Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of
original purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center
for calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our
option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to
Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if
the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized
Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE
EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration
of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since
some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages,
this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or
other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
7/07
Page 3
Notices
All Rights Reserved
Copyright 2013, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or
translated into any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service
training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke
Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking
the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking
instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or
broken parts, dents, or scratches.
Technical Support
For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800- 850-4608 or 1-440248-9300. In Europe, email techsupport.emea@flukebiomedical.com or call +31-40-2675314.
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in
their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not
operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local
sales representative.
Page 4
Returns and Repairs
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an
instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that
you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are
received in damaged condition due to improper packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
Use a double–walled carton of sufficient strength for the weight being shipped.
Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts.
Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order
Entry Group at 1-440-498-2560.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service or
In the U.S.A.:
Cleveland Calibration Lab
Tel: 1-800-850-4608 x2564
Email: globalcal@flukebiomedical.com
To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product be calibrated at least once
every 12 months. Calibration must be done by qualified personnel. Contact your local Fluke Biomedical representative for calibration.
In Europe, Middle East, and Africa:
Eindhoven Calibration Lab
Tel: +31-40-2675300
Email: servicedesk@fluke.nl
In Asia:
Everett Calibration Lab
Tel: +425-446-6945
Email: service.international@fluke.com
Page 5
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was
shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for
which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper
operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the
information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical
for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The ESA615 Electrical Safety Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.
To prevent possible electrical shock, fire, or
personal injury, read all “safety information”
before you use the Product.
The Fluke Biomedical ESA615 Electrical Safety Analyzer
(the Product) is a full-featured, compact, portable
analyzer, designed to verify the electrical safety of
medical devices. The Product tests to domestic
(ANSI/AAMI ES1, NFPA 99) and international (IEC62353,
AN/NZS 3551, and parts of IEC 60601-1) electrical-safety
standards. The Product simulates ECG to do
performance tests on ECG monitors.
The integrated ANSI/AAMI ES1 and IEC60601-1 patient
loads are easily selectable.
The Product does these tests:
• Line (Mains) voltage
• Ground Wire (Protective Earth) resistance
• Equipment current
• Insulation resistance
• Ground (Earth) leakage
• Chassis (Enclosure) leakage
• Lead to Ground (Patient) and Lead to Lead (Patient
Auxiliary) leakage
• Lead isolation (Mains on applied parts leakage)
• Differential leakage
1
Page 16
ESA615
Users Manual
• Direct equipment leakage
• Direct applied part leakage
• Alternative equipment leakage
• Alternative applied part patient leakage
• Point to point leakage, voltage, and resistance
• ECG simulation and performance waveforms
Table is a list of the symbols used on the Product and in
this manual.
Table 1. Symbols
Symbol Description
Important information. Refer to manual.
Hazardous voltage.
Conforms to relevant Canadian and US
standards.
Conforms to relevant Australian EMC
requirements.
Conforms to European Union directives.
Symbol Description
This product complies with the WEEE
Directive (2002/96/EC) marking
requirements. The affixed label indicates
that you must not discard this
electrical/electronic product in domestic
household waste. Product Category:
CAT II
With reference to the equipment types in
the WEEE Directive Annex I, this product
is classed as category 9 "Monitoring and
Control Instrumentation" product. Do not
dispose of this product as unsorted
municipal waste. Go to Fluke’s website
for recycling information.
IEC Measurement Category II – CAT II
equipment designed to protect against
transients from energy-consuming
equipment supplied from fixed
installations.
Accessible Functional Earth Terminal.
2
Page 17
Electrical Safety Analyzer
Intended Use
Intended Use
The Product is an electronic signal source and
measurement device for verifying the electrical safety of
medical devices. The Product also provides ECG
simulation and performance waveforms to verify patient
monitors are performing within their operating
specifications.
The Product provides the following function categories:
• ECG Functions
• ECG-Performance Testing
The intended user is a trained biomedical equipment
technician who performs periodic preventative
maintenance checks on patient monitors in service. Users
can be associated with hospitals, clinics, original
equipment manufacturers and independent service
companies that repair and service medical equipment.
The end user is an individual, trained in medical
instrumentation technology.
This Product is intended to be used in the laboratory
environment, outside of the patient care area, and is not
intended for use on patients, or to test devices while
connected to patients. This Product is not intended to be
used to calibrate medical equipment. It is intended for
over the counter use.
Safety Information
In this manual, a Warning identifies conditions and
procedures that are dangerous to the user. A Caution
identifies conditions and procedures that can cause
damage to the Product or the equipment under test.
Warning
To prevent possible electrical shock, fire, or
personal injury, follow these guidelines:
•Use the Product only as specified, or the
protection supplied by the Product can
be compromised.
•Use only the mains power cord and
connector approved for the voltage and
plug configuration in your country and
rated for the Product.
•Do not connect the Product to a patient
or equipment connected to a patient. The
Product is intended for equipment
evaluation only. The Product must not be
used in diagnostics, treatment, or other
capacities where the Product could
touch a patient.
3
Page 18
ESA615
Users Manual
•Do not use test leads if they are
damaged. Examine the test leads for
damaged insulation, exposed metal, or if
the wear indicator shows. Check test
lead continuity.
•Examine the case before you use the
Product. Look for cracks or missing
plastic. Carefully look at the insulation
around the terminals.
•Make sure the ground conductor in the
mains power cord is connected to a
protective earth ground. Disruption of
the protective earth could put voltage on
the chassis that could cause death.
•Do not use an extension cord or adapter
plug.
•Do not use and disable the Product if it
is damaged.
•Do not use the Product around explosive
gas, vapor, or in damp or wet
environments.
•Do not touch voltages > 30 V ac rms, 42
V ac peak, or 60 V dc.
•Measure a known voltage first to make
sure that the Product operates correctly.
•Do not apply more than the rated
voltage, between the terminals or
between each terminal and earth ground.
• Remove the null post adapter from the
∅/Null jack after a test lead zero is
performed. The ∅/Null jack becomes
potentially hazardous during some of the
test conditions. Use only cables with
correct voltage ratings.
•Remove all probes, test leads, and
accessories that are not necessary for
the measurement.
•Keep fingers behind the finger guards on
the probes.
•Connect the common test lead before
the live test lead and remove the live test
lead before the common test lead.
•Use only current probes, test leads, and
adapters supplied with the Product.
•Do not use the 15-20 A adapter to supply
power to devices rated more than 15 A.
This can overload the installation.
4
Page 19
Electrical Safety Analyzer
Unpack the Product
•Replace the mains power cord if the
insulation is damaged or if the insulation
shows signs of wear.
•Comply with local and national safety
codes. Use personal protective
equipment (approved rubber gloves, face
protection, and flame-resistant clothes)
to prevent shock and arc blast injury
where hazardous live conductors are
exposed.
•Do not touch metal parts of the device
under test (DUT) while you do a test.
Some tests apply high voltage and high
current to the DUT with the DUT earth
connection open or closed.
Unpack the Product
Carefully unpack all items from the box and check that
you have these items:
• ESA615
• Getting Started Manual
• Users Manual CD
• Carrying case
• Power cord
• 15 – 20 A Adapter (USA only)
• ESA USA Accessory Kit (USA, Australia, and Israel
only) or ESA EUR Accessory Kit
• Ansur Demo CD
• Null Post Adapter
• 5-to-5 Banana to ECG Adapter (BJ2ECG)
• USB Transfer Cable
5
Page 20
ESA615
Users Manual
Instrument Familiarization
Figure 1 and Table 2 show the front-panel controls and connections of the Product.
6
9
8
7
6
1
2
3
4
5
gtv116.eps
Figure 1. Front-Panel Controls and Connections
Page 21
Electrical Safety Analyzer
Instrument Familiarization
Table 2. Top-Panel Controls and Connections
Item Name Description
Equipment Outlet
1
Configuration Buttons
2 High Voltage Indicator
3 Test Function Buttons Selects the Product test functions.
4 Navigation Buttons Cursor control buttons for navigating menus and lists.
5 Test Button Starts selected tests.
6 Enter Button Sets the highlighted function.
7 Input Jacks Test lead connectors.
8 Nulling Jack Connection to zero test lead resistance.
9 Function Softkeys
Controls the configuration of the equipment outlet. Opens and closes the neutral and
ground connection and reverses the polarity of the neutral and hot connection.
