Fluke Fluke 572-2 Manual

Page 1
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
572-2
Infrared Thermometer

Users Manual

March 2013, Rev. 1, 08/13
©2013 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
Page 2
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is two years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
11/99
Page 3

Table of Contents

Title Page
Introduction ........................................................................................................ 1
Contact Fluke ..................................................................................................... 1
Safety Information ............................................................................................. 2
Features .............................................................................................................. 4
Display ............................................................................................................... 5
Menu Overview ............................................................................................. 5
Save ............................................................................................................... 6
Light .............................................................................................................. 6
Memory ......................................................................................................... 6
Emissivity Menu ............................................................................................ 6
°C and °F ....................................................................................................... 7
Min, Max, Avg, Differential .......................................................................... 8
Alarm ............................................................................................................. 8
Trigger Lock .................................................................................................. 8
Laser .............................................................................................................. 9
Setup .............................................................................................................. 9
Language ................................................................................................... 9
Backlight ................................................................................................... 9
Time/Date .................................................................................................. 9
Delete Data .................................................................................................... 10
Delete All Data .......................................................................................... 10
Delete Individual Data Records ................................................................ 10
How the Product Works ..................................................................................... 11
Product Operation .............................................................................................. 11
Temperature Measurement ............................................................................ 11
Locate a Hot or Cold Spot ............................................................................. 12
Distance and Spot Size .................................................................................. 12
Field of View ................................................................................................. 13
Emissivity ...................................................................................................... 13
HOLD ............................................................................................................ 13
Data Storage .................................................................................................. 14
Data Download .............................................................................................. 14
External Contact Probe ...................................................................................... 15
Troubleshooting ................................................................................................. 15
Maintenance ....................................................................................................... 16
Battery Charge ............................................................................................... 16
i
Page 4
572-2
Users Manual
Battery Replacement...................................................................................... 16
Clean the Lens ............................................................................................... 17
Clean the Case ............................................................................................... 17
User-Replaceable Parts and Accessories ........................................................... 17
User-Replaceable Parts .................................................................................. 17
Accessories .................................................................................................... 18
Recommended Temperature Probes .............................................................. 18
Specifications ..................................................................................................... 19
K-Type Thermocouple Bead Probe Specifications ....................................... 20
ii
Page 5

List of Tables

Table Title Page
1. Symbols .................................................................................................................. 3
2. Top-Level Menu Description ................................................................................. 5
3.
Nominal Surface Emissivity ................................................................................... 7
4. Troubleshooting ..................................................................................................... 15
5. Replaceable Parts ................................................................................................... 17
6. Recommended Temperature Probes ....................................................................... 18
iii
Page 6
572-2
Users Manual
iv
Page 7

List of Figures

Figure Title Page
1. Safety Markings ..................................................................................................... 3
Menu Navigation
2.
3.
How the Product Works ......................................................................................... 11
4. Locating a Hot or Cold Spot .................................................................................. 12
5. Distance and Spot Size ........................................................................................... 12
6. Field of View .......................................................................................................... 13
7. Thermocouple and USB Connections .................................................................... 14
8. Battery Replacement .............................................................................................. 17
.................................................................................................... 5
v
Page 8
572-2
Users Manual
vi
Page 9

Introduction

The 572-2 Infrared Thermometer (the Product) is for non-contact temperature measurement. The Product determines the surface temperature of an object by measuring the amount of infrared energy radiated by the object surface. The Product also uses a K-type thermocouple for contact temperature measurement.
1
Page 10
572-2
Users Manual

