Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for
calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our
option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to
Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if
the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized
Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE
EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration
of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since
some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages,
this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or
other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provi sion.
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and ot her printed ma terials for use in ser vice training programs
and other technical publications. If you would like other reproducti ons or distrib utions, submit a written re quest to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrum ent.
Notify the carrier and ask for an agent to be present while the instrum ent is unpacked. There ar e no special unpacking in structions, but be careful no t to damage the instrument when unpacking it. Inspect the instrument for phy sical damage such as bent or broken parts, dent s, or scratc hes.
Technical Support
For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800- 64 8-7942 or 1- 425-446- 6945.
Claims
Our routine method of shipment is via comm on carrier, FOB origin. Upo n delivery, if physical dam age is found, re tain all packing m aterials in their original
condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by ship ping dam age, please contact Fluke Biom edical or your local s ales representative.
Standard Terms and Conditions
Refunds and Credits
Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for
partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for return or refund. Only products returned within 90 days from the date of ori ginal purchase are eligible for refund/credit. In order to receive a partial refund/credit of a product purchase price on a serialized produc t, the product must n ot have been dam aged by the customer or by the carrier chosen by the customer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and resalable condition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condit ion, are not eligible for credit return and
will be returned to the customer. The Return Procedure (see below) must be followed to assure prompt refund/credit.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or miss ing parts and accessories will be applied to all returns.
Return Procedure
All items being returned (including all warranty- claim shipments) m ust be sent freight-pre paid to our factory locati on. When yo u return an instrument to
Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its
actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition d ue to im proper
packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
Use a double–walled carton of sufficient strength for the weight be ing shipped.
Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts.
Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at
1-800-648-7952 or 1-425-446-6945.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service
In the U.S.A.:
In Europe, Middle East, and Africa:
Eindhoven Calibration Lab
Tel: +31-402-675300
Email: ServiceDesk@fluke.com
In Asia:
Everett Calibration Lab
Tel: +425-446-6945
Email: service.international@fluke.com
or
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications w hen it was shipped fr om the
factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedure s.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in
this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability
of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The Impulse 6000D and 7000DP Defibrillator/Transcutaneous Analyzers are manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA,
U.S.A.
29. Analyzer Information Screen................................................................................................. 26
30. Ansur Test Guide Window .................................................................................................... 42
31. Graph of Discharge Curve .................................................................................................... 44
32. Test Template with Selected Test Element........................................................................... 54
33. User-Definable Parts of the General Setup Tab.................................................................... 55
34. Expected Results Options for User Input.............................................................................. 56
35. Changing the Operand in Expected Results......................................................................... 57
36. Add or Delete Limits Pop-up Menu....................................................................................... 58
37. Custom Setup Page for Pacer Parameter Test Element....................................................... 59
viii
Defibrillator Analyzer
Introduction
The Impulse 6000D and 7000DP (hereafter the Analyzer)
are portable, battery-powered precision instruments for
testing external defibrillators. The 7000DP has the added
capability of testing trancutaneous pacemakers. Where
the additional pacemaker testing capability is applicable,
this manual qualifies the description with “7000DP only.”
The model 7000DP appears in all product illustrations.
Intended Use
The Analyzer is used to determine that defibrillators and
transcutaneous pacemakers are performing within their
performance specifications through measurement of
energy output.
Unpacking the Analyzer
Carefully unpack all items from the box and check that
you have the following items:
• Impulse 6000D or 7000DP
• Battery charger
• Getting Started Manual
• Users Manual CD
• Defib paddle contact plates
• Impulse 6000D 7000DP Ansur Software CD (demo)
1
Impulse 6000D, 7000DP
Users Manual
Table 1. Symbols
Symbol Description
WImportant information; refer to manual.
Do not dispose of this product as
~
;
)
unsorted municipal waste. Go to Fluke’s
website for recycling information.
Conforms to relevant Australian EMC
requirements
Conforms to relevant Canadian and US
standards
X Hazardous voltage
P Conforms to European Union directives
IEC Measurement Category I – CAT I
equipment designed to protect against
transients in equipment on circuits not
CAT I
directly connected to
circumstances should the terminals of
the Analyzer be connected to any
MAINS voltage.