Illuminates when high voltage is applied to the ECG/Applied Parts posts or L1 and
L2 of the Test Receptacle.
Keys F1 through F4 are used to select from a number of selections that show in the
LCD display above each function softkey.
7
Page 22
ESA615
Users Manual
Figure 2 and Table 3 describe the side and top-panel connections of the Product.
RA
R
LLLA
FL
RL
N
V1
C1
3
9
8
2
5
1
4
10
6
7
8
gtv110.eps
Figure 2. Side and Top-Panel Connections
Page 23
Electrical Safety Analyzer
Instrument Familiarization
Table 3. Side and Top-Panel Connections
Item Name Description
1 Equipment Outlet
2 USB A Controller Port For external keyboard or barcode reader.
USB Device Port
3
(Mini B-style connector)
4 Fuse Access Door Equipment outlet fuse access.
5 Tilt Stand Holds the Product in a tilted position.
6 SD Card Slot SD Memory Card access
7 AC Power Switch Turns ac power on and off.
8 Power Input Connector
9 ECG/Applied Parts Jacks
10 Banana Jack to ECG Adapter Adapter to connect ECG snap leads to the Product.
Equipment outlet, specified to the version of the Product, which supplies a DUT
connection.
Digital connection to control the Product from a PC or instrument controller.
A grounded male three-prong (IEC 60320 C19) connector that accepts the
line-power cord.
Connection posts for Device Under Test (DUT) applied parts, such as ECG
leads. Used to test for leakage current through leads and to supply ECG
signals and performance waveforms to a DUT.
9
Page 24
ESA615
Users Manual
How to Hold the Product
When you move the Product, use the handle on the
bottom case to hold it. See Figure .
Figure 3. Product Handle
gtv122.eps
How to Connect to Line Power
Warning
To prevent possible electrical shock, fire, or
personal injury:
•Do not use an extension cord or adapter
plug.
•Make sure the ground conductor in the
mains power cord is connected to a
protective earth ground. Disruption of
the protective earth could put voltage on
the chassis that could cause death.
•Replace the mains power cord if the
insulation is damaged or if the insulation
shows signs of wear.
•Use only the mains power cord and
connector approved for the voltage and
plug configuration in your country and
rated for the Product.
10
Page 25
Electrical Safety Analyzer
How to Connect a DUT to the Product
The Product is intended for use with single-phase,
grounded power. It is not intended for dual, split-phase or
three-phase power configurations. It can be used with a
power system that supplies the correct voltages for
single-phase and is grounded, or is an isolated power
system.
Use the power cord for your country mains supply that is
not more than the voltage or power rating of the product.
Connect the cord into the power input connector and then
to the mains outlet.
How to Connect a DUT to the Product
You can connect a Device Under Test (DUT) a number of
different ways for a full electrical safety test. Figure
shows a DUT connected to the test receptacle, applied
parts posts, and a connection to the enclosure or
protective earth ground of the DUT.
Figure 4. Product Ready for Operation
gtv125.bmp
How to Turn On the Product
Note
To make sure the high voltage indicator works,
look for it to illuminate at the power-up.
Push the power switch found on the left-side panel so the
“I” side of the ac power switch is down. The Product does
a series of self tests and then shows the message in
Figure when the self test has completed successfully.
11
Page 26
ESA615
Users Manual
Connect ESA615
to grounded
mains socket.
To protective earth or
any exposed conductive
surface on the enclosure
RA
R
LL
F
LARLV1
L
NC1
Connect the DUT ac power cord to the
equipment outlet on the Analyzer
12
Figure 5. DUT Connections to the Product
gtv113.eps
Page 27
Electrical Safety Analyzer
How to Access the Product Functions
The self-test measures the ac mains input for correct
polarity, ground integrity and voltage level. The high
voltage indicator illuminates briefly during the self test. If
the polarity is reversed, the Product shows this condition
and sets the polarity to be reversed internally. If the
ground is open, the Product shows this fault. If the mains
voltage is too high or too low, the Product shows this fault
and does not continue until the supply voltage is
corrected and the Product power turned off and then on
again.
How to Access the Product Functions
For each test and setup function, the Product uses a
series of menus to access different Product tests and
setup variables. The example shown in Figure , the
Product shows different leakage current tests along the
bottom of the display. A More softkey lets you access
more menus related to the test. When you push a softkey
(F1 through F4) below a test name, the Product sets up
for or does the selected test.
For some tests, it will be necessary to set parameters
with the navigation buttons. In the example above, the
leakage parameter has next to it. This icon shows you
must push or to set its value. In this example, the
leakage current measurement changes between AC+DC,
AC only, or DC only. The applied parts indicator has on
the left end and on the right end. These icons show that
you must push and to set an applied part.
13
Page 28
ESA615
Users Manual
gtv102.bmp
Figure 6. Leakage Current Menu
The three buttons along the right side of the display
() control the wiring of the Product’s
test receptacle for some electrical tests. The current
condition of these three buttons is shown along the right
edge of the display when these controls are active.
Figure 5 shows polarity is settable between normal,
reversed, and off. Neutral is also settable to closed or
open. Earth condition is not shown, which means it can
not be changed. Earth is internally opened while the
Product does this test.
How to Connect a PC to the Product
To connect the Product to a PC:
Connect a USB port on your PC or laptop to the Mini B
USB device port on the Product.
Or
Plug in an XStick wireless USB dongle to your PC USB
port. Products available over wireless will be listed by
serial number. Connect to a single Product. See Figure 7.
14
Page 29
Electrical Safety Analyzer
RA LLLARLV1
RFLNC1
How to Connect a PC to the Product
RA LL LA RL
RF
L
N
V1
C1
gtv129.bmp
Figure 7. Product to PC Connection
15
Page 30
ESA615
Users Manual
How to Set Up the Product
There are a number of Product parameters that are
adjusted through the setup function. To access the first
Setup menu shown in Figure , push .
gtv124.bmp
Figure 8. Setup Menu
Set the Operator Name
You can associate an operators name with electrical test
results. To set the operator name:
1. From the setup menu, push or until the
Operator name is highlighted.
2. Push .
An operator list shows in the display. See Figure 9.
The Default User name “None” is used when an
actual user name is not set.
gtv123.jpg
Figure 9. Operator List Screen
Set the Operator Name from a List
1. From the Operator list, push or to highlight one
of the operator names.
2. Push .
16
Page 31
Electrical Safety Analyzer
How to Set Up the Product
Set a New Operator Name
1. From the Operator list, push the New softkey.
Note
The Product keeps a maximum of 20 operator
names. If you push the New softkey and there
are already 20 names, the Product will show an
error. You will have to delete one or more names
to add a new name.
2. In the keyboard screen, push , , , or to move
the highlight to a character.
3. Push to add the highlighted character to the
name field.
Note
You can push the F2 softkey to set the keyboard
between upper and lower case letters. Highlight
айоцз and push to toggle between accent
characters and standard alpha-numeric letters.
4. Do steps 2 and 3 again until the operator name is
complete.
5. Push the Done softkey.
Note
Push the Backspace softkey to delete the last
character in the name field.
Delete an Operator Name
To delete an operator name:
1. From the setup menu, push or until the
Operator name is highlighted.
2. Push .
3. From the Operator list, push or to highlight one
of the operator names.
4. Push the Delete softkey.
5. A delete confirmation screen will show in the display.
6. Push the Delete softkey.
17
Page 32
ESA615
Users Manual
Set the Date
From the setup menu, push or until the Date value
is highlighted.
1. Push .
2. In the keyboard screen, push or to move the
highlight to a character.
3. Push .
4. Do steps 2 and 3 again until the date is complete.
5. Push the Done softkey.
Note
Push the Backspace softkey to delete the last
character in the date field.
Set the Time
1. From the setup menu, push or until the Time
value is highlighted.
2. Push .
3. In the Keyboard screen, push or to move the
highlight to a character.
4. Push .
5. Do steps 3 and 4 again until the time is complete.
Note
When you set the time for 12-hour format, push
the am/pm softkey to set AM or PM.
6. Push the Done softkey.
Note
Push the Backspace softkey to delete the last
character in the date field.
Set the Test Standard
1. From the setup menu, push or until the
Standard variable is highlighted.
2. Push .
3. Push or to highlight one of the standards.
4. Push .
Set the GFCI Limit
To set the GFCI current limit:
5. From the setup menu, push or until the GFCI
Limit variable is highlighted.