Safety Information

A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product
or the equipment under test.
Symbols used on the Product and in this manual are explained in Table 1. Laser safety markings are shown in Figure 1.
Warning
To prevent possible electrical shock, fire, eye damage, and personal injury:
Read all safety Information before you use the Product.
Use the Product only as specified, or the protection
supplied by the Product can be compromised.
Do not use the Product around explosive gas, vapor, or
in damp or wet environments.
Do not use the Product if it operates incorrectly.
See emissivity information for actual temperatures.
Reflective objects result in lower than actual temperature measurements. These objects pose a burn hazard.
Do not look directl
example, binoculars, telescopes, microscopes). Optical tools can focus the laser and be dangerous to the eye.
Do not look into the laser. Do not point laser directly
persons or animals or indirectly off reflective surfaces.
Use the product only as specified, or hazardous laser
radiation exposure can occur.
Do not use laser viewing glasses as laser protection
glasses. Laser viewing glasses are used only for better visibility
Do not open the Product. The laser beam is dangerous
to eyes. Have the Product repaired only through an approved technical site.
Have an approved technician repair the product.
For safe operation and maintenance of the Product:
Remove the batteries if the Product is not used for an
extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
Replace the batteries when the low battery indicator
shows to prevent incorrect measurements.
Have the Product repaired before use if the batteries
leak.
Be sure that the battery polarity is correct to prevent
battery leakage.
Do not short the battery terminals together.
the laser in bright light.
of
y into the laser with optical
Caution
tools (for
at
2
Page 11
Infrared Thermometer
Do not keep cells or batteries in a container where the
terminals can be shorted.
Do not put battery cells and battery packs near heat or
fire. Do not put in sunlight.
Caution
To avoid damage to the Product or the equipment under test, protect them from:
EMF (electro-magnetic fields) from arc welders,
induction heaters, etc.
Static electricity
Thermal shock (caused by large or abrupt ambient
temperature changes. For highest accuracy, allow 30 minutes for Thermometer to become stable before use).
Table 1. Symbols
Symbol Explanation
Safety Information
Hazardous voltage. Risk of electrical shock.
Risk of danger. Important information. See manual.
Warning. Laser.
Conforms to European Union directives.
°C Celsius
°F Fahrenheit
Battery
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product
in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
Battery
Conforms to relevant South Korean EMC Standards.
Conforms to China Metrology Certification.
3
Page 12
572-2
Users Manual
Laser
Aperture

Features

Dual laser sighting
Backlit display
Figure 1. Safety Markings
hdp01.eps
Current Temperature plus MAX, MIN, DIF, AVG temperature displays
Two AA batteries
Hard case
80PK-1 K-type thermocouple probe
Adjustable emissivity and predefined emissivity table
Infrared and thermocouple temperature display
Celsius or Fahrenheit temperature display
Tripod mount
Auto off
Standard miniature K-type thermocouple connector input
Printed 572-2 Getting Started
572-2 Manuals CD
12 or 24 hour clock
Last reading Hold (20 seconds)
Multi-language interface
High and low temperature alarm
Data storage and review
4
Trigger lock
USB 2.0 computer interface cable
FlukeView
®
Forms Documenting Software
Page 13
Infrared Thermometer

Display

Display
The Product display can show data in these languages:
English
Spanish
French
German
Portuguese
Simplified Chinese
To change the displayed language, refer to “Setup”.

Menu Overview

There are many settings that can be changed with the menu. Figure 2 shows the LCD and menu interface. Selecting the Menu softkey advances the menu to the next level. Table 2 is a top-level description of the menu.
Figure 2. Menu Navigation
Table 2. Top-Level Menu Description
Level
1 Save Save reading to memory Menu Light
2 Mem
3 MnMx Enables Min/Max Menu Avg Enable Avg/Diff
4 °C/°F Toggle between C and F Menu Alarm Set and enable alarms
5
6 Setup
Left
Softkey
Review/delete memory entries
Lock the Product on Menu Laser Toggle the laser on/off
(Lock)
- Turn off/on backlight
- Change Time/Date
- Change Language
Description
Center
Softkey
Menu
Menu - -
Right
Softkey
Description
Adjust backlight brightness
Set emissivity
hdp02.eps
Each menu item and function is explained in greater detail in the following sections.
5
Page 14
572-2
Users Manual

Save

To save readings:
1. Pull the trigger to take a measurement.
2. Release the trigger to stop taking the measurement.
3. Push the Save softkey to enter the Save menu.
4. Push the Yes softkey to save the reading.
The reading is assigned a memory location and a time and date stamp. The reading includes:
IR temperature
Emissivity
Date/Time
You can also push the Cancel softkey to abort saving the reading.

