Fluke Calibration 6054 User Manual

Via Acquanera, 29 22100 Como tel. 031.526.566 (r.a.) fax 031.507.984
6054
Calibration Bath
October 2013 © 2013 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
User Manual
WARNING
To ensure the safety of operating personnel, and to avoid damage to this unit:
DO NOT operate this unit without a properly grounded, properly polarized power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
DO use a ground fault interrupt device.
WARNING
HIGH TEMPERATURES PRESENT
in this equipment.
FIRES and SEVERE BURNS
may result if personnel fail to observe safety precautions.
WARNING
HIGH VOLTAGE
is used in the operation of this equipment.
SEVERE INJURY or DEATH
may result if personnel fail to observe safety precautions.
Before working inside the equipment, turn power off and disconnect power cord.
Consult the fluid manufacturer’s MSDS (Material Safety Data Sheet).
SAFETY PRECAUTIONS MUST BE OBSERVED.
WARNING
Fluids used in this bath may produce
NOXIOUS OR TOXIC FUMES
under certain circumstances.
PROPER VENTILATION AND
c
d
TABLE OF CONTENTS
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Two Modes of Temperature Control . . . . . . . . . . . . . . . . . . . . . . 11
3.1.1 The Temperature Control Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1.2 The Temperature Drift Mode: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1.3 The Fluid system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Description Of Features and Controls . . . . . . . . . . . . . . . . . . . . . 12
3.2.1 The Controller Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2.2 The Power Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.3 The Rear Panel: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1 Receiving and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Installation Location Requirements . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2 Electrical Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.3 Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
4.3 Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3.1 Filling the Bath . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3.1.1 Filling the Bath With Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3.2 Draining the Oil from the Bath . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5 Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . 18
5.1 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6 Controller Operation . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.1 Bath temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.2 Reset Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.3 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.3.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.3.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.3.3 Set-point vernier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.4 Temperature Scale Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.5 Secondary Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.6 Heater Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.7 Proportional Band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.8 Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
v
6.9 Controller Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.10 Probe Parameters Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.10.1 R0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.10.2 ALPHA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.11 Operating Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.11.1 Cutout Reset Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
6.11.2 Stirrer Mode Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.11.3 Stirrer set-point. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.12 Serial Interface Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.12.1 BAUD Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.12.2 Sample Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.12.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.12.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.13 IEEE-488 Parameters Menu . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.13.1 IEEE-488 Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.14 Calibration Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.14.1 CTO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.14.2 CO and CG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.14.3 H and L. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.15 Operation Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7 Digital Communication Interface . . . . . . . . . . . . . . . . . . . 28
7.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.1.2 Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.1.2.1 BAUD rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.1.2.2 Sample Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.1.2.3 Duplex Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.1.2.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.1.3 Serial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.2 IEEE-488 Communication (optional) . . . . . . . . . . . . . . . . . . . . . . 29
7.2.1 Setup and Address Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.2.2 IEEE-488 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
7.3 Interface Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.4 Power Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.5 Heater Settings for Control. . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8 Bath Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.2 Measuring the Set-point Error . . . . . . . . . . . . . . . . . . . . . . . . . 34
8.3 Computing R0 and ALPHA . . . . . . . . . . . . . . . . . . . . . . . . . . 34
8.4 Calibration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9 Bath Heat Transfer Fluid . . . . . . . . . . . . . . . . . . . . . . . 35
vi
9.1 Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.2 Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.3 Specific heat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.4 Thermal Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.5 Thermal Expansion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.6 Electrical Resistivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.7 Fluid lifetime . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.8 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.9 Cost . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.9.1 Commonly used fluids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
9.9.2 Water. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.9.3 Ethylene Glycol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.9.4 Mineral Oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.9.5 Silicone oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.9.6 Heat Transfer Salt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.10 Fluid Characteristics Charts . . . . . . . . . . . . . . . . . . . . . . . . . . 39
9.10.1 Limitations and Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
9.10.2 About the Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
10 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
10.1 The Heater Indicator LED Stays Red But The Temperature Does Not In-
crease. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
10.2 The Controller Display Flashes “CUT-OUT” And The Heater Does Not Oper-
ate. 41
10.3 The Display Flashes “CUT-OUT” And An Incorrect Process Temperature. . . 41
10.4 The Displayed Process Temperature Is In Error And The Controller Remains In
The Cooling Or Heating State At Any Set-point Value. . . . . . . . . . . . . . . . . 42
10.5 The Controller Controls Or Attempts To Control At An Inaccurate Tempera
ture. 42
10.6 The Controller Shows That The Output Power Is Steady But The Process Temper
ature Is Unstable.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
10.7 The Controller Alternately Heats For A While Then Cools. . . . . . . . . . . 42
10.8 The Controller Erratically Heats Then Cools, Control Is Unstable. . . . . . . . 42
-
-
vii
1 Introduction
Introduction
The Hart Scientific Model 6054 Calibration Bath is a highly stable constant temperature liquid bath. It has been designed for calibrating liquid and glass ther mometers or other types of long thermometers against a known temperature standard such as a Standard Platinum Resistance Thermometer (SPRT).
The 6054 calibration bath provides the following fea tures:
A deep fluid tank (test well is 7.5 inches in di
ameter and has 24 inches of fluid depth).
It provides a highly temperature stable low gra
dient environment typically a nominal stability of ±0.005°C with oils. The fluid is well stirred
and environmentally protected to minimize gradients.
-
-
-
-
The fluid level is near the top of the test well to
facilitate calibration of liquid and glass ther mometers without needing to compensate for stem effect.
The bath provides two calibration modes. An
electronically controlled mode uses a hybrid digital and analog PI controller with lock in am plifier design. The temperature and other func tions are selected with the four button keypad to a hundredth of a degree and finer with a digi tal vernier adjustment. The Drift mode by passes the controller with heater power manually controlled with variable transformer.
-
-
-
-
-
9
Specifications
2 Specifications
Table 1
Power Required . . . . . . . . . 230 VAC; 60 Hz; 15 Amps max.
Bath Temperature Range . . . . 50°C to 325°C
Temperature Stability. . . . . . .
Temperature Uniformity . . . . .
Controller Accuracy . . . . . . .
Test Well Area . . . . . . . . . . 7.5-inches dia X 24-inches deep
Heater . . . . . . . . . . . . . . Electrical, 5 positions; 1=off, 2=250, 3=500, 4=750 & 5=1,000 watts at
Boost Heater . . . . . . . . . . . 750watts
Heat Transfer Liquid . . . . . . . Fluids compatible with stainless steel may be used. (e.g. oils such as
Exterior Dimensions . . . . . . . Height 56" X Width 31" X FB 23"
Weight . . . . . . . . . . . . . . 156 Lbs.
Specifications
±0.005°C to 200°C, ±0.010°C to 325°C
±0.005°C max
±0.5°C
240 VAC line
Dow Corning 710 and 210H or Hart Salt)
10
3 Theory of Operation
Theory of Operation
The components, features and operational theory of the Model 6054 calibration bath are described in this section.
3.1 Two Modes of Temperature Control
Two modes of temperature control are available with the model 6054. The TEMPERATURE CONTROL mode or the TEMPERATURE DRIFT mode may be selected by a switch on the front panel.
3.1.1 The Temperature Control Mode
The control mode uses a hybrid digital/analog PI tem­perature controller with lock-in-amplifier. The bath stability is very high with this controller. The tempera­ture is selected using a four button keypad on the front panel. Temperatures from 0.00 to 325.00°C may be selected directly to a hundredth of a degree.Finer ad­justment is available using the vernier adjustment. Accuracy of the setting is typically ±0.5°C or better.
The controller pulses AC current to the control heat­ers in a time modulated fashion to compensate for heat gains and losses to the system. A two color LED on the control panel glows red when the heaters are on and glows green when they are off and cooling is taking place. (Note: The cooling required for control is supplied through heat loss to ambient.)
The Temperature control probe uses a 100 ohm PRT. It is a totally separateunit for ease of replacement.It is inserted into the top of the bath near the stirring motor as shown and plugs into the rear of the control unit.
The heaters are external to the tank. They are ar ranged electrically to provide the 4 control heating po sitions plus an off position. The additional boost heating position is accessed from a separate switch on the control panel.
3.1.2 The Temperature Drift Mode:
In the Temperature Drift mode the heater may be set manually to allow the temperature to drift very slowly
(a few milli-°C per minute) over the desired range. This allows the control noise to be eliminated al though greater skill is required in making calibrations. The heater power is adjusted by means of a variable transformer located on the control panel.It allows po sition one of the heater selection switch to be continu ally variable from 0 to 100% Positions 2, 3, 4 and Boost add their full value of heat incrementally to the adjusted value of position 1.
-
-
-
3.1.3 The Fluid system
The fluid system consists of the insulated tank, the stirrer assembly, the condensate drain, the overflow test well, and the fluid itself. The heaters and probe, which are part of the control system, are physically external to the tank.
The tank and other wettedparts are made of stainless steel for compatibility with most practical thermostating fluids.The stirrer is attached to the tank top plate of the bath and its motor receives additional cooling from a fan to keep from overheating and in­crease lifetime at high bath temperatures. The stirrer directly drives four 2-inch diameter stirring propellers. The down draft from the propellers forces the bath fluid through the overflow test well.The stirring motor plugs into the rear of the control unit.(See Figure 3 on page 14.)
The over-flow test well serves to provide a constant depth of fluid at an essentially constant height near the well opening. Variations in fluid volume due to thermal expansion and volatilization will not effect measurements within reasonable volume ranges. The fluid expelled from the tank is controlled to flow past the control heaters first before entering the main
-
-
tank for thermal management.
A drain is provided for convenience in changing the bath fluid. (See draining the tank.)
The condensate drain collects condensed oil vapor and oil that has expanded over the top of the tank walls.Tubes on either side of the drainat the bottom of the bath allow the liquid to be collected into a pan. Be sure this pan is in place at all times to prevent oil from draining onto the floor The condensate pan must be
11
Theory of Operation
EXIT — Used to exit from a menu. When EXIT is pressed any changes made to the displayed value will be ignored.
3) The control indicator is a two color light emitting di ode (LED).This indicatorlets theuser visually see the ratio of heating to cooling. When the indicator is red the heater is on, and when it is green the heater is off and the bath is cooling.
