
28
Calibration Furnace
Operation
Temperature (TEMP) Setup Menu
Push the softkey under TEMP SETUP to set the scan rate and cutout parameters.
Scan Rate
The SCAN RATE parameter sets the rate at which the Product increases or decreases the temperature to a new setpoint. The scan rate can be set between 0.1 °C to
80.0 °C/minute.
Note
The maximum scan rate is the natural rate the Product heats and cools. This is less than the maximum scan rate parameter value.
Cutout
In the cutout menu, there are two parameters: soft and hard cutout. The Product has an adjustable cutout device that turns off power to the heat source if the temperature is higher than a set value.
Use the soft cutout parameter to cut out at the high temperature limit of the probes under calibration. Fluke Calibration recommends you set the cutout 5 °C to 10 °C below the high temperature limit of the probe under test.
When the cutout opens, the Product shows CUTOUT in the display, turns off power to the heater, and sets the Product to cool. The Product stays in the cutout condition until the temperature drops below the cutout temperature and the Product is reset by the operator.
To reset the Product after the temperature drops below the cutout value:
1. Push .
2. Push . If it is necessary to change the setpoint
temperature, use the cursor buttons to change the value.
3. Push  again.
The HARD CUTOUT parameter is set at the factory and is not user settable. This menu selection only lets you see the cutout value.

Program (PROG) Menu

Push the softkey under PROG to show the RAMP/SOAK and RUN PROG menus.

RAMP/SOAK Menu

The RAMP/SOAK feature lets the user program setpoint temperatures and set the time the Product holds the temperature at each setpoint. The Product automatically heats or cools to the programmed setpoint and holds that temperature for the specified time.
The NO. SETPOINTS parameter is used to set the number of different temperatures the ramp/soak function will use in the program. You can set a maximum of eight temperature setpoints.
29
9118A
Users Manual
The SOAK TIME parameter is used to set the number of minutes each setpoint temperature is held before it changes to the subsequent setpoint. Soak time starts when the temperature is stable.
The NO. CYCLES parameter is used to set how many times the setpoint sequence is to be done again.
The DIRECTION parameter is used to set how the setpoints sequence before the sequence is done again. When set to →, the Product does the setpoints in one direction from 1 to 8. If set to , it does the setpoints from 1 to 8 and then back to 1 before the sequence is done again.
RUN PROG Function
Use the RUN PROG function to start and stop the test set up through the RAMP/SOAK menu. With the TEST STATUS parameter highlighted, push or to change the test status and push . When set to RUN, the furnace uses the values set through the RAMP/SOAK menu to control the furnace. When the furnace does all the values in the RAMP/SOAK parameters, the status changes to OFF. To stop a test, set the parameter to OFF.

Calibration (CALIB) Menu

Calibration (CAL) Points Menu
When you push the CAL POINTS softkey, the product prompts you for a numeric password. You must know the password before you can change a calibration point.
There are 12 calibration points: Zone 1 (300), Skew 1 (300), Temp 1 (300), Zone 2 (700), Skew 2 (700), Temp 2 (700), Zone 3 (1000), Skew 3 (1000), Temp 3 (1000), Zone 4 (1200), Skew 4 (1200), and Temp 4 (1200).
The TEMP parameter adjusts the actual temperature at the midpoint location for the applicable set-point temperature, but does not change the temperature gradient. The ZONE parameter increases or decreases the temperatures in the outer areas of the access tube for an applicable setpoint temperature. The curvature of the temperature profile is adjusted, and does not have an effect on the temperature at the midpoint location or the slope of the temperature profile at the midpoint location. The SKEW parameter adjusts the slope of the temperature profile at the midpoint location for the applicable setpoint temperature. This increases the temperature of one end of the access tube and decreases the temperature of the opposite end.
Note
Calibration parameters must be correct for the Product to operate correctly.
The CALIB menu lets you get access to the calibration points constants and the controller parameters.
30
Calibration Furnace
Operation

PID Setup Menu

The PID SETUP menu is used to set controller parameters.
The TEMP PB parameter is the main proportional band and the gain that the proportional-integral-derivative (PID) controller uses for control.
The TEMP INT parameter lets you set the time the controller uses for control. The time is in seconds.
The TEMP DER parameter lets you set the derivative time the controller uses for control. The time is in seconds.

Furnace Setup Menu

To operate the Product with or without an isothermal block.
1. From the calibration menu, push the Furnace Setup softkey.
2. Enter the password.
3. Push .
4. Push  or  to scroll through the display status of “YES” and “NO”. Select “YES” to set the Product to work with the isothermal block installed. Select “NO” to set the Product to work without the isothermal block installed.
5. Push  to save the selected mode.
Caution
The Product can be used with or without an isothermal block installed. To prevent damage to the Product, unique calibration parameters are necessary for each mode of operation. Before Product use, select the appropriate mode of operation in the calibration menu. When the mode of operation is changed, calibration for the new mode of operation may be required.

System Menu

The SYSTEM menu lets you get access to the display menu to set the language, set the type of decimal character, and turn off and turn on key audio. The communication setup parameters, password, and system information are also available through this menu.

Display (DISP) Menu

The LANGUAGE parameter is used to set the language used in the display.
The DECIMAL parameter is used to set the decimal character to a period or a comma.
The KEY AUDIO parameter is used to turn off and turn on the audio feedback when a key is pushed.

Communications (COMM) Setup Menu

The BAUD RATE parameter lets you set the baud rate for the RS-232 communications port.
31
9118A
Users Manual
The LINEFEED parameter turns on and turns off the line feed character sent across the RS-232 communications port.
Password Setup Menu
The PASSWORD menu lets you set the user password and set the level of protection for specified parameters. A password is necessary to access the password menu.
The USER PASSWORD parameter is used to set the password for the system that is used to access some menus. The password is a four-digit number where each digit is a number from 1 to 9. The default password is “1234”.
The PROTECTION parameter sets the level of password protection in the Product. If set to HIGH, then a password is necessary to access the Cutout menu, Ramp/Soak menu, CALIB (Calibration) menu, and PROG menu. When set to LOW, only CALIB and PROG menus are password protected.

System Information (INFO) Menu

The SYSTEM INFO menu is a view only menu. This menu shows the Product model number, serial number, firmware version, and the calibration date.

Adjust Furnace Temperature

There are two methods to adjust the furnace temperature. You can recall a preset temperature or set a temperature setpoint manually. The term “setpoint” refers to the temperature the furnace goes to and holds.

Set the Furnace Temperature with a Preset

The Product stores eight preset setpoint temperatures. You can set each preset setpoint as necessary. To set the setpoint to a preset temperature:
1. Push .
2. Push the softkey below SELECT PRESET.
3. Push  or  to move between presets (1 through 8). In
Figure 21, the temperature of the preset is shown in the display.
gxs111.eps
Figure 21. Preset Selection
4. With a preset number highlighted, push .
Note
If the Product is in standby mode, you must push again to start the furnace.
32
Calibration Furnace
Operation
Set the Furnace Temperature Manually
To adjust the furnace temperature manually:
1. Push .
2. Push .
3. Push  to increase the temperature or push  to decrease the setpoint temperature.
Note
When you push the up or down key momentarily, the setpoint temperature moves by one step. Push and hold the key and the setpoint temperature increases or decreases until the key is released.
4. Push  to save the setpoint temperature and start the furnace.
Note
After a maximum of 5 seconds, the furnace starts to heat or cool to the setpoint temperature.
The actual temperature is the measured temperature in the furnace and lags the setpoint temperature. The lag interval is the time necessary to increase or decrease the temperature of the center of the furnace.

