The Victoreen® Model 943-27 Current Mode Beta Scintillation detector is a radiation detection device that
detects beta radiation. The detector is housed in a 1.5-inch diameter stainless steel tube. The detector is
identified as the Model 943-27 and is manufactured as a P/N 943-27-31 detector assembly. Both
identification numbers are used interchangeably throughout this manual. The detector is designed for
accident range operation.
The detector operates at a potential of approximately (-) 500 VDC and provides a nominal DC output
current of 1 x 10
The detector consists of a 0.002-inch thick Titanium end window, a calcium fluoride scintillation crystal
optically coupled to a quartz light pipe and a photomultiplier tube.
The detector is designed to operate with a Model 943-227-15 Series Digital preamplifier and a Model 960
Digital Process Radiation monitoring system or a Model 942A-200C Series Digital Ratemeter. The Model
943-227-15 Digital Preamplifier is a self-contained electrometer housed in a NEMA 4 enclosure. The
preamplifier processes the DC input signal, digitizes the signal, and then transmits the digitized radiation
value to the readout device via a RC232C compatible serial data communication. For additional
information on the preamplifier and readout device, refer to the applicable product or system level
instruction manual.
Interconnection between the detector and preamplifier is accomplished using two (2) six (6) foot coaxial
cables. One cable supplies the detector high voltage and the other cable provides the signal output
For additional information on the detector preamplifier or readout device, refer to the instruction manual
for the Model 943-227-15 Preamplifier, the Model 942A-200C UDR, or the applicable Model 960 System
manual.
-12
amperes at 1 x 10-3 to 1 x 10-5 uCi/cc. The maximum current output is 1 x 10-4 amperes.
1.2 Application
The Model 943-27-31 Detector is used with the Victoreen 942A-200C or Model 960 Digital Process
Radiation monitor and the Model 943-227-15 Series Digital preamplifier. The Model 942A-200C or 960
system readouts provide the +15 VDC preamplifier electronics power, (-) 500 VDC detector power and
the RS232C serial communications support. The digital preamplifier converts the detector output into a
digital signal for processing by the digital readout. Table 1-1 is a list of available configurations and
compatible 9XX series equipment. Please contact Fluke Biomedical Radiation Management Services for
additional information.
Table 1-1 943-27 Variations and Compatible Equipm ent
Detector Preamplifier
Variations
943-27-31 Beta
Detector,
Current Mode
Operation
943-227-15 Digital
Preamplifier
943-227-15VL Digital
Preamplifier
Application Readout
Transmission
Distances
Less than 50 feet
Transmission
Distances up to 1
mile
942A-200C Digital Ratemeter
or Model 960 Digital Process
Control System
942A-200C Digital Ratemeter
or Model 960 Digital Process
Control System
1-1
Page 6
943-27 Beta Scintillation Detector
Operator Manual
1.3 Specifications
General specifications for the Model 943-27 Detector are listed below. The 943-27 series products are
designed for nuclear applications, and any repairs to it by personnel not qualified to ANSI 45.2.6 1978,
Skill Level II may void the nuclear rating.
If a problem develops, the preamplifier may either be returned to the factory for service, or repaired by a
qualified technician.
Table 1-2. General Specifications for the 943-27 Detector
Feature: Specification:
Radiation Detected
End Window
Scintillation Crystal
Detector Housing
Detector Constant
Detector Range
Energy Dependence
and Efficiency
Power Requirements
Operating Temperature
Relative Humidity
Operating Pressure
Dimensions
Weight
Preamplifier Interface
Modification
Information
Beta rays
Titanium, 1.125 in dia. x 0.0002 in thick, designed to maximize beta
transmission and minimize noble gas adsorption
Calcium Fluoride, CaF
(Eu)
2
1.125 in dia. X 0.005 in thick interfaced to 1.125 in dia. x 0.125 in thick
quartz light pipe
Stainless Steel
1.0 x 10
sampling geometry
1.0 x 10
-3
uCi/cc/picoampere, typical, based on calibration method and
-12
to 1.0 x 10-4 amperes (approximately 1.0 x 10-3 to 1.0 x 105
uCi/cc, based on calibration method and sampling geometry
Dependent on sampling geometry; Refer to Appendix C for applicable
isotopic calibration reports
(-) 500 VDC @ 400 uAmperes
0° F to 122 °F (-16 °C to 50 °C)
0 – 95 %, non-condensing
12 psig max
1.5 in dia. x 9.0 in long (3.8 cm x 22.9 cm)
3 lbs (1.4 kg)
High Voltage: Six (6) foot coaxial cable with pre-terminated SHV connector
and “Seal Grip”;
Signal: Six (6) foot coaxial cable with pre-terminated BNC connector and
“Seal Grip”;
None
1-2
Page 7
General Information
Inspection & Storage
1.4 Auxiliary Equipment
Model
None See Table 1-1 for compatible equipment
Description
1.5 Recommended Spare Parts
Table 1-9 Recommended Spare Parts List for the Model 943-27 Detector
Part Number Description Used On
There are no replaceable parts in the 943-27 detector
1
1.6 Receiving Inspection
Upon receipt of the unit:
1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and
notify the Fluke Biomedical Radiation Management Services Customer Service Department.
