Fluke 9260 User guide

Page 1
Hart Scientific
9260
Mini Fixed-Point Cell Furnace
Users’ Guide
Rev. 611902
Page 2
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com
www.hartscientific.com
Subject to change without notice. • Copyright © 2005 • Printed in USA
Rev. 611902
Page 3

Table of Contents

1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.1 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.2 CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Hart Scientific Authorized Service Centers . . . . . . . . . . . . . 5
2 Specifications and Environmental Conditions . . . . . . . . . . 7
2.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 “Dry-out” Period . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Setting the Temperature . . . . . . . . . . . . . . . . . . . . . . . 12
3.6 Changing Display Units . . . . . . . . . . . . . . . . . . . . . . . 12
4 Parts and Controls . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Bottom Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3 Top Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.4 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.5 Thermal Block Assembly . . . . . . . . . . . . . . . . . . . . . . 21
4.5.1 Thermal Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.5.2 Heaters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.5.3 Basket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.5.4 Thermal Shunt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.5.5 Insulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.5.6 Temperature Control Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.6 Mini Fixed-point Cells . . . . . . . . . . . . . . . . . . . . . . . 22
4.7 Comparison Blocks . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.7.1 Block Well Sizes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
i
Page 4
4.7.2 Comparison Block Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.7.2.1 Comparison Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.7.2.2 Thermal Shunt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
4.7.2.3 Top Insulation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 Controller Operation . . . . . . . . . . . . . . . . . . . . . . . 27
5.1 Well Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.2 Reset Cut-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.3 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . 29
5.3.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.3.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.4 Temperature Scale Units . . . . . . . . . . . . . . . . . . . . . . 30
5.5 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.5.1 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.5.2 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.6 Program Advance . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.7 Temperature Scale Units . . . . . . . . . . . . . . . . . . . . . . 32
5.8 Secondary Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.9 Heater Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.10 Set-point Resistance . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.11 Proportional Band . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.12 Controller Configuration . . . . . . . . . . . . . . . . . . . . . . 34
5.13 Operating Parameters . . . . . . . . . . . . . . . . . . . . . . . . 34
5.13.1 High Limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.13.2 Soft Cut-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.13.3 Cut-out Reset Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.14 Program Parameters . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.14.1 Fixed-point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.14.2 Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.14.3 Curve Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.15 Serial Interface Parameters . . . . . . . . . . . . . . . . . . . . . 38
5.15.1 Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.15.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.15.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.15.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.16 Calibration Parameters . . . . . . . . . . . . . . . . . . . . . . . 39
5.16.1 Hard Cut-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.16.2 R
5.16.3 ALPHA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.16.4 DELTA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.16.5 Top and Bottom Zone Percent Heating . . . . . . . . . . . . . . . . . . . . . 40
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
0
5.16.5.1 Bottom Zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.16.5.2 Top Zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6 Digital Communication Interface . . . . . . . . . . . . . . . . 43
ii
Page 5
6.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . 43
6.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.1.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.1.2.1 Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.1.2.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.1.2.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.1.2.4 Linefeed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.1.3 Serial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.2 Interface Commands . . . . . . . . . . . . . . . . . . . . . . . . 45
7 Fixed-Point Realization . . . . . . . . . . . . . . . . . . . . . 49
7.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.2 Installing a Sealed Cell into the Basket . . . . . . . . . . . . . . . 49
7.3 Melting Point Realization . . . . . . . . . . . . . . . . . . . . . . 51
7.3.1 Melting Point Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.3.1.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.3.1.2 Setting Up The Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.3.1.3 Program Initiation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.4 Freezing Point Realization . . . . . . . . . . . . . . . . . . . . . 54
7.4.1 Freezing Point Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.4.1.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.4.1.2 Setting Up The Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.4.1.3 Program Initiation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.5 Test Probe Calibration . . . . . . . . . . . . . . . . . . . . . . . . 57
7.5.1 Calibrating a Single Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.5.2 Furnace Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7.5.2.1 Stabilization and Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8 Furnace Calibration . . . . . . . . . . . . . . . . . . . . . . . 61
8.1 Temperature Profile Adjustment . . . . . . . . . . . . . . . . . . 61
8.1.1 Step 1: Measure the profile . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
8.1.2 Step 2: Adjust the end zones . . . . . . . . . . . . . . . . . . . . . . . . . . 62
8.1.3 Repeat Step 1 and Step 2 if necessary . . . . . . . . . . . . . . . . . . . . . 62
8.2 Temperature Calibration. . . . . . . . . . . . . . . . . . . . . . . 62
8.2.1 One-point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.2.2 Three-point Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.2.3 Compute DELTA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
8.2.4 Computer R0 and Alpha . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
8.2.5 Accuracy and Repeatability. . . . . . . . . . . . . . . . . . . . . . . . . . . 65
9 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
10 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 69
10.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 69
10.1.1 Incorrect Temperature Reading . . . . . . . . . . . . . . . . . . . . . . . . . 69
10.1.2 The unit will not heat or heats at half rate . . . . . . . . . . . . . . . . . . . 69
10.1.3 The unit heats slowly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
10.1.4 If the display flashes any of the following:. . . . . . . . . . . . . . . . . . . 69
iii
Page 6
10.1.5 If the display flashes any of the following:. . . . . . . . . . . . . . . . . . . 70
10.1.6 If the display flashes any of the following:. . . . . . . . . . . . . . . . . . . 70
10.2 CE Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
10.2.1 EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
10.2.2 Low Voltage Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
iv
Page 7

Figures and Tables

Table 1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . 2
Table 2 9260 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Table 3 Mini Cell Specifications . . . . . . . . . . . . . . . . . . . . . . . . 8
Figure 1 Bottom Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Figure 2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Figure 3 Top Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Figure 4 Rear Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Figure 5 Thermal Block Assembly . . . . . . . . . . . . . . . . . . . . . . . 20
Figure 6 Typical Sealed Mini Fixed-Point Cell. . . . . . . . . . . . . . . . . 23
Figure 7 Comparison Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Figure 8 Flow Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Figure 9 Serial Cable Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Table 4 Communications Command Summary . . . . . . . . . . . . . . . . 46
Table 4 Communications Command Summary cont. . . . . . . . . . . . . . 47
Figure 10 Melting Point Realization. . . . . . . . . . . . . . . . . . . . . . . 52
Figure 11 Freezing Point Realization . . . . . . . . . . . . . . . . . . . . . . 55
Figure 12 9260 Comparison Block Heating Up. . . . . . . . . . . . . . . . . 59
Figure 13 9260 Comparison Block Cool Down . . . . . . . . . . . . . . . . . 59
v
Page 8
1 Before You Start
1.1 Introduction
The 9260 is a specialized furnace for the realization of certain defining fixed-points of the International Temperature Scale of 1990 (ITS-90). This fur nace is intended for Calibration Laboratory use and not for field applications. The 9260 permits simplified realization of either freezing or melting curves. In ternal programming of the micro-processor controller provides preprogrammed scan rates, set-points for each step in the process, dwell timing, and indication that the next step is ready. ITS-90 points including Indium, Tin, Zinc, and Alu minum are included. In addition, one other user-defined point may be selected. User defined non-ITS 90 fixed-points are useful for specific applications.
The 9260 furnace is available in 115 VAC (±10%) 60 Hz or 230 VAC(±10%) 50 Hz models.
The 9260 furnace may also be used as a dry-well calibrator or as a temperature comparator. Pre-drilled inserts are available from Hart Scientific for this application.
Built in programmable features include:
Temperature scan rate control
Fixed-point programming
Eight set-point memory
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s hybrid analog/digital con­troller. The controller uses a precision, platinum RTD as a sensor and controls the well temperature with a solid state relay (triac) driven heater.
The LED front display panel continuously shows the current well temperature. The temperature may be easily set with the control buttons to any desired tem perature within the specified range. The furnace’s multiple fault protection de vices insure user and instrument safety and protection.
The 9260 furnace was designed for high accuracy calibrations using compari son measurements or fixed-point calibration methods and for ease of operation. Through proper use, the instrument will continuously provide accurate calibra tion of temperature sensors and devices. The user should be familiar with the safety guidelines and operating procedures of the furnace as described in this user manual.
1 Before You Start
Introduction
-
-
-
-
-
-
-
1
Page 9
9260 Mini Fixed Point Cell Furnace
User’s Guide
1.2 Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
Table 1 International Electrical Symbols
Symbol Description
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union Directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
Canadian Standards Association
2
Page 10
Symbol Description
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re fers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances.
C-TIC Australian EMC Mark
1.3 Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
“WARNING” identifies conditions and actions that may pose hazards to
the user.
“CAUTION” identifies conditions and actions that may damage the in-
strument being used.
1 Before You Start
Safety Information
-
1.3.1
WARNINGS
To avoid personal injury, follow these guidelines.
DO NOT operate this unit without a properly grounded, properly polar-
ized power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
DO USE a ground fault interrupt device.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE INJURY OR DEATH may result if personnel fail to observe safety pre
cautions. Before working inside the equipment, turn power off and dis connect power cord.
HIGH TEMPERATURES PRESENT in this equipment FIRES AND SEVERE BURNS may result if personnel fail to observe safety precau
tions.
DO NOT use this unit in environments other than those listed in the user’s manual.
Continuous use of this equipment at high temperatures for extended peri ods of time requires caution. Completely UNATTENDED HIGH TEM
PERATURE OPERATION IS NOT RECOMMENDED.
Components and heater lifetimes can be shortened by continuous high temperature operation.
The instrument can generate extreme temperatures. Precautions must be taken to prevent personal injury or damage to objects. Probes may be ex tremely hot when removed from the instrument. Cautiously handle probes
-
-
-
-
-
-
3
Page 11
9260 Mini Fixed Point Cell Furnace
User’s Guide
to prevent personal injury. Always use the special comparison block tongs that are supplied with the furnace to remove the comparison block or cell basket. Carefully place probes on a heat resistant surface or rack until they are at room temperature. Never place any objects other than the com parison blocks, cell basket, or cells supplied with the furnace into the well.
Use only grounded AC mains supply of the appropriate voltage to power
the instrument. The furnace requires 11 amps maximum at 115 VAC (±10%),60 Hz and 6 amps maximum at 230VAC (±10%), 50 Hz.
Follow all safety guidelines listed in the user’s manual.
Calibration Equipment should only be used by Trained Personnel.
-
1.3.2
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
Operate the instrument in room temperatures between 5-45°C (41-113°F).
Allow sufficient air circulation by leaving at least 6 inches of space be­tween the furnace and nearby objects. Overhead clearance needs to allow for safe and easy insertion and removal of probes for calibration.
The furnace is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with care. Always carry the unit in an upright position to prevent the compari­son blocks from falling out. Keep the well of the instrument clean and clear of any foreign matter. Do not operate near flammable materials.
DO NOT use fluids to clean out the well.
DO NOT move the furnace with the fixed-point cells inside. They can be
easily broken.
Before initial use, after transport, and anytime the instrument has not been energized for more than 7 days, the instrument needs to be energized for a “dry-out” period of 1-2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-1.
The instrument is equipped with operator accessible system fuses. If a fuse blows, it may be due to a power surge or failure of a component. Re place the fuse once. If the fuse blows a second time, it is likely caused by failure of a component. If this occurs, contact an Authorized Service Center. Always replace the fuse with one of the same rating, voltage, and type. Never replace the fuse with one of a higher current rating.