3.2.2 The Power Panel
The Power Panel (Figure 2) controls include; 1) The
Mode Select switch and indicators, 2) the Drift Ad just control 3) the control Heating select switch, 4) the Power switch and indicator, and 5) the Boost Heater and indicator.
1) The MODE SELECT switch selects between the Temperature Control and Drift Adjust modes. Lights show which mode is functioning. The Temperature Control position selects the temperature controller to operate and the desired temperature is selected on the controller panel. In the Temperature Drift mode
heater control is via the Drift Adjust control and the Heating select switch.
2) The DRIFT ADJUST control is a variable trans
-
former that adjusts the Low control heater through 0 to 100% of its power range. The additional power re
-
­quired for higher temperatures may be added in steps by selecting heating positions Medium, Medium High, and High, as required.
3) The control HEATING switch selects control heater power positions 1 through 5. Select the lowest rea sonable value for normal control conditions depend
-
­ing on bath temperature. The switch simply adds
-
more heaters into the circuit until the desired power is attained. Position 2 is variable using the Drift Adjust control.
4) The POWER (On-Off) switch powers up the bath. The switch is a DPST type that opens both legs of the 230 volt power source. A red indicator light shows that power is on.
5) The BOOST HEATING provides an additional 750 watts for slewing between temperatures. The Boost Heating Indicator shows whether the boost heater is on or off. The boost heater is powered through the
Boost Heating
Indicator
Power
Indicator
Figure 2
BOOST
HEATER
OFF
ON
OFF
Power panel
Boost Heating
Switch
5
4
Power
Switch
Mode Select
Switch
1
Temperature Drift
Mode Indicator
TEMPERATURE DRIFT
MODE SELECT
TEMPERATURE CONTROL
OFF
3
Heating
Switch
LOW
HEATINGPOWER
HIGH
Temperature Control
Mode Indicator
20
Drift Adjust
Control
2
DRIFT ADJUST
40
30
10
0
50
PERCENT
60
70
80
90
100
13
Theory of Operation
1
2
3
8
4
7
E
F
S
U
E
S
U
F
F
U
S
U
F
E
S
E
6
Figure 3
Back panel
14
5
Theory of Operation
temperature controller triac to prevent exceeding the desired set temperature. The boost heater indicator will flash when the set temperature has been reached as a reminder to turn it off for control.
3.2.3 The Rear Panel:
The Rear Panel has many features (see Figure 3). 1) The PROBE connector, 8) STIRRER POWER outlet,
7) FUSES, 6) Power Connection, 5) ELECTRONICS FUSE INTERNAL notation, 2) Unit SERIAL NO. no tation, 3) optional SERIAL PORT and, 4) optional IEEE-488 PORT.
1) The PROBE connector on the back panel is used for the temperature controller probe.
2) The unit SERIAL NO.is located at the top right cor ner of the back panel. When consulting with the fac
-
­tory, refer to the serial number.
3) SERIAL PORT (optional)
4) IEEE-488 PORT (optional)
5) The ELECTRONICS FUSE is located inside the cabinet and is a 1/4 amp fuse. To replace the elec
­tronic fuse, remove the cover over the controller.
-
6) The POWER CORD connector
7) The SYSTEM FUSES are 15 amps slow blow (one fuse for each leg of the 230 VAC). The heater fuses are located internally and are 10 amp, 250 VAC.
8) The STIRRER POWER connection provides 230 VAC to the stirring motor and its cooling fan.
15
Installation
4 Installation
The Model 6054 bath can be readily installed given due care to the following instructions.
4.1 Receiving and Inspection
Upon receipt of the bath, inspect it to see that there is no obvious damage from shipment. If any damage is observed, notify the carrier at once for an inspector to check out the damage.
Verify that all of the items ordered have been shipped. Notify Hart Scientific immediately if there are any dis crepancies.
4.2 Installation Location Requirements
4.2.1 Environment
The model 6054 bath is a precision instrument that must be located in an appropriate environment. The location should be free from drafts, extreme tempera­tures and temperature changes, dirt, etc. The bath must be level, use the levelers. Allow free air space around the bath to allow surface heat to convect away freely. Ventilation ofoil fumes will require a fumehood.
4.2.2 Electrical Power
Install the bath in an inherently fire safe area. There should not be any material around that will ignite by setting hot probes and thermometers on it or by spill age of the hot bath fluid.
The best floor surface is concrete. If concrete is un available, the surface should be protected in some way from inadvertent spillage.
Do Not install thebath near flammable wall materials.
We recommend installing the bath under a fume hood to safely remove oil fumes. It also will help to remove
-
excess heat.It is best to direct oilfumes away from the user when using.
Keep all flammable liquids, fumes, gasses, etc. away from the bath to prevent ignition.
Keep fire safety equipment specific to the type of me­dium handy in case they are needed.
Safe handling equipment such as leather gloves (such as welding gloves), face shields, long apron etc. are required for reasonable safety.
-
-
4.3 Setup
Inspect the probe.It should not be bent or damaged in any way.The probe used with the Model 6054is apre cision PRT sensor.
The probe is to be plugged into its connector on the rear of the bath and inserted fully into the hole located near the stirrer motor inside the motor cover.
-
The bath requires 208-240 VAC single phase power. The higher voltage is used forrunning the heaters and is required to reduce the current to reasonable levels. The power connected at the junction box on the rear of the bath. The two hot legs are wired to the brown and blue wires, and the ground to the green/yellow wire according to standard convention.
4.2.3 Safety Considerations
The Hart model 6054 bath is a high temperature bath. Although safety has been a concern in its design, there are several installation and operational consid erations to prevent fire and burns.
16
Plug the stirrer into the receptacle located on the rear panel of the bath marked STIRRER POWER. This re ceptacle is switched on with the unit’s main power.
Attach a drain line to the overflow drain tube at the rear of the bath behind the strirer.Use a line compati ble with the selected fluid and maximum bath temper ature to be used. Run the line to a similarly adequate sump.
-
-
-
4.3.1 Filling the Bath
-
The bath is shipped dry. Check inside of the test well for foreign matter and remove it to avoid interference with operation.
Installation
4.3.1.1 Filling the Bath With Oil
To fill the bath with oil, first be sure the drain is fully closed off. Check the drain assembly for tight fittings. Pour in the heat transfer fluid until it reaches a 1-inch depth from the top of the top plate.Note: Actual depth depends on the thermal expansion of the oil and the anticipated temperature range. The fluid volume of the tank is approximately 2800 cubic inches to the level of the primary overflow. The flowing fluid level
can be adjusted somewhat by adding to or removing fluid.The fluid levelduring pumping should be flowing over the rim.
The bath is now ready for operation.
4.3.2 Draining the Oil from the Bath
17
Operating Instructions
5 Operating Instructions
Operating the model 6054 constant temperature bath is not complex, but must be done according to the fol lowing instructions.
The system must be installed and the bath filled ac cording to the instructions in the previous section.
5.1 Quick Start
With the bath fluid in the bath and the control probe in place, the bath is ready to be turned on. When switched on, the stirring motor, the controller display ing the bath temperature, and the heater will come on. Now set the bath to the desired temperature using the buttons to set the temperature controller. This is ac complished by pressing the SET button and then us­ing the UP and DOWN buttons to reach the desired set temperature. Once the set-point desired is dis­played, press the SET button to set the bath to the new temperature and then press the EXIT button to return to the temperature display (refer to the Temper­ature Controller User Flow chart Figure 4). The bath will heat to the set temperature and begin to control. Allow several minutes for the bath to stabilize at the control set-point.
The heater power switch should be set to the lowest position necessary to provide adequate power to con trol. Obviously, higher bath temperatures will require higher heater settings. The boost heater may be
switched on to bring the bath up to higher tempera
-
-
-
-
-
tures quickly. It mustbe switched off when the temper ature is reached.
To achieve optimum control stability the proportional band may require adjustment. The ideal proportional band setting varies with temperature, heater setting, and fluid type.
It is advised that you operate the 6054 bath with an access cover or thermometer holder in place. At most temperatures, control stability is improved with the cover in place. At high temperatures, the use of the cover also improves safety.
To accelerate bath cooling after operation at a higher temperature, the access cover may be removed to al­low greater air cooling. It also may be possible to make use of a dip chiller or auxiliary cooling coil to de­crease cooling times.
If the Temperature Drift mode is desired, select that position after adjusting the Drift Adjust to 0% to elim­inate jumps in temperature, then adjust to the antici­pated heating position. Use a bridge and strip chart recorder or other adequate means to establish the desired temperature drift rate as the Drift Adjustcon­trol is adjusted upward. Select the LOW control heater position for the minimum amount of heat and add heat in increments with positions MEDIUM, ME DIUM HIGH and HIGH as needed.
-
-
-
18
6 Controller Operation
Controller Operation
This Section discusses in detail how to operate the bath temperature controller using the front control panel. Using the front panel key switches and LED display the user may monitor the bath temperature, set the temperature set-point in degrees C or F, moni tor the heater output power, adjust the controller pro portional band, set the cutout set-point, and program the probe calibration parameters, operating parame ters, serial and IEEE-488 interface configuration, and controller calibration parameters. Operation ofthe pri mary functions is summarized in Figure 4.
6.1 Bath temperature
The digital LED display on the front panel allows di rect viewing of the actual bath temperature. This tem­perature value is what is normally shown on the display. The units, C or F, of the temperature value are displayed at the right. For example,
25.00 C
The temperature display function may be accessed from any other function by pressing the “EXIT” button.
Bath temperature in degrees Celsius
6.2 Reset Cutout
If the over-temperature cutout has been triggered then the temperature display will alternately flash,
Cut-out
The message will continue to flash until the tempera ture is reduced and the cutout is reset.
Indicates cut-out condition
until the user resets the cutout. To access the reset cutout function press the “SET” button.
-
-
-
-
-
S
The display will indicate the reset function.
Press “SET” once more to reset the cutout.
This will also switch the display to the set temperature function. To return to displaying the temperature press the “EXIT” button. If the cutout is still in the over-temperature fault condition the display will con­tinue to flash “cutout”. The bath temperature must drop a few degrees below the cutout set-point before the cutout can be reset.