Light

The Product has a backlit display with two brightness levels. The Light softkey is used to adjust the backlight brightness. The backlight is on each time the trigger is pulled.
To toggle the backlight brightness, push the Light softkey. The backlight can be disabled using the Setup menu. See “Setup” for more
information.

Memory

The Product can store measurement records including time, date, emissivity, and measurement record numbers (see “Save” for more information). The Product can store 99 records.
To access records stored in memory:
1. Push the Menu softkey until Mem shows as the left softkey function.
2. Push the Mem softkey to access the Memory menu. Saved readings can be read.

Emissivity Menu

The Emissivity menu includes a list of pre-defined materials and lists their typical emissivity values, see Table 3. For further information, see “Emissivity”.
Note
Default emissivity is 0.95.
To access the Emissivity menu:
1. Push the Menu softkey until  (emissivity) shows as the right softkey function.
2. Push the  softkey.
6
Page 15
Infrared Thermometer
Display
You can push the Table softkey to access the Emissivity list. You can also push the No. softkey to manually enter the typical emissivity of a material.
If the Emissivity table is accessed, a list of materials and their suggested
emissivity is shown.
1. Use the down arrow to navigate through the list.
2. Push the Enter softkey to choose the desired material.
To enter an emissivity value manually:
1. Push the No. softkey.
2. Use the down or up arrow softkey to change the entry. Hold down the up or down arrow softkey to increase the rate of change.
3. Push the Done softkey when finished to return to the main menu.
Table 3. Nominal Surface Emissivity
Material Value Material Value
Default**** 0.95 Glass (plate) 0.85
Aluminum* 0.30 Iron* 0.70
Asbestos 0.95 Lead* 0.50
Asphalt 0.95 Oil 0.94
Brass* 0.50 Paint 0.93
Ceramic 0.95 Plastic** 0.95
Concrete 0.95 Rubber 0.95
Copper* 0.60 Sand 0.90
Food - Frozen 0.90 Steel* 0.80
Food - Hot 0.93 Water 0.93
Wood *** 0.94
* Oxidized ** Opaque, over 20 mils *** Natural **** Factory Setting Highlighted items may also be found in the emissivity table built into the Product.
°
C and °F
To toggle between °C (Celsius) and °F (Fahrenheit) measurements, push the Menu softkey until °C or °F shows as the left softkey function. Push the corresponding softkey to change between the measurements.
7
Page 16
572-2
Users Manual

Min, Max, Avg, Differential

The Product can measure minimum (MIN), maximum (MAX), average (AVG), or differential (Δ) temperatures each time a reading is taken. These values are not shown if a thermocouple is plugged into the Product.
To turn on the Min Max mode:
1. Push the Menu softkey until MnMx (Min Max) shows as the left softkey function.
2. Push the MnMx softkey.
The display shows the present reading, maximum and minimum readings, and the emissivity setting.
To turn on the Avg/Dif mode:
1. Push the Menu softkey until Avg shows as the right softkey function.
2. Push the Avg softkey.
The display shows the present reading, average reading, the differential reading between max and min (designated by Δ), and the emissivity setting.

Alarm

The Product has a programmable high and low alarm to designate high or low readings, depending on the thresholds entered. When the alarm level is reached, an alarm will sound and the display will flash orange and white. To set either the high or low alarm:
1. Push the Menu softkey until Alarm shows as the right softkey function.
2. Push the Alarm softkey to access the Alarm menu.
3. Push either the Hi or Lo softkey, depending on which alarm will be set.
4. Push the ON softkey to turn the alarm on.
5. Push the OFF softkey to turn the alarm off.
6. Use the Set softkey to access the Hi or Lo Alarm Set menu.
7. Use the down or up softkeys to change the alarm setting.
8. Once the necessary settings have been entered, push the Done softkey.

