Fluke FI-525, FI-500 User manual

Page 1
FI-500
FiberInspector™ Micro
Video Microscope
The FI-500 FiberInspector Micro video microscope is a hand-held probe and display used to inspect fiber optic endfaces on patch cord and bulkhead connectors.
WSafety Information
WWarningX
To avoid possible fire, electric shock, personal injury, or damage to the video microscope:
Carefully read all instructions.
Read all safety information before you use the Product.
Use the Product only as specified, or the protection supplied by the Product can be compromised.
Use this Product indoors only.
Do not use the Product around explosive gas, vapor, or in damp or wet environments.
Do not use the Product if it operates incorrectly.
Do not use the Product if it is damaged.
Disable the Product if it is damaged.
Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
The battery door must be closed and locked before you operate the Product.
Do not put metal objects into connectors.
September 2016, Rev. 1 9/2016
2016 Fluke Corporation All product names are trademarks of their respective companies.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Safety and Compliance Symbols
Warning or Caution: Risk of damage to or destruction of equipment or
W
software.
X
P
Ã
~
See the user documentation.
Warning: Risk of electrical shock.
Conforms to relevant European Union directives
Conforms to relevant Australian standards
40 year Environment Friendly Use Period (EFUP) under China Regulation - Administrative Measure on the Control of Pollution Caused by Electronic Information Products. This is the period of time before any of the identified hazardous substances are likely to leak out, causing possible harm to health and the environment.
KCC-REM-FKN-012001001: EMC approval for Korea
Class A Equipment (Industrial Broadcasting & Communication Equipment)
This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.
This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/ electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste.
To return unwanted products, contact the manufacturer’s web site shown on the product or your local sales office or distributor.
Features and Use
Refer to Figure 1.
2
Find Quality Products Online at: sales@GlobalTestSupply.com
Charge the batteries for about 4 hours before you use them for the first time. See “Charge the Batteries” on page 7. To ensure continuous operation, connect the ac adapter whenever possible.
www.GlobalTestSupply.com
Page 3
SC LC
D
C
B
A
E
W
F1
F2 F3
F4
C
F
D
G
H
SC (U25) LC (U125)
Cap holder
Screw an adapter tip onto the fiber probe to match the connection you
will inspect. Tips with blue rings are for PC and UPC connectors. Tips with green rings are for APC connectors.
To avoid damage to the probe, use only tips specified for the FI-500. Do not use tips from other video microscopes.
Figure 1. Features and Use
WCaution
AM00.EPS
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
In live mode when the Auto Focus setting is On, press AF to automatically focus the
image.
In live mode when the Auto Focus setting is Off, press + and - to focus the image manually.
Keep the probe still when you use the auto
focus function.
The blue, vertical bar is the range of focus. The yellow mark shows
where the focus is within that range. The green mark shows where the image will be focused.
Switch the probe between live and paused modes. Figure 2 shows the button functions in paused mode.
Turns the probe light on and off.
See “Settings” on page 6.
The probe is connected.
The probe is not connected.
The display does not recognize the connected probe.
Shows you when the display is live or paused.
Shows you the display’s focus mode. AF is auto focus. MF is manual focus. See “Settings” on page 6.
A USB drive is connected to the display.
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
AM04.EPS
Zoom
Horizontal
Vertical
Live image
Focus automatically or manually. See
Auto Focus under "Settings".
Paused image
Magnify and move the image.
Green
ring
Angled
ferrule
Green
housing
SC bulkhead tip
Green
ring
1.25 mm
connector tip
SC
connector
Figure 2. Image Controls
View APC Endfaces
Probe tips for APC (angled physical contact) connectors have a green ring (Figure 3). When you view an APC endface, rotate the probe or the connector while you use the auto focus function until you get the best image. If you cannot get a good image, focus the image manually.
AM08.EPS
Figure 3. APC Tip and Connector Examples
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5
Page 6
Settings
Select a setting Change the setting
Refer to Figure 4.
Display and probe software versions: If a software update is available, you can download it from the Fluke Networks website. See "Upgrade the Software".
Power Down: The display turns off after the selected period of inactivity.
Standby: The display goes into low-power mode after the selected period of inactivity.
Dim Display: The display dims after the selected period of inactivity.
Probe Light: The probe light stays on for the selected time.
Brightness: Changes the brightness of the display. Keeping the display at less than 100% brightness increases the battery life.
Dimmed Brightness: Changes the brightness of the display when the display is dim.
Auto Focus
On: Press on the probe or on the display to focus the image
automatically. The display will also focus the image automatically when
you press
Off: Press to focus the image manually.
Save and Exit: Saves the settings and exits the settings menu.
on the probe or on the display.
AM02.EPS
Figure 4. Settings
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
Charge the Batteries
WWarning
To avoid possible fire, electric shock, personal injury, damage to the video microscope, or incorrect operation:
Use only the ac adapter/charger provided to power the display and charge the batteries.
Disconnect the battery charger and move the Product or batteries to a cool, non-flammable location if the rechargeable batteries become hot (>50 °C, >122 °F) during the charge period.
Do not recharge non-rechargeable batteries.
Batteries contain hazardous chemicals that can cause burns or explode. If exposure to chemicals occurs, clean with water and get medical aid.
Repair the Product before use if the battery leaks.
Do not keep cells or batteries in a container where the terminals can be shorted.
Do not put battery cells and battery packs near heat or fire. Do not put in sunlight.
When the battery charges, the LED is red. Charge the batteries for about 4 hours before you use them for the first time.
X
The charger is connected and no batteries are installed.
When the battery is fully-charged, the LED is green. Fully­charged batteries last about 3 hours when the display is on continuously.
