Fluke Biomedical Impulse 7000DP, Impulse 6000D User Manual

© 2007 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies.
Impulse 6000D
Defibrillator Analyzer
Impulse 7000DP
Defibrillator/Transcutaneous Pacer Analyzer
Users Manual

Warranty and Product Support

Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of origi­nal purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSE­QUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provi sion.
7/07

Notices

All Rights Reserved
© Copyright 2007, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, st ored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and ot her printed ma terials for use in ser vice training programs and other technical publications. If you would like other reproducti ons or distrib utions, submit a written re quest to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrum ent. Notify the carrier and ask for an agent to be present while the instrum ent is unpacked. There ar e no special unpacking in structions, but be careful no t to dam­age the instrument when unpacking it. Inspect the instrument for phy sical damage such as bent or broken parts, dent s, or scratc hes.
Technical Support
For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800- 64 8-7942 or 1- 425-446- 6945.
Claims
Our routine method of shipment is via comm on carrier, FOB origin. Upo n delivery, if physical dam age is found, re tain all packing m aterials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifica­tions, or if there are any other problems not caused by ship ping dam age, please contact Fluke Biom edical or your local s ales representative.
Standard Terms and Conditions
Refunds and Credits
Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for re­turn or refund. Only products returned within 90 days from the date of ori ginal purchase are eligible for refund/credit. In order to receive a partial re­fund/credit of a product purchase price on a serialized produc t, the product must n ot have been dam aged by the customer or by the carrier chosen by the cus­tomer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and resalable con­dition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condit ion, are not eligible for credit return and will be returned to the customer. The Return Procedure (see below) must be followed to assure prompt refund/credit.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days af­ter purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or miss ing parts and accesso­ries will be applied to all returns.
Return Procedure
All items being returned (including all warranty- claim shipments) m ust be sent freight-pre paid to our factory locati on. When yo u return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition d ue to im proper packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
Use a double–walled carton of sufficient strength for the weight be ing shipped. Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts. Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at 1-800-648-7952 or 1-425-446-6945.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service In the U.S.A.:
Cleveland Calibration Lab Tel: 1-800-850-4606 Email: globalcal@flukebiomedical.com
Everett Calibration Lab Tel: 1-888-99 FLUKE (1-888-993-5853) Email: service.status@fluke.com
In Europe, Middle East, and Africa: Eindhoven Calibration Lab Tel: +31-402-675300 Email: ServiceDesk@fluke.com
In Asia: Everett Calibration Lab Tel: +425-446-6945 Email: service.international@fluke.com
or
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications w hen it was shipped fr om the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibra­tion standards are measured against in-house performance standards using accepted test procedure s.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Bio­medical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The Impulse 6000D and 7000DP Defibrillator/Transcutaneous Analyzers are manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.