Trigger Lock

The Product trigger can be locked on for continuous measurement.
To enable continuous measurement mode, hold down the trigger for a few seconds until the symbol shows.
To lock the trigger:
1. Push the Menu softkey until the lock symbol () shows as the left softkey
function.
2. Push the  softkey to lock the trigger. The lock symbol shows on the display to designate a locked trigger. When the trigger is locked, the softkey changes to . Push this softkey to unlock the trigger.
8
Page 17
Infrared Thermometer
Display

Laser

Warning
To prevent eye damage and personal injury, do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
The laser of the Product is to be used for aiming purposes only. The laser turns off when the trigger is released.
To enable or disable the laser:
1. Push the Menu softkey until Laser shows as the right softkey function.
2. Push the Laser softkey to enable or disable the laser.  shows on the display
when the laser is enabled.

Setup

From the Setup menu, the backlight, time and date, and display language can be changed.

Language

To change the display language:
1. From the main menu, push the Menu softkey until Setup shows as the left softkey function.
2. Push the Setup softkey.
3. Use the down arrow softkey to move the indicator to Language.
4. Push the Enter softkey.
5. Use the down arrow to move the indicator to the desired language.
6. Push the Enter softkey to complete the language selection or push the Back softkey to return to the Setup menu.

Backlight

In normal use, the backlight is always on. Use this menu to change the backlight setting from on to off. Turn the backlight off to conserve battery power.
1. Push the Menu softkey until Setup shows as the left softkey function.
2. Push the Setup softkey.
3. Backlight has an indicator next to it. To enter the backlight menu, push the Enter softkey.
4. Push the OFF softkey to turn the backlight off, push the ON softkey to turn it on.
5. Push the Back softkey to return to the Setup menu.

Time/Date

To change the time on the Product:
1. Push the Menu softkey until Setup shows as the left softkey function.
2. Push the Setup softkey to enter the Setup menu.
3. Push the down arrow softkey to select Time/Date.
4. Push the Enter softkey.
5. Push the Time softkey.
9
Page 18
572-2
Users Manual
6. Push the necessary time format softkey (24hr or 12hr).
7. Use the up and down softkeys to change the selected hour to the correct hour.
8. Push the Next softkey to select the minutes.
9. Use the up and down softkeys to change the selection to the desired minute.
10. When in 12-hour mode, push the Next softkey to highlight the am/pm parameter.
11. Use the up and down softkey to change to am or pm.
12. Push the Done softkey when finished. The display reverts to the first page of the Time/Date menu.
To change the date on the Product:
1. From the main menu, push the Menu softkey until Setup shows as the left softkey function.
2. Push the Setup softkey.
3. Push the down arrow softkey to select Time/Date.
4. Push the Enter softkey.
5. Push the Date softkey.
6. Select the date format: Day/Month/Year (dmy) or Month/Day/Year (mdy).
7. Use the up and down softkeys to change the selected parameter.
8. Push the Next softkey and the arrow softkeys to select the month, date, or year parameters.
9. Use the up and down softkeys to change the selected parameter.
10. Push the Next softkey to move through each parameter.
11. Push the Done softkey when finished. The display reverts to the start of the Time/Date menu.

Delete Data

To delete stored data from the Product, from the main menu, push the menu softkey until Mem shows as the left softkey function. The last memory location shows on the display. To access the Delete menu, push the Delete softkey. From here, all of the Product memory or individual records can be deleted.

Delete All Data

To delete all of the records:
1. Push the All softkey.
2. At the confirmation screen, push the Yes softkey.