MAINS. Under no
Safety Information
In this manual, a Warning identifies hazardous conditions
and actions that could cause bodily harm or death. A
Caution identifies conditions and actions that could
damage the Analyzer, the equipment under test, or cause
permanent loss of data.
XW Warning
To avoid possible electrical shock or personal
injury, follow these guidelines:
• Use this Analyzer only in the manner
specified by the manufacturer or the
protection provided may be impaired.
• Read the Users Manual before operating the
Analyzer.
• Do not use the product if it operates
abnormally.
• Do not use the product in wet locations,
around explosive gases or dust.
2
Defibrillator/Transcutaneous Pacemaker AnalyzerSafety Information
• Do not operate the Analyzer with the battery
eliminator connected, unless connected
directly to mains power. During battery
operation, completely remove the battery
eliminator/charger from both the Analyzer
and wall socket.
• Do not connect the Analyzer to a patient or
equipment connected to a patient. The
Analyzer is intended for equipment
evaluation only and should never be used
in diagnostics, treatment or in any other
capacity where the Analyzer would come in
contact with a patient.
• Observe all precautions noted by the
Device Under Test (DUT) equipment
manufacturer when analyzing the DUT.
• Use extreme caution when working with
voltages above 30 volts.
• Use the proper terminals, functions and
ranges for the test being performed.
3
Impulse 6000D, 7000DP
Users Manual
Instrument Familiarization
Figure 1 and Table 2 describes the top-panel controls and connections of the Analyzer.
2 Backlight button Turns the LCD display backlight on and off.
3 Power button Turns the Analyzer on and off.
4 Navigation buttons Cursor control buttons for navigating menus and lists.
5 Defib connectors Defibrillator connections (Shown with removable defib paddle contact plates installed).
6 Function softkeys
7 Setup button Opens the setup menu.
8 ECG button Opens the main menu for ECG test functions.
9 Pacemaker inputs Input for low-level Pacer signal (7000DP only).
10 Pacer button Opens the main menu for pacer test functions (7000DP only).
11 Defibrillator button Opens the main menu for defibrillator test functions.
Outputs of low-level ECG signals (RA/R, LL/F, LA/L, RL/N, V1/C1, V2/C2, V3/C3, V4/C4,
V5/C5, and V6/C6).
Keys F1 through F5 are used to select from a number of selections that appear in the LCD
display above each function softkey.
5
Impulse 6000D, 7000DP
Users Manual
Figure 2 and Table 3 describes the rear-panel connections of the Analyzer.
6
CHARGE STATUS
21
SN
15VDC / 1.5 A
SERIAL NUMBER
FLUKE BIOMEDICAL
6920 SEAWAY BLVD
EVERETT, WA 98203
www.flukebiomedical.com
MADE IN USA
Figure 2. Rear-Panel Connections
HIGH LEVEL
SCOPE
ECG OUTPUT
OUTPUT
345
COMPUTER
PORT
fak08.eps
Defibrillator/Transcutaneous Pacemaker AnalyzerTurning the Analyzer On and Off
Table 3. Rear-Panel Connections
Item Name Description
1 Charge Status LED
Battery Charger
2
connector
3 Scope output Output signal jack for displaying the defib playback wave on an oscilloscope.
4 Hi-level ECG output High-level ECG signal output jack for oscilloscope viewing.
5 Computer Port Device Port (B-style USB) for controlling the Analyzer from a PC or instrument controller.
Indicates RED while battery is charging. Indicates GREEN when the battery is fully
charged and the charger is still connected.
Input connector for attaching the battery charger to the Analyzer.
Turning the Analyzer On and Off
Note
Before using the Analyzer for the first time, plug
the battery charger into the Analyzer and a power
outlet and charge the Analyzer for at least 4
hours.
Press the power button (O) on the top panel to turn the
Analyzer on. After a short self-test period, the Analyzer will
display the screen shown in Figure 3 to indicate it is ready
for operation.
Battery condition is displayed in the upper right-hand
corner of the display (S) when a top-level menu is
displayed. When a low battery is indicated, attach the
battery charger to the Analyzer and plug it into a power
outlet.
Figure 3. Analyzer Ready Display
7
fak01.eps
Impulse 6000D, 7000DP
Users Manual
Accessing the Analyzer Tests
The Analyzer uses a series of menus to access various
Analyzer functions and setup variables. As shown in
Figure 4, the Analyzer indicates three different defibrillator
tests (Energy, Sync, and Charge Time) along the bottom
of the display. An Exit selection is also indicated as a way
of backing out of the defibrillator tests. Pressing a softkey
(F1 through F5) under a specific test will cause that test to
be selected.