Table of Contents

Title Page
Defibrillator Analyzer...............................................................................................................1
Introduction ....................................................................................................................1
Intended Use..................................................................................................................1
Unpacking the Analyzer .................................................................................................1
Safety Information..........................................................................................................2
Instrument Familiarization..............................................................................................4
Turning the Analyzer On and Off....................................................................................7
Accessing the Analyzer Tests ........................................................................................8
Analyzing Defibrillators...................................................................................................8
Testing Energy Levels...............................................................................................8
Testing Defibrillator Synchronization.........................................................................10
Testing Defibrillator Charge Time..............................................................................11
Analyzing Pacemakers (7000DP only)...........................................................................12
Setting Up the Analyzer for Pacer Testing.................................................................12
Performing a Pacer Asynchronous Test....................................................................14
i
7000DP, 6000D
Users Manual
Performing a Pacer Demand Test............................................................................. 15
Performing a Pacer Sensitivity Test..........................................................................16
Performing a Pacer Refractory Period Test ..............................................................17
Simulating ECG Signals.................................................................................................18
Connecting to the ECG Terminals.............................................................................19
Setting a Normal Sinus Rhythm ECG Signal ............................................................19
Setting a Performance ECG Signal........................................................................... 19
Setting Pacer Interactive ECG Waves (7000DP only)...............................................21
Selecting ECG Arrhythmias ......................................................................................22
Selecting TV Paced...................................................................................................23
Testing R Wave Detection ........................................................................................23
Performing a Noise Immunity Test............................................................................24
Setting Analyzer Setup Functions..................................................................................24
Setting Up the Battery...............................................................................................24
Setting Up the Display...............................................................................................25
Setting Up Sound......................................................................................................26
Displaying Instrument Information............................................................................. 26
Controlling the Analyzer Remotely.................................................................................26
Maintenance..................................................................................................................27
Cleaning the Analyzer............................................................................................... 27
Maintaining Peak Battery Condition..........................................................................28
Accessories ...................................................................................................................29
Specifications.................................................................................................................30
General Specifications..............................................................................................30
Defibrillator Analyzer Specifications.......................................................................... 31
Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only).......... 38
ii
Contents
Appendix A - Impulse 6000D/7000DP Remote Operation .....................................................41
Ansur Test Guide ...........................................................................................................41
Defibrillator Tests ...........................................................................................................43
Energy Measurement Test ........................................................................................43
Charge Time Test ......................................................................................................45
Sync Time Test..........................................................................................................45
Pacemaker Tests ...........................................................................................................46
Pacer Parameter Test ...............................................................................................46
Pacer Refractory Test................................................................................................46
Pacer Sensitivity Test................................................................................................47
Pacer Demand Mode Test.........................................................................................47
Asynchronous Mode Test..........................................................................................48
ECG Pacer Interactive Test.......................................................................................48
ECG Waveform Simulation Tests...................................................................................48
Normal Sinus Wave Simulation Test.........................................................................48
Arrhythmia Wave Test...............................................................................................49
Performance Wave Simulation..................................................................................49
ECG R-Wave Test .....................................................................................................49
ECG Noise Immunity Test.........................................................................................50
Battery Performance Tests.............................................................................................50
Battery Capacity Test................................................................................................50
Defib Pulse Repetition Test.......................................................................................51
Appendix B - Impulse 6000D/7000DP Test Templates..........................................................53
Introduction ....................................................................................................................53
Creating Test Templates................................................................................................53
Using Defibrillator Test Elements...................................................................................60
(continued)
iii
7000DP, 6000D
Users Manual
Energy Measurement Test........................................................................................60
Charge Time Test .....................................................................................................62
Synchronization Time Test........................................................................................63
Using Pacemaker Test Elements (Impulse 7000DP only) ............................................. 64
Pacer Parameter Test...............................................................................................64
Pacer Refractory Test...............................................................................................66
Pacer Sensitivity Test................................................................................................67
ECG Pacer Interactive Test ......................................................................................68
Pacer Demand Mode Test........................................................................................70
Asynchronous Mode Test .........................................................................................71
Using ECG Waveform Simulation Test Elements..........................................................71
Normal Sinus Wave Simulation.................................................................................71
Arrhythmia Wave Test...............................................................................................72
Performance Wave Simulation..................................................................................73
ECG R-Wave Test ....................................................................................................74
ECG Noise Immunity Test.........................................................................................75
Using Battery Performance Test Elements.................................................................... 75
Battery Capacity Test................................................................................................ 75
Defib Pulse Repetition Test.......................................................................................77
iv

List of Tables

Table Title Page
1. Symbols................................................................................................................................. 2
2. Top-Panel Controls and Connections.................................................................................... 5
3. Rear-Panel Connections ....................................................................................................... 7
4. Accessories........................................................................................................................... 29
5. Energy Measurement Test Measurements............................................................................ 61
6. Energy Measurement Test Custom Parameters ................................................................... 61
7. Charge Time Test Measurements......................................................................................... 62
8. Charge Time Test Custom Parameters................................................................................. 63
9. Synchronization Time Test Measurements ........................................................................... 63
10. Synchronization Time Test Custom Parameters................................................................... 64
11. Pacer Parameter Test Measurements................................................................................... 65
12. Pacer Parameter Test Custom Parameters........................................................................... 65
13. Pacer Refractory Test Measurements................................................................................... 66
14. Pacer Refractory Test Custom Parameters........................................................................... 67
15. Pacer Sensitivity Test Measurements ................................................................................... 67
16. Pacer Sensitivity Test Custom Parameters........................................................................... 68
17. ECG Pacer Interactive Test Custom Parameters.................................................................. 69
v
7000DP, 6000D
Users Manual
18. Pacer Demand Mode Test Custom Parameters ................................................................... 70
19. Normal Sinus Wave Simulation Test Custom Parameters.................................................... 71
20. Arrhythmia Wave Advisory Test Custom Parameters........................................................... 72
21. Performance Wave Simulation Test Custom Parameters..................................................... 73
22. ECG R-Wave Test Custom Parameters ............................................................................... 74
23. ECG Noise Immunity Test Custom Parameters.................................................................... 75
24. Battery Capacity Test Measurements................................................................................... 76
25. Battery Capacity Test Custom Parameters........................................................................... 76
26. Defib Pulse Repetition Test Measurements.......................................................................... 77
27. Defib Pulse Repetition Test Custom Parameters.................................................................. 77
vi

