MEDTESTER 5000C FIELD UPGRADE INSTALLATION INSTRUCTIONS
medTester 5000C Field
Upgrade Installation
Instructions
These instructions tell you how to perform an instrument
firmware upgrade to a medTester 5000C and how to install
optional modules into a Fluke Biomedical medTester 5000C.
Note
If you ordered more than one module and/or upgrade, you have
received more than one set of these instructions. But you can
perform all installations at the same time with one set of instructions.
Before You Begin
In these instructions, upgrade means that you are bringing up to date the firmware
in your medTester 5000. Upgrading involves replacing firmware integrated circuits
(ICs). Enabling modules means that you are installing software from diskettes to your
medTester in order to make medTester modules available for use.
In some cases you need to install additional ICs that work with specific
medTester 5000C modules. These are the modules for Expanded Memory, and
medCheck. You have received all of the parts necessary to perform the installation
yourself and to enable individual medTester 5000C modules.
If you have already had Fluke Biomedical upgrade your medTester or install
modules at the factory, you are receiving the module installation disks in the event
that you need to reinstall the modules later. Please save these instructions with
your installation disks. Your instrument is ready to use.
1
![](/html/13/1360/1360d82064191c9df45be22e48713e4c466e767546067df8f099ac36ff4d3b3a/bg4.png)
MEDTESTER 5000C FIELD UPGRADE INSTALLATION INSTRUCTIONS
NOTE: IMPORTANT
When upgrading your 5000C firmware, all memory for test
records and checklists will be erased. Process all current
medTester test records first. For example, upload them to
your compatible CMMS or print them.
Note
If you are installing a medTester 5000C firmware Upgrade; and
Modules 4 or 7; install your ICs before enabling the module(s).
Module Identification
Modules are identified by number according to the following table:
Module Number Module Name
2 RS-232/PRINTER
3 100 RECORDS
4 EXPANDED MEMORY
5 WAVES/EXTENDED TEST
6 DATA TRANSFER
7 MEDCHECK
8 DEFIBRILLATOR
9 IV PUMP
10 CMMS INTERFACE
11 ESU
12 SPO2
13 TRANSCUTANEOUS PACER
14 NIBP
2