Figure 4. Defibrillator Menu
Some menu selections reveal a list of selectable items by
displaying K to the right of the presently selected item.
See Figure 5. To change the selection, press either G or
H to scroll through the possible selections. Once the
desired selection appears, press the function softkey and
K disappears from the display.
fak02.eps
Figure 5. Cursor Navigation Example
fak03.eps
Analyzing Defibrillators
There are three main defibrillator test functions to evaluate
a defibrillator’s performance: Energy, Synchronization, and
Charge Time. To set the Analyzer for defibrillator testing,
press M.
The Analyzer’s defibrillator input connectors are designed
to be used with test leads or adapter plates when testing
defibrillators using external paddles.
Connect the defibrillator to the Analyzer as shown in
Figure 6.
Testing Energy Levels
Press the softkey labeled Energy to enter the energy test
procedure. As shown in Figure 7, the Analyzer has a
waveform selection already set. Either the waveform
characteristic is off or it was the last one setup from a
previous defibrillator test.
If the waveform characteristics are correct, then charge
the defibrillator using one of the energy settings, and with
the defib paddles on the Analyzer’s input, press the
discharge button. The Analyzer senses the discharge and
the energy delivered appears in the display in Joules.
Changing Waveform Characteristics
If the waveform characteristics are not the desired ones,
press the softkey labeled Set Wave. The waveform, its
amplitude, and frequency are new softkey selections.
Press the softkey under the signal attribute you want to
change. Use G or H to scroll through all the values. Once
the desired value is set, press the softkey under the
adjusted characteristic. This same process applies to
Amplitude and Rate selections as well. With the three
parameters set, press the softkey labeled Done to return
to the discharge ready state.
The softkey labeled Summary provides additional
information about the current discharge waveform
depending on the defibrillator type tested. For dc
Monophasic: peak voltage, peak current and pulse width.
For dc bi-phasic: peak and average voltage, peak and
average current, pulse width, interphase delay, and overall
tilt. For ac bi-phasic: all dc bi-phasic data and ac carrier
base frequency and duty cycle.
Note
AC Pulsed Bi-Phasic waveform has not been
approved in the United States.
Testing Defibrillator Synchronization
From the Defibrillator main menu, press the softkey
labeled Sync. As shown in Figure 8, the waveform
selection is already set.
Figure 8. Defibrillator Synchronization Test
The test measures the response of the defibrillator in its
synchronous (sync) mode. Place the defibrillator in this
mode. The defibrillator will now synchronize its discharge
pulse with the ECG heart rate. The sync time measured is
the time from the ECG heart beat ‘R’ wave to the onset of
the defibrillator pulse.
If the waveform characteristics are not correct, then
change the characteristics as explained in the “Changing
Waveform Characteristics” section earlier in this manual.
With the waveform set to the desired characteristics,
charge the defibrillator and discharge it into the Analyzer’s
defibrillator inputs. The Analyzer senses the discharge and
the measured delay appears in the display.
The Analyzer can automatically identify the correct
defibrillator waveforms delivered by the defibrillator under
test. The softkey labeled Summary provides information
about the current discharge waveform depending on the
defibrillator type tested. For dc monophasic: peak voltage,
peak current and pulse width. For dc bi-phasic: peak and
average voltage, peak and average current, pulse width,
interphase delay, and overall tilt. For ac bi-phasic: all dc
bi-phasic data and ac carrier base frequency and duty
cycle.
Testing Defibrillator Charge Time
Before starting the charge time test, ensure the
defibrillator is not charged. This test measures the amount
of time it takes the defibrillator to go from a full discharge
to charge at the desired charge level (typically maximum
setting) and then discharge into the Analyzer’s test load.
From the Defibrillator main menu, press the softkey
labeled Charge Time. As shown in Figure 9, the
waveform selection is already set and Measure Charge Time… is displayed.
fak06.eps
Figure 9. Defibrillator Charge Time Test
In a few seconds after pressing the softkey labeled
Measure, a Charge Defib in: countdown begins. When
the countdown reaches zero and sounds the beeper,
press the charge button on the defibrillator. The Analyzer
begins a Charge Time count up. When the defibrillator
reaches full charge, discharge the defibrillator into the
Analyzer.