Fluke Biomedical
Radiation Management Service
6045 Cochran Road
Cleveland, Ohio 44139
Phone: 440.248.9300
Fax: 440.542.3682
2. Remove the contents from the packing material.
3. Verify that all items listed on the packing list have been received and are in good condition.
NOTE
Management Service Customer Service Department If any of the
listed items are missing or damaged, notify the Fluke Biomedical
Radiation
1-3
Page 8
943-27 Beta Scintillation Detector
Operator Manual
1.7 Storage
Storage of Fluke Biomedical instruments must comply with Level B storage requirements as outlined in
ANSI N45.2.2 (1972) Section 6.1.2(.2). The storage area shall comply with ANSI N45.2.2 (1972) Section
6.2 Storage Area, Paragraphs 6.2.1 through 6.2.5. Housekeeping shall conform to ANSI N45.2.3 (1972).
Level B components shall be stored within a fire resistant, tear resistant, weather tight enclosure, in a
well-ventilated building or equivalent.
Storage of Fluke Biomedical instruments must comply with the following:
1. Inspection and examination of items in storage must be in accordance with ANSI N45.2.2 (1972)
Section 6.4.1.
2. Requirements for proper storage must be documented and written procedures or instructions must
be established.
3. In the event of fire, post-fire evaluation must be in accordance with ANSI N45.2.2 (1972), Section
6.4.3.
4. Removal of items from storage must be in accordance with ANSI N45.2.2 (1972), Sections 6.5 and
6.6.
1-4
Page 9
Theory of Operation
Functional Description
2
Section 2
Theory of Operation
2.1 Functional Description
The detector is comprised of a thin calcium fluoride scintillation disc located behind an ultra-thin (0.0002
in) Titanium end window. The end window permits the penetration of Beta rays as low as 80 Kev. The
calcium fluoride disc is optically coupled to the active cathode of a one-inch diameter photomultiplier tube
(PMT). A schematic representation of the detector is shown on drawing 943-27-32, located in Appendix
A. As shown in the schematic, the cathode of the PMT is tied to a negative high voltage through a ten
(10) megohm current limiting resistor, R1. The PMT dynodes are biased by the divider string formed by
resistors R2 through R13. The output signal is taken directly from the PMT anode, which is held at a
virtual ground potential by the input of the associated integrating electrometer.
2.2 Theory of Operation
Beta particles that have enough energy to penetrate the end window and impinge upon the disc will
produce light pulses, proportional to the energy deposited in the disc. The disc is thin so that gamma rays
will not have a high incidence of interaction with the disc. This yields a high rejection of gamma rays while
still possessing good sensitivity to beta particles.
A photomultiplier tube, optically coupled to the scintillation disc, detects visible light emitted from the disc
and converts this light to electrical energy that is proportional to the energy deposited by the beta particle.
The electrical charge is sent to an electrometer in an external preamplifier that converts the charge into a
digital value for display and processing.
2-1
Page 10
(Blank Page)
Page 11
Operation
Installation
3
Section 3
Operation
3.1 Installation
Installation consists of mounting the detector in its associated sampling geometry and making the
required electrical connections. Refer to the applicable sampling geometry in struction manual for
detector installation instructions.
CAUTION
Remove all power prior to installing the Detector.
Detector Mounting
The detector is designed to be installed within a sampling geometry. The detector is normally part of the
process pressure boundary, and is normally installed using to mounting flanges and 2 quad-lobed seal orings. The basic mounting procedure is provided below. Refer to the specific sampling geometry
installation procedure for further information.
