The 9260 is a specialized furnace for the realization of certain defining
fixed-points of the International Temperature Scale of 1990 (ITS-90). This fur
nace is intended for Calibration Laboratory use and not for field applications.
The 9260 permits simplified realization of either freezing or melting curves. In
ternal programming of the micro-processor controller provides preprogrammed
scan rates, set-points for each step in the process, dwell timing, and indication
that the next step is ready. ITS-90 points including Indium, Tin, Zinc, and Alu
minum are included. In addition, one other user-defined point may be selected.
User defined non-ITS 90 fixed-points are useful for specific applications.
The 9260 furnace is available in 115 VAC (±10%) 60 Hz or 230 VAC(±10%)
50 Hz models.
The 9260 furnace may also be used as a dry-well calibrator or as a temperature
comparator. Pre-drilled inserts are available from Hart Scientific for this
application.
Built in programmable features include:
• Temperature scan rate control
• Fixed-point programming
• Eight set-point memory
• Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s hybrid analog/digital controller. The controller uses a precision, platinum RTD as a sensor and controls
the well temperature with a solid state relay (triac) driven heater.
The LED front display panel continuously shows the current well temperature.
The temperature may be easily set with the control buttons to any desired tem
perature within the specified range. The furnace’s multiple fault protection de
vices insure user and instrument safety and protection.
The 9260 furnace was designed for high accuracy calibrations using compari
son measurements or fixed-point calibration methods and for ease of operation.
Through proper use, the instrument will continuously provide accurate calibra
tion of temperature sensors and devices. The user should be familiar with the
safety guidelines and operating procedures of the furnace as described in this
user manual.
1 Before You Start
Introduction
-
-
-
-
-
-
-
1
Page 9
9260 Mini Fixed Point Cell Furnace
User’s Guide
1.2Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols
may be used on the instrument or in this manual.
Table 1 International Electrical Symbols
SymbolDescription
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union Directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
Canadian Standards Association
2
Page 10
SymbolDescription
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re
fers to the level of Impulse Withstand Voltage protection provided. Equipment of
OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from
the fixed installation. Examples include household, office, and laboratory appliances.
C-TIC Australian EMC Mark
1.3Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
“WARNING” identifies conditions and actions that may pose hazards to
•
the user.
• “CAUTION” identifies conditions and actions that may damage the in-
strument being used.
1 Before You Start
Safety Information
-
1.3.1
WARNINGS
To avoid personal injury, follow these guidelines.
• DO NOT operate this unit without a properly grounded, properly polar-
ized power cord.
• DO NOT connect this unit to a non-grounded, non-polarized outlet.
• DO USE a ground fault interrupt device.
•
HIGH VOLTAGE is used in the operation of this equipment. SEVERE
INJURY OR DEATH may result if personnel fail to observe safety pre
cautions. Before working inside the equipment, turn power off and dis
connect power cord.
•
HIGH TEMPERATURES PRESENT in this equipment FIRES AND
SEVERE BURNS may result if personnel fail to observe safety precau
tions.
•
DO NOT use this unit in environments other than those listed in the
user’s manual.
•
Continuous use of this equipment at high temperatures for extended peri
ods of time requires caution. Completely UNATTENDED HIGH TEM
PERATURE OPERATION IS NOT RECOMMENDED.
•
Components and heater lifetimes can be shortened by continuous high
temperature operation.
•
The instrument can generate extreme temperatures. Precautions must be
taken to prevent personal injury or damage to objects. Probes may be ex
tremely hot when removed from the instrument. Cautiously handle probes
-
-
-
-
-
-
3
Page 11
9260 Mini Fixed Point Cell Furnace
User’s Guide
to prevent personal injury. Always use the special comparison block tongs
that are supplied with the furnace to remove the comparison block or cell
basket. Carefully place probes on a heat resistant surface or rack until
they are at room temperature. Never place any objects other than the com
parison blocks, cell basket, or cells supplied with the furnace into the
well.
Use only grounded AC mains supply of the appropriate voltage to power
•
the instrument. The furnace requires 11 amps maximum at 115 VAC
(±10%),60 Hz and 6 amps maximum at 230VAC (±10%), 50 Hz.
Follow all safety guidelines listed in the user’s manual.
•
Calibration Equipment should only be used by Trained Personnel.
•
-
1.3.2
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
Operate the instrument in room temperatures between 5-45°C (41-113°F).
•
Allow sufficient air circulation by leaving at least 6 inches of space between the furnace and nearby objects. Overhead clearance needs to allow
for safe and easy insertion and removal of probes for calibration.
• The furnace is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with
care. Always carry the unit in an upright position to prevent the comparison blocks from falling out. Keep the well of the instrument clean and
clear of any foreign matter. Do not operate near flammable materials.
• DO NOT use fluids to clean out the well.
• DO NOT move the furnace with the fixed-point cells inside. They can be
easily broken.
•
Before initial use, after transport, and anytime the instrument has not been
energized for more than 7 days, the instrument needs to be energized for a
“dry-out” period of 1-2 hours before it can be assumed to meet all of the
safety requirements of the IEC 1010-1.
•
The instrument is equipped with operator accessible system fuses. If a
fuse blows, it may be due to a power surge or failure of a component. Re
place the fuse once. If the fuse blows a second time, it is likely caused by
failure of a component. If this occurs, contact an Authorized Service
Center. Always replace the fuse with one of the same rating, voltage, and
type. Never replace the fuse with one of a higher current rating.
•
If a main supply power fluctuation occurs, immediately turn off the in
strument. Power bumps from brown-outs and black-outs can damage the
instrument. Wait until the power has stabilized before re-energizing the
instrument.
-
-
4
Page 12
Hart Scientific Authorized Service Centers
1 Before You Start
1.4Hart Scientific Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Hart Scientific, Inc.
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
5
Page 13
9260 Mini Fixed Point Cell Furnace
User’s Guide
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following
information available:
Model Number
•
Serial Number
•
Voltage
•
• Complete description of the problem
6
Page 14
2 Specifications and Environmental Conditions
2Specifications and Environmental
Conditions
2.1Specifications
The 9260 specifications are detailed in Table 2 and the Mini Cell specifications
are detailed in Table 3.
Table 2 9260 Specifications
Operating Range50°C to 680°C (122°F to 1256°F)
Ambient Temperature5°C to 45°C (41°F to 113°F)
Accuracy±0.2°C 50°C to 300°C
±0.3°C 300°C to 450°C
±0.5°C 450°C to 680°C
Stability± 0.03°C to 300°C
± 0.05°C above 300°C
Well-to-Well Gradient±0.02°C (±0.036°F)
Melting/Freezing Point Duration6 to 10 hours typical
Vertical GradientsTop and bottom zones adjustable by offset
Resolution0.01°C or °F
Display Scale°C or °F, switchable
Comparison BlockBlank block, two multi-hole blocks, and custom blocks available
Fault ProtectionSensor burnout and short protection, over temperature thermal
cut-out
Heater1200 W maximum - adjustable top and bottom zones
Heating TimeApproximately 1.25 hours, 25°C to 680°C
Cooling TimeApproximately 10.5 hours, 680°C to 100°C
Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
Expanded Uncertainty – Using
9260 Furnace
• temperature range: 5 - 45°C (41 - 113°F)
• ambient relative humidity: 15 - 50%
• pressure: 75kPa - 106kPa
• mains voltage within ± 10% of nominal
•
vibrations in the calibration environment should be minimized
•
altitude does not effect the performance or safety of the unit
2.3Warranty
Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be
free from defects in material and workmanship under normal use and service
for a period as stated in our current product catalog from the date of shipment.
This warranty extends only to the original purchaser and shall not apply to any
product which, in Hart’s sole opinion, has been subject to misuse, alteration,
abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc
tions on appropriate Hart products. It is not warranted to be error free.
Hart’s obligation under this warranty is limited to repair or replacement of a
product which is returned to Hart within the warranty period and is determined,
upon examination by Hart, to be defective. If Hart determines that the defect or
malfunction has been caused by misuse, alteration, abuse or abnormal condi
-
-
8
Page 16
2 Specifications and Environmental Conditions
Warranty
tions or operation or handling, Hart will repair the product and bill the pur
-
chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling
or writing an Authorized Service Center for authorization. Service Centers as
-
sume NO risk for in-transit damage.
For service or assistance, please contact an Authorized Service Center.
Hart Scientific, Inc.
799 East Utah Valley Drive
American Fork, UT 84003-9775
Phone: (801) 763-1600
Fax: (801) 763-1010
E-mail: support@hartscientific.com
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLU
SIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EX
-
PRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY
PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR
ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
9
Page 17
3Quick Start
3.1Unpacking
Unpack the instrument carefully and inspect it for any damage that may have
occurred during shipment. If there is shipping damage, notify the carrier
immediately.
Verify that the following components are present:
9260 Furnace
•
Inner-Melt Heater
•
Power Cord
•
Manual
•
Comparison Blocks (optional)
•
Cell Basket
•
• Cell Basket Lid
• Basket Removal Tool
• Thermal Shunt
• Top Insulation
• Cell Pad Insulation
3 Quick Start
Unpacking
3.2Setup
Place the furnace on a flat surface with at least 6 inches of free space around
and 18 inches above the instrument. Install the power cord into the power entry
module on the underside of the furnace. Plug the power cord into a grounded
mains outlet. Verify that the nominal voltage corresponds to that indicated on
the back of the instrument.
Carefully insert the comparison blocks or cell baskets into the well. (DO NOT
drop them into the well.) Comparison block holes should be of the smallest di
ameter possible while still allowing the probe to slide in and out easily. Various
hole sizes are available from Hart Scientific. The well must be clear of any for
eign objects, dirt and grit before the comparison block is inserted. See Section
4.7 for more details.
Turn on the power to the instrument by toggling the switch on the power entry
module located underneath the front of the furnace. The fan should begin qui
etly blowing air through the instrument and the controller display should illu
minate after 3 seconds. After a brief self-test the controller should begin normal
operation. If the unit fails to operate please check the power connection.
The display will begin to show the well temperature and the well heater will
start operating to bring the temperature of the well to the set-point temperature.
-
-
-
-
11
Page 18
9260 Mini Fixed Point Cell Furnace
User’s Guide
3.3“Dry-out” Period
Before initial use, after transport, and anytime the instrument has not been en
ergized for more than 10 days, the furnace will need to be energized for a
“dry-out” period of 1–2 hours before it can be assumed to meet all of the safety
requirements of IEC 1010-1.