List of Figures

Figure Title Page
1. Top-Panel Controls and Connections.................................................................................... 4
2. Rear-Panel Connections ....................................................................................................... 6
3. Analyzer Ready Display ........................................................................................................ 7
4. Defibrillator Menu .................................................................................................................. 8
5. Cursor Navigation Example................................................................................................... 8
6. Defibrillator Test Connections ............................................................................................... 9
7. Defibrillator Energy Test........................................................................................................ 10
8. Defibrillator Synchronization Test.......................................................................................... 10
9. Defibrillator Charge Time Test............................................................................................... 11
10. Pacemaker Brand Selection.................................................................................................. 12
11. Connecting a Pacemaker to the Analyzer ............................................................................. 13
12. Displayed Pacer Parameters ................................................................................................. 14
13. Pacer Async Overdrive Mode................................................................................................ 15
14. Pacer Demand Overdrive Test.............................................................................................. 16
15. Pacer Sensitivity Test Display............................................................................................... 16
16. Paced Refractory Period (PRP)............................................................................................. 17
17. Sensed Refractory Period (SRP)........................................................................................... 18
vii
7000DP, 6000D
Users Manual
18. ECG Main Menu ................................................................................................................... 19
19. Normal Sinus Rhythm Rate Selection................................................................................... 19
20. ECG Connections................................................................................................................. 20
21. Performance Wave Selection................................................................................................ 21
22. Pacer Simulation Interactive Setup Screen........................................................................... 21
23. Ventricular Parameter Selection ........................................................................................... 22
24. TV Paced Selection .............................................................................................................. 23
25. AV Sequential Screen........................................................................................................... 23
26. R Wave Detection Screen..................................................................................................... 23
27. Pacer Noise Immunity Test................................................................................................... 24
28. Battery Setup Screen............................................................................................................ 25
29. Analyzer Information Screen................................................................................................. 26
30. Ansur Test Guide Window .................................................................................................... 42
31. Graph of Discharge Curve .................................................................................................... 44
32. Test Template with Selected Test Element........................................................................... 54
33. User-Definable Parts of the General Setup Tab.................................................................... 55
34. Expected Results Options for User Input.............................................................................. 56
35. Changing the Operand in Expected Results......................................................................... 57
36. Add or Delete Limits Pop-up Menu....................................................................................... 58
37. Custom Setup Page for Pacer Parameter Test Element....................................................... 59
viii

Defibrillator Analyzer

Introduction

The Impulse 6000D and 7000DP (hereafter the Analyzer) are portable, battery-powered precision instruments for testing external defibrillators. The 7000DP has the added capability of testing trancutaneous pacemakers. Where the additional pacemaker testing capability is applicable, this manual qualifies the description with “7000DP only.” The model 7000DP appears in all product illustrations.

Intended Use

The Analyzer is used to determine that defibrillators and transcutaneous pacemakers are performing within their performance specifications through measurement of energy output.

Unpacking the Analyzer

Carefully unpack all items from the box and check that you have the following items:
Impulse 6000D or 7000DP
Battery charger
Getting Started Manual
Users Manual CD
Defib paddle contact plates
Impulse 6000D 7000DP Ansur Software CD (demo)
1
Impulse 6000D, 7000DP
Users Manual
Table 1. Symbols
Symbol Description
W Important information; refer to manual.
Do not dispose of this product as
~
;
)
unsorted municipal waste. Go to Fluke’s website for recycling information.
Conforms to relevant Australian EMC requirements
Conforms to relevant Canadian and US standards
X Hazardous voltage P Conforms to European Union directives
IEC Measurement Category I – CAT I equipment designed to protect against transients in equipment on circuits not
CAT I
directly connected to circumstances should the terminals of the Analyzer be connected to any MAINS voltage.
MAINS. Under no

