Table 1 lists the International Electrical Symbols. Some or all of these symbols
may be used on the instrument or in this manual.
Tabl e1 International Electrical Symbols
SymbolDescription
AC
AC-DC
Battery
CE
1 Before You Start
Symbols Used
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface
Read the User’s Manual
Off
On
1
Page 8
9230 Maintenance System
User’s Guide
Canadian Standards Association
C-TIC Australian EMC Mark
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re
fers to the level of Impulse Withstand Voltage protection provided. Equipment of
OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from
the fixed installation. Examples include household, office, and laboratory appliances.
The European Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) mark.
1.2Safety Information
Use of this instrument other than as specified in this manual may impair the
protection provided by the instrument.
The following definitions apply to the terms “Warning” and “Caution”.
• “WARNING” identifies conditions and actions that may pose hazards to
the user.
• “CAUTION” identifies conditions and actions that may damage the instrument being used.
1.2.1
WARNINGS
To avoid possible electric shock or personal injury, follow these guidelines.
BURN HAZARD – DO NOT leave the melt heater outside of the unit and en
ergized. The melt heater can reach approximately 110°C if not properly in
serted into the unit. Place the melt heater in the storage well located at the back
of the instrument when not in use.
-
-
-
DO NOT operate this unit without a properly grounded, properly polarized
power cord.
Use only grounded AC mains supply of the appropriate voltage to power the in
-
strument. Refer to Section 3.1, Specifications, for power details.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE IN
-
JURY OR DEATH may result if personnel fail to observe safety precautions.
Before working inside the equipment, turn the power off and disconnect the
power cord.
Always replace the fuse with one of the same rating, voltage, and type.
2
Page 9
1 Before You Start
Safety Information
Overhead clearance is required. DO NOT place this instrument under a cabinet
or other structure. Additionally, at least 6 inches of space between the instru
-
ment and nearby objects is necessary for sufficient air circulation.
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the user’s
guide.
DO NOT turn the unit upside down with the cell in place; the cell will fall out
of the unit.
DO NOT operate near flammable materials.
DO use a ground fault interrupt device.
Before initial use, after transport, and anytime the dry-well has not been ener
gized for more than 10 days, the calibrator must be energized for a dry-out pe
-
riod of 1 to 2 hours before it can be assumed to meet all of the safety
requirements of the IEC1010-1.
Follow all safety guidelines listed in the user’s manual.
Calibration Equipment should only be used by Trained Personnel.
When the melt heater switch is in the MANUAL ON position, the melt heater is
continuously energized and can get hot!
1.2.2
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
DO NOT use fluids to clean out the well.
DO NOT slam the probe stems into the entrant well of the gallium cell. The
probe could be damaged.
DO use the cell removal tool to remove the cell from the well of the instrument.
Grasp the cell with a hand as soon as the cell is removed far enough from the
well.
DO NOT use the removal tool to carry the cell. The removal tool is not de
signed for transporting the cell.
Never introduce foreign material into the well. Fluids, etc. can leak into the cal
ibrator causing damage.
DO use silicon oil or other appropriate heat transfer medium in the entrant well
of the gallium cell. Be sure the transfer medium does not leak out into the well
of the instrument.
DO NOT change the value of the calibration constant from the factory set
value. The correct setting of this parameter is important to the safety and proper
operation of the calibrator.
-
-
3
Page 10
9230 Maintenance System
User’s Guide
The Auto Program configuration parameters should not be adjusted under nor
mal use. These parameters directly affect the melt plateau duration time and
proper freezing of the gallium cell.
Insert only frozen gallium cells into the 9230. DO NOT insert a melted cell
into the unit and then initiate the Auto Program. An improper freeze may occur
resulting in damage to the cell.
1.3Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
-
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
4
Page 11
1 Before You Start
Authorized Service Centers
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following
information available:
• Model Number
• Serial Number
• Voltage
• Complete description of the problem
5
Page 12
2Introduction
The 9230 is a specialized instrument for the realization of the gallium melting
point (29.7646°C). The instrument works with Hart’s 5943 stainless steel Ga
cell to provide melting plateaus that can last a week (or longer). This instru
ment is intended for Calibration Laboratory use and not for field applications.
Internal programming of the micro-processor controller provides prepro
grammed scan rates, set-points for each step in the process, dwell timing, and
indication that the next step is ready. The Model 9230 is available in 100-230
VAC (±10%), 50-60 Hz.
