Fluke 9230 User guide

Page 1
9230
Hart Scientific
Gallium Cell
Maintenance System
User’s Guide
Rev. 992801
Page 2
Fluke Corporation, Hart Scientific Division
www.hartscientific.com
Subject to change without notice. • Copyright © 2005 • Printed in USA
Rev. 992801
Page 3

Table of Contents

1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Authorized Service Centers. . . . . . . . . . . . . . . . . . . . . . 4
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Specifications and Environmental Conditions . . . . . . . . . . 9
3.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 Quick Start in Auto Program Mode. . . . . . . . . . . . . . . 11
4.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.3 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.4 Auto Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.4.1 Starting the Auto Program . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.4.2 Generating the Melt Plateau . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.4.3 Refreezing the Gallium Cell . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.4.4 Return to Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5 Parts and Controls . . . . . . . . . . . . . . . . . . . . . . . . 15
5.1 Bottom Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5.2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5.3 Top Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.4 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.5 Thermal Block Assembly . . . . . . . . . . . . . . . . . . . . . . 18
6 Controller Operation . . . . . . . . . . . . . . . . . . . . . . . 21
6.1 Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.2 Starting the Auto Program . . . . . . . . . . . . . . . . . . . . . 21
6.3 Auto Program – Initiating and Maintaining the Gallium
Melt Plateau . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
i
Page 4
6.4 Auto Program – Ending the Gallium Melt Plateau and
Refreezing the Cell . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.5 Return to Standby . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.6 Program Advance . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.7 Manual Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.8 Starting the Manual Mode. . . . . . . . . . . . . . . . . . . . . . 26
6.9 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . 27
6.9.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.9.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.10 Temperature Scale Units . . . . . . . . . . . . . . . . . . . . . . 28
6.11 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.11.1 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.12 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.13 Temperature Scale Units . . . . . . . . . . . . . . . . . . . . . . 30
6.14 Secondary Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.15 Peltier Device Power . . . . . . . . . . . . . . . . . . . . . . . . 30
6.16 Set-point Resistance . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.17 Proportional Band . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.18 Controller Configuration, Operating, and Calibration Parameters . 32
6.19 Auto Program Configuration Parameters . . . . . . . . . . . . . . 33
6.20 Prog 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.20.1 Ready Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.20.2 Prep Melt Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.20.3 Prep Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.20.4 Beeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.20.5 Prep 1 Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.20.6 Prep 2 Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.20.7 Prep 3 Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.21 Prog 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.21.1 Maintain Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.21.2 Maintain Time-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.21.3 Maintain Time-out Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.21.4 Freeze Hot Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.21.5 Freeze Hot Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.21.6 Freeze Cold Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.21.7 Freeze Cold Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.21.8 Freeze Cold Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.22 Prog 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.22.1 Freeze Melt State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.23 Serial Interface Parameters . . . . . . . . . . . . . . . . . . . . . 45
6.23.1 BAUD Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.23.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.23.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
ii
Page 5
6.23.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.24 Calibration Parameters . . . . . . . . . . . . . . . . . . . . . . . 47
6.24.1 R0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7 Digital Communication Interface . . . . . . . . . . . . . . . . 49
7.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . 49
7.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1.2.1 Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.1.2.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.1.2.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.1.2.4 Linefeed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.1.3 Serial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.2 Interface Commands . . . . . . . . . . . . . . . . . . . . . . . . 51
8 Melting Point of Gallium Realization . . . . . . . . . . . . . . 55
8.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.2 Preparing the Model 9230 to Receive the Gallium Cell. . . . . . . 56
8.3 Preparing the Frozen Model 5943 Gallium Cell for
Installation in the Model 9230 . . . . . . . . . . . . . . . . . . . 57
8.4 Installing the Frozen Model 5943 Gallium Cell into the
Model 9230 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.5 Inserting the Top Insulation Plugs . . . . . . . . . . . . . . . . . . 58
8.6 Inserting the Inner Melt Heater . . . . . . . . . . . . . . . . . . . 58
8.7 Auto Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.7.1 Starting the Auto Program . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
8.7.2 Generating the Melt Plateau and Removing the Inner Melt Heater . . . . . . 59
8.7.3 Calibrating Thermometers . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
8.7.4 Advancing the Auto Program and Refreezing the Gallium Cell . . . . . . . . 60
8.7.5 Return to Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
9 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . 61
9.1 Calibration Equipment . . . . . . . . . . . . . . . . . . . . . . . 61
9.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 61
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.1 Troubleshooting Problems, Possible Causes, and Solutions . . . . 67
11.2 Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
11.2.1 EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
11.2.2 Low Voltage Directive (Safety) . . . . . . . . . . . . . . . . . . . . . . . . . 68
iii
Page 6

Figures and Tables

Table1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1
Figure 1 Auto Program Diagram . . . . . . . . . . . . . . . . . . . . . . . . . 13
Figure 2 Bottom Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Figure 3 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Figure 4 Top Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Figure 5 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Figure 6 Controller Flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Table 2 Default Programmable Set-points (Factory Set) . . . . . . . . . . . . . 28
Table 3 Auto Program Parameters (Prog 1, Prog 2, Prog 3) . . . . . . . . . . . 34
Figure 7 Serial Cable Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Table 4 Serial Interface Commands . . . . . . . . . . . . . . . . . . . . . . . 52
iv
Page 7
1 Before You Start
1.1 Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
Tabl e1 International Electrical Symbols
Symbol Description
AC
AC-DC
Battery
CE
1 Before You Start
Symbols Used
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface
Read the User’s Manual
Off
On
1
Page 8
9230 Maintenance System
User’s Guide
Canadian Standards Association
C-TIC Australian EMC Mark
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re fers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances.
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) mark.
1.2 Safety Information
Use of this instrument other than as specified in this manual may impair the protection provided by the instrument.
The following definitions apply to the terms “Warning” and “Caution”.
“WARNING” identifies conditions and actions that may pose hazards to the user.
“CAUTION” identifies conditions and actions that may damage the in­strument being used.
1.2.1
WARNINGS
To avoid possible electric shock or personal injury, follow these guidelines.
BURN HAZARD DO NOT leave the melt heater outside of the unit and en ergized. The melt heater can reach approximately 110°C if not properly in serted into the unit. Place the melt heater in the storage well located at the back of the instrument when not in use.
-
-
-
DO NOT operate this unit without a properly grounded, properly polarized power cord.
Use only grounded AC mains supply of the appropriate voltage to power the in
-
strument. Refer to Section 3.1, Specifications, for power details.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE IN
-
JURY OR DEATH may result if personnel fail to observe safety precautions.
Before working inside the equipment, turn the power off and disconnect the power cord.
Always replace the fuse with one of the same rating, voltage, and type.
2
Page 9
1 Before You Start
Safety Information
Overhead clearance is required. DO NOT place this instrument under a cabinet or other structure. Additionally, at least 6 inches of space between the instru
-
ment and nearby objects is necessary for sufficient air circulation.
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the user’s
guide.
DO NOT turn the unit upside down with the cell in place; the cell will fall out of the unit.
DO NOT operate near flammable materials.
DO use a ground fault interrupt device.
Before initial use, after transport, and anytime the dry-well has not been ener gized for more than 10 days, the calibrator must be energized for a dry-out pe
-
­riod of 1 to 2 hours before it can be assumed to meet all of the safety requirements of the IEC1010-1.
Follow all safety guidelines listed in the user’s manual.
Calibration Equipment should only be used by Trained Personnel.
When the melt heater switch is in the MANUAL ON position, the melt heater is continuously energized and can get hot!
1.2.2
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
DO NOT use fluids to clean out the well.
DO NOT slam the probe stems into the entrant well of the gallium cell. The
probe could be damaged.
DO use the cell removal tool to remove the cell from the well of the instrument. Grasp the cell with a hand as soon as the cell is removed far enough from the well.
DO NOT use the removal tool to carry the cell. The removal tool is not de signed for transporting the cell.
Never introduce foreign material into the well. Fluids, etc. can leak into the cal ibrator causing damage.
DO use silicon oil or other appropriate heat transfer medium in the entrant well of the gallium cell. Be sure the transfer medium does not leak out into the well of the instrument.
DO NOT change the value of the calibration constant from the factory set value. The correct setting of this parameter is important to the safety and proper operation of the calibrator.
-
-
3
Page 10
9230 Maintenance System
User’s Guide
The Auto Program configuration parameters should not be adjusted under nor mal use. These parameters directly affect the melt plateau duration time and proper freezing of the gallium cell.
Insert only frozen gallium cells into the 9230. DO NOT insert a melted cell into the unit and then initiate the Auto Program. An improper freeze may occur resulting in damage to the cell.
1.3 Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
-
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
4
Page 11
1 Before You Start
Authorized Service Centers
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following
information available:
Model Number
Serial Number
Voltage
Complete description of the problem
5
Page 12
2 Introduction
The 9230 is a specialized instrument for the realization of the gallium melting point (29.7646°C). The instrument works with Hart’s 5943 stainless steel Ga cell to provide melting plateaus that can last a week (or longer). This instru ment is intended for Calibration Laboratory use and not for field applications. Internal programming of the micro-processor controller provides prepro grammed scan rates, set-points for each step in the process, dwell timing, and indication that the next step is ready. The Model 9230 is available in 100-230 VAC (±10%), 50-60 Hz.
Built in programmable features include:
Temperature scan rate control
Fixed-point programming with default settings for the gallium cell
Eight set-point memories
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s special proportional-inte­gral-derivative (PID)/digital controller. The controller uses a platinum RTD as a sensor and controls the well temperature with thermoelectric Peltier devices.
After properly inserting a frozen Ga cell and powering the unit on, the Model 9230 controller sets the block temperature to 25°C and the LED front panel dis­play shows auto. The user then makes sure that the inner melt heater is at­tached to the back of the unit, the melt heater switch is in the Auto position, and inserts the melt heater into the entrant well of the gallium cell. When the user selects auto by pressing the SET button, the Model 9230 Auto Program begins a pre-determined program to initiate and maintain a gallium melt pla­teau. While in the Auto Program mode, the display shows the current process in the gallium melt cycle. During this mode the user may not change those pa rameters, which could negatively impact the gallium melt. At the end of the plateau the unit properly refreezes the cell and returns the temperature to 25°C (standby) to await the next user initiated melt.
While in the manual mode of operation, the LED front display panel continu ously shows the current well temperature. The temperature may be easily set with the control buttons to any desired temperature within the specified range. Certain parameters of the Auto Program itself may be adjusted in the manual mode of operation.
The Model 9230 was designed for high accuracy fixed-point calibration meth ods and for ease of operation. Through proper use, the instrument will continu ously provide accurate calibration of temperature sensors and devices. The user should be familiar with the safety guidelines and operating procedures of the instrument as described in this User’s Guide.
2 Introduction
-
-
-
-
-
-
7
Page 13
3 Specifications and Environmental Conditions
3 Specifications and Environmental
Conditions
3.1 Specifications
Operating Range 15°C to 35°C (59°F to 95°F)
Gallium Cell Uncertainty ±0.08mK, typical
Gallium Melting Plateau Duration
Display Accuracy ±0.05°C at 29.76°C
Stability ±0.02°C
Display Resolution 0.01°C or °F (0.001°C or °F in program mode)
Display Scale °C or °F, switchable
Preheat Wells 2 wells (8.2 mm diameter)
Immersion Depth 152 mm (6 inches) in gallium cell
Vertical Gradient < 0.03°C over 6 inches during cell maintenance
Fault Protection Firmware protection
Communications RS-232 included
Ambient Temperature 18°C to 28°C (65°F to 82°F)
Power Requirements 115 VAC (±10%), 60 Hz, 1.0 A or 230 VAC (±10%), 50 Hz , 0.65 A,
Exterior Dimensions 8.75” W x 10.25” D x 19.25” H (22 x 26 x 49 cm)
Weight 18 lb (8.2 kg) without cell
5days,typical
175 W maximum
Specifications
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and clean ing recommendations can be found in the Maintenance section of this manual.
The instrument operates safely under the following conditions:
temperature range: 18–28°C (65–82°F)
ambient relative humidity: 15-50%
pressure: 75kPa – 106kPa
mains voltage within ±10% of nominal
vibrations in the calibration environment should be minimized
altitudes less than 2,000 meters
-
-
9
Page 14
9230 Maintenance System
User’s Guide
3.3 Warranty
Hart Scientific, Inc. (Hart) warrants this product to be free from defects in ma terial and workmanship under normal use and service for a period as stated in the current product catalog from the date of shipment. This warranty extends only to the original purchaser and shall not apply to any product which, in Hart’s sole opinion, has been subject to misuse, alteration, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc
-
tions on appropriate Hart products. It is not warranted to be error free.
Hart’s obligation under this warranty is limited to repair or replacement of a product which is returned to Hart within the warranty period and is determined, upon examination by Hart, to be defective. If Hart determines that the defect or malfunction has been caused by misuse, alteration, abuse or abnormal condi tions or operation or handling, Hart will repair the product and bill the pur
-
-
chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling or writing a Hart Scientific Authorized Service Center for authorization. Hart assumes NO risk for in-transit damage.