The following materials are normally needed to remove or install the Model 943-27 detector:
O-Ring Lubricant, Dow Corning #55, or equivalent
1ea, 3/16 Hex head bit or T-handle tool
1ea, P/N 943-27-47, Detector Mounting Flange
1ea, P/N 46-103, Quad lobed seal (Detector mounting flange to Adapter flange)
6ea, P/N 5-610, 10-32 x 0.62 Socket head cap screws (varies with application)
6ea, P/N 5-795, #10 split washes, Stn Stl (varies with application)
1ea, P/N 943-27-48, Adaptor Flange
1ea, P/N 46-104, Quad lobed seal (Adapter flange to sampler body)
6ea, P/N 5-1080, 10-32 x 0.88 Socket head cap screws (varies with application)
6ea, P/N 5-795, #10 split washes, Stn Stl (varies with application)
6ea, P/N 5-795, #10 flat washers, Stn Stl (varies with application)
CAUTION
Due to the potential for high activity radioactive
gasses in or around the vicinity of the detector,
protective Clothing should be worn while installing
or removing the detector
Detector Installation
Detector installation includes installing the adaptor flange and quad-lobed seal o-ring, installing the
detector mounting flange, detector flange quad-lobed seal o-ring, inserting the detector into the detector
mounting tube and securing the detector mounting flange. The high voltage power supply for the detector,
if operational, must be turned off by turning the controlling Digital Ratemeter or 960 System power OFF.
3-1
Page 12
943-27 Beta Scintillation Detector
Operator Manual
Adapter Flange Installation: Lubricate the P/N 46-104 Quad lobed seat, the larger of the 2 seals, with Oring lubricant, Dow Corning #55, or equivalent. Insert the seal into the groove on the P/N 943-27-48
adaptor flange, the larger of the 2 flanges, with 2 sets of mounting holes. Place the adaptor flange over
the mating mounting surface. The quad lobed seal must be located between the adaptor flange and the
mating stationery mounting surface. Secure the adaptor flange to the mounting surface using the
hardware provided with the sampling geometry. Tighten the nuts securely (50 inch-lbs).
Detector Installation: Locate the P/N 943-27-47 detector-mounting flange, the smaller of the 2 flanges,
with 1 set of mounting holes. From the bottom of the detector, the end with the titanium window, slide the
detector flange over the body of the detector, up to the cables at the opposite end of the detector. The
groove cut into the flange must face the bottom of the detector. Lubricate the P/N 46-103 Quad lobed
seat, the smaller of the 2 seals, with O-ring lubricant, Dow Corning #55, or equivalent. Slide the quad
lobed seal over the body of the detector, up to the detector flange. Insert the seal into the groove on the
P/N 943-27-47 detector flange. The quad lobed seal must face the bottom, or sensitive end of the
detector to permit sealing to the adaptor flange.
Install the Model 943-27 detector, with detector flange and quad lobed seal, into the hole in the adaptor
flange and into the detector housing. Insert the detector into the detector housing until the sensitive end
of the detector rests on the support ring on the bottom of the detector tube. Use care to not damage the
thin, beta sensitive end window on the detector.
Detector Flange Installation: Slide the detector mounting flange and quad lobed seal down the body of
the detector until it stops on the detector adaptor flange. Install the 6, 10-32 x 0.62 inch long, stainless
steel, Socket head cap screws and lock washers into the holes provided in the detector flange. Line up
the holes in the detector flange with the threaded holes in the adaptor flange. Screw the cap screws into
the threaded mounting holes in the adaptor flange. Tighten the nuts securely (50 inch-lbs ).
This completes the detector installation. The sampling geometry may now be pressure tested. The
maximum pressure for the system is 12 psig.
Detector Removal:
CAUTION
Due to the potential for high activity radioactive gasses in or
around the vicinity of the detector, protective Clothing should
be worn while installing or removing the detector.
WARNING
The sampling geometry is an accident range noble gas monitor,
and radioactive gasses may be present within the sampling
volume. The local plant Health Physicist should be consulted
prior to removing the Detector or opening the Grab sample valve.