6. Push .
7. Push or to highlight one of the preset GFI
current values.
8. Push .
18
Page 33
Electrical Safety Analyzer
How to Set Up the Product
Set Polarity Switching Delay
When the test receptacle of the Product is switched, a
delay can be set to control the actual switch time. To set
the polarity delay:
1. From the setup menu, push or until the
Polarity Delay variable is highlighted.
2. Push .
3. Push or to highlight one of the preset delay
values.
4. Push .
Set the Date Format
1. From the setup menu, push the Instrument Setup
softkey.
2. Push or until the Date Format variable is
highlighted.
3. Push .
4. Push or to highlight DD/MM/YYYY, MM/DD/YYYY, or YYYY/MM/DD.
5. Push .
Set the Time Format
1. From the setup menu, push the Instrument Setup
softkey.
2. Push or until the Time Format variable is
highlighted.
3. Push .
4. Push or to highlight 12 Hour or 24 Hour.
5. Push .
Set the Language
The Product can display data in English, French,
German, Spanish, Italian, or Portuguese. To change the
language:
1. From the setup menu, push the Instrument Setup
softkey.
2. Push or until the Language variable is
highlighted.
3. Push .
4. Push or to highlight one of the languages.
5. Push .
19
Page 34
ESA615
Users Manual
Set the Beeper
To enable or disable the beeper:
1. From the setup menu, push the Instrument Setup
softkey.
2. Push or until the Beeper variable is
highlighted.
3. Push .
4. Push or to highlight Off or On.
5. Push .
Set the Display Contrast
There are two procedures to set the display contrast.
From the Test Sequence start-up menu or through the
setup menu.
When the Product shows its start-up menu (Select a
test…), push or to increase or decrease the display
contrast respectively. Push the Done softkey to exit
contrast setup.
To adjust the contrast through the setup menu:
1. From the setup menu, push the Instrument Setup
softkey.
2. Push the Display Contrast softkey.
3. Push or to increase or decrease the display
contrast respectively.
4. Push the Done softkey to exit contrast setup.
How to View Instrument Information
To show model number, serial number, firmware version,
and last calibration date of the Product, push . Next
push the Instrument Information softkey.
How to Do Electrical Safety Tests
The Product does a number of different electrical and
performance tests on biomedical equipment. The sections
that follow are test descriptions with instructions on how
to do them with the Product.
20
Page 35
Electrical Safety Analyzer
How to Do Electrical Safety Tests
Set the Test Standard
The electrical safety tests in the Product are specified by
different safety standards: AAMI ES1/NFPA99,
IEC62353, IEC60601-1, and AN/NZS 3551. AAMI is set
as the default standard. To select different standard:
1. From the setup menu, push until the Standard
variable is highlighted.
2. Push .
3. Push or to highlight one of the standards.
4. Push .
Not all electrical tests are applicable for all standards. In
these cases, the menu only shows the tests that are
specified in the set standard.
Note
The standard set in the setup menu applies to all
manual measurement modes. The automated
tests use the standard set for the selected test
sequence.
Figure 10. Mains Voltage Test Menu
Push each function softkey to do each of the three
measurements: Live to neutral, neutral to ground, and live
to ground.
Note
Power is removed from the test receptacle while
the Product does a Mains Voltage test.
gtv104.bmp
Mains Voltage Test
The Mains Voltage test measures the voltage on the
mains input through three measurements. To access the
Mains Voltage test, push . The Mains Voltage test
menu is shown in Figure .
21
Page 36
ESA615
Users Manual
Ground Wire (Protective Earth) Resistance Test
The Ground Wire (Protective Earth) Resistance test
measures the impedance between the PE terminal of the
test receptacle and the exposed conductive parts of the
DUT that are connected to the Protective Earth of the
DUT.
Note
Before you do leakage tests with the Product, it
is best to make sure the ground connection is
good.
Do this test between the test receptacle ground and the
Protective earth ground of the DUT or DUT enclosure.
To access the Ground Wire (Protective Earth) ∅/Null
Resistance Test menu push .
Note
The DUT is turned off for this test.
To do a ground wire resistance test:
1. Make sure the power cord from the DUT is
connected into the test receptacle.
2. Push to reveal the resistance function menu.
3. Connect one end of a test lead to the V/Ω/A jack.
4. If you use an accessories probe, connect it to the
other end of the test lead and put the probe tip into
the ∅/Null jack. If you use an alligator clip accessory,
connect it to the other end of the test lead, put the
null post adapter in the ∅/Null jack, and clamp the
alligator clip to the null post adapter.
Note
∅
/Null jack does not accept the test leads
The
supplied with the Product.
5. Push the Zero Leads softkey. The Product zeroes
out the measurement to cancel the test lead
resistance.
6. Connect the test lead from the V/Ω/A jack to the DUT
enclosure or protective earth connection.
22
Page 37
Electrical Safety Analyzer
How to Do Electrical Safety Tests
7. After you make the connections to the DUT, the
measured resistance shows in the display. See
Figure .
gtv105.jpg
Figure 11. DUT Ground Resistance Measurement
Warning
To prevent electrical shock, remove the null
post adapter from the ∅/Null jack after a test
lead zero is performed. The ∅/Null jack
becomes potentially hazardous during some
of the test conditions.
A low resistance measurement is necessary to make sure
there is a good ground connection through the power
cord. Refer to the applicable electrical safety standard for
the specified limit value to be followed.
Figure shows the electrical connections between the
Product and the DUT. Table 4 is a list of the abbreviations
used in the schematics and their descriptions.
The five Insulation Resistance Tests measure mains (L1 &
L2) to Protective earth, applied parts to Protective earth,
mains to applied parts, mains to non-earthed conductive
points, and applied parts to non-earthed conductive points.
To access the Insulation Resistance Test menu,
push .
All Insulation Resistance Tests can be done with 500 V dc
or 250 V dc. To change the test voltage from the Insulation
Resistance Test menu, push the More softkey. Push the
Change Voltage softkey to toggle the test voltage
between 250 V dc and 500 V dc.
Note
When you exit and re-enter the Insulation
Resistance Test menu, the test voltage is set to
its default value of 500 V dc.
As shown in Figure , three of the five tests are shown
above function softkeys F1 through F3. To access the
other two tests or test voltage selection, push the More
softkey. The Back softkey will move the menu up to the
top-level Insulation Resistance test menu.
Figure 14. Insulation Resistance Measurement
After you push a test softkey, push to apply the test
voltage to the DUT and take the resistance measurement.
Figures through shows the electrical connections
between the Product and DUT for the five Insulation
Resistance tests.
Note
The DUT is powered off for this test.
gtv106.jpg
27
Page 42
ESA615
Users Manual
DEVICE UNDER TEST
APPLIED
PART
FE
CONDUCTIVE PART
MD
DUT_L1
DUT_L2
DUT_PE
Figure 15. Mains to Protective-Earth Insulation Resistance Test Schematic
faw17.eps
28
Page 43
Electrical Safety Analyzer
How to Do Electrical Safety Tests
DUT_L1
DUT_L2
DUT_PE
MD
DEVICE UNDER TEST
APPLIED
PART
FE
CONDUCTIVE PART
Figure 16. Applied Parts to Protective-Earth Insulation Test Schematic
faw18.eps
29
Page 44
ESA615
Users Manual
MD
DUT_L1
DUT_L2
DUT_PE
DEVICE UNDER TEST
APPLIED
PART
FE
CONDUCTIVE PART
Figure 17. Mains to Applied-Parts Insulation Test Schematic
faw19.eps
30
Page 45
Electrical Safety Analyzer
How to Do Electrical Safety Tests
DUT_L1
DUT_L2
MD
DUT_PE
DEVICE UNDER TEST
APPLIED
PART
FE
CONDUCTIVE PART
TEST
LEAD
Figure 18. Mains to Non-Earth Accessible Conductive Points Schematic
faw20.eps
31
Page 46
ESA615
Users Manual
DUT_L1
DUT_L2
DUT_PE
DEVICE UNDER TEST
APPLIED
PART
FE
CONDUCTIVE PART
TEST
LEAD
MD
Figure 19. Applied Parts to Non-Earth Conductive Points Schematic
faw21.eps
32
Page 47
Electrical Safety Analyzer
How to Do Electrical Safety Tests
Equipment Current Test
To measure the current consumed by the DUT, push
. The Product shows the current that flows through
the mains connections of the test receptacle.