For service or assistance, please contact a Hart Scientific Authorized Service Center (see Section 1.3).
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLU­SIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EX­PRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAM
-
AGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
-
10
Page 15
4 Quick Start in Auto Program Mode
4 Quick Start in Auto Program Mode
Unpacking
The user should read this user’s guide and be familiar with the functional per formance of the Model 9230 before using the instrument to achieve a gallium melt plateau.
4.1 Unpacking
Unpack the instrument carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier imme diately. Verify that the following components are present:
Model 9230 Gallium Maintenance Apparatus
Melt heater
Power cord
User’s Guide
Top Insulation Plugs
Cell Removal Tool
If you are missing any item, please call a Hart Scientific Authorized Service Center.
4.2 Set Up
Place the instrument on a flat surface with at least 6 inches of free space around and 18 inches above the instrument. Avoid drafty areas. Install the power cord into the power entry module on the back of the instrument. Plug the power cord into a grounded mains outlet. Verify that the nominal voltage corresponds to that indicated on the back of the instrument.
Using the cell removal tool, carefully insert the frozen gallium cell, prepared as described in Section 9, Melting Point of Gallium Realization, into the well. (DO NOT drop it into the well.) The well must be clear of any foreign objects, dirt and grit before the cell is inserted.
Firmly place the two top insulation plugs into the well such that they rest on top of the cell. Make sure that the inner melt heater is connected to the inner melt heater connector on the back of the instrument and the melt heater switch is in the auto position. Carefully insert the melt heater through the hole in the top insulation plugs and into the entrant well of the frozen gallium cell. Make sure the inner melt heater extends to the bottom of the gallium cell entrant well.
-
-
4.3 Power
Plug the instrument power cord into a mains outlet of the proper voltage, fre quency and current capability. Refer to Section 3.1, Specifications, for power
-
11
Page 16
9230 Maintenance System
User’s Guide
details. Turn the instrument ON using the power switch on the back of the unit. The instrument turns on and begins to approach 25°C. The front panel LED display shows auto.
4.4 Auto Program
The Model 9230 is designed to automate the gallium cell melt plateau. This is accomplished through the use of the Auto Program; a microprocessor-con trolled program that determines set temperatures, scan rates, duration times, Peltier module configurations, and melt heater controls. By controlling all of these parameters in the proper order, a temperature profile can be maintained in the block assembly that assures a uniform melt, maintains the melt over long periods of time, properly refreezes the cell and keeps the cell ready for the next melt cycle. The Auto Program is illustrated in Figure 1.
4.4.1 Starting the Auto Program
After the frozen gallium cell is properly installed in the instrument (See Section 9, Melting Point of Gallium Realization, for more details) and the unit is pow­ered ON, the front panel LED display shows Auto. The instrument goes to 25°C and remains there until the SET button on the front panel is pressed which starts the Auto Program. The display changes to uuAit and the Auto Program begins. The unit resets to 29.27°C at a scan rate of 0.2°C/minute in ac­cordance with the Auto Program (see Figure 1).
-
12
4.4.2 Generating the Melt Plateau
Once the unit has stabilized at 29.27°C, it waits for a minimum of 30 minutes to ensure that the gallium cell has equalized in temperature. At the end of the equalization time, the Auto Program begins to initiate the melt. The display changes to PrEP as the set temperature of the block is changed to 30.77°C. This initiates a melt on the outer surfaces of the gallium within the cell. A few minutes into the program, the unit beeps four times and the display changes to a message that flashes between prep and htr on to indicate that the inner melt heater has been turned ON. The inner melt heater remains ON for a few minutes and then shuts off as the display changes back to prep and the unit beeps eight times. The user should remove the inner melt heater at this time. Note: No damage occurs to the gallium cell if the heater is not removed, how ever, a shortening of the melt duration may result.
After the inner melt heater is powered OFF, the instrument remains at 30.77°C for a few more minutes to complete the outer melt of the gallium cell. The unit then beeps 16 times, the display changes from prep to nnaint and the block temperature resets to 29.86°C to maintain the melting gallium cell.
The user may then insert a preheated monitor thermometer to monitor the tem perature of the gallium cell. Due to the heat applied to the cell by the inner melt heater, the temperature of the cell’s entrant well is higher than the gallium melt ing point (29.7646°C) for a short period of time. Generally, after 30 to 60 min utes, the temperature of the cell’s entrant well is sufficiently stable to begin
-
-
-
-
Page 17
4 Quick Start in Auto Program Mode
35
4
0
3
0.77 2
9.86
2
9.27
20
10
0
4 beeps
melthtr=0
(ON)
8min
PREP1
6min
PREP3
User T
est
T
ime Duration
(galliu
mc
e
ll at 29.7646
°C )
up to
6hrs
2.5 hrs
comple
tel
y
freeze
gallium
cell
SR = 0.2
°
C/m
in
T
set = 29.
86°C
25°C
16 beeps
25.00
T (°C) - block sensor
User in
ser
ts
frozen
gal
li
um
cell and p
owers
on u
nit
T
set=25°C
S
R=OFF
frz
mlt=0
(MELT/M
A
INT
)
melthtr=
1
(OFF
)
Unit st
ays
a
t 25°C until U
ser
sele
cts AU
TO or MAN
using
UP/D
Nan
d
presses SE
T
Un
it goes
into manu
al mode
if
MAN is se
lected
User s
elec
t
s
A
UTO
T
set=29.2
7°C
SR
=0.2°C
/min
fr
zmlt=0
(MELT/M
A
IN
T)
melthtr
=1
(OF
F)
Unit re
ma
in
sat
29.27°
Cfo
r
30 min
to allow
ga
lliu
m
cell to
stab
ilize
30 min
minimum
Unit
begins
m
elt
Tset=30.7
7°C
SR=0.2°C
/min
frzmlt=0
(ME
LT/MA
INT)
me
lthtr=1
(OFF)
4min
PREP
2
DISPL
AY
AU
TO
DISPL
AY
WA
IT
Display
PREP
Display
PRE
P/
HTR ON
Di
splay
PRE
P
8 beep
s
melthtr=1
(OFF)
remove
m
elt heate
r
DISPL
AY
MA
INT
Un
it
be
gins
freeze
H
Tse
t=35.0
0°C
SR
=OFF
frzm
lt=0
(MELT/MA
INT)
melthtr=1
(OFF)
Un
it
be
gins
freeze
C
Ts
et=0.00
°C
SR=
0.5°C
/min
frzm
lt=1
(FREEZ
E
)
melthtr=1
(OFF
)
Unit
returns
to
25°C
Tset=
25°C
SR=O
FF
frzmlt=0
(M
ELT/MA
INT)
m
elthtr=1
(OFF)
DISPL
AY
FREEZEH
(OPTIONA
L)
DISPL
AY
FREE
ZC
Unit rem
a
ins
at 25°C
unti
l User
selects
AUTO
or M
AN
(be
gin aga
in)
begin
again
DISPL
AY
AUTO
If in
MAN
mode, User
can re
turn
to AU
TO m
ode by adva
ncin
g
t
he progra
m(SET+
DN)
25°C
29
.27°C
30.77°C
29.86°C
Time - as indicated above
User c
an
ab
ort the
Auto
Prog
ram and
return t
oS
ta
ndby
Mode
(25°
C
)during
the "WAI
T
"tim
eby
adva
ncing
the
program (
SET+
DN)
Defaul
t-U
s
er initiates F
reez
e
by
advancing
prog
ram (SET+D
N)
Option - U
ser pr
e-sets MAIN
T
Duration
T
ime
after which
program in
itiates
Freeze
Default -
G
od
irectly
to FreezC
(no Fr
eezH)
O
ption - U
ser pre-
sets
FreezH Te
mper
ature
a
nd Durat
ion Time
to melt ga
llium ce
ll
be
fore Fre
eze
up to
36°C
Auto Program
Figure 1 Auto Program Diagram
13
Page 18
9230 Maintenance System
User’s Guide
calibrations of thermometer sensors. The duration of the melt plateau varies de pending on usage, but lengths of 5 days or longer are achievable.
4.4.3 Refreezing the Gallium Cell
At the end of the melt plateau (or when the user wishes to end the plateau), the Auto Program is advanced by pressing the SET and DOWN buttons (press the SET button and keep it pressed and then press the DOWN button and release both) on the front panel (program advance). The dislay will flash “AdU”and display “nnAint.” Use the UP and DOWN buttons to select the program state “FrEE2C” then press the ENTER button. See Section 6.21.2, Maintain Time-out, for optional user defined auto time-out. Gallium, like water, expands in volume when freezing. Therefore, the cell must be frozen from the bottom up or there is a potential for damage to the cell. The Auto Program process en sures that the cell is properly refrozen.
The Auto Program initiates the freeze and the display shows free2E at this point. The Peltier devices are reconfigured and controlled such that a large tem perature gradient is generated between the top and bottom of the block. The in­strument slowly scans to 0°C to maintain the gradient over several hours. See Section 6.21.4, Freeze Hot Temperature, for optional user defined freeze hot temperature.
4.4.4 Return to Standby
Upon completion of the freeze, the Auto Program returns the unit to 25°C, the front panel LED display shows auto and the unit remains there until the user initiates the Auto Program again by pressing the SET button.
-
-
-
14
Page 19
5 Parts and Controls
115V
Front
Foot
Fan
1
2
The user should become familiar with the Model 9230 parts. Successful use of the instrument is dependent upon knowledge of important components and theirproperuse.
5.1 Bottom Panel
On the bottom panel are the fan and feet. See Figure 2.
1. The cooling fan inlet is at the bottom of the unit. The cooling air circulating through the instrument keeps the electronics and the chassis cool. Keep the area immediately around the instrument clear to allow adequate ventilation.
2. Three feet support the chassis permitting air space for the fan.
5 Parts and Controls
Bottom Panel
Figure 2 Bottom Panel
15
Page 20
9230 Maintenance System
SET
ADV
°C/°F
DOWN
UP EXIT
0.020 C
1
2
9230
GALLIUM MAINTENANCE APPARATUS 15°C to 30°C
User’s Guide
5.2 Front Panel
The front panel contains the digital display and the controller keypad. See Fig
-
ure 3.
1. The digital display is an important part of the temperature controller because it not only displays set and actual temperatures but also displays various instru ment functions, settings, and constants. The display shows temperatures in units according to the selected scale °C or °F. When in the Auto Program mode, the display indicates the current status of the unit in the gallium melt process.
2. The four button controller keypad allows easy setting of the set-point tem
­perature. The control buttons (SET, DOWN, UP, and EXIT) are used to set the temperature set-point, access and set other operating parameters, and access and set calibration parameters.
Setting the control temperature is done directly in degrees of the current scale and can be set to 0.01 of a degree Celsius or Fahrenheit.
During the Manual mode of operation, the functions of the buttons are as follows:
SET – Used to display the next parameter in the menu and to store parameters to the displayed value.
DOWN – Used to decrement the displayed value of parameters.
UP –Usedtoincrementthedisplayedvalue.
EXIT – Used to exit a function and to skip to the next function. Any changes
made to the displayed value are ignored. Holding the EXIT button for about ½ second returns control to the main display.
-
16
Figure 3 Front Panel
Page 21
5.3 Top Panel
Inner Melt Heater Storage Well
Cooling Air Vents
Preheat Wells
Constant Temperature Block Assembly with Gallium Cell
2
4
5
3
Top Insulation Plugs
1
SET
ADV
°C/°F
DOWN
UP EXIT
9230
GALLIUM MAINTENANCE APPARATUS 15°C to 30°C
Auto
5 Parts and Controls
Top Panel
The primary feature of the top of the unit is to provide access to the tempera
­ture controlled block. The top panel consists of the constant temperature block assembly; two preheat wells, top insulation plug, cooling air vents, the melt heater and melt heater storage well. See Figure 4.
1. The constant temperature block assembly is where the gallium cell is in
­serted. Peltier devices surround the cell in order to provide a uniform tempera ture. These devices can be configured by the controller to provide a large vertical gradient during the freezing of the gallium cell, or a very uniform tem perature when maintaining a melting cell.
-
-
Figure 4 Top Panel
17
Page 22
9230 Maintenance System
User’s Guide
2. The inner melt heater is a low-power heater that creates an inner liquid layer next to the reentrant tube of the gallium cell during the melting process. This inner melt heater is controlled automatically by the microprocessor when using the Auto Program mode, or may be used manually with the switch in back. When not in use, the inner melt heater is stored in the well at the back of the unit to prevent it from causing any damage.
3.The thermometer preheat wells are located at the back of the block access well. Thermometers are preheated in these wells prior to insertion into the cell in order to conserve its latent energy.
4. The cooling air vents in the top of the unit permit heated air to exit the unit.
5. The top insulation plugs fit into the well over the gallium cell. See Section
8.5, Inserting the Top Insulation Plugs.