How to Stop the Furnace Heat Cycle

To stop the furnace heat cycle:

Change a Preset Temperature

You can set the temperature of the eight presets. To change a preset temperature:
1. Push .
2. Push the softkey below SELECT PRESET.
3. Push  or  to scroll to the preset to change.
4. Push the softkey below EDIT PRESET. As shown in Figure 22 the temperature of the selected preset is highlighted in the display.
gxs112.eps
Figure 22. Change Preset Temperature
1. With the main screen in the display, push .
2. Push the softkey below SAVE/DISABLE.
The heat or cool procedure stops and the Product is set to standby mode.
33
9118A
Users Manual
5. Push .
6. The highlight moves to the most significant digit of the temperature.
7. Push  or  to change the digit.
8. Push  to move the highlight to the next digit.
9. Do steps 7 and 8 again until all digits are set.
10. Push  to set the preset value.
Note
If the preset you edited is not number 8, then the highlight moves to the subsequent preset. You can push to edit that preset. If the edited preset is number 8, then the screen goes back to the select preset screen.

Probe Insertion

Caution
To prevent possible damage to probes, make sure all probes put into the furnace are rated for the temperature range used in the calibration procedure.
Note
Fluke Calibration recommends you use a metal and/or ceramic fiber surface or container to set hot probes on.
For the best stable temperature and minimum gradient, put sample probes into the full depth of the holes in the isothermal block or the center of the furnace tube if the
isothermal block is not installed. Variations in equipment, probe dimension, and configuration can have an effect on temperature stability and gradients.
Note
A solid (unstirred) mass, as in a furnace, can have heat loss through the probe stem. The loss changes between probes and temperatures.

Cutouts

The Product has five cutout functions: soft cutout, hard cutout, chassis thermostat cutout, TC and heater cutout, and fan fault cutout. Only the soft cutout can be set through the user interface.

Soft Cutout

The soft cutout prevents damage to the Product and temperature probes. Use the soft cutout to turn off power to the furnace so the probes do not become too hot.
The temperature at which the soft cutout turns off power to the furnace can be set through the user interface. See the Temperature Setup Menu section to learn how to set the soft cutout temperature. Set the cutout temperature to ±10 °C of the temperature limit of the equipment under calibration.
34
Calibration Furnace
Remote Operation
When the soft cutout turns off power to the furnace, CUTOUT shows in the display. The Product stays in this condition until the temperature of the furnace goes below the cutout temperature and the setpoint is set again.
Hard Cutout
The Hard Cutout disables the heaters if the control thermocouples sense that the furnace temperature is above a factory-set limit. It also cuts out if a control thermocouple is open-circuit or disconnected.
The hard cutout temperature is set at the factory and can be seen through the user interface. This is not a user­accessible parameter. When the hard cutout turns off power to the furnace, the temperature must decrease a few degrees below the hard cutout temperature to reset.

Chassis Thermostat Cutout

The Chassis Thermostat Cutout turns off power if the chassis temperature is too hot. The power stays off until the temperature of the chassis decreases a few degrees below the cutout temperature. This is not a user­accessible adjustment.

Fan Fault Cutout

The Fan Fault Cutout turns off power to the furnace when a fan fails. The furnace stays off until the fan is repaired.
probes opens. This cutout also turns off the furnace if it is in the heat mode and the temperature of the furnace does not increase. When this cutout turns off furnace power, TC OR HEATER FAULT! shows in the display. The probe or heater must be repaired or replaced to reset this cutout.

Remote Operation

Remote operation of the Product from a PC or computer is done with commands through the remote interface. The command set will let you set setpoint temperatures, monitor furnace temperature, get measurement data, and control furnace operation. This section contains the procedures to set up, configure, and operate the Product through the remote interface.
The Product is controlled with Standard Commands and Programmable Instruments (SCPI) commands. The SCPI command set and how the Product implements those commands is included in the SCPI Commands section of this manual.

TC or Heater Fault Cutout

The TC Fault Cutout turns off power to the furnace when the product senses there is a problem with a control probe or a heater. This protects the Product when a control probe comes out of its socket or a wire in one of the
35
9118A
Users Manual

Digital Communication Interface

There is an RS-232C and USB interface on the rear panel of the Product. See Figure 5. To prevent noise on the signal wires, use a shielded low-resistance cable between the Product and PC.
Figure 23 shows the cable connections that connects the PC to the Product. The connections shown in the figure are for a DB 9-pin connector and DB 25-pin connector on the PC.
Instrument
Connector
(DB 9-Pin)
1 NC 2 RxD 3 TxD 4 ND 5 GND 6 NC 7 RTS 8 CTS 9 NC
Instrument
Connector
(DB 9-Pin)
1 NC 2 RxD 3 TxD 4 ND 5 GND 6 NC 7 RTS 8 CTS 9 NC
Computer (DTE)
Connector
(DB 9-Pin)
Computer (DTE)
Connector
(DB 25-Pin)
Figure 23. RS-232 Cable Connections
1 DCD 2 RxD 3 TxD 4 DTR 5 GND 6 DSR 7 RTS 8 CTS 9 NC
1 DCD 2 RxD 3 TxD 4 DTR 5 GND 6 DSR 7 RTS 8 CTS 9 NC
gxs005.eps
36
To use the USB interface for remote operation, you must install a driver on the PC. The driver is found on the 9118A CD-ROM.
To get an up-to-date USB driver, go to
http://www.ftdichip.com/Drivers/VCP.thm.
Calibration Furnace
Remote Operation
How to Configure the Digital Interface
To control the Product with a PC, you must set the parameters of the digital interface in the Product. To set the interface parameters:
1. Push .
2. Push the softkey below SYSTEM MENU.
3. Push the softkey below COMM SETUP.
4. Set the RS-232 parameters to the values as the
parameter values for the PC interface.
Input Terminators
An input terminator is a character sent by the PC that identifies the end of a string. When the Product receives an input terminator, it does the command preceding the terminator character.
Applicable terminators are:
LF (Line Feed)
CR (Carriage Return)
CR LF (Carriage Return/Line Feed)
In some instances, a terminator is automatically sent at the end of an output string from the PC.

SCPI Commands

This section gives all SCPI commands recognized by the Product with their description and syntax rules.

