Fluke Fiber OneShot PRO Quick Reference Manual

Page 1
PN 4805467, May 2016
©2016 Fluke Corporation All product names are trademarks of their respective companies.
FIBER ONESHOT™ PRO
Singlemode Fiber Troubleshooter
Accessing the Users Manual
This guide provides basic information to help you get started using the FIBER QNESHOT PRO Singlemode Troubleshooter. For additional information, see the FIBER Q
NESHOT PRO Users
Manual on the Fluke Networks website.
Battery Installation
fjy03.eps
AA
IEC LR6
Note: Fluke Networks recommends alkaline batteries.
Page 2
2
FIBER ONESHOT PRO Quick Reference Guide
Display Features
gmh01.eps
When you press , CheckActive
shows and the troubleshooter looks for an optical signal on the fiber. If there is a singlemode signal stronger than approximately
-29 dBm on the fiber, ACTIVE LINE blinks and the troubleshooter will not do a test.
WCaution
If ACTIVE LINE blinks, immediately disconnect the troubleshooter from the fiber. Optical power levels more than +7 dBm can cause damage to the detector in the troubleshooter.
When the low battery symbol shows, replace the batteries soon. See page 1.
The digits show the fiber length in feet, kilofeet, meters, or kilometers.
Page 3
3
Display Features
sec: Shows when you look at the setting for the backlight timer. The setting is in seconds.
dB: Shows when you look at the setting for the reflection limit. The setting is in decibels.
dB: Shows when the display shows the reflectance of an incident on the fiber. The measurement is in decibels.
: Shows when the display shows the power loss of an
incident on the fiber. The measurement is in decibels.
RL: Result loss.
MORE : Shows when the troubleshooter finds more than one incident. Press   to see more incidents. The numbers show the number of the incident and the total number of incidents. The troubleshooter shows up to 9 incidents. If there are more than 9 incidents on the fiber, the last
blinks when you look at the ninth incident.
SAVED: Shows after a test. The troubleshooter automatically saves the test results.
TESTING: Shows as the troubleshooter does a test.
BREAK OR END: Shows when the troubleshooter shows the distance to a break or the end of the fiber.
: The measurement is above or below the range that the troubleshooter can show, or the troubleshooter cannot give a more accurate measurement.
RESULT: Shows when you look at saved results.
Settings for the troubleshooter. LOSS LIMITW and REFLECTION LIMITW blink if a measurement is equal to or
greater than the limit you select.
Page 4
4
FIBER ONESHOT PRO Quick Reference Guide
Settings
To change the settings on the troubleshooter:
1
Hold down until the settings menu shows.
2
To select a setting to change, press , then press .
3
Use to change the setting.
4
Press or to save the setting.
5
To exit the settings menu, hold down for 4 seconds. To exit and do a test, press
.
BACKLIGHT: Timer for the display backlight.
I.O.R.: Index of refraction.
ft/m/kft/km: Unit for length measurements.
LOSS LIMIT: Sets the minimum level at which the troubleshooter shows the warning LOSS LIMITW for an incident.
REFLECTION LIMIT: Sets the minimum size at which the
troubleshooter shows the warning REFLECTION LIMITW for an incident.
Page 5
5
The Connector Adapter
The Connector Adapter
You can change the SC connector adapter to connect to LC, ST, or FC fiber connectors.
WCaution
Turn only the collar on the adapter. Do not use tools to remove or install the adapters.
gmh08.eps
How to Remove and Install the Connector Adapter
Key
Slot
Keep extra adapters in the containers provided.
Page 6
6
FIBER ONESHOT PRO Quick Reference Guide
How to Clean Connectors
Fluke Networks recommends that you use a mechanical device to clean connectors. If you do not have such a device, use other optical-grade supplies to clean connectors. See the Users Manual.
Use a video microscope, such as the Fluke Networks FI-7000 FiberInspector
Pro, to inspect connectors.
