The Victoreen® Model 943-27 Current Mode Beta Scintillation detector is a radiation detection device that
detects beta radiation. The detector is housed in a 1.5-inch diameter stainless steel tube. The detector is
identified as the Model 943-27 and is manufactured as a P/N 943-27-31 detector assembly. Both
identification numbers are used interchangeably throughout this manual. The detector is designed for
accident range operation.
The detector operates at a potential of approximately (-) 500 VDC and provides a nominal DC output
current of 1 x 10
The detector consists of a 0.002-inch thick Titanium end window, a calcium fluoride scintillation crystal
optically coupled to a quartz light pipe and a photomultiplier tube.
The detector is designed to operate with a Model 943-227-15 Series Digital preamplifier and a Model 960
Digital Process Radiation monitoring system or a Model 942A-200C Series Digital Ratemeter. The Model
943-227-15 Digital Preamplifier is a self-contained electrometer housed in a NEMA 4 enclosure. The
preamplifier processes the DC input signal, digitizes the signal, and then transmits the digitized radiation
value to the readout device via a RC232C compatible serial data communication. For additional
information on the preamplifier and readout device, refer to the applicable product or system level
instruction manual.
Interconnection between the detector and preamplifier is accomplished using two (2) six (6) foot coaxial
cables. One cable supplies the detector high voltage and the other cable provides the signal output
For additional information on the detector preamplifier or readout device, refer to the instruction manual
for the Model 943-227-15 Preamplifier, the Model 942A-200C UDR, or the applicable Model 960 System
manual.
-12
amperes at 1 x 10-3 to 1 x 10-5 uCi/cc. The maximum current output is 1 x 10-4 amperes.
1.2 Application
The Model 943-27-31 Detector is used with the Victoreen 942A-200C or Model 960 Digital Process
Radiation monitor and the Model 943-227-15 Series Digital preamplifier. The Model 942A-200C or 960
system readouts provide the +15 VDC preamplifier electronics power, (-) 500 VDC detector power and
the RS232C serial communications support. The digital preamplifier converts the detector output into a
digital signal for processing by the digital readout. Table 1-1 is a list of available configurations and
compatible 9XX series equipment. Please contact Fluke Biomedical Radiation Management Services for
additional information.
Table 1-1 943-27 Variations and Compatible Equipm ent
Detector Preamplifier
Variations
943-27-31 Beta
Detector,
Current Mode
Operation
943-227-15 Digital
Preamplifier
943-227-15VL Digital
Preamplifier
Application Readout
Transmission
Distances
Less than 50 feet
Transmission
Distances up to 1
mile
942A-200C Digital Ratemeter
or Model 960 Digital Process
Control System
942A-200C Digital Ratemeter
or Model 960 Digital Process
Control System
1-1
943-27 Beta Scintillation Detector
Operator Manual
1.3 Specifications
General specifications for the Model 943-27 Detector are listed below. The 943-27 series products are
designed for nuclear applications, and any repairs to it by personnel not qualified to ANSI 45.2.6 1978,
Skill Level II may void the nuclear rating.
If a problem develops, the preamplifier may either be returned to the factory for service, or repaired by a
qualified technician.
Table 1-2. General Specifications for the 943-27 Detector
Feature: Specification:
Radiation Detected
End Window
Scintillation Crystal
Detector Housing
Detector Constant
Detector Range
Energy Dependence
and Efficiency
Power Requirements
Operating Temperature
Relative Humidity
Operating Pressure
Dimensions
Weight
Preamplifier Interface
Modification
Information
Beta rays
Titanium, 1.125 in dia. x 0.0002 in thick, designed to maximize beta
transmission and minimize noble gas adsorption
Calcium Fluoride, CaF
(Eu)
2
1.125 in dia. X 0.005 in thick interfaced to 1.125 in dia. x 0.125 in thick
quartz light pipe
Stainless Steel
1.0 x 10
sampling geometry
1.0 x 10
-3
uCi/cc/picoampere, typical, based on calibration method and
-12
to 1.0 x 10-4 amperes (approximately 1.0 x 10-3 to 1.0 x 105
uCi/cc, based on calibration method and sampling geometry
Dependent on sampling geometry; Refer to Appendix C for applicable
isotopic calibration reports
(-) 500 VDC @ 400 uAmperes
0° F to 122 °F (-16 °C to 50 °C)
0 – 95 %, non-condensing
12 psig max
1.5 in dia. x 9.0 in long (3.8 cm x 22.9 cm)
3 lbs (1.4 kg)
High Voltage: Six (6) foot coaxial cable with pre-terminated SHV connector
and “Seal Grip”;
Signal: Six (6) foot coaxial cable with pre-terminated BNC connector and
“Seal Grip”;
None
1-2
General Information
Inspection & Storage
1.4 Auxiliary Equipment
Model
None See Table 1-1 for compatible equipment
Description
1.5 Recommended Spare Parts
Table 1-9 Recommended Spare Parts List for the Model 943-27 Detector
Part Number Description Used On
There are no replaceable parts in the 943-27 detector
1
1.6 Receiving Inspection
Upon receipt of the unit:
1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and
notify the Fluke Biomedical Radiation Management Services Customer Service Department.
