Manual Supplement
Manual Title: ESA609 Users Supplement Issue: 4
Print Date: February 2014 Date: 12/14
Revision/Date: 1 Page Count: 2
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
CD Title: ESA609
CD Rev. & Date: 2/2014
CD PN: 4370089
© 2014 Fluke Corporation. All rights reserved.

ESA609 Users Manual Supplement
Change #1
On page 13, under the Performing a Ground-Wire (Protective-Earth) Resistance Test,
replace the two bullets with.
• If using an accessories probe, connect it to the other end of the test lead and place the probe tip into the
Ground Pin of the Analyzer’s test receptacle.
• If using an alligator clip accessory, connect it to the other end of the test lead, place the null post adapter in the
Ground Pin of the Analyzer’s test receptacle, and clamp the alligator clip to the null post adapter.
Change #2, 184, 185
On page 2, add the following to the Symbols table:
CAT II
MEASUREMENT CAT II is applicable to test and measuring circuits connected directly to
utilization points (socket outlets and similar points) of the low-voltage mains installation.
On page 13, under the Performing a Ground-Wire (Protective-Earth) Resistance Test, add
the following for Products shipped with a US outlet:
As a temporary deviation, this unit includes a 15 A to 20 A adapter. Use the adapter to accommodate
devices with 20 A power plugs. When this adapter is installed, use the included ground lug and zero
out the resistance for Ground Wire (Protective Earth) Resistance measurements. If you do not zero the
resistance, you must add an additional factor of 5 mΩ to the readings obtained Ground Wire
Resistance mode.
7/14 1

Manual Supplement ESA609 Users
To zero the resistance for units with the adapter, connect the adapter to the test receptacle. Follow the
procedure on page 13, under Performing a Ground-Wire (Protective-Earth) Resistance Test, and
replace the two bullets with:
• If using an accessories probe, connect it to the other end of the test lead and place the probe tip into the
Ground Pin of the 15 A to 20 A adapter.
• If using an alligator clip accessory, connect it to the other end of the test lead, place the null post adapter in
the Ground Pin of the 15 A-to-20 A adapter, and clamp the alligator clip to the null post adapter.
On page 29, under Specifications, replace the Power section with:
Power
115 Volt test outlet .............................. 100/120 V ac rms, 50Hz/60Hz, 20 A maximum
230 Volt test outlet .............................. 200/220/230/240 V ac rms, 50Hz/60Hz, 16 A maximum
Power input ......................................... 115V 20A - 2.6 kVA and 230V at 16A - 4.2 kVA
Change #3, 187
On page 30, under Leakage Current, replace the Accuracy with:
Accuracy
DC to 1 kHz ............................... ±(1 % of reading + 1 μA)
1 kHz to 100 kHz ........................ ±(2.5 % of reading + 1 μA)
100 kHz to 1 MHz ...................... ±(5 % of reading + 1 μA)
Change #4, 188
On page 29, under Specifications, remove the Agency approvals.
2 12/14