WCaution
To prevent damage to the device and to connectors, read all instructions and obey all safety precautions given by the manufacturer of the device you use to clean connectors.
ghm09.eps
How to Use a Mechanical Device to Clean Connectors
Protect Connectors
Always put protective caps on connectors that you do not use. Clean the caps periodically with a swab or wipers and fiber optic solvent.
CLICK!
CLICK!
Page 7
7
About Launch and Receive Fibers
About Launch and Receive Fibers
Launch and receive fibers give the troubleshooter a better view of the first and last connectors in the link. If you do not use a launch fiber, the troubleshooter cannot detect the loss of the first connector in the link. If you do not use a receive fiber, the troubleshooter cannot detect the loss of the last connector in the link. Also, the reflectance measurement for the first and last connectors will be inaccurate.
Usually, you do not need to use a launch or receive fiber with the FIBER O
NESHOT PRO tester. But you must use them in these
situations:
You must use a launch fiber if you want to detect a loss incident (such as a connector, splice, or macrobend) in the first 30 m of the fiber link.
You must use a receive fiber if you want to detect the loss of the last connector in the fiber link.
If you use a launch or receive fiber, it must have a minimum length of 30 m (98 ft).
When you use launch or receive fibers, be sure to subtract their lengths from the length measurement to get the actual length of the fiber you are testing.
Page 8
8
FIBER ONESHOT PRO Quick Reference Guide
How to Use the Troubleshooter
Notes
Always use patch cords that comply with GR-326-CORE specifications and have a UPC connector at the end you will connect to the troubleshooter. Other patch cords can cause unreliable measurements and damage to the troubleshooter.
Do not use the troubleshooter to do tests on fibers that have PC connectors. PC connectors cause large reflections that the troubleshooter shows as the end of the fiber.
1
Clean all fiber connectors.
2
Connect the fiber to the troubleshooter.
3
Turn on the troubleshooter, then press . The troubleshooter automatically saves the results.
Note
After you turn on the troubleshooter, it shows for a short time to show that it operates correctly.
WCaution
If ACTIVE LINE blinks, immediately disconnect the troubleshooter from the fiber. Optical power levels more than +7 dBm can cause damage to the detector in the troubleshooter.
Page 9
9
How to Use the Troubleshooter
gmh05.eps
How to Make Connections
Note: Always use patch cords that comply with GR-326-CORE specifications and have UPC connectors at one end.
Patch panel Patch panel
Launch
fiber
Receive
fiber
Mechanical splice
Unterminated
spool of fiber
Installed fiber
WUPC connector only
WUPC connector only
Page 10
10
FIBER ONESHOT PRO Quick Reference Guide
gmh02.eps
Examples of Measurement Results
The reflectance of the first connection is -52 dB. The length of the launch fiber is 130 m.
The end of the link. The length of the link (without the launch fiber) is 8463 m.
A bad splice on the fiber at 2450 m caused a reflectance incident that is larger than
-45 dB (the default limit).
To see the loss measurement, press
.
Blinking
Incident
is the connection to the
troubleshooter.
The result loss of the fiber (
RL) is 2.6 dB.
The receive fiber adds 130 m to the length measurement.
Page 11
11
How to See Saved Results
How to See Saved Results
gmh03.eps
When RESULT shows on the results display, use
to scroll
through results.
To use   to scroll through results again, press
to put the
bar back on RESULT.
To open the selected result, press
to remove the bar. Use
and to examine the result.
Page 12
12
FIBER ONESHOT PRO Quick Reference Guide
How to Delete Saved Results
Note
This procedure deletes all saved results. You cannot delete individual results.
1
Hold down until the settings menu shows.
2
With RESULT selected, press to see the results display.
3
Hold down for 4 seconds. blinks.
4
To delete all tests, press . The display shows to show you that the tests were deleted.
5
To exit and not delete tests, press or 
Contact Fluke Networks
Fluke Networks operates in more than 50 countries worldwide. For more contact information, go to our website.
Warranty
1-year limited warranty. See the website for details.
www.flukenetworks.com
support@flukenetworks.com +1-425-446-5500
Fluke Networks 6920 Seaway Boulevard, MS 143F Everett, WA 98203 USA