Commands by Function or Group

Table 5 is a list of commands grouped by Calibration, Main Screen, Program, Setup, System, and Temperature.
37
9118A
Users Manual
Table 5. Commands by Function or Group
Group Screen Parameter Command Password Group Read/Write
Calibration – Controller TEMP PB SOUR:LCON:PBAN Unconditional R/W
TEMP INT SOUR:LCON:INT Unconditional R/W
TEMP DER SOUR:LCON:DER Unconditional
Calibration – Heat Source TEMP 1 SOUR:SENS:CAL:PAR1 Unconditional R/W
TEMP 2 SOUR:SENS:CAL:PAR2 Unconditional R/W
TEMP 3 SOUR:SENS:CAL:PAR3 Unconditional R/W
TEMP 4 SOUR:SENS:CAL:PAR4 Unconditional R/W
GRAD 1 SOUR:SENS:CAL:ZONE1 Unconditional R/W
GRAD 2 SOUR:SENS:CAL:ZONE2 Unconditional R/W
GRAD 3 SOUR:SENS:CAL:ZONE3 Unconditional R/W
GRAD 4 SOUR:SENS:CAL:ZONE4 Unconditional R/W
SKEW 1 SOUR:SENS:CAL:SKEW1 Unconditional R/W
SKEW 2 SOUR:SENS:CAL:SKEW2 Unconditional R/W
SKEW 3 SOUR:SENS:CAL:SKEW3 Unconditional R/W
SKEW 4 SOUR:SENS:CAL:SKEW4 Unconditional R/W
Calibration Furnace THERMAL BLOCK SYST:CONF:BLC Unconditional R/W
R/W
38
Calibration Furnace
Remote Operation
Table 5. Commands by Function or Group (cont.)
Group Screen Parameter Command Password Group Read/Write
Main Screen (none) SOUR:SENS:DATA N/A R
SETPT SOUR:SPO N/A R/W
HEAT % OUTP1:DATA N/A R
(none) OUTP2:DATA N/A R
ENABLE OUTP1:STAT N/A R/W
Program – Run TEST STATUS PROG:STAT N/A R/W
Program - Setup TEST ID PROG:IDEN N/A R/W
Ramp & Soak RAMP/SOAK SETUP PROG:SEQ:CAT N/A R
SETPOINT n PROG:SEQ:PAR SPOn Conditional R/W
SOAK TIME PROG:SEQ:PAR DWEL Conditional R/W
SETPOINTS PROG:SEQ:PAR POIN Conditional R/W
NO CYCLES PROG:SEQ:PAR CYCL Conditional R/W
DIRECTION PROG:SEQ:PAR DIR Conditional R/W
SETPOINT 1 SOUR:LIST:SPO1 N/A R/W
SETPOINT 2 SOUR:LIST:SPO2 N/A R/W
SETPOINT 3 SOUR:LIST:SPO3 N/A R/W
SETPOINT 4 SOUR:LIST:SPO4 N/A R/W
39
9118A
Users Manual
Table 5. Commands by Function or Group (cont.)
Group Screen Parameter Command Password Group Read/Write
Ramp & Soak (cont.) SETPOINT 5 SOUR:LIST:SPO5 N/A R/W
SETPOINT 6 SOUR:LIST:SPO6 N/A R/W
SETPOINT 7 SOUR:LIST:SPO7 N/A R/W
SETPOINT 8 SOUR:LIST:SPO8 N/A R/W
Setup – Communications BAUD RATE SYST:COMM:SER:BAUD N/A R/W
LINEFEED SYST:COMM:SER:LIN N/A R/W
Setup – Display LANGUAGE SYST:LANG N/A R/W
DECIMAL SYST:DEC:FORM N/A R/W
KEY AUDIO SYST:BEEP:KEYB N/A R/W
Setup – Password PASSWORD (Disable) SYST:PASS:CDIS Unconditional W
PASSWORD (Enable) SYST:PASS:CEN Unconditional W
Status (none) SYST:PASS:CEN:STAT N/A R
USER PASSWORD SYST:PASS:NEW Unconditional W
PROTECTION SYST:PASS:PROT N/A R/W
Setup – Date/Time DATE SYST:DATE Unconditional R/W System – Setup °C/°F UNIT:TEMP N/A R/W
Heat Enable (none) OUTP:STAT N/A R/W
(none) SYST:KLOC Conditional R/W
40
Calibration Furnace
Remote Operation
Table 5. Commands by Function or Group (cont.)
Group Screen Parameter Command Password Group Read/Write
System – Information (none) SYST:ERR N/A R
(all) *IDN N/A R
(none) *CLS N/A W
(none) *OPT N/A R
FW VER SYST:CODE:VERS N/A R
Temperature – Cutout HARD CUTOUT SOUR:PROT:HCUT N/A R
SOFT CUTOUT SOUR:PROT:SCUT:LEV Conditional R/W
Reset (none) SOUR:PROT:CLEA N/A W
Trip State (none) SOUR:PROT:TRIP N/A R
Temperature – Setup SCAN RATE SOUR:RATE N/A R/W
Monitor Temperature CJC Temperature SOUR:SENS:DATA N/A R
Monitor TC SOUR:SENS:DATA N/A R
41
9118A
Users Manual