3.4Power
Plug the instrument power cord into a mains outlet of the proper voltage, fre
quency, and current capability. Typically this will be 11 amps maximum at 115
VAC (±10%), 60 Hz [6 amps maximum at 230 VAC (±10%), 50 Hz]. Turn the
instrument on using the “POWER” switch underneath the unit. The instrument
will turn on and begin to heat to the previously programmed temperature
set-point. The front panel LED display will indicate the actual instrument
temperature.
3.5Setting the Temperature
Section 5.3 explains in detail how to set the temperature set-point on the furnace using the front panel keys. The procedure is summarized here.
(1) Press “SET” twice to access the set-point value.
(2) Press “UP” or “DOWN” to change the set-point value.
(3) Press “SET” to program in the new set-point.
(4) Press “EXIT” to return to the temperature display.
When the set-point temperature is changed the controller will switch the well
heater on or off to raise or lower the temperature. The displayed well tempera
ture will gradually change until it reaches the set-point temperature. The well
may require 10 to 75 minutes to reach the set-point depending on the span and
the scan rate. Another 15 minutes is required to stabilize within ±0.1°C of the
set-point. Ultimate stability may take 15 to 20 minutes of stabilization time.
-
-
-
12
3.6Changing Display Units
The instrument can display temperature in Celsius or Fahrenheit. The instru
ment is shipped from the factory set to Celsius. To change to Fahrenheit or
back to Celsius there are two ways:
Press “SET” and “UP” simultaneously. This will change the display units.
or
Press the “SET” key three times from the temperature display to show
Un = CPress the “UP” or “DOWN” key to change units.
-
Page 19
Press “SET” to store changes.
Changing Display Units
3 Quick Start
13
Page 20
4Parts and Controls
The user should become familiar with the 9260 furnace parts. Successful use of
the instrument is dependent upon knowledge of important components and
theirproperuse.
4.1Bottom Panel
The bottom panel consists of the removable power cord inlet, the power entry
module (PEM) and power switch, and the fan. See Figure 1.
1.The removable power cord inlet is located underneath the furnace and
plugs into an IEC grounded socket.
4 Parts and Controls
Bottom Panel
Figure 1 Bottom Panel
15
Page 21
9260 Mini Fixed Point Cell Furnace
User’s Guide
2.The power switch is located on the power entry module (PEM). The
PEM also houses the fuses. Models are available for either 115 VAC
(±10%) 60 Hz or 230 VAC (±10%) 50 Hz operation.
3.The cooling fan inlet is at the bottom of the unit. The cooling air circulat
ing through the furnace keeps the electronics and the chassis cool. Keep
the area immediately around the furnace clear to allow adequate
ventilation.
4.Three feet support the chassis permitting air space for the fan and access
to the power entry module and power switch.
4.2Front Panel
-
The front panel contains the digital display and the controller keypad. See Fig
ure 2.
1.The digital display is an important part of the temperature controller be
cause it not only displays set and actual temperatures but also displays
various instrument functions, settings, and constants. The display shows
temperatures in units according to the selected scale °C or °F.
2.The four button controller keypad allows easy setting of the set-point
temperature. The control buttons (SET, DOWN, UP, and EXIT) are used
to set the instrument temperature set-point, access and set other operating
parameters, and access and set calibration parameters.
Setting the control temperature is done directly in degrees of the current scale
and can be set to 0.01 of a degree Celsius or Fahrenheit.
The functions of the buttons are as follows:
-
16
Figure 2 Front Panel
Page 22
4 Parts and Controls
Front Panel
SET - Used to display the next parameter in the menu and to store parameters
to the displayed value.
DOWN - Used to decrement the displayed value of parameters.
UP - Used to increment the displayed value.
EXIT - Used to exit a function and to skip to the next function. Any changes
made to the displayed value are ignored. Holding “EXIT” for about 1/2 a sec
ond returns control to the main display.
-
Figure 3 Top Panel
17
Page 23
9260 Mini Fixed Point Cell Furnace
User’s Guide
4.3Top Panel
The primary feature of the top of the unit is the access to the temperature-con
trolled block. The top panel consists of the constant temperature block assem
bly, the pre-heat wells, the inner-melt heater, and cooling air vents. See Figure
3.
1.The constant temperature block assembly is where the cell basket con
taining the fixed-point cell is inserted. Or, when the furnace is used as a
temperature comparator, where the pre-drilled inserts are placed for in
serting the thermometers. The block assembly is made of a special alu
minum-bronze alloy that is resistant to the temperatures that the furnace
is capable of reaching.
Heaters surround the cell in order to provide uniform heat. The 9260 fea
tures adjustable top and bottom zone heaters that help to keep the tem
perature uniform over the entire fixed-point cell. These zone heaters add
heat to each end of the block where more heat is lost to ambient.
A high-temperature platinum resistance thermometer is imbedded into
the wall of the block to sense and control the temperature of the block.
This entire assembly is insulated and suspended in the airflow of the fan
to remove lost heat and to keep the chassis cool.
2.The thermometer pre-heat wells are located on either side of the block
access well. Thermometers are pre-heated in these wells prior to insertion into the cell in order to conserve its latent energy.
3.The inner-melt heater is a low-power heater that creates an inner liquid
layer next to the reentrant tube of the fixed-point cell during the melting
curve process. This inner-melt heater is controlled automatically by the
microprocessor when using the program mode, or may be used manually
with the switch in back. When not in use, the inner-melt heater is stored
in the well at the back of the unit to prevent it from causing any damage.
4.The cooling air vents in the top of the unit permit heated air to exit the
unit. Care must be taken not to touch these vents while the furnace is at
high temperatures or burns may result.
-
-
-
-
-
-
-
18
CAUTION: Areas on the top of the furnace may be very hot due to hot air
blowing upward. Please use caution.
4.4Rear Panel
The rear panel consists of the inner-melt heater connector, the inner-melt heater
switch, and the serial port. See Figure 4.
1.The inner-melt heater plugs into the rear of the furnace into the connec
tor provided. Be sure it is plugged in during operation.
-
Page 24
4 Parts and Controls
Rear Panel
Figure 4 Rear Panel
19
Page 25
9260 Mini Fixed Point Cell Furnace
User’s Guide
20
Figure 5 Thermal Block Assembly
Page 26
2.The inner-melt heater switch can be set to “MANUAL ON” position or
“AUTO” position. The microprocessor has control when the switch is in
the “AUTO” position.
3.The serial port is a DB-9 male connector for interfacing the instrument to
a computer or terminal RS-232 communications.
4.5Thermal Block Assembly
The thermal block assembly is shown in Figure 5 and described below.
4.5.1Thermal Block
The thermal block is specifically designed to contain the sealed cell and basket
containment assembly. This design permits the uniformity to be tuned carefully
for best performance and cell safety. Heaters in the perimeter of the alumi
num-bronze cylinder provide heat as dictated by the temperature controller. A
PRT sensor in the block monitors the block temperature and provides feedback
to the controller. Forced airflow around the assembly keeps the outside of the
furnace cool.
4.5.2Heaters
4 Parts and Controls
Thermal Block Assembly
-
The block assembly is heated by up to 1200 watts of heat. There are 3 heated
zones; the main zone heaters heat the entire length of the block and the top and
bottom zones heat their respective ends. The end zone heating compensates for
losses out of the unheated ends. These heaters are adjusted to provide the required temperature uniformity within the cell itself. The adjustments are made
by way of the keypad.
4.5.3Basket
The sealed cell is positioned within the basket which allows the cell to be easily
inserted and removed and contains all materials should the cell be broken. The
top and bottom parts of the basket are made of aluminum-bronze and provide
thermal shunting across the top and bottom as well. The lid and basket are re
moved separately with the tongs provided. A cell-pad in the bottom of the bas
ket helps to cushion the cell from the metal bottom.
4.5.4Thermal Shunt
A disk of aluminum-bronze in the top of the thermal well above the basket pro
vides heat transfer across the top of the cell and to the thermometer itself. The
effectiveness of the heat transfer to the thermometer is dependent on its fit to
the shunt. Measurements at the aluminum point may be affected by heat con
ducted up the thermometer stem.
-
-
-
-
21
Page 27
9260 Mini Fixed Point Cell Furnace
User’s Guide
4.5.5Insulation
The entire block assembly is surrounded by fiber-ceramic insulation. A remov
able portion above the cell permits the cell to be inserted and removed.
4.5.6Temperature Control Sensor
The temperature control sensor is a high quality PRT with 4 leads. Accuracy is
calibrated into the unit. Zero resistance, alpha, and delta coefficients of the Cal
endar-Van Dusen equation permit linearization over the desired temperature
range.
4.6Mini Fixed-point Cells
Mini Fixed-point Cells (Figure 6) utilize physical properties of a substance to
provide well established temperatures. The sample in the cell is placed into a
condition of multiple phases at a melting or freezing temperature or at a tri
ple-point temperature. While the sample substances are in this condition they
can exhibit very stable constant temperatures for long periods of time. Properly
used, the temperatures provided by these constants of nature are extremely precise and repeatable. The International Temperature Scale of 1990 (ITS-90) is
based on these principles. The ITS-90 temperatures defined at the freezing
points of Indium, Tin, Zinc, and Aluminum are among these and are achievable
with the Hart Scientific 9260 Furnace.
In order to achieve the ITS-90 temperatures and maintain long flat plateaus, the
substance samples (metals in this case) must be very pure. Typically 6 nines
purity is best. Sometimes 5 nines purity is used for the lower price but at the
cost of higher uncertainty and shorter plateaus. The 9260 furnace utilizes a
sealed mini-cell construction. The construction of the mini-cell follows the pattern of the full size cell. The high-purity sample is contained within a graphite
crucible. The graphite is free of contaminants and will not react with the metal
maintaining the metal purity. This material is all hermetically sealed within a
silica glass (quartz) envelope. The internal atmosphere is high purity argon. In
order to immerse the test thermometer into the high accuracy temperature zone
of the cell, a reentrant well is provided in the center of the cell.
These cells are manufactured in the same careful manner as their larger full size
counterparts. The certified high purity metal sample must maintain its purity
throughout the process of manufacture and use of the cell. To accomplish this,
high purity materials must be used for all the other components of the cell. Af
ter fabrication of each component, it must be treated to remove any impurities
that may have been added during the process. The components are assembled
in a clean environment and never touched directly by hand. After the compo
nents have been assembled and the silica glass permanently sealed, all of the air
is evacuated out while the cell is melted. Numerous cycles of vacuum and purg
ing with high purity argon are finally completed when the evacuation port is
sealed leaving approximately one atmosphere pressure of argon in the cell at
the freezing point of the sample.