Safety Information

In this manual, a Warning identifies hazardous conditions and actions that could cause bodily harm or death. A Caution identifies conditions and actions that could damage the Analyzer, the equipment under test, or cause permanent loss of data.
XW Warning
To avoid possible electrical shock or personal injury, follow these guidelines:
Use this Analyzer only in the manner
specified by the manufacturer or the protection provided may be impaired.
Read the Users Manual before operating the
Analyzer.
Do not use the product if it operates
abnormally.
Do not use the product in wet locations,
around explosive gases or dust.
2
Defibrillator/Transcutaneous Pacemaker Analyzer Safety Information
Do not operate the Analyzer with the battery
eliminator connected, unless connected directly to mains power. During battery operation, completely remove the battery eliminator/charger from both the Analyzer and wall socket.
Do not connect the Analyzer to a patient or
equipment connected to a patient. The Analyzer is intended for equipment evaluation only and should never be used in diagnostics, treatment or in any other capacity where the Analyzer would come in contact with a patient.
Observe all precautions noted by the
Device Under Test (DUT) equipment manufacturer when analyzing the DUT.
Use extreme caution when working with
voltages above 30 volts.
Use the proper terminals, functions and
ranges for the test being performed.
3
Impulse 6000D, 7000DP
Users Manual

Instrument Familiarization

Figure 1 and Table 2 describes the top-panel controls and connections of the Analyzer.
1
11
10
2
3
4
4
9
8
275 V MAX
DEFIBRILLATOR
V
5000
p
MAX
7
5
5
6
fak07.eps
Figure 1. Top-Panel Controls and Connections
Defibrillator/Transcutaneous Pacemaker Analyzer Instrument Familiarization
Table 2. Top-Panel Controls and Connections
Item Name Description
1 ECG lead connectors
2 Backlight button Turns the LCD display backlight on and off. 3 Power button Turns the Analyzer on and off. 4 Navigation buttons Cursor control buttons for navigating menus and lists. 5 Defib connectors Defibrillator connections (Shown with removable defib paddle contact plates installed).
6 Function softkeys
7 Setup button Opens the setup menu. 8 ECG button Opens the main menu for ECG test functions.
9 Pacemaker inputs Input for low-level Pacer signal (7000DP only). 10 Pacer button Opens the main menu for pacer test functions (7000DP only). 11 Defibrillator button Opens the main menu for defibrillator test functions.
Outputs of low-level ECG signals (RA/R, LL/F, LA/L, RL/N, V1/C1, V2/C2, V3/C3, V4/C4, V5/C5, and V6/C6).
Keys F1 through F5 are used to select from a number of selections that appear in the LCD display above each function softkey.
5
Impulse 6000D, 7000DP
Users Manual
Figure 2 and Table 3 describes the rear-panel connections of the Analyzer.
6
CHARGE STATUS
21
SN
15VDC / 1.5 A
SERIAL NUMBER
FLUKE BIOMEDICAL
6920 SEAWAY BLVD
EVERETT, WA 98203
www.flukebiomedical.com
MADE IN USA
Figure 2. Rear-Panel Connections
HIGH LEVEL
SCOPE
ECG OUTPUT
OUTPUT
3 4 5
COMPUTER
PORT
fak08.eps
Defibrillator/Transcutaneous Pacemaker Analyzer Turning the Analyzer On and Off
Table 3. Rear-Panel Connections
Item Name Description
1 Charge Status LED
Battery Charger
2
connector 3 Scope output Output signal jack for displaying the defib playback wave on an oscilloscope. 4 Hi-level ECG output High-level ECG signal output jack for oscilloscope viewing. 5 Computer Port Device Port (B-style USB) for controlling the Analyzer from a PC or instrument controller.
Indicates RED while battery is charging. Indicates GREEN when the battery is fully charged and the charger is still connected.
Input connector for attaching the battery charger to the Analyzer.

Turning the Analyzer On and Off

Note
Before using the Analyzer for the first time, plug the battery charger into the Analyzer and a power outlet and charge the Analyzer for at least 4 hours.
Press the power button (O) on the top panel to turn the Analyzer on. After a short self-test period, the Analyzer will display the screen shown in Figure 3 to indicate it is ready for operation.
Battery condition is displayed in the upper right-hand corner of the display (S) when a top-level menu is displayed. When a low battery is indicated, attach the battery charger to the Analyzer and plug it into a power outlet.
Figure 3. Analyzer Ready Display
7
fak01.eps
Impulse 6000D, 7000DP
Users Manual