When the battery is very low, the battery icon shows a red bar.
The LED blinks red and green and the yellow battery icon shows if you connect the power adapter when the wrong type of batteries are installed. See the specifications for the correct battery types.
The microscope will operate with alkaline batteries, but they will drain quickly. The display will not try to charge alkaline batteries.
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Maintenance
W
Rechargeable NiMH,
1.2 V, 2200 mAh*
* Fluke Networks recommends Gold Peak GH230AAHC batteries. Use only
NiMH batteries. Do not use other types of batteries.
WWarningX
To prevent possible electrical shock, fire, or personal injury:
Do not operate the Product with covers removed or the case open. Hazardous voltage exposure is possible.
Use only specified replacement parts.
Have an approved technician repair the Product.
Clean the LCD display with a soft, lint-free cloth.
Clean the display and probe housings with a soft cloth dampened with water or a mild detergent. Do not use solvents or abrasive cleansers.
If the probe lens is dirty, clean it with optical-grade cleaning supplies.
Replace the Batteries
WWarningX
Replacement cells used shall have closely matched capacities, be of the same design, be of the same chemistry and be from the same manufacturer and age.
Use only the specified type of batteries, correctly installed, to power the display. See the specifications for battery types.
Replace the rechargeable batteries after 5 years of moderate use or 2 years of heavy use. Moderate use is defined as recharged twice a week. Heavy use is defined as discharged to cutoff and recharged daily.
Replace the NiMH batteries at intervals described under “Charge the Batteries” on page 7. Figure 5 shows how to remove and install the batteries.
AM01.EPS
Figure 5. Replace the Batteries
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Upgrade the Software
B
C
A
D
If new software is available, you can download it from the Fluke Networks website at no charge. Figure 6 shows how to install new software in the display and probe.
Refer to Figure 6.
Download the upgrade file from the Fluke Networks website to the root
folder of a USB drive. If the file is a .zip file, unzip the files before you put them on the USB drive.
Connect the drive to the display. The software update icon shows in the status
bar, and the display imports the upgrade file. The screen shows the software versions loaded from the USB flash drive (Figure 7).
If there is not a software update file in the root folder of the drive or if you
press
Exit, the USB drive icon shows:
Connect the probe to the display. To upgrade the software, press OK.
To cancel the upgrade, press
When the upgrade is completed, press
Figure 6. Upgrade the Software
Exit.
Exit to restart the display.
AM06.EPS
Get Service
Contact Fluke Networks for information on authorized Fluke Networks service centers.
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
AM07.EPS
Figure 7. Software Update Screen with the USB Drive Connected
Register Your Product
Registering your product with Fluke Networks gives you access to valuable information on product updates, troubleshooting tips, and other support services. To register, fill out the online registration form on the Fluke Networks website.
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
Open Source Software
This product uses freeRTOS and FatFS software. The software license statements and files that contain the binary and source code for the open source software are on the CD supplied with the product and on the Fluke Networks website. For more
Specifications
General Specifications
Temperature range*
Humidity range
Altitude Operating: 4,000 meters; Storage: 12,000 meters
Vibration 2 g, 5 Hz to 500 Hz
Shock 1 meter drop test
Safety IEC 61010-1: Pollution Degree 2
EMC IEC 61326-1: Controlled Electromagnetic Environment; IEC 61326-2-1
Operating: 32 °F to 122 °F (0 °C to +50 °C) Storage: -22 °F to +140 °F (-30 °C to +60 °C)
Operating: 0 % to 95 % (32 °F to 95 °F, 0 °C to 35 °C) RH non-condensing Storage: 0 % to 95 % (95 °F to 113 °F, 35 °C to 45 °C) RH non-condensing
CISPR 11: Group 2, Class A Group 2: Equipment contains ISM RF equipment in which radio­frequency energy in the frequency range 9 kHz to 400 GHz is intentionally generated and used or only used, in the form of
electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purposes. Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted and radiated disturbances. USA (FCC): 47 CFR 15 Intentional Radiators: This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. (15.19). Changes or modifications not expressly approved by Fluke could void the user's authority to operate the equipment. (15.21) Emissions which exceed the levels required by CISPR 11 can occur when the Product is connected to a test object.
* Without the power adapter. With the power adapter, operating range is 0 °C to
+40 °C and storage range is -20 °C to +60 °C.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
Display
Frames per second 12
Battery type* Rechargeable NiMH, 2 x 1.2 V, 2200 mAh
Battery life* 3 hours of continuous probe use
6 hours of typical probe use
Charging time and temperature
Power adapter Input: 100 to 240 VAC ±10%, 50/60Hz
Display 3.2 inch TFT LCD, 320 x 240
Software upgrades Upgrades can be installed from a USB drive
Input USB 2.0, Type A
Dimensions 5.5in x 3.2in x 1.5in (14.0 cm x 8.0cm x 3.9cm)
Weight 9.7 oz (275 g)
* Tested using Gold Peak GH230AAHC batteries. Use only NiMH batteries. Do not use
other types of batteries.
4 hours minimum 50 °F to 113 °F (10 °C to 45 °C)
Output: 6 VDC, 3 A maximum
LPS
Probe
Camera type 5 Megapixel 1/4-inch CMOS sensor
Field of view 610 μm x 460 μm
Resolution 1 μm
Light source LED, >100,000 hr life
Endface illumination
Port illumination 2 white LEDs
Power Supplied through the USB interface
Output Video output through USB 2.0 interface
Dimensions 4.6 in x 2 in x 0.95 in (117 mm x 51 mm x 23 mm)
Weight 4.4 oz (125 g) (with no adapter tip)
12