Delete Individual Data Records

To delete individual records:
1. Push the View softkey.
2. Use the down and up arrow softkeys to access the desired record.
3. Once the desired record is shown, push the Yes softkey to delete the record.
4. To cancel data deletion, pull the trigger.
10
Page 19
Infrared Thermometer

How the Product Works

How the Product Works
Infrared thermometers measure the surface temperature of an object. The Product optics sense emitted, reflected, and transmitted energy, which is collected and focused onto a detector. The Product electronics translate the signal into a temperature measurement and shows the measurement on the display (see Figure 3).
Reflected energy
Emitted
Energy
Transmitted
Target
Figure 3. How the Product Works
energy
hdp03.eps

Product Operation

Temperature Measurement

To measure temperature, point the Product at an object and pull the trigger. You can use the laser pointer to aim the Product. You can also insert the K-type thermocouple probe for contact measurement. Be sure to consider distance-to-spot size ratio and field of view (see “Distance and Spot Size” and “Field of View”). The temperature is shown on the display.
Note
The laser is used for aiming purposes only and is not related to temperature measurement.
The Product features an auto off function that automatically powers down the Product after 20 seconds of inactivity. To turn the Product on, pull the trigger.
11
Page 20
572-2
Users Manual

Locate a Hot or Cold Spot

To find a hot or cold spot, aim the Product outside the desired area. Then, slowly scan across the area with an up and down motion until you locate the hot or cold spot (see Figure 4).
Figure 4. Locating a Hot or Cold Spot
hdp04.eps

Distance and Spot Size

As the distance (D) from the object being measured increases, the spot size (S) of the area measured by the Product becomes larger. The relationship between distance and spot size (D:S) for each Product is shown in Figure 5. The spot sizes indicate 90 % encircled energy.
31.70 mm @ 1500 mm
D:S=60:1
20.77 mm @ 500 mm
(.82 ” @ 19.69 ”)
21.38 mm @ 250 mm
(.84 ” @ 9.84 ”)
19.17 mm @ 1150 mm
(.75 ” @ 45.28 ”)
(1.25 ” @ 59.05 ”)
S
D
12
Figure 5. Distance and Spot Size
hdp05.eps
Page 21
Infrared Thermometer
Product Operation

Field of View

For accurate measurements, make sure that the target is larger than the Product spot size. The smaller the target, the closer you should be to it (see Figure 6).
Yes No
Figure 6. Field of View
hdp06.eps

Emissivity

Emissivity describes the energy-emitting characteristics of materials. Most organic materials and painted or oxidized surfaces have an emissivity of approximately
0.95, the default setting for the Product.
To compensate for inaccurate readings that may result from measuring shiny metal surfaces, you can cover the surface to be measured with electrical tape or flat black paint (<148 °C/300 °F) with emissivity set to 0.95. Allow time for the tape or paint to reach the same temperature as the surface beneath it. Measure the temperature of the tape or painted surface.
If you cannot paint or use tape, then you can improve the accuracy of your measurements by either numerically adjusting emissivity or by using the Emissivity Menu to access a table of some common materials. The Product has numerically­adjustable emissivity from 0.10 to 1.00 that can be used in conjunction with the emissivity in Table 3. Also, the Product has a built-in table of the most common materials that can be set from the Emissivity menu. The built-in values are indicated in Table 3.

HOLD

The display retains its last infrared measurement for twenty seconds when the trigger is released, when HOLD is shown on the display. With the probe inserted, the contact thermometer remains on. To freeze the infrared temperature when a probe is not inserted, release the trigger until HOLD shows on the display.
13
Page 22
572-2
Users Manual

Data Storage

The Product can store up to 99 data records. The information below is stored in each record:
Record number
IR and probe temperature in °F or °C
Date/Time
Emissivity
For more information, see “Save”.

Data Download

The stored data can be downloaded to a personal computer (PC) with the included USB cable and the included FlukeView FlukeView
®
Forms documentation for details. The USB input is located on the top
Forms Documenting Software. See the
of the Product next to the external probe input (see Figure 7).
Note
To prevent incorrect readings, do not do a temperature measurement of an earthed conductor while the Product is connected to a PC that is earthed by a three-phase grounding plug.
1
Figure 7. Thermocouple and USB Connections
2
hdp07.eps
14
Page 23
Infrared Thermometer

External Contact Probe

External Contact Probe
Warning
To avoid electrical shock or personal injury, do not connect the external contact probe to live electrical circuits.
The Product comes with a bead K-type thermocouple probe. The probe attaches to the Product via the probe input located on the top of the Product (see Figure 7).
With the probe installed, the probe symbol () shows on the display. The probe can be used simultaneously while the Product is taking non-contact measurements. The probe measurements are shown below the non-contact measurements. The Product will stay on when there is a probe inserted. When used with a personal computer and FlukeView Forms, the auto off feature is disabled.