5.4 Rear Panel
The rear panel consists of the power connection, power switch, melt heater switch, inner melt heater connection, and serial port. See Figure 5.
1. The power connection is where the power cord plugs into the back of the instrument.
2. Set the power switch “ON” to operate the instrument and “OFF” to discon­nect the power.
3. The melt heater plugs into the rear of the instrument into the inner melt heater connector provided. Be sure it is plugged in during operation.
4. The melt heater switch can be set to MANUAL ON position or AUTO posi­tion. The instrument’s microprocessor has control when the switch is in the AUTO position.
18
Warning: When the melt heater switch is in the MANUAL ON position,
the melt heater is continuously energized and can get hot!
5. The serial port is a DB-9 male connector for interfacing the instrument to a computer or terminal RS-232 communications.
5.5 Thermal Block Assembly
The thermal block assembly holds the gallium cell and heats or cools to main tain the proper temperature. The thermal block assembly consists of the thermal block, thermoelectric Peltier devices, the gallium cell, and the top insulation plugs.
1. The thermal block is specifically designed to contain the gallium cell. The thermal block also has 2 preheat wells for the test thermometers.
-
Page 23
5 Parts and Controls
RS-232
115V
RS-232
Power Switch
Powe r Connection
RS-232 Interface
MANUAL ON
AUTO
INNER MELT
HEATER
Inner Melt Heater Connector
Melt Heater Switch
Thermal Block Assembly
Figure 5 Rear Panel
19
Page 24
9230 Maintenance System
User’s Guide
2. The block assembly is heated or cooled as necessary to maintain the proper temperature with thermoelectric Peltier devices.
3. The gallium cell is placed into the block. Use the removal tool to carefully insert or remove the cell.
WARNING: DO NOT use the removal tool to carry the cell.
4. The top insulation plugs are placed over the gallium cell to help maintain the proper temperature. See Section 8.5, Inserting the Top Insulation Plugs.
20
Page 25
6 Controller Operation
6 Controller Operation
Standby
This section discusses in detail how to operate the instrument temperature con troller using the front control panel. By using the front panel key-switches and LED display, the user may initiate the Auto Program mode; a microproces sor-controlled program that determines set temperatures, scan rates, duration times, Peltier module configurations, and melt heater controls. By controlling all of these parameters in the proper order, a temperature profile can be main tained in the block assembly that assures a uniform melt, maintains the melt over long periods of time, properly refreezes the cell and keeps the cell ready for the next melt cycle. After selecting the Manual mode of operation, the user may monitor the well temperature, adjust the set-point temperature in degrees C or F, monitor the Peltier device output power, adjust the controller propor tional band, and program the operating parameters, program parameters, serial interface configuration, and the controller calibration parameters. Operation of the functions and parameters is shown in the flowchart in Figure 6 on page 22. This chart may be copied for reference.
In the following discussion a button with the word SET, UP, DOWN, or EXIT inside indicates the panel button while the dotted box indicates the display reading. Explanation of the button or display reading are to the right of each button or display value.
When two buttons are shown to initiate an action, for example,
S+D
6.1 Standby
After the instrument is powered ON or at the end of a gallium cell refreeze pro cess, the unit automatically is set to 25°C and the digital LED display reads:
-
-
-
-
Press and hold the first button (e.g., SET) and then press the second button (e.g., DOWN) and release both
-
Auto Ready to Start Auto Program
The unit is in a Standby condition and remains at 25°C indefinitely until the user initiates a change by pressing a button sequence.
6.2 Starting the Auto Program
The user begins the Auto Program by pressing the SET button.
S
Start the Auto Program
uuAlt WAIT - Auto Program started
21
Page 26
9230 Maintenance System
Powe r ON
Display
“Auto”
T =25°C
SET
Select “nnAn”
Select Set-Point
UP/DN
SET
SET
Adjust Set-Point
Adjust Units °C/°F
Adjust Scan ON/OFF
Display
Temperature
Adjust Scan Rate
UP/DN
UP/DN
UP/DN
UP/DN
Toggles °C/°F
Display Set-Point
Resistance
SET
Adjust Proportional
Band
Configuration
Menu
Display Power
Display
“uuAIt”
SET
Display “PREP”
Display “nnaint”
Display “FrEE2H”
Program Advance
SET + DN
PRE-SET TIMEOUT
Display “frEE2C”
Optional
SET+EXIT
SET + UP
SET
SET
EXIT
Adjust
Ready Temperature
SET
Program 1
Parameters
Menu
UP
DN
Adjust Prep Melt
Temperature
Adjust Prep
Scan Rate
Adjust
Beeper
Adjust
Prep 1Time
Adjust
Prep 2 Time
Adjust
Prep 3 Time
Adjust Maintain
Temperature
SET
Program 2
Parameters
Menu
UP
DN
Adjust Maintain
Time-Out
Adjust Maintain Time-Out Time
Adjust Freeze
Hot Temperature
Adjust Freeze
Hot Time
Adjust Freeze
Cold Temperature
Adjust Freeze
Cold Scan Rate
Adjust Freeze
Melt State
SET
Program 3
Parameters
Menu
UP
DN
Adjust
Baud Rate
SET
Serial
Interface
Menu
UP
DN
Adjust
Sample Period
Adjust
Duplex Mode
Adjust
Linefeed
Adjust
R0
SET
Cal
Menu
Adjust Freeze
Cold Time
X5
EXIT EXIT EXIT EXIT EXIT
Menu Legend: While in MANUAL Mode:
Press the SET button to step through the menu and to access the parameter. Use the UP and DOWN buttons to adjust the parameter value. Press the SET button again to accept and store the new value, and to advance to the next parameter.
Press the EXIT button briefly to skip a parameter without storing the parameter value.
Hold the EXIT button to exit the menu and to display the temperature.
Press the SET and DOWN buttons at the same time to advance the program to AUTO mode.
While in AUTO Mode:
Press the EXIT and UP buttons to momentarily display the block temperature.
The AUTO Mode parameters may be viewed, but no changes are allowed except for the beeper and serial interface parameters.
Manual Mode
SET + DOWN
SET + DOWN
Auto Mode
UP
SET
User’s Guide
(see Figure
(see Figure
Figure 6 Controller Flowchart
22
Page 27
Plateau
Auto Program – Initiating and Maintaining the Gallium Melt
6 Controller Operation
At this point the Auto Program takes control of the instrument and initiates a melt in the gallium cell in accordance with Figure 1 on page 13 and Figure 6 on page 22.
At any time the user may press the program advance button sequence to access the program advance mode.
S+D
Access program advance
toAuto Exit from auto mode
From here, the set button can be pressed to go back to the Auto state or press the up or down buttons to choose a different program state. Once the desired program state is chosen press SET to accept the chosen state.
S
The chosen program state will now start.
Accept the program state
6.3 Auto Program – Initiating and Maintaining the Gallium Melt Plateau
Approximately 55-65 minutes after starting the Auto Program, the gallium cell reaches 29.27°C and stabilizes enough to start the melt plateau. The display shows:
prep PrEP – gallium cell melt initiated
The Auto Program takes the block assembly to 30.77°C, which begins melting the outer surface of the gallium inside the cell.
After approximately 8 minutes, the Auto Program energizes the inner melt heater inside the gallium cell entrant well. The instrument sounds an audible beep 4 times and the display message alternates as indicated:
PrEP Alternating message between PrEP and Htr on
Htr on
The inner melt heater is energized for approximately four minutes after which the Auto Program turns it OFF and sounds an audible beep eight times. At this point the display indicates:
PrEP Inner melt heater OFF
The user should remove the inner melt heater at this time. Note: No damage will occur to the gallium cell if the heater is not removed, however, a shorten
-
23
Page 28
9230 Maintenance System
User’s Guide
ing of the melt plateau may result. The inner melt heater may be stored in the melt heater well at the back of the unit.
After the inner melt heater has been powered OFF, the instrument remains at
30.77°C for a few more minutes to complete the outer melt of the gallium cell. The unit then sounds an audible beep 16 times, the display changes from PrEP to nnAint and the block temperature is reset to 29.86°C to maintain the melt ing gallium cell.
nnAint Maintain – instrument maintaining melt plateau
-
The user may then insert a preheated monitor thermometer to monitor the tem perature of the gallium cell. Due to the heat applied to the cell by the inner melt heater, the temperature of the cell’s entrant well is higher than the gallium melt ing point (29.7646°C) for a short period of time. Generally, after 30 to 60 min utes, the temperature of the cell’s entrant well is sufficiently stable to begin calibrations of thermometer sensors. The duration of the melt plateau varies de pending on usage, but lengths of 5 days or longer are achievable.
6.4 Auto Program – Ending the Gallium Melt Plateau and Refreezing the Cell
At the end of the melt plateau (or when the user wishes to end the plateau), the Auto Program is advanced by pressing the program advance button sequence.
S+D
Display will flash advance for 1 second then show the current program state.
AdU Program advance
nnAint Maintain - instrument maintaining melt plateau
U S
Access program advance
Advances to next program state
Accepts new program state
-
-
-
-
24
FrEE2C Start freezing the gallium cell
Gallium, like water, expands in volume when freezing. Therefore, the cell must be frozen from the bottom up or there is a potential for damage to the cell.
The Peltier devices are reconfigured and controlled such that a large tempera ture gradient is generated between the top and bottom of the block. The instru ment slowly scans from 29.86°C to 0°C to maintain the gradient over several hours thus ensuring the cell freezes properly, beginning at the bottom and con tinuing up to the top of the cell.
-
-
-
Page 29
See Section 6.21.2, Maintain Time-out, for optional user defined auto time-out (an optional method of ending the melt plateau and initiating the refreeze pro cess). See Section 6.21.4, Freeze Hot Temperature, for optional user defined freeze hot temperature (a method to totally melt the cell before freezing it).
6.5 Return to Standby
Upon completion of the freeze, the Auto Program returns the unit to 25°C, the digital LED display shows:
Auto Instrument returning to 25°C
6 Controller Operation
Return to Standby
-
After reaching 25°C, the unit remains there until the user initiates the Auto Pro gram again by pressing the SET button or goes into the Manual mode of operation.
6.6 Program Advance
From the manual mode, the program advance function allows the user to return to the standby condition.
S+D
AdU ” flashes and then displays program mode
Auto Program mode
Press the UP or DOWN button to toggle between the Auto Program and Man ual mode.
S
At any time during the auto program the user may press the program advance button sequence to access the program advance function.
S+D
The current program state will be displayed.
Access program advance
Accepts the new operation
Access program advance
-
-
toAuto Return to main menu
PrEP Initiate cell melt with melt heater
nnAint Maintain melt plateau
25
Page 30
9230 Maintenance System
User’s Guide
FrEE2H Initiate total melt
or
FrEE2C Start freezing the cell
From the current program state, the up or down buttons can be pressed to choose a different program state.
UorD
Once the desired program state is chosen, press “SET” to accept and start that program.
S
To get to the wait state the user must select the Auto state then press “set” to start the program in the wait state.
Choose the desired program state
Accept new program state and start the program
6.7 Manual Mode
6.8 Starting the Manual Mode
After the instrument is powered ON or at the end of a gallium cell refreeze pro­cess, the unit automatically is set to 25°C and the digital LED display reads:
Auto Ready to start Auto Program
The unit is in a Standby condition and remains at 25°C indefinitely until the user initiates a change by pressing a button sequence. To begin the Manual mode of operation, the user presses the UP or DOWN buttons to change the display reading as follows:
DorU
Toggle display between Auto and Manual
26
nnAn Ready to go into Manual mode
By pressing the SET button while this is displayed, the instrument goes into the Manual mode and begins to display the well temperature:
S
25.00C Well temperature in degrees Celsius
Accept Manual mode
Page 31
Once in the Manual mode, the temperature display function may be accessed from any other function by pressing and holding the EXIT button a few seconds.
6.9 Temperature Set-point
6 Controller Operation
Temperature Set-point
The temperature set-point can be set to any value within the range and resolu tion as given in the specifications. Be careful not to exceed the safe temperature limit of any device inserted into the well.
Setting the temperature involves selecting one of the eight (8) set-points in memory and then adjusting the set-point value.
6.9.1 Programmable Set-points
The controller stores eight (8) set-point temperatures in memory. The set-points can be quickly recalled to conveniently set the instrument to a previously pro grammed temperature set-point.
To set the temperature, first select the set-point memory. This function is ac­cessed from the temperature display function by pressing the SET button. The number of the set-point memory currently being used is shown at the left on the display followed by the current set-point value.
25.00C Well temperature in degrees Celsius
S
Access set-point memory
1 25. Set-point memory 1, 25.0°C currently used
To change the set-point memory to another preset value press the UP or DOWN button.
-
-
4 30.C New set-point memory 4, 30.0°C
Press the SET button to display the new selection and access the set-point value.
