Fluke Ti40FT-20, Ti45FT-20, FlexCam Ti40FT, FlexCam Ti45FT, FlexCam Ti40 Datasheet

...
The Fluke Ti4x models feature everything needed for virtually every thermography task. With a 160 x 120 detector and a temperature sensitivity to 0.08 °C (NETD) they deliver high resolution images where even the smallest temperature differences can be seen. The units are extremely easy to use through the Windows® CE menu structure and offer an extended troubleshooting feature set to allow on the spot analysis in the field.
The versatile choice for maintenance and production engineers and technicians.
Features
High resolution, low noise VOx detector for high quality images Temperature range to cover broad industrial applications High temperature option High thermal sensitivity for viewing even the smallest
temperature differences 180° articulating fl exible lens to view images in every situation Choice of 3 interchangeable lenses to cover every application Large 5” high contrast color LCD for a clear picture independent
of lighting conditions Fully radiometric for detailed temperature analysis and tracking SmartFocus for best image quality and accurate temperature
measurements Windows CE based menu structure for ease of use Personalized instrument set-up for multiple use CompactFlash memory cards to store over 1000 IR images plus
fully radiometric temperature data SmartView reporting and analysis software included AutoCapture for making intermittent problems visible On-board analysis functions User defi ned text annotations for simplifi ed reporting Built-in visible light (digital) camera IR-Fusion blending thermal and visible light images to easily
pinpoint suspect components IR/Visible Alarm function Laser pointer for easy targeting Flash and torch light for high quality images in dark
environments
Ti45FT Ti45 Ti40FT Ti40
160 x 120
-20 to +600 °C (-4 to 1112 ºF) -20 to +350 °C (-4 ºF to 662 ºF) 1200 °C (2192 °F) 0.08 °C 0.09 °C
Picture-in-picture
only
Typical applications:
Predictive maintenance – Identify electrical and mechanical problems before they
cause failure
Industrial maintenance – Check whether repairs have been performed successfully
Quality control – Examine prototypes and refine thermal management designs
Process monitoring – Real-time observation to ensure efficient and safe operation
Motor
3-phase system
Predictive maintenance
Ti40 and Ti45 IR FlexCam® Thermal Imagers
To fi nd out more about Fluke thermal imagers go to www.fl uke.com/thermography or in the United States call
1-800-760-4523.
Imaging performance
Temperature measurement
Image presentation
Optional lenses
Image and data storage
Interfaces and software
Laser (IR-Fusion models only)
Controls and adjustments
Power
Environmental and mechanical design
Other
Fluke Ti45 Fluke Ti40
Thermal Field of view (FOV)* 23° horizontal x 17° vertical Spatial resolution (IFOV)* 2.60 mrad Min focus distance* 0.15 m Thermal sensitivity (NET D) 0.08 °C at 30 °C 0.09 °C at 30 °C Detector data acquisition / 30 Hz/30 Hz
Image frequency Focus SmartFocus; one fi nger continuous focus IR digital zoom 2x ­Detector type 160 x 120 Focal Plane Array, Vanadium Oxide (VOx) Uncooled Microbolometer Spectral band 8 µm to 14 µm Digital image enhancement Automatic full-time enhanced
Visual (IR-Fusion models only)
On camera operating modes Full thermal, full visual light or merged Full thermal or full v isual light. thermal-visual images. Merge thermal-visual images in SmartView Picture-in-Picture software. Picture-in-Picture
Visible light camera 1280 x 1024 pixels, full color Visible light digital zoom 2x -
Calibrated temperature range -20 °C to 600 °C (-4 ºF to 1112 ºF) in 3 ranges -20 °C to 350 °C (-4 ºF to 662 ºF) in 2 ranges Range 1 = -20 °C to 100 °C (-4 ºF to 212 ºF) Range 1 = -20 °C to 100 °C (-4 ºF to 212 ºF) Range 2 = -20 °C to 350 °C (32 ºF to 662 ºF) Range 2 = -20 °C to 350 °C (32 ºF to 662 ºF) Range 3 = 250 °C to 600 °C (482 ºF to 1112 ºF) ­Optional - High temperature Up to 1200 °C (2192 ºF) ­ Range 4 = 500 °C to 1200 °C (932 ºF to 2192 ºF) ­Accuracy ±2 °C or 2% (whichever is greater) Measurement modes Centerpoint, center box (area min/max,
average), moveable spots/boxes, user defi ned fi eld/text annotations, isotherms, Centerpoint, center box automatic hot and cold point detection, (area min/max, average) visible color alarm above and below
Emissivity correction 0.1 to 1.0 (0.01 increments)
Digital display 5” large high-resolution digital display LCD backlight Sunlight readable color LCD Video output RS170 EIA/NTSC or CCIR/ PAL composite video Palettes Grayscale, grayscale inverted, blue red, high contrast, hot metal, ironbow, amber, amber inverted
54 mm Telephoto lens High precision Germanium lens Field of view (FOV) 9° horizontal x 6° vertical Spatial resolution (IFOV) 0.94 mrad Min focus distance 0.6 m
10.5 mm wide angle lens High precision Germanium lens Field of view (FOV) 42° horizontal x 32° vertical Spatial resolution (IFOV) 4.9 mrad Min focus distance 0.3 m
Storage medium Compact fl ash card stores over 1000 IR images (512MB card standard) File formats supported 14 bit measurement data included. Exportable J PEG, BMP, PCX, PNG, PS D.
Interface Compact fl ash card reader included Software SmartView; Full analysis and reporting software included
Classifi cation Class II Laser targeting Laser dot v isible on screen when blending thermal and visible image
Set-up controls Date/time, temperature units C/ F, language, scale, LCD intensity (high/normal/low) Image controls Level, span, auto adjust (continuous/manual) On-screen indicators Battery status, target emissivity, background temperature and realtime clock
Battery type Li-Ion smart battery, rechargeable, fi eld-replaceable Battery operating time 3 hours continuous operation (2 hours for models w ith IR-Fusion) Battery charging 2 bay intelligent charger powered via AC outlet AC operation AC adapter 110/220 VAC, 50/60 Hz ­Power saving Automatic shutdown and sleep modes (user specifi ed)
Operating temperature -10 °C to +50 °C (14 °F to 122 °F) Storage temperature -40 °C to +70 °C (-40 °F to 158 °F) Relative humidity Operating and storage 10% to 95%, non-condensing Water and dust resistant IP54 Weight (including batteries) 1.95 kg (4.3 lbs) Camera size (HxWxD) 162 x 262 x 101mm (6.5” x 10.5” x 4.0”)
Warranty 2 years
Included accessories
Heavy duty carrying case 2 rechargeable battery packs Battery charger AC adapter (for Ti45 only) Video cable 512 MB compact flash card Compact flash card reader and USB cable PCMCIA compact flash card reader Neck strap SmartView reporting and analysis software on CD User manual on CD
Ordering information*
FLK Ti40-20 IR FlexCam Thermal
Imager
FLK Ti40FT-20 IR FlexCam Thermal
Imager with Fusion
FLK Ti45-20 IR FlexCam Thermal
Imager
FLK Ti45FT-20 IR FlexCam Thermal
Imager with Fusion
*For ordering information of optional lenses check the Fluke web
Specifications
*standard 20 mm Germanium lens
Visit the Fluke web site for complete specifi cations
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved. Printed in U.S.A. 5/2006 2674273 D-US-N Rev A
Fluke. Keeping your world up and running.
Loading...