Alphabetic List of Commands

In this section, commands are shown alphabetically. Each command is shown with it long and short form. For each command there is a description, one or more examples, and a parameter range when applicable.
Note
In this section, square brackets ([ ]) show optional keywords or parameters. Braces ({ }) enclose parameters inside a command string. Triangle brackets (< >) show that you must replace the parameter with a value. Some commands use upper and lower case letters. You must use all the uppercase letters in a command. The lowercase letters are optional. The “|” is used as a separator between optional parameter values in a command.
All commands used with the Product are sequential. Each command is completed before a subsequent command is processed.
All commands that are not recognized by the Product cause an error message to show in the error queue. A command can be rejected with an error if the command is misspelled, has incorrect syntax, and/or has invalid parameters.
*CLS
Clears all status registers.
Example: *CLS
Response: None
*IDN?
Gets manufacturer, model number, serial number and firmware version of the product.
Example: *IDN?
Response: FLUKE,9118A,0,4.03
*OPT?
Gets product configuration or reference functionality.
Example: *OPT?
Response: 1 Reference hardware is enabled. (0 = disabled)
42
Calibration Furnace
Alphabetic List of Commands
OUTPut:STATus <n>
Sets or gets the main heat output on or off.
Parameters: n = ON or 1 Turn on main heat.
OFF or 0 Turn off main heat.
Example: OUTP:STAT 1 Turn on main heat.
OUTPut:STATus?
Gets the main heat output on or off.
Example: OUTP:STAT?
Response: 0 Main heat output is off. (1 = on)
OUTP[n]:DATA?
Gets the duty cycle of a heater duty cycle in percent.
Parameters: n 1 to 3 1 = main heater output (default), 2 = zone heater output, 3 = balance heater output.
Example: OUTP:DATA?
Response: 18.0 Main heater output duty cycle is 18 %.
Example: OUTP2:DATA?
Response: 57.0 Zone heater output duty cycle is 57.0 %.
PROGram:IDEN [quoted string]
Sets the program identifier.
Parameters: string = 1 to 12 characters (0 - 9, A – Z, ‘-‘). Default = “0”
Example: IDEN? “TEST-1”
43
9118A
Users Manual
PROGram:IDEN? <quoted string>
Gets the program identifier.
Example: IDEN?
Response: TEST-1
PROGram:SEQ:CAT?
Gets a parameters list for ramp and soak tests.
Example: PROG:SEQ:CAT?
Response: “SPOn”,”DWELL”,”DIR”,”POIN”,”CYCL”
PROGram:SEQ:PAR <Par>[, n]
Set a parameter in the ramp/soak test.
Parameters: Par = SPO[x] x = 1 to 8 setpoint number. Default = 1. n = 300.00 to 1200.00. Default = 300.00.
DWEL 1 to 100 seconds of soak time. Default = 15.
POIN 1 to 8 setpoints in the ramp/soak test. Default = 8.
CYCL 1 to 999 times the ramp/soak test is to be repeated. Default = 1.
DIR 0 or 1. Sets the direction of how the setpoints will be done in the test. See DIRECTION
in the Ramp/Soak Menu section. 0 = Up. 1 = Up then Down. Default = 0.
Example: To make a ramp/soak test with 3 setpoint temperatures of 450 °, 1145 °, and 825 ° that soak for 20 minutes:
PROG:SEQ:PAR POIN,3 Sets three temperature setpoints for the test.
PROG:SEQ:PAR SPO1,450.00 Sets setpoint #1 to 450.00 degrees.
PROG:SEQ:PAR SPO2,1145.00 Sets setpoint #2 to 1145.00 degrees
PROG:SET:PAR SPO3,825 Sets setpoint #3 to 825.00 degrees
PROG:SET:PAR DWEL,20 Sets soak time to 20 minutes.
44
Calibration Furnace
Alphabetic List of Commands
PROGram:SEQ:PAR? <Par>
Get a parameter from the ramp/soak test.
Parameters: Par = SPO[x] x = 1 to 8 setpoint number. Default = 1. Gets setpoint temperature for the setpoint
number.
DWEL Gets soak time
POIN Gets the number of setpoints in the ramp/soak test.
CYCL Gets the number of times the ramp/soak test is to be repeated.
DIR Gets the direction of how the setpoints will be done in the test. See DIRECTION in the
Ramp/Soak Menu section.
Example: PROG:SEQ:PAR? SPO2 Gets the setpoint temperature for setpoint number 2.
PROGram:STATus <n>
Sets the execution state for the selected program.
Parameters: n = RUN or 1 Start the selected program.
OFF or 0 Stop the selected program. (Default)
Example: PROG:STAT 1 Starts the selected program.
PROGram:STATus?
Sets or gets the execution state for the selected program.
Example: PROG:STAT?
Response: 0 The selected program is stopped.
45
9118A
Users Manual
SOURce:LCON:DER <n>
Sets the main control derivative time in seconds.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameters: n 0.0 to 99.9 seconds.
Example: SOUR:LCON:DER 5
SOURce:LCON:DER?
Gets the main control derivative time in seconds.
Example: SOUR:LCON:DER?
Response: 1.5 The main zone derivative time is 1.5 seconds.
SOURce:LCON:INT <n>
Sets the main control loop integral time in seconds.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameter: n = 10.0 to 999.9 seconds.
Example: SOUR:LCON:INIT 10
SOURce:LCON:INT?
Gets the main control loop integral time in seconds.
Example: SOUR:LCON:INIT?
Response: 20.0 The main control loop integral time is 20.0 seconds.
46
Calibration Furnace
Alphabetic List of Commands
SOURce:LCON:PBAN <n>
Sets the main control proportional band in °C which the PID controller uses for main zone control. Parameters: n = 1.0 to 199.9 °C
Example: SOUR:LCON:PBAN 7
SOURce:LCON:PBAN?
Gets the main control proportional band in °C which the PID controller uses for main zone control.
Example: SOUR:LCON:PBAN? Response: 1.5 The main control loop proportional band is 1.5 °C.
SOURce:LIST:SPO<i> <n>
Sets the temperature value for a preset setpoint in the RAMP/SOAK menu.
Parameters: i = Setpoint number (1 to 8)
n = Setpoint temperature.
Example: SOUR:LIST:SPO6 700.00 Sets setpoint number 6 to 700.00 degrees.
SOURce:LIST:SPO<i>?
Gets the temperature value for a preset setpoint in the RAMP/SOAK menu.
Example: SOUR:LIST:SPO4?
Response: 400.0 The setpoint temperature for setpoint 4 is 400.0 degrees.
SOURce:PROT:HCUT?
Gets the hard cutout temperature setpoint in °C or °F.
Example: SOUR:PROT:HCUT? Response: 140 The hard cutout temperature is set to 140 °C.
47
9118A
Users Manual
SOURce:PROT:CLEA
Resets the cutout to enable the Product.
Example: SOUR:PROT:CLEA
Response: None
If the furnace goes above the maximum temperature, the cutout circuit will trip. The Product does not heat or cool until the cutout condition is cleared.
SOURce:PROT:SCUT:LEV <n>
Sets the soft cutout temperature.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameters: n = 25.00 to 1240.00 degrees.
Example: SOUR:PROT:SCUT:LEV 450 Sets the soft cutout to 450 degrees.
SOURce:PROT:SCUT:LEV?
Gets the soft cutout temperature.
Example: SOUR:PROT:SCUT:LEV?
Response: 1000 Soft cutout temperature is set to 1000 degrees.
SOURce:PROT:TRIP?
Gets the cutout trip state.
Example: SOUR:PROT:TRIP?
Response: 0 Cutout not tripped (1 = cutout tripped).
48
Calibration Furnace
Alphabetic List of Commands
SOURce:RATE <n>
Sets the control temperature rate of change (scan rate) in °C or °F per minute.
Parameters: n = 0.10 to 500.00 degrees per minute. Default = 100.00
Example: SOUR:RATE 1.26 Sets the scan rate to 1.26 degrees per minute.
SOURce:RATE?
Gets the control temperature rate of change (scan rate) in °C or °F per minute.
Example: SOUR:RATE?
Response: 0.531 The scan rate is set to 0.531 degrees per minute. The response to this command starts out high and
decreases as the temperature gets nearer to the setpoint temperature.
SOURce:SENS:CAL:ZONE <x> <n>
Sets the axial gradient adjustment parameter.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameters: x = 1 to 4 (1 = ZONE1, 2 = ZONE2, 3 = ZONE3, 4 = ZONE4) n = -10.0 to 10.0 as a temperature offset in °C.
Example: SOUR:SENS:CAL:ZONE2 0.08
SOURce:SENS:CAL:ZONE <x>?
Gets the axial gradient adjustment parameter.
Example: SOUR:SENS:CAL:ZONE2?
Response: 0.05 Axial gradient adjustment is 0.05.
49
9118A
Users Manual
SOURce:SENS:CAL:PAR<x> [n]
Sets a control temperature calibration parameter.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameters: x = 1 to 3 (1 = PAR1 = TEMP 1, 2 = PAR2 = TEMP 2, 3 = PAR3 = TEMP 3, 4 = TEMP 4)
n = -50.00 to +50.00 (Default = 0.000)
Example: SOUR:SENS:CAL:PAR2 0.02 Set the calibration parameter to 0.02 for main control.
SOURce:SENS:CAL:PAR<x>?
Gets a control temperature calibration parameter.
Example: SOUR:SENS:CAL:PAR2?
Response: 0.0 The calibration parameter #2 is set to 0.0.
SOURce:SENS:CAL:TEMP[n]?
Gets the temperature of a calibration parameter. This temperature is fixed and is in degrees Celsius.
Parameter: n = 1 to 3 (1 = TEMP1, 2 = TEMP2, 3 = TEMP3). Default = 1
Example: SOUR:SENS:CAL:TEMP1? Response: 300 The temperature value of the TEMP1 parameter is 300 °C.
SOURce:SENS:CAL:SKEW<x> <n>
Sets an axial compensation adjustment parameter.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameters: x = 1 to 4 (1 = SKEW1, 2 = SKEW2, 3 = SKEW3, 4 = SKEW4)
50
Calibration Furnace
Alphabetic List of Commands
n = -10.00 to 10.00
Example: SOUR:SENS:CAL:SKEW2 0.08
SOURce:SENS:CAL:SKEW<x>?
Gets an axial compensation adjustment parameter.
Example: SOUR:SENS:CAL:SKEW2?
Response: 0.05 The axial compensation for SKEW2 is 0.05 of the main heater power.
SYST:CONF:BLC [n]
Sets the furnace work mode with the isothermal block inside or without the isothermal block inside.
Parameters: n =0 determines if the furnace will work without the isothermal block inside
=1 determines if the furnace will work with the isothermal block inside
Example: SYST:CONF:BLC 1 Determines if the furnace is to work with the isothermal block inside.
SYST:CONF:BLC?
Gets the furnace work mode with the isothermal block inside or without the isothermal block inside.
Example: SYST:CONF:BLC?
Response: 0 The furnace works without the isothermal block inside
SOURce:SENS:DATA? [heater part]
Gets the control temperature of the parts of the furnace.
Parameters: heater part MAIN – gets main heater temperature (Default)
ZONE – gets zone heater temperature
FULL – gets main and zone heater temperature
Example: SOUR:SENS:DATA? MAIN
51
9118A
Users Manual
Response: 700.00 Main temperature is 700.00 degrees.
Example: SOUR:SENS:DATA? FULL
Response: 700.00,600.00 Main temperature is 700.00 degrees and zone temperature is 600.00 degrees.
SOURce:SENS:DATA? [sensor function]
Gets sensor temperature in °C or °F.
Parameters: function = CJC – CJC Senor temperature
MTC – Monitor TC sensor
Example: SOUR:SENS:DATA? CJC Response: 36.12 The CJC sensor is at 36.12 °C.
SOURce:SPO [n]
Sets the control setpoint.
Parameters: n = 300.00 to 1200.00 degrees Celsius (Default = 300.00).
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Example: SOUR:SPO 400.00 Sets the setpoint temperature to 400.00 degrees.
SOURce:SPO?
Gets the control setpoint.
Example: SOUR:SPO?
Response: 300.00 Setpoint temperature is set to 300.00 degrees.
SYSTem:BEEP:KEYB [n]
Sets the state of the keyboard beep function.
52
Calibration Furnace
Alphabetic List of Commands
Parameters: n = ON or 1 Turn on keyboard beep. (Default)
OFF or 0 Turn off keyboard beep.
Example: SYST:BEEP 1 Turn on keyboard beep.
SYSTem:BEEP:KEYB?
Gets the state of the keyboard beep function.
Example: SYST:BEEP?
Response: 0 keyboard beep is off. (1 = on)
SYSTem:CODE:VERS?
Gets the version of the firmware.
Example: SYST:CODE:VERS?
Response: 1.10 The version of firmware is 1.10.
SYSTem:COMM:SER:BAUD [baud]
Sets the baud rate of the serial interface.
Parameters: baud = 1200, 2400, 4800, 9600, 19200, 38400 (Default = 2400)
Example: SYST:COMM:BAUD 2400 Sets the baud rate of the serial interface to 2400.
SYSTem:COMM:SER:BAUD?
Gets the baud rate of the serial interface.
Example: SYST:COMM:BAUD?
Response: 1200 The baud rate of the serial interface is set to 1200.
SYSTem:COMM:SER:LIN [n]
Sets the use of a line feed character at the end of data sent out the serial interface.
53
9118A
Users Manual
Parameters: n = ON or 1 Turn on line feed character.
OFF or 0 Turn off line feed character. (Default)
Example: SYST:COMM:SER:LIN 1 Turn on line feed character.
SYSTem:COMM:SER:LIN?
Gets the use of a line feed character at the end of data sent out the serial interface.
Example: SYST:COMM:SER:LIN?
Response: 0 Line feed character is not used at the end of data form the serial interface.
SYSTem:DATE <date>
Sets the system date.
Parameters: date = yyyy,mm,dd yyyy = four digit number for year, mm = two digit number for month, dd = two digit number for day.
Example: SYST:DATE 2012,05,24 Sets the system date to May 24, 2012.
SYSTem:DATE?
Gets the system date.
Example: SYST:DATE?
Response: 2012,11,30 The system date is set to November 30, 2012.
SYSTem:DEC:FORM [n]
Sets the decimal format.
Parameters: n = 0 – period (Default)
1 – comma
Example: SYST:DEC:FORM 1 Sets the decimal format to comma.
54
Calibration Furnace
Alphabetic List of Commands
SYSTem:DEC:FORM?
Gets the decimal format.
Example: SYST:DEC:FORM?
Response: 0 Decimal format is set to period.
SYSTem:KLOC <n>
Sets the status of the front-panel keypad lock.
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameters: n = 1 – Lock keypad.
0 – Unlock keypad. (Default)
Example: SYST:KLOC 1 Locks the keyboard.
SYSTem:KLOC?
Gets the status of the front-panel keypad lock.
Example: SYST:KLOC?
Response: 0 Keypad is unlocked.
SYSTem:LANG <lang>
Sets the display language.
Parameters: lang = 1 – English
7 – Chinese
Example: SYST:LANG 7 Sets the display language to Chinese.
55
9118A
Users Manual
SYSTem:LANG?
Gets the display language.
Example: SYST:LANG?
Response: 1 The display language is set to English.
SYSTem:PASS:CDIS
Disable access to commands with password protection.
Example: SYST:PASS:CDIS
This command has no response.
SYSTem:PASS:CEN [n]
Enables access to commands with password protection.
Parameters: n = 0000 to 9999. If the n parameter is left out, then the default password of 1234 is used.
Example: SYST:PASS:CEN 1045 Disables password protection on commands where a password is necessary to set
them. Password protection remains disabled until SYST:PASS:CDIS is used or you turn off power to the Product.
This command has no response.
SYSTem:PASS:CEN:STAT?
Gets the access state of password protected commands.
Example: SYST:PASS:CEN:STAT?
Response: 0 Cannot change commands with password protection. A response of 1 means password commands
can be set by the applicable command. Use the SYST:PASS:CEN command to set the password protection parameter.
SYSTem:PASS:NEW <n>|DEF
Sets the system password.
56
Calibration Furnace
Alphabetic List of Commands
Note
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to set this value.
Parameters: n = 0000 to 9999
DEF Sets the password to 1234.
Example: SYST:PASS:NEW 1045 Sets the password to 1045.
This command has no response.
SYSTem:PASS:PROT <n>
Sets the password protection level.
Parameters: n = 0 – Low
1 – High
Example: SYST:PASS:PROT 1 Sets the password protection level to high.
SYST:PASS:PROT?
Gets the password protection level.
Example: SYST:PASS:PROT?
Response: 0 The password protection level is low.
57
9118A
Users Manual
UNIT:TEMP [n]
Sets the display temperature units.
Parameters: n = C – Centigrade (Default)
F - Fahrenheit
Example: UNIT:TEMP C Sets the display temperature units to Centigrade.
UNIT:TEMP?
Gets the display temperature units.
Example: UNIT:TEMP?
Response: F Display temperature units is set to Fahrenheit.
58
Calibration Furnace
Maintenance