Leakage Current Test
The Product measures leakage current for different DUT
configurations. The Product measures leakage found on
the enclosure and the earth connection as well as leakage
Touch or Enclosure Leakage Current Chassis Leakage Current
Patient Leakage Current Lead to Ground Leakage Current
Patient Auxiliary Leakage Current Lead to Lead Leakage Current
Mains on Applied Part (MAP) Leakage Current Isolation Leakage Current
on each connected applied part and combinations of
connected applied parts.
The available leakage tests are set by the standard you
set in setup. See the How to Set the Test Standard section
to change the test standard.
Table is a list of six leakage current tests. Their names
are different when the standard is changed in the Product.
Push to access the leakage current main menu
shown in Figure .
33
Page 48
ESA615
Users Manual
gtv102.bmp
Figure 20. Leakage Current Main Menu
Note
The display shown in Figure is the main leakage
current menu when AAMI is the standard.
All leakage currents but Lead Isolation (Mains on Applied
parts), are shown in as AC+DC, AC Only, or DC only. The
initial result is shown in the applicable parameter for the
set test standard. To change the parameter, push or .
The measurement method is shown to the right of the
current measurement while leakage current tests are
done.
Earth Leakage Current
Note
The Ground Wire (Earth) Leakage test is
available for AAMI, 60601, and not IEC 62353.
To measure the current that flows in the protective earth
circuit of the DUT, push the Ground Wire softkey from the
leakage current main menu. Figure shows the electrical
connections between the Product and the DUT when you
do a Ground Wire Leakage Current Test.
The Ground Wire Leakage Current test has some
combination measurements that can be done. Push
to switch the polarity of the mains voltage applied
to the test receptacle between Normal, Off, Reverse, and
Off. Push to open and close the neutral connection
to the test receptacle. It is not necessary to open up the
test receptacle earth (ground), since this is done internally
by the measurement.
34
Page 49
Electrical Safety Analyzer
How to Do Electrical Safety Tests
The outlet conditions below apply when you do this test:
• Normal Polarity
• Normal Polarity, Open Neutral
• Reversed Polarity
• Reversed Polarity, Open Neutral
IEC60601-1 specifies that the applied parts must be
connected for this measurement. Push or to ground
and unground all applied parts connection posts.
Note
A ground connection on an applied part is shown
as a box around the applied part in the display.
35
Page 50
ESA615
Users Manual
DEVICE UNDER TEST
CONDUCTIVE PART
L1
MAINS
L2
MAINS
ON
PE
L2
MD
APPLIED PART
DUT_L1
DUT_L2
REV
POL
DUT_PE
Figure 21. Earth Leakage Current Test Schematic
Note
Ground wire leakage is the same schematic without the Applied Parts switch.
APPLIED
PART
FE
faw27.eps
36
Page 51
Electrical Safety Analyzer
How to Do Electrical Safety Tests
Chassis (Enclosure) Leakage Test
Note
The Chassis (Enclosure) Leakage test is only
available for the IEC60601 or ANSI/AAMI ES1
1993 standard selections.
The Chassis (Enclosure) Leakage Test measures the
current that flows between the enclosure of the DUT and
protective earth. Figure shows the electrical connections
between the Product and the DUT.
To do a Chassis (Enclosure) Leakage Test:
1. Connect a lead between the V/Ω/A jack and the DUT
enclosure.
2. Push .
3. Push the Chassis softkey from the Leakage Current
Test menu.
4. The measured current shows in the display.
The Chassis Leakage test can be done with different fault
conditions on the test receptacle. Push to switch
the test receptacle between Normal, Off, Reverse, and
Off. Push to open and close the neutral connection
to the receptacle. Push to open and close the earth
connection of the receptacle.
The outlet conditions below apply when do this test:
• Normal Polarity
• Normal Polarity, Open Earth
• Normal Polarity, Open Neutral
• Reversed Polarity
• Reversed Polarity, Open Earth
• Reversed Polarity, Open Neutral
IEC60601-1 specifies that the applied parts must be
connected for this measurement. Push or to ground
and unground all applied parts connection posts.
Note
Chassis leakage for ANSI/AAMI is the same schematic
without the Applied Parts switch.
37
Page 52
ESA615
Users Manual
L1
MAINS
L2
MAINS
ON
PE
DEVICE UNDER TEST
CONDUCTIVE PART
L2
EARTH
MD
APPLIED PART
DUT_L1
DUT_L2
DUT_PE
Figure 22. Enclosure Leakage Current Test Schematic
APPLIED
PART
TEST LEAD
FE
faw28.eps
38
Page 53
Electrical Safety Analyzer
How to Do Electrical Safety Tests
Lead-to-Ground (Patient) Leakage Test
Note
The Lead-to-Ground (Patient) Leakage Current
Test is not available for IEC 62353 standard
selections.
The Lead-to-Ground (Patient) Leakage Current test
measures the current that flows between in one applied
part, a group of applied parts, or all applied parts, and the
Mains PE. Figure shows the electrical connections
between the Product and the DUT.
To do a Lead-to Ground (Patient) leakage test:
1. Push .
2. Push the More softkey.
3. Push or to set one of the applied part groupings.
Note
Refer to the test standard when you know the
type of the applied parts and how they must be
grouped for the test.
4. Push the Select softkey.
5. Push or to advance through each applied part
grouping, or the individual applied parts, to ground.
These are selected and measured.
The Lead-to-Ground Leakage test can be done with
different fault conditions on the test receptacle. Push
to switch the test receptacle between Normal, Off,
Reverse, and Off. Push to open and close the
neutral connection to the receptacle. Push to open
and close the earth connection in the receptacle.
The outlet conditions below apply when you do this test:
• Normal Polarity
• Normal Polarity, Open Neutral
• Normal Polarity, Open Earth
• Reversed Polarity
• Reversed Polarity, Open Neutral
• Reversed Polarity, Open Earth
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
39
Page 54
ESA615
Users Manual
L1
MAINS
L2
MAINS
ON
DEVICE UNDER TEST
APPLIED
PART
CONDUCTIVE PART
SELECT
RELAY*
PE
LEAD GND
SELECT RELAY
(REMOTE ONLY)
LEAD
PE
L2
EARTH
DUT_L1
DUT_L2
REV
POL
DUT_PE
MD
Figure 23. Lead-to-Ground (Patient) Leakage Current Test Schematic
FE
*Leads not selected are open.
gtv29.eps
40
Page 55
Electrical Safety Analyzer
How to Do Electrical Safety Tests
Lead-to-Lead (Patient Auxiliary) Leakage Tests
Note
The Lead-to-Lead (Patient Auxiliary) leakage test
is available when the IEC60601 or ANSI/AAMI
ES1-1993 standard is selected.
To measure the leakage current through each applied part
or lead and combination of lead connections (all other or
between two), push the Lead to Lead softkey from the
Leakage Test main menu shown in Figure . Figure shows
the electrical connections between the Product and the
DUT when it does a Lead-to-Lead (Patient Auxiliary)
Leakage Current Test.
The Lead-to-Lead (Patient Auxiliary) Leakage test adds a
diagram of the applied parts connection posts to the
display, as shown in Figure . In the figure, the applied
parts post RA/R is shown above the other posts. This
shows that the leakage measurement is from RA/R to all
others. To move to the subsequent applied part post, push
. The first post will show inline with the other posts while
the LL/F post shows above all others. This shows the
second leakage measurement is from LL/F to all others.
Continue to push or to move from one connection post
to another.
After each post is isolated individually, the Lead-to-Lead
(Patient Auxiliary) Leakage test measures current of three
different combinations of posts tied together: RA/R and
LL/F, RA/R and LA/L, or LL/F and LA/L.
fis107.eps
Figure 24. Applied Parts Connection Posts Display
The Lead-to-Lead (Patient Auxiliary) Leakage test can
make different fault measurements. Push to switch
the polarity of the mains voltage applied to the test
receptacle between Normal, Off, Reverse, and Off. Push
to open and close the neutral connection to the test
receptacle. Push to open and close the earth or
ground connection to the test receptacle.