-
-
-
-
-
-
22
Page 28
4 Parts and Controls
Mini Fixed-point Cells
Figure 6 Typical Sealed Mini Fixed-Point Cell
23
Page 29
9260 Mini Fixed Point Cell Furnace
User’s Guide
Temperature corrections must be made to the reading to account for actual cell
pressure and for hydrostatic pressure. During manufacturing, the cell is sealed
with argon near the 1 atmosphere pressure. At that time the actual pressure is
measured. With that pressure a small temperature correction can be calculated.
The hydrostatic pressure, created by the mass of the sample itself, depresses the
temperature of the reading. Since different cell designs and thermometer de
signs translate to different immersion depths, the practice is to calculate and
correct for the error. Figure 6 illustrates the maximum immersion depth within
the mini-cell. The actual immersion depth is taken to the center of the sensor el
ement of the thermometer. This depth will vary and the thermometer manufac
turer may need to be consulted. Approximations can be made for typical types
of thermometers since the hydrostatic error is small anyway and may be negli
gible for some requirements. Refer to the cell manual for the equations and
constants that need to be applied.
Due to the fragile nature of the fixed-point cells, extra care must be taken dur
ing use and handling. Do not handle it with bare hands, use clean cotton gloves
or equivalent. Make sure anything that comes in contact with the cell is clean.
To remove contaminants, wipe the cell down with a clean cloth and pure alcohol. Quartz glass is subject to a process called devitrification. The glass will
break down at high temperatures during this process. Oils in the skin and other
contaminants can initiate or accelerate this process.
Contaminants introduced to the reentrant well of the cell from unclean thermometers can cause the same problem. In addition, some types of metals can
contaminate the platinum sensor in a quartz SPRT at high temperatures (650°C
and up). Clean all thermometers prior to testing.
-
-
-
-
-
24
4.7Comparison Blocks
The 9260 furnace can function either as a calibrator or as a comparator. As a
calibrator, the calibration of the controller provides the reference temperature.
As a comparator, a reference thermometer value is compared to the values of
the units under test. A smaller uncertainty is obtainable with the comparison
method. Comparison blocks are available as options to the furnace.
4.7.1Block Well Sizes
Three standard comparison blocks are available. See Figure 7. Model
3160-2-provides 9 wells with clearance for 1/4-inch diameter thermometers.
Model 3160-3 is a combination of wells providing access for a variety of popu
lar sizes. Model 3160-1 is a blank block that can be drilled by the user to any
desired sizes.
4.7.2Comparison Block Assembly
Comparison blocks provide a uniform temperature between multiple thermom
eters. For accurate results, the thermometers must fit closely inside the well.
-
-
Page 30
4 Parts and Controls
Comparison Blocks
Figure 7 Comparison Blocks
25
Page 31
9260 Mini Fixed Point Cell Furnace
User’s Guide
The comparison block assembly is comprised of three components, the com
parison block, the thermal shunt, and the top insulation.
4.7.2.1Comparison Block
The comparison block is carefully lowered to the bottom of the thermal block
well with the tongs provided. DO NOT drop the block into the well. Damage
to the furnace may result. Small changes to the furnace calibration may re
sult as well.
4.7.2.2Thermal Shunt
Just above the comparison block is a ledge in the thermal block itself. The ther
mal shunt is lowered into the thermal well until it rests on this ledge. The inser
tion wells must match the comparison block. This block conducts heat from the
thermal block of the furnace in order to reduce stem conduction along the ther
mometer. Heat loss from stem conduction will reduce the accuracy of the
measurement.
4.7.2.3Top Insulation
A fiber ceramic pad of insulation is provided with the comparison block. This
insulation helps the thermal shunt by keeping it closer to the furnace temperature. After it is inserted, poke holes through it to match the comparison block.
-
-
-
-
-
26
Page 32
5Controller Operation
5 Controller Operation
Well Temperature
This section discusses in detail how to operate the furnace temperature control
ler using the front control panel. By using the front panel key-switches and
LED display, the user may monitor the well temperature, adjust the set-point
temperature in degrees C or F, monitor the heater output power, adjust the con
troller proportional band, and program the operating parameters, program pa
rameters, serial interface configuration, and the controller calibration
parameters. Operation of the functions and parameters is shown in the
flowchart in Figure 8 on page 28. This chart may be copied for reference.
In the following discussion a button with the word SET, UP, DOWN, or EXIT
inside indicates the panel button while the dotted box indicates the display
reading. Explanation of the button or display reading are to the right of each
button or display value.
5.1Well Temperature
The digital LED display on the front panel allows direct viewing of the actual
well temperature. This temperature value is normally shown on the display. The
units, C or F, of the temperature value are displayed at the right. For example,
100.00 CWell temperature in degrees Celsius
The temperature display function may be accessed from any other function by
pressing the “EXIT” button.
5.2Reset Cut-out
-
-
-
If the over-temperature cut-out has been triggered then the temperature display
will alternately flash,
Cut-outIndicates cut-out condition
The message continues to flash until the temperature is reduced and the cut-out
is reset. The cut-out has two modes - automatic reset and manual reset. The
mode determines how the cut-out is reset which allows the instrument to heat
up again. When in automatic mode, the cut-out will reset itself as soon as the
temperature is lowered below the cut-out set-point. With manual reset mode the
cut-out must be reset by the operator after the temperature falls below the
set-point.
When the cut-out is active and the cut-out mode is set to manual (“reset”) then
the display will flash “cut-out” until the user resets the cut-out. To access the
reset cut-out function press the “SET” button.
S
Access cut-out reset function
27
Page 33
9260 Mini Fixed Point Cell Furnace
User’s Guide
Figure 8 Flow Chart
28
Page 34
The display will indicate the reset function.
rESEt ?Cut-out reset function
Press “SET” once more to reset the cut-out.
5 Controller Operation
Temperature Set-point
S
This will also switch the display to the set temperature function. To return to
displaying the temperature press the “EXIT” button. If the cut-out is still in the
over-temperature fault condition the display will continue to flash “cut-out”.
The well temperature must drop a few degrees below the cut-out set-point be
fore the cut-out can be reset.
Reset cut-out
5.3Temperature Set-point
The temperature set-point can be set to any value within the range and resolu
tion as given in the specifications. Be careful not to exceed the safe upper temperature limit of any device inserted into the well.
Setting the temperature involves selecting one of the eight (8) set-points in
memory and then adjusting the set-point value.
5.3.1Programmable Set-points
The controller stores eight (8) set-point temperatures in memory. The set-points
can be quickly recalled to conveniently set the instrument to a previously programmed temperature set-point.
To set the temperature, first select the set-point memory. This function is accessed from the temperature display function by pressing “SET”. The number
of the set-point memory currently being used is shown at the left on the display
followed by the current set-point value.
-
-
100.00 CWell temperature in degrees Celsius
S
Access set-point memory
1. 100.Set-point memory 1, 100.0°C currently used
To change the set-point memory to another preset value press “UP” or
“DOWN”.
4. 300.New set-point memory 4, 300.0°C
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
29
Page 35
9260 Mini Fixed Point Cell Furnace
User’s Guide
NOTE: Pressing “SET” at this point turns off the program mode if it is
on.
5.3.2Set-point Value
The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”.
4 200.Set-point value in °C
If the set-point value is correct, hold “EXIT” to resume displaying the well
temperature. Press “UP” or “DOWN” to adjust the set-point value.
220.00New set-point value
When the desired set-point value is reached press “SET” to accept the new
value and to access the temperature scale units. If “EXIT” is pressed, any
changes made to the set-point are ignored.
S
Accept new set-point value
5.4Temperature Scale Units
The temperature scale units of the controller can be set by the user to degrees
Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well temperature, set-point, and proportional band.
Press “SET” after adjusting the set-point value to change display units.
Un= CScale units currently selected
Press “UP” or “DOWN” to change the units.
Un= FNew units selected
5.5Scan
The scan rate can be set and enabled so that when the set-point is changed the
furnace heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. With the scan disabled the furnace heats or cools at the maxi
mum possible rate.
5.5.1Scan Control
-
30
The scan is controlled with the scan on/off function that appears in the main
menu after the set-point function.
Page 36
Sc=OFFScan function off
Press “UP” or “DOWN” to toggle the scan on or off.
Sc=OnScan function on
Press “SET” to accept the present setting and continue.
5 Controller Operation
Program Advance
S
Accept scan setting
5.5.2Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set
from 0.1 to 99.9°C/minute. The maximum scan rate, however, is actually lim
ited by the natural heating or cooling rate of the instrument. This is often less
than 100 °C/minute, especially when cooling.
The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees per minute, degrees C or F depending on the
selected units.
Sr= 10.0Scan rate in °C/min.
Press“UP”or“DOWN”tochangethescanrate.
Sr= 2.0New scan rate
Press “SET” to accept the new scan rate and continue.
S
Accept scan rate
5.6Program Advance
The program advance function allows the user to step through the maintain,
freeze, and melt operations of the fixed-point realization. They are explained in
detail in Section 8, 7 on Fixed-Point Realization.
-
S+D
Access program advance
Adv“Adv” flashes
MAINTDisplays one of the functions MAINT, FREEZE, MELT, or
STOP
Press “UP” or “DOWN” to view the desired function.
S
Accepts the new the operation
31
Page 37
9260 Mini Fixed Point Cell Furnace
User’s Guide
5.7Temperature Scale Units
To toggle between °C and °F, press the “SET” and “UP” keys simultaneously
when the temperature is displayed.
5.8Secondary Menu
Functions which are used less often are accessed within the secondary menu.
Pressing “SET” and “EXIT” simultaneously and then releasing accesses the
secondary menu. The first function in the secondary menu is the heater power
display. (See Figure 8 on page 28.)
5.9Heater Power
The temperature controller controls the temperature of the well by pulsing the
heater on and off. The total power being applied to the heater is determined by
the duty cycle or the ratio of heater on time to the pulse cycle time. By knowing
the amount of heating the user can tell if the instrument is heating up to the
set-point, cooling down, or controlling at a constant temperature. Monitoring
the percent heater power lets the user know the stability of the well temperature. With good control stability the percent heating power should not fluctuate
more than ±1% within one minute.
The heater power display is accessed in the secondary menu. Press “SET” and
“EXIT” simultaneously and release. The heater power is displayed as a percentage of full power.
32
100.00Well temperature
S+E
Access heater power in secondary menu
SECFlashes
12.0 PHeater power in percent
To exit out of the secondary menu press “EXIT” and hold for a brief moment.
To continue on to the proportional band setting function press “EXIT” momen
tarily or “SET”.