Troubleshooting

See Table 4 for solutions to possible problems when using the Product.
Table 4. Troubleshooting
Symptom Cause Action
--- (on display)
Low batteries Replace batteries
Blank display
Laser does not work
Failed communication via USB
Inaccuracy
Settings such as emissivity, date/time, F/C, and saved data lost
Target temperature is over or under range.
Thermometer is asleep
Possible dead batteries
Low or dead batteries
Ambient temperature is above 40 °C (104 °F)
Unit is off
FlukeView Forms is not running
Possible incorrect emissivity setting, field of view, or spot size
Battery dead or not replaced in <1 minute of removal
Select target within specifications
Pull trigger
Replace batteries
Replace batteries
Use in area with lower ambient temperature
Pull the trigger
Start FlukeView Forms
See “Emissivity”, “Field of View” and “Distance and Spot Size” sections.
Reset settings. Replace batteries as soon as low battery indicated; Exchange the batteries within one minute of removal. Refer to “Changing the Batteries”.
15
Page 24
572-2
Users Manual

Maintenance

Battery Charge

Use the battery charge symbols to gauge the approximate level of charge left on the batteries.
Note
When in low battery mode, the Product does not store values. “Err” shows on the display if attempted.

Battery Replacement

To prevent possible electrical shock, fire, eye damage, and personal injury:
Do not open the Product. The laser beam is dangerous
to eyes. Have the Product repaired only through an approved technical site.
Have an approved technician repair the Product.
For safe operation and maintenance of the Product:
Repair the Product before use if the batteries leak.
Be sure that the battery polarity is correct to prevent
battery leakage.
Do not short the battery terminals together.
Do not keep cells or batteries in a container where the
terminals can be shorted.
Batteries are at 5 %. Before taking further readings, the batteries must be changed.
Warning
Caution
16
Do not put battery cells and battery packs near heat or
fire. Do not put in sunlight.
Note
The batteries should be replaced in less than one minute after removal to avoid manually re-initializing the Product clock and date.
To install or change the two AA batteries (see Figure 8):
1. Slide the locking tab on the battery door to the “” position and then open the handle.
2. Insert the batteries noting their correct polarity.
3. Close and lock the handle.
Page 25
Infrared Thermometer
AA
Figure 8. Battery Replacement

User-Replaceable Parts and Accessories

hdp08.eps

Clean the Lens

Blow off loose particles. Carefully wipe the surface with a water-moistened cotton swab.

Clean the Case

Use soap and water on a damp sponge or soft cloth.
Caution
To avoid damaging the Product, do NOT submerge it in water. Do not use abrasive cleaners, they will damage the case.
User-Replaceable Parts and Accessories

User-Replaceable Parts

See Table 5 for a list of replaceable parts.
Table 5. Replaceable Parts
Part Part Number
572-2 Manuals CD 4307031
572-2 Getting Started
(English, French, German, Spanish, Portuguese, Russian, Simplified Chinese)
4307046
17
Page 26
572-2
Users Manual

Accessories

Optional accessories for the Product are:
Soft Carrying Case (H6)
Calibration Certification
All K-type thermocouple probes with standard mini-connector. See Table 7 for
more information.