If a main supply power fluctuation occurs, immediately turn off the in strument. Power bumps from brown-outs and black-outs can damage the instrument. Wait until the power has stabilized before re-energizing the instrument.
-
-
4
Page 12
Hart Scientific Authorized Service Centers
1 Before You Start
1.4 Hart Scientific Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate service on your Hart product:
Hart Scientific, Inc.
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
5
Page 13
9260 Mini Fixed Point Cell Furnace
User’s Guide
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following information available:
Model Number
Serial Number
Voltage
Complete description of the problem
6
Page 14
2 Specifications and Environmental Conditions
2 Specifications and Environmental
Conditions
2.1 Specifications
The 9260 specifications are detailed in Table 2 and the Mini Cell specifications are detailed in Table 3.
Table 2 9260 Specifications
Operating Range 50°C to 680°C (122°F to 1256°F)
Ambient Temperature 5°C to 45°C (41°F to 113°F)
Accuracy ±0.2°C 50°C to 300°C
±0.3°C 300°C to 450°C ±0.5°C 450°C to 680°C
Stability ± 0.03°C to 300°C
± 0.05°C above 300°C
Well-to-Well Gradient ±0.02°C (±0.036°F)
Melting/Freezing Point Duration 6 to 10 hours typical
Vertical Gradients Top and bottom zones adjustable by offset
Resolution 0.01°C or °F
Display Scale °C or °F, switchable
Comparison Block Blank block, two multi-hole blocks, and custom blocks available
Fault Protection Sensor burnout and short protection, over temperature thermal
cut-out
Heater 1200 W maximum - adjustable top and bottom zones
Heating Time Approximately 1.25 hours, 25°C to 680°C
Cooling Time Approximately 10.5 hours, 680°C to 100°C
Stabilization Time 15 minutes nominal
Immersion Depth 229 mm (9”)
Power Requirements 115 VAC (±10%), 60 Hz, 11 amps maximum, 10 amps nominal
230 VAC (±10%), 50 Hz, 6 amps maximum, 5 amps nominal
Exterior Dimensions 250 mm L x 203 mm W x 489 mm H
(10” x 8” x 19.25”)
Weight 20.5 kg with comparison block
(45 lb)
Specifications
7
Page 15
9260 Mini Fixed Point Cell Furnace
User’s Guide
Table 3 Mini Cell Specifications
Model Fixed Point Temperature Six 9’s Cell Five 9’s Cell
5914 Indium 156.5985°C 2 mK 5 mK
5915 Tin 231.928°C 3 mK 5 mK
5916 Zinc 419.527°C 4 mK 7 mK
5917 Aluminum 660.323°C 10 mK 12 mK
2.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou­ble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and clean­ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
Expanded Uncertainty – Using
9260 Furnace
temperature range: 5 - 45°C (41 - 113°F)
ambient relative humidity: 15 - 50%
pressure: 75kPa - 106kPa
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitude does not effect the performance or safety of the unit
2.3 Warranty
Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be free from defects in material and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment. This warranty extends only to the original purchaser and shall not apply to any product which, in Hart’s sole opinion, has been subject to misuse, alteration, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc tions on appropriate Hart products. It is not warranted to be error free.
Hart’s obligation under this warranty is limited to repair or replacement of a product which is returned to Hart within the warranty period and is determined, upon examination by Hart, to be defective. If Hart determines that the defect or malfunction has been caused by misuse, alteration, abuse or abnormal condi
-
-
8
Page 16
2 Specifications and Environmental Conditions
Warranty
tions or operation or handling, Hart will repair the product and bill the pur
-
chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling or writing an Authorized Service Center for authorization. Service Centers as
-
sume NO risk for in-transit damage.
For service or assistance, please contact an Authorized Service Center.
Hart Scientific, Inc.
799 East Utah Valley Drive
American Fork, UT 84003-9775
Phone: (801) 763-1600
Fax: (801) 763-1010
E-mail: support@hartscientific.com
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLU SIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EX
-
­PRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAM­AGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
9
Page 17
3 Quick Start
3.1 Unpacking
Unpack the instrument carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
9260 Furnace
Inner-Melt Heater
Power Cord
Manual
Comparison Blocks (optional)
Cell Basket
Cell Basket Lid
Basket Removal Tool
Thermal Shunt
Top Insulation
Cell Pad Insulation
3 Quick Start
Unpacking
3.2 Setup
Place the furnace on a flat surface with at least 6 inches of free space around and 18 inches above the instrument. Install the power cord into the power entry module on the underside of the furnace. Plug the power cord into a grounded mains outlet. Verify that the nominal voltage corresponds to that indicated on the back of the instrument.
Carefully insert the comparison blocks or cell baskets into the well. (DO NOT drop them into the well.) Comparison block holes should be of the smallest di ameter possible while still allowing the probe to slide in and out easily. Various hole sizes are available from Hart Scientific. The well must be clear of any for eign objects, dirt and grit before the comparison block is inserted. See Section
4.7 for more details.
Turn on the power to the instrument by toggling the switch on the power entry module located underneath the front of the furnace. The fan should begin qui etly blowing air through the instrument and the controller display should illu minate after 3 seconds. After a brief self-test the controller should begin normal operation. If the unit fails to operate please check the power connection.
The display will begin to show the well temperature and the well heater will start operating to bring the temperature of the well to the set-point temperature.
-
-
-
-
11
Page 18
9260 Mini Fixed Point Cell Furnace
User’s Guide
3.3 “Dry-out” Period
Before initial use, after transport, and anytime the instrument has not been en ergized for more than 10 days, the furnace will need to be energized for a “dry-out” period of 1–2 hours before it can be assumed to meet all of the safety requirements of IEC 1010-1.
3.4 Power
Plug the instrument power cord into a mains outlet of the proper voltage, fre quency, and current capability. Typically this will be 11 amps maximum at 115 VAC (±10%), 60 Hz [6 amps maximum at 230 VAC (±10%), 50 Hz]. Turn the instrument on using the “POWER” switch underneath the unit. The instrument will turn on and begin to heat to the previously programmed temperature set-point. The front panel LED display will indicate the actual instrument temperature.
3.5 Setting the Temperature
Section 5.3 explains in detail how to set the temperature set-point on the fur­nace using the front panel keys. The procedure is summarized here.
(1) Press “SET” twice to access the set-point value.
(2) Press “UP” or “DOWN” to change the set-point value.
(3) Press “SET” to program in the new set-point.
(4) Press “EXIT” to return to the temperature display.
When the set-point temperature is changed the controller will switch the well heater on or off to raise or lower the temperature. The displayed well tempera ture will gradually change until it reaches the set-point temperature. The well may require 10 to 75 minutes to reach the set-point depending on the span and the scan rate. Another 15 minutes is required to stabilize within ±0.1°C of the set-point. Ultimate stability may take 15 to 20 minutes of stabilization time.
-
-
-
12
3.6 Changing Display Units
The instrument can display temperature in Celsius or Fahrenheit. The instru ment is shipped from the factory set to Celsius. To change to Fahrenheit or back to Celsius there are two ways:
Press “SET” and “UP” simultaneously. This will change the display units.
or
Press the “SET” key three times from the temperature display to show
Un = C Press the “UP” or “DOWN” key to change units.
-
Page 19
Press “SET” to store changes.
Changing Display Units
3 Quick Start
13
Page 20
4 Parts and Controls
The user should become familiar with the 9260 furnace parts. Successful use of the instrument is dependent upon knowledge of important components and theirproperuse.
4.1 Bottom Panel
The bottom panel consists of the removable power cord inlet, the power entry module (PEM) and power switch, and the fan. See Figure 1.
1. The removable power cord inlet is located underneath the furnace and
plugs into an IEC grounded socket.
4 Parts and Controls
Bottom Panel
Figure 1 Bottom Panel
15
Page 21
9260 Mini Fixed Point Cell Furnace
User’s Guide
2. The power switch is located on the power entry module (PEM). The PEM also houses the fuses. Models are available for either 115 VAC (±10%) 60 Hz or 230 VAC (±10%) 50 Hz operation.
3. The cooling fan inlet is at the bottom of the unit. The cooling air circulat ing through the furnace keeps the electronics and the chassis cool. Keep the area immediately around the furnace clear to allow adequate ventilation.
4. Three feet support the chassis permitting air space for the fan and access to the power entry module and power switch.
4.2 Front Panel
-
The front panel contains the digital display and the controller keypad. See Fig ure 2.
1. The digital display is an important part of the temperature controller be
­cause it not only displays set and actual temperatures but also displays various instrument functions, settings, and constants. The display shows temperatures in units according to the selected scale °C or °F.
2. The four button controller keypad allows easy setting of the set-point temperature. The control buttons (SET, DOWN, UP, and EXIT) are used to set the instrument temperature set-point, access and set other operating parameters, and access and set calibration parameters.
Setting the control temperature is done directly in degrees of the current scale and can be set to 0.01 of a degree Celsius or Fahrenheit.
The functions of the buttons are as follows:
-
16
Figure 2 Front Panel
Page 22
4 Parts and Controls
Front Panel
SET - Used to display the next parameter in the menu and to store parameters to the displayed value.
DOWN - Used to decrement the displayed value of parameters.
UP - Used to increment the displayed value.
EXIT - Used to exit a function and to skip to the next function. Any changes
made to the displayed value are ignored. Holding “EXIT” for about 1/2 a sec ond returns control to the main display.
-
Figure 3 Top Panel
17
Page 23
9260 Mini Fixed Point Cell Furnace
User’s Guide
4.3 Top Panel
The primary feature of the top of the unit is the access to the temperature-con trolled block. The top panel consists of the constant temperature block assem bly, the pre-heat wells, the inner-melt heater, and cooling air vents. See Figure
3.
1. The constant temperature block assembly is where the cell basket con taining the fixed-point cell is inserted. Or, when the furnace is used as a temperature comparator, where the pre-drilled inserts are placed for in serting the thermometers. The block assembly is made of a special alu minum-bronze alloy that is resistant to the temperatures that the furnace is capable of reaching.
Heaters surround the cell in order to provide uniform heat. The 9260 fea tures adjustable top and bottom zone heaters that help to keep the tem perature uniform over the entire fixed-point cell. These zone heaters add heat to each end of the block where more heat is lost to ambient.
A high-temperature platinum resistance thermometer is imbedded into the wall of the block to sense and control the temperature of the block. This entire assembly is insulated and suspended in the airflow of the fan to remove lost heat and to keep the chassis cool.
2. The thermometer pre-heat wells are located on either side of the block access well. Thermometers are pre-heated in these wells prior to inser­tion into the cell in order to conserve its latent energy.
3. The inner-melt heater is a low-power heater that creates an inner liquid layer next to the reentrant tube of the fixed-point cell during the melting curve process. This inner-melt heater is controlled automatically by the microprocessor when using the program mode, or may be used manually with the switch in back. When not in use, the inner-melt heater is stored in the well at the back of the unit to prevent it from causing any damage.
4. The cooling air vents in the top of the unit permit heated air to exit the unit. Care must be taken not to touch these vents while the furnace is at
high temperatures or burns may result.
-
-
-
-
-
-
-
18
CAUTION: Areas on the top of the furnace may be very hot due to hot air
blowing upward. Please use caution.