S
Display selected set-point memory
29.86 New set-point value 29.86°C
6.9.2 Set-point Value
If the set-point value is correct, hold the EXIT button to accept the new value and resume displaying the well temperature. Otherwise, press the UP or DOWN button to adjust the set-point value.
27
Page 32
9230 Maintenance System
User’s Guide
22.00 New set-point value
When the desired set-point value is reached press the SET button to accept the new value and to access the temperature scale units. If the EXIT button is pressed, any changes made to the set-point are ignored.
S
Accept new set-point value
The default programmable set-points (factory set) are given in Table 2.
Tabl e 2 Default Programmable Set-points (Factory Set)
Set-point Memory
Number
1 25.00
2 29.27
3 30.77
4 29.86
5 35.00
6 00.00
7 10.00
8 20.00
NOTE: Setpoint 1 will always be set back to 25.00°C af-
ter the power has been cycled off then on.
Default Set-point Temperature (°C)
6.10 Temperature Scale Units
The temperature scale units of the controller can be set by the user to degrees Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well tem perature, set-point, scan rate, and proportional band.
Press the SET button after adjusting the set-point value to change display units.
-
28
Un= C Scale units currently selected
Press the UP or DOWN button to change the units.
Un = F New units selected
The units can also be changed quickly from the main display by pressing the SET and UP buttons.
Page 33
6.11 Scan
The scan rate can be set and enabled so that when the set-point is changed the instrument heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the instrument heats or cools at the maximum possible rate.
6.11.1 Scan Control
The scan is controlled with the scan ON/OFF function that appears in the main menu after the set-point and unit’s functions.
ScAn=OFF Scan function OFF
Press the UP or DOWN button to toggle the scan ON or OFF.
ScAn=On Scan function ON
Press the SET button to accept the present setting and continue.
6 Controller Operation
Scan
S
Accept scan setting
6.12 Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set from 0.1 to 5.0°C/minute (0.2 to 9.0°F/min). The maximum scan rate, however, is actually limited by the natural heating or cooling rate of the instrument. This is typically less than 2°C/minute.
The scan rate function appears in the main menu after the scan control function. The scan rate units are in degrees per minute, degrees C or F depending on the selected units.
Sr= 0.2 Scan rate in °C/min
Press the UP or DOWN button to change the scan rate.
Sr=2.0 New scan rate
Press the SET button to accept the new scan rate and continue.
S
The display returns to the current well temperature.
Accept scan rate
29
Page 34
9230 Maintenance System
User’s Guide
6.13 Temperature Scale Units
To toggle between °C and °F, press and hold the SET button and then press the UP button when the temperature is displayed.
6.14 Secondary Menu
Functions used less often are accessed within the secondary menu. Press and hold the SET and EXIT buttons to access the secondary menu. The first func tion in the secondary menu is the power display. (See Figure 6 on page 22.)
6.15 Peltier Device Power
The temperature controller controls the temperature of the well by pulsing the Peltier devices ON and OFF. The total power being applied to the Peltier de vices is determined by the duty cycle or the ratio of device on time to the pulse cycle time. By knowing the amount of heating the user can tell if the instrument is heating up to the set-point, cooling down, or controlling at a constant temper­ature. Monitoring the percent power lets the user know the stability of the well temperature. With good control stability the percent heating power should not fluctuate more than ±1% within one minute.
The device power display is accessed in the secondary menu. Press and hold the SET and EXIT buttons. The Peltier device power is displayed as a percent­age of full power.
-
-
30
25.00 Well temperature
S+E
Access Peltier device power in secondary menu
SEC Flashes and then percent power flashes
Pct Po Flashes prior to displaying Peltier device power
6.0 P Peltier device power in percent
To exit out of the secondary menu press the EXIT button and hold for a brief moment. To continue on to the proportional band setting function, press the EXIT button momentarily or the SET button.
6.16 Set-point Resistance
The set-point resistance is the resistance associated with the set-point tempera ture. Allow the temperature to stabilize at the desired set-point before using the
-
Page 35
6 Controller Operation
Proportional Band
displayed resistance value. The set-point resistance function is available for di agnostic purposes.
Press the SET and EXIT buttons to enter the secondary menu and show the heater power. Then press the SET button twice to access the set-point resistance.
S+E
Access Peltier device power in secondary menu
SEC Flashes and then percent power flashes
Pct Po Flashes prior to displaying Peltier device power
6.0 Peltier device power in percent
S
Access set-point resistance
Sr Flashes “ ” (Set-point Resistance) and then displays the
setting
109.715 Resistance in ohms
6.17 Proportional Band
-
In a proportional controller such as this, the Peltier device output power is pro­portional to the well temperature over a limited range of temperatures around the set-point. This range of temperature is called the proportional band. At the bottom of the proportional band, the Peltier device output is ±100%. At the top of the proportional band, the Peltier device output is 0. Thus as the temperature rises the Peltier device power is reduced, which consequently tends to lower the temperature back down. In this way the temperature is maintained at a fairly constant value.
The temperature stability of the well and the response time depend on the width of the proportional band. If the band is too wide, the well temperature deviates excessively from the set-point due to varying external conditions. This devia tion is because the power output changes very little with temperature and the controller does not respond well to changing conditions or noise in the system. If the proportional band is too narrow the temperature may swing back and forth because the controller overreacts to temperature variations. For best con trol stability the proportional band must be set for the optimum width.
The proportional band width is set at the factory to about 8.0°C. The propor tional band width may be altered by the user to optimize the control character istics for a particular application.
-
-
-
-
31
Page 36
9230 Maintenance System
User’s Guide
The proportional band width is easily adjusted from the front panel. The width may be set to discrete values in degrees C or F depending on the selected units. The proportional band adjustment can be accessed within the secondary menu. Press the SET and EXIT buttons to enter the secondary menu and show the de vice power. Then press the SET button twice to access the proportional band.
-
S+E
SEC Flashes “ ” and then percent power flashes
Pct Po Flashes prior to displaying Peltier device power
6.0 Peltier device power in percent
S
Sr Flashes “ ” (Set-point Resistance) and then displays the
109.715 Resistance in ohms
S
Prop Flashes “ ” and then displays the setting
8.0 Proportional band setting
To change the proportional band press the UP or DOWN button.
Access Peltier device power in secondary menu
Access set-point resistance
setting
Access proportional band
32
9.0 New proportional band setting
To store the new setting press the SET button. Press the EXIT button to con tinue without storing the new value.
S
Accept the new proportional band setting
6.18 Controller Configuration, Operating, and Calibration Parameters
The controller has a number of configuration, operating, and calibration param eters that are programmable via the front panel. The first set of parameters is associated with the Auto Program and is located under the configuration submenus called prog 1, prog 2,andprog 3. The next set of parameters is
-
-
Page 37
Auto Program Configuration Parameters
6 Controller Operation
associated with the serial RS-232 interface and is located under the configura tion submenu called serIAL. The last set of parameters is associated with the calibration of the instrument and is located under the configuration submenu called CAL.
6.19 Auto Program Configuration Parameters
CAUTION: The Auto Program configuration parameters should not be adjusted under normal use. These parameters directly affect the melt pla teau duration time and proper freezing of the gallium cell.
The first set of parameters is associated with the Auto Program (see Figure 1, page 13). The Auto Program is a microprocessor-controlled program that deter mines set temperatures, scan rates, duration times, Peltier module configura tions and melt heater controls. By controlling all of these parameters in the proper order, a temperature profile can be maintained in the block assembly that assures a uniform melt, maintains the melt over long periods of time, prop­erly refreezes the cell and keeps the cell ready for the next melt cycle. The de-
-
-
-
-
33
Page 38
9230 Maintenance System
User’s Guide
fault programmable set-points (factory set) are given in Table 3, along with minimum and maximum set values.
Tabl e 3 Auto Program Parameters (Prog 1, Prog 2, Prog 3)
Parameter Name Parameter
Ready Temperature (WAIT)
Prep Melt Temperature (PREP)
Prep Scan Rate (PREP)
Beeper (PREP)
Prep 1 Time (PREP)
Prep 2 Time (PREP)
Prep 3 Time (PREP)
Maintain Temperature (MAINT)
Maintain Time-out (MAINT)
Maintain Time-out Time (MAINT)
Freeze Hot Temperature (FREEZH)
Freeze Hot Time (FREEZH)
Freeze Cold Temperature (FREEZC)
Freeze Cold Scan Rate (FREEZC)
Freeze Cold Time (FREEZC)
Freeze Melt State (Manual Mode Only)
*Factory Set
rdu t
PrEPt
prEPSr
bEEPEr
dur 1
dur 2
dur 3
nnAint
ti out
dur 4
FrEE2H
dur 5
FrEE2C
FrECSr
dur 6
frnELt
Low Value
29.000 29.270 29.300 °C 1 xx.xxx
30.000 30.770 35.000 °C 1 xx.xxx
0.1 0.2 0.5 °C/min 1 x.x
-ON--1ON/OFF
360 480 600 sec 1 xxx
120 240 360 sec 1 xxx
240 360 480 sec 1 xxx
29.790 29.860 35.000 °C 2 xx.xxx
-OFF--2ON/OFF
1 7200 43,200 min 2 xx,xxx
29.860 29.860 36.000 °C 2 xx.xxx
0 0 360 min 2 xxx
-1.000 0.000 10.000 °C 2 xx.xxx
0.4 0.5 0.6 °C/min 2 x.x
120 150 180 min 2 xxx
-MELT--3MELT/
Default Value*
High Value Units Prog Set Resolution
FREEZE
34
6.20 Prog 1
The parameters within the Prog 1 submenu are accessed from the secondary menu after the proportional band function by pressing the SET button. ConfIg flashes and then the name of the first Auto Program parameter menu ProG 1 is displayed:
8.0 Proportional band setting
S
ConFIG Flashes before entering configuration menus
Select the configuration menus
Page 39
ProG 1 Prog 1 submenu
Pressing the SET button accesses the parameters within the Prog 1 submenu. Pressing the UP button accesses Prog 2. The well temperature is displayed by pressing the EXIT button.
6.20.1 Ready Temperature
The Ready Temperature parameter is the temperature at which the instrument stabilizes before initiating a melt in the gallium cell. The default value is
29.27°C and can be seen on Figure 1 as the temperature that the well goes to when the display shows uuAit. The instrument holds the well at this tempera ture for 30 minutes minimum to assure that the gallium cell has stabilized, be fore allowing the melt to proceed.
When Prog 1 is displayed, press the SET button to access the Ready Tempera ture parameter. The display flashes rdu t and then displays the current Ready Temperature value:
ProG 1 Prog 1 submenu
6 Controller Operation
Prog 1
-
-
-
S
Access the ready temperature parameter
rdu t Flashes before entering ready temperature
29.270 Current ready temperature
Use the UP or DOWN buttons to adjust the value.
29.300 New ready temperature
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new ready temperature value and continue to Prep Melt Tem perature
6.20.2 Prep Melt Temperature
The Prep Melt Temperature parameter is the temperature at which the instru ment begins to melt the outer surfaces of the gallium within the gallium cell. The default value is 30.77°C and can be seen on Figure 1 as the temperature that the well goes to when the display shows Prep. The instrument holds the well at this temperature for approximately 18 minutes (the sum of Prep 1, Prep 2 and Prep 3 duration times) to begin melting the outer surfaces of the gallium.
-
-
-
35
Page 40
9230 Maintenance System
User’s Guide
After accessing the Ready Temperature parameter, the display flashes PrEP t and displays the current value of the Prep Melt Temperature.
PrEP t Flashes before entering prep melt temperature parameter
30.770 Current prep melt temperature
Use UP or DOWN buttons to adjust the value.
35.000 New prep melt temperature
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new Prep Melt Temperature value and continue to the Prep Scan Rate
6.20.3 Prep Scan Rate
The Prep Scan Rate parameter is the rate in °C/minute or °F/minute at which the instrument scans from the Ready Temperature to the Prep Melt Tempera­ture. The default value is 0.2°C/minute and can be seen on Figure 1, where the unit begins the melt.
After accessing the Prep Melt Temperature parameter, the display flashes
PrEP sr and displays the current value of the Prep Scan Rate.
PrEP Sr Flashes before entering prep scan rate parameter
Sr = 0.2 Current prep scan rate
Use the UP or DOWN buttons to adjust the value.
Sr = 0.5 New prep scan rate
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
-
-
36
S
6.20.4 Beeper
The Beeper parameter is used to enable or disable the audible warnings that may be generated during the initiation of the melt. The default value is ON.
Accept new prep scan rate value and continue to the inner melt heater Beeper
Page 41
6 Controller Operation
As seen on Figure 1, the Auto Program sounds an audible beep 4 times when it energizes the inner melt heater and the display shows an alternating message from PrEP to Htr on. After approximately 4 minutes, the Auto Program sounds an audible beep 8 times and turns the inner melt heater OFF (the display returns PrEP). After approximately 6 more minutes, the Auto Program sounds an audible beep 16 times and displays nnaint, an indication that the gallium melt has begun.