Access cutout reset function
rESEt ?
Reset cutout
Cutout reset function
6.3 Temperature Set-point
The bath temperature can be set to any value within the range as given inthe specifications with ahigh de­gree of resolution. The temperature range of the par ticular fluid used in the bath must be known by the operator and the bath should only be operated well below the upper temperature limit of the liquid. In ad dition, the cutout temperature should also be set be low the upper limit of the fluid.
Setting the bath temperature involves three steps: (1)
-
select the set-point memory, (2) adjust the set-point value, and (3) adjust the vernier, if desired.
-
-
-
The cutout has two modes — automatic reset and manual reset.The mode determines how the cutout is reset which allows the bath to heat up again. When in automatic mode, the cutout will reset itself as soon as the temperature is lowered below the cutout set-point. With manual reset mode the cutout must be reset by the operator after the temperature falls below the set-point.
When the cutout is active and the cutout mode is set to manual (“reset”) then the display will flash “cutout”
6.3.1 Programmable Set-points
The controller stores 8 set-point temperatures in memory. The set-points can be quickly recalled to conveniently set the bath to a previously programmed temperature.
To set the bath temperature one must first select the set-point memory. This function is accessed from the temperature display function by pressing “SET”. The number of the set-point memory currently being used
19
Controller Operation
EXIT
Cutout Active
EXIT
EXIT
Display
Temperature
SET
Reset Cutout
SET
Select Setpoint
SET
Adjust Setpoint
EXIT
EXIT
EXIT
Secondary Functions
+
SET
EXIT
Display Power
SET
Set Proportional Band
SET
Set Cutout Temp.
SET
SET
EXIT
Adjust Vernier
EXIT
SET
EXIT
Set Scale °C/°F
SET
Configuration Menu
SET
EXITEXITEXIT EXIT EXIT
UP
Probe Menu
SET
R0
SET SET SET
SET
Adj. ALPHA
SET/EXIT SET/EXIT
EXIT EXIT EXIT
Adj. R0
SET/EXIT SET/EXIT SET/EXIT SET/EXIT
ALPHA
EXIT
DOWN
Operating
Parameters
Menu
SET SET SET SET
Cutout
Reset Mode
SET
Adj. Cutout Reset Mode
Stirrer
Mode Select
SET
Adj. Stirrer
Mode
Stirrer
Setpoint
SET
Adj. Stirrer
Setpoint
SET/EXIT
UP UP UP
DOWN DOWN DOWN
EXIT
EXIT
EXIT
Serial
Interface
Menu
BAUD
Rate
Adjust
BAUD Rate
Sample
Period
SET
Adj. Sample
Period
SET/EXIT
Duplex
Mode
SET
Adj. Duplex
Mode
SET/EXIT
Linefeed
SET
Adjust
Linefeed
SET/EXIT
EXIT
EXIT
EXIT
IEEE-488
Interface
Menu
Device
Address
SET
Adj. Device
Address
SET/EXIT
IEEE-488
Option Installed
EXIT
Calibration
Menu
Adjust CTO
SET
Adjust C0
SET/EXIT
SET
DO NOT CHANGE THESE VALUES. SEE MANUALDO NOT CHANGE THESE VALUES. SEE MANUAL
Adjust CG
SET/EXIT
SET
Adjust H
SET/EXIT
Interface
Option Installed
SET
X5
CTO
C0
EXIT
CG
EXIT
H
EXIT
L
EXIT
Figure 4
Controller Operation Flowchart
20
Adjust L
SET/EXIT
Controller Operation
is shown at the left on the display followed by the cur rent set-point value.
25.00 C
S
Access set-point memory
1. 25.0
To change the set-point memory press “UP” or “DOWN”.
U
Increment memory
4. 40.0
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
Bath temperature in degrees Celsius
Set-point memory 1, 25.0°C currently used
New set-point memory 4,
40.0°C
-
6.3.3 Set-point vernier
The set-point value can be set with a resolution of
0.01°C. The user may want to adjust the set-point slightly to achieve a more precise bath temperature. The set-point vernier allows one to adjust the temper ature below or above the set-point by a small amount with very high resolution. Each of the 8 stored set-points has an associated vernier setting.The ver nier is accessed from the set-point by pressing “SET”. The vernier setting is displayed as a 6 digit number with five digits after the decimal point. This is a tem perature offset in degrees of the selectedunits, C or F.
0.00000
To adjust the vernier press “UP” or “DOWN”. Unlike most functions the vernier setting has immediate ef fect as the vernier is adjusted. “SET” need not be pressed.This allows one to continuallyadjust the bath temperature with the vernier as it is displayed.
U
Increment display
Current vernier value in°C
-
-
-
-
6.3.2 Set-point Value
The set-point value may be adjusted after selecting the set-point memory and pressing “SET”. The set-point value is displayed with the units, C or F, at the left.
C 40.00
If the set-point value need not be changed then press “EXIT” to resume displaying the bath temperature.To adjust the set-point value press “UP” or “DOWN”.
U
Increment display
C 42.50
When the desired set-point value is reached press “SET” to accept the new value and access the set-point vernier.If “EXIT”is pressed instead then any changes made to the set-point will be ignored.
S
Accept new set-point value
Set-point 4 value in°C
New set-point value
0.00090
Next press “EXIT”to return to the temperature display or “SET”to access the temperature scale units selec­tion.
S
Access scale units
New vernier setting
6.4 Temperature Scale Units
The temperature scale units of the controller may be set by the user to degrees Celsius (°C) or Fahrenheit (°F). The units will be used in displaying the bath tem perature, set-point, vernier, proportional band, and cutout set-point.
The temperature scale units selection is accessed af ter the vernier adjustment function by pressing “SET”. From the temperature display function access the units selection by pressing “SET” 4 times.
Bath temperature
S
25.00 C
Access set-point memory
-
-
1. 25.0
Set-point memory
21
Controller Operation
S
Access set-point value
C 25.00
S
Access vernier
Set-point value
percent heating power should not fluctuate more than ±1% within one minute.
The heater power display is accessed in the second ary menu. Press “SET” and “EXIT” simultaneously and release. The heater power will be displayed as a percentage of full power.
-
0.00000
S
Access scale units selection
Un= C
Press “UP” or “DOWN” to change the units.
U
Change units
Un= F
Press “SET” to accept the new selection and resume displaying the bath temperature.
S
Set the new units and resume temperature display
Vernier setting
Scale units currently selected
New units selected
6.5 Secondary Menu
Functions which are used less often are accessed within the secondary menu. The secondary menu is accessed by pressing “SET” and “EXIT” simulta neously and then releasing. The first function in the secondary menu is the heater power display. (See
Figure 4.)
6.6 Heater Power
The temperature controller controls the temperature of the bath by pulsing the heater on and off. The total power being applied to the heater is determined by the duty cycle or the ratio of heater on time to the pulse cycle time. This value may be estimated by watching the red/green control indicator light or read directly from the digital display. By knowing the amount of heating the user can tell if the bath is heat ing up to the set-point, cooling down, or controlling at a constant temperature. Monitoring the percent heater power will let the user know how stable the bath temperature is. With good control stability the
S+E
12 Pct
To exit out of the secondary menu press “EXIT”. To continue on to the proportional band setting function press “SET”.
E
Return to temperature display
Access heater power in secondary menu
Heater power in percent
6.7 Proportional Band
In a proportional controller such as this the heater output power is proportional to the bath temperature over a limited range of temperatures around the set-point. This range of temperature is called the pro­portional band.At the bottom of the proportional band the heater output is 100%. At the top of the propor­tional band the heater output is 0. Thus as the bath temperature rises the heater power is reduced, which consequently tends to lower the temperature back down. In this way the temperature is maintained at a
-
-
fairly constant temperature.
The temperature stability of the bath depends on the width of the proportional band. See Figure 5. If the band is too wide the bath temperature will deviate ex cessively from the set-point due to varying external conditions.This is because the power output changes very little with temperature and the controller cannot respond very well to changing conditions or noise in the system. If the proportional band is too narrow the bath temperature may swing back and forth because the controller overreacts to temperature variations. For best control stability the proportional band must be set for the optimum width.
The optimum proportional band width depends on several factors among which are fluid volume, fluid characteristics (viscosity, specific heat, thermal con ductivity), operating temperature, and stirring. Thus the proportional band width may require adjustment for best bath stability when any of these conditions
-
-
22
Controller Operation
Proportional Band too Narrow Proportional Band too Wide
Optimum Proportional Band
Figure 5
change. Of these, the most significant factors affect­ing the optimum proportional band width are the fluid viscosity and thermal noise due to difference in tem­perature between the fluid and ambient. The propor­tional band should be wider when the fluid viscosity is higher because of the increased response time and also when noise is greater.
The proportional band width is easily adjusted from the bath front panel.The width may be set to discrete values in degrees C or F depending on the selected units. The optimum proportional band width setting may be determined by monitoring the stability with a high resolution thermometer or with the controller percent output powerdisplay. Narrow the proportional band width to the point at which the bath temperature begins to oscillate and then increase the band width from this point to 3 or 4 times wider.
The proportional band adjustment may be accessed within the secondary menu. Press “SET” and “EXIT” to enter the secondary menu and show the heater power. Then press “SET” to access the proportional band.
S+E
Bath temperature flucuations at various proportional band settings.
Access heater power in secondary menu
12 Pct
Heater power in percent
S
To change the proportional band press “UP” or “DOWN”.
D
To accept the new setting and access the cutout set-point press “SET”. Pressing “EXIT” will exit the secondary menu ignoring any changes just made to the proportional band value.
S
Access proportional band
Pb=0.101C
Decrement display
Pb=0.060C
Accept the new proportional band setting
Proportional band setting
New proportional band setting
6.8 Cutout
As a protection against software or hardware fault, shorted heater triac, or user error, the bath is equipped with an adjustable heater cutout device that will shut off power to the heater if the bath tempera ture exceeds a set value.This protects the heater and bath materials from excessive temperatures and,
-
23
Controller Operation
most importantly, protects the bath fluids from being heated beyond the safe operating temperature pre venting hazardous vaporization, breakdown, or igni tion of the liquid. The cutout temperature is programmable by the operator from the front panel of the controller. It must always be set below the upper temperature limit of the fluid and no more than 10 de grees above the upper temperature limit of the bath.