4.4 Rear Panel
The rear panel consists of the inner-melt heater connector, the inner-melt heater switch, and the serial port. See Figure 4.
1. The inner-melt heater plugs into the rear of the furnace into the connec tor provided. Be sure it is plugged in during operation.
-
Page 24
4 Parts and Controls
Rear Panel
Figure 4 Rear Panel
19
Page 25
9260 Mini Fixed Point Cell Furnace
User’s Guide
20
Figure 5 Thermal Block Assembly
Page 26
2. The inner-melt heater switch can be set to “MANUAL ON” position or “AUTO” position. The microprocessor has control when the switch is in the “AUTO” position.
3. The serial port is a DB-9 male connector for interfacing the instrument to a computer or terminal RS-232 communications.
4.5 Thermal Block Assembly
The thermal block assembly is shown in Figure 5 and described below.
4.5.1 Thermal Block
The thermal block is specifically designed to contain the sealed cell and basket containment assembly. This design permits the uniformity to be tuned carefully for best performance and cell safety. Heaters in the perimeter of the alumi num-bronze cylinder provide heat as dictated by the temperature controller. A PRT sensor in the block monitors the block temperature and provides feedback to the controller. Forced airflow around the assembly keeps the outside of the furnace cool.
4.5.2 Heaters
4 Parts and Controls
Thermal Block Assembly
-
The block assembly is heated by up to 1200 watts of heat. There are 3 heated zones; the main zone heaters heat the entire length of the block and the top and bottom zones heat their respective ends. The end zone heating compensates for losses out of the unheated ends. These heaters are adjusted to provide the re­quired temperature uniformity within the cell itself. The adjustments are made by way of the keypad.
4.5.3 Basket
The sealed cell is positioned within the basket which allows the cell to be easily inserted and removed and contains all materials should the cell be broken. The top and bottom parts of the basket are made of aluminum-bronze and provide thermal shunting across the top and bottom as well. The lid and basket are re moved separately with the tongs provided. A cell-pad in the bottom of the bas ket helps to cushion the cell from the metal bottom.
4.5.4 Thermal Shunt
A disk of aluminum-bronze in the top of the thermal well above the basket pro vides heat transfer across the top of the cell and to the thermometer itself. The effectiveness of the heat transfer to the thermometer is dependent on its fit to the shunt. Measurements at the aluminum point may be affected by heat con ducted up the thermometer stem.
-
-
-
-
21
Page 27
9260 Mini Fixed Point Cell Furnace
User’s Guide
4.5.5 Insulation
The entire block assembly is surrounded by fiber-ceramic insulation. A remov able portion above the cell permits the cell to be inserted and removed.
4.5.6 Temperature Control Sensor
The temperature control sensor is a high quality PRT with 4 leads. Accuracy is calibrated into the unit. Zero resistance, alpha, and delta coefficients of the Cal endar-Van Dusen equation permit linearization over the desired temperature range.
4.6 Mini Fixed-point Cells
Mini Fixed-point Cells (Figure 6) utilize physical properties of a substance to provide well established temperatures. The sample in the cell is placed into a condition of multiple phases at a melting or freezing temperature or at a tri ple-point temperature. While the sample substances are in this condition they can exhibit very stable constant temperatures for long periods of time. Properly used, the temperatures provided by these constants of nature are extremely pre­cise and repeatable. The International Temperature Scale of 1990 (ITS-90) is based on these principles. The ITS-90 temperatures defined at the freezing points of Indium, Tin, Zinc, and Aluminum are among these and are achievable with the Hart Scientific 9260 Furnace.
In order to achieve the ITS-90 temperatures and maintain long flat plateaus, the substance samples (metals in this case) must be very pure. Typically 6 nines purity is best. Sometimes 5 nines purity is used for the lower price but at the cost of higher uncertainty and shorter plateaus. The 9260 furnace utilizes a sealed mini-cell construction. The construction of the mini-cell follows the pat­tern of the full size cell. The high-purity sample is contained within a graphite crucible. The graphite is free of contaminants and will not react with the metal maintaining the metal purity. This material is all hermetically sealed within a silica glass (quartz) envelope. The internal atmosphere is high purity argon. In order to immerse the test thermometer into the high accuracy temperature zone of the cell, a reentrant well is provided in the center of the cell.
These cells are manufactured in the same careful manner as their larger full size counterparts. The certified high purity metal sample must maintain its purity throughout the process of manufacture and use of the cell. To accomplish this, high purity materials must be used for all the other components of the cell. Af ter fabrication of each component, it must be treated to remove any impurities that may have been added during the process. The components are assembled in a clean environment and never touched directly by hand. After the compo nents have been assembled and the silica glass permanently sealed, all of the air is evacuated out while the cell is melted. Numerous cycles of vacuum and purg ing with high purity argon are finally completed when the evacuation port is sealed leaving approximately one atmosphere pressure of argon in the cell at the freezing point of the sample.
-
-
-
-
-
-
22
Page 28
4 Parts and Controls
Mini Fixed-point Cells
Figure 6 Typical Sealed Mini Fixed-Point Cell
23
Page 29
9260 Mini Fixed Point Cell Furnace
User’s Guide
Temperature corrections must be made to the reading to account for actual cell pressure and for hydrostatic pressure. During manufacturing, the cell is sealed with argon near the 1 atmosphere pressure. At that time the actual pressure is measured. With that pressure a small temperature correction can be calculated.
The hydrostatic pressure, created by the mass of the sample itself, depresses the temperature of the reading. Since different cell designs and thermometer de signs translate to different immersion depths, the practice is to calculate and correct for the error. Figure 6 illustrates the maximum immersion depth within the mini-cell. The actual immersion depth is taken to the center of the sensor el ement of the thermometer. This depth will vary and the thermometer manufac turer may need to be consulted. Approximations can be made for typical types of thermometers since the hydrostatic error is small anyway and may be negli gible for some requirements. Refer to the cell manual for the equations and constants that need to be applied.
Due to the fragile nature of the fixed-point cells, extra care must be taken dur ing use and handling. Do not handle it with bare hands, use clean cotton gloves or equivalent. Make sure anything that comes in contact with the cell is clean. To remove contaminants, wipe the cell down with a clean cloth and pure alco­hol. Quartz glass is subject to a process called devitrification. The glass will break down at high temperatures during this process. Oils in the skin and other contaminants can initiate or accelerate this process.
Contaminants introduced to the reentrant well of the cell from unclean ther­mometers can cause the same problem. In addition, some types of metals can contaminate the platinum sensor in a quartz SPRT at high temperatures (650°C and up). Clean all thermometers prior to testing.
-
-
-
-
-
24
4.7 Comparison Blocks
The 9260 furnace can function either as a calibrator or as a comparator. As a calibrator, the calibration of the controller provides the reference temperature. As a comparator, a reference thermometer value is compared to the values of the units under test. A smaller uncertainty is obtainable with the comparison method. Comparison blocks are available as options to the furnace.
4.7.1 Block Well Sizes
Three standard comparison blocks are available. See Figure 7. Model 3160-2-provides 9 wells with clearance for 1/4-inch diameter thermometers. Model 3160-3 is a combination of wells providing access for a variety of popu lar sizes. Model 3160-1 is a blank block that can be drilled by the user to any desired sizes.
4.7.2 Comparison Block Assembly
Comparison blocks provide a uniform temperature between multiple thermom eters. For accurate results, the thermometers must fit closely inside the well.
-
-
Page 30
4 Parts and Controls
Comparison Blocks
Figure 7 Comparison Blocks
25
Page 31
9260 Mini Fixed Point Cell Furnace
User’s Guide
The comparison block assembly is comprised of three components, the com parison block, the thermal shunt, and the top insulation.
4.7.2.1 Comparison Block
The comparison block is carefully lowered to the bottom of the thermal block well with the tongs provided. DO NOT drop the block into the well. Damage
to the furnace may result. Small changes to the furnace calibration may re sult as well.
4.7.2.2 Thermal Shunt
Just above the comparison block is a ledge in the thermal block itself. The ther mal shunt is lowered into the thermal well until it rests on this ledge. The inser tion wells must match the comparison block. This block conducts heat from the thermal block of the furnace in order to reduce stem conduction along the ther mometer. Heat loss from stem conduction will reduce the accuracy of the measurement.
4.7.2.3 Top Insulation
A fiber ceramic pad of insulation is provided with the comparison block. This insulation helps the thermal shunt by keeping it closer to the furnace tempera­ture. After it is inserted, poke holes through it to match the comparison block.
-
-
-
-
-
26
Page 32
5 Controller Operation
5 Controller Operation
Well Temperature
This section discusses in detail how to operate the furnace temperature control ler using the front control panel. By using the front panel key-switches and LED display, the user may monitor the well temperature, adjust the set-point temperature in degrees C or F, monitor the heater output power, adjust the con troller proportional band, and program the operating parameters, program pa rameters, serial interface configuration, and the controller calibration parameters. Operation of the functions and parameters is shown in the flowchart in Figure 8 on page 28. This chart may be copied for reference.
In the following discussion a button with the word SET, UP, DOWN, or EXIT inside indicates the panel button while the dotted box indicates the display reading. Explanation of the button or display reading are to the right of each button or display value.
5.1 Well Temperature
The digital LED display on the front panel allows direct viewing of the actual well temperature. This temperature value is normally shown on the display. The units, C or F, of the temperature value are displayed at the right. For example,
100.00 C Well temperature in degrees Celsius
The temperature display function may be accessed from any other function by pressing the “EXIT” button.
5.2 Reset Cut-out
-
-
-
If the over-temperature cut-out has been triggered then the temperature display will alternately flash,
Cut-out Indicates cut-out condition
The message continues to flash until the temperature is reduced and the cut-out is reset. The cut-out has two modes - automatic reset and manual reset. The mode determines how the cut-out is reset which allows the instrument to heat up again. When in automatic mode, the cut-out will reset itself as soon as the temperature is lowered below the cut-out set-point. With manual reset mode the cut-out must be reset by the operator after the temperature falls below the set-point.
When the cut-out is active and the cut-out mode is set to manual (“reset”) then the display will flash “cut-out” until the user resets the cut-out. To access the reset cut-out function press the “SET” button.
S
Access cut-out reset function
27
Page 33
9260 Mini Fixed Point Cell Furnace
User’s Guide
Figure 8 Flow Chart
28
Page 34
The display will indicate the reset function.
rESEt ? Cut-out reset function
Press “SET” once more to reset the cut-out.
5 Controller Operation
Temperature Set-point
S
This will also switch the display to the set temperature function. To return to displaying the temperature press the “EXIT” button. If the cut-out is still in the over-temperature fault condition the display will continue to flash “cut-out”. The well temperature must drop a few degrees below the cut-out set-point be fore the cut-out can be reset.
Reset cut-out
5.3 Temperature Set-point
The temperature set-point can be set to any value within the range and resolu tion as given in the specifications. Be careful not to exceed the safe upper tem­perature limit of any device inserted into the well.
Setting the temperature involves selecting one of the eight (8) set-points in memory and then adjusting the set-point value.
5.3.1 Programmable Set-points
The controller stores eight (8) set-point temperatures in memory. The set-points can be quickly recalled to conveniently set the instrument to a previously pro­grammed temperature set-point.