Changing the Beeper parameter to OFF disables the beeper during this phase of the Auto Program.
After accessing the Prep Scan Rate parameter, the display flashes beeper and displays the current value of the Beeper parameter.
Prog 1
bEEPEr Flashes before entering beeper parameter
On Current beeper setting (enabled)
Use the UP or DOWN buttons to adjust the value.
OFF New beeper setting
If no changes are desired, press the EXIT button to continue to the next param­eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new beeper setting and continue to the prep 1 time
6.20.5 Prep 1 Time
After beginning the gallium melt, see Figure 1, the Auto Program waits a short time before energizing the inner melt heater. The Prep 1 Time parameter speci fies this time in seconds. The default value is 480 seconds (8 minutes). At the end of this time, the Auto Program energizes the inner melt heater and sounds 4 audible beeps.
After accessing the Beeper parameter, the display flashes dur 1 and displays the current value of the Prep 1 Time parameter.
dur 1 Flashes before entering the prep 1 time parameter
SEC480 Current prep 1 time value (8 minutes)
Use the UP or DOWN buttons to adjust the value.
SEC600 New prep 1 time value (10 minutes)
-
37
Page 42
9230 Maintenance System
User’s Guide
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new prep 1 time value and continue to prep 2 time
6.20.6 Prep 2 Time
After energizing the inner melt heater during the initialization of the gallium melt, see Figure 1, the Auto Program waits a short time before turning OFF the inner melt heater. The Prep 2 Time parameter specifies this time in seconds. The default value is 240 seconds (4 minutes). At the end of this time, the Auto Program turns the inner melt heater OFF and sounds 8 audible beeps.
After accessing the Prep 1 Time parameter, the display flashes dur 2 and dis plays the current value of the Prep 2 Time parameter.
dur 2 Flashes before entering the prep 2 time parameter
SEC240 Current prep 2 time value (4 minutes)
Use the UP or DOWN buttons to adjust the value.
SEC360 New prep 2 time value (6 minutes)
If no changes are desired, press the EXIT button to continue to the next param­eter. Press the SET button to accept the new value and continue on to the next parameter.
-
-
38
S
Accept new prep 2 time value and continue to prep 3 time
6.20.7 Prep 3 Time
After turning OFF the inner melt heater during the initialization of the gallium melt, see Figure 1, the Auto Program waits a short time before beginning the maintenance process of the gallium melt plateau. During this time the instru ment remains at 30.77°C to complete the melting of the outer surfaces of the gallium in the cell. The Prep 3 Time parameter specifies this time in seconds. The default value is 360 seconds (6 minutes). At the end of this time, the Auto Program begins the maintenance process and sounds 16 audible beeps.
After accessing the Prep 2 Time parameter, the display flashes dur 3 and dis plays the current value of the Prep 3 Time parameter.
dur 3 Flashes before entering the prep 3 time parameter
SEC360 Current prep 3 time value (6 minutes)
-
-
Page 43
Use the UP or DOWN buttons to adjust the value.
SEC480 New prep 3 time value (8 minutes)
6 Controller Operation
Prog 2
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and to return to the Prog 1 submenu.
S
ProG 1 Prog 1 submenu
Press the UP button to access Prog 2. The well temperature is displayed by pressing the EXIT button. Press the SET button to access the parameters within the Prog 1 submenu.
6.21 Prog 2
Prog 2 is displayed by pressing the UP button while the display shows ProG 1 . Pressing the SET button then accesses the parameters within the Prog 2
submenu.
ProG 1 Prog 1 submenu
U
ProG 2 Prog 2 submenu
Pressing the DOWN button may access Prog 1. Pressing the UP button may ac cess Prog 3. The well temperature is displayed by pressing the EXIT button. Pressing the SET button accesses the parameters within the Prog 2 submenu.
-
Accept new prep 3 time value and return to the prog 1 submenu
Select the Prog 2 submenu
-
6.21.1 Maintain Temperature
The Maintain Temperature parameter is the temperature at which the instru ment stabilizes and remains during the melt plateau in the gallium cell. The de fault value is 29.86°C and can be seen on Figure 1 as the temperature that the well goes to when the display shows nnAint. The instrument holds the well at this temperature until the melt plateau is ended.
ProG 2 Prog 2 submenu
S
Access the maintain temperature parameter
nnaint Flashes before entering the maintain temperature
parameter
-
-
39
Page 44
9230 Maintenance System
User’s Guide
29.860 Current maintain temperature value
Use the UP or DOWN buttons to adjust the value.
29.790 New maintain temperature value
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new maintain temperature value and continue to the main tain time-out
6.21.2 Maintain Time-out
The Maintain Time-out setting is used to enable a preset delay time after which the gallium melt plateau is terminated. The default setting is OFF and at this setting the instrument remains at the Maintain Temperature until the user ad vances the Auto Program by depressing the SET and DOWN buttons and then selecting the FrEE2C program state. A refreeze of the gallium cell is then started.
If the Maintain Time-out setting is ON, the Auto Program automatically ends the gallium melt plateau after the preset delay time has been reached and refreezes the gallium cell. It can be seen on Figure 1 under the Maintain portion of the Auto Program.
After accessing the Maintain Temperature parameter, the display flashes ti
out and displays
ti out Flashes before entering the maintain time-out parameter
OFF Current maintain time-out setting
-
-
-
40
Use UP or DOWN buttons to adjust the value.
On New maintain time-out setting
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new maintain time-out setting and continue to the maintain time-out time
6.21.3 Maintain Time-out Time
The Maintain Time-out Time parameter is the time that the instrument main tains the gallium cell at the Maintain Temperature before terminating the gal
-
-
-
Page 45
6 Controller Operation
lium melt plateau. The default value is 7200 minutes (5 days). It is only used if the Maintain Time-out parameter is set to ON. If the Time-out parameter is set to ON, the instrument ends the Maintain portion of the Auto Program after the Maintain Time-out Time has been reached and refreezes the gallium cell. It can be seen on Figure 1 under the Maintain portion of the Auto Program.
After accessing the Maintain Time-out parameter, the display flashes dur 4 and displays the current value of the Maintain Time-out Time.
Prog 2
dur 4 Flashes before maintain time-out time parameter
7200 Current maintain time-out time value (7200 minutes)
Use the UP or DOWN buttons to adjust the value.
1440 New maintain time-out time value (1440 minutes)
If no changes are desired, press the EXIT button to continue to the next param­eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new maintain time-out time value and continue to the freeze hot temperature
6.21.4 Freeze Hot Temperature
After the melt plateau is finished or the user has ended the melt, the instrument normally proceeds to refreeze the cell. If desired, the Freeze Hot Temperature and Freeze Hot Time parameters may be used to preset the instrument to totally melt the cell before refreezing.
The Freeze Hot Temperature parameter is the temperature at which the instru ment totally melts the gallium in the cell. The value can be set up to 36.00°C and can be seen on Figure 1 as the temperature that the well goes to when the display shows FrEE2H. The instrument holds the well at this temperature for up to 6.0 hours to ensure that all of the gallium has been melted before refreezing the cell.
After accessing the Maintain Time-out Time parameter, the display flashes
FrEE2H and displays the current value of the Freeze Hot Temperature.
FrEE2H Flashes before entering the freeze hot temperature
parameter
29.86 Current freeze hot temperature value
Use the UP or DOWN buttons to adjust the value.
-
41
Page 46
9230 Maintenance System
User’s Guide
36.00 New freeze hot temperature value
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new freeze hot temperature value and continue to the freeze hot time
6.21.5 Freeze Hot Time
The Freeze Hot Time parameter is the time that the instrument maintains the gallium cell at the Freeze Hot Temperature to totally melt the gallium in the cell. The value can be set up to 360 minutes (6 hours) and can be seen on Fig ure 1 as the time duration when the display shows FrEEZH. The instrument re mains at the Freeze Hot Temperature for this time to ensure that all of the gallium has been melted before refreezing the cell.
Note: The default program does not use the Freeze Hot option and thus the Freeze Hot time is set to 0 seconds.
After accessing the Freeze Hot Temperature parameter, the display flashes dur
5 and displays the current value of the Freeze Hot Time.
dur 5 Flashes before entering the freeze hot time parameter
0 Current freeze hot time value (0 minutes)
Use the UP or DOWN buttons to adjust the value.
-
-
-
42
360 New freeze hot time value (360 minutes)
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new freeze hot time value and continue to the freeze cold temperature
6.21.6 Freeze Cold Temperature
The Freeze Cold Temperature parameter is the temperature to which the instru ment scans during the refreezing of the gallium cell. The default value is
0.00°C and can be seen on Figure 1 as the temperature the unit scans to while the display shows FrEE2C.
After accessing the Freeze Hot Time parameter, the display flashes Free2C and displays the current value of the Freeze Cold Temperature.
-
-
Page 47
6 Controller Operation
FrEE2C Flashes before entering the freeze cold temperature
parameter
0.000 Current freeze cold temperature value
Use the UP or DOWN buttons to adjust the value.
5.000 New freeze cold temperature value
Prog 2
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
S
Accept new freeze cold temperature value and continue to the freeze cold scan rate
6.21.7 Freeze Cold Scan Rate
The Freeze Cold Scan Rate parameter is the rate in °C/minute or °F/minute at which the instrument scans from the Freeze Hot Temperature to the Freeze Cold Temperature. The default value is 0.5°C/minute and can be seen on Figure 1 where the unit begins to refreeze the gallium cell. This parameter is very im­portant for the proper gradient to be maintained during the refreeze process.
After accessing the Freeze Cold Temperature parameter, the display flashes
frECsr and displays the current value of the Freeze Cold Scan Rate.
FrECSr Flashes before entering the freeze cold scan rate parameter
Sr = 0.5 Current freeze cold scan rate
Use the UP or DOWN buttons to adjust the value.
Sr = 0.6 New freeze cold scan rate
If no changes are desired, press the EXIT button to continue to the next param eter. Press the SET button to accept the new value and continue on to the next parameter.
-
-
S
Accept new freeze cold scan rate value and continue to the Freeze Cold Time
6.21.8 Freeze Cold Time
The Freeze Cold Time parameter is the time that the instrument scans to and maintains the gallium cell at the Freeze Cold Temperature to totally refreeze the gallium in the cell. The default value is 150 minutes (2.5 hours). At the end of
43
Page 48
9230 Maintenance System
User’s Guide
this time, the Auto Program returns the well to 25°C to wait for the next melt process.
After accessing the Freeze Cold Scan Rate parameter, the display flashes dur
6 and displays the current value of the Freeze Cold Time parameter.
dur 6 Flashes before entering the freeze cold time parameter
150 Current freeze cold time value (150 minutes)
Use the UP or DOWN buttons to adjust the value.
180 New freeze cold time value (180 minutes)
If no changes are desired, press the EXIT button to Prog 1 submenu. Press the SET button to accept the new value and continue on to the next parameter.
S
Prog 1 may be accessed by pressing the DOWN button. Prog 3 may be ac­cessed by pressing the UP button. The well temperature is displayed by press­ing the EXIT button. Pressing the SET button accesses the parameters within the Prog 2 submenu.
Accept new freeze cold time value and continue to the Prog 2 submenu
6.22 Prog 3
Prog 3 is displayed by pressing the UP button while the display shows ProG
2. Pressing the SET button then accesses the parameters within the Prog 3
submenu.
ProG 2 Prog 2 submenu
U
ProG 3 Prog 3 submenu
Prog 2 may be accessed by pressing the DOWN button. The serial interface menu may be accessed by pressing the UP button. The well temperature is dis played by pressing the EXIT button. Pressing the SET button accesses the pa rameters within the Prog 3 submenu.
6.22.1 Freeze Melt State
Select the Prog 3 submenu
-
-
44
The Freeze Melt State parameter is used to control the configuration of the Peltier devices and is only usable in the manual mode of operation. During nor mal operation of the instrument this parameter should not be altered.
-
Page 49
6 Controller Operation
Serial Interface Parameters
During the melt and maintain processes, the Peltier devices are configured to apply a uniform heat source to the well. This minimizes vertical gradients in the well and gallium cell to maximize the length of the melt plateau. During a Freeze process, the Peltier devices are configured to generate a large vertical gradient in the well to ensure the proper refreezing of the gallium cell.
The default setting for this parameter is MELT.
The Freeze Melt State parameter is accessed from the Prog 3 submenu.
ProG 3 Prog 3 submenu
S
Access the Freeze Melt State parameter
FrnELT Flashes before entering the freeze melt state parameter
nnELT Current freeze melt state setting
Use the UP or DOWN buttons to adjust the setting.
FrEE2E New freeze melt state setting
If no changes are desired, press the EXIT button to exit the Prog 3 submenu. Press the SET button to accept the new value and return to the Prog 3 submenu.
S
The display then shows ProG 3 and returns to the Prog 3 submenu.