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
41
Page 56
ESA615
Users Manual
L1
MAINS
L2
MAINS
ON
PE
L2
EARTH
DUT_L1
DUT_L2
REV
POL
DUT_PE
DEVICE UNDER TEST
CONDUCTIVE PART
MD
PE
APPLIED
PART
+
LEAD
SELECT
RELAY*
- LEAD
SELECT
RELAY*
LEAD GND
SELECT RELAY*
(REMOTE ONLY)
FE
Figure 25. Lead-to-Lead (Patient Auxiliary) Leakage Current Test Schematic
*Leads not selected are open.
gtv30.eps
42
Page 57
Electrical Safety Analyzer
How to Do Electrical Safety Tests
The outlet conditions below apply when you do this test:
• Normal Polarity
• Normal Polarity, Open Neutral
• Normal Polarity, Open Earth
• Reversed Polarity, Open Neutral
• Reversed Polarity, Open Earth
Lead Isolation (Mains on Applied Part) MAP
Leakage Test
Note
The Lead Isolation (Mains on Applied Part)
leakage test is available when the IEC60601 &
ANSI/AAMI standard is selected.
The Lead Isolation (Mains On Applied Parts) Leakage
Current test measures the current that flows in response
to an isolated AC voltage applied between a selected
applied part, group of applied parts, or all applied parts,
and Earth (and any conductive part connected to the RED
terminal). Figure shows the electrical connections
between the Product and the DUT when it does a Mains
on Applied Part Leakage Current Test.
Note
With 60601 standard selected, the MAP test
voltage is available in Normal and Reverse (180
degrees out of phase with mains).
To do a Lead Isolation (Mains on Applied Part) test:
1. Push .
2. Push the More softkey.
3. Set the applied part groupings with and .
Note
Refer to the test standard when you decide the
type of the applied parts and how they must be
grouped for the test.
4. Push the Select softkey.
5. Push the Lead Isolation softkey.
6. Push or to set the desired applied part connection.
7. Push to apply the voltage and read the leakage
current in the display.
43
Page 58
ESA615
Users Manual
Push and to scroll through the applied part
connections or groupings. Push for each connection
configuration to thoroughly test the DUT.
The outlet conditions below apply when you do this test:
• Normal Polarity
• Reverse Polarity
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
44
Page 59
Electrical Safety Analyzer
How to Do Electrical Safety Tests
L1
MAINS
L2
MAINS
On
PE
MAP (ISOLATION)
TRANSFORMER
DEVICE UNDER TEST
CONDUCTIVE PART
TEST
LEAD
MD
APPLIED
PART
LEAD
SELECT
RELAY*
FE
*Leads not selected are open.
REV
POL
MAP
REV
DUT_L1
DUT_L2
DUT_PE
Figure 26. Lead Isolation (Mains On Applied Parts) Leakage Test Schematic
gtv31.eps
45
Page 60
ESA615
Users Manual
Alternative Equipment Leakage Test
Note
The alternative equipment leakage test is
available when the EN62353 standard is
selected.
The Alternative Equipment Leakage test, applies the
voltage source between short-circuited equipment outlet
mains live, neutral, and equipment outlet earth, the
exposed conductive surface on the housing, and all
applied parts short-circuited together. The test disconnects
equipment from mains. The current which flows over the
insulation of the DUT is measured.
This test is not applicable for equipment with internal
electrical power source. The switches in mains part must
be closed for this measurement.
To do an alternative equipment leakage test:
1. Push .
2. Push the Alternative Equipment softkey.
3. Push to apply the voltage and read the current
in the display.
Figure shows the electrical connections between the
Product and the DUT for the Alternative Equipment
Leakage Test.
The outlet conditions below apply when you do this test:
• Closed Earth
• Open Earth
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
Alternative Applied Part Leakage Test
Note
The Alternative applied part leakage test is
available when the EN62353 standard is
selected.
The Alternative Applied Part Leakage test applies the test
voltage between short-circuited applied parts of a single
function and the short-circuited equipment outlet mains
live, neutral, earth, and exposed conductive surface on the
housing. This test should only be done for equipment with
F-Type applied parts. For equipment with multiple applied
parts, test each group of applied parts of a single function
in turn with all others float during the test. All applied parts
can be connected to the applied parts jacks and the lead
selection will float those not selected.
46
Page 61
Electrical Safety Analyzer
How to Do Electrical Safety Tests
DUT_L1
DUT_L2
DUT_PE
MD
DEVICE UNDER TEST
APPLIED
PART
FE
CONDUCTIVE PART
TEST
LEAD
Figure 27. Alternative Equipment Leakage Current Test Schematic
faw22.eps
47
Page 62
ESA615
Users Manual
To do an alternative applied part leakage test:
1. Push .
2. Push the More softkey.
3. Set the applied part groupings with and .
4. Push the Select softkey.
5. Push the Alternative A.P. softkey.
6. Push to apply the test voltage and read the
current in the display.
7. Push or to advance to the next applied part
group(s) of a single function if applicable. Push
to read leakage current for each group.
Figure shows the electrical connections between the
Product and the DUT for an Alternative Applied Part
Leakage current test.
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
Direct Equipment Leakage Test
Note
The Direct Equipment Leakage test is available
when the EN62353 standard is selected.
The Direct Equipment Leakage Current test measures the
leakage current between all applied parts and the exposed
conductive surface on the housing to mains earth.
To do a direct equipment test:
1. Push .
The direct equipment test is the default test and must
already be selected.
2. Push to apply the voltage and read the leakage
current in the display
Figure shows the electrical connections between the
Product and the DUT for a Direct Equipment Leakage
Current Test.
The outlet conditions below apply when you do this test:
• Normal Polarity, Closed Earth
• Normal Polarity, Open Earth
• Reversed Polarity, Closed Earth
• Reversed Polarity, Open Earth
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
48
Page 63
Electrical Safety Analyzer
How to Do Electrical Safety Tests
DUT_L1
DUT_L2
DUT_PE
DEVICE UNDER TEST
APPLIED
PART
CONDUCTIVE PART
TEST
LEAD
LEAD
SELECT
RELAY*
*Leads not selected
are open.
FE
Figure 28. Alternative Applied Part Leakage Test Schematic
MD
gtv23.eps
49
Page 64
ESA615
Users Manual
Direct Applied Part Leakage Test
Note
The Direct Applied Part Leakage test is available
when the EN62353 standard is selected.
The Direct Applied Part Leakage Current test measures
the leakage current between all applied parts of one
function and the exposed conductive surface on the
housing, to mains earth. For equipment with multiple
applied parts, you must do the test on each group of one
function each in turn while all others float. This test must
only be done for equipment with F-Type applied parts.
For Type B applied part, see direct equipment leakage
schematic in Figure .
To do a direct applied part leakage test:
1. Push .
2. Push the More softkey.
3. Set the desired applied part groupings with and .
4. Push the Select softkey. The Direct A.P. test must
already be set.
5. Push or to set the applied part test configuration.
6. Push to apply the test voltage and read the
current in the display.
7. Push or to advance to the next group of applied
parts, if applicable.
Figure shows the electrical connections between the
Product and the DUT for a Direct Applied Part Leakage
Current Test.
The outlet conditions below apply when you do this test:
• Normal Polarity
• Reversed Polarity
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
50
Page 65
Electrical Safety Analyzer
How to Do Electrical Safety Tests
L1
MAINS
L2
MAINS
ON
PE
DEVICE UNDER TEST
CONDUCTIVE PART
TEST
LEAD
L2
MD
REV
POL
PE
DUT_L1
DUT_L2
DUT_PE
Figure 29. Direct Equipment Leakage Test Schematic
APPLIED
PART
FE
faw24.eps
51
Page 66
ESA615
Users Manual
L1
MAINS
L2
MAINS
ON
PE
DEVICE UNDER TEST
APPLIED
PART
CONDUCTIVE PART
L2
DUT_L1
DUT_L2
REV
POL
DUT_PE
Figure 30. Direct Applied Parts Leakage Current Test Schematic
LEAD
SELECT
RELAY*
*Leads not selected
are open.
FE
MD
gtv25.eps
52
Page 67
Electrical Safety Analyzer
How to Do Electrical Safety Tests
Differential Leakage Current Test
Note
The Differential Leakage Current test is available
when the EN62353 standard is selected.
The differential leakage current test measures the
magnitudes of the differential current that flows in the
Equipment Outlet live and neutral, with power applied to
the equipment outlet. All applied parts must be connected
for this test, if equipment has applicable applied parts.
To do a differential leakage current test:
1. Push .
2. Push the Differential softkey.
Figure shows the electrical connections between the
Product and the DUT for a Differential Leakage Current
test.