To set the temperature, first select the set-point memory. This function is ac­cessed from the temperature display function by pressing “SET”. The number of the set-point memory currently being used is shown at the left on the display followed by the current set-point value.
-
-
100.00 C Well temperature in degrees Celsius
S
Access set-point memory
1. 100. Set-point memory 1, 100.0°C currently used
To change the set-point memory to another preset value press “UP” or “DOWN”.
4. 300. New set-point memory 4, 300.0°C
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
29
Page 35
9260 Mini Fixed Point Cell Furnace
User’s Guide
NOTE: Pressing “SET” at this point turns off the program mode if it is
on.
5.3.2 Set-point Value
The set-point value may be adjusted after selecting the set-point memory and pressing “SET”.
4 200. Set-point value in °C
If the set-point value is correct, hold “EXIT” to resume displaying the well temperature. Press “UP” or “DOWN” to adjust the set-point value.
220.00 New set-point value
When the desired set-point value is reached press “SET” to accept the new value and to access the temperature scale units. If “EXIT” is pressed, any changes made to the set-point are ignored.
S
Accept new set-point value
5.4 Temperature Scale Units
The temperature scale units of the controller can be set by the user to degrees Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well tem­perature, set-point, and proportional band.
Press “SET” after adjusting the set-point value to change display units.
Un= C Scale units currently selected
Press “UP” or “DOWN” to change the units.
Un= F New units selected
5.5 Scan
The scan rate can be set and enabled so that when the set-point is changed the furnace heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the furnace heats or cools at the maxi mum possible rate.
5.5.1 Scan Control
-
30
The scan is controlled with the scan on/off function that appears in the main menu after the set-point function.
Page 36
Sc=OFF Scan function off
Press “UP” or “DOWN” to toggle the scan on or off.
Sc=On Scan function on
Press “SET” to accept the present setting and continue.
5 Controller Operation
Program Advance
S
Accept scan setting
5.5.2 Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set from 0.1 to 99.9°C/minute. The maximum scan rate, however, is actually lim ited by the natural heating or cooling rate of the instrument. This is often less than 100 °C/minute, especially when cooling.
The scan rate function appears in the main menu after the scan control function. The scan rate units are in degrees per minute, degrees C or F depending on the selected units.
Sr= 10.0 Scan rate in °C/min.
Press“UP”or“DOWN”tochangethescanrate.
Sr= 2.0 New scan rate
Press “SET” to accept the new scan rate and continue.
S
Accept scan rate
5.6 Program Advance
The program advance function allows the user to step through the maintain, freeze, and melt operations of the fixed-point realization. They are explained in detail in Section 8, 7 on Fixed-Point Realization.
-
S+D
Access program advance
Adv “Adv” flashes
MAINT Displays one of the functions MAINT, FREEZE, MELT, or
STOP
Press “UP” or “DOWN” to view the desired function.
S
Accepts the new the operation
31
Page 37
9260 Mini Fixed Point Cell Furnace
User’s Guide
5.7 Temperature Scale Units
To toggle between °C and °F, press the “SET” and “UP” keys simultaneously when the temperature is displayed.
5.8 Secondary Menu
Functions which are used less often are accessed within the secondary menu. Pressing “SET” and “EXIT” simultaneously and then releasing accesses the secondary menu. The first function in the secondary menu is the heater power display. (See Figure 8 on page 28.)
5.9 Heater Power
The temperature controller controls the temperature of the well by pulsing the heater on and off. The total power being applied to the heater is determined by the duty cycle or the ratio of heater on time to the pulse cycle time. By knowing the amount of heating the user can tell if the instrument is heating up to the set-point, cooling down, or controlling at a constant temperature. Monitoring the percent heater power lets the user know the stability of the well tempera­ture. With good control stability the percent heating power should not fluctuate more than ±1% within one minute.
The heater power display is accessed in the secondary menu. Press “SET” and “EXIT” simultaneously and release. The heater power is displayed as a percent­age of full power.
32
100.00 Well temperature
S+E
Access heater power in secondary menu
SEC Flashes
12.0 P Heater power in percent
To exit out of the secondary menu press “EXIT” and hold for a brief moment. To continue on to the proportional band setting function press “EXIT” momen tarily or “SET”.
5.10 Set-point Resistance
The set-point resistance is the resistance of the temperature sensor at the cur rent temperature. Allow the temperature to stabilize at the desired set-point be fore taking its resistance. In order to calibrate the furnace temperature, the set-point resistance must be displayed.
-
-
-
Page 38
5 Controller Operation
Proportional Band
Press “SET” and “EXIT” to enter the secondary menu and show the heater power. Then press “SET” twice to access the set-point resistance
S+E
Access heater power in secondary menu
SEC Flashes “ ” and then displays the heater power setting
12.0 P Heater power in percent
S
Access set-point resistance
rS Flashes “ ” (Set-point Resistance) and then displays the
setting
160.095 Resistance in ohms
5.11 Proportional Band
In a proportional controller such as this, the heater output power is proportional to the well temperature over a limited range of temperatures around the set-point. This range of temperature is called proportional band. At the bottom of the proportional band the heater output is 100%. At the top of the propor­tional band the heater output is 0. Thus as the temperature rises the heater power is reduced, which consequently tends to lower the temperature back down. In this way the temperature is maintained at a fairly constant temperature.
The temperature stability of the well and response time depend on the width of the proportional band. If the band is too wide the well temperature deviates ex cessively from the set-point due to varying external conditions. This deviation is because the power output changes very little with temperature and the con troller does not respond well to changing conditions or noise in the system. If the proportional band is too narrow the temperature may swing back and forth because the controller overreacts to temperature variations. For best control sta bility the proportional band must be set for the optimum width.
The proportional band width is set at the factory to about 5.0°C. The propor tional band width may be altered by the user to optimize the control character istics for a particular application.
The proportional band width is easily adjusted from the front panel. The width may be set to discrete values in degrees C or F depending on the selected units. The proportional band adjustment can be accessed within the secondary menu. Press “SET” and “EXIT” to enter the secondary menu and show the heater power. Then press “SET” twice to access the proportional band.
-
-
-
-
-
S+E
Access heater power in secondary menu
33
Page 39
9260 Mini Fixed Point Cell Furnace
User’s Guide
SEC Flashes “ ” and then displays the heater power setting
12.0 P Heater power in percent
S
Access set-point resistance
rS Flashes “ ” (Set-point Resistance) and then displays the
setting
160.095 Resistance in ohms
S
ProP Flashes “ ” and then displays the setting
5.0 Proportional band setting
To change the proportional band press “UP” and “DOWN”.
4.0 New proportional band setting
To store the new setting press “SET”. Press “EXIT” to continue without storing the new value.
S
5.12 Controller Configuration
The controller has a number of configuration and operating options and calibra tion parameters that are programmable via the front panel. These are accessed from the secondary menu after the proportional band function by pressing “SET”. “ConFiG” flashes and then the name of the first parameter menu “PAR” is displayed.Pressing “SET” again enters the first of four groups of con figuration parameters: operating parameters, program parameters, serial inter face parameters, and calibration parameters. The groups are selected using the “UP” and “DOWN” keys and then pressing “SET”. (See Figure 8 on page 28)
Accept the new proportional band setting
-
-
-
34
5.13 Operating Parameters
The operating parameters menu is indicated by,
PAr Operating parameters menu
Page 40
Press “SET” to enter the menu. The operating parameters menu contains the High Limit (HL) parameter, the Soft Cut-out parameter, and the Cut-out Reset mode parameter.
5.13.1 High Limit
The High Limit parameter adjusts the upper set-point temperature. The factory default and maximum are set to 680°C. For safety, a user can adjust the High Limit parameter down so the maximum temperature set-point is restricted.
HL Flashes “ ” (High Limit parameter) and then displays
H=680 Current HL setting
Adjust the HL parameter using “UP” or “DOWN”
H=600 New High Limit setting
Press “SET” to accept the new High Limit parameter and to access the Soft Cut-out parameter.
5 Controller Operation
Operating Parameters
the setting
5.13.2 Soft Cut-out
The next parameter in this menu is the Soft Cut-out. The Soft Cut-out parame­ter is used by the controller to shut down the unit during over-temperature con­ditions. If the temperature of the unit is ever greater than the Soft Cut-out temperature the controller shuts itself down and displays, alternately, “SCtOut”and“Err 8".
SoFtCo Flashes “ ” (Soft Cut-out parameter) and then dis
plays the setting
705 Current value
Adjust this parameter by using “UP” or “DOWN”.
700 New Soft Cut-out setting
Press “SET” to accept the new parameter and to access the Cut-out Reset
Mode.
5.13.3 Cut-out Reset Mode
The final parameter in this menu is the Cut-out Reset Mode. The Cut-out Reset Mode determines whether the cut-out resets automatically when the well tem perature drops to a safe value or must be manually reset by the operator.
-
-
35
Page 41
9260 Mini Fixed Point Cell Furnace
User’s Guide
CtorSt Flashes “ ” (Cut-out reset mode parameter) and
Auto Current setting
To change to manual reset mode press “UP” or “DOWN”.
rSt New Cut-out reset for manual reset
Press “SET” to accept the new parameter.
5.14 Program Parameters
The program parameters menu is indicated by,
Prog Program parameters menu
Press “SET” to enter the menu. The Program parameters menu contains the fixed-point parameter, the curve parameter, and the curve temperature parameter.
then displays the setting
5.14.1 Fixed-point
The first parameter in this menu is the Fixed-point parameter.The Fixed-point parameter allows the user to select the fixed-point metal. The available options are In (Induim), Sn (Tin), Zn (Zinc), Al (Aluminum), or Other.
FP Flashes “ ” (Fixed-point parameter) and then displays
In Current Fixed-point setting (Indium)
Adjust the Fixed-point parameter by using “UP” or “DOWN”.
2n New Fixed-point setting (Zinc)
Press “SET” to accept the new Fixed-point parameter and to access the Curve parameter.
5.14.2 Curve
The next parameter in this menu is the Curve parameter. The Curve parameter is selected as either melting or freezing.
Curve Flashes “ ” (Curve parameter) and then displays
the setting
the setting
36
Page 42
MeLt Current Curve setting
Adjust this parameter by using “UP” or “DOWN”.
FrEE2E New Curve setting
5 Controller Operation
Program Parameters
Press “SET” to accept the new Curve parameter and to access the Curve Tem perature parameter.
5.14.3 Curve Temperature
The final set of three parameters in this menu are the Curve Temperature pa rameters. These parameters set the temperature of the curve set-points.
TemP Flashes “ ” (Curve Temperature parameter) and then
displays the melt parameter
MELt Flashes “ ” and then displays the value
100.00 Current melt temperature setting
Adjust this parameter by using “UP” or “DOWN”. Press “SET” to accept the new value and to display the freeze parameter.
S
FREEZE Flashes “ ” and then displays the value
25.00 Current freeze temperature setting
Adjust this set-point by using “UP” or “DOWN”. Press “SET” to accept the new value and to display the maintain parameter
-
-
S
MAint Flashes “ ” and then displays the value
148.59 Current maintain temperature setting.
Adjust this set-point by using “UP” or “DOWN”.
Press “SET” to accept the new value.