Maintenance

 Warning
To prevent possible electrical shock, fire, or personal injury:
Do not operate the Product with covers
removed or the case open. Hazardous voltage exposure is possible.
Use only specified replacement parts.
Have an approved technician repair the
Product.
Caution
To prevent possible damage to the Product:
Do not use fluids to clean the well.
Do not use harsh chemicals on the Product
surfaces.
Do an over-temperature cutout test a
minimum of one time in a 6 month period.
Make sure the well is clean and clear of
unwanted material.
Never introduce any foreign material into
the probe hole of the Isothermal Block. Fluids, etc., can leak into the instrument causing damage.
Do not lift or move this instrument with the
isothermal block in place. The isothermal block and rear plug may fall out of the Product.
Fluke Calibration recommends that you do an axial uniformity test when you receive the Product and at 6 month intervals. See the Furnace Calibration section.
Use the National Safety Council
decontamination steps when hazardous material is spilt on or in the Product.
59
9118A
Users Manual

Cutout Test

Fluke Calibration recommends you do a minimum of one cutout test in 6 months. To do a cutout test:
1. Push .
2. Set the setpoint temperature to 400 °C and let the furnace temperature become stable.
3. Push .
4. Push .
5. Push the softkey below TEMP SETUP.
6. Push the softkey below CUTOUT.
7. Set the soft cutout to 300 °C.
The cutout indicator on the front panel illuminates and the furnace stops the heat cycle. If the cutout does not stop the heat cycle, the Product must be repaired.
8. Set the soft cutout temperature to 1240 °C.

How to Clean the Product

Clean the Product surfaces with a damp cloth and mild detergent. Make sure liquids do not go into the Product.
Note
Before you clean or decontaminate the Product with a procedure not recommended by Fluke Calibration, speak with a service center associate for more information.