Detector removal is the opposite of the typical detector installation described above. If the detector has
not been replaced for over 18 months, the 46-103 Quad lobed seal should be inspected for wear or
damage. If the seal is worn, it should be replaced at this time. When replacing the seal, the new seal
should be lubricated with O-ring lubricant, Dow Corning #55, or equivalent. If the detector adaptor flange
has not been replaced for over 6 years, the 46-104 Quad lobed adaptor flange seal should be inspected
for wear or damage. If the seal is worn, it should be replaced at this time. When replacing the seal, the
new seal should be lubricated with O-ring lubricant, Dow Corning #55, or equivalent
3-2
Page 13
Operation
Installation
3
CAUTION
Remove all power prior to connecting field wiring.
Electrical Interface
Electrical connections from the detector include a high voltage and a signal coaxial cable, with preterminated connectors. The connectors are designed for termination within the associated preamplifier.
The detector high voltage and signal cables enter the preamplifier through “Seal Grip” type penetrations
on the enclosure body. The “Seal Grips” contain a rubber gland that is intended to grip the coaxial cables
to prevent moisture entry into the enclosure. The “Seal Grips” are an integral part of the detector cables.
The “Seal Grip” enclosure nut is removed from the fitting. The insulating washer stays with the fitting.
The detector signal and high voltage cables, with connectors and “Seal Grip” fitting nuts are then inserted
into the holes preamplifier enclosure. The “Seal Grip: end fittings are then re-installed securely against
the wall of the enclosure.
The detector Signal (BNC) and High Voltage (SHV) cables are then terminated on the bulkhead
connectors located on the enclosure inner panel.
NOTE
The electrometer circuitry in the preamplifier is highly sensitive to
moisture and physical damage. Use extreme care when the preamplifier
case is opened as the high impedance electrometer circuitry is easily
damaged. Do not leave the preamplifier enclosure in the open position
for any extended period of time.
3.2 Set-up
Once the detector is installed and electrically connected to the preamplifier, the detector must be
calibrated. This includes exposing the detector to a series of standard radioactive sources and adjusting
the high voltage supplied by the readout device to obtain the same output current (adjusted for
radioactive decay) that was obtained during the factory calibration. This data is provided on the
Calibration Data sheet provided with the detector. To place the system in operation, the following steps
should be performed:
NOTE
Ensure you have read and fully understand
Section 1, 2 and 3 prior to continuing.
1. Adjust the High Voltage. Refer to the applicable preamplifier and readout/controller device manual
prior to powering up the controller and adjusting the high voltage.
3-3
Page 14
943-27 Beta Scintillation Detector
Operator Manual
3.3 Operation
Once the detector/preamplifier/readout devices have been installed and the high voltage has been
adjusted, operation of the Model 943-27 Current Mode Beta Scintillation Detector is automatic.
Operation of the 943-27 Detector is controlled by the Model 943-227-17 Digital Preamplifier and the
Model 942A-200C Digital Ratemeter or Model 960 Digital Process Radiation monitoring system. Once
the power to the controlling device is turned on, detector operation begins. No operator interaction is
required.
An operational check source is normally provided with the sampling geometry. The check source is
actuated from the controlling device and is used to verify operation of the detector by producing an
upscale reading on the controlling device.
Normal Operation
If the measured radiation field is within the range of the detector during power-up, the uCi/cc value will be
displayed on the controller. The digital display will update once per second.
Alarms
The radiation alarms are initiated by the applicable controller. Refer to the applicable controller manual.
3-4
Page 15
Maintenance, Calibration and Troubleshooting
Maintenance
4
Section 4
Maintenance, Calibration and Troubleshooting
4.1 Maintenance
The 943-27-15 Detector is designed to operate for extended periods of time with no scheduled
maintenance required. Operation may be verified by periodically actuating the check source (provided
with the detector and sampling assembly) and observing the response of the unit. If the response varies
by more than 50% of the normal value, further troubleshooting may be required. If a problem develops,
troubleshoot the unit per Section 4.3 and the drawings in Appendix B.
Periodically, approximately every 18 months, it is suggested that the detector be re-calibrated.
4.2 Calibration
Electronic Adjustments
There are no user adjustments available on the 943-27 Detector. The detector high voltage is determined
during calibration and is provided by the associated readout/controller.
Point source Calibration
WARNING
The calibration procedures require the use of high activity
beta emitting button sources. These sources are intended
solely to provide a beta radiation field. They should be
used only by persons who have been trained in the
appropriate safety procedures to be followed in the
presence of beta radiation and the proper interpretation of
the data. Notify the plant Health Physicist prior to the use
of these sources.