-
-
-
S
The next function is the configuration menu. Press “EXIT” to resume displaying the bath temperature.
Accept cutout set-point
6.9 Controller Configuration
If the cutout is activated because of excessive bath temperature then power to the heater will be shut off and the bath will cool. The bath will cool until it reaches a few degrees below the cutout set-point temperature. At this point the action of the cutout is determined by the setting of the cutout mode parame ter. The cutout has two modes — automatic reset or manual reset. If the mode is set to automatic, then the cutout will automatically reset itself when the bath temperature falls below the reset temperature allow ing the bath toheat up again.If the mode isset to man ual, then the heater will remain disabled until the user manually resets the cutout.
The cutout set-point may be accessed within the sec­ondary menu. Press “SET” and “EXIT” to enter the secondary menu and show the heater power. Then press “SET” twice to access the cutout set-point.
S+E
12 Pct
S
S
To change the cutout set-point press “UP” or “DOWN”.
Access proportional band
Pb=0.101C
Access cutout set-point
CO= 210C
Access heater power in secondary menu
Heater power in percent
Proportional band setting
Cutout set-point
The controller has a number of configuration and op erating options and calibration parameters which are programmable via the front panel. These are ac cessed from the secondary menu after the cutout set-point function by pressing “SET”. The display will
-
-
-
prompt with “COnFIG”. Press “SET” once more. There are 5 sets ofconfiguration parameters — probe parameters, operating parameters, serial interface parameters, IEEE-488 interface parameters, and controller calibration parameters. The menus are se lected using the “UP” and “DOWN” keys and then pressing “SET”. See Figure 4.
-
-
-
6.10 Probe Parameters Menu
The probe parameter menu is indicated by,
PrObE
Press “SET” to enter the menu. The probe parame­ters menu contains the parameters, R0 and ALPHA, which characterize the resistance-temperature rela tionship of the platinum control probe.These parame ters may be adjusted to improve the accuracy of the bath.This procedure is explained in detail inSection.
The probe parameters are accessed by pressing “SET” after the name of the parameter is displayed. The value of theparameter may be changed using the “UP” and “DOWN” buttons. After the desired value is reached press “SET” to set the parameter to the new value.Pressing “EXIT” will cause the parameter to be skipped ignoring any changes that may have been made.
Probe parameters menu
-
-
D
To accept the new cutout set-point press “SET”.
24
Decrement display
CO= 95C
New cutout set-point
6.10.1 R0
This probe parameter refers to the resistance of the control probe at 0°C. Normally this is set for 100.000 ohms.
Controller Operation
6.10.2 ALPHA
This probe parameter refers to the average sensitivity of the probe between 0 and 100°C. Normally this is set for 0.00385°C
-1
.
6.11 Operating Parameters
The operating parameters menu is indicated by,
PAr
Press “SET” to enter the menu. The operating param eters menu contains the cutout reset mode parame ter.
6.11.1 Cutout Reset Mode
The cutout reset mode determines whether the cutout resets automatically when the bath temperature drops to a safe value or must be manually reset by the operator.
Operating parameters menu
Str Act
Press “SET” to access the parameter setting.
Str=Auto
To change the setting to always on press the “UP” or “DOWN” buttons and then “SET”. When set to “Str=ON” the stirrer motor comes on with the bath power regardless of the temperature set in the stirrer set-point parameter.
-
-
6.11.3 Stirrer set-point
Allows setting of the temperature above which the stirrer motor will activate when the stirrer activation is set to automatic.
To access the paramter press “SET” from the stirrer activation parameter. set-point mode selection pa­rameter is indicated by,
Stirrer mode selection parame ter
Stirrer is set for automatic acti vation at the stirrer set-point temperature.
-
-
The parameter is indicated by,
CtorSt
Press “SET” to access the parameter setting. Normally the cutout is set for automatic mode.
Cto=Auto
To change to manual reset mode press “UP” and then “SET”.
Cto=rSt
Cutout reset mode parameter
Cutout set for automatic reset
Cutout set for manual reset
6.11.2 Stirrer Mode Select
This parameter along with the Stirrer set-point allow the user to set the temperature at which the stirrer motor is activated. This setting is generally used when salt is used for the bath medium. For example, you can set the mode to “auto” and the temperature to 200°C. This allows the stirrer motor to shut off and turn on only when the salt is a liquid (>200°C) prevent ing the stirrer motor from overheating and damage.
The parameter is indicated by,
Str SEt
Press “SET” to access the parameter value.
Str=200
Press “UP” or “DOWN” to change the value and then “SET” to enter the new value.
Stirrer motor activation set-point parameter
Stirrer motor activation set-point
6.12 Serial Interface Parameters
The serial RS-232 interface parameters menu is indi cated by,
SErIAL
The serial interface parameters menu contains pa rameters which determine the operation of the serial interface. These controls only apply to baths fitted
-
with the serial interface. The parameters in the menu are — BAUD rate, sample period, duplex mode, and linefeed.
Serial RS-232 interface param eters menu
-
-
-
25
Controller Operation
6.12.1 BAUD Rate
The BAUD rate is the first parameter in the menu.The BAUD rate setting determines the serial communica tions transmission rate.
The BAUD rate parameter is indicated by,
BAUd
Press “SET”to choose to set the BAUD rate. The cur rent BAUD rate value will then be displayed.
1200 b
The BAUD rate of the bath serial communications may be programmed to 300,600,1200, or 2400 BAUD. Use “UP”or “DOWN” to change the BAUD rate value.
2400 b
Press “SET” to set the BAUD rate to the new value or “EXIT” to abort the operation and skip to the next pa­rameter in the menu.
Serial BAUD rate parameter
Current BAUD rate
New BAUD rate
6.12.2 Sample Period
The sample period is the next parameter in the serial interface parameter menu. The sample period is the time period in seconds between temperature mea surements transmitted from the serial interface. If the sample rate is set to 5 for instance then the bath will transmit the current measurement over the serial in terface approximately every five seconds. The auto matic sampling is disabled with a sample period of 0. The sample period is indicated by,
SAmPLE
Press “SET” to choose to set the sample period. The current sample period value will be displayed.
Serial sample period parameter
6.12.3 Duplex Mode
The next parameter is the duplex mode. The duplex
-
-
mode may be set to full duplex or half duplex.With full duplex any commands received by the bath via the serial interface will be immediately echoed or trans mitted back to the device of origin. With half duplex the commands will be executed but not echoed. The duplex mode parameter is indicated by,
dUPL
Press “SET” to access the mode setting.
dUP=FULL
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
dUP=HALF
Serial duplex mode parameter
Current duplex mode setting
New duplex mode setting
-
6.12.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This parameter enables (on) or dis­ables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return. The linefeed parameter is indicated by,
-
Press “SET” to access the linefeed parameter.
-
-
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF
LF= On
LF= OFF
Serial linefeed parameter
Current linefeed setting
New linefeed setting
SA= 1
Adjust the value with “UP” or “DOWN” and then use “SET” to set the sample rate to the displayed value.
SA= 60
26
Current sample period (sec onds)
New sample period
-
6.13 IEEE-488 Parameters Menu
Baths may optionally be fitted with an IEEE-488 GPIB interface. In this case the user may set the interface address within the IEEE-488 parameter menu. This menu does not appear on baths not fitted with the in terface. The menu is indicated by,
-
Controller Operation
IEEE
Press “SET” to enter the menu.
IEEE-488 parameters menu
6.13.1 IEEE-488 Address
The IEEE-488 interface must be configured to use the same address as the external communicating device. The address is indicated by,
AddrESS
Press “SET” to access the address setting.
Add= 22
Adjust the value with “UP” or “DOWN” and then use “SET” to set the address to the displayed value.
Add= 15
IEEE-488 interface address
Current IEEE-488 interface ad dress
New IEEE-488 interface ad­dress
-
The calibration parameters menu is indicated by,
CAL
Press “SET” five times to enter the menu.
Calibration parameters menu
6.14.1 CTO
Parameter CTO sets the calibration of the over-tem perature cutout. This is not adjustable by software but is adjusted with an internal potentiometer. For the 6054 bath this parameter should read between 560 and 570.
6.14.2 CO and CG
These parameters calibrate the accuracy of the bath set-point. These are programmed at the factory when the bath is calibrated. Do not alter the value of these parameters.If the user desires to calibrate the bath for improved accuracy then calibrate R0 and ALPHA ac­cording to the procedure given in Section 10.
-
6.14 Calibration Parameters
The operator of the bath controller has access to a number of the bath calibration constants, namely CTO, C0, CG, H, and L. These values are set at the factory and must not be altered. The correct values are important to the accuracy and proper and safe op­eration of the bath. Access to these parameters is available to the user only so that in the event that the controller’s memory fails the user may restore these values to the factory settings.The user should have a list of these constants and their settings with the man ual.
DO NOT change the values of the bath calibration constants from the factory set values.The correct setting of these parameters is important to the safety and proper operation of the bath.
6.14.3 H and L
These parameters set the upper and lower set-point limits of the bath. DO NOT change the values of these parameters from the factory set values. To do so may present danger of the bath overheating and causing damage or fire.
6.15 Operation Summary
­A complete flowchart of controller operation is shown
in Figure 4. This chart may be reproduced and used as a reference and operating guide.
27
Digital Communication Interface
9NC
9NC
7 Digital Communication Interface
If supplied with the option, the 6025/6035 bath is ca pable of communicating with and being controlled by other equipment through the digital interface. Two types of digital interface are available — the RS-232 serial interface and the IEEE-488 GPIB interface.
7.1 Serial Communications
The bath may be installed with an RS-232 serial inter face that allows serial digital communications over fairly long distances. With the serial interface the user may access any of the functions, parameters and set tings discussed in Section 6 with the exception of the BAUD rate setting. The serial interface operates with 8 data bits, 1 stop bit, and no parity. The use of a shielded communications cable is recommended.
