Model:
Fluke 9500 and 9500B
Product Name:
Instrument Security Procedures
9500 Oscilloscope Calibrator
Instrument Description:
The Model 9500/9500B is a calibrator offering oscilloscope test and calibration
capabilities from a single source, providing wide functionality.
Memory Description:
The Fluke 9500/9500B has the following memory devices:
1. SRAM 256 kB, Contains run time data and temporary configuration data.
2. ROM 1024 kB FLASH Memory. Contains embedded program and associated
storage area. It contains no user data.
3. ROM 256 kB FLASH Memory. Contains characterization data for the
instruments digital to analog converter.
4. RAM 256-bytes battery backed. Used to hold information about the active
head(s) that have been plugged in.
5. EEPROM 16 kB. This contains user defined data such as bus address,
Reference Frequency, safety voltage limits.
6. EEPROM 48 kB. This contains calibration constants.
7. EEPROM 16 kB. Located in 9510, 9530, 9550, 9560 active heads. This
contains calibration constants and factory data such as serial number.
8. Removable PCMCIA cards. The instrument has two slots for memory cards.
In this instrument the left slot (when viewed from the front) is used for cards
containing procedures. The slot on the right is used to store data obtained
when the procedure is executed.
& 9500B Oscilloscope Calibrator
Memory Cleaning Instructions:
1. SRAM 256 kB, This area is volatile and contents are lost on power down.
2. ROM 1024 kB FLASH Memory. This area is programmed at manufacture and
cannot be cleared.
3. ROM 256 kB FLASH Memory. This area is programmed at manufacture and
cannot be cleared.
4. RAM 256-bytes battery backed. This is data for internal use and cannot be
accessed.
9500/9500B SOMV 1 of 2 3/27/2007

5. EEPROM 16 kB. There is no single method to clear the data held in this area,
each item has to be reset manually. This can be achieved by pressing the
‘MODE’ key, then ‘CONFIG’ softkey. Select ‘MORE’ softkey. There will then
be a prompt for the password. The factory set password is ‘12321’. Each
configuration can then be selected in turn using the softkeys and set to the
desired value.
6. EEPROM 48 kB. There are no user defined items of data stored in this area
and cannot be cleared.
7. EEPROM 16 kB. Located in 9510, 9530, 9550, 9560 active heads. This area
is programmed a manufacture and cannot be cleared.
8. Removable PCMCIA cards. SDRAM Cards that have either Procedure or
Data on them may be erased using the following: Select the ‘MODE’ key,
Then the ‘TEST’ softkey. Press ‘EXIT’ softkey if there is a prompt for a
password. Select the ‘Interface’ softkey. Then the ‘MEMORY CARD’ softkey.
Insert the card to be erased into the left hand slot. Then select ‘SLOT 1’
softkey. This will overwrite the card.
The 9500/9500B does not have the ability to erase FLASH cards. This must be
done using a FLASH read/write device. These devices are not supplied by Fluke.
9500/9500B SOMV 2 of 2 3/27/2007