SPECIAL NOTE
Due to the activity of these sources, possession of these
sources requires a USNRC Byproduct Material License or
Agreement State License, as defined in Title 10 of the
Code of Federal Regulations.
4-1
Page 16
943-27 Beta Scintillation Detector
Operator Manual
Source and Source Holder Description
WARNING
USE PRECAUTIONS FOR HANDLING
RADIOACTIVE MATERIAL SOURCES
WARNING
HANDLE WITH CAUTION
WARNING
”HIGH BETA RADIATION FIELDS”
Handle the sources with caution.
Direct the open end of the source holder away from personnel.
DO NOT Look at the Open Source.
Wear safety glasses.
DO NOT touch the source window.
DO NOT remove the large end of the holder. It contains the source.
Table 4-1. Current Mode Beta Detector Calibration Source Set
Part Number: Isotope: Activity, microCuries
844-130-1-12
844-130-1-13
844-130-1-14
Refer to figure 4-1 for a pictorial view of the source holder assembly. The source holder
consists of the following components:
Shielding: The source is housed in an Aluminum cylinder, 2.5-inch diameter x 4.5inch length. The source is sealed in one end of the cylinder. Refer to drawing844131-1-TAB, included in Appendix A.
90
Sr
137
Cs
36
Cl
CAUTION
200
200
200
4-2
The activity of the sources are a nominal 200
microCuries. The cesium source holder has no lead
shielding and readings on the surface of the holder are
approximately 25 mR/hr (0.25 mGy/hr). Consult the
local Health Physicist for precautions that must be
followed when handling the sources.
Page 17
Maintenance, Calibration and Troubleshooting
Calibration
Source: The activity is deposited on a 40 mm diameter Whatman #1 qualitative filter paper that is
bonded to a 0.51 mm thick stainless steel disc. The disc is sealed with a 2-mil Kapton window and
sealed into an aluminum mount. The overall dimensions on the source disc are 3.2 mm in height
and 47 mm in diameter
.
Calibration: To calibrate the detector, the small end cap on the source holder is removed,
exposing the internal radioactive source.
WARNING
”HIGH BETA RADIATION FIELDS”
Handle the sources with caution.
Direct the open end of the source holder away from personnel.
DO NOT Look at the Open Source.
Wear safety glasses.
DO NOT touch the source window.
Follow the following precautions when handling the source holder:
1. Remember that high Beta radiation fields exist when the end cap is
removed
2. Direct the open end of the source holder away from people.
3. Do not look directly into the end of the open source holder.
4. Wear safety glasses when using the source.
5. Do Not Touch the source window. The thin window is easily
punctured, directly exposing the source material.
6. The U.S. Nuclear Regulatory Commission requires a leak test of the
source every six (6) months.
7. The large end cap of the holder contains the radioactive source
capsule. DO NOT REMOVE.
4
4-3
Page 18
943-27 Beta Scintillation Detector
Operator Manual
Figure 4-1 Source Holder Assembly and Operation
4-4
Page 19
Maintenance, Calibration and Troubleshooting
Calibration
4
Source Wipe Test
Due to the amount of activity in each source, the Current Mode Beta source set requires a USNRC or
Agreement State Byproduct Material license. To guard against leaks of radioactive material, or possible
contamination, the regulatory agency generally requires a wipe test be performed at an interval not to
exceed six (6) months for each source.
WARNING
The wipe test of the source is made with the end cap
removed. Any part of the body in front of the open
source may be exposed to high radiation fields. Do
not test quickly to minimize exposure. Direct the open
end away from personnel and wear safety glasses.
The activity is covered with a 2 mil Kapton window and
is sealed into the aluminum mount. DO NOT TOUCH
the source window with anything, including the wipes.
Wipe Test Procedure:
1. Procure a long handled Q-tip type cotton swab and moisten it with alcohol.
2. Remove the small end cap. Direct the open end away from any personnel.
WARNING
Do not swab the source window. It is a 2-mil Kapton window
that may easily be broken.
3. Holding the swab, quickly swab the internal surface of the cylinder.
4. Replace the small end cap. Return the holder to storage.
5. Measure the swab using a detector capable of detecting 0.005 microCuries or
less and record the data.
6. If more than 0.005 microCuries of activity is detected, notify the cognizant plant
Radiation Safety Officer or Health Physicist.