Note
For this test the Analyzer is timing operator
actions. The measurement depends on the user
accurately starting the defibrillator as soon as it is
charged. Any operator time delay is included in
the measurement result. The user should repeat
any tests that have not been timed accurately.
11
Impulse 6000D, 7000DP
Users Manual
The Analyzer senses the discharge and the charge time
appears in the display. Press the softkey labeled Measure
to perform another charge time test.
The softkey labeled Summary provides additional
information about the current discharge waveform
depending on the defibrillator type tested. For dc
monophasic: peak voltage, peak current and pulse width.
For dc bi-phasic: peak and average voltage, peak and
average current, pulse width, interphase delay, and overall
tilt. For ac bi-phasic: all dc bi-phasic data and ac carrier
base frequency and duty cycle.
Analyzing Pacemakers (7000DP only)
The Analyzer is designed to work with a variety of
pacemaker brands. See the specifications section later in
this manual for a list of pacemaker brands. The Analyzer
measures and displays pacemaker pulse amplitude, rate,
and width. It also performs demand sensitivity tests,
measures and displays refractory periods, and tests the
pacemaker’s susceptibility to 50/60 Hz interference.
Setting Up the Analyzer for Pacer Testing
W Caution
To avoid damage to the Analyzer or
defibrillator, do not apply defibrillator pulses
to the pacer inputs.
Connect the pacemaker to be tested to the Analyzer
through either the pacer input jacks or defibrillator jacks as
shown in Figure 11.
For tests where the pacemaker interacts with the
simulated heart beat (Async, Demand, Sensitivity, and
Refractory Period tests), the pacemaker senses the heart
beat on its ECG leads. Connect the ECG leads to the
Analyzer ECG posts as shown in Figure 20.
In preparation for testing a pacemaker, the Analyzer will
have to be set to the specific brand of the pacemaker
under test. Press N to enter the top-level pacer menu
shown in Figure 10.
Analyzer or defibrillator, do
not apply defibrillator pulses
to the pacer inputs.
7000DP Only
50 - 1500 Ω
+
Figure 11. Connecting a Pacemaker to the Analyzer
50 Ω
Only
+
fak10.eps
13
Impulse 6000D, 7000DP
Users Manual
Press the softkey labeled Brand to activate the brand list
and scroll through the list using G or H. When the correct
brand is displayed, enter the selection in one of three
ways. Press the softkey labeled Brand, press one of the
other two setup function softkeys (Load or Input Jacks)
or press the softkey labeled Done.
The load the pacer is working into through the Analyzer’s
pacer inputs is set through the Load softkey. If the load
value needs to be changed, press the softkey labeled
Load and then use G or H to select a value between 50
and 1500 Ω in 50 Ω steps. Set the load value by pressing
the Load softkey again, press one of the other two pacer
variable softkeys, or press the softkey labeled Done.
Note
The load value is only selectable when the pacer
Ω
input selection is set to input jacks. Only a 50
load is available when the input selection is set to
Defib.
The third pacer variable is the selection of the jacks where
the pacemaker has been attached to the Analyzer. The
input jacks softkey toggles between two settings: Input Jacks or Defib. The Input Jacks selection monitors the
jack just to the left of the function and setup buttons. When
Defib is selected, the Analyzer monitors the pacemaker
through the defibrillator jacks.
When all three pacer setup variables are set to their
desired values, press the softkey labeled Done. The
Analyzer begins to monitor the pacer signal through the
selected input jacks. When the pacer signal is detected
the display indicates the pacemakers pulse rate, pulse
width, energy, and amplitude. In addition, pacer test
function labels appear above the softkeys indicating the
Analyzer is ready to perform one of the pacer tests. See
Figure 12.
fak13.eps
Figure 12. Displayed Pacer Parameters
Performing a Pacer Asynchronous Test
This qualitative test verifies the continuous (or nondemand) mode pacemaker’s ability to interact with a
simulated ECG signal. The Analyzer first measures the
pacemaker’s applied pulse rate then computes
“underdrive” and “overdrive” rates for the simulated ECG
signal. Initially, the “underdrive” rate is 85 % of the applied
pacemaker rate and the “overdrive” rate is 115 % of the
applied pacemaker rate.
14
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.