37
Page 43
9260 Mini Fixed Point Cell Furnace
User’s Guide
5.15 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SEriAL Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the operation of the serial interface. These controls only apply to instruments fitted with the serial interface. The parameters in the menu are: BAUD rate, sample period, duplex mode, and linefeed.
5.15.1 Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting deter mines the serial communications transmission rate. The baud rate of the serial communications may be programmed to 300, 600, 1200, 2400, 4800, or 9600 baud. 2400 baud is the default setting.
bAUd Flashes “ ” (Serial baud rate parameter) and then
2400 b Current baud rate
Adjust the baud rate by using “UP” or “DOWN”.
4800 b New baud rate
Press “SET” to store the baud rate to the new value and to access the Sample Period.
5.15.2 Sample Period
The sample period is the next parameter in the serial interface parameter menu. The sample period is the time period in seconds between temperature measure ments transmitted from the serial interface. If the sample rate is set to 5, the in strument transmits the current measurement over the serial interface approximately every five seconds. The automatic sampling is disabled with a sample period of 0.
-
displays the setting
-
-
38
SPEr Flashes ” ” (Serial sample period parameter) and
then displays the setting
SP=1 Current sample period (seconds)
Adjust the value with “UP” or “DOWN”.
SP=60 New sample period
Page 44
5 Controller Operation
Calibration Parameters
Press “SET” to store the sample period to the new value and to access the Du plex Mode.
5.15.3 Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du plex or half duplex. With full duplex any commands received by the instrument via the serial interface are immediately echoed or transmitted back to the device of origin. With half duplex the commands are executed but not echoed.
dUPL Flashes “ ” (Serial duplex mode parameter) and then
d=FULL Current duplex mode setting
The mode may be changed using “UP” or DOWN".
d=HALF New duplex mode setting
Press “SET” to store the duplex mode to the new value and to access the the Linefeed.
5.15.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa­rameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return.
-
-
displays the setting
LF Flashes “ ” (Serial linefeed parameter) and then dis-
plays the setting
LF=On Current linefeed setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF=OFF New linefeed setting
Press “SET” to store the new linefeed value.
5.16 Calibration Parameters
The operator of the instrument has access to a number of the calibration con stants namely R0, ALPHA, DELTA, top and bottom zone percent heat, and the hard cut-out. These values are set at the factory and must not be altered. The correct values are important to the accuracy and proper and safe operation of the furnace. Access to these parameters is available to the user only so that in the event that the controller memory fails the user may restore these values to
-
39
Page 45
9260 Mini Fixed Point Cell Furnace
User’s Guide
the factory settings. The user should have a list of these constants and their set tings with manual.
CAUTION: DO NOT change the values of the instrument calibration con
stants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the instrument.
The calibration parameters menu is indicated by,
CAL Calibration parameters menu
Press “SET” five times to enter the menu.
The calibration parameters R0, ALPHA, and DELTA characterize the resis tance-temperature relationship of the platinum control sensor. These parameters may be adjusted by an experienced user to improve the accuracy of the furnace.
CAUTION: Care should be exercised when adjusting these parameters
since they affect the accuracy of the set-point value. This procedure is ex­plained in detail in Section 8.
5.16.1 Hard Cut-out
This parameter is the temperature above which the unit shuts down automati­cally. The parameter is set at the factory to approximately 700°C and cannot be changed by the user.
-
-
-
40
5.16.2 R
0
This probe parameter refers to the resistance of the control probe at 0°C. The value of this parameter is set at the factory for best instrument accuracy.
5.16.3 ALPHA
This probe parameter refers to the average sensitivity of the probe between 0 and 100°C. The value of this parameter is set at the factory for best instrument accuracy.
5.16.4 DELTA
This probe parameter characterizes the curvature of the resistance-temperature relationship of the sensor. The value of this parameter is set at the factory for best instrument accuracy. should have a list of these constants and their settings with manual.
5.16.5 Top and Bottom Zone Percent Heating
The top and bottom zone heaters of the furnace start heating each time the main zone heater turns on. These end zone heaters are adjusted to a percentage of on
Page 46
5 Controller Operation
Calibration Parameters
time compared to the main zone. Each zone is pre-adjusted according to a cali bration procedure that provides a low vertical temperature gradient in the ther mal block.
5.16.5.1 Bottom Zone
botPCt is the percentage of the main heater value for the bottom zone. Do not adjust this value unless you are following the procedure in Section 8 Calibra tion Procedure, of this manual. Adjustment values range from 0 to 200 percent.
5.16.5.2 Top Zone
toPPCt is the percentage of the main heater value for the top zone. Do not ad just this value unless you are following the procedure in Section 8 Calibration Procedure, of this manual. Adjustment values range from 0 to 200 percent.
-
-
-
-
41
Page 47
6 Digital Communication Interface
6 Digital Communication Interface
The furnace is capable of communicatingwithandbeingcontrolledbyother equipment through the digital serial interface.
With a digital interface the instrument may be connected to a computer or other equipment. This allows the user to set the set-point temperature, monitor the temperature, and access any of the other controller functions, all using remote communications equipment. Communications commands are summarized in Table 4 on page 46.
6.1 Serial Communications
The instrument is installed with an RS-232 serial interface that allows serial digital communications over fairly long distances. With the serial interface the user may access any of the functions, parameters and settings discussed in Sec tion 5 with the exception of the BAUD rate setting.
6.1.1 Wiring
The serial communications ca­ble attaches to the instrument through the DB-9 connector at the back of the instrument. Fig­ure 9 shows the pin-out of this connector and suggested cable wiring. To eliminate noise the serial cable should be shielded with low resistance between the connector (DB-9) and the shield. If the unit is used in a heavy industrial setting, the se rial cable must be limited to ONE METER in length.
-
Serial Communications
-
6.1.2 Setup
Before operation the serial in terface must first be set up by programming the BAUD rate and other configuration parame ters. These parameters are pro grammed within the serial interface menu.
To enter the serial parameter programming mode first press “EXIT” while pressing “SET”
-
-
-
Figure 9 Serial Cable Wiring
43
Page 48
9260 Mini Fixed Point Cell Furnace
User’s Guide
and release to enter the secondary menu. Press “SET” repeatedly until the dis play reads “PAr”. Press “UP” until the serial interface menu is indicated with “SErIAL”. Finally press “SET” to enter the serial parameter menu. In the serial interface parameters menu are the BAUD rate, the sample rate, the duplex mode, and the linefeed parameter.
6.1.2.1 Baud Rate
The baud rate is the first parameter in the menu. The display will prompt with the baud rate parameter by showing “bAUd”. Press “SET” to choose to set the baud rate. The current baud rate value will then be displayed. The baud rate of the 9260 serial communications may be programmed to 300, 600, 1200, 2400, 4800, or 9600 baud. The baud rate is pre-programmed to 2400 baud. Use “UP” or “DOWN” to change the baud rate value. Press “SET” to set the baud rate to the new value or “EXIT” to abort the operation and skip to the next parameter in the menu.
6.1.2.2 Sample Period
The sample period is the next parameter in the menu and prompted with “SPEr”. The sample period is the time period in seconds between temperature measurements transmitted from the serial interface. If the sample rate is set to 5, the instrument transmits the current measurement over the serial interface ap­proximately every five seconds. The automatic sampling is disabled with a sample period of 0. Press “SET” to choose to set the sample period. Adjust the period with “UP” or “DOWN” and then use “SET” to set the sample rate to the displayed value.
-
44
6.1.2.3 Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full duplex any commands received by the instrument via the serial interface are immediately echoed or transmitted back to the device of origin. With half du plex the commands are executed but not echoed. The default setting is full du plex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
6.1.2.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac ter (LF, ASCII 10) after transmission of any carriage-return. The default setting is with linefeed on. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
6.1.3 Serial Operation
Once the cable has been attached and the interface set up properly the control ler immediately begins transmitting temperature readings at the programmed rate. The serial communications uses 8 data bits, one stop bit, and no parity.
-
-
-
-
-
Page 49
The set-point and other commands may be sent via the serial interface to set the temperature set-point and view or program the various parameters. The inter face commands are discussed in Section 6.2. All commands are ASCII charac ter strings terminated with a carriage-return character (CR, ASCII 13).
6.2 Interface Commands
6 Digital Communication Interface
Interface Commands
-
-
The various commands for accessing the instrument functions via the digital in terface are listed in this section (see Table 4). These commands are used with the RS-232 serial interface. The commands are terminated with a car
­riage-return character. The interface makes no distinction between upper and lower case letters, hence either may be used. Commands may be abbreviated to the minimum number of letters that determines a unique command. A com
­mand may be used to either set a parameter or display a parameter depending on whether or not a value is sent with the command following a “=” character. For example, “s” <cr> will return the current set-point and “s=150.0" will set the set-point to 150.0 degrees.
In the following list of commands, characters or data within brackets, “[” and “]”, are optional for the command. A slash, “/”, denotes alternate characters or data. Numeric data, denoted by “n”, may be entered in decimal or exponential notation. Characters are shown in lower case although upper case may be used. Spaces may be added within command strings and will simply be ignored. Backspace (BS, ASCII 8) may be used to erase the previous character. A termi­nating CR is implied with all commands.
-
45
Page 50
9260 Mini Fixed Point Cell Furnace
User’s Guide
Table 4 Communications Command Summary
Command Description
Command Format
Command Example Returned
Returned Example
Acceptable Values
Display Temperature
Read units u u u:{C or F} u: C
Read current set-point s[etpoint] s set: 9999.99 {C or F} set: 150.00 C
Set current set-point to n s[etpoint]=n s=450 Instrument Range
Read temperature t t t: 9999.99{C or F} t: 478.03 C
Set temperature units: u[nits]=c/f
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to Fahrenheit u[nits]=f u=f
Read scan function sc[an] sc scan: {ON or OFF} scan: ON
Set scan function: sc[an]=on/of[f]
Turn scan function on sc[an]=on sc=on
Turn scan function off sc[an]=of[f] sc=of
Read scan rate sr[ate] sr srat: 999.99 {C or F}/min srat: 10.0 C/min
Setscanrateto
Read program control pc pc pc:{[STOP/FREEZE/ME[LT]/MA[
Set program control pc={stop/freeze/me[lt]/ma[intain]}pc=melt STOP, FREEZE,
n
degrees per minute sr[ate]=
n
sr=5 .1 to 99.9
INTAIN]}
pc:FREEZE
CorF
ON or OFF
MELT or MAINTAIN
Secondary Menu
Read heater power
(duty cycle)
Read proportional band setting pr[op-band] pr pb: 999.9 pb: 15.9
Set proportional band to
Read set-point resistance *sr *sr 999.999 121.091
n
po[wer] po po: 999.9 po: 1
pr[op-band]=n pr=8.83 0.1 to 100
Configuration Menu
Operating Parameters Menu
Read high limit hl hl hl: 9999 hl: 925
Set high limit hl=n hl=900 100–1200
Read soft cut-out cu cu cu: 9999.9 cu: 1150
Set soft cut-out setting: cu[tout]=n
Set soft cut-out to
Read cut-out mode cm[ode] cm cm:{xxxx} cm: AUTO
Set cut-out mode cm[ode]=r[eset]/a[uto] Reset or Auto
Set cut-out to be reset manually cm[ode]=r[eset] cm=r
Set cut-out to be reset automatically cm[ode]=a[uto] cm=a
Program Parameters Menu
Read fixed-point parameter fp fp fp:{INDIUM, TIN,
n
degrees cu[tout]=n cu=500 0.0 to 1150.0
ZINC,ALUMINUM}
fp:INDIUM
46
Page 51
Communications Command Summary cont.