Recommended Temperature Probes

Warning
To avoid electrical shock or personal injury, do not connect the external contact probe to live electrical circuits.
See Table 6 for a list of recommended temperature probes.
Table 6. Recommended Temperature Probes
Probe Usage
80PK-1
80PK-8
80PK-9
80PK-11
80PK-25
80PK-26
The general purpose bead probe is for quick, accurate surface temperatures and air temperatures within ducts, vent temperatures.
Pipe clamp probes (2) are essential for tracking continuously changing temperature differentials on hydronic tubing and pipe loops, and good for quick, accurate refrigerant temperatures.
The insulation-piercing probe provides a sharp tip to pierce pipe insulation and flat probe tip for good surface thermal contact, air temperatures within ducts, and vent temperatures.
Flexible cuff thermocouple temperature probe is a convenient way to attach a thermocouple to a pipe while keeping hands free.
The piercing probe is the most versatile option. Good for checking air temperature in ducts, surface temperature under carpets/pads, liquids, thermometer wells, vent temperatures, and for penetrating pipe insulation.
The tapered probe is a good general-purpose gas and surface probe, with a good length and low mass tip casing for faster reaction to surface and air temperatures.
18
Page 27
Infrared Thermometer

Specifications

Specifications
Thermal
IR Temperature Range -30 °C to 900 °C (-22 °F to 1652 °F)
IR Accuracy (Calibration geometry with ambient temperature 23 °C±2 °C)
IR Repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater. Display Resolution 0.1 °C / 0.1 °F Spectral Response 8 μm to 14 μm
Response Time (95 %) <500 ms
K-Type Thermocouple Input Temperature Range
K-Type Thermocouple Input Accuracy (with ambient temperature 23 °C±2 °C)
K-Type Thermocouple Repeatability
K-Type Thermocouple Resolution
Optical
Distance:Spot (90 % energy)
Laser sighting Offset dual laser, output <1 mW
Minimum spot size 19 mm
Operational
Emissivity
Probe input KTC (Should display simultaneous probe and IR temperature)
Data storage 99 points
Communication USB 2.0
Operating Altitude 2000 meters
Storage Altitude 12,000 meters
Relative Humidity 10 % to 90 % RH non-condensing up to 30 °C (86 °F)
Operating Temperature 0 °C to 50 °C (32 °F to 122 °F) Storage Temperature -20 °C to 60 °C (-4 °F to 140 °F)
Vibration 2.5 G, IEC 60068-2-6
Weight 0.322 kg (0.7099 lb)
Dimensions 17.69 cm (6.965 in) H x 16.36 cm (6.441 in) L x 5.18 cm (2.039 in) W
Power
Battery Life
Standards and Agency Approval
0 °C: ± 1 °C or ± 1 % of reading, whichever is greater (32 °F: ± 2 °F or ± 1 % of reading, whichever is greater) -10 °C to <0 °C: ±2 °C(14 °F to <32 °F: ±4 °F) <-10°C: ±3 °C (<14 °F: ±6 °F)
-270 °C to 1372 °C (-454 °F to 2501 °F)
<-40 °C: ±(1 °C + 0.2 °/1 °C) -40 °C: ±1 % or 1 °C, whichever is greater <-40 °F: ±(2 °F + 0.2 °/1 °F)
-40 °F: ±1 % or 2 °F, whichever is greater
±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater
0.1 °C/0.1 °F
60:1(calculated at 90 % energy)
Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials (see Table 3 for list of materials and value)
2 AA Batteries
8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100 % duty cycle (thermometer continuously on)
EMC: meets IEC61326-1: Portable Safety Compliance: IEC 60825-1, Class 2
Implement Standard: Q/SXAV 16
19
Page 28
572-2
t
Users Manual
Electromagnetic Compatibility
Applies to use in Korea only. Class A equipment; (Industrial Broadcasting & Communication Equipment)
[1] This product meets the requirement for industrial (Class A) electromagnetic wave
equipment and seller or user should take notice of it. This equipment is intended use in business environment and is not to be used in homes.
[1]

K-Type Thermocouple Bead Probe Specifications

Measurement Range -40 °C to 260 °C (-40 °F to 500 °F)
Accuracy
Cable Length
±1.1 °C (±2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F). Typically within 1.1 °C (2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F)
1 m (40 in) K-type thermocouple cable with standard miniature thermocouple connector and bead termination
20