5.10Set-point Resistance
The set-point resistance is the resistance of the temperature sensor at the cur
rent temperature. Allow the temperature to stabilize at the desired set-point be
fore taking its resistance. In order to calibrate the furnace temperature, the
set-point resistance must be displayed.
-
-
-
Page 38
5 Controller Operation
Proportional Band
Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” twice to access the set-point resistance
S+E
Access heater power in secondary menu
SECFlashes “” and then displays the heater power setting
12.0 PHeater power in percent
S
Access set-point resistance
rSFlashes “” (Set-point Resistance) and then displays the
setting
160.095Resistance in ohms
5.11Proportional Band
In a proportional controller such as this, the heater output power is proportional
to the well temperature over a limited range of temperatures around the
set-point. This range of temperature is called proportional band. At the bottom
of the proportional band the heater output is 100%. At the top of the proportional band the heater output is 0. Thus as the temperature rises the heater
power is reduced, which consequently tends to lower the temperature back
down. In this way the temperature is maintained at a fairly constant
temperature.
The temperature stability of the well and response time depend on the width of
the proportional band. If the band is too wide the well temperature deviates ex
cessively from the set-point due to varying external conditions. This deviation
is because the power output changes very little with temperature and the con
troller does not respond well to changing conditions or noise in the system. If
the proportional band is too narrow the temperature may swing back and forth
because the controller overreacts to temperature variations. For best control sta
bility the proportional band must be set for the optimum width.
The proportional band width is set at the factory to about 5.0°C. The propor
tional band width may be altered by the user to optimize the control character
istics for a particular application.
The proportional band width is easily adjusted from the front panel. The width
may be set to discrete values in degrees C or F depending on the selected units.
The proportional band adjustment can be accessed within the secondary menu.
Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” twice to access the proportional band.
-
-
-
-
-
S+E
Access heater power in secondary menu
33
Page 39
9260 Mini Fixed Point Cell Furnace
User’s Guide
SECFlashes “” and then displays the heater power setting
12.0 PHeater power in percent
S
Access set-point resistance
rSFlashes “” (Set-point Resistance) and then displays the
setting
160.095Resistance in ohms
S
ProPFlashes “” and then displays the setting
5.0Proportional band setting
To change the proportional band press “UP” and “DOWN”.
4.0New proportional band setting
To store the new setting press “SET”. Press “EXIT” to continue without storing
the new value.
S
5.12Controller Configuration
The controller has a number of configuration and operating options and calibra
tion parameters that are programmable via the front panel. These are accessed
from the secondary menu after the proportional band function by pressing
“SET”. “ConFiG” flashes and then the name of the first parameter menu
“PAR” is displayed.Pressing “SET” again enters the first of four groups of con
figuration parameters: operating parameters, program parameters, serial inter
face parameters, and calibration parameters. The groups are selected using the
“UP” and “DOWN” keys and then pressing “SET”. (See Figure 8 on page 28)
Accept the new proportional band setting
-
-
-
34
5.13Operating Parameters
The operating parameters menu is indicated by,
PArOperating parameters menu
Page 40
Press “SET” to enter the menu. The operating parameters menu contains the
High Limit (HL) parameter, the Soft Cut-out parameter, and the Cut-out Reset
mode parameter.
5.13.1High Limit
The High Limit parameter adjusts the upper set-point temperature. The factory
default and maximum are set to 680°C. For safety, a user can adjust the High
Limit parameter down so the maximum temperature set-point is restricted.
HLFlashes “” (High Limit parameter) and then displays
H=680Current HL setting
Adjust the HL parameter using “UP” or “DOWN”
H=600New High Limit setting
Press “SET” to accept the new High Limit parameter and to access the Soft
Cut-out parameter.
5 Controller Operation
Operating Parameters
the setting
5.13.2Soft Cut-out
The next parameter in this menu is the Soft Cut-out. The Soft Cut-out parameter is used by the controller to shut down the unit during over-temperature conditions. If the temperature of the unit is ever greater than the Soft Cut-out
temperature the controller shuts itself down and displays, alternately,
“SCtOut”and“Err 8".
SoFtCoFlashes “” (Soft Cut-out parameter) and then dis
plays the setting
705Current value
Adjust this parameter by using “UP” or “DOWN”.
700New Soft Cut-out setting
Press “SET” to accept the new parameter and to access the Cut-out Reset
Mode.
5.13.3Cut-out Reset Mode
The final parameter in this menu is the Cut-out Reset Mode. The Cut-out Reset
Mode determines whether the cut-out resets automatically when the well tem
perature drops to a safe value or must be manually reset by the operator.
-
-
35
Page 41
9260 Mini Fixed Point Cell Furnace
User’s Guide
CtorStFlashes “” (Cut-out reset mode parameter) and
AutoCurrent setting
To change to manual reset mode press “UP” or “DOWN”.
rStNew Cut-out reset for manual reset
Press “SET” to accept the new parameter.
5.14Program Parameters
The program parameters menu is indicated by,
ProgProgram parameters menu
Press “SET” to enter the menu. The Program parameters menu contains the
fixed-point parameter, the curve parameter, and the curve temperature
parameter.
then displays the setting
5.14.1Fixed-point
The first parameter in this menu is the Fixed-point parameter.The Fixed-point
parameter allows the user to select the fixed-point metal. The available options
are In (Induim), Sn (Tin), Zn (Zinc), Al (Aluminum), or Other.
FPFlashes “” (Fixed-point parameter) and then displays
InCurrent Fixed-point setting (Indium)
Adjust the Fixed-point parameter by using “UP” or “DOWN”.
2nNew Fixed-point setting (Zinc)
Press “SET” to accept the new Fixed-point parameter and to access the Curve
parameter.
5.14.2Curve
The next parameter in this menu is the Curve parameter. The Curve parameter
is selected as either melting or freezing.
CurveFlashes “” (Curve parameter) and then displays
the setting
the setting
36
Page 42
MeLtCurrent Curve setting
Adjust this parameter by using “UP” or “DOWN”.
FrEE2ENew Curve setting
5 Controller Operation
Program Parameters
Press “SET” to accept the new Curve parameter and to access the Curve Tem
perature parameter.
5.14.3Curve Temperature
The final set of three parameters in this menu are the Curve Temperature pa
rameters. These parameters set the temperature of the curve set-points.
TemPFlashes “” (Curve Temperature parameter) and then
displays the melt parameter
MELtFlashes “” and then displays the value
100.00Current melt temperature setting
Adjust this parameter by using “UP” or “DOWN”. Press “SET” to accept the
new value and to display the freeze parameter.
S
FREEZEFlashes “” and then displays the value
25.00Current freeze temperature setting
Adjust this set-point by using “UP” or “DOWN”. Press “SET” to accept the
new value and to display the maintain parameter
-
-
S
MAintFlashes “” and then displays the value
148.59Current maintain temperature setting.
Adjust this set-point by using “UP” or “DOWN”.
Press “SET” to accept the new value.
37
Page 43
9260 Mini Fixed Point Cell Furnace
User’s Guide
5.15Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SEriALSerial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to instruments fitted
with the serial interface. The parameters in the menu are: BAUD rate, sample
period, duplex mode, and linefeed.
5.15.1Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting deter
mines the serial communications transmission rate. The baud rate of the serial
communications may be programmed to 300, 600, 1200, 2400, 4800, or 9600
baud. 2400 baud is the default setting.
bAUdFlashes “” (Serial baud rate parameter) and then
2400 bCurrent baud rate
Adjust the baud rate by using “UP” or “DOWN”.
4800 bNew baud rate
Press “SET” to store the baud rate to the new value and to access the Sample
Period.
5.15.2Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure
ments transmitted from the serial interface. If the sample rate is set to 5, the in
strument transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0.
-
displays the setting
-
-
38
SPErFlashes ”” (Serial sample period parameter) and
then displays the setting
SP=1Current sample period (seconds)
Adjust the value with “UP” or “DOWN”.
SP=60New sample period
Page 44
5 Controller Operation
Calibration Parameters
Press “SET” to store the sample period to the new value and to access the Du
plex Mode.
5.15.3Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du
plex or half duplex. With full duplex any commands received by the instrument
via the serial interface are immediately echoed or transmitted back to the device
of origin. With half duplex the commands are executed but not echoed.
dUPLFlashes “” (Serial duplex mode parameter) and then
d=FULLCurrent duplex mode setting
The mode may be changed using “UP” or DOWN".
d=HALFNew duplex mode setting
Press “SET” to store the duplex mode to the new value and to access the the
Linefeed.
5.15.4Linefeed
The final parameter in the serial interface menu is the linefeed mode. This parameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return.
-
-
displays the setting
LFFlashes “” (Serial linefeed parameter) and then dis-
plays the setting
LF=OnCurrent linefeed setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF=OFFNew linefeed setting
Press “SET” to store the new linefeed value.
5.16Calibration Parameters
The operator of the instrument has access to a number of the calibration con
stants namely R0, ALPHA, DELTA, top and bottom zone percent heat, and the
hard cut-out. These values are set at the factory and must not be altered. The
correct values are important to the accuracy and proper and safe operation of
the furnace. Access to these parameters is available to the user only so that in
the event that the controller memory fails the user may restore these values to
-
39
Page 45
9260 Mini Fixed Point Cell Furnace
User’s Guide
the factory settings. The user should have a list of these constants and their set
tings with manual.
CAUTION: DO NOT change the values of the instrument calibration con
stants from the factory set values. The correct setting of these parameters
is important to the safety and proper operation of the instrument.
The calibration parameters menu is indicated by,
CALCalibration parameters menu
Press “SET” five times to enter the menu.
The calibration parameters R0, ALPHA, and DELTA characterize the resis
tance-temperature relationship of the platinum control sensor. These parameters
may be adjusted by an experienced user to improve the accuracy of the furnace.
CAUTION: Care should be exercised when adjusting these parameters
since they affect the accuracy of the set-point value. This procedure is explained in detail in Section 8.
5.16.1Hard Cut-out
This parameter is the temperature above which the unit shuts down automatically. The parameter is set at the factory to approximately 700°C and cannot be
changed by the user.
-
-
-
40
5.16.2R
0
This probe parameter refers to the resistance of the control probe at 0°C. The
value of this parameter is set at the factory for best instrument accuracy.
5.16.3ALPHA
This probe parameter refers to the average sensitivity of the probe between 0
and 100°C. The value of this parameter is set at the factory for best instrument
accuracy.
5.16.4DELTA
This probe parameter characterizes the curvature of the resistance-temperature
relationship of the sensor. The value of this parameter is set at the factory for
best instrument accuracy. should have a list of these constants and their settings
with manual.