7.1.1 Wiring
The serial communications cable attaches to the bath through the D-9 connector on the back panel.Figure 6 shows the pin-out of this connector and suggested cable wiring.
7.1.2 Setup
Before operation, the serial interface of the bath must first be set up by programming the BAUD rate and other configuration parameters. These parameters are programmed within the serial interface menu. The serial interface parameters menu is outlined in Figure
4 on page 20.
-
showing “BAUd”. Press “SET” to choose to set the BAUD rate. The current BAUD rate value will then be displayed. The BAUD rate of the serial communica
­tions may be programmed to 300, 600, 1200, or 2400 BAUD. The BAUD rate is pre-programmed to 1200 BAUD. Use “UP”or “DOWN” to change the BAUD rate value. Press “SET” to set the BAUD rate to the new value or “EXIT” to abort the operation and skip to the next parameter in the menu.
-
7.1.2.2 Sample Period
The sample period is the next parameter in the menu
-
and prompted with “SAmPLE”. The sample period is the time period in seconds between temperature measurements transmitted from the serial interface.If the sample rate is set to 5 for instance then the bath will transmit the current measurement over the serial
RS-232 Cable Wiring for
IBM PC and Compatibles
6054 Controller
Connector
(DB 9-Pin)
1NC 2 RxD 3 TxD 4NC 5 GND 6NC 7RTS 8 CTS 9NC
Computer (DTE)
Connector
(DB 25-Pin)
2 TxD 3 RxD 4RTS 5 CTS 6 7 GND 8 20
To enter the serial parameter programming mode first press “EXIT” while pressing “SET” and release to en ter the secondary menu. Press “SET” repeatedly until the display reads “ProbE”.This is the menu selection. Press “UP” repeatedly until the serial interface menu is indicated with “SErIAL”. Finally press “SET” to enter the serial interface parameters menu. In the serial in terface parameters menu are the BAUD rate, sample rate, duplex mode, and linefeed parameters.
7.1.2.1 BAUD rate
The BAUD rate is the first parameter in the menu.The display will prompt with the BAUD rate parameter by
28
-
-
6054 Controller
Connector
(DB 9-Pin)
1NC 2 RxD 3 TxD 4NC 5 GND 6NC 7RTS 8 CTS
Figure 6
Computer (DTE)
Connector
(DB 9-Pin)
1NC 2 RxD 3 TxD 4NC 5 GND 6NC 7RTS 8 CTS
Serial Cable Wiring Diagram
Digital Communication Interface
interface approximately every five seconds.The auto matic sampling is disabled with a sample period of 0. Press “SET” to choose to set the sample period. Ad just the period with “UP” or “DOWN” and then use “SET” to set the sample rate to the displayed value.
7.1.2.3 Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full duplex any commands received by the bath via the serial inter face will be immediately echoed or transmitted back to the device of origin.With half duplex the commands will be executed but not echoed. The default setting is full duplex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
7.1.2.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This parameter enables (“On”) or dis ables (“OFF”) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-re­turn. The default setting is with linefeed on. The mode may be changed using “UP”or “DOWN”and pressing “SET”.
7.1.3 Serial Operation
Once the cable has been attached and the interface set up properly the controller will immediately begin transmitting temperature readings at the pro grammed rate. The serial interface operates with 8 data bits, 1 stop bit, and no parity. The set-point and other commands may be sent to thebath via the serial interface to set the bath and view or program the vari ous parameters. The interface commands are dis cussed in Section 7.3. All commands are ASCII character strings terminated with a carriage-return character (CR, ASCII 13).
7.2 IEEE-488 Communication (optional)
The IEEE-488 interface is available as an option. Baths supplied with this option may be connected to a GPIB type communication bus which allows many in struments to be connected and controlled simulta neously.
-
-
-
-
7.2.1 Setup and Address Selection
To use the IEEE-488 interface first connect an IEEE-488 standard cable to the back of the bath.
Next set the device address. This parameter is pro grammed within the IEEE-488 interface menu. To en ter the IEEE-488 parameter programming menu first press “EXIT” while pressing “SET” and release to en ter the secondary menu. Press “SET” repeatedly until the display reaches “PrObE”.This is the menu selec tion. Press “UP” repeatedly until the IEEE-488 inter face menu is indicated with “IEEE”. Press “SET” to enter the IEEE-488 parameter menu. The IEEE-488 menu contains the IEEE-488 address parameter.
The IEEE-488 address is prompted with “AddrESS”. Press “SET” to program the address. The default ad dress is 22. Change the device address of the bath if necessary to match the address used by the commu nication equipment by pressing “UP” or “DOWN” and then “SET”.
-
-
-
-
-
-
-
7.2.2 IEEE-488 Operation
Commands may now be sent via the IEEE-488 inter­face to read or set the temperature or access other controller functions. All commands are ASCII charac­ter strings and are terminated with a carriage-return (CR, ASCII 13). Interface commands are listed below.
-
7.3 Interface Commands
-
-
-
-
The various commands for accessing the bath con troller functions via the digital interfaces are listed in this section (see Ta bl e 2 ).These commands are used with both the RS-232 serial interface and the IEEE-488 GPIB interface. In either case the com mands are terminated with a carriage-return charac ter.The interface makes no distinction between upper and lower case letters, hence either may be used. Commands may be abbreviated tothe minimum num ber of letters which determines a unique command. A command may be used to either set a parameter or display a parameter depending on whether or not a value is sent with the command following a “=” charac ter. For example “s”<CR> will return the current set-point and “s=50.00”<CR> will set the set-point (set-point 1) to 50.00 degrees.
In Ta bl e 2 , characters or data within brackets, “[” and “]”, are optional for the command.A slash, “/”, denotes
-
-
-
-
-
29
Digital Communication Interface
Table 2
Command Description
Communications Commands
Command Format
Command Example Returned
Returned Example
Acceptable Values
Display Temperature
Read current set-point s[etpoint] s set: 9999.99 {C or F} set: 150.00 C
Set current set-point to n s[etpoint]=n s=450 Instrument Range
Read vernier v[ernier] v v: 9.99999 v: 0.00000
Set vernier to n v[ernier]=n v=.00001 Depends on
Read temperature t[emperature] t t: 9999.99 {C or F} t: 55.69 C
Read temperature units u[nits] u u: x u: c
Set temperature units: u[nits]=c/f
Set temperature units to
u[nits]=c u=c
Configuration
C or F
Celsius Set temperature units to
u[nits]=f u=f
Fahrenheit
Secondary Menu
Read proportional band setting pr[op-band] pr pr: 999.9 pr: 15.9
Set proportional band to
Read cutout setting c[utout] c c: 9999 {x},{xxx} c: 620 C, in
Set cutout setting: c[utout]=n/r[eset]
n
Set cutout tondegrees
Reset cutout now
Read heater power
(duty cycle)
pr[op-band]=n pr=8.83 Depends on
Configuration
c[utout]=n c=500 Temperature
Range
c[utout]=r[eset] c=r
po[wer] po po: 9999 po: 1
Configuration Menu
Probe Menu
Read R0 calibration parameter r[0] r r0: 999.999 r0: 100.578
n
Set R0 calibration parameter to
Read ALPHA calibration parameter al[pha] al al: 9.9999999 al: 0.0038573
Set ALPHA calibration parameter to
Operating Parameters Menu
Read cutout mode cm[ode] cm cm: {xxxx} cm: AUTO
Set cutout mode: cm[ode]=r[eset]/a[uto]
Set cutout to be reset manually-
Set cutout to be reset automatically
Read Stirrer Mode smod smod smod: {xxxx} smod: AUTO
Set Stirrer Mode: smod=o/a
Set striier activation mode to on
Set striier activation mode to auto
Read Stirrer Set-Point sset sset sset: 9999.99 {C or F} sset: 160.00C
Set Stirrer Set-Point to
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 9 sa: 1
Set serial sampling setting to seconds
Set serial duplex mode: du[plex]=f[ull]/h[alf]
n
n
Set serial duplex mode to full Set serial duplex mode to half
Set serial linefeed mode: lf[eed]=on/of[f]
r[0]=n r=100.324 98.0 to 104.9
n
al[pha]=n al=0.0038433 .00370 to .00399
RESET or AUTO
cm[ode]=r[eset] cm=r
cm[ode]=a[uto] cm=a
ON or AUTO
smod=o smod=o
smod=a smode=a
sset=n sset=160 instrument range
sa[mple]=n sa=0 0 to 4000
FULL or HALF
du[plex]=f[ull] du=f
du[plex]=h[alf] du=h
ON or OFF
30
Communications Commands, cont.
Digital Communication Interface
Command
Command Description
Set serial linefeed mode to on Set serial linefeed mode to off
Calibration Menu
Read C0 calibration parameter *c0 *c0 c0: 9 c0: 0
Set C0 calibration parameter to
Read CG calibration parameter *cg *cg cg: 999.99 cg: 406.25
Set CG calibration parameter to
Read low set-point limit value *tl[ow] *tl tl: 999 tl: –80
Set low set-point limit to
Read high set-point limit value *th[igh] *th th: 999 th: 205
Set high set-point limit to
Miscellaneous (not on menus)
Read firmware version number *ver[sion] *ver ver.9999,9.99 ver.2100,3.56
Read structure of all commands h[elp] h list of commands
Read Heater 1 f1 f1 f1:9 f1:1
Set Heater 1 f1=1/0 1 or 0
Set heater 1 to on Set heater 1 to off
Read Heater 2 f2 f2 f2:9 f2:0
Set Heater 2 f2=1/0 1 or 0
Set Heater 2 to on Set Heater 2 to off
Read Heater 3 f3 f3 f3:9 f3:1
Set Heater 3 f3=1/0 1 or 0
Set Heater 3 to on Set Heater 3 to off
Read Heater 4 f4 f4 f4:9 f4:1
Set Heater 4 f4=1/0 1 or 0
Set Heater 4 to on Set Heater 4 to off
Read Boost Heater f5 f5 f5:9 f5:1
Set Boost Heater f5=1/0 1 or 0
Set Boost Heater to on Set Boost Heater to off Legend:
Note:
n
n
Format
lf[eed]=on lf=on
lf[eed]=of[f] lf=of
n
*c0=n *c0=0 –999.9 to 999.9
n
*cg=n *cg=406.25 –999.9 to 999.9
*tl[ow]=n *tl=-80 –999.9 to 999.9
*th[igh]=n *th=205 –999.9 to 999.9
f1=
n
f1=
n
f2=
n
f2=
n
f3=
n
f3=
n
f4=
n
f4=
n
f5=
n
f5=
n
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Command Example Returned
f1=1
f1=0
f2=1
f2=0
f3=1
f3=0
f4=1
f4=0
f5=1
f5=0
Returned Example
Acceptable Values
alternate characters or data. Numeric data, denoted by “n”, may be entered in decimal or exponential nota tion. Bold type characters are literal characters while normal type symbolizes data. Characters are shown in lower case although upper case may be used. Spaces may be added within command strings and will simply be ignored. Backspace (BS, ASCII 8) may be used to erase the previous character. A terminat ing CR is implied with all commands.