Built in programmable features include:
Temperature scan rate control
•
Fixed-point programming with default settings for the gallium cell
•
Eight set-point memories
•
Adjustable readout in °C or °F
•
The temperature is accurately controlled by Hart’s special proportional-integral-derivative (PID)/digital controller. The controller uses a platinum RTD as a
sensor and controls the well temperature with thermoelectric Peltier devices.
After properly inserting a frozen Ga cell and powering the unit on, the Model
9230 controller sets the block temperature to 25°C and the LED front panel display shows auto. The user then makes sure that the inner melt heater is attached to the back of the unit, the melt heater switch is in the Auto position,
and inserts the melt heater into the entrant well of the gallium cell. When the
user selects auto by pressing the SET button, the Model 9230 Auto Program
begins a pre-determined program to initiate and maintain a gallium melt plateau. While in the Auto Program mode, the display shows the current process
in the gallium melt cycle. During this mode the user may not change those pa
rameters, which could negatively impact the gallium melt. At the end of the
plateau the unit properly refreezes the cell and returns the temperature to 25°C
(standby) to await the next user initiated melt.
While in the manual mode of operation, the LED front display panel continu
ously shows the current well temperature. The temperature may be easily set
with the control buttons to any desired temperature within the specified range.
Certain parameters of the Auto Program itself may be adjusted in the manual
mode of operation.
The Model 9230 was designed for high accuracy fixed-point calibration meth
ods and for ease of operation. Through proper use, the instrument will continu
ously provide accurate calibration of temperature sensors and devices. The user
should be familiar with the safety guidelines and operating procedures of the
instrument as described in this User’s Guide.
2 Introduction
-
-
-
-
-
-
7
Page 13
3 Specifications and Environmental Conditions
3Specifications and Environmental
Conditions
3.1Specifications
Operating Range15°C to 35°C (59°F to 95°F)
Gallium Cell Uncertainty±0.08mK, typical
Gallium Melting Plateau
Duration
Display Accuracy±0.05°C at 29.76°C
Stability±0.02°C
Display Resolution0.01°C or °F (0.001°C or °F in program mode)
Display Scale°C or °F, switchable
Preheat Wells2 wells (8.2 mm diameter)
Immersion Depth152 mm (6 inches) in gallium cell
Vertical Gradient< 0.03°C over 6 inches during cell maintenance
Fault ProtectionFirmware protection
CommunicationsRS-232 included
Ambient Temperature18°C to 28°C (65°F to 82°F)
Power Requirements115 VAC (±10%), 60 Hz, 1.0 A or 230 VAC (±10%), 50 Hz , 0.65 A,
Exterior Dimensions8.75” W x 10.25” D x 19.25” H (22 x 26 x 49 cm)
Weight18 lb (8.2 kg) without cell
5days,typical
175 W maximum
Specifications
3.2Environmental Conditions
Although the instrument has been designed for optimum durability and trou
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean
ing recommendations can be found in the Maintenance section of this manual.
The instrument operates safely under the following conditions:
•
temperature range: 18–28°C (65–82°F)
•
ambient relative humidity: 15-50%
•
pressure: 75kPa – 106kPa
•
mains voltage within ±10% of nominal
•
vibrations in the calibration environment should be minimized
•
altitudes less than 2,000 meters
-
-
9
Page 14
9230 Maintenance System
User’s Guide
3.3Warranty
Hart Scientific, Inc. (Hart) warrants this product to be free from defects in ma
terial and workmanship under normal use and service for a period as stated in
the current product catalog from the date of shipment. This warranty extends
only to the original purchaser and shall not apply to any product which, in
Hart’s sole opinion, has been subject to misuse, alteration, abuse or abnormal
conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc
-
tions on appropriate Hart products. It is not warranted to be error free.
Hart’s obligation under this warranty is limited to repair or replacement of a
product which is returned to Hart within the warranty period and is determined,
upon examination by Hart, to be defective. If Hart determines that the defect or
malfunction has been caused by misuse, alteration, abuse or abnormal condi
tions or operation or handling, Hart will repair the product and bill the pur
-
-
chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling
or writing a Hart Scientific Authorized Service Center for authorization. Hart
assumes NO risk for in-transit damage.
For service or assistance, please contact a Hart Scientific Authorized Service
Center (see Section 1.3).
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY
PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR
ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAM
-
AGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
-
10
Page 15
4 Quick Start in Auto Program Mode
4Quick Start in Auto Program Mode
Unpacking
The user should read this user’s guide and be familiar with the functional per
formance of the Model 9230 before using the instrument to achieve a gallium
melt plateau.