5.16.5Top and Bottom Zone Percent Heating
The top and bottom zone heaters of the furnace start heating each time the main
zone heater turns on. These end zone heaters are adjusted to a percentage of on
Page 46
5 Controller Operation
Calibration Parameters
time compared to the main zone. Each zone is pre-adjusted according to a cali
bration procedure that provides a low vertical temperature gradient in the ther
mal block.
5.16.5.1Bottom Zone
botPCt is the percentage of the main heater value for the bottom zone. Do not
adjust this value unless you are following the procedure in Section 8 Calibra
tion Procedure, of this manual. Adjustment values range from 0 to 200 percent.
5.16.5.2Top Zone
toPPCt is the percentage of the main heater value for the top zone. Do not ad
just this value unless you are following the procedure in Section 8 Calibration
Procedure, of this manual. Adjustment values range from 0 to 200 percent.
-
-
-
-
41
Page 47
6 Digital Communication Interface
6Digital Communication Interface
The furnace is capable of communicatingwithandbeingcontrolledbyother
equipment through the digital serial interface.
With a digital interface the instrument may be connected to a computer or other
equipment. This allows the user to set the set-point temperature, monitor the
temperature, and access any of the other controller functions, all using remote
communications equipment. Communications commands are summarized in
Table 4 on page 46.
6.1Serial Communications
The instrument is installed with an RS-232 serial interface that allows serial
digital communications over fairly long distances. With the serial interface the
user may access any of the functions, parameters and settings discussed in Sec
tion 5 with the exception of the BAUD rate setting.
6.1.1Wiring
The serial communications cable attaches to the instrument
through the DB-9 connector at
the back of the instrument. Figure 9 shows the pin-out of this
connector and suggested cable
wiring. To eliminate noise the
serial cable should be shielded
with low resistance between the
connector (DB-9) and the
shield. If the unit is used in a
heavy industrial setting, the se
rial cable must be limited to
ONE METER in length.
-
Serial Communications
-
6.1.2Setup
Before operation the serial in
terface must first be set up by
programming the BAUD rate
and other configuration parame
ters. These parameters are pro
grammed within the serial
interface menu.
To enter the serial parameter
programming mode first press
“EXIT” while pressing “SET”
-
-
-
Figure 9 Serial Cable Wiring
43
Page 48
9260 Mini Fixed Point Cell Furnace
User’s Guide
and release to enter the secondary menu. Press “SET” repeatedly until the dis
play reads “PAr”. Press “UP” until the serial interface menu is indicated with
“SErIAL”. Finally press “SET” to enter the serial parameter menu. In the serial
interface parameters menu are the BAUD rate, the sample rate, the duplex
mode, and the linefeed parameter.
6.1.2.1Baud Rate
The baud rate is the first parameter in the menu. The display will prompt with
the baud rate parameter by showing “bAUd”. Press “SET” to choose to set the
baud rate. The current baud rate value will then be displayed. The baud rate of
the 9260 serial communications may be programmed to 300, 600, 1200, 2400,
4800, or 9600 baud. The baud rate is pre-programmed to 2400 baud. Use “UP”
or “DOWN” to change the baud rate value. Press “SET” to set the baud rate to
the new value or “EXIT” to abort the operation and skip to the next parameter
in the menu.
6.1.2.2Sample Period
The sample period is the next parameter in the menu and prompted with
“SPEr”. The sample period is the time period in seconds between temperature
measurements transmitted from the serial interface. If the sample rate is set to
5, the instrument transmits the current measurement over the serial interface approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. Press “SET” to choose to set the sample period. Adjust the
period with “UP” or “DOWN” and then use “SET” to set the sample rate to the
displayed value.
-
44
6.1.2.3Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex
mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full
duplex any commands received by the instrument via the serial interface are
immediately echoed or transmitted back to the device of origin. With half du
plex the commands are executed but not echoed. The default setting is full du
plex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
6.1.2.4Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa
rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac
ter (LF, ASCII 10) after transmission of any carriage-return. The default setting
is with linefeed on. The mode may be changed using “UP” or “DOWN” and
pressing “SET”.
6.1.3Serial Operation
Once the cable has been attached and the interface set up properly the control
ler immediately begins transmitting temperature readings at the programmed
rate. The serial communications uses 8 data bits, one stop bit, and no parity.
-
-
-
-
-
Page 49
The set-point and other commands may be sent via the serial interface to set the
temperature set-point and view or program the various parameters. The inter
face commands are discussed in Section 6.2. All commands are ASCII charac
ter strings terminated with a carriage-return character (CR, ASCII 13).
6.2Interface Commands
6 Digital Communication Interface
Interface Commands
-
-
The various commands for accessing the instrument functions via the digital in
terface are listed in this section (see Table 4). These commands are used with
the RS-232 serial interface. The commands are terminated with a car
riage-return character. The interface makes no distinction between upper and
lower case letters, hence either may be used. Commands may be abbreviated to
the minimum number of letters that determines a unique command. A com
mand may be used to either set a parameter or display a parameter depending
on whether or not a value is sent with the command following a “=” character.
For example, “s” <cr> will return the current set-point and “s=150.0" will set
the set-point to 150.0 degrees.
In the following list of commands, characters or data within brackets, “[” and
“]”, are optional for the command. A slash, “/”, denotes alternate characters or
data. Numeric data, denoted by “n”, may be entered in decimal or exponential
notation. Characters are shown in lower case although upper case may be used.
Spaces may be added within command strings and will simply be ignored.
Backspace (BS, ASCII 8) may be used to erase the previous character. A terminating CR is implied with all commands.
-
45
Page 50
9260 Mini Fixed Point Cell Furnace
User’s Guide
Table 4 Communications Command Summary
Command Description
Command
Format
Command
ExampleReturned
Returned
Example
Acceptable
Values
Display Temperature
Read unitsuuu:{C or F}u: C
Read current set-points[etpoint]sset: 9999.99 {C or F}set: 150.00 C
Set current set-point to ns[etpoint]=ns=450Instrument Range
Read temperaturettt: 9999.99{C or F}t: 478.03 C
Set temperature units:u[nits]=c/f
Set temperature units to Celsiusu[nits]=cu=c
Set temperature units to Fahrenheitu[nits]=fu=f
Read scan functionsc[an]scscan: {ON or OFF}scan: ON
Set scan function:sc[an]=on/of[f]
Turn scan function onsc[an]=onsc=on
Turn scan function offsc[an]=of[f]sc=of
Read scan ratesr[ate]srsrat: 999.99 {C or F}/minsrat: 10.0 C/min
Setscanrateto
Read program controlpcpcpc:{[STOP/FREEZE/ME[LT]/MA[
Set program controlpc={stop/freeze/me[lt]/ma[intain]}pc=meltSTOP, FREEZE,
n
degrees per minutesr[ate]=
n
sr=5.1 to 99.9
INTAIN]}
pc:FREEZE
CorF
ON or OFF
MELT or MAINTAIN
Secondary Menu
Read heater power
(duty cycle)
Read proportional band settingpr[op-band]prpb: 999.9pb: 15.9
Set proportional band to
Read set-point resistance*sr*sr999.999121.091
n
po[wer]popo: 999.9po: 1
pr[op-band]=npr=8.830.1 to 100
Configuration Menu
Operating Parameters Menu
Read high limithlhlhl: 9999hl: 925
Set high limithl=nhl=900100–1200
Read soft cut-outcucucu: 9999.9cu: 1150
Set soft cut-out setting:cu[tout]=n
Set soft cut-out to
Read cut-out modecm[ode]cmcm:{xxxx}cm: AUTO
Set cut-out modecm[ode]=r[eset]/a[uto]Reset or Auto
Set cut-out to be reset manuallycm[ode]=r[eset]cm=r
Set cut-out to be reset automaticallycm[ode]=a[uto]cm=a
Program Parameters Menu
Read fixed-point parameterfpfpfp:{INDIUM, TIN,
n
degreescu[tout]=ncu=5000.0 to 1150.0
ZINC,ALUMINUM}
fp:INDIUM
46
Page 51
Communications Command Summary cont.
6 Digital Communication Interface
Interface Commands
Command Description
Set fixed-point parameterfp=[INDIUM/TIN/ZINC/
Read curve parametercrvcrvcrv:{MELT or FREEZE}crv:MELT
Set curve parametercrv={melt/freeze}crv=meltMELT or FREEZE
Read top zone % heatingtpcttpcttpct: 999.9tpct: 200.0
Set top zone % heatingtpct=
Read bottom zone % heatingbpctbpctbpct: 999.9bpct: 150.0
Set bottom zone % heatingbpct=
These commands are only used for factory testing.
Miscellaneous (not on menus)
Read firmware version number*ver[sion]*verver.9999,9.99ver.9260,v1.1
Read structure of all commandsh[elp]hlist of commands
Legend:[] Optional Command data
Note:When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return and linefeed.
n
secondssa[mple]=
n
Format
ALUMINUM]
n
n=n
n
r[0]=
n
n
al[pha]=
n
n
n
n
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Then the value is returned as indicated in the RETURNED column.
Command
Command
ExampleReturned
fp=InINDIUM, TIN, ZINC,
ps3psn: 999.99{C or F}ps1: 480.00 C
ps3=1001 to 3, dependent on
sa=00 to 4000
r=100.798.0 to 104.9
al=0.003865.002 to .006
de=1.370to3
tpct=1000to200
bpct=150.00to200
Returned
Example
Acceptable
Values
ALUMINUM
program control
FULL or HALF
ON or OFF
47
Page 52
7Fixed-Point Realization
7.1General
Either a freezing or melting plateau may be realized. The melting point is faster
and easier than the freezing point method and the plateau can last longer. Prop
erly done, accuracy of measurements is nearly the same.
Realizing the fixed-point temperature is a matter of achieving the ITS-90 de
fined temperature through a careful process. The Hart Scientific Model 9260
furnace features an internally programmed method or a manual method to per
form this process.
The internal programming procedure requires the user to:
1.Select the point (Indium, Tin, Zinc, Aluminum or a user programmed
point) to be realized.
2.Select either the freezing curve or melting curve mode.
3.The controller then provides pre-programmed scan rates, set-points for
each step in the process, dwell times, and an indication that the next step
is ready. Some of these pre-programmed conditions can be modified by
the user.
Temperature calibration should be checked occasionally to verify the set-point
temperatures.
7 Fixed-Point Realization
General
-
-
-
7.2Installing a Sealed Cell into the Basket
The sealed cell is a very delicate instrument. The quartz glass on the outside of
the cell can be easily broken and easily scratched by the harder metal materials
into which it is inserted.