Accept new Freeze Melt State setting
ProG 3 Prog 3 submenu
Prog 2 may be accessed by pressing the DOWN button. The serial interface menu may be accessed by pressing the UP button. The well temperature is dis played by pressing the EXIT button. Pressing the SET button accesses the pa rameters within the Prog 3 submenu.
-
-
6.23 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SEriAL Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters, which determine the operation of the serial interface. The parameters in the menu are – BAUD rate, sample period, duplex mode, and linefeed.
45
Page 50
9230 Maintenance System
User’s Guide
6.23.1 BAUD Rate
The BAUD rate is the first parameter in the menu. The BAUD rate setting de termines the serial communications transmission rate. The BAUD rate of the serial communications may be programmed to 300, 600, 1200, 2400, 4800, or
9600. The default setting is 2400 baud.
bAUd Flashes before displaying the serial baud rate setting
2400 b Current BAUD rate
Use the UP or DOWN buttons to adjust the setting.
1200 b New BAUD rate
Press the SET button to store the new setting as the baud rate or press the EXIT button briefly to skip to the next parameter without storing the new setting.
6.23.2 Sample Period
The sample period is the next parameter in the serial interface parameter menu. The sample period is the time period in seconds between temperature measure­ments transmitted from the serial interface. If the sample rate is set to 5, the in­strument transmits the current measurement over the serial interface approximately every five seconds. The automatic sampling is disabled with a sample period of 0.
SPer Flashes before displaying serial sample period setting
-
46
SP=1 Current sample period (seconds)
Use the UP or DOWN buttons to adjust the setting.
SP=60 New sample period
Press the SET button to store the new setting as the sample period or press the EXIT button briefly to skip to the next parameter without storing the new setting.
6.23.3 Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du plex or half duplex. With full duplex any commands received by the instrument via the serial interface are immediately echoed or transmitted back to the device of origin. With half duplex the commands are executed but not echoed.
dUPL Flashes before displaying serial duplex mode setting
-
Page 51
d=FULL Current duplex mode
Use the UP or DOWN buttons to adjust the setting.
d=HALF New duplex mode setting
Press the SET button to store the new setting as the duplex mode or press the EXIT button briefly to skip to the next parameter without storing the new setting.
6.23.4 Linefeed
6 Controller Operation
Calibration Parameters
The final parameter in the serial interface menu is the linefeed mode. This pa rameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return.
LF Flashes before displaying serial linefeed setting
LF=On Current linefeed setting
Use the UP or DOWN buttons to adjust the setting.
LF=OFF New linefeed setting
Press the SET button to store the new setting as the linefeed.
Prog 3 may be accessed by pressing the DOWN button. The calibration menu may be accessed by pressing the UP button. The well temperature is displayed by pressing the EXIT button. Pressing the SET button accesses the parameters within the serial interface menu.
6.24 Calibration Parameters
The operator of the instrument has access to one of the calibration constants, R0. This value is set at the factory and must not be altered. The correct value is important to the accuracy and proper and safe operation of the instrument. This parameter should not be adjusted except in the event the controller’s memory fails. The user may then restore this value to the factory setting. This constant is supplied to the user on the Report of Calibration with the manual.
-
CAUTION: DO NOT change the value of the instrument calibration con stant from the factory set value. The correct setting of this parameter is important to the safety and proper operation of the instrument.
CAL Calibration parameters menu
-
47
Page 52
9230 Maintenance System
User’s Guide
Press the SET button five times to enter the menu.
6.24.1 R0
The calibration parameter R0 characterizes the resistance temperature relation ship of the platinum control sensor. This parameter may be adjusted by an ex perienced user to improve the accuracy of the instrument.
To properly maintain the gallium cell during the melt plateau and during the refreezing process, the accuracy of the Model 9230 near the gallium melting point (29.7646°C) must be very precise, ±0.05°C or better. The value of R0 for this instrument refers to the resistance of the control probe at 0°C. For calibra tion purposes, the value of R0 is adjusted to compensate for the error at
29.27°C. The value of this parameter is set at the factory for best instrument ac curacy. This parameter may be adjusted by an experienced user to improve the accuracy of the instrument (see Section 10, Calibration Procedure).
From the CAL submenu, pressing the SET button 5 times accesses the R0 parameter:
-
CAL Calibration parameters menu
Press the SET button five times to enter the menu.
r0 Flashes before displaying the R0 setting
99.788 Current R0 value (ohms)
Use the UP or DOWN button to adjust the setting.
99.999 New R0 value (ohms)
-
-
-
48
If no changes are desired, press the EXIT button to return to the CAL submenu. Press the SET button to accept the new value.
The display then shows CAL and returns to the CAL submenu. The serial inter face menu may be accessed by pressing the DOWN button. The well tempera ture is displayed by pressing the EXIT button. Pressing the SET button five times accesses the R0 parameter within the CAL submenu.
-
-
Page 53
7 Digital Communication Interface
7 Digital Communication Interface
The Model 9230 is capable of communicating with and being controlled by other equipment through the digital interface.
With a digital interface the instrument may be connected to a computer or other equipment. This allows the user to set the set-point temperature, monitor the temperature, and access any of the other controller functions, all using remote communications equipment. Communications commands are summarized in Table 4 on page 52.
7.1 Serial Communications
The instrument is installed with an RS-232 serial interface that allows serial digital communications over fairly long distances. With the serial interface the user may access any of the functions, parameters and settings discussed in Sec tion 7, Controller Operation with the exception of the BAUD rate setting. Addi tionally, when the instrument is in Standby mode (well at 25°C and display shows Auto), the user is required to use the front panel buttons, to enter Man­ual mode or to initiate the Auto Program.
The serial interface operates with eight data bits, one stop bit, and no parity.
Serial Communications
-
-
7.1.1 Wiring
The serial communications ca­ble attaches to the instrument through the DB-9 connector at the back of the instrument. Fig­ure 7 shows the pin-out of this connector and suggested cable wiring. To eliminate noise, the serial cable should be shielded with low resistance between the connector (DB-9) and the shield. If the instrument is used in a heavy industrial setting, the serial cable must be limited to one meter in length.
7.1.2 Setup
Before operation, the serial in terface must first be set up by programming the BAUD rate and other configuration parame ters. These parameters are pro
-
-
-
Figure 7 Serial Cable Wiring
49
Page 54
9230 Maintenance System
User’s Guide
grammed within the serial interface menu.
To enter the serial interface menu from the standby mode do the following:
Auto Instrument in standby mode
Use the UP or DOWN buttons to select the manual mode.
nnAn Instrument in manual mode
S
25.00 C Well temperature displayed
S+E
Pct Po Flashes prior to displaying Peltier device power
18.0 P Peltier device power in percent
S
ConFIG Flashes configuration menu
ProG 1 Program 1 submenu
U
SErIAL Serial interface menu
Enter the manual mode
Enter the secondary menu
SEC Flashes and then
Continue pressing the SET button until the display shows
Press the UP button three times until is displayed
50
S
7.1.2.1 Baud Rate
The baud rate is the first parameter in the menu. The display prompts with the baud rate parameter by showing “bAUd”. Press the SET button to choose to set the baud rate. The current baud rate value is displayed. The baud rate may be programmed to 300, 600, 1200, or 2400 baud. The baud rate defaults to 2400 baud.
Access the serial interface parameters
Page 55
Use the UP or DOWN buttons to change the baud rate value. Press the SET button to set the baud to the new value or the EXIT button to abort the opera tion and skip to the next parameter in the menu.
7.1.2.2 Sample Period
The sample period is the next parameter in the menu and prompted with “SPEr”. The sample period is the time period, in seconds, between tempera ture measurements transmitted from the serial interface. If the sample rate is set to 5, the instrument transmits the current measurement over the serial interface approximately every five seconds. The automatic sampling is disabled with a sample period of 0. Press the SET button to choose to set the sample period. Adjust the period with the UP or DOWN button and then use the SET to set the sample rate to the displayed value.
7.1.2.3 Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full duplex any commands received by the instrument via the serial interface are immediately echoed or transmitted back to the device of origin. With half du­plex the commands are executed but not echoed. The default setting is full du­plex. The mode may be changed using the UP or DOWN buttons and pressing the SET button.
7 Digital Communication Interface
Interface Commands
-
-
7.1.2.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa­rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac- ter (LF, ASCII 10) after transmission of any carriage-return. The default setting is with linefeed on. The mode may be changed using the UP or DOWN buttons and pressing the SET button.
7.1.3 Serial Operation
Once the cable has been attached and the interface set up properly the control ler immediately begins transmitting temperature readings at the programmed rate. The set-point and other commands may be sent to the instrument via the serial interface to set the temperature set-point and to view or program the vari ous parameters. The interface commands are discussed in Section 8.3, Interface Commands.
The serial interface operates with eight data bits, one stop bit, and no parity.
7.2 Interface Commands
The various commands for accessing the instrument functions via the digital in terface are listed in this section. These commands are used with the RS-232 se rial interface. The commands are terminated with a carriage-return character. The interface makes no distinction between upper and lower case letters, hence
-
-
-
-
51
Page 56
9230 Maintenance System
User’s Guide
either may be used. Commands may be abbreviated to the minimum number of letters, which determines a unique command. A command may be used to ei ther set a parameter or display a parameter depending on whether or not a value is sent with the command following an “=” character. For example, an “s”<cr> returns the current set-point and an “s=28.00”<cr> sets the set-point to 28.00 degrees.
In the following list of commands, characters or data within brackets, “[“ and “]”, are optional. A slash, “/”, denotes alternate characters or data. Numeric data, denoted by “n”, may be entered in decimal or exponential notation. Char acters are shown in lower case although upper case may be used. Spaces may be added within command strings and are simply ignored. Backspace (BS, ASCII 8) may be used to erase the previous character. A terminating CR is im plied with all commands.
Tabl e 4 Serial Interface Commands
-
-
-
Command Description Command Format
Display Temperature
Read current set-point s[etpoint] s set: 9999.99{C or F} set: 25.00 C
Set current set-point to n s[etpoint]=n s=28 -5.00C to
Read units u u u:{C or F} u: C
Read temperature t t t: 9999.99{C or F} t: 29.86 C
Set temperature units: u[nits]=c/f CorF
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to
Fahrenheit
Read scan function sc[an] sc scan:{ON or OFF} scan:ON
Set scan function: sc[an]=on/of[f] ON or OFF
Turn scan function on sc[an]=on sc=on
Turn scan function off sc[an]=of[f] sc=of
Read scan rate sr[ate] sr srat: 999.99{C or F}/min srat: 0.5C/min 0.1C/min to
Setscanrateto minute
Read program state adv adv adv:{OFF, WAIT, PREP,
Advance program state adv={adv/auto} adv=adv ADV, AUTO
Secondary Menu
Read heater power (duty cycle)
n
degrees per
u[nits]=f u=f
sr[ate]=n]sr=5
po[wer] po po: 999.9 po: 12.3
Command Example Returned
MAINTAIN, FREEZHOT, FREEZCOLD}
Returned Example
WAIT
Acceptable Values
40.00C
5.0C/min
52
Page 57
7 Digital Communication Interface
Interface Commands
Command Description Command Format
Read proportional band pr[op-band] pr pb: 999.9 pb: 8
Set proportional band pr[op-band]=n pr=8.83 0.1C to 100C
Read set-point resistance *sr *sr 999.999 111.153
Configuration Menu
Program 1
Read ready temp rd[y] rd readytemp :99.99(C orF)readytemp:29.2
Set ready temp rd[y]=n rd=29.27 28.0 to 29.30 C
Read prep melt temperature me me Preptemp:999.99(C orF)Preptemp:30.77
Set prep melt temperature me=n me=30.77 30.0 to 35.0 C
Read prep scan rate ps[ra] ps Prepsrate :99.9(C or
Set prep scan rate ps[ra]=n ps=0.2 0.1 to 0.5 C/min
Read beeper setting bee[p] bee beep: ON/OFF beep: ON
Set beeper bee[p]=on/of[f] bee=off ON or OFF
Read prep 1 duration prea prea Prep1dur :999 sec Prep1dur :360
Set prep 1 duration prea=n prea=400 360 to 600 sec
Read prep 2 duration preb preb Prep2dur :999 sec Prep2dur :360
Set prep 2 duration preb=n preb=300 120 to 360 sec
Read prep 3 duration prec prec Prep3dur :999 sec Prep3dur :360
Set prep 3 duration prec=n prec=400 240 to 480 sec
Program 2
Read maintain temperature ma ma ma:99.999{C or F} ma:29.86 C
Set maintain temperature ma=n ma=29.86 29.79 to 35.00
Read maintain duration dm dm dm:99999 dm:7200
Set maintain duration dm=n/OFF dm=1000 OFF,1 to 43200
Read freeze hot temperature freh freh freezHtemp:99.99(C orF)freezHtemp:99.