-
7.4 Power Commands
The digital interface is capable of controlling the heat ing functions so that the bath can be remotely oper
-
-
ated at any temperature within the range of the bath.
To allow the interface to control the heating, the front
-
panel controls are disabled by 1) switching the heater control to “OFF”, and 2) switching the boost heater
31
Digital Communication Interface
switch to “OFF”
able to switch these functions off. The 6054 bath has five control functions with the digital interface. These are controls for heaters 1, 2, 3, and 4, each 250 W, and the boost heater which is 1000 W. Heater 1 power is variable with the DRIFT control on the front panel. The boost heater should only be used for quickly heating the bath up to a high temperature and not for controlling at a constant temperature.
Serial commands “F1” through “F5” control the heat ers 1 through 4 and the boost heater individually. These commands are used to turn the heaters on or off or to read thestates of the heaters.Sending a com mand with parameter “1” turns the heater on.Parame ter “0” turns the heater off. No parameter returns the state, “1” for on or “0” for off.For example “F1=1”<RE TURN> turns on heater 1. “F1”<RETURN> (no pa
.Otherwise, the interface would not be
rameter) will return “f1:1” or “f1:0” depending on whether heater 1 is on or off respectively.
Since unlike the front panel heater control, which turns on multiple heaters to achieve the desired power level, the interface commands control the heat ers individually, multiple commands mustbe issued to set the desired amount of power.Ta b le 4 lists the com mands which should be given to set various power levels. Power is variable with the front panel DRIFT
-
-
-
-
-
control when heater 1 (F1) is turned on.
7.5 Heater Settings for Control
Suggested heater control settings for various ranges are suggested in Ta bl e 5 below.
-
-
Table 4
POWER F1 F2 F3 F4 F5
OFF 00000
0–250 W LOW 10000
250–500 W MEDIUM
500–750 W MEDIUM HIGH
750–1000 W HIGH
‘750 W BOOST 11111
32
Heater Power Settings
11000
11100
11110
Table 5
Range Controls 50°C–125°C LOW - Heater 1 on, heaters 2–4
125°C–200°C MEDIUM - Heaters 1 and 2 on,
200°C–275°C MEDIUM HIGH - Heaters 1–3 on
275°C–325°C HIGH - Heaters 1–4 and boost
Suggested Heater Settings
and boost heater off
heater 3, 4, and boost heater off
and heater 4 and boost heater off
heater on
8 Bath Calibration
Bath Calibration
In some instances the user may want to calibrate the bath to improve the temperature set-point accuracy. Calibration is done by adjusting the controller probe calibration constants R0 and ALPHA so that the tem perature of the bath as measured with a standard thermometer agrees more closely with the bath set-point. The thermometer used must be able to measure the bath fluid temperature with higher accu racy than the desired accuracy of the bath.By using a good thermometer and carefully following procedure the bath can be calibrated to an accuracy of better than 0.02°C over a range of 100 degrees.
R0= 100.000
ALPHA = 0.0038500
t
= 80.00°C
L
measured t = 79.843°C
t
= 120.00°C
H
measured t = 119.914°C
8.1 Calibration Points
In calibrating the bath R0 and ALPHA are adjusted to
-
-
minimize the set-point error at each of two different bath temperatures. Any two reasonably separated bath temperatures may be used for the calibration however best results will be obtainedwhen using bath temperatures which are just within the most useful operating range of the bath.The further apart the cali bration temperatures the larger will be the calibrated temperature range but the calibration error will also be greater over the range. If for instance 50°C and 150°C are chosen as the calibration temperatures then the bath may achieve an accuracy of say ±0.03°C over the range 40 to 160°C. Choosing 80°C
-
Compute errors,
err
= 79.843 - 80.00°C = -0.157°C
L
err
= 119.914 - 120.00°C = -0.086°C
H
Compute R
0086 800 0157 1200
−−
R
(. ) . (. ) .
0
⎢ ⎣
Compute ALPHA,
ALPHA
Figure 7
′=
Calibration Example
,
0
xx
1200 80 0
..
+−+(. .)(.)(. .)(1 0 00385 120 0 0157 1 0 00385 80 0xx
⎢ ⎣
1200 80 0
000385′=
.
..
+
1 100 000 100115
⎥ ⎦
=..
0086
.)
1 0 00385 0 0038387
..
+
⎥ ⎦
=
33
Bath Calibration
and 120°C may allow the bath to have a better accu racy of maybe ±0.01°C over the range75 to 125°C but outside that range theaccuracy may be only±0.05°C.
8.2 Measuring the Set-point Error
The first step in the calibration procedure is to mea sure the temperature errors (including sign) at the two calibration temperatures. First set the bath to the lower set-point which we will callt
.Wait forthe bath to
L
reach the set-point and allow 15 minutes to stabilize at that temperature. Check the bath stability with the thermometer.When both the bath and the thermome ter have stabilized measure the bath temperature with the thermometer and compute the temperature error err
which is the actual bath temperature minus
L
the set-point temperature. If for example the bath is set for a lower set-point of t
=50°C and the bath
L
reaches a measured temperature of 49.7°C then the error is -0.3°C.
Next, set the bath for the upper set-point t
and after
H
stabilizing measure the bath temperature and com­pute the error err
. For our example we will suppose
H
the bath was set for 150°C and the thermometer mea­sured 150.1°C giving an error of +0.1°C.
8.3 Computing R0 and ALPHA
Before computing the new values for R0 and ALPHA the current values must be known.The values may be found by either accessing the probe calibration menu from the controller panel or by inquiring through the digital interface. The user should keep a record of these values in case they may need to be restored in
-
the future. The new values R0and ALPHAare com
­puted by entering the old values for R0 and ALPHA, the calibration temperature set-points t the temperature errors err
and errHinto the following
L
and tH, and
L
equations,
err t err t
R
010⎡=
⎡ ⎢ ⎣
HL LH
tt
HL
ALPHA R
+
⎥ ⎦
-
ALPHA
⎡=
⎡ ⎢ ⎣
HL LH
tt
HL
+
1
ALPHA
⎥ ⎦
ALPHA t err ALPHA t err
+−+
()()11
If for example R0 and ALPHA were previously set for
-
100.000 and 0.0038500 respectively and the data for t
, errL, and errHwere as given above then the new
L,tH
values R0and ALPHAwould be computed as
100.193 and 0.0038272 respectively. Program the new values R0 and ALPHA into the controller. Check the calibration by setting the temperature to t
and t
L
H
and measuring the errors again.If desired the calibra­tion procedure may be repeated again to further im­prove the accuracy.
8.4 Calibration Example
The bath is to be used between 75 and 125°C and it is desired to calibrate the bath as accurately as possible for operation within this range. The current values for R0 and ALPHA are 100.000 and 0.0038500 respec tively. The calibration points are chosen to be 80.00 and 120.00°C. The measured bath temperatures are
79.843 and 119.914°C respectively. Refer to Figure 7 for applying equations to the example data and com puting the new probe constants.
-
-
34
9 Bath Heat Transfer Fluid
Bath Heat Transfer Fluid
Many fluids will work with 6054 bath.Choosing a fluid requires consideration of many important character istics of the fluid. Among these are temperature range, viscosity, specific heat, thermal conductivity, thermal expansion, electrical conductivity, fluid life time, safety, and cost.
9.1 Temperature Range
One of the most important characteristics to consider is the temperature range of the fluid. Few fluids work well throughout the entire temperature range of the bath. The temperature at which the bath is operated must always be within the safe and useful tempera ture range of the fluid used. The lower temperature range of the fluid is determined either by the freeze point of the fluid or the temperature at which the vis­cosity becomes too great. The upper temperature is usually limited by vaporization, flammability, or chem­ical breakdown of the fluid. Vaporization of the fluid at higher temperatures may adversely affect tempera­ture stability because of cool condensed fluid dripping into the bath from the lid.
compensate for the reduced response time. Other
-
-
-
wise the temperature may begin to oscillate.
9.3 Specific heat
Specific heat is the measure of the heat storage abil ity of the fluid. Specific heat, to a small degree, affects the control stability. It also affects the heating and cooling rates. Generally, a lower specific heat means quicker heating and cooling. The proportional band may require some adjustment depending on the spe cific heat of the fluid.
-
-
-
9.4 Thermal Conductivity
Thermal conductivity measures how easily heat flows through the fluid. Thermal conductivity of the fluid af­fects the control stability, temperature uniformity, and temperature settling time. Fluids with higher conduc­tivity distribute heat more quickly and evenly improv­ing bath performance.
The bath temperature should be limited by setting the safety cutout so that the bath temperature cannot ex ceed the safe operating temperature limit of the fluid.
9.2 Viscosity
Viscosity is a measure of the thickness of a fluid or how easily it can be poured and mixed. Viscosity af fects the temperature uniformity and stability of the bath. With lower viscosity, fluid mixing is better there fore creating a more uniform temperature throughout the bath. This improves the bath response time which allows it to maintain a more constant temperature. For good control the viscosity should be less than 10 centistokes. The practical upper limit of allowable vis cosity is about 50 centistokes. Viscosity greater than this causes very poor control stability because of poor stirring and may also overheat or damage the stirring motor. Viscosity may vary greatly with temperature, especially with oils.