CAUTION: Sealed cells are designed to be kept upright. They should be
stored in this position as well. Before inserting the cell be sure it is clean
and free of finger oils. Use pure or reagent grade alcohol to clean all
pieces. Quartz glass will devitrify breaking down the glass and eventually
causing breakage or air leakage to the cell. Clean the basket surfaces as
well. Use cotton gloves to handle the basket and cell after cleaning.
Each cell should have its own basket assembly. Repeated insertion and re
moval poses an unnecessary risk to the cell. A cell-pad of fiber ceramic fi
ber or quartz wool should be in the bottom of the basket to pad the cell.
This cell-pad should not be too thick so as to prevent proper fitting of the
basket cover.
Follow these steps to properly insert the sealed cell. Refer to Figure 5 for loca
tion of component detail.
-
-
-
49
Page 53
9260 Mini Fixed Point Cell Furnace
User’s Guide
1.Install the cell-pad into the bottom of the basket. Typically a 1/4" (6.35
mm) thick pad is used. It will crush to a thinner dimension as the cell is
installed. It SHOULD NOT be so thick that the basket cover touches the
cell at the top. If the pad contains an organic binder, heat the pad to re
move it before use.
2.Cut a strip of clean paper approximately 1-inch wide and 1-inch longer
than the basket.
3.Insert this paper into the bottom of the basket while it is lying down. At
least 1-inch of the paper should be extending out of the basket for easy
removal.
4.Temporarily hold the sealed cell sideways with the spherical end toward
the basket opening.
5.Carefully insert the cell into the basket sliding it on the paper instead of
on the metal basket. When the cell reaches the bottom, return the basket
carefully to the upright position.
6.Remove the paper completely.
7.Install a thin insulating pad on top of the cell providing a clearance hole
for the thermometer.
8.Check the basket cover fit to ensure that the evacuation port on the top of
the cell does not prevent it from fitting flush against the top of the basket.
9.Remove the lid for now.
10. Using the tool provided, carefully lower the basket into the thermal well.
11. Install the basket lid on top of the basket being sure that it is flush
against the basket.
12. Install a thin insulation pad on top of the lid with a clearance hole for the
thermometer.
13. Install the thermal shunt over the top of the basket fitting it onto the
ledge above.
14. Check the alignment.
15. Fit the top insulation above the shunt.
16. Carefully check the thermometer fit all the way to the bottom of the cell.
There must not be any significant resistance to insertion.
17. Place a small pad of quartz wool at the bottom of the reentrant well of
the cell which will, to some extent, pad the bottom of the well against the
thermometers when they are inserted.
-
50
CAUTION: Careless insertion of a thermometer into the well can break
the quartz glass at the bottom or even break quartz thermometers.
Page 54
CAUTION: Occasionally remove the thermal shunt and the basket. If
there is any resistance to removal, remove oxides with fine grit sandpaper.
This is generally only a problem at the aluminum point.
7.3Melting Point Realization
Recent improvements of the melting point method have shown that excellent
accuracy can also be achieved with this method. With this method, the cell tem
perature is carefully raised to a temperature just below the melting point. The
furnace and cell are allowed to dwell at this temperature for a time to allow
temperature equalization throughout the system. The furnace is then raised to a
point a few degrees above the melting point for a short time to give melting a
good start. The temperature is then dropped to a temperature just above the
melting point for the duration of the measuring period. A “melt heater” inside
the reentrant well is turned on for a short time creating a second zone of melted
sample which is next to the thermometer during measurements. This action per
mits more accurate measurements and is similar to the technique used with a
Triple-point of Water cell.
Measurements are made until the plateau begins to deviate and then the melting
must be completed and the process started over.
7 Fixed-Point Realization
Melting Point Realization
-
-
7.3.1Melting Point Procedure
The following procedure illustrates the steps required to successfully realize a
51
Page 55
9260 Mini Fixed Point Cell Furnace
User’s Guide
fixed-point temperature through melting the cell sample. Figure 10 illustrates
the process graphically.
Figure 10 Melting Point Realization
7.3.1.1Preparation
The cell should be carefully loaded into the basket and then into the furnace ac
cording to the procedure outlined in Section 8.2, 7.2. Thermometers that are to
be inserted into the cell should first be cleaned with pure alcohol. The in
ner-melt heater should be inserted carefully into the cell.
7.3.1.2Setting Up The Controller
Setup of the controller consists of selecting the fixed-point and the curve to be
used. The setup is in the secondary menu of the controller menus. Access the
setup by pressing “SET” and “EXIT”. Then press “SET” 4 times and “UP” un
til “ProG” is displayed. Press “SET”. “FP” (fixed-point) flashes and then the
current setting is displayed. This setting is the one used on the previous test.
Press “UP” or “DOWN” repeatedly to toggle through Indium, Tin, Zinc, Alu
minum and Other. When you reach the desired fixed-point press “SET” to se
lect it. Immediately after pressing “SET”, “CURVE” flashes and the current
setting appears. The available curves are melt and freeze. Use the “UP” or
“DOWN” key to select the “MELT” curve. Press “SET” to select it. Following
the curve selection, the term “TEMP” flashes on the display followed by adjust
ments that can be made by the user to the MELT, FREEZE, and MAINTAIN
temperatures. These adjustments are only for experienced users. Press
52
-
-
-
-
-
-
Page 56
“EXIT” to exit the program menu and to return to the display temperature.
Carefully insert the melt heater into the cell (see Figure 3 ). Set the heater
switch on the rear of the furnace to AUTO. The heater must be clean and care
fully inserted into the cell to prevent damage. This should be done before
program initiation.
7.3.1.3Program Initiation
Now that the fixed-point and the melt curve have been selected, the furnace is
ready to initiate the program. Advance to the program from the display temper
ature by pressing “SET” and “DOWN”. The term “StOP” appears unless the
program is running. The following three steps include “MELt”, “MAInt”and
“FrEEzE”. The step that first appears will be the last step utilized and not nec
essarily the first step desired. Press the “UP” or “DOWN” keys to view the first
step needed which is “MAInt” for the preliminary heating of the cell to just be
low the melting point. The cell is held at that temperature for a period of time
to allow everything to equilibrate to that temperature. Press “SET” to select it.
The controller immediately starts heating the cell with its metal sample. The
furnace heats at a preprogrammed rate that automatically slows down before
the sample reaches the “MAINT” point. Once the furnace has reached that temperature (about 1°C below the melting point), it dwells there for 60 minutes to
permit the furnace and metal sample to equilibrate. After the 60 minute period,
the display flashes on and off alternately indicating that the cell is now ready to
initiate the melt function.
To initiate the melting of the sample, press “SET” and “DOWN” again. Press
the“UP”or“DOWN”keysuntiltheterm“MELt” is on the display. Press
“SET” to select that step and the controller selects a set-point temperature a few
degrees above the melting point to begin the sample melting process. The temperature of the furnace slowly scans to about 4°C above the melting point and
dwells there for about 8 minutes. At that time the melt heater turns on automat
ically. This heater melts a thin layer of the sample next to the measurement
thermometer increasing the accuracy of this technique. After a few minutes the
melt heater automatically turns off and the furnace temperature drops to a tem
perature just above the melting point of the sample. Remove the melt heater
and insert a pre-heated monitor thermometer. When the temperature has stabi
lized, calibrations may begin. Use the pre-heat wells to heat up the thermome
ters before inserting them into the cell. This action preserves the latent energy
and permits more calibrations during the melting plateau. Allow the thermome
ters to equilibrate for 20 minutes before making readings. The plateau can last
several days depending on how it is used.
When the plateau has ended, the furnace may be turned to about 25°C and
turned off or, if desired, the process may be repeated. To repeat the process,
press “SET” and “DOWN” and use the “UP” or “DOWN” keys to find the
“FrEEzE” step. Select by pressing “SET”. The furnace temperature will drop
to a temperature 8°C below the freezing point and dwell there as before. The
sample will freeze and the temperature will automatically rise to a temperature
just below the melting point. The furnace will hold there for a time until the
cell and furnace have fully equilibrated. At that time, the display will flash indi
7 Fixed-Point Realization
Melting Point Realization
-
-
-
-
-
-
-
-
-
-
53
Page 57
9260 Mini Fixed Point Cell Furnace
User’s Guide
cating that the cell is ready to begin the melting process again. It is initiated as
before by selecting “MELt”step.
7.4Freezing Point Realization
The freezing point is an established ITS-90 method. This method first entirely
melts the metal sample. The furnace temperature is then carefully reduced to a
temperature below the freezing point just low enough to induce freezing of the
sample. This temperature must be several degrees below the freezing point in
order to overcome the sub-cooling of the sample. After recalescense, (initiation
of the freeze as observed on a monitor thermometer) the furnace temperature is
increased to a temperature just slightly below the freezing point. This action re
stricts the amount of heat that can flow from the cell, which controls the rate of
freezing. High temperature stability and good temperature uniformity permit
the temperature to be adjusted closely achieving long freezing plateaus. Long
freezing plateaus translate into a larger number of calibrations that can be done
during one freeze.
Measurements are made until the plateau begins to deviate and then the freezing must be completed and the process started over.
7.4.1Freezing Point Procedure
-
The following procedure illustrates the steps required to successfully realize a
54
Page 58
7 Fixed-Point Realization
Freezing Point Realization
fixed-point temperature through freezing the cell sample. Figure 11 illustrates
the process graphically.
Figure 11 Freezing Point Realization
7.4.1.1Preparation
The cell should be carefully loaded into the basket and then into the furnace ac
cording to the procedure previously outlined in Section 8.2, 7.2. Thermometers
that are to be inserted into the cell should first be cleaned with pure alcohol. A
monitor thermometer should be inserted into the cell to monitor the process
from the beginning.
7.4.1.2Setting Up The Controller
Setup of the controller consists of selecting the mode and fixed-point to be
used. The setup is in the secondary menu of the controller menus. Access the
program parameters by pressing “SET” and “EXIT” simultaneously. Then
press “SET” 4 times and “UP” until the program menu is reached. The word
“ProG” is displayed. Press “SET”. “FP” (fixed-point) flashes and then the
current setting is displayed. (The metal sample displayed is from the previous
test.) Press “UP” or “DOWN” repeatedly to toggle through Indium, Tin, Zinc,
Aluminum and Other. When you reach the desired fixed-point press “SET” to
select the metal sample to use. Immediately after pressing “SET”, “CURVE”
-
55
Page 59
9260 Mini Fixed Point Cell Furnace
User’s Guide
flashes and the current setting “MELt”or“FrEEZE” appears. The terms indi
cate the first of the two curves, melt or freeze. Use the “UP” or “DOWN” key
to view the desired curve, “FrEEZE”. Press “SET” to select this curve. Press
“EXIT” to exit the program menu and to return to the display temperature.