6 Digital Communication Interface
Interface Commands
Command Description
Set fixed-point parameter fp=[INDIUM/TIN/ZINC/
Read curve parameter crv crv crv:{MELT or FREEZE} crv:MELT
Set curve parameter crv={melt/freeze} crv=melt MELT or FREEZE
Read curve temperatures ps
Set curve temperatures ps
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 9 sa: 1
Set serial sampling setting to
Set serial duplex mode: du[plex]=f[ull]/h[alf]
Set serial duplex mode to full du[plex]=f[ull] du=f
Set serial duplex mode to half du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f]
Set serial linefeed mode to on lf[eed]=on lf=on
Set serial linefeed mode to off lf[eed]=of[f] lf=of
Cal Menu
Read R
calibration parameter r[0] r r0: 999.999 r0: 100.7
0
Set R
calibration parameter to
0
Read Alpha calibration parameter al[pha] al al: 9.999999 al: 0.003865
Set Alpha calibration parameter to
Read Delta calibration parameter de[lta] de de: 9.99 de: 1.50
Set Delta calibration parameter de[lta]=
Read top zone % heating tpct tpct tpct: 999.9 tpct: 200.0
Set top zone % heating tpct=
Read bottom zone % heating bpct bpct bpct: 999.9 bpct: 150.0
Set bottom zone % heating bpct=
These commands are only used for factory testing.
Miscellaneous (not on menus)
Read firmware version number *ver[sion] *ver ver.9999,9.99 ver.9260,v1.1
Read structure of all commands h[elp] h list of commands
Legend: [] Optional Command data
Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return and linefeed.
n
seconds sa[mple]=
n
Format
ALUMINUM]
n
n=n
n
r[0]=
n
n
al[pha]=
n
n
n
n
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Then the value is returned as indicated in the RETURNED column.
Command
Command Example Returned
fp=In INDIUM, TIN, ZINC,
ps3 psn: 999.99{C or F} ps1: 480.00 C
ps3=100 1 to 3, dependent on
sa=0 0 to 4000
r=100.7 98.0 to 104.9
al=0.003865 .002 to .006
de=1.37 0to3
tpct=100 0to200
bpct=150.0 0to200
Returned Example
Acceptable Values
ALUMINUM
program control
FULL or HALF
ON or OFF
47
Page 52
7 Fixed-Point Realization
7.1 General
Either a freezing or melting plateau may be realized. The melting point is faster and easier than the freezing point method and the plateau can last longer. Prop erly done, accuracy of measurements is nearly the same.
Realizing the fixed-point temperature is a matter of achieving the ITS-90 de fined temperature through a careful process. The Hart Scientific Model 9260 furnace features an internally programmed method or a manual method to per form this process.
The internal programming procedure requires the user to:
1. Select the point (Indium, Tin, Zinc, Aluminum or a user programmed
point) to be realized.
2. Select either the freezing curve or melting curve mode.
3. The controller then provides pre-programmed scan rates, set-points for
each step in the process, dwell times, and an indication that the next step is ready. Some of these pre-programmed conditions can be modified by the user.
Temperature calibration should be checked occasionally to verify the set-point temperatures.
7 Fixed-Point Realization
General
-
-
-
7.2 Installing a Sealed Cell into the Basket
The sealed cell is a very delicate instrument. The quartz glass on the outside of the cell can be easily broken and easily scratched by the harder metal materials into which it is inserted.
CAUTION: Sealed cells are designed to be kept upright. They should be
stored in this position as well. Before inserting the cell be sure it is clean and free of finger oils. Use pure or reagent grade alcohol to clean all pieces. Quartz glass will devitrify breaking down the glass and eventually causing breakage or air leakage to the cell. Clean the basket surfaces as well. Use cotton gloves to handle the basket and cell after cleaning.
Each cell should have its own basket assembly. Repeated insertion and re moval poses an unnecessary risk to the cell. A cell-pad of fiber ceramic fi ber or quartz wool should be in the bottom of the basket to pad the cell. This cell-pad should not be too thick so as to prevent proper fitting of the basket cover.
Follow these steps to properly insert the sealed cell. Refer to Figure 5 for loca tion of component detail.
-
-
-
49
Page 53
9260 Mini Fixed Point Cell Furnace
User’s Guide
1. Install the cell-pad into the bottom of the basket. Typically a 1/4" (6.35 mm) thick pad is used. It will crush to a thinner dimension as the cell is installed. It SHOULD NOT be so thick that the basket cover touches the cell at the top. If the pad contains an organic binder, heat the pad to re move it before use.
2. Cut a strip of clean paper approximately 1-inch wide and 1-inch longer than the basket.
3. Insert this paper into the bottom of the basket while it is lying down. At least 1-inch of the paper should be extending out of the basket for easy removal.
4. Temporarily hold the sealed cell sideways with the spherical end toward the basket opening.
5. Carefully insert the cell into the basket sliding it on the paper instead of on the metal basket. When the cell reaches the bottom, return the basket carefully to the upright position.
6. Remove the paper completely.
7. Install a thin insulating pad on top of the cell providing a clearance hole for the thermometer.
8. Check the basket cover fit to ensure that the evacuation port on the top of the cell does not prevent it from fitting flush against the top of the basket.
9. Remove the lid for now.
10. Using the tool provided, carefully lower the basket into the thermal well.
11. Install the basket lid on top of the basket being sure that it is flush against the basket.
12. Install a thin insulation pad on top of the lid with a clearance hole for the thermometer.
13. Install the thermal shunt over the top of the basket fitting it onto the ledge above.
14. Check the alignment.
15. Fit the top insulation above the shunt.
16. Carefully check the thermometer fit all the way to the bottom of the cell. There must not be any significant resistance to insertion.
17. Place a small pad of quartz wool at the bottom of the reentrant well of the cell which will, to some extent, pad the bottom of the well against the thermometers when they are inserted.
-
50
CAUTION: Careless insertion of a thermometer into the well can break
the quartz glass at the bottom or even break quartz thermometers.
Page 54
CAUTION: Occasionally remove the thermal shunt and the basket. If
there is any resistance to removal, remove oxides with fine grit sandpaper. This is generally only a problem at the aluminum point.
7.3 Melting Point Realization
Recent improvements of the melting point method have shown that excellent accuracy can also be achieved with this method. With this method, the cell tem perature is carefully raised to a temperature just below the melting point. The furnace and cell are allowed to dwell at this temperature for a time to allow temperature equalization throughout the system. The furnace is then raised to a point a few degrees above the melting point for a short time to give melting a good start. The temperature is then dropped to a temperature just above the melting point for the duration of the measuring period. A “melt heater” inside the reentrant well is turned on for a short time creating a second zone of melted sample which is next to the thermometer during measurements. This action per mits more accurate measurements and is similar to the technique used with a Triple-point of Water cell.
Measurements are made until the plateau begins to deviate and then the melting must be completed and the process started over.
7 Fixed-Point Realization
Melting Point Realization
-
-
7.3.1 Melting Point Procedure
The following procedure illustrates the steps required to successfully realize a
51
Page 55
9260 Mini Fixed Point Cell Furnace
User’s Guide
fixed-point temperature through melting the cell sample. Figure 10 illustrates the process graphically.
Figure 10 Melting Point Realization
7.3.1.1 Preparation
The cell should be carefully loaded into the basket and then into the furnace ac cording to the procedure outlined in Section 8.2, 7.2. Thermometers that are to be inserted into the cell should first be cleaned with pure alcohol. The in ner-melt heater should be inserted carefully into the cell.
7.3.1.2 Setting Up The Controller
Setup of the controller consists of selecting the fixed-point and the curve to be used. The setup is in the secondary menu of the controller menus. Access the setup by pressing “SET” and “EXIT”. Then press “SET” 4 times and “UP” un til “ProG” is displayed. Press “SET”. “FP” (fixed-point) flashes and then the current setting is displayed. This setting is the one used on the previous test. Press “UP” or “DOWN” repeatedly to toggle through Indium, Tin, Zinc, Alu minum and Other. When you reach the desired fixed-point press “SET” to se lect it. Immediately after pressing “SET”, “CURVE” flashes and the current setting appears. The available curves are melt and freeze. Use the “UP” or “DOWN” key to select the “MELT” curve. Press “SET” to select it. Following the curve selection, the term “TEMP” flashes on the display followed by adjust ments that can be made by the user to the MELT, FREEZE, and MAINTAIN temperatures. These adjustments are only for experienced users. Press
52
-
-
-
-
-
-
Page 56
“EXIT” to exit the program menu and to return to the display temperature. Carefully insert the melt heater into the cell (see Figure 3 ). Set the heater switch on the rear of the furnace to AUTO. The heater must be clean and care fully inserted into the cell to prevent damage. This should be done before program initiation.
7.3.1.3 Program Initiation
Now that the fixed-point and the melt curve have been selected, the furnace is ready to initiate the program. Advance to the program from the display temper ature by pressing “SET” and “DOWN”. The term “StOP” appears unless the program is running. The following three steps include “MELt”, “MAInt”and “FrEEzE”. The step that first appears will be the last step utilized and not nec essarily the first step desired. Press the “UP” or “DOWN” keys to view the first step needed which is “MAInt” for the preliminary heating of the cell to just be low the melting point. The cell is held at that temperature for a period of time to allow everything to equilibrate to that temperature. Press “SET” to select it. The controller immediately starts heating the cell with its metal sample. The furnace heats at a preprogrammed rate that automatically slows down before the sample reaches the “MAINT” point. Once the furnace has reached that tem­perature (about 1°C below the melting point), it dwells there for 60 minutes to permit the furnace and metal sample to equilibrate. After the 60 minute period, the display flashes on and off alternately indicating that the cell is now ready to initiate the melt function.
To initiate the melting of the sample, press “SET” and “DOWN” again. Press the“UP”or“DOWN”keysuntiltheterm“MELt” is on the display. Press “SET” to select that step and the controller selects a set-point temperature a few degrees above the melting point to begin the sample melting process. The tem­perature of the furnace slowly scans to about 4°C above the melting point and dwells there for about 8 minutes. At that time the melt heater turns on automat ically. This heater melts a thin layer of the sample next to the measurement thermometer increasing the accuracy of this technique. After a few minutes the melt heater automatically turns off and the furnace temperature drops to a tem perature just above the melting point of the sample. Remove the melt heater and insert a pre-heated monitor thermometer. When the temperature has stabi lized, calibrations may begin. Use the pre-heat wells to heat up the thermome ters before inserting them into the cell. This action preserves the latent energy and permits more calibrations during the melting plateau. Allow the thermome ters to equilibrate for 20 minutes before making readings. The plateau can last several days depending on how it is used.