When using fluids with higher viscosities the control ler proportional band may need to be increased to
-
-
-
9.5 Thermal Expansion
Thermal expansion describes how much the volume of the fluid changes with temperature. Thermal ex pansion of the fluid must be considered since the in crease in fluid volume as the bath temperature increases may cause overflow. Excessive thermal ex pansion may also be undesirable in applications where constant liquid level is important. Many fluids including oils have significant thermal expansion.
-
-
-
9.6 Electrical Resistivity
-
-
Electrical resistivity describes how well the fluid insu lates against the flow of electric current. In some ap plications, such as measuring the resistance of bare temperature sensors, it may be important that little or no electrical leakage occur through the fluid. In such conditions choose a fluid with very high electrical re sistivity.
-
-
-
35
Bath Heat Transfer Fluid
9.7 Fluid lifetime
Many fluids degrade over time because of vaporiza tion, water absorption, gelling, or chemical break down. Often the degradation becomes significant near the upper temperature limit of the fluid, substan tially reducing the fluid’s lifetime.
9.8 Safety
When choosing a fluid always consider the safety is sues associated. Obviously where there are extreme temperatures there can be danger to personnel and equipment. Fluids may also be hazardous for other reasons. Some fluids may be considered toxic. Con tact with eyes, skin, or inhalation of vapors may cause injury.A proper fume hood must be used if hazardous or bothersome vapors are produced.
Fluids may be flammable and require special fire safety equipment and procedures.
An important characteristic of the fluid to consider is the flash point. The flash point is the temperature at which there is sufficient vapor given off so that when there is adequate oxygen present and an ignition source is applied the vapor will ignite. This does not necessarily mean that fire will be sustained at the flash point. The flash point may be either of the open cup or closed cup type. Either condition may occur in a bath situation.
The open cup flash point is measured under the con dition of vapors escaping the tank. The closed cup flash point is measured with the vapors being con tained within the tank. Since oxygen and an ignition source is less available inside the tank the closed cup flash point will belower than the opencup flash point.
Remember also that environmentally hazardous flu ids require special disposal according to applicable federal or local laws after use.
9.9 Cost
Cost of bath fluids may vary greatly, from cents per gallon for water to hundreds of dollars per gallon for synthetic oils. Cost may be an important consider ation when choosing a fluid.
9.9.1 Commonly used fluids
-
-
-
-
-
Below is a description of some of the more commonly used fluids and their characteristics.
9.9.2 Water
Water is often used because of its very low cost, its availability, and its excellent temperaturecontrol char acteristics. Water has very low viscosity and good thermal conductivity and heat capacity which makes it among the best fluids for good control stability at lower temperatures. Temperature stability is much poorer at higher temperatures because water con denses on the lid, cools and drips into the bath.Water is safe and relatively inert. The electrical conductivity of water may prevent its use in some applications. Water has a limited temperature range, from a few de grees above 0°C to a few degrees below 100°C. At higher temperatures evaporation becomes signifi­cant. Water used in the bath should be distilled or deionized to prevent mineral deposits. Consider using an algicide chemical in the water to prevent contami­nation.
-
-
-
9.9.3 Ethylene Glycol
The temperature range of water may be extended by using a solution of 1 part water and 1 part ethylene glycol (antifreeze).The characteristics of the ethylene
-
-
glycol-water solution are similar to water but with higher viscosity. Use caution with ethylene glycol since this fluid is very toxic. Ethylene glycol must be disposed of properly.
9.9.4 Mineral Oil
­Mineral oil or paraffin oil is often used at moderate
temperatures above the range of water. Mineral oil is relatively inexpensive. At lower temperatures mineral oil is quite viscous and control may be poor. At higher temperatures vapor emission becomes significant. The vapors may be dangerous and use of a fume hood is highly recommended. As with most oils ,min
-
eral oil will expand as temperature increases, there fore, be careful not to fill the bath too full that it overflows when heated. The viscosity and thermal characteristics of mineral oil is poorer than water so temperature stability will not be as good. Mineral oil has very low electrical conductivity. Use caution with
-
-
36
Bath Heat Transfer Fluid
Table 6
Halocarbon 0.8 #5019
Methanol –96°C (fr) 60°C (b) 54°C 1.3 @ –35°C
Water 0°C (fr) 95°C (b) NONE 1 @ 25°C
Ethylene Glycol—50% #5020
Mineral Oil 40°C (v) 190°C (fl) 190°C 15 @ 75°C
Dow Corning
200.5 Silicone Oil
Dow Corning
200.10 #5012
Dow Corning
200.20 #5013
Dow Corning
200.50 Silicone Oil
Dow Corning 550 #5016
Dow Corning 710 #5017
Dow Corning 210-H Silicone Oil
Table of Bath Fluids
Fluid
(# = Hart
Part No.)
Lower
Temperatur
e Limit*
–90°C (v)** 70°C (e) NONE 5.7 @ –50°C
–35°C (fr) 110°C (b) NONE 7 @ 0°C
–40°C (v)** 133°C (fl, cc) 133°C 5 @ 25°C 0.92 @ 25°C 0.4 0.00028 @
–35°C (v)** 165°C (fl, cc) 165°C 10 @ 25°C
7°C (v) 230°C (fl, cc) 230°C 20 @ 25°C 0.949 @ 25°C 0.370 @ 40°C
25°C (v) 280°C (fl, cc) 280°C 20 @ 25°C 0.96 @ 25°C 0.4 0.00037 @
70°C (v) 232°C (fl, cc)
80°C (v) 302°C (fl, oc) 302°C 50 @ 80°C
66°C (v) 315°C (fl, oc) 315°C 50 @ 66°C
Upper
Temperatur
e Limit*
300°C (fl, oc)
Flash Point
232°C 50 @ 70°C
Viscosity
(centistokes)
0.8 @ 40°C
0.5 @ 70°C
0.66 @ 0°C
0.45 @ 20°C
0.4 @ 75°C
2 @ 50°C
0.7 @ 100°C
5 @ 125°C
3 @ 135°C
10 @ 104°C
7 @ 204°C
14 @ 204°C
mineral oil since it is flammable and may also cause serious injury if inhaled or ingested.
Specific
Specific
Gravity
1.71 @ 40°C 0.2 0.0004 0.0011
0.810 @ 0°C
0.792 @ 20°C
1.00 1.00 0.0014 0.0002 @
1.05 0.8 @ 0°C 0.001
0.87 @ 25°C
0.84 @ 75°C
0.81 @ 125°C
0.934 @ 25°C 0.43 @ 40°C
1.07 @ 25°C 0.358 @ 40°C
1.11 @ 25°C 0.363 @ 40°C
0.96 @ 25°C 0.34 @
Heat
(cal/g/°C)
0.6 0.0005 @
0.48 @ 25°C
0.53 @ 75°C
0.57 @ 125°C
0.45 @ 100°C
0.482 @ 200°C
0.393 @ 100°C
0.420 @ 200°C
0.386 @ 100°C
0.433 @ 200°C
0.454 @ 100°C
0.505 @ 200°C
100°C
Thermal Conductivity (cal/s/cm/°C)
20°C
0.00025 @ 25°C
25°C
0.00032 @ 25°C
0.00034 @ 25°C
25°C
0.00035 @ 25°C
0.00035 @ 25°C
0.0003 0.00095 100 @ 25°C
Thermal
Expansion
(cm/cm/°C)
0.0014 @ 25°C
25°C
0.0007 @ 50°C
0.00105 1000 @ 25°C
0.00108 1000 @ 25°C
0.00107 1000 @ 25°C
0.00104 1000 @ 25°C
0.00075 100 @ 25°C
0.00077 100 @ 25°C
very high electrical resistivity. Silicone oils are fairly safe and non-toxic. Silicone oils are fairly expensive.
Resistivity
12
(10
-cm )
5 @ 25°C
10 @ 150°C
50 @ 150°C
50 @ 150°C
50 @ 150°C
1 @ 150°C
1 @ 150°C
1 @ 150°C
9.9.5 Silicone oil
Silicone oils are available which offer a much wider operating temperature range than mineral oil. Like most oils, silicone oils have temperature control char acteristics which are somewhat poorer than water. The viscosity changes significantly with temperature and thermal expansion also occurs. These oils have
9.9.6 Heat Transfer Salt
Heat transfer salt is often used at high bath tempera tures.Salt has a very highupper temperature limit and
-
a wide useful temperature range. Salt may be used from its freezing point at 145°C up to the upper tem perature limit of thebath. Viscosity islow, especially at higher temperatures. Salt when melted has very low electrical resistivity. Salt may be corrosive to some
-
-
37
Bath Heat Transfer Fluid
–100°C 0°C 100°C 200°C 300°C 400°C 500°C 600°C
Silicone Oil
710
Silicone Oil
200.50
Silicone Oil
200.20
Silicone Oil
200.10
Silicone Oil
200.05
Silicone Oil
550
Halocarbon
0.8
Mineral Oil
Water
Methanol
Ethylene Glycol
50/50 with H O
2
Bath Salt
PP
PP
PP
PP
PP
10 CS
FR (Pure)
FR
FR
10 CS
FL 12°C
10 CS
10 CS
10 CS
10 CS
EP
BP
BP
BP
FL 133°C
FR
10 CS
10 CS
FL 163°C
10 CS
FL 302°CPP
FL 280°C
FL 232°C
FL 230°C
FL 177°C
Shaded area represents usable range of fluid starting at 50 centistokes. Lighter shading represents decreasing viscosity, while vaporization and decomposition increase.
Black area represents liquid range with excessive viscosity.
Range over which a fume hood is recommended.
Figure 8
Chart showing usable range of various fluids.
materials. Salt will quickly oxidize the coating on gal vanized metal. Carbon steel may be used with salt up to 450°C. Beyond this temperature stainless steel is recommended.
Decomposition Starts
BP - Boiling Point CS - Centistokes EP - Evaporation Point, fluid loss
due to evaporation too high. FL - Flash Point FR - Freeze Point PP - Pour Point
-
CAUTION: Keep all combustible materials away from the bath when using salt. Operate the bath on a heat-proof surface such as concrete.Provide a means of safely containing any spills which may occur.