7.4.1.3Program Initiation
Now that the fixed-point and the freeze curve have been selected, the furnace is
ready to initiate the program. Advance to the program from the display temper
ature by pressing “SET” and “DOWN”. One of three steps of the selected pro
gram curve or “StOP” is displayed. The three steps include “MELt”, “MAInt”
and “FrEEzE”. The step that first appears will be the last step utilized and not
necessarily the first step desired. Press the “UP” or “DOWN” keys to view the
first step needed which is “MELt” for the freezing process. Press “SET” to
select it.
NOTE: Select stop to turn off the program function. Manually adjusting
the set-point also stops the program.
The controller immediately starts heating the cell. The furnace heats at a preprogrammed rate that automatically slows down before the sample reaches the
melting point. Once the furnace has reached the melting temperature (about
8°C above the melting point), it dwells there for 90 minutes to permit the metal
sample to melt. The furnace then ramps down automatically to a temperature
just above the melting point. The controller holds at that temperature for approximately 20 minutes while everything equilibrates after which the display
flashes on and off alternately indicating that the cell is now ready to initiate the
freeze. Note: The monitor thermometer temperature can be checked while the
sample is melting providing a good check of its accuracy. The control tempera
ture can then be accurately checked during this dwell period.
To initiate the freezing of the metal, press “SET” and “DOWN” again. Press the
up and down keys until the term “FrEEzE” is on the display. Press “SET” to
select that mode and the controller selects a set-point temperature a few degrees
below the freezing point to begin the sample freezing process. This lower tem
perature is to overcome the super-cooling of the metal sample. The cell temper
ature must be monitored with the monitor thermometer at this time in order to
see when the cell starts to freeze. The freeze is established when the cell tem
perature begins a sharp rise toward the freezing point (recalesence). Immedi
ately insert a quartz glass rod that is at room temperature for approximately 2
minutes to create a thin layer of frozen metal. This action positions the change
of phase right next to the thermometer which is necessary for high accuracy.
After removing the tube, press “SET” and “DOWN” again and select “MAint”
to maintain the furnace at a temperature just below the freezing temperature.
With the freeze plateau underway, calibrations can now be made. Use the
pre-heat wells to heat up the thermometers before inserting them into the cell
preserving the latent energy and permitting more calibrations during the freeze.
Allow the thermometers to equilibrate for 20 minutes before making readings.
-
-
-
-
-
-
-
-
56
Page 60
7 Fixed-Point Realization
Test Probe Calibration
Calibration must end before the end of the freezing plateau. To continue use,
re-melt the sample. Press “SET” and “DOWN”. Select the “MELt” function
and the controller will re-melt the sample and the process may be repeated.
This function is convenient if the furnace is to be used again early the next day.
The sample can be melted overnight and be ready to freeze in the morning.
CAUTIONS: Use care in handling the metal fixed-point cells. They can
be easily broken or contaminated. Keep all items clean. Clean with pure
alcohol. Handle with clean cotton gloves or clean paper. Remove any in
advertent finger prints or other contaminants.
-
DO NOT drop the cell or basket into the furnace. Do not drop a thermom
eter into the cell even a short distance.
DO NOT leave quartz thermometers in the preheat wells any longer than
necessary to heat to temperature, especially at the aluminum point. Con
tamination may result.
DO NOT insert cells into a hot furnace. Start from near ambient temperatures.
When all calibrations are complete, manually set the temperature to about
25°C to prevent the furnace from heating unexpectedly the next time it is
used. The furnace can be turned off while hot and permitted to cool on its
own.
7.5Test Probe Calibration
For optimum accuracy and stability, allow the furnace to warm up for 10 min
utes after power-up and then allow adequate stabilization time after reaching
the set-point temperature. After completing operation of the furnace, allow the
well to cool by setting the temperature to 25°C and allowing the unit to cool off
before switching the power off.
7.5.1Calibrating a Single Probe
Insert the probe to be calibrated into the well of the instrument. The probe
should fit snugly into the comparison block yet should not be so tight that it
cannot be easily removed. Avoid any dirt or grit that may cause the probe to
jam into the comparison block. Best results are obtained with the probe inserted
to the full depth of the well. Once the probe is inserted into the well, allow ade
quate stabilization time to allow the test probe temperature to settle as de
scribed above. Once the probe has settled to the temperature of the well, it may
be compared to the furnace display temperature. The display temperature
should be stable to within 0.01°C degree for best results.
-
-
-
-
-
57
Page 61
9260 Mini Fixed Point Cell Furnace
User’s Guide
Never introduce any foreign material into the probe hole of the insert. Fluids
etc. can leak into the furnace causing damage to the instrument or binding and
damage to your probe.
7.5.2Furnace Characteristics
There is a temperature gradient vertically in the test well. Heat is applied to the
block in such a way as to compensate for nominal heat losses out of the top of
the furnace. However, actual heat losses will vary with the design of the ther
mometer probes inserted into the instrument and the temperature. For best re
sults, insert the probe to the full depth of the well.
CAUTION: DO NOT heat the thermometer hub or handle above the
rated temperature.
7.5.2.1Stabilization and Accuracy
The stabilization time of the instrument depends on the conditions and temper
atures involved. Typically the test well will be stable to 0.1°C within 10 minutes of reaching the set-point temperature as indicated by the display. Ultimate
stability will be achieved 15 to 20 minutes after reaching the set temperature.
Inserting a cold probe into the well requires another period of stabilization depending on the magnitude of the disturbance and the required accuracy. For example, inserting a 0.25 inch diameter room temperature probe into a
comparison block at 300°C takes 5 minutes to be within 0.1°C of its settled
point and takes 10 minutes to achieve maximum stability.
Speeding up the calibration process can be accomplished by knowing how soon
to make the measurement. Test measurements should be made at the desired
temperatures with the desired test probes to establish these times.
-
-
-
58
Page 62
Figure 12 9260 Comparison Block Heating Up
7 Fixed-Point Realization
Test Probe Calibration
Figure 13 9260 Comparison Block Cool Down
59
Page 63
8Furnace Calibration
For optimum performance in realizing melting or freezing points of fixed-point
cells, calibration of the 9260 must be maintained. Excessive temperature gradi
ents and inaccurate furnace temperatures can make it difficult to realize melting
or freezing points, reduce the length of the melting or freezing plateau, and
even possibly damage a fixed-point cell. For best results the vertical tempera
ture uniformity should be kept within 0.2°C over the lower 12 cm (5 in.) and
the temperature accuracy should be kept within 0.5°C. The temperature profile
and accuracy should be checked often and adjusted as necessary. While the fur
nace is new consider checking these at least once a month. The following sec
tions explain the temperature profile and temperature accuracy calibration
procedures.
8.1Temperature Profile Adjustment
The 9260 Furnace is adjusted at the factory for a flat vertical temperature profile. The temperature profile should be regularly checked and adjusted as necessary to maintain it within about 0.2°C over the bottom 12 cm (5 in.) with a
slightly higher temperature toward the top. The following steps explain how to
test and adjust the temperature profile. The temperature accuracy should subsequently be calibrated according to Section 8.2 below whenever the zones are
adjusted.
8 Furnace Calibration
Temperature Profile Adjustment
-
-
-
-
8.1.1Step 1: Measure the profile
The temperature profile should be measured at 660°C or at the highest fixed
point with which the furnace is intended to be operated. The furnace should
have the fixed-point cell installed and the cell should be completely frozen or
melted. Incorrect data will result if the cell is only partially melted or frozen.
Use a quartz SPRT and thermometer readout that together provide adequate
resolution and stability to allow relative temperature measurements to 0.02°C
or better. You must wait until the temperature of the SPRT is stable before re
cording readings. It may take 20 to 30 minutes after inserting the SPRT before
it becomes stable. Measure the temperature at three locations: t1 with the SPRT
fully inserted into the well, t2 with the SPRT withdrawn 6 cm (2.5 in.), and t3
with the SPRT withdrawn 12 cm (5 in.). To reduce possible effects of tempera
ture drift repeat the measurements in the following order: t1, t2, t3, t2 again,
then t1 again. Wait until the SPRT is stable (about two minutes) before record
ing each reading. Average the two t1 measurements and the two t2 measure
ments then compare the temperatures. Temperature t2 relative to t1 should be
less than 0.1°C higher or lower. Temperature t3 relative to t1 should be between
0 and 0.2°C higher. If the gradient exceeds these limits adjust the end zones as
explained in the next step.
-
-
-
-
61
Page 64
9260 Mini Fixed Point Cell Furnace
User’s Guide
8.1.2Step 2: Adjust the end zones
The temperature profile can be adjusted by altering the calibration parameters
"toPPCt"and"botPCt" in the controller (see Section 5.16.5). Increasing or
decreasing the top zone heat by 1% typically increases or decreases t3 relative
to t
by about 0.1°C at 660°C and increases or decreases t2relative to t1by
1
about 0.03°C. Likewise, increasing or decreasing the bottom zone heat by 1%
typically increases or decreases temperature t
660°C and increases or decreases temperature t
From these relationships, the following formulas are derived and these can be
used to determine new settings for the top and bottom zones to produce a flat
gradient with the top only 0.1°C hotter than the bottom.
relative to t3by about 0.1°C at
1
relative to t3by about 0.03°C.
2
topPcttopPct
botPctbotPct
%
⎛
⎞
tt
−−
⎜
⎝
⎛
⎜
⎝
()()
⎟
2131
⎠
C
°
%
⎞
tt
−−
()()
⎟
2131
⎠
C
°
%
⎛
⎜
⎝
⎛
⎜
⎝
⎞
tt′=+
.
.
−2517 5
⎟
⎠
C
°
%
⎞
tt′=+
−+257 5
⎟
⎠
C
°
For example, t1is measured in Step 1 to be 657.71, t2is measured to be 657.83,
and t
is measured to be 658.41. The bottom zone is set at 94% and the top
3
zone is set at 91%. The new settings are calculated as follows.
topPct
botPct
′=+
912565783657 7117 5%
′=+
942565783657 717 5%
%
⎛
⎞
⎜
⎝
⎛
⎜
⎝
()
⎟
⎠
C
°
%
⎞
()
⎟
⎠
C
°
CC
°−° −
.. .
CC
°−° −
...