Set freeze hot temperature freh=n freh=35.00 29.86 to 36.00
Read freeze hot duration dfrh dfrh freezHdur : 999 min freezHdur : 150
Set freeze hot duration dfrh=n dfrh=300 0 to 360 min
Read freeze cold temperature fr[ec] fr freezCtemp :99.99(C orF)freezCtemp
Command Example Returned
F)/min
Returned Example
7C
C
Prepsrate:0.2 C/min
Sec
Sec
Sec
99(C or F)
:99.99(C or F)
Acceptable Values
C
min
C
53
Page 58
9230 Maintenance System
User’s Guide
Command Description Command Format
Set freeze cold temperature fr[ec]=n fr=0.01 -0.01 to 10.00 C
Read freeze cold scan rate fc[sr] fc freezCsrate: 0.2(C or
Set freeze cold scan rate fc[sr]=n fc=0.6 0.4 to 0.6 C/min
Read freeze cold duration d[frc] dfrc freezCdur : 140 min freezCdur : 140
Set freeze cold duration d[frc]=n df=170 120 to 180 min
Program 3
Read freeze/melt state frm[t] frm FreezeMelt: MELT Mode FreezeMelt:
Set freeze/melt state frm[t]=n frm=melt MELT or
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 9 sa: 1
Set serial sampling setting to seconds
Set serial duplex mode: du[plex]=f[ull]/h[alf] FULL or HALF
Set serial duplex mode to full du[plex]=f[ull] du=f
Set serial duplex mode to half du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f] ON or OFF
Set serial linefeed mode to on lf[eed]=on lf=on
Set serial linefeed mode to off lf[eed]=of[f] lf=of
Calibration Menu
Read R0 calibration parameter r[0] r r0: 999.999
Set R0 calibration parameter r[0]=
These commands are only used for factory testing
Miscellaneous (not on menus)
Read firmware version number *ver[sion] *ver ver: 999,9.99 ver.9230,v1.00
Read structure of all commands h[elp] h list of commands
Legend: [] Optional Command data
NOTE:
n
sa[mple]=
n
n
{} Returns either information n Numeric data supplied by user 9 Numeric data returned to user x Character data returned to user
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Command Example Returned
F)/min
sa=0 0 to 10,000
r=100.713 98.0 to 102.0
Returned Example
freezCsrate:
0.2(C or F)/min
MELT Mode
Acceptable Values
FREEZE
ohms
54
Page 59
8 Melting Point of Gallium Realization
8 Melting Point of Gallium Realization
This section explains the procedure for realizing the melting point of gallium using a Model 5943 Stainless Steel Gallium Cell and the Model 9230 Mainte nance System.
8.1 General
The gallium melting point (29.7646°C) is a critical temperature. Thermometers used in life science, environmental monitoring, at room or body temperatures, and many other applications depend on it for accurate calibrations. Laboratory standards rely on it as an ITS-90 check standard and as a means of measuring drift between calibrations. Use of the Model 5943 Stainless Steel Gallium Cell and the Model 9230 Maintenance System provides an easy way to generate and maintain the gallium melting point. Plateaus lasting up to 5 days are typical.
The Model 5943 Stainless Steel Gallium Cell holds a gallium sample that is
99.99999+% pure. The gallium is sealed in a Teflon envelope in a high purity argon atmosphere, which is then sealed inside a stainless steel housing. This double-sealing method reduces leaching of the gallium sample and ensures a life of ten years or longer for the cell. Gallium expands on freezing by 3.1% and, to compensate for this, the inner construction of the cell incorporates a flexible Teflon wall. Although the cell is more rugged than other gallium cells, care must be taken to prevent damage to the cell, particularly during the refreeze process. Because gallium expands on freezing (like water), the cell must be frozen from the bottom up or there is potential for damage to the cell. The Auto Program process of the Model 9230 ensures that the cell is frozen properly. The cell’s entrant well dimensions are 8.7” (length) x 0.324” (diame­ter) with an immersion depth of 6.58” in gallium.
The Model 9230 Gallium Maintenance Apparatus is specifically designed to maintain the Model 5943 Gallium Cell. Once properly installed in the Model 9230, the Auto Program may be started and subsequently initiates the melt pla teau, maintains the plateau for up to 5 days typical, and properly refreezes the cell for the next melt. When not actively melting or refreezing a cell, the Model 9230 automatically maintains the cell at 25.00°C (Standby), well below the melting point. To initiate a melt, the inner melt heater is inserted into the cell and the Auto Program is started through the Model 9230’s front-panel buttons. The Model 9230 increases its temperature to just above the gallium point and powers the 8-watt inner melt heater. Approximately 30 minutes later, the gal lium cell is ready to be used (see Figure 1).
The Model 9230 uses a Proportional Integral Derivative PID controller with specialized analog circuitry to control several Peltier devices. The controlled Peltier devices provide the appropriate amount of heating or cooling to the block assembly into which the gallium cell is inserted. The controller also con trols the inner melt heater, cooling fan and beeper alarm. Two preheat wells en sure that thermometers are properly warmed when inserted into the gallium cell. The preheat wells are oversized so thermometers fit loosely and heat loss
General
-
-
-
-
-
55
Page 60
9230 Maintenance System
User’s Guide
from the gallium cell is neither too sudden nor too severe when cold thermome ters are initially placed into the wells.
At the end of the melting plateau (or when the user wishes to end the plateau), the Auto Program is advanced. When totally melted, the Peltier devices are re configured and controlled such that a large temperature gradient is generated between the top and bottom of the block. The instrument slowly scans to 0°C to maintain the gradient over several hours. This ensures that the gallium cell is frozen from the bottom up.
In summary, the following steps are involved in the realization of the melting point of gallium:
Prepare the Model 9230 to receive the gallium cell.
Prepare the frozen Model 5943 gallium cell to be installed in the Model
9230.
Install the frozen Model 5943 Stainless Steel Gallium Cell into the Model
9230.
Insert the top insulation plugs into the well on top of the cell.
Insert the inner melt heater into the entrant well of the gallium cell.
Start the Auto Program.
Generate the melt plateau and remove the inner melt heater when
prompted by the Model 9230 beeper alarm.
Calibrate thermometers during the gallium melt plateau.
Advance the Auto Program to refreeze the cell.
The details of operating the Model 9230 to realize the melting point of gallium are explained for each step in the following sections.
-
-
56
8.2 Preparing the Model 9230 to Receive the Gallium Cell
The Model 9230 is designed to be used in a controlled laboratory environment (see Section 3.2, Environmental Conditions). Control the ambient temperature to a fixed temperature within the specified operating range (18°C to 28°C). Rapid changes in ambient temperature of more than ±1.0°C reduces the stabil ity of the instrument. Avoid drafty areas.
Place the Model 9230 on a flat surface with at least 6 inches of free space around and 18 inches above the instrument.
Make sure that the well of the instrument is clean and clear of any foreign ob jects, debris, dirt or grit.
Connect the inner melt heater to the back of the instrument making sure that it is securely connected. Place the inner melt heater cartridge into the melt heater storage well at the top back of the instrument.
-
-
Page 61
Installation in the Model 9230
Preparing the Frozen Model 5943 Gallium Cell for
8 Melting Point of Gallium Realization
Verify that the POWER switch is in the OFF position (“0”). Verify that the In ner Melt Heater switch is in the AUTO position. Install the power cord into the power entry module on the back of the instrument and plug the cord into a mains outlet of the proper voltage, frequency, and current capability. Turn on the power to the instrument by toggling the switch on the power entry module located on the back of the instrument. The fan should begin blowing air through the instrument, the beeper sounds one beep, and the display illuminates after approximately 3 seconds. After a brief self-test the controller should begin normal operation, the instrument will begin to approach 25°C, and the display will show Auto.
8.3 Preparing the Frozen Model 5943 Gallium Cell for Installation in the Model 9230
Maintain the Model 5943 gallium cell at temperatures below the melting point (29.7646°C). The cell must be frozen when installed in the Model 9230.
Make sure that the outer surfaces and entrant well of the frozen Model 5943 gallium cell are clean and that the entrant well of the cell is clear of any foreign objects, debris, dirt or grit. Use pure or reagent grade alcohol and clean, soft cotton cloths to clean all surfaces if necessary. Thoroughly dry the cell before continuing.
Add a small amount of silicone oil 200.05, or equivalent transfer fluid, to the entrant well of the cell. Do not overfill the entrant well. The oil level in the well should be approximately the same level (52 mm below the top of the cell) as that of the gallium in the cell when the thermometer is in place in the well.
-
8.4 Installing the Frozen Model 5943 Gallium Cell into the Model 9230
The instrument comes with a cell removal tool to aid in installing and removing the gallium cell from the Model 9230 well. The tool has a spring-loaded handle with a compressible rubber tip. When the handle is squeezed, thereby com pressing the spring, the rubber tip is extended and the tip may be inserted into the entrant well of the cell. Releasing the handle causes the rubber tip to com press and become firmly positioned in the entrant well. Once lodged in the en trant well, the cell can be gently lowered into the Model 9230 well. Wet the rubber tip of the tool with transfer fluid (e.g. silicone oil 200.05) for ease in in serting into the cell’s entrant well.
CAUTION: Do not use the cell removal tool to carry the cell. The tool is to be used only to aid in installing and removing the cell from the Model 9230 well. Do not drop the cell into the well. Damage to the cell and Model 9230 may occur if the cell is dropped.
-
-
-
-
57
Page 62
9230 Maintenance System
User’s Guide
With one hand holding the outside of the cell, insert the cell removal tool tip into the cell’s entrant well as described above and verify that the rubber tip is firmly lodged after releasing the tool handle. Hold onto the removal tool with the other hand and gently lower the cell with the tool the rest of the way while carefully releasing the grip on the outside of the cell. When the cell is securely lowered into the Model 9230 well, compress the removal tool handle to loosen the rubber tip and gently remove the tool from the cell’s entrant well. The gal lium cell is now installed in the Model 9230.
8.5 Inserting the Top Insulation Plugs
The instrument comes with 4 rubber insulation plugs to be inserted into the Model 9230 well on top of the cell (2 spares). The plugs have a hole in the cen ter through which thermometers can be inserted. The plugs serve two purposes. First, they help ensure an even temperature profile in the block and cell allow ing a longer lasting and more accurate melting-point realization. Second, they protect against excessive moisture condensation during the cell refreezing pro cess. Condensation can lead to corrosion over the long term and freezing water can damage the block.
Firmly insert two insulation plugs into the Model 9230 well on top of the cell. Make sure that they are positioned such that the center holes are aligned with the entrant well of the gallium cell.
-
-
-
-
58
8.6 Inserting the Inner Melt Heater
The inner melt heater’s function is to melt a small volume of gallium around the entrant well of the cell when initiating the melt. The recommended practice for realizing the gallium melt by national standards laboratories requires two solid/liquid interfaces. An inner liquid interface exists at the reentrant well with an outer liquid interface completely surrounding the inner interface at the con tainer wall. For a detailed description of the liquid/solid/liquid interfaces in the realization of the gallium melt see NIST Note 1265.
Verify that the inner melt heater is securely connected to the back of the instru ment. Remove the inner melt heater from the storage well and carefully insert it through the holes in the top insulation plugs and into the entrant well of the gal lium cell. Make sure that the heater cartridge is completely inserted to the bot tom of the cell’s entrant well.
8.7 Auto Program
The Model 9230 is designed to automate the gallium cell melt plateau. This is accomplished through the use of the Auto Program, a microprocessor-con trolled program that determines set temperatures, scan rates, duration times, Peltier module configurations and melt heater controls. By controlling all of these parameters in the proper order, a temperature profile can be maintained in the block assembly that assures a uniform melt, maintains the melt over long
-
-
-
-
-
Page 63
8 Melting Point of Gallium Realization
Auto Program
periods of time, properly refreezes the cell and keeps the cell ready for the next melt cycle. The Auto Program is illustrated in Figure 1.
8.7.1 Starting the Auto Program
After the frozen gallium cell has been properly installed in the instrument and the unit has been powered ON, the front panel LED display shows Auto.The instrument goes to 25°C and remains there until the SET button is pressed on the front panel starting the Auto Program. After pressing the SET button, the display changes to uuait and the Auto Program begins. The instrument auto matically resets to 29.27°C at a scan rate of 0.2°C/minute in accordance with the Auto Program (see Figure 1).
8.7.2 Generating the Melt Plateau and Removing the Inner Melt Heater
Once the unit has stabilized at 29.27°C, it waits for a minimum of 30 minutes to ensure that the gallium cell has equalized in temperature. At the end of the equalization time, the Auto Program begins to initiate the melt. The display changes to PrEP as the set temperature of the block is changed to 30.77°C. This initiates a melt on the outer surfaces of the gallium within the cell. A few minutes into the program, the unit beeps four times and the display changes to a message that flashes between PrEP and Htr on to indicate that the inner melt heater has been turned ON. The inner melt heater remains ON for a few minutes, then shuts off as the display changes back to PrEP , and the unit beeps 8 times. The user should remove the inner melt heater at this time (Note: Damage does not occur to the gallium cell if the heater is not removed, how­ever, a shortening of the melt duration may result).