4-5
Page 20
943-27 Beta Scintillation Detector
Operator Manual
Source Holder Data Sheet:
The following summarizes the data provided with the source holders:
CURRENT MODE BETA SOURCE SET
P/N 844-130-1-12
P/N 844-130-1-13
P/N 844-130-1-14
USE PRECAUTIONS FOR HANDLING SOURCES
RADIOACTIVE MATERIAL
90
137
36
Sr
Cs
Cl
WARNING: HANDLE WITH CAUTION
HIGH BETA RADIATION FIELDS EXIST WHEN END CAP IS REMOVED
DIRECT THE OPEN SOURCE AWAY FROM PEOPLE
WEAR SAFETY GLASSES WHEN USING THE SOURCES
DO NOT TOUCH THE SOURCE WINDOW – THE THIN WINDOW WILL
PUNCTURE EASILY
U. S. NUCLEAR REGULATORY COMMISSION REQUIRES SIX MONTH LEAK TESTING
NOTICE: THE LARGE END CAP CONTAINS THE RADIOACTIVE SOURCE
CAPSULE
DO NOT REMOVE
PART
NUMBER
844-130-1-12 XXXX XXXX
SERIAL
NUMBER,
FLUKE
SERIAL
NUMBER,
MANUFACTURER
RADIOACTIVE
MATERIAL
90
Sr
ACTIVITY,
microcuries
200 XX/XX/XX
DATE OF
MEASUREMENT
844-130-1-13 XXXX XXXX
844-130-1-14 XXXX XXXX
137
36
Cs
Cl
200 XX/XX/XX
200 XX/XX/XX
Detector Calibration Procedure
The following procedure is recommended to assure an accurate calibration of the Model 943-27 Current
Mode Beta scintillation detector.
Equipment Required:
1. Detector to be calibrated
2. Electronics to be used with the detector
3. Calibration sources P/N 844-130-1-12, 844-1-130-1-13 and 844-1 30-1-14.
4. Data Sheet: The data sheets are used to delineate the minimum recording requirements. The
baseline calibration data is supplied at the time of installation. Record all identification
information on the data sheet prior to doing any measurements. The expected calibration
values must be corrected for decay prior to doing the calibration. Refer to the calibration
procedures in Appendix B for sample data sheets.
4-6
Page 21
Maintenance, Calibration and Troubleshooting
Calibration
Detector Leakage:
1. With the detector connected to the readout device, apply power and allow a 15-minute warmup.
2. Record the readout with no radiation sources present. The maximum background current is
approximately 4.0 x 10
3. Expose the window of the detector directly to light. Read detector leakage. There should be
little difference between this and the previous reading. If there is a significant increase in
readings, the detector window has a light leak and must be repaired before p roceeding.
Calibration Source Measurements, General:
1. Remove the detector from the sampling geometry and note the background radiation reading.
In a normal background environment, the background should be less than 4.0 x 10
If the background is greater than this, the detector should be replaced. Record the ambient
background radiation value.
-11
amps.
-11
WARNING
The following procedure exposes the operator to beta
radiation. Safety glasses must be worn and the amount
of “line of sight” exposure time must be minimized.
2. Remove the source holder end cap. This is the smaller of the two end caps.
3. Insert the detector into the P/N 844-211-1-13,
4. With the source in place, allow approximately two minutes for the reading to stabilize. Record
the reading displayed on the readout device. The reading should yield a net value , within +/2%, of the decay corrected, background subtracted target value provided on the original
calibration data sheet. If the reading is not within +/- 2%, the high voltage may be increased or
decreased as necessary. Once +/-2% agreement is reached, NO FURTHER ADJUSTMENTS
of the high voltage are permitted.
5. Remove the detector from the source holder and re-install the end cap.
6. Re-run the background measurement. The background should remain less than 4.0 x 10
amperes
7. Repeat steps 2 through 5 using the remaining two sources. The reading should yield a net
value, within +/- 6%, of the decay corrected, background subtracted target value provided on
the original calibration data sheet.
Calibration Source Measurements, using Model 960 Digital Radiation Monitoring System:
1. Select the Display Option on the 960 front panel keypad that converts the display units to
raw output from the detector (in amperes).
2. Remove the detector from the sampler.
3. Decay to the present day the “net current” value printed on the calibration data sheet,
under the heading of customer sources, for the Cesium-137 source.
137
Cs reference source.
WARNING
4
amperes.