Furnace Calibration

The furnace calibration procedure in this manual is a general guideline. Each calibration laboratory must write a procedure for their calibration equipment and quality program. Do an uncertainty analysis with the laboratory equipment and environment.
Note
To get answers to furnace calibration questions, talk to a Fluke Calibration associate. See the How to Contact Fluke section.
 Warning
To prevent possible electrical shock, fire, or personal injury, have an approved technician calibrate the Product.
60
A technician who does the furnace calibration procedure must know the data in this manual.
Calibration Furnace
Maintenance
Calibration Terminology
As Found Data: Data measured on the UUT before it is adjusted.
As Left Data: Data measured on the UUT after it is adjusted.
Environment Conditions
Furnace calibration must be done in an environment with a temperature of 23 °C ±4 °C and relative humidity below 60 %.
Calibration Equipment
Table 6 is a list of equipment necessary for a furnace calibration.
Table 6. Calibration Equipment
Equipment Specifications
Ice Point
Thermocouple Readout 0.25 °C accuracy
Thermocouples
0.05 °C accuracy
0.02 °C stability
S or R type 1 °C @ 1200 °C

Calibration

Product calibration uses a stability test, uniformity test, and adjustment, followed by an accuracy test and adjustment. All tests and adjustments are done at different temperatures across the range of the Product. Figure 24 is a flow chart of the calibration procedure for the tube furnace.
61
9118A
Users Manual
Prepare Test Equipment and Power
Fail
On
Set setpoint
Heat On
Test the stabiltiy
Take As Found Data
for stability
Pass/Fail
Pass
Calculate the uniformity
Take As Found Data for
axial uniformity
Pass/Fail
Calculate the difference
of +/- 30 mm
Set the unifomity align
parameter
Fail
Pass
Take As Found Data
for stability
Pass/Fail
Take As Found Data for
Axial Uniformity
Fail
Pass/Fail
Calculate the gradient
Set the gradient align
parameter
Take As Found Data
for stability
Pass/Fail
Take As Found Data
for Axial Gradient
Pass/Fail
Pass
Pass
Pass
Fail
Fail
Fail
Calculate the accuracy
Set the accuracy
parameter
Take As Found Data
for stability
Pass/Fail
Take As Found Data
for Accuracy
Pass/Fail
Pass
All
Calibration Point
Finished
Yes
Finished
Fail
Fail
No
62
Figure 24. Tube Furnace Calibration Flow Chart
gxs114.eps
Calibration Furnace
Maintenance
Table 7 is a list of Furnace calibration tests and adjustments.
Table 7. Furnace Calibration Tests and Adjustments
No. Test Name Test Points (°C) Notes
As Found/As Left data to verify the temperature
1 Stability Test 300, 700, 1000, 1200
stability – can be done with other tests at same nominal setpoints.
2 Axial Uniformity Test 300, 700, 1000, 1200
3 Axial Uniformity Adjustment 300, 700, 1000, 1200
4 Axial Uniformity Test 300, 700, 1000, 1200 As Left data to verify the temperature uniformity.
5 Accuracy Test 300, 700, 1000, 1200 As Found test data – use to adjust the accuracy
6 Accuracy Adjustment 300, 700, 1000, 1200
7 Accuracy Test 300, 700, 1000, 1200 As Left data to verify the accuracy
As Found data – use data for initial uniformity adjustment
Adjust the zone and skew parameters to set axial uniformity to the specification
Adjusts the TEMP calibration parameters to set the accuracy to the specification
63
9118A
Users Manual

Calibration Procedure for Tube Furnace

Caution
The Product can be used with or without an isothermal block installed. Unique calibration parameters are necessary for each mode of operation. To prevent Product damage, before using the Product, select the correct mode of operation in the calibration menu (see Furnace Setup Menu). If necessary, a calibration for the new mode of operation may be required.
1. Connect the test equipment as shown in Figure 25.
Note
In the steps that follow, start the calibration procedure at the lowest temperature (300
°
move up the range to 1200
2. Set the Product to one of the setpoint temperatures in Table 8.
C.
°
C) and
64
Calibration Furnace
Calibration Procedure for Tube Furnace
TC2
TC1
Ice Point
Figure 25. Test Equipment Connections
gxs120.eps
65
9118A
Users Manual

Stability Test

The 7 steps that follow find the As Found and As Left data for stability.
3. Push  to start the furnace.
4. Put the test thermocouple (TC1) at the center of the furnace (zero reference).
5. Stabilization time starts after one hour when the stability indicator shows the furnace temperature is stable.
6. After the stabilization time, record the temperature on the readout for TC1 at 1 minute intervals for 30 minutes.
7. Calculate 2σ (2 times the standard deviation) of the 30 recorded measurements.
8. Compare the calculated 2σ with the value in the stability column of Table 8.
Table 8. Stability Test Data
Setpoint
(°C)
300 60 30 0.2
700 60 30 0.2
1000 60 30 0.2
1200 60 30 0.2
If the test fails, do steps 6 through 8 again. If it fails a second time, repair the instability.
9. Record the result.
Stabilization
Time (min)
Samples
Stability 2σ Limit
(±°C)

Axial Uniformity

The steps that follow make up the axial uniformity test and the formulae to calculate the SKEW and ZONE parameter values. The new values are put into the Product to correct for axial temperature gradients.
Note
The axial uniformity test must be done before the axial uniformity adjustment.
10. Move test thermocouple TC2 to one of the positions shown in Table 9.
66
11. Let the thermocouple soak for a minimum of 5 minutes at this position.
12. Record the temperature for TC1 and TC2 in Table 9.
Calibration Furnace
Calibration Procedure for Tube Furnace
13. Do steps 10 through 12 again for each position in Table 9.
Note
Let the Product soak for 5 minutes after you move TC2.
Table 9. Axial Uniformity Test
TC2 Position from Furnace Center
(mm)
-30 -10 0 10 30
TC1
TC2
Difference
14. Calculate the temperature difference relative to the center position for each position of TC2 in Table 9. and record it in the table.
Difference
= (TC2
(x)
– TC1
(x)
) – ( TC2
(x)
– TC1
(0)
)
(0)
15. Calculate a gradient value for the 30 and -30 positions of TC2 with:
Gradient
= TC2
(x)
– TC1
(x)
(x)
X = TC2 Position at 30 mm and -30 mm
16. Calculate the slope value with:
Slope = Gradient
– Gradient
(+30)
(-30)
17. Calculate the SKEW value with:
SKEW
n(new)
= SKEW
+ A x (Slope)
n(old)
18. Table 10 shows the value for A at each of the temperature setpoints.
Table 10. Skew Correction Values
Setpoint (°C) A Value
300 1.02
700 1.35
1000 4.67
1200 5.30
19. Put the SKEW
value into the Product through the
n(new)
front panel or RS-232 port. Table 11 is a list of temperature setpoints and the RS-232 command and front-panel name for the SKEW parameters.
Table 11. SKEW Parameter
Setpoint
(°C)
RS-232 Command Front Panel
Parameter
300 SOUR:SENS:CAL:SKEW1 SKEW (300)
700 SOUR:SENS:CAL:SKEW2 SKEW (700)
1000 SOUR:SENS:CAL:SKEW3 SKEW (1000)
1200 SOUR:SENS:CAL:SKEW4 SKEW (1200)
20. Let the Product soak for 60 minutes after you put in the SKEW value.
67
9118A
(0)
(
)
(0)
(
)
Users Manual
21. Do steps 10 through 13 again to make sure axial uniformity is equal to or less than the specification. The difference between the maximum and minimum values in the difference row of Table 9 must be between ±0.5 °C.
22. Calculate a curve value for the 0, 30 and -30 positions of TC2 with:
Curve =
(Gradient
- Gradient
) + (Gradient
+30
2
– Gradient
)
-30
23. Calculate the ZONE value with:
ZONE
n(new)
= ZONE
+ B x (Curve)
n(old)
Table 12 shows the value for B at each temperature setpoint.
Table 12. Zone Correction Values
Setpoint B Value
300 1.5
700 2.4
1000 3.3
1200 4.2
Table 13. ZONE Parameter
Setpoint
(°C)
RS-232 Command Front Panel
Parameter
300 SOUR:SENS:CAL:ZONE1 ZONE (300)
700 SOUR:SENS:CAL:ZONE2 ZONE (700)
1000 SOUR:SENS:CAL:ZONE3 ZONE (1000)
1200 SOUR:SENS:CAL:ZONE4 ZONE (1200)
25. Let the Product soak for 60 minutes after you put in the ZONE value.
26. Do steps 10 through 14 again to make sure axial uniformity is equal to or less than the specification. If the Product does not satisfy the specification, do steps 15 through 25 again with the new axial uniformity values.
24. Put the ZONE
value into the Product through the
n(new)
front panel or RS-232 port. Table 13 a list of temperature setpoints and the RS-232 command and front-panel name for the ZONE parameters.
68
Calibration Furnace
Calibration Procedure for Tube Furnace