The outlet conditions below apply when you do this test:
• Normal Polarity, Closed Earth
• Normal Polarity, Open Earth
• Reversed Polarity, Closed Earth
• Reversed Polarity, Open Earth
Note
If there are more than five applied parts to
connect to the Product, see How to Use the 1-to10 Adapter later in this manual.
53
Page 68
ESA615
Users Manual
L1
MAINS
L2
MAINS
ON
PE
DEVICE UNDER TEST
APPLIED
PART
CONDUCTIVE PART
TEST
LEAD
REV
POL
EARTH
DUT_L1
MD
DUT_L2
DUT_PE
Figure 31. Differential Leakage Current Test Schematic
FE
gtv32.eps
54
Page 69
Electrical Safety Analyzer
How to Use the 1-to-10 Adapter
How to Use the 1-to-10 Adapter
The 1-to-10 Adapter, an optional accessory, increases the
number of lead or applied parts connections to the Product
from 5 to 14. The adapter connects a maximum of 10
leads together into a one lead that is connected to one of
the input jacks of the Product. The other four Product input
jacks can also be used in conjunction with the Adapter.
More leads can be added with multiple 1-to-10 Adapters.
The example in Figure shows one application of the
Adapter. The Defibrillator/Monitor in the example has 10
ECG leads, two pacer leads, and two defibrillator paddles
which must be connected together, and in groups if single
function, for current leakage per IEC62353. The example
shows the ECG leads to be snap type connectors and two
BJ2ECG adapters are shown connected to the Adapter. If
the ECG leads do not have snap connectors, then the
Universal Snap to Banana Adapter can be used to make
the connections to the Adapter.
The common lead from the Adapter is plugged into the RA
st
jack (1
leads with alligator clips, connect the two defibrillator
paddles into the LL and LA Product jacks and the two
pacer leads into the RL and V1 jacks. Set the connection
that ties all five Product jacks together. This will measure
leakage current in all fourteen leads. The applied part
group of 1, 2, and 2 lets you test groups of applied parts of
one function.
jack) of the Product. Using four sheathed test
55
Page 70
ESA615
Users Manual
Defibrillator/Pacer
Use leads shorted
for electrical safety
testing only. Do not
short leads for ECG
simulation.
Caution: To prevent damage to the Product,
do not apply more than 30 V dc to
DEFIB
OFF
PACER
the AP (ECG).
Check unkown connections using a
ESA615
DMM before connecting to ESA615.
1-to-10 and
Alligator Clips
BJ2ECG
snap adapters
gtv120.eps
Figure 32. 1-to-10 Adapter Connections
56
Page 71
Electrical Safety Analyzer
How to Use the 1-to-10 Adapter
When you do an applied parts test with the AAMI/NFPA99 standard, the normal connections of RA, LL, LA, and
RL are made to their related input jacks. Four adapters
from the Universal Snap to Banana Adapter set will be
necessary for the first four connections. The other chest
leads are connected to the Adapter and the common lead
from the Adapter is connected to the V1 jack (5
the Product. See Figure . This configuration lets you
isolate the RA, LL, LA, and RL leads from each other and
the other chest leads, which are shorted together, while
the Product does the leakage tests.
th
jack) of
57
Page 72
ESA615
Users Manual
ECG Monitor
Use leads shorted
for electrical safety
testing only. Do not
short leads for ECG
simulation.
1-to-10 and
BJ2ECG
snap adapters
Figure 33. ECG Lead Connection with 1-to-10 Adapter
ESA615
gtv121.eps
58
Page 73
Electrical Safety Analyzer
How to Do Point-To-Point Measurements
How to Do Point-To-Point Measurements
The Product can make voltage, resistance, and low
current measurements through its Point-to-Point function.
To access the Point-to-Point function menu shown in
Figure , push . Softkeys F1 through F3 are used to
set the measurement function.
gtv128.jpg
Figure 34. Point-To-Point Function Menu
Measure Voltage
To make a voltage measurement:
1. Push the Voltage softkey from the Point-To-Point
menu.
2. Put test leads in the RED (V/Ω/A ) and BLACK jacks.
3. Put the probe tips across the unknown voltage and
read the measurement in the display.
The Product measures a maximum of 300 volts ac.
Measure Resistance
To make a resistance measurement:
1. Push the Resistance softkey from the Point-To-Point
menu.
2. Put test leads in the RED (V/Ω/A ) and BLACK jacks.
3. Null lead resistance by shorting the leads together
and Push the Zero Leads softkey.
4. Put the probes across the unknown resistance and
read the measurement in the display.
The Product measures resistances to a maximum of
2.0 Ω.
59
Page 74
ESA615
Users Manual
Measure Current
The Product can make dc only, ac only, and ac+dc current
measurements to a maximum of 10 mA. To do a current
measurement:
1. Push the Leakage softkey from the Point-To-Point
menu.
2. Push or set the ac only, dc only, or ac+dc
measurement mode.
3. Put test leads in the RED (V/Ω/A ) and BLACK jacks.
4. Put the probe tips on the two points the unknown
current may flow and read the measurement in the
display.
How to Simulate ECG Waveforms
The Product can put different waveforms on the applied
parts connection posts. These signals are used to
measure the performance parameters of ECG monitors
and ECG strip printers. See Figure for the connections
between the Product and an ECG monitor. For monitors
that use the snap style connectors, put the BJ2ECG
adapter into the connectors at the top of the Product and
connect the monitor leads to the snap connectors on the
adapter.
Note
If the ECG monitor/interpreter has banana posts,
use the optional universal snap to banana
adapter to connect to the Product.
To access the ECG Simulation Waveform menu shown in
Figure , push . From this menu, a number of
waveforms are set through F1, and the rate or frequency
of the waveform is set through F2.
gtv109.bmp
Figure 35. ECG Waveform Simulation Menu
To output one of the predefined waveforms, push the
Wave Form softkey. A scroll box with next to it appears
above the softkey label. Use or to scroll through the
different waveforms.
60
Page 75
Electrical Safety Analyzer
How to Simulate ECG Waveforms
For all waveforms but VFIB and Triangle, the rate or
frequency of the waveform is adjusted through the
Frequency or Rate softkey. For some waveforms, there
are more than two frequency or rate selections. For those
waveforms, push the Frequency or Rate softkey to open
a scroll box above the softkey label with next to it. Use
or to select the frequency or rate. For those
waveforms that have only two frequencies or rates, the
Frequency or Rate softkey is a toggle, where each push
of the softkey switches to the other value.
61
Page 76
ESA615
Users Manual
ECG Monitor
RALL
RF
LA
L
RLV1
NC1
R
RA
FLN
LLLARL
C1
V1
BJ2ECG
Adapter
62
Figure 36. ECG Monitor Connections
gtv115.eps
Page 77
Electrical Safety Analyzer
Memory
Memory
The Product keeps test results data and test sequences
on an SD memory card. The memory card holds a
minimum of 100 test sequences and 1000 test results.
Each test result can be recalled to the display of the
Product or exported to a PC.
Note
The Product can show the last 200 test results.
All results in the Product can be exported to a
PC.
To remove the memory card:
1. Push in on the memory card and release it.
2. The memory card will eject out of the slot.
3. Hold the memory card with your fingers and remove it
from the product.
Note
With the memory card removed, no test
sequences will show in the test library list. You
cannot make new test sequence without the card
installed in the Product.
To install the memory card:
1. Insert the card with the contacts toward the back.
2. Push the card all the way in until you hear it click.
3. Release the card.
Test Sequences
The test sequence feature automates the tests you do on
a DUT. You make test sequences with the built-in test
sequence wizard. A different procedure is to start with a
test sequence that is already in the test library and change
it to make a new test sequence. The test sequences and
test results are kept on the memory card.
Factory Supplied Test Sequences
Table is a list of factory supplied test sequences that are
on the memory card that ships with the Product. Each
factory supplied test sequence is prefixed with the test
standard number. For example, the 60601-1 Monitor test
sequence is based on the 60601-1 test standard.