-11
The following procedure exposes the operator to beta
radiation. Safety glasses must be worn and the amount
of “line of sight” exposure time must be minimized.
4. Remove the cap marked “Removing this cap will expose user to beta radiation”.
5. Place the Cesium-137 source assembly on the ground with the open side up.
6. Insert the detector into the source assembly.
4-7
Page 22
943-27 Beta Scintillation Detector
Operator Manual
7. Wait a minimum of 1 minute, and then observe the 960 display. Compare the value
displayed to the calculated value in step 3 above. The observed value must be within +/2% of the calculated value.
8. If the value is within the required tolerance, proceed to step 10.
9. If the value is not within the required tolerance, adjust the scaler high voltage (96SF, R8)
until the observed value falls within the required tolerance. If the observed value is too
low, increase the high voltage. If the observed value is too high, decrease the high
voltage.
DO NOT adjust the high voltage after this point
10. Remove the detector from the source holder and lay it on the ground. Replace the end
cap on the source holder.
11. Wait 5 minutes, and then observe the 960 display. The displayed value is the
background current for this detector. The value must be less than 4.0 x 10
greater than 4.0 x 10
replacement.
12. Decay to the present day the “net current” value printed on the calibration data sheet,
under the heading of customer sources, for the Strontium-90 source.
13. Remove the cap from the Strontium-90 source holder. Place the source holder on the
ground. Insert the detector into the holder.
14. Wait a minimum of 1 minute, and then observe the 960 display. Compare the value
displayed to the calculated value in step 12 above. The observed value must be within
+/-6% of the calculated value.
15. Remove the detector from the source holder and lay it on the ground. Replace the end
cap on the source holder.
16. Decay to the present day the “net current” value printed on the calibration data sheet,
under the heading of customer sources, for the Chlorine-36 source.
17. Remove the cap from the Chlorine-36 source holder. Place the source holder on the
ground. Insert the detector into the holder.
18. Wait a minimum of 1 minute, and then observe the 960 display. Compare the value
displayed to the calculated value in step 16 above. The observed value must be within
+/-6% of the calculated value.
19. Remove the detector from the source holder and lay it on the ground. Replace the end
cap on the source holder.
20. If all 4 values fall within the required tolerances, the detector is calibrated. Record the
values obtained as well as the detector high voltage. The high voltage may be measured
at the 1000:1 test jack on the 960 front panel.
21. If any of the 4 values does not fall within the required tolerances, return the detector for
repair or replacement.
22. Re-install the detector into the sampler.
23. This concludes the field detector calibration procedure.
NOTE
-11
amperes, the det4ector should be returned for repair or
-11
amperes. If
4-8
Page 23
Maintenance, Calibration and Troubleshooting
Troubleshooting
4
4.3 Troubleshooting
WARNING
Extreme care must be used when troubleshooting a
system that has power applied. All standard
troubleshooting precautions apply.
WARNING
Once a problem has been located, remove all power
before continuing with the repair.
CAUTION
Personnel performing the troubleshooting/repair
must be qualified to ANSI 45.2.6, 1978, Skill Level II.
Personnel performing the following procedure must
be familiar with the operation of the monitoring
system and the location of each piece of equipment
used in the system.
If a problem develops, verify that the voltages at connection point inputs and outputs are present and that
all wiring is secure. Refer to Appendix B and C for drawings and parts lists. If a PROM requires
replacement, specify the board revision level when ordering the part.
Troubleshooting / Adjustments
Troubleshooting includes a verification of detector response by actuating the radioactive check source
and verification of the high voltage value. The check source is actuated from the associated
readout/controller. The high voltage may be measured at the connection point in the digital preamplifier
enclosure.
Specialized test equipment and firmware are required to adjust the auto-zero and PGA subsystems in the
electrometer. If the test equipment and firmware is not available, the unit should be returned to Fluke
Biomedical RMS for re-alignment.
B.2 Related Manuals (Not Supplied with this document)
B
SXXXXXX-1 Applicable 960 Series system Manual
942A-200C-M1 Digital Ratemeter
943-227-15-1 Digital Preamplifier Manual
B.3 Related Calibration Reports
(Not supplied with this document)
The available detector calibration procedures and primary isotopic calibration reports
applicable to the detector are listed below. The calibration reports each describe a
different sampling geometry. Copies of these procedures are available from Fluke
Biomedical RMS Customer Service Department at (440) 498-2564.