The greatest safety concern with liquid salt is with its high temperature. The fluid provides a source of heat
38
Bath Heat Transfer Fluid
which can ignite or destroy materials. Especially dan gerous are spills since the hot fluid is difficult to con tain. Salt may also cause steam explosions if it comes into contact with water. Using salt requires special caution. Read carefully the information and safety data sheets provided with thesalt. Use ofa fume hood is recommended to remove any products of decom position or oxidation.
Because salt is solid atroom temperature special pro cedures are required in using the bath with salt. The bath has been designed so that the stir motor will not turn on until the bath has reached the temperature for the salt to liquify.
Solid salt has poor thermal conductivity. Only operate the bath with the “HEATER” switch on position “LOW” until the salt is melted. Using higher heater power may over-heat and damage the bath heaters.The heaters will only run at 25% power until the bath reaches 200°C. This helps to protect the heaters during melt ing of the salt. This is especially possible when heat­ing salt which has solidified and contracted leaving gaps between the bath walls and the salt.
Salt is supplied in a pink granular form. Fill the bath gradually as the salt heats and melts. Because of the heat required to melt the salt this is a slow process and may take 10 hours or more. Fill the bath until the liquid is a few inches below the lid. The level will rise as much as two inches as it is heated to the upper temperature limit.
9.10 Fluid Characteristics
-
-
-
-
-
9.10.1 Limitations and Disclaimer
The information given in this manual regarding fluids is intended only to be used as a general guide in choosing a fluid.Though every effort has been made to provide correct information we cannot guarantee accuracy of data or assure suitability of a fluid for a particular application.Specifications may change and sources sometimes offer differing information. Hart Scientific cannot be liable for any personal injury or damage to equipment, product or facilities resulting from the use of these fluids.The user of the bath is re sponsible for collecting correct information, exercis ing proper judgement, and insuring safe operation. Operating near the limits ofcertain properties such as the flash point or viscosity can compromise safety or performance. Your company’s safety policies regard ing flash points,toxicity, and such issues must be con sidered. You are responsible for reading the MSDS (material safety data sheets)and acting accordingly.
-
-
-
-
9.10.2 About the Chart
The fluid chart (Figure 8 on page 38) visually illus­trates some of the important qualities of the fluids shown.
Temperature Range: The temperature scale is shown in degrees Celsius.The fluids’general range of appli­cation is indicated by the shaded bands. Qualities in­cluding pour point, freeze point, important viscosity points, flash point, boiling point and others may be shown.
Charts
Ta bl e 6 and Figure 8 on pages 37 and 38 have been
created to provide help in selecting a heat exchange fluid media foryour constant temperature bath.These charts provide both a visual and numerical represen tation of most of the physical qualities important in making a selection. The list is not all inclusive. There may be other useful fluids not shown in this listing.
The charts include information on a variety of fluids which are often used as heat transfer fluid in baths. Because of the temperature range some fluids may not be useful with your bath.
Freezing Point: The freezing point of a fluid is an obvi ous limitation to stirring. As the freezing point is ap proached high viscosity may also limit performance.
Pour Point: This represents a handling limit for the
-
fluid.
Viscosity: Points shown are at 50 and 10 centistokes viscosity. When viscosity is greater than 50 centi stokes stirring is very poor and the fluid is unsatisfac tory for bath applications. Optimum stirring generally occurs at 10 centistokes and below.
Fume Point: This is the point at which a fume hood is recommended. This point is very subjective in nature and is impacted by individual tolerance to different fumes and smells, how well the bath is covered, the surface area of the fluid in the bath, the size and venti lation of the facility where the bath is located and other
-
-
-
-
-
39
Bath Heat Transfer Fluid
conditions.We assume the bath is well covered at this point. This is also subject to company policy.
Flash Point: This is the point at which ignition may oc cur.The point shown may be either the open or closed cup flash point. Refer to the flash point discussion in
Section.
Boiling Point: At or near the boiling point of the fluid the temperature stability is difficult to maintain.
Fuming or evaporation is excessive. Large amounts of heater power may be required because of the heat
-
of vaporization.
Decomposition: The temperature may reach a point at which decomposition of the fluid begins. Further in creasing the temperature may accelerate decompo
-
-
sition to the point of danger or impracticality.
40
10 Troubleshooting
Troubleshooting
In the event that the user of the 6054 bath encounters difficulty in operation this section may help to find and solve the problem. Several possible problem condi tions are described along with likely causes and solu tions. If a problem arises please read this section carefully and attempt to understand and solve the problem. If the bath seems faulty or the problem can not otherwise be solved, then contact the factory for assistance.
10.1 The Heater Indicator LED Stays
Red But The Temperature Does
Not Increase.
Here the display does not show “cut-out” nor displays an incorrect bath temperature, but the controller oth­erwise appears to operate normally. The problem may be either insufficient heating or no heating at all. Insufficient heating may be caused by the heater power setting being too low, especially at higher oper­ating temperatures. Switching to a higher heater power switch setting may solve the problem. If one or more of the heaters are burned out, it also may cause this problem. If this seems the case contact the fac­tory for assistance.
If the heater seems not to be receiving power at all, then first check the heater fuse. If the fuse is burned out, then try replacing the fuse with a new one (of the same rating) and then check to see if the bath re sumes normal operation. If the fuse blows gain then there may be a shorted heater. Contact the factory. If the fuse was not blown in the first place then the prob lem may be a faulty solid-state relay or mechanical re lay. Contact the factory for assistance.
10.2 The Controller Display Flashes
“CUT-OUT” And The Heater
Does Not Operate.
cut-out disconnects power to the heater when the bath temperature exceeds the cut-out set-point. This
-
-
-
will cause the temperature to drop back down to a safe value. If the cut-out mode is set to “AUTO” then the heater will switch back on when the temperature drops. If the mode is set to “RESET” then the heater will only come on again when the temperature is re duced and the cut-out is manually reset by the opera tor. See Section 6.8. Check that the cut-out set-point is adjusted to 10 or 20°C above the desired bath oper ating temperature and that the cut-out mode is set as desired.
If the cut-out activates when the bath temperature is well belowthe cut-out set-point or the cut-out does not reset when the bath temperature drops and it is man ually reset, then the cut-out circuitry may be faulty or the cut-out thermocouple sensor may be faulty or dis­connected. Contact the factory for assistance.
-
-
-
-
10.3 The Display Flashes
“CUT-OUT” And An Incorrect
Process Temperature.
The problem may be that the controller’s voltmeter cir cuit is not functioning properly. This could be a prob lem with the memory back-up battery. If the battery power is insufficient to maintain the memory then data may become scrambled causing problems. A
-
-
-
nearby large static discharge may also affect data in memory. The memory may be reset by holding the “SET” and “EXIT” keys down while power to the con troller is switched on. The display will show “—init—” indicating the memory is being initialized. At this point, each of the controller parameters and calibra tion constants must be reprogrammed into memory. If this solves the problem but the same problem reoc curs then the battery should be replaced. Contact the factory. If initializing the memory does not remedy the problem then there may be a failed electronic compo nent. Contact the factory.
-
-
-
-
-
-
The display will flash “CUT-OUT” alternately with the process temperature. If the process temperature dis played seems grossly in error then also consult Sec
tion 6.8 after this procedure is followed. Normally the
-
-
41
Troubleshooting
10.4 The Displayed Process
Temperature Is In Error And
The Controller Remains In The
Cooling Or Heating State At
Any Set-point Value.
Possible causes of this problem may be either a faulty control probe or erroneous data in memory. The probe may be disconnected, burned out, or shorted. Check that the probe is connected properly. The probe may be checked with an ohmmeter to see if it is open or shorted. The probe is a platinum 4-wire DIN 43760 type therefore the resistance should read 0.2 to 2.0 ohms between pins 1 and 2 on the probe con­nector and between pins 3 and 4. The resistance should read 100 to 300 ohms between pins 1 and 4. If the probe is defective contact the factory.
If the problem is not the probe then it may be caused by erroneous data in memory. Re-initialize the mem­ory as discussed in Section 10.3 above. If the prob­lem remains then it may be caused by a defective electronic component. Contact the factory.
10.5 The Controller Controls Or
Attempts To Control At An
gram the constants. It may be that the memory backup battery is weak causing errors in data as de scribed in Section 10.3.
Check that the control probe has not been struck, bent, or damaged. If the causeof the problem remains unknown then contact the factory.
10.6 The Controller Shows That The
Output Power Is Steady But
The Process Temperature Is
Unstable.
If the bath temperature does not achieve the expected degree of stability when measured using a thermom­eter, try adjusting the proportional band to a narrower width as discussed in Section 6.7.
10.7 The Controller Alternately
Heats For A While Then Cools.
This is typically oscillation caused by the proportional band being too narrow. Increase the width of the pro­portional band until the temperature stabilizes as dis­cussed in Section 6.7.
-
Inaccurate Temperature.
Here the controller operates normally except when the controller has reached stable control of the bath temperature, the temperature does not agree with that measured with the user’s reference thermometer to within the specified accuracy. This problem may be caused by an actual difference in temperature be­tween the points where the control probe and ther­mometer probe measure temperature or by erroneous bath calibration parameters or a damaged control probe.
Check that the bath has an adequate amount of fluid in the tank and that the pump is operating causing fluid to circulate completely. Check that the thermom­eter probe and control probe are both fully inserted into the bath tominimize temperature gradient errors.
Check that the calibration parameters are all correct according to the certification sheet. If not then repro
42
10.8 The Controller Erratically Heats
Then Cools, Control Is
Unstable.
If both the bath temperature and output power do not vary periodically but in a very erratic manner, then the problem may be excess noise in the system. Noise due to the control sensor should be less than
0.001°C. However, if the probe has been damaged or has developed an intermittent short, then this may cause erratic behavior. Check for a damaged probe or poor connection between the probe and bath. Inter­mittent shorts in the heater or controller electronic cir­cuitry may also be a possible cause. Contact the factory if this seems to be the case.
-
Via Acquanera, 29 22100 Como tel. 031.526.566 (r.a.) fax 031.507.984
Loading...