After the inner melt heater has been powered OFF, the instrument remains at
30.77°C for a few more minutes to complete the outer melt of the gallium cell. The unit then beeps 16 times, the display changes from PrEP to nnAinT and the block temperature resets to 29.86°C to maintain the melting gallium cell.
-
8.7.3 Calibrating Thermometers
After the melt has been initiated, a preheated monitor thermometer may be in serted to monitor the temperature of the gallium cell. Due to the heat applied to the cell by the inner melt heater, the temperature of the cell’s entrant well is higher than the gallium melting point (29.7646°C) for a short period of time. Generally, after 30 to 60 minutes, the temperature of the cell’s entrant well is sufficiently stable to begin calibrations of thermometer sensors. The duration of the melt plateau varies depending on usage, but lengths of 5 days or longer are achievable.
The two preheat wells in the block may be used to heat up the thermometers before inserting them into the cell. The preheat wells are slightly oversized so thermometers fit loosely and heat loss from the gallium cell is neither too sud den nor too severe when cold thermometers are initially placed in the preheat wells.
-
-
59
Page 64
9230 Maintenance System
User’s Guide
8.7.4 Advancing the Auto Program and Refreezing the Gallium Cell
At the end of the melt plateau (or when the user wishes to end the plateau), the Auto Program is advanced by pressing the SET and DOWN buttons and then using the UP and DOWN buttons to select the next program state. Press SET to start the program. See Section 6.21.2, Maintain Time-out, for optional user de fined Auto Time-Out. Gallium, like water, expands in volume when freezing. Therefore, the cell must be frozen from the bottom up or there is a potential for damage to the cell. The Auto Program process ensures that the cell is properly refrozen.
The Auto Program initiates the freeze and the display shows FrEE2C at this point. The Peltier devices are reconfigured and controlled such that a large tem perature gradient is generated between the top and bottom of the block. The in strument slowly scans to 0°C to maintain the gradient over several hours.
8.7.5 Return to Standby
Upon completion of the freeze, the Auto Program returns the unit to 25°C, the front panel LED display shows Auto , and the unit remains there until the user initiates the Auto Program again by pressing the SET key.
-
-
-
60
Page 65
9 Calibration Procedure
9 Calibration Procedure
Calibration Equipment
Note:This procedure is to be considered a general guideline. Each labora tory should write their own procedure based on their equipment and their quality program. Each procedure should be accompanied by an uncer tainty analysis also based on the laboratory’s equipment and environment.
To properly maintain the gallium cell during the melting plateau and during the refreezing process, the accuracy of the Model 9230 near the gallium melting point (29.7646°C) must be very precise, ±0.05°C or better. To achieve this, the instrument must be calibrated near this temperature. The value of R0 for this instrument refers to the resistance of the control probe at 0°C. For calibration purposes, the value of R0 is adjusted to compensate for the error at 29.27°C. The value of this parameter is set at the factory for best instrument accuracy and is located on the Report of Calibration shipped with the instrument. This parameter may be adjusted by an experienced user to improve the accuracy of the instrument. A calibration should be done at regular intervals, such as once a year, to ensure that it is operating properly.
9.1 Calibration Equipment
Calibration of the Model 9230 requires a standard thermometer and readout de­vice that are adequately accurate. The combined accuracy of the thermometer and readout should be ±0.0125°C or better. Recommended equipment includes a laboratory grade PRT or thermistor probe with a length of 230 to 300 mm (9 to 12 inches). The calibration may be performed in a frozen Model 5943 gal­lium cell or in an aluminum test block that simulates the gallium cell. A small amount of silicone oil 200.05 (or equivalent) should be used in the entrant well for good thermal transfer from the cell to the thermometer.
-
-
9.2 Calibration Procedure
The accuracy of the Model 9230 at the gallium melting point (29.7646°C) is determined by the value of the calibration parameter R0. The calibration proce dure involves measuring the error between the Model 9230 and the reference thermometer at a temperature near the gallium melting point and adjusting the calibration parameter R0 as necessary to reduce the errors to within acceptable limits. The stated accuracy of the Model 9230 is given in Section 3.1, Specifi cations. For calibration purposes, the temperature is measured at 29.27°C and R0 is adjusted as applicable. The following steps may be followed to calibrate the Model 9230:
1. Enter the Manual mode of operation and set the temperature of the Model 9230 to 29.270°C with a frozen gallium cell installed. Use a small amount of silicone oil 200.05 (or equivalent) in the entrant well for good thermal transfer from the cell to the thermometer. Be sure to install the 3 top insulation plugs
-
-
61
Page 66
9230 Maintenance System
User’s Guide
on top of the cell. Insert the reference thermometer and monitor the tempera ture of the cell with an appropriate readout.
2. Allow the system to stabilize at 29.270°C for approximately 1 to 2 hours. When the stability of the reading is less than ±0.02 over a 20 to 30 minute pe riod record the average reading of the reference thermometer. This reading gives the error of the Model 9230 and may be used for “as found” data if re quired. No further action is required if the unit is found to be within acceptable limits.
3. If step 2 above shows that the Model 9230 is out of specification or accept able limits, use the following equation to calculate a new R0 value:
R
= R1– (T2– T1) * (0.3850)
2
Where,
R
= the new calculated value for R0
2
R
= the old value for R0 currently programmed in the Model 9230 mem
1
ory
T
= the temperature measured in step 2 above
2
T
= the set temperature of the 9230 from step 1 above (29.270°C)
1
The value of R
9230. Use the following process with the front panel buttons to obtain the cur­rently programmed value of R0:
(old R0) should be obtained from the memory of the Model
1
-
-
-
-
25.00 C Instrument in manual mode; well temperature displayed
-
62
S+E
Press the SET button and then the EXIT button to enter the secondary menu
SEC Flashes secondary menu
Pct Po Flashes percent power
18.0P Peltier device power in percent
S
Press the SET button until is displayed
ConFIG Flashes configuration menu
ProG 1 Prog 1 submenu
U
Press the UP button until is displayed
Page 67
CAL Calibration parameter menu
9 Calibration Procedure
Calibration Procedure
S
Press the SET button 5 times to access the R0 value
R0 Flashes before entering the R0 setting
99.983 Current R0 value (ohms)
4. Program the new value for R0 (R memory (see Section 6.24.1, R0).
5. Repeat steps 1 and 2 to verify that the newly calibrated Model 9230 is now within acceptable limits. If not, perform steps 3, 4, and 5 again as necessary.
Note: If it is known that the Model 9230 is more than 0.5°C out of specifica tion, such that, if the unit was calibrated at 29.270°C the cell would begin melt ing, then the above procedure should be performed below the 29.270°C set point as applicable (e.g., 28.5°C). After adjusting the unit at the lower tempera­ture (e.g., 28.5°C), and verifying that the error had been reduced to less than
0.5°C, then the above procedure may be safely used.
) from step 3 above into the Model 9230
2
-
-
63
Page 68
10 Maintenance
The calibration instrument has been designed with the utmost care. Ease
of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in an oily, wet, dirty, or dusty environment.
If the outside of the instrument becomes soiled, it may be wiped clean
with a damp cloth and mild detergent. Do not use harsh chemicals on the surface, which may damage the paint.
It is important to keep the well of the instrument clean and clear of any
foreign matter.
The instrument should be handled with care. Avoid knocking or dropping
the instrument.
Examine the gallium cell for deformities before inserting it in the well. If
there is any chance of jamming the cell in the well, file or grind off the protuberance.
Do not drop the probes into the well. This type of action can cause dam­age to the entrant well of the gallium cell.
If a hazardous material is spilled on or inside the equipment, the user is responsible for taking the appropriate decontamination steps as outlined by the national safety council with respect to the material.
If the mains supply cord becomes damaged, replace it with a cord with the appropriate gauge wire for the current of the instrument. If there are any questions, call Hart Scientific Customer Service for more informa­tion.
Before using any cleaning or decontamination method except those rec ommended by Hart, users should check with Hart Scientific Customer Service to be sure that the proposed method does not damage the equip ment.
If the instrument is used in a manner not in accordance with the equip ment design, the operation of the instrument may be impaired or safety hazards may arise.
Do not drop the gallium cell into the instrument well. This type of action can cause a shock to the control sensor, thermoelectric modules, or dam age the cell.
10 Maintenance
-
-
-
-
65
Page 69
11 Troubleshooting
11 Troubleshooting
This section contains information on troubleshooting, CE Comments, and a wiring diagram.
11.1 Troubleshooting Problems, Possible Causes, and Solutions
In the event that the instrument appears to function abnormally, this section may help to find and solve the problem. Several possible problem conditions are described along with likely causes and solutions. If a problem arises, please read this section carefully and attempt to understand and solve the problem. If the problem cannot otherwise be solved, contact Hart Scientific Customer Ser vice for assistance (1-801-763-1600). Be sure to have the model number and serial number of your instrument available.
Problem Possible Causes and Solutions
Incorrect temperature reading
Incorrect R0 parameter. Find the value for R0 on the Report of Calibration that was shipped with the instrument (or from subsequent calibrations of the in­strument). Reprogram the parameter into the Model 9230 memory (see Section
6.24.1, R0). Allow the instrument to stabilize and verify the accuracy of the temperature reading.
-
Blank display after mains power applied
The Instrument heats or cools too quickly or too slowly
Controller locked up. The controller may have locked up due to a power surge or other aberration. Initialize the system by performing the Factory Reset Sequence.
Factory Reset Sequence. Hold the SET and EXIT buttons down at the same time while powering on the instrument. After the instrument displays release the buttons. The display shows plays the firmware version, and then displays ter reset sequence, all of the configuration parameters are reset to their default values. Reprogram the R0 parameter into the Model 9230 memory (see Sec tion 6.24.1, R0) and any other applicable configuration parameters. Allow the instrument to stabilize and verify the accuracy of the temperature reading.
Blown fuse. A fuse may have blown due to a power surge or failure of a com ponent. Replace the fuse once. If the fuse blows a second time, it is likely caused by the failure of a component. Always replace the fuse with one of the same rating, voltage, and type. Never replace the fuse with one of a higher cur rent rating.
Incorrect scan and scan rate settings. The scan and scan rate settings may be set to unwanted values. Check the Scan and Scan Rate settings. The scan may be off (if the unit seems to be responding too quickly). The scan may be on with the Scan Rate set low (if unit seems to be responding too slowly).
-init-
, then displays
Auto
. After performing the mas
9230
-init-
, then dis
,
-
67
-
-
-
-
Page 70
9230 Maintenance System
User’s Guide
Problem Possible Causes and Solutions
The display shows any of the following:
err 2 , err 3 , err 4,orerr5
The display shows
6
The display shows
8
After starting the Auto Program the display shows
uuAit
n’t go into
err 1 ,
PrEP
err
err
and does
.
Controller problem. The error messages signify the following problems with the controller.
Err 1
- a RAM error
Err 2
- a NVRAM error
Err 3
- a Structure error
Err 4
- an ADC setup error
Err 5
- an ADC ready error
Initialize the system by performing the Factory Reset Sequence describe above.
Defective control sensor. The control sensor may be shorted, open or other wise damaged. Disconnect the power cord from the instrument. Looking at the front of the instrument (display), remove the front panel and right side panel. Locate the analog control board (on the block side of the electronics panel). The Molex connector at the top of the board connects the control probe to the controller (see wiring diagram, connector J4). Remove the connector and mea sure the resistance of the control sensor. It should be approximately 110 ohms at room temperature. If it measures close to this, make sure that it is securely re-connected to the controller, replace the panels and reapply power to the in strument. If the error message returns, there is a problem with the controller.
Improper freezing. The Auto Program of the instrument is designed to monitor power levels and scan rates during the freezing cycle. If an unacceptable level or rate is detected, the unit will show To prevent any damage from occurring to the cell, remove it immediately and freeze it manually. To freeze manually, place the lower 1/3 of the cell in an ice bath with the top 2/3 exposed to the ambient air for approximately one hour un­til the cell freezes. The instrument may be reset by pressing any of the front panel buttons, but if the error message continues, the unacceptable freeze con­ditions are still present. Power the unit OFF and wait a few minutes. Then power the unit ON and reset the instrument by pressing any of the front panel buttons to see if the problem has been eliminated.
Ambient temperature may be changing too rapidly. Rapid changes in the ambient temperature of more than ±1.0°C reduce the stability of the instrument.
-
The Auto Program monitors the stability of the block at 29.27°C and if it is too unstable the melt does not proceed. Check the ambient temperature around the instrument for rapid changes. Avoid drafty areas.
err 8
and the beeper alarm will sound.
-
-
-
68
11.2 Comments
11.2.1 EMC Directive
Hart Scientifics’ equipment has been tested to meet the European Electromag netic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specific standards to which the unit was tested.
11.2.2 Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (73/23/EEC), Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and the IEC 1010-2-010 (EN 61010-2-010) standards.
-
Loading...