63
Page 78
ESA615
Users Manual
Table 6. Factory Supplied Test Sequences
Test Sequence Description
60601 3rd Patient Monitor 3rd Ed., Class I, 5 ECG
60601 3rd Defibrillator 3rd Ed., Class I, 2 Paddles and 3 ECG
60601 3rd Infusion Device 3rd Ed., Class II, 1 No AP
60601 3rd Ultrasound Device 3rd Ed., Class I, 1 Probe
60601 3rd Generic Device 3rd Ed., Class I, No AP
60601 3rd System 3rd Ed., Class I, No AP
62353-Alt. Patient Monitor Class I, 5 ECG
62353-Alt. Defibrillator Class I, 2 Paddles and 3 ECG
62353-Alt. Infusion Device Class II, 1 No-AP
62353-Alt. Ultrasound Device Class I, 1 Probe
62353-Alt. Generic Device Class I, No AP
NFPA99 Patient Monitor Class I, 5 ECG
[1]
NFPA99 Defibrillator Class I, 2 Paddles and 3 ECG
NFPA99 Infusion Device Class II, 1 No AP
64
Page 79
Electrical Safety Analyzer
Test Sequences
Table 6. Factory Supplied Test Sequences (cont.)
Test Sequence Description
NFPA99 Ultrasound Device Class I, 1 Probe
NFPA99 Generic Device Class I, No AP
ANSI/AAMI ES-1 Patient Monitor Class I, 5 ECG
ANSI/AAMI ES-1 Defibrillator Class I, 2 Paddles and 3 ECG
ANSI/AAMI ES-1 Infusion Device Class II, 1 No AP
ANSI/AAMI ES-1 Ultrasound Device Class I, 1 Probe
ANSI/AAMI ES-1 Generic Device Class I, No AP
[1] Class designations refer to applicable electrical safety standard definitions for devices, not FDA medical device definitions.
[1]
65
Page 80
ESA615
Users Manual
How to Make a Test Sequence
You can make a new test sequence from a different test
sequence or make a new one.
Make a New Test Sequence
To make a new test:
1. Push .
2. Push the Test Library softkey.
3. Push the New softkey.
A test sequence wizard steps you through the
configuration of the test sequence. There are five setup
steps.
1. Push or to highlight a test standard and push .
2. Push or to highlight a device class and push .
3. Configure the Applied Parts (A.P.).
Note
A USB keyboard or barcode reader can be used
to type in the A.P. name.
If the DUT does not have applied parts, push the Next
Step softkey to skip the applied parts configuration.
If the DUT has applied parts, push New A.P.. You
must name the applied part and show how many DUT
applied parts are in the configuration.
The Product has five applied parts inputs. If you set
more than five applied parts, an error message shows
in the display. If the DUT has more than five applied
parts, you can use the 1-to-10 adapter. See the How
to Use the 1-to-10 Adapter section to learn more.
If you attach multiple applied parts to one input, then
you must set the Tied Setting to Tied. The Product
sets the subsequent available input for the applied
parts that are tied together. The position icon in the
display shows inputs that are configured as dots and
inputs not configures as circles. Figure shows the
position icon for one input configured and four inputs
not configured.
Figure 37. Input Position ICON
To put a DUT applied part on each of the Product
inputs, set Tied Settings to Not Tied. The Product
will show the subsequent available inputs as
configured for this configuration.
The last applied parts variable is the Type. Highlight
the Type variable and push . Highlight a type
from the type list and push .
gtv127.bmp
66
Page 81
Electrical Safety Analyzer
Test Sequences
Push the Done softkey to complete the applied part
configuration.
To edit an applied part configuration, use or to
highlight the A.P. configuration and push .
4. Edit test settings.
Table 7 is a list of test settings with their descriptions
and default values.
To set one of the test setting, push or to
highlight a test setting and push . After you
change the setting, push the Done softkey.
5. Name the test sequence. When you push the Next Step softkey in the edit test settings step, the Product
automatically names the sequence Test Sequence
followed by the date and time. To accept the default
name, push the Next Step softkey. To change the
name, push the Edit softkey.
Note
A USB keyboard or barcode reader can be used
to edit the test sequence name.
The test sequence is complete. To put the test sequence
on the memory card, push the Save softkey. To see the
test sequence parameters before you save it, push the
Edit softkey.
67
Page 82
ESA615
Users Manual
Table 7. Test Settings for Test Sequences
Test Setting Description
Pause after power on If set to yes, delays the start of the test by the time set in the Power on delay
parameter when power is applied to the DUT.
Pause before power off If set to yes, delays the start of the test by the time set in the Power off delay
parameter when power is removed from the DUT.
Power on delay The time the Product will wait before it does the subsequent step after power is
applied to the DUT. The range is 0 to 9999 seconds.
Power off delay The time the Product will wait before it does the subsequent step after power is
removed from the DUT. The range is 0 to 9999 seconds.
Test Speed
Test Mode When set to Automatic, the Product does each test sequence step automatically.
Halt on Test Failure When set to Yes, the Product stops the test sequence when a fault is sensed. Yes
[1]
When set to Normal, the Product does a leakage current measurement in
5 seconds and an insulation resistance test in 1 minute.
When set to Rapid, the Product measures leakage current as fast as possible and
an insulation resistance test in 3 seconds.
When set to Step-by-Step, you must push the Next Step softkey to move to the
subsequent step in the sequence.
Default
Value
No
No
2 sec
0 sec
Normal
Automatic
68
Page 83
Electrical Safety Analyzer
Test Sequences
Table 7. Test Settings for Test Sequences (cont.)
Test Setting Description
Default
Value
Multiple PE Tests If set to Yes, the Product prompts you to repeat PE tests or continue. No
Multiple Non-Earth
Leakage
Patient Lead Records
storage
[2]
If set to Yes, the Product prompts you to repeat non-earth leakage tests or
continue.
When set to Store all, all results are put into the test results. If set to Store
worst/last, only the worst measured value is put into the test results. If set to store
worst/last and all measured values are the same, then the last value is put into
No
Store
worst/last
the test results.
Insulation Test Voltage Sets the insulation test voltage to 250 V dc or 500 V dc. 500 V dc
Reverse Polarity If set to No, reverse polarity tests will be skipped. Yes
[1] You can change the test speed parameter on a test sequence to decrease the time necessary to do a test sequence.
[2] Applies to leakage current and multiple PE tests only.
69
Page 84
ESA615
Users Manual
Make a Test Sequence from a Test Sequence in the
Test Library
To make a test sequence from a test sequence already in
the test library:
1. Push .
2. Push the Test Library softkey.
3. Push or to highlight a test sequence name.
Note
Push softkey F1 to toggle between an A to Z or Z
to A sort of the test sequences.
4. Push the View/Edit softkey.
5. Push the Save as New softkey.
6. Type in a new name with the screen keyboard.
Note
A USB keyboard or barcode reader can be used
to edit the test sequence name.
7. Push the Done softkey.
The new test sequence steps are filled in with the
steps from the initial test sequence.
8. Push the Edit softkey.
Go through each setup step of the test sequence and
accept the parameters or change them for the new
test sequence.
9. Push the Save softkey to save the test sequence and
exit the wizard.
Edit a Test Sequence
To change one or more steps of a test sequence:
1. Push .
2. Push the Test Library softkey.
3. Push or to highlight a test sequence name.
Note
Push softkey F1 to toggle between an A to Z or Z
to A sort of the test sequences.
4. Push the View/Edit softkey.
5. Push the Edit softkey.
You go through each step of the test sequence and accept
the step parameters or change them.
70
Page 85
Electrical Safety Analyzer
Test Sequences
Do a Test Sequence
To do a test sequence:
1. Push to show the test sequence screen in
Figure 38.
Figure 38. Test Sequence Screen
2. Push the Test Library softkey.
3. Push or to highlight a test sequence name.
Note
Push softkey F1 to toggle between an A to Z, Z to
A, by date sort of the test sequences.
4. Push .
gtv125.bmp
Note
To see the details of the test sequence you are
about to start, push the Sequence Details
softkey.
5. Push the Next softkey.
The equipment information screen in Figure 39 shows
in the display.
gtv126.bmp
Figure 39. Equipment Information Screen
The number, serial number, manufacturer, model, and
location of the Device Under Test can be typed in to help
identify the DUT on the test results.
71
Page 86
ESA615
Users Manual
Note
A USB keyboard or barcode reader can be used
to type in the DUT information. You can use a
barcode reader to automatically move through
each field if it automatically sends a carriage
return.
When the test sequence step is a visual inspection, you
must push the Pass, Fail, or N/A softkey to move on to
the subsequent test step.
When the Product measures a value(s) for a test, Testing
shows in the display. When the test step is done, the
results are shown with PASS or FAIL in the display.