Accessing the Analyzer Tests

The Analyzer uses a series of menus to access various Analyzer functions and setup variables. As shown in Figure 4, the Analyzer indicates three different defibrillator tests (Energy, Sync, and Charge Time) along the bottom of the display. An Exit selection is also indicated as a way of backing out of the defibrillator tests. Pressing a softkey (F1 through F5) under a specific test will cause that test to be selected.
Figure 4. Defibrillator Menu
Some menu selections reveal a list of selectable items by displaying K to the right of the presently selected item. See Figure 5. To change the selection, press either G or H to scroll through the possible selections. Once the desired selection appears, press the function softkey and K disappears from the display.
fak02.eps
Figure 5. Cursor Navigation Example
fak03.eps

Analyzing Defibrillators

There are three main defibrillator test functions to evaluate a defibrillator’s performance: Energy, Synchronization, and Charge Time. To set the Analyzer for defibrillator testing, press M.
The Analyzer’s defibrillator input connectors are designed to be used with test leads or adapter plates when testing defibrillators using external paddles.
Connect the defibrillator to the Analyzer as shown in Figure 6.

Testing Energy Levels

Press the softkey labeled Energy to enter the energy test procedure. As shown in Figure 7, the Analyzer has a waveform selection already set. Either the waveform characteristic is off or it was the last one setup from a previous defibrillator test.
8
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Defibrillators
+
+
DEFIB
OFF
PACER
Defibrillator/Pacer
fak11.eps
Figure 6. Defibrillator Test Connections
9
Impulse 6000D, 7000DP
Users Manual
fak04.eps
Figure 7. Defibrillator Energy Test
If the waveform characteristics are correct, then charge the defibrillator using one of the energy settings, and with the defib paddles on the Analyzer’s input, press the discharge button. The Analyzer senses the discharge and the energy delivered appears in the display in Joules.
Changing Waveform Characteristics
If the waveform characteristics are not the desired ones, press the softkey labeled Set Wave. The waveform, its amplitude, and frequency are new softkey selections. Press the softkey under the signal attribute you want to change. Use G or H to scroll through all the values. Once the desired value is set, press the softkey under the adjusted characteristic. This same process applies to Amplitude and Rate selections as well. With the three parameters set, press the softkey labeled Done to return to the discharge ready state.
The softkey labeled Summary provides additional information about the current discharge waveform
depending on the defibrillator type tested. For dc Monophasic: peak voltage, peak current and pulse width. For dc bi-phasic: peak and average voltage, peak and average current, pulse width, interphase delay, and overall tilt. For ac bi-phasic: all dc bi-phasic data and ac carrier base frequency and duty cycle.
Note
AC Pulsed Bi-Phasic waveform has not been approved in the United States.

Testing Defibrillator Synchronization

From the Defibrillator main menu, press the softkey labeled Sync. As shown in Figure 8, the waveform selection is already set.
Figure 8. Defibrillator Synchronization Test
The test measures the response of the defibrillator in its synchronous (sync) mode. Place the defibrillator in this mode. The defibrillator will now synchronize its discharge pulse with the ECG heart rate. The sync time measured is
fak05.eps
10
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Defibrillators
the time from the ECG heart beat ‘R’ wave to the onset of the defibrillator pulse.
If the waveform characteristics are not correct, then change the characteristics as explained in the “Changing Waveform Characteristics” section earlier in this manual.
With the waveform set to the desired characteristics, charge the defibrillator and discharge it into the Analyzer’s defibrillator inputs. The Analyzer senses the discharge and the measured delay appears in the display.
The Analyzer can automatically identify the correct defibrillator waveforms delivered by the defibrillator under test. The softkey labeled Summary provides information about the current discharge waveform depending on the defibrillator type tested. For dc monophasic: peak voltage, peak current and pulse width. For dc bi-phasic: peak and average voltage, peak and average current, pulse width, interphase delay, and overall tilt. For ac bi-phasic: all dc bi-phasic data and ac carrier base frequency and duty cycle.