When the plateau has ended, the furnace may be turned to about 25°C and turned off or, if desired, the process may be repeated. To repeat the process, press “SET” and “DOWN” and use the “UP” or “DOWN” keys to find the “FrEEzE” step. Select by pressing “SET”. The furnace temperature will drop to a temperature 8°C below the freezing point and dwell there as before. The sample will freeze and the temperature will automatically rise to a temperature just below the melting point. The furnace will hold there for a time until the cell and furnace have fully equilibrated. At that time, the display will flash indi
7 Fixed-Point Realization
Melting Point Realization
-
-
-
-
-
-
-
-
-
-
53
Page 57
9260 Mini Fixed Point Cell Furnace
User’s Guide
cating that the cell is ready to begin the melting process again. It is initiated as before by selecting “MELt”step.
7.4 Freezing Point Realization
The freezing point is an established ITS-90 method. This method first entirely melts the metal sample. The furnace temperature is then carefully reduced to a temperature below the freezing point just low enough to induce freezing of the sample. This temperature must be several degrees below the freezing point in order to overcome the sub-cooling of the sample. After recalescense, (initiation of the freeze as observed on a monitor thermometer) the furnace temperature is increased to a temperature just slightly below the freezing point. This action re stricts the amount of heat that can flow from the cell, which controls the rate of freezing. High temperature stability and good temperature uniformity permit the temperature to be adjusted closely achieving long freezing plateaus. Long freezing plateaus translate into a larger number of calibrations that can be done during one freeze.
Measurements are made until the plateau begins to deviate and then the freez­ing must be completed and the process started over.
7.4.1 Freezing Point Procedure
-
The following procedure illustrates the steps required to successfully realize a
54
Page 58
7 Fixed-Point Realization
Freezing Point Realization
fixed-point temperature through freezing the cell sample. Figure 11 illustrates the process graphically.
Figure 11 Freezing Point Realization
7.4.1.1 Preparation
The cell should be carefully loaded into the basket and then into the furnace ac cording to the procedure previously outlined in Section 8.2, 7.2. Thermometers that are to be inserted into the cell should first be cleaned with pure alcohol. A monitor thermometer should be inserted into the cell to monitor the process from the beginning.
7.4.1.2 Setting Up The Controller
Setup of the controller consists of selecting the mode and fixed-point to be used. The setup is in the secondary menu of the controller menus. Access the program parameters by pressing “SET” and “EXIT” simultaneously. Then press “SET” 4 times and “UP” until the program menu is reached. The word “ProG” is displayed. Press “SET”. “FP” (fixed-point) flashes and then the current setting is displayed. (The metal sample displayed is from the previous test.) Press “UP” or “DOWN” repeatedly to toggle through Indium, Tin, Zinc, Aluminum and Other. When you reach the desired fixed-point press “SET” to select the metal sample to use. Immediately after pressing “SET”, “CURVE
-
55
Page 59
9260 Mini Fixed Point Cell Furnace
User’s Guide
flashes and the current setting “MELt”or“FrEEZE” appears. The terms indi cate the first of the two curves, melt or freeze. Use the “UP” or “DOWN” key to view the desired curve, “FrEEZE”. Press “SET” to select this curve. Press “EXIT” to exit the program menu and to return to the display temperature.
7.4.1.3 Program Initiation
Now that the fixed-point and the freeze curve have been selected, the furnace is ready to initiate the program. Advance to the program from the display temper ature by pressing “SET” and “DOWN”. One of three steps of the selected pro gram curve or “StOP” is displayed. The three steps include “MELt”, “MAInt” and “FrEEzE”. The step that first appears will be the last step utilized and not necessarily the first step desired. Press the “UP” or “DOWN” keys to view the first step needed which is “MELt” for the freezing process. Press “SET” to select it.
NOTE: Select stop to turn off the program function. Manually adjusting
the set-point also stops the program.
The controller immediately starts heating the cell. The furnace heats at a pre­programmed rate that automatically slows down before the sample reaches the melting point. Once the furnace has reached the melting temperature (about 8°C above the melting point), it dwells there for 90 minutes to permit the metal sample to melt. The furnace then ramps down automatically to a temperature just above the melting point. The controller holds at that temperature for ap­proximately 20 minutes while everything equilibrates after which the display flashes on and off alternately indicating that the cell is now ready to initiate the freeze. Note: The monitor thermometer temperature can be checked while the sample is melting providing a good check of its accuracy. The control tempera ture can then be accurately checked during this dwell period.
To initiate the freezing of the metal, press “SET” and “DOWN” again. Press the up and down keys until the term “FrEEzE” is on the display. Press “SET” to select that mode and the controller selects a set-point temperature a few degrees below the freezing point to begin the sample freezing process. This lower tem perature is to overcome the super-cooling of the metal sample. The cell temper ature must be monitored with the monitor thermometer at this time in order to see when the cell starts to freeze. The freeze is established when the cell tem perature begins a sharp rise toward the freezing point (recalesence). Immedi ately insert a quartz glass rod that is at room temperature for approximately 2 minutes to create a thin layer of frozen metal. This action positions the change of phase right next to the thermometer which is necessary for high accuracy. After removing the tube, press “SET” and “DOWN” again and select “MAint” to maintain the furnace at a temperature just below the freezing temperature.
With the freeze plateau underway, calibrations can now be made. Use the pre-heat wells to heat up the thermometers before inserting them into the cell preserving the latent energy and permitting more calibrations during the freeze. Allow the thermometers to equilibrate for 20 minutes before making readings.
-
-
-
-
-
-
-
-
56
Page 60
7 Fixed-Point Realization
Test Probe Calibration
Calibration must end before the end of the freezing plateau. To continue use, re-melt the sample. Press “SET” and “DOWN”. Select the “MELt” function and the controller will re-melt the sample and the process may be repeated. This function is convenient if the furnace is to be used again early the next day. The sample can be melted overnight and be ready to freeze in the morning.
CAUTIONS: Use care in handling the metal fixed-point cells. They can
be easily broken or contaminated. Keep all items clean. Clean with pure alcohol. Handle with clean cotton gloves or clean paper. Remove any in advertent finger prints or other contaminants.
-
DO NOT drop the cell or basket into the furnace. Do not drop a thermom eter into the cell even a short distance.
DO NOT leave quartz thermometers in the preheat wells any longer than necessary to heat to temperature, especially at the aluminum point. Con tamination may result.
DO NOT insert cells into a hot furnace. Start from near ambient tempera­tures.
When all calibrations are complete, manually set the temperature to about 25°C to prevent the furnace from heating unexpectedly the next time it is used. The furnace can be turned off while hot and permitted to cool on its own.
7.5 Test Probe Calibration
For optimum accuracy and stability, allow the furnace to warm up for 10 min utes after power-up and then allow adequate stabilization time after reaching the set-point temperature. After completing operation of the furnace, allow the well to cool by setting the temperature to 25°C and allowing the unit to cool off before switching the power off.
7.5.1 Calibrating a Single Probe
Insert the probe to be calibrated into the well of the instrument. The probe should fit snugly into the comparison block yet should not be so tight that it cannot be easily removed. Avoid any dirt or grit that may cause the probe to jam into the comparison block. Best results are obtained with the probe inserted to the full depth of the well. Once the probe is inserted into the well, allow ade quate stabilization time to allow the test probe temperature to settle as de scribed above. Once the probe has settled to the temperature of the well, it may be compared to the furnace display temperature. The display temperature should be stable to within 0.01°C degree for best results.
-
-
-
-
-
57
Page 61
9260 Mini Fixed Point Cell Furnace
User’s Guide
Never introduce any foreign material into the probe hole of the insert. Fluids etc. can leak into the furnace causing damage to the instrument or binding and damage to your probe.
7.5.2 Furnace Characteristics
There is a temperature gradient vertically in the test well. Heat is applied to the block in such a way as to compensate for nominal heat losses out of the top of the furnace. However, actual heat losses will vary with the design of the ther mometer probes inserted into the instrument and the temperature. For best re sults, insert the probe to the full depth of the well.
CAUTION: DO NOT heat the thermometer hub or handle above the
rated temperature.
7.5.2.1 Stabilization and Accuracy
The stabilization time of the instrument depends on the conditions and temper atures involved. Typically the test well will be stable to 0.1°C within 10 min­utes of reaching the set-point temperature as indicated by the display. Ultimate stability will be achieved 15 to 20 minutes after reaching the set temperature. Inserting a cold probe into the well requires another period of stabilization de­pending on the magnitude of the disturbance and the required accuracy. For ex­ample, inserting a 0.25 inch diameter room temperature probe into a comparison block at 300°C takes 5 minutes to be within 0.1°C of its settled point and takes 10 minutes to achieve maximum stability.
Speeding up the calibration process can be accomplished by knowing how soon to make the measurement. Test measurements should be made at the desired temperatures with the desired test probes to establish these times.
-
-
-
58
Page 62
Figure 12 9260 Comparison Block Heating Up
7 Fixed-Point Realization
Test Probe Calibration
Figure 13 9260 Comparison Block Cool Down
59
Page 63
8 Furnace Calibration
For optimum performance in realizing melting or freezing points of fixed-point cells, calibration of the 9260 must be maintained. Excessive temperature gradi ents and inaccurate furnace temperatures can make it difficult to realize melting or freezing points, reduce the length of the melting or freezing plateau, and even possibly damage a fixed-point cell. For best results the vertical tempera ture uniformity should be kept within 0.2°C over the lower 12 cm (5 in.) and the temperature accuracy should be kept within 0.5°C. The temperature profile and accuracy should be checked often and adjusted as necessary. While the fur nace is new consider checking these at least once a month. The following sec tions explain the temperature profile and temperature accuracy calibration procedures.
8.1 Temperature Profile Adjustment
The 9260 Furnace is adjusted at the factory for a flat vertical temperature pro­file. The temperature profile should be regularly checked and adjusted as neces­sary to maintain it within about 0.2°C over the bottom 12 cm (5 in.) with a slightly higher temperature toward the top. The following steps explain how to test and adjust the temperature profile. The temperature accuracy should subse­quently be calibrated according to Section 8.2 below whenever the zones are adjusted.
8 Furnace Calibration
Temperature Profile Adjustment
-
-
-
-
8.1.1 Step 1: Measure the profile
The temperature profile should be measured at 660°C or at the highest fixed point with which the furnace is intended to be operated. The furnace should have the fixed-point cell installed and the cell should be completely frozen or melted. Incorrect data will result if the cell is only partially melted or frozen. Use a quartz SPRT and thermometer readout that together provide adequate resolution and stability to allow relative temperature measurements to 0.02°C or better. You must wait until the temperature of the SPRT is stable before re cording readings. It may take 20 to 30 minutes after inserting the SPRT before it becomes stable. Measure the temperature at three locations: t1 with the SPRT fully inserted into the well, t2 with the SPRT withdrawn 6 cm (2.5 in.), and t3 with the SPRT withdrawn 12 cm (5 in.). To reduce possible effects of tempera ture drift repeat the measurements in the following order: t1, t2, t3, t2 again, then t1 again. Wait until the SPRT is stable (about two minutes) before record ing each reading. Average the two t1 measurements and the two t2 measure ments then compare the temperatures. Temperature t2 relative to t1 should be less than 0.1°C higher or lower. Temperature t3 relative to t1 should be between 0 and 0.2°C higher. If the gradient exceeds these limits adjust the end zones as explained in the next step.