Accuracy Test

Note
Make sure TC1 is at the midpoint in the access tube (0 MM) when you do an accuracy test.
To do an accuracy test:
1. Set the Product to a setpoint temperature in Table 14.
2. Wait for the stability indicator to show the furnace temperature is stable.
3. Record TC1 measurements for 1 minute.
4. Calculate the average of all measurements.
5. Calculate accuracy with:
Accuracy = TC1 – setpoint temperature
6. The accuracy must be less than or equal to ±5 °C. If not, do an accuracy adjustment.
7. Do steps 1 through 6 for each setpoint temperature in Table 14.

Accuracy Adjustment

To do an accuracy adjustment:
1. Calculate the new TEMP
Temp
2. Put the TEMP
= Temp
n(new)
value into the Product through the
n(new)
front panel or RS-232 port. Table 14 is a list of temperature setpoints and the RS-232 command and front-panel name for the TEMP parameters.
parameter with:
n
- Accuracy
n(old)
Table 14. Accuracy Parameter
Setpoint
(°C)
RS-232 Command Front Panel
Parameter
300 SOUR:SENS:CAL:PAR1 TEMP (300)
700 SOUR:SENS:CAL:PAR2 TEMP (700)
1000 SOUR:SENS:CAL:PAR3 TEMP (1000)
1200 SOUR:SENS:CAL:PAR4 TEMP (1200)
3. Let the Product soak for 60 minutes after you put in the TEMP value.
4. Do steps 1 through 6 of the accuracy test.
To complete the calibration of the Product, do a stability test, axial uniformity test, and accuracy test to make sure the Product is equal to or less than the specifications for each setpoint.
69
9118A
Users Manual

Calibration procedure for 9118A-ITB (with Isothermal Block)

Caution
The furnace can be used with or without an isothermal block installed. Unique calibration parameters are necessary for each mode of operation. To prevent damage to the Product, before Product use, select the appropriate mode of operation in the calibration menu (see Furnace Setup Menu). When the mode of operation is changed, a calibration for the new mode of operation may be required.
Figure 27 shows a brief overview of the calibration procedure with a thermo block.

Calibration Equipment Setup

1. Mark scales on TC1 and TC2 as shown in Figure 26. Scale interval is 10 mm.
2. Connect equipment as shown in Figure 25.
3. Put TC1 and TC2 in the position of 0 and 60 as shown in Figure 26.
Note
In the steps that follow, start the calibration procedure at the lowest temperature (300
°
move up the range to 1200 one of the setpoint temperatures in Table 15.
C. Set the Product to
°
C) and
70
Calibration Furnace
Calibration Procedure for Tube Furnace
365.50 mm
(14.39 inch)
305.50 mm
(12.03 inch)
Front Plug Guide Tube Insulator
TC1
0 10 20 30 40 50 60
TC2
Isothermal Block
0 10 20 30 40 50 60
Figure 26. Calibration Equipment Setup
gxs134.eps
71
9118A
Users Manual
Prepare Test Equipment and Power
On
Set setpoint
Heat On
Test the stabiltiy
Take As Found Data
for stability
Pass/Fail
Pass
Calculate the accuracy
Pass/Fail
Fail
Set the accuracy
align parameter
Take As Found Data
for stability
Pass/Fail
Pass
Take As Found Data
for accuracy
Fail
Fail
Figure 27. Calibration Procedure for Furnace with Thermo Block
Pass
Calculate the Axial
uniformaty
Take As Found Data for
Axial Uniformity
Pass/Fail
Fail
Calculate the Slope of
0 mm and 60 mm
Slope <0.1 C
Pass
Set the uniformity align
parameter
Take As Found Data
for stability
Pass/Fail
Calculate the Zone
Set the Zone align
parameter
Yes
Pass
Fail
No
Take As Found Data
for stability
Pass/Fail
Pass
Take As Found axial
uniformity
All
Calibration Point
Finished
Yes
Finished
Fail
gxs135.eps
72
Calibration Furnace
Calibration Procedure for Tube Furnace

Stability Test

The steps that follow find the As Found and As Left data for stability.
1. Push to start the furnace.
2. Put TC1 at 0 mm and TC2 at 60 mm.
3. Stabilization time starts after one hour when the stability indicator shows the furnace temperature is stable.
4. After the stabilization time, record the temperature on the readout for TC1 at 1-minute intervals for 30 minutes.
5. Calculate 2σ (2 times the standard deviation) of the 30 recorded measurements.
6. Compare the calculated 2σ with the value in the stability column of Table 15.
Table 15. Stability Test Data with Isothermal block
Setpoint
(°C)
300 60 30 0.1
700 60 30 0.1
1000 60 30 0.1
Stabilization
Time (min)
Samples
Stability 2σ
Limit (±°C)
7. Record the result.
After the stability of the set temperature is verified, it is recommended that you do the accuracy test that follows.