Testing Defibrillator Charge Time

Before starting the charge time test, ensure the defibrillator is not charged. This test measures the amount of time it takes the defibrillator to go from a full discharge to charge at the desired charge level (typically maximum setting) and then discharge into the Analyzer’s test load.
From the Defibrillator main menu, press the softkey labeled Charge Time. As shown in Figure 9, the waveform selection is already set and Measure Charge Time… is displayed.
fak06.eps
Figure 9. Defibrillator Charge Time Test
In a few seconds after pressing the softkey labeled Measure, a Charge Defib in: countdown begins. When the countdown reaches zero and sounds the beeper, press the charge button on the defibrillator. The Analyzer begins a Charge Time count up. When the defibrillator reaches full charge, discharge the defibrillator into the Analyzer.
Note
For this test the Analyzer is timing operator actions. The measurement depends on the user accurately starting the defibrillator as soon as it is charged. Any operator time delay is included in the measurement result. The user should repeat any tests that have not been timed accurately.
11
Impulse 6000D, 7000DP
Users Manual
The Analyzer senses the discharge and the charge time appears in the display. Press the softkey labeled Measure to perform another charge time test.
The softkey labeled Summary provides additional information about the current discharge waveform depending on the defibrillator type tested. For dc monophasic: peak voltage, peak current and pulse width. For dc bi-phasic: peak and average voltage, peak and average current, pulse width, interphase delay, and overall tilt. For ac bi-phasic: all dc bi-phasic data and ac carrier base frequency and duty cycle.

Analyzing Pacemakers (7000DP only)

The Analyzer is designed to work with a variety of pacemaker brands. See the specifications section later in this manual for a list of pacemaker brands. The Analyzer measures and displays pacemaker pulse amplitude, rate, and width. It also performs demand sensitivity tests, measures and displays refractory periods, and tests the pacemaker’s susceptibility to 50/60 Hz interference.

Setting Up the Analyzer for Pacer Testing

W Caution
To avoid damage to the Analyzer or defibrillator, do not apply defibrillator pulses to the pacer inputs.
Connect the pacemaker to be tested to the Analyzer through either the pacer input jacks or defibrillator jacks as shown in Figure 11.
For tests where the pacemaker interacts with the simulated heart beat (Async, Demand, Sensitivity, and Refractory Period tests), the pacemaker senses the heart beat on its ECG leads. Connect the ECG leads to the Analyzer ECG posts as shown in Figure 20.
In preparation for testing a pacemaker, the Analyzer will have to be set to the specific brand of the pacemaker under test. Press N to enter the top-level pacer menu shown in Figure 10.
Figure 10. Pacemaker Brand Selection
fak12.eps
12
Defibrillator/Transcutaneous Pacemaker Analyzer Analyzing Pacemakers (7000DP only)
Defibrillator/Pacer
DEFIB
OFF
PACER
Caution To avoid damage to the
Analyzer or defibrillator, do not apply defibrillator pulses to the pacer inputs.
7000DP Only
50 - 1500 Ω
+
Figure 11. Connecting a Pacemaker to the Analyzer
50 Ω
Only
+
fak10.eps
13
Impulse 6000D, 7000DP
Users Manual
Press the softkey labeled Brand to activate the brand list and scroll through the list using G or H. When the correct brand is displayed, enter the selection in one of three ways. Press the softkey labeled Brand, press one of the other two setup function softkeys (Load or Input Jacks) or press the softkey labeled Done.
The load the pacer is working into through the Analyzer’s pacer inputs is set through the Load softkey. If the load value needs to be changed, press the softkey labeled Load and then use G or H to select a value between 50 and 1500 Ω in 50 Ω steps. Set the load value by pressing the Load softkey again, press one of the other two pacer variable softkeys, or press the softkey labeled Done.
Note
The load value is only selectable when the pacer
Ω
input selection is set to input jacks. Only a 50 load is available when the input selection is set to Defib.
The third pacer variable is the selection of the jacks where the pacemaker has been attached to the Analyzer. The input jacks softkey toggles between two settings: Input Jacks or Defib. The Input Jacks selection monitors the jack just to the left of the function and setup buttons. When Defib is selected, the Analyzer monitors the pacemaker through the defibrillator jacks.
When all three pacer setup variables are set to their desired values, press the softkey labeled Done. The Analyzer begins to monitor the pacer signal through the selected input jacks. When the pacer signal is detected the display indicates the pacemakers pulse rate, pulse width, energy, and amplitude. In addition, pacer test function labels appear above the softkeys indicating the Analyzer is ready to perform one of the pacer tests. See Figure 12.
fak13.eps
Figure 12. Displayed Pacer Parameters

Performing a Pacer Asynchronous Test

This qualitative test verifies the continuous (or non­demand) mode pacemaker’s ability to interact with a simulated ECG signal. The Analyzer first measures the pacemaker’s applied pulse rate then computes “underdrive” and “overdrive” rates for the simulated ECG signal. Initially, the “underdrive” rate is 85 % of the applied pacemaker rate and the “overdrive” rate is 115 % of the applied pacemaker rate.
14
Loading...
+ 64 hidden pages