%
⎛
⎜
⎝
⎛
⎜
⎝
⎞
⎟
⎠
C
°
%
⎞
()
⎟
⎠
C
°
°−° +=658 41657 711 7584...%%CC
()
°−° +=658 41657 710 7593...%%CC
8.1.3Repeat Step 1 and Step 2 if necessary
After adjusting the end zones repeat Step 1 to check the temperature profile
again. If the temperature differences still exceed the recommended limits repeat
Step2again.
8.2Temperature Calibration
For best results in realizing freezing or melting points the furnace should be ac
curate to at least 0.5°C and preferably to 0.3°C. The accuracy should be
checked periodically and adjusted if necessary. For small adjustments at one
temperature the simplified one-point calibration procedure explained in Section
8.2.1 may be used. For calibration over the entire range of the furnace the
three-point calibration procedure explained in Section 8.2.2 should be used. For
each procedure a quartz SPRT is needed in a addition to a thermometer readout
that together provide an accuracy of 0.1°C or better.
+175.%
075.%
-
62
Page 65
8.2.1One-point Calibration
This one-point calibration procedure improves the temperature accuracy of the
furnace at one point by adjusting the R0 calibration parameter (see Section
5.16.2). It may be used often to ensure the quality of a specific fixed-point
realization.
The first step is to measure the furnace temperature at the temperature of inter
est. The furnace should have the fixed-point cell installed and the cell must be
entirely frozen or melted. Measure the temperature with the SPRT fully in
serted into the well. Make sure the furnace temperature is stable and make care
ful and accurate measurements.
The next step is to adjust R0 to make the temperature closer to the set-point. In
creasing or decreasing R0 by 0.01Ω increases or decreases the temperature by
about 0.025°C at 0°C and about 0.09°C at 650°C. Use the following formula to
calculate a new value for R0 to make the temperature closer to the set-point.
tt
−
RR
00
′=−
For example, with the set-point at 655°C the furnace temperature is measured
and found to actually be 655.65. R0 is set to 100.124 (ohms). The new value
for R0 is calculated as follows.
meassp
t
C
°
25
+.
100
ΩΩ
sp
8 Furnace Calibration
Temperature Calibration
-
-
-
-
CC
655 65655 00
°−°
R
0 100124
..
Ω−
CC
°
25
.
ΩΩ
655
+
100
8.2.2Three-point Calibration
The first time the furnace is calibrated at the factory it is calibrated over the full
range by adjusting the three calibration parameters R0, ALPHA, and DELTA.
This full calibration may be repeated periodically to ensure that the furnace is
accurate to 0.5°C or better at all temperatures in its range.
Full calibration requires making measurements of the furnace temperature at
three widely separated set-points throughout the range of the furnace. Sug
gested set-points are 50°C, 400°C, and 650°C. The actual temperature of the
furnace is measured at these temperatures. As before, the furnace should have
the fixed-point cell installed and the cell must be entirely frozen or melted.
Measure the temperature with the SPRT fully inserted into the well. Make sure
the furnace temperature is stable (wait about 20 minutes after inserting the
SPRT) and make careful and accurate measurements. Once the temperatures
are obtained, new values for R0, ALPHA, and DELTA can be calculated. The
procedure is summarized as follows.
1.Choose three set-points to use in the calibration of the R0 , ALPHA, and
DELTA parameters. These set-points are generally 50°C, 400°C, and
650°C but other set-points may be used if desired or necessary.
100′=
=.
°
.052Ω
-
63
Page 66
9260 Mini Fixed Point Cell Furnace
User’s Guide
2.Set the furnace to the low set-point. When the furnace reaches the
set-point and the display is stable, wait approximately 15 minutes and
then take a reading from the thermometer. Sample the set-point resis
tance. Write these values down as T
3.Repeat step 2 for the other two set-points recording them as T
and R
respectively.
3
4.Using the recorded data, calculate new values for the R0 , ALPHA, and
DELTA parameters using the equations given below.
8.2.3Compute DELTA
ATT=−
32
BTT=−
21
TTTT
⎡
⎤
⎡
3322
C
=
D
=
−
⎢
⎥
⎢
10011001001100
⎣
⎦
⎣
TTTT
⎡
⎤
⎡
2211
−
⎢
⎥
⎢
10011001001100
⎣
⎦
⎣
-
and R1respectively.
1
2,R2,T3
⎤
⎡
⎤
⎡
−
⎥
⎢
⎥
⎦
⎤
⎥
⎦
⎢
⎣
⎦
⎣
⎡
⎤
⎡
−
⎢
⎥
⎢
⎣
⎦
⎣
⎤
−
⎥
⎦
⎤
−
⎥
⎦
,
ERR=−
32
FRR=−
21
AF BE
=
DE CF
−
−
delta
T
are the measured temperatures using the thermometer. R
1-3
of R from the display of the 9260. (Press SET and DOWN at the same time.)
T
and R1are the measured temperature and resistance at 50°C.
1
T
and R2are the measured temperature and resistance at 400°C
2
T
and R3are the measured temperature and resistance at 650°C
3
8.2.4Computer R0 and Alpha
TT
⎡
⎤
⎡
=+
aTdelta
11
=+
aTdelta
33
Ra Ra
rzero
31 13
=
aa
13
11
⎢
⎥
1001100
⎣
⎦
TT
⎡
⎤
33
⎢
⎥
1001100
⎣
⎦
−
−
⎤
−
⎢
⎥
⎣
⎦
⎡
⎤
−
⎢
⎥
⎣
⎦
are the values
1-3
64
Page 67
RR
−
alpha
13
=
Ra Ra
31 13
−
Delta is the new value of DELTA computed above.
Program the new values for DELTA (delta), R0(rzero), and ALPHA (alpha)
into the furnace by: pressing the “SET” and “EXIT” keys simultaneously and
then pressing “SET” until R0 is displayed. Press “SET” then use the UP or
DOWN keys until the correct numerical setting is displayed. Press “SET” to ac
cept the new value. Continue this process for ALPHA and DELTA.
8.2.5Accuracy and Repeatability
Check the accuracy of the furnace at various points over the calibrated range. If
the furnace does not pass specification at all set-points, repeat the Calibration
Procedure.
8 Furnace Calibration
Temperature Calibration
-
65
Page 68
9Maintenance
The calibration instrument has been designed with the utmost care. Ease
•
of operation and simplicity of maintenance have been a central theme in
the product development. Therefore, with proper care the instrument
should require very little maintenance. Avoid operating the instrument in
an oily, wet, dirty, or dusty environment.
If the outside of the instrument becomes soiled, it may be wiped clean
•
with a damp cloth and mild detergent. Do not use harsh chemicals on the
surface which may damage the paint.
It is important to keep the well of the instrument clean and clear of any
•
foreign matter. Do not use fluid to clean out the well.
The furnace should be handled with care. Avoid knocking or dropping the
•
instrument.
If the comparison block is dropped, examine the comparison block for de
•
formities before inserting it in the well. If there is any chance of jamming
the comparison block in the well, file or grind off the protuberance.
• Do not slam the probe stems into the well. This type of action can cause a
shock to the sensor or break the entrant well of a fixed-point cell.
• If a hazardous material is spilt on or inside the equipment, the user is re-
sponsible for taking the appropriate decontamination steps as out-lined by
the national safety council with respect to the material.
• If the mains supply cord becomes damaged, replace it with a cord with
the appropriate gauge wire for the current of the instrument. If there are
any questions, call Hart Scientific Customer Service for more information.
•
Before using any cleaning or decontamination method except those rec
ommended by Hart, users should check with Hart Scientific Customer
Service to be sure that the proposed method will not damage the equip
ment.
•
If the instrument is used in a manner not in accordance with the equip
ment design, the operation of the furnace may be impaired or safety haz
ards may arise.
9 Maintenance
-
-
-
-
-
67
Page 69
10Troubleshooting
10 Troubleshooting
Troubleshooting
If problems arise while operating the 9260, this section provides some sugges
tions that may help you solve the problem. A wiring diagram is also included.
10.1Troubleshooting
Below are several situations that may arise followed by suggested actions to
take for fixing the problem.
10.1.1Incorrect Temperature Reading
Power the unit on and watch the display. If the first number displayed is less
than “-0005-”, the unit has been re-initialized. The unit needs to be repro
grammed for R0, ALPHA, and DELTA. These numbers can be found on the
Report of Calibration that was shipped with the unit.
10.1.2The unit will not heat or heats at half rate
• Check the fuse. If the fuse is blown the display should be out.
• If the problem continues, contact Hart Scientific Customer Support.
10.1.3The unit heats slowly
• Check the Scan and Scan Rate settings. The Scan may be on with the
Scan Rate set low.
NOTE: When in program mode, the scan rate is automatically set.
-
-
10.1.4If the display flashes any of the following:
“err 1” - This error means there is a RAMerror
“err 2” - This error means there is a NVRAM error
“err 3” - This error means there is a RAMerror
“err 4” - This error means there is an ADC set up error
“err 5” - This error means there is an ADC ready error
•
Initialize the system by performing the master reset sequence. If the unit
repeats the error code, contact Hart Scientific Customer Sup-port for a re
turn authorization and for instructions on returning the unit.
•
Master Reset Sequence - Hold the “SET” and “EXIT” keys down at the
same time while powering up the unit. The screen will display “-init-” ,
“9260" and the version of the software. The unit will need to be repro
grammed for R0, ALPHA, and DELTA in the calibration menu. These
-
-
69
Page 70
9260 Mini Fixed Point Cell Furnace
User’s Guide
numbers can be found on the Report of Calibration that was shipped with
the unit.
10.1.5If the display flashes any of the following:
“err 6" - This error means there is a SENSOR error
The sensor is disconnected or shorted. Please contact Hart Scientific Cus
•
tomer Support for further instructions.
10.1.6If the display flashes any of the following:
“err 7" - This error means there is a HtrCTL error
The fan will go on high speed. Initialize the system by performing the master
reset sequence. If the unit repeats the error code, turn the unit off and contact
Hart Scientific Customer Support for a return authorization and for instructions
on returning the unit.
10.2CE Comments
10.2.1EMC Directive
Hart Scientific's equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). Selection of Light
Industrial or Heavy Industrial compliance has been based on the intended use
of the instrument. Units designed for use in a calibration laboratory have been
tested to Light Industrial Standards. Units designed to be used in the "field"
have been tested to both Light Industrial and Heavy Industrial Standards. The
Declaration of Conformity for your instrument lists the specific standards to
which the unit was tested.
-
70
10.2.2Low Voltage Directive
(Safety) In order to comply with the European Low Voltage Directive
(73/23/EEC), Hart Scientific equipment has been designed to meet the IEC
1010-1 (EN 61010-1) and IEC 1010-2-010 (EN 61010-2-010)standards.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.