Accuracy Test

Note
When you do an accuracy test, make sure TC1 is at 0 mm inside isothermal block
To do an accuracy test:
1. Set the Product to a setpoint temperature in Table 16.
2. Wait for the furnace temperature to become stable.
Note
Use the “Stability Test” above to judge if the furnace is stable.
3. Record TC1 measurements for 1 minute.
4. Calculate the average of all measurements.
5. Calculate accuracy with:
Accuracy = TC1 – setpoint temperature
6. The accuracy must be ±5 °C. If not, do an accuracy
adjustment.
7. Do steps 1 through 6 until the accuracy is ±5 °C.
1200 60 30 0.1
If the test fails, do steps 4 through 6 again. If it fails a second time, repair the instability.
73
9118A
Users Manual
Accuracy Adjustment
To do an accuracy adjustment:
1. Calculate the new TEMP
Temp
(new) = Tempn (old) - Accuracy
n
2. Put the Temp
(new) value into the Product through
n
the front panel or RS-232 port. Table 16 is a list of temperature setpoints and the RS-232 command and front-panel name for the TEMP parameters.
Table 16. Accuracy Parameter for Isothermal Block
Setpoint
(°C)
RS-232 Command Front Panel
300 SOUR:SENS:CAL:PAR1 TEMP (300)
700 SOUR:SENS:CAL:PAR2 TEMP (700)
1000 SOUR:SENS:CAL:PAR3 TEMP (1000)
1200 SOUR:SENS:CAL:PAR4 TEMP (1200)
3. After you put in the TEMP value, let the Product soak for 60 minutes.
4. Do steps 1 through 6 of the Accuracy Test.
parameter with:
n
Parameter
Axial Uniformity
The steps that follow make up the axial uniformity test and the formulae to calculate the SKEW and ZONE parameter values. The new values are put into the Product to correct for axial temperature gradients.
Note
The axial uniformity test must be done before the axial uniformity adjustment.
1. Respectively, move TC2 to one of the positions shown in Table 17. TC1 is always fixed at the position 0 mm as a reference in the period of moving TC2 at different locations.
2. Let TC2 soak for a minimum of 5 minutes at this position.
3. Record the temperature for TC1 and TC2 in below Table 17.
After the accuracy is verified or adjusted, it is recommended that you do the axial uniformity verification and adjustment that follows.
74
Calibration Furnace
Calibration Procedure for Tube Furnace
Table 17. Axial Uniformity Test with Isothermal Block
60 mm 50 mm 40 mm 30 mm 20 mm 10 mm 0 mm
TC1(fixed @0 mm)
TC2(test in each position from 60 mm to 0 mm)
Difference(x),
x=60,50,40,…,0 mm
4. Do steps 1 through 3 again for each position in Table 17.
Note
After you move TC2, let the Product soak for 5 minutes.
5. Calculate the temperature difference relative to the 0 mm position for each position of TC2 in Table 17 and record it in the table.
Difference(x) = (TC2(x) – TC1(x)) – ( TC2(0) – TC1(0)) X=0,10,..60mm
Differencep-p=Max(Difference(0),Difference(10),…,Difference(60))­Min(Difference(0),Difference(10),…,Difference(60))
Note
If the Differencep-p already meets the axial uniformity tolerances, it is not necessary to make further adjustments to the SKEW and ZONE calibration values.
6. Calculate a gradient value for the 0 mm and 60 mm positions of TC2 with:
Gradient(x) = TC2(x) – TC1(x)
X = Position at 0 mm and 60 mm
7. Calculate the slope value with:
Slope = Gradient(60) – Gradient(0)
8. Calculate the SKEW value with:
SKEWn(new) = SKEWn(old) + A x (Slope)
9. Table 18 shows the value for A at each of the temperature setpoints.
75
9118A
Users Manual
Table 18. Skew Correction Values for Isothermal Block
Calibration Point A value
300 0.78
700 2.14
1000 3.67
1200 4.90
10. Put the SKEWn(new) value into the Product through the front panel or RS-232 port. Table 19 is a list of temperature setpoints and the RS-232 command and front-panel name for the SKEW parameters.
Table 19. SKEW Parameter for Isothermal Block
Setpoint
(°C)
RS-232 Command Front Panel
Parameter
300 SOUR:SENS:CAL:SKEW1 SKEW (300)
700 SOUR:SENS:CAL: SKEW 2 SKEW (700)
1000 SOUR:SENS:CAL: SKEW 3 SKEW (1000)
1200 SOUR:SENS:CAL: SKEW 4 SKEW (1200)
11. Let the Product soak for 90 minutes for the calibration points of 700 °C, 1000 °C, and 1200 °C. Let the Product soak for 180 minutes for calibration points of 300 °C after you put in the SKEW value.
12. Do steps 1 through 5 again to make sure axial uniformity is equal to or less than the specification or requirement.
Note
If the axial uniformity already meets the specification or requirement, it is not necessary to continue with the following zone adjustment. It is recommended that you do steps 6 through 12 again if the result cannot meet the specification or the necessary requirement. If the slope is
<
0.1 °C and the axial uniformity still cannot meet specification, then go the following zone adjustment. If the absolute value of slope of every adjustment does not show the trend of decreasing, contact an authorized Fluke Calibration Service Center.
13. Calculate a curve as the peak to peak value from the table of difference(x) from 0 mm to 60 mm. For example, the maximum value is 0.1 °C and the minimum value is -0.1 °C, then the curve is 0.2 in this case.
Curve = Maximum(Difference(0), … Difference(60)) – Minimum (Difference(0), … Difference(60))
14. Calculate the ZONE value with:
ZONEn(new) = ZONEn(old) + B x (Curve)
76
Calibration Furnace
Calibration Procedure for Tube Furnace
Table 20 shows the value for B at each temperature setpoint.
Setpoint
Table 20. Zone Correction Values For Isothermal Block
Calibration Point B Value
300 SOUR:SENS:CAL:ZONE1 SKEW (300)
300 30
700 30
700 SOUR:SENS:CAL: ZONE 2 SKEW (700)
1000 SOUR:SENS:CAL: ZONE 3 SKEW (1000)
1000 10
1200 SOUR:SENS:CAL: ZONE 4 SKEW (1200)
1200 10
15. Put the ZONE
n(new) value into the Product through the
16. Let the Product soak for 60 minutes for the calibration
front panel or RS-232 port. Table 21 is a list of temperature setpoints and the RS-232 command and front-panel name for the ZONE parameters.
17. Do steps 1 through 5 again to make sure axial
To complete the calibration of the Product, do a stability test, axial uniformity test, and accuracy test to make sure the Product is equal to or less than the specifications for each setpoint.
Table 21. Zone Parameter for Isothermal Block
Parameter
(°C)
RS-232 Command Front Panel
points of 700 °C, 1000 °C, and 1200 °C. Let the Product soak for 120 minutes for calibration points of 300 °C after you put in the ZONE value.
uniformity is equal to or less than the specification or requirement. If the Product does not satisfy the specification or requirement, do steps 6 through 16 again with the new axial uniformity values.
77
9118A
Users Manual
Specifications for 9118A and 9118A-ITB
General Specifications
Operating Conditions
Operating Temperature ............................................................... 5 °C to 40 °C
Storage Temperature ................................................................... -20 °C to 70 °C
Humidity ....................................................................................... 80 % maximum for temperatures <31 °C,
Altitude ......................................................................................... <2000 m
Power Requirements ....................................................................... 230 V ac (±10 %), 50/60 Hz, 20 A
Heater Power ................................................................................... 4000 watts at 230 V ac
Over-current Protection
System ......................................................................................... 20 A, 250 V resettable circuit breaker
Main Heater Fuse ........................................................................ F 12 A, 250 V
Zone Heater Fuse ........................................................................ F 12 A, 250 V
Computer Interface .......................................................................... RS-232 and USB
Display ............................................................................................. Monochrome LCD, °C or °F user-selectable
Display Resolution ........................................................................... 0.1 °C or °F
Size (H x W x L) ............................................................................... 400 mm x 337 mm x 700 mm
Net Weight ....................................................................................... 29 kg (without isothermal block)
Isothermal Block (optional)
Isothermal Block Construction ..................................................... Alumina
Block Outer Diameter .................................................................. 37 mm
Block Length ................................................................................ 380 mm
Well Diameter (four wells total) .................................................... 6.7 mm
Well Depth ................................................................................... 365 mm
Net Weight ................................................................................... 0.84 kg (isothermal block only)
decreasing linearly to 50 % at 40 °C
78
Calibration Furnace
Specifications for 9118A and 9118A-ITB

Accuracy Specifications

All accuracy specifications except temperature are for a period of 1 year after calibration, at 13 °C to 33 °C.
Note
The furnace can be used with or without an isothermal block installed. Unique calibration parameters are necessary for each configuration. Calibration does NOT come standard from the factory for each configuration. Ensure that the furnace has been properly calibrated for the desired mode of operation.
Furnace Temperature Range .......................................................... 300 °C to 1200 °C
Set-Point Accuracy .......................................................................... ±5 °C

Radial Uniformity

Temperature
300 °C 700 °C ±0.20 °C 1200 °C ±0.25 °C
(14 mm from geometry central point)
9118A
±0.5 °C
9118A-ITB
(hole to hole)
±0.1 °C

Axial Uniformity

Temperature
Full Range ±0.25 °C ±0.2 °C
(±30 mm axial length from geometry central point)
9118A
(at 60 mm from full immersion)
9118A-ITB

Temperature Stability

Specification 9118A 9118A-ITB
Stability ±0.2 °C ±0.1 °C
Stabilization Time
Note: Temperature stability measured as 2-sigma over 30 minutes
2 hours, full range
3 hours at or below 700 °C
2 hours above 700 °C
79
9118A
Users Manual
Heating Time (23 °C to 1200 °C)
9118A .............................................................................................. 40 minutes
9118A-ITB ........................................................................................ 45 minutes
Cooling Time (1200 °C to 300 °C)
9118A .............................................................................................. 180 minutes
9118A-ITB ........................................................................................ 200 minutes
80
Loading...