Fluke Biomedical
Radiation Management Service
6045 Cochran Road
Cleveland, Ohio 44139
Phone: 440.248.9300
Fax: 440.542.3682
2. Remove the contents from the packing material.
3. Verify that all items listed on the packing list have been received and are in good condition.
NOTE
Management Service Customer Service Department If any of the
listed items are missing or damaged, notify the Fluke Biomedical
Radiation
1-3
943-27 Beta Scintillation Detector
Operator Manual
1.7 Storage
Storage of Fluke Biomedical instruments must comply with Level B storage requirements as outlined in
ANSI N45.2.2 (1972) Section 6.1.2(.2). The storage area shall comply with ANSI N45.2.2 (1972) Section
6.2 Storage Area, Paragraphs 6.2.1 through 6.2.5. Housekeeping shall conform to ANSI N45.2.3 (1972).
Level B components shall be stored within a fire resistant, tear resistant, weather tight enclosure, in a
well-ventilated building or equivalent.
Storage of Fluke Biomedical instruments must comply with the following:
1. Inspection and examination of items in storage must be in accordance with ANSI N45.2.2 (1972)
Section 6.4.1.
2. Requirements for proper storage must be documented and written procedures or instructions must
be established.
3. In the event of fire, post-fire evaluation must be in accordance with ANSI N45.2.2 (1972), Section
6.4.3.
4. Removal of items from storage must be in accordance with ANSI N45.2.2 (1972), Sections 6.5 and
6.6.
1-4
Theory of Operation
Functional Description
2
Section 2
Theory of Operation
2.1 Functional Description
The detector is comprised of a thin calcium fluoride scintillation disc located behind an ultra-thin (0.0002
in) Titanium end window. The end window permits the penetration of Beta rays as low as 80 Kev. The
calcium fluoride disc is optically coupled to the active cathode of a one-inch diameter photomultiplier tube
(PMT). A schematic representation of the detector is shown on drawing 943-27-32, located in Appendix
A. As shown in the schematic, the cathode of the PMT is tied to a negative high voltage through a ten
(10) megohm current limiting resistor, R1. The PMT dynodes are biased by the divider string formed by
resistors R2 through R13. The output signal is taken directly from the PMT anode, which is held at a
virtual ground potential by the input of the associated integrating electrometer.
2.2 Theory of Operation
Beta particles that have enough energy to penetrate the end window and impinge upon the disc will
produce light pulses, proportional to the energy deposited in the disc. The disc is thin so that gamma rays
will not have a high incidence of interaction with the disc. This yields a high rejection of gamma rays while
still possessing good sensitivity to beta particles.
A photomultiplier tube, optically coupled to the scintillation disc, detects visible light emitted from the disc
and converts this light to electrical energy that is proportional to the energy deposited by the beta particle.
The electrical charge is sent to an electrometer in an external preamplifier that converts the charge into a
digital value for display and processing.
2-1
(Blank Page)
Operation
Installation
3
Section 3
Operation
3.1 Installation
Installation consists of mounting the detector in its associated sampling geometry and making the
required electrical connections. Refer to the applicable sampling geometry in struction manual for
detector installation instructions.
CAUTION
Remove all power prior to installing the Detector.
Detector Mounting
The detector is designed to be installed within a sampling geometry. The detector is normally part of the
process pressure boundary, and is normally installed using to mounting flanges and 2 quad-lobed seal orings. The basic mounting procedure is provided below. Refer to the specific sampling geometry
installation procedure for further information.
The following materials are normally needed to remove or install the Model 943-27 detector:
O-Ring Lubricant, Dow Corning #55, or equivalent
1ea, 3/16 Hex head bit or T-handle tool
1ea, P/N 943-27-47, Detector Mounting Flange
1ea, P/N 46-103, Quad lobed seal (Detector mounting flange to Adapter flange)
6ea, P/N 5-610, 10-32 x 0.62 Socket head cap screws (varies with application)
6ea, P/N 5-795, #10 split washes, Stn Stl (varies with application)
1ea, P/N 943-27-48, Adaptor Flange
1ea, P/N 46-104, Quad lobed seal (Adapter flange to sampler body)
6ea, P/N 5-1080, 10-32 x 0.88 Socket head cap screws (varies with application)
6ea, P/N 5-795, #10 split washes, Stn Stl (varies with application)
6ea, P/N 5-795, #10 flat washers, Stn Stl (varies with application)
CAUTION
Due to the potential for high activity radioactive
gasses in or around the vicinity of the detector,
protective Clothing should be worn while installing
or removing the detector
Detector Installation
Detector installation includes installing the adaptor flange and quad-lobed seal o-ring, installing the
detector mounting flange, detector flange quad-lobed seal o-ring, inserting the detector into the detector
mounting tube and securing the detector mounting flange. The high voltage power supply for the detector,
if operational, must be turned off by turning the controlling Digital Ratemeter or 960 System power OFF.
3-1
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.