-
-
-
-
61
Page 64
9260 Mini Fixed Point Cell Furnace
User’s Guide
8.1.2 Step 2: Adjust the end zones
The temperature profile can be adjusted by altering the calibration parameters "toPPCt"and"botPCt" in the controller (see Section 5.16.5). Increasing or decreasing the top zone heat by 1% typically increases or decreases t3 relative to t
by about 0.1°C at 660°C and increases or decreases t2relative to t1by
1
about 0.03°C. Likewise, increasing or decreasing the bottom zone heat by 1% typically increases or decreases temperature t 660°C and increases or decreases temperature t From these relationships, the following formulas are derived and these can be used to determine new settings for the top and bottom zones to produce a flat gradient with the top only 0.1°C hotter than the bottom.
relative to t3by about 0.1°C at
1
relative to t3by about 0.03°C.
2
topPct topPct
botPct botPct
%
tt
−−
⎜ ⎝
⎛ ⎜
() ()
21 31
C
°
%
tt
−−
() ()
21 31
C
°
%
⎛ ⎜
⎛ ⎜
tt′= +
.
.
25 17 5
⎟ ⎠
C
°
%
tt′= +
−+25 7 5
⎟ ⎠
C
°
For example, t1is measured in Step 1 to be 657.71, t2is measured to be 657.83, and t
is measured to be 658.41. The bottom zone is set at 94% and the top
3
zone is set at 91%. The new settings are calculated as follows.
topPct
botPct
′= +
91 25 65783 657 71 17 5%
′= +
94 25 65783 657 71 7 5%
%
⎜ ⎝
⎛ ⎜
()
⎟ ⎠
C
°
%
()
⎟ ⎠
C
°
CC
°− ° −
.. .
CC
°− ° −
...
%
⎛ ⎜
⎛ ⎜
⎞ ⎟
C
°
%
()
⎟ ⎠
C
°
°− ° + =658 41 657 71 1 75 84...%%CC
()
°− ° + =658 41 657 71 0 75 93...%%CC
8.1.3 Repeat Step 1 and Step 2 if necessary
After adjusting the end zones repeat Step 1 to check the temperature profile again. If the temperature differences still exceed the recommended limits repeat Step2again.
8.2 Temperature Calibration
For best results in realizing freezing or melting points the furnace should be ac curate to at least 0.5°C and preferably to 0.3°C. The accuracy should be checked periodically and adjusted if necessary. For small adjustments at one temperature the simplified one-point calibration procedure explained in Section
8.2.1 may be used. For calibration over the entire range of the furnace the
three-point calibration procedure explained in Section 8.2.2 should be used. For each procedure a quartz SPRT is needed in a addition to a thermometer readout that together provide an accuracy of 0.1°C or better.
+175.%
075.%
-
62
Page 65
8.2.1 One-point Calibration
This one-point calibration procedure improves the temperature accuracy of the furnace at one point by adjusting the R0 calibration parameter (see Section
5.16.2). It may be used often to ensure the quality of a specific fixed-point realization.
The first step is to measure the furnace temperature at the temperature of inter est. The furnace should have the fixed-point cell installed and the cell must be entirely frozen or melted. Measure the temperature with the SPRT fully in serted into the well. Make sure the furnace temperature is stable and make care ful and accurate measurements.
The next step is to adjust R0 to make the temperature closer to the set-point. In creasing or decreasing R0 by 0.01Ω increases or decreases the temperature by
about 0.025°C at 0°C and about 0.09°C at 650°C. Use the following formula to calculate a new value for R0 to make the temperature closer to the set-point.
tt
RR
00
′=
For example, with the set-point at 655°C the furnace temperature is measured and found to actually be 655.65. R0 is set to 100.124 (ohms). The new value for R0 is calculated as follows.
meas sp
t
C
°
25
+.
100
ΩΩ
sp
8 Furnace Calibration
Temperature Calibration
-
-
-
-
CC
655 65 655 00
°− °
R
0 100124
..
Ω−
CC
°
25
.
ΩΩ
655
+
100
8.2.2 Three-point Calibration
The first time the furnace is calibrated at the factory it is calibrated over the full range by adjusting the three calibration parameters R0, ALPHA, and DELTA. This full calibration may be repeated periodically to ensure that the furnace is accurate to 0.5°C or better at all temperatures in its range.
Full calibration requires making measurements of the furnace temperature at three widely separated set-points throughout the range of the furnace. Sug gested set-points are 50°C, 400°C, and 650°C. The actual temperature of the furnace is measured at these temperatures. As before, the furnace should have the fixed-point cell installed and the cell must be entirely frozen or melted. Measure the temperature with the SPRT fully inserted into the well. Make sure the furnace temperature is stable (wait about 20 minutes after inserting the SPRT) and make careful and accurate measurements. Once the temperatures are obtained, new values for R0, ALPHA, and DELTA can be calculated. The procedure is summarized as follows.
1. Choose three set-points to use in the calibration of the R0 , ALPHA, and DELTA parameters. These set-points are generally 50°C, 400°C, and 650°C but other set-points may be used if desired or necessary.
100′=
=.
°
.052Ω
-
63
Page 66
9260 Mini Fixed Point Cell Furnace
User’s Guide
2. Set the furnace to the low set-point. When the furnace reaches the set-point and the display is stable, wait approximately 15 minutes and then take a reading from the thermometer. Sample the set-point resis tance. Write these values down as T
3. Repeat step 2 for the other two set-points recording them as T and R
respectively.
3
4. Using the recorded data, calculate new values for the R0 , ALPHA, and DELTA parameters using the equations given below.
8.2.3 Compute DELTA
ATT=−
32
BTT=−
21
TTTT
3322
C
=
D
=
1001100 1001100
TTTT
2211
1001100 1001100
-
and R1respectively.
1
2,R2,T3
⎤ ⎥
⎥ ⎦
⎥ ⎦
,
ERR=−
32
FRR=−
21
AF BE
=
DE CF
delta
T
are the measured temperatures using the thermometer. R
1-3
of R from the display of the 9260. (Press SET and DOWN at the same time.)
T
and R1are the measured temperature and resistance at 50°C.
1
T
and R2are the measured temperature and resistance at 400°C
2
T
and R3are the measured temperature and resistance at 650°C
3
8.2.4 Computer R0 and Alpha
TT
=+
a T delta
11
=+
a T delta
33
Ra Ra
rzero
31 13
=
aa
13
11
1001100
TT
33
1001100
are the values
1-3
64
Page 67
RR
alpha
13
=
Ra Ra
31 13
Delta is the new value of DELTA computed above.
Program the new values for DELTA (delta), R0(rzero), and ALPHA (alpha) into the furnace by: pressing the “SET” and “EXIT” keys simultaneously and then pressing “SET” until R0 is displayed. Press “SET” then use the UP or DOWN keys until the correct numerical setting is displayed. Press “SET” to ac cept the new value. Continue this process for ALPHA and DELTA.
8.2.5 Accuracy and Repeatability
Check the accuracy of the furnace at various points over the calibrated range. If the furnace does not pass specification at all set-points, repeat the Calibration Procedure.
8 Furnace Calibration
Temperature Calibration
-
65
Page 68
9 Maintenance
The calibration instrument has been designed with the utmost care. Ease
of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in an oily, wet, dirty, or dusty environment.
If the outside of the instrument becomes soiled, it may be wiped clean
with a damp cloth and mild detergent. Do not use harsh chemicals on the surface which may damage the paint.
It is important to keep the well of the instrument clean and clear of any
foreign matter. Do not use fluid to clean out the well.
The furnace should be handled with care. Avoid knocking or dropping the
instrument.
If the comparison block is dropped, examine the comparison block for de
formities before inserting it in the well. If there is any chance of jamming the comparison block in the well, file or grind off the protuberance.
Do not slam the probe stems into the well. This type of action can cause a
shock to the sensor or break the entrant well of a fixed-point cell.
If a hazardous material is spilt on or inside the equipment, the user is re-
sponsible for taking the appropriate decontamination steps as out-lined by the national safety council with respect to the material.
If the mains supply cord becomes damaged, replace it with a cord with
the appropriate gauge wire for the current of the instrument. If there are any questions, call Hart Scientific Customer Service for more informa­tion.
Before using any cleaning or decontamination method except those rec ommended by Hart, users should check with Hart Scientific Customer Service to be sure that the proposed method will not damage the equip ment.
If the instrument is used in a manner not in accordance with the equip ment design, the operation of the furnace may be impaired or safety haz ards may arise.
9 Maintenance
-
-
-
-
-
67
Page 69
10 Troubleshooting
10 Troubleshooting
Troubleshooting
If problems arise while operating the 9260, this section provides some sugges tions that may help you solve the problem. A wiring diagram is also included.
10.1 Troubleshooting
Below are several situations that may arise followed by suggested actions to take for fixing the problem.
10.1.1 Incorrect Temperature Reading
Power the unit on and watch the display. If the first number displayed is less than “-0005-”, the unit has been re-initialized. The unit needs to be repro grammed for R0, ALPHA, and DELTA. These numbers can be found on the Report of Calibration that was shipped with the unit.
10.1.2 The unit will not heat or heats at half rate
Check the fuse. If the fuse is blown the display should be out.
If the problem continues, contact Hart Scientific Customer Support.
10.1.3 The unit heats slowly
Check the Scan and Scan Rate settings. The Scan may be on with the
Scan Rate set low.
NOTE: When in program mode, the scan rate is automatically set.
-
-
10.1.4 If the display flashes any of the following:
err 1” - This error means there is a RAMerror
err 2” - This error means there is a NVRAM error
err 3” - This error means there is a RAMerror
err 4” - This error means there is an ADC set up error
err 5” - This error means there is an ADC ready error
Initialize the system by performing the master reset sequence. If the unit repeats the error code, contact Hart Scientific Customer Sup-port for a re turn authorization and for instructions on returning the unit.
Master Reset Sequence - Hold the “SET” and “EXIT” keys down at the same time while powering up the unit. The screen will display “-init-” , “9260" and the version of the software. The unit will need to be repro grammed for R0, ALPHA, and DELTA in the calibration menu. These
-
-
69
Page 70
9260 Mini Fixed Point Cell Furnace
User’s Guide
numbers can be found on the Report of Calibration that was shipped with the unit.
10.1.5 If the display flashes any of the following:
err 6" - This error means there is a SENSOR error
The sensor is disconnected or shorted. Please contact Hart Scientific Cus
tomer Support for further instructions.
10.1.6 If the display flashes any of the following:
err 7" - This error means there is a HtrCTL error
The fan will go on high speed. Initialize the system by performing the master reset sequence. If the unit repeats the error code, turn the unit off and contact Hart Scientific Customer Support for a return authorization and for instructions on returning the unit.
10.2 CE Comments
10.2.1 EMC Directive
Hart Scientific's equipment has been tested to meet the European Electromag­netic Compatibility Directive (EMC Directive, 89/336/EEC). Selection of Light Industrial or Heavy Industrial compliance has been based on the intended use of the instrument. Units designed for use in a calibration laboratory have been tested to Light Industrial Standards. Units designed to be used in the "field" have been tested to both Light Industrial and Heavy Industrial Standards. The Declaration of Conformity for your instrument lists the specific standards to which the unit was tested.
-
70
10.2.2 Low Voltage Directive
(Safety) In order to comply with the European Low Voltage Directive (73/23/EEC), Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and IEC 1010-2-010 (EN 61010-2-010)standards.
Loading...