To add comments when a test step ends, push the Add Comments softkey. When you see individual results of a
Test Summary, you can push the Add Comments softkey
to type in more information about the test step.
When the test sequence is done, a name is made from the
equipment number, the date, and the time. To change the
name, push or the Edit softkey. To put the test
sequence on the memory card, push the Save softkey.
Show Test Results
To show test results in the display:
1. Push .
2. Push or to highlight a test results name.
Note
Push softkey F1 to toggle between sort results by
date or sort A-Z.
3. Push to show the test and its results in the
display.
4. Push or to highlight a test result.
5. Push to show the test results in the display.
6. Push the Back softkey to go back to the test results
list.
Note
A USB keyboard or barcode reader can be used
to edit the test results name.
When you push the Start Test Sequence softkey, the
complete test sequence starts.
72
Page 87
Electrical Safety Analyzer
Maintenance
Delete a Set of Test Results
To delete a test result:
1. Push .
2. Push or to highlight a test result name.
Note
Push softkey F1 to toggle between sort results by
date or sort A-Z.
3. Push the Delete softkey.
A delete confirmation screen will show in the display.
4. Push the Delete softkey.
Maintenance
Warning
To prevent possible electrical shock, fire, or
personal injury:
•Have an approved technician repair the
Product.
• Use only specified replacement fuses.
• Use only specified replacement parts.
• Remove the input signals before you
clean the Product.
•Disconnect the mains power cord before
you remove the Product covers.
•Turn off the Product and remove the
mains power cord. Stop for two minutes
to let the power assemblies discharge
before you open the fuse door.
•Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.
The Product is a calibrated measurement instrument. Use
the necessary precautions to prevent mechanical abuse
that could change the calibrated adjustments.
Fuse Test and Fuse Replacement
Warning
To prevent electric shock, remove all power
cords and test leads from the Product before
you open the fuse door.
For electrical protection of the equipment outlet, the
Product uses two fuses, one in the live (L1) line and one in
the neutral (L2) line. To do a fuse test:
1. Turn the Product so the case bottom is up. See
Figure .
2. Flip up the tilt stand.
3. Remove the screw in the fuse door with a #2 Phillips
head screwdriver and lift the fuse door from the
Product.
4. Remove the two fuses from the Product.
73
Page 88
ESA615
Users Manual
F1 - F2
gtv111.eps
Figure 40. Fuse Access
5. Use a multimeter to measure the continuity of each
fuse.
If a fuse does not show continuity, replace the fuse(s)
with a fuse of the same current and voltage rating.
Applicable fuse ratings are posted on the case bottom
label of the Product. Table is a list of available fuses
with Fluke Biomedical part numbers.
6. Reinstall the fuse door and secure it with the screw.
How to Clean the Product
Warning
To prevent electric shock, do not clean the
Product plugged into mains or attached to a
DUT.
Caution
Do not pour fluid onto the Product surface.
Fluid in the electrical circuitry can cause the
Product to fail.
Caution
Do not use spray cleaners on the Product.
This can cause fluid to leak into the Product
and damage electronic components.
Clean the Product occasionally with a damp cloth and
weak detergent. Try to prevent the entrance of liquids.
Clean the adapter cables with the same precautions.
Examine them for damage and deterioration of the
insulation. Examine the connections for integrity before
each use.
74
Page 89
Electrical Safety Analyzer
Replaceable Parts
Replaceable Parts
Table is a list of replaceable parts for the Product.
Table 8. Replaceable Parts
Item Fluke Biomedical Part Number
ESA615 Getting Started Manual 4105845
ESA615 Users Manual CD 4105850
USA/Japan 2238680
UK 2238596
Australia/China 2238603
Europe 2238615
Power Cord
USA to Brazil Outlet Adapter 4151242
Null Post Adapter 3326842
Ansur, CD with demo version 2795488
France/Belgium 2238615
Thailand 2238644
Israel 2434122
Switzerland 3379149
Brazil 3841358
75
Page 90
ESA615
Users Manual
Table 8. Replaceable Parts (cont.)
Item Fluke Biomedical Part Number
5-to-5 Banana jack to ECG (BJ2ECG) adapter 3359538
Carrying Case 2248650
Data Transfer Cable 4034393
USA-115, Japan
T20A 250V Fuse (Time Lag), 1¼ in x ¼ in
2183691
Australia, China,
Fuse
15 – 20 A Adapter 2195732
ESA USA/AUS/ISR Accessory Kit:
Test Lead Set
TP1 Test Probe Set
AC285 Alligator Clip Set
ESA EUR Accessory Kit:
Test Lead Set
TP74 Test Probe Set
AC285 Alligator Clip Set
To ensure safety, use exact replacement only.
Switzerland
Europe, UK, USA-220,
France/Belgium, Thailand,
Brazil, Israel
T10A 250V Fuse (Time Lag), 1¼ in x ¼ in
T16A 250V Fuse (Time Lag), 6.3 mm x 32 mm
109298
3321245
3111008
3111024
76
Page 91
Electrical Safety Analyzer
Accessories
Accessories
Table is a list of available accessories for the Product.
Table 9. Accessories
Item Fluke Biomedical Part Number
Test Leads with Retractable Sheath 1903307
Ground Pin Adapters 2242165
1–to–10 ECG Adapter 3392119
Universal Snap to Banana Adapter 2462072
Ultrasound Test Cable Adapter 3472633
USB Wireless Dongle 3341333
77
Page 92
ESA615
Users Manual
Specifications
Temperature
Operating ............................................................ 10 °C to 40 °C (50 °F to 104 °F)
Storage ............................................................... -20 °C to 60 °C (-4 °F to 140 °F)
Humidity ................................................................. 10 % to 90 % non-condensing
Altitude
100 V/115 V ac mains supply voltage ................. 5000 m
230 V ac mains supply voltage ........................... 2000 m
10 A to 15 A, 7 min on/3 min off
0 A to 10 A, continuous
AC only
DC only
* For tests that do not use MAP voltage, AC+DC, AC ONLY, and DC ONLY modes are
available for all leakages. MAP voltages are available only in True-rms (shown as
AC+DC)
Ranges ................................................................ 0.0 μA to 199.9 μA
Accuracy
DC to 1 kHz ..................................................... ±(1 % of reading + (1 μA or 1 LSD, whichever is greater))
1 kHz to 100 kHz ............................................. ±(2 % of reading + (1 μA or 1 LSD, whichever is greater))
1 kHz to 5 kHz (current > 1.6 mA) .................. ±(4 % of reading + (1 μA or 1 LSD, whichever is greater))
100 kHz to 1 MHz ........................................... ±(5 % of reading + (1 μA or 1 LSD, whichever is greater))
IEC 60601: Fig. 15
200 μA to 1999 μA
2.00 mA to 10.00 mA
Note
Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are:
μ
• At 120 V ac + (2.5
• At 230 V ac additional
A or 1 LSD, whichever is greater)
±
3.0 % and + (2.5 μA or 1 LSD, whichever is greater)
For Alternative equipment, Alternative AP, and Direct AP leakage tests, the leakage values are compensated for
nominal mains as per 62353. Therefore, the accuracy specified for other leakages is not applicable.
Mains on applied part test voltage ...................... 100 % ±7 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI
Differential leakage
Ranges ................................................................ 75 μA to 199 μA
Accuracy ............................................................. ±(10 % of reading + (2 counts or 20 μA, whichever is greater))
100 % ±7 % of Mains for IEC 62353 current limited to 3.5 mA ±25 % per IEC 62353
100 % ±7 % of Mains for IEC 60601-1 current limited to 7.5 mA ±25 % per IEC 60601-1
200 μA to 1999 μA
2.00 mA to 20.00 mA
80
Page 95
Electrical Safety Analyzer
Detailed Specifications
Insulation resistance
Ranges ................................................................ 0.5 to 20.0 MΩ
Accuracy
20 MΩ Range .................................................. ±(2 % of reading + 0.2 MΩ)
100 MΩ Range ................................................ ±(7.5 % of reading + 0.2 MΩ)
Source test voltage ............................................. 500 or 250 V dc (+20 %, -0 %) 2.0 ±0.25 mA short-circuit current
Maximum load capacitance ................................. 1 μF
ECG Performance Waveforms
Accuracy
Frequency ....................................................... ±2 %
Amplitude ........................................................ ±5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration
Waveforms
ECG Complex ................................................. 30, 60, 120, 180, and 240 BPM