4.1Unpacking
Unpack the instrument carefully and inspect it for any damage that may have
occurred during shipment. If there is shipping damage, notify the carrier imme
diately. Verify that the following components are present:
Model 9230 Gallium Maintenance Apparatus
•
Melt heater
•
Power cord
•
User’s Guide
•
• Top Insulation Plugs
• Cell Removal Tool
If you are missing any item, please call a Hart Scientific Authorized Service
Center.
4.2Set Up
Place the instrument on a flat surface with at least 6 inches of free space around
and 18 inches above the instrument. Avoid drafty areas. Install the power cord
into the power entry module on the back of the instrument. Plug the power cord
into a grounded mains outlet. Verify that the nominal voltage corresponds to
that indicated on the back of the instrument.
Using the cell removal tool, carefully insert the frozen gallium cell, prepared as
described in Section 9, Melting Point of Gallium Realization, into the well.
(DO NOT drop it into the well.) The well must be clear of any foreign objects,
dirt and grit before the cell is inserted.
Firmly place the two top insulation plugs into the well such that they rest on top
of the cell. Make sure that the inner melt heater is connected to the inner melt
heater connector on the back of the instrument and the melt heater switch is in
the auto position. Carefully insert the melt heater through the hole in the top
insulation plugs and into the entrant well of the frozen gallium cell. Make sure
the inner melt heater extends to the bottom of the gallium cell entrant well.
-
-
4.3Power
Plug the instrument power cord into a mains outlet of the proper voltage, fre
quency and current capability. Refer to Section 3.1, Specifications, for power
-
11
Page 16
9230 Maintenance System
User’s Guide
details. Turn the instrument ON using the power switch on the back of the unit.
The instrument turns on and begins to approach 25°C. The front panel LED
display shows auto.
4.4Auto Program
The Model 9230 is designed to automate the gallium cell melt plateau. This is
accomplished through the use of the Auto Program; a microprocessor-con
trolled program that determines set temperatures, scan rates, duration times,
Peltier module configurations, and melt heater controls. By controlling all of
these parameters in the proper order, a temperature profile can be maintained in
the block assembly that assures a uniform melt, maintains the melt over long
periods of time, properly refreezes the cell and keeps the cell ready for the next
melt cycle. The Auto Program is illustrated in Figure 1.
4.4.1Starting the Auto Program
After the frozen gallium cell is properly installed in the instrument (See Section
9, Melting Point of Gallium Realization, for more details) and the unit is powered ON, the front panel LED display shows Auto. The instrument goes to
25°C and remains there until the SET button on the front panel is pressed
which starts the Auto Program. The display changes to uuAit and the Auto
Program begins. The unit resets to 29.27°C at a scan rate of 0.2°C/minute in accordance with the Auto Program (see Figure 1).
-
12
4.4.2Generating the Melt Plateau
Once the unit has stabilized at 29.27°C, it waits for a minimum of 30 minutes
to ensure that the gallium cell has equalized in temperature. At the end of the
equalization time, the Auto Program begins to initiate the melt. The display
changes to PrEP as the set temperature of the block is changed to 30.77°C.
This initiates a melt on the outer surfaces of the gallium within the cell. A few
minutes into the program, the unit beeps four times and the display changes to
a message that flashes between prep and htr on to indicate that the inner
melt heater has been turned ON. The inner melt heater remains ON for a few
minutes and then shuts off as the display changes back to prep and the unit
beeps eight times. The user should remove the inner melt heater at this time.
Note: No damage occurs to the gallium cell if the heater is not removed, how
ever, a shortening of the melt duration may result.
After the inner melt heater is powered OFF, the instrument remains at 30.77°C
for a few more minutes to complete the outer melt of the gallium cell. The unit
then beeps 16 times, the display changes from prep to nnaint and the block
temperature resets to 29.86°C to maintain the melting gallium cell.
The user may then insert a preheated monitor thermometer to monitor the tem
perature of the gallium cell. Due to the heat applied to the cell by the inner melt
heater, the temperature of the cell’s entrant well is higher than the gallium melt
ing point (29.7646°C) for a short period of time. Generally, after 30 to 60 min
utes, the temperature of the cell’s entrant well is sufficiently stable to begin
-
-
-
-
Page 17
4 Quick Start in Auto Program Mode
35
4
0
3
0.77
2
9.86
2
9.27
20
10
0
4 beeps
melthtr=0
(ON)
8min
PREP1
6min
PREP3
User T
est
T
ime Duration
(galliu
mc
e
ll at 29.7646
°C )
up to
6hrs
2.5 hrs
comple
tel
y
freeze
gallium
cell
SR = 0.2
°
C/m
in
T
set = 29.
86°C
25°C
16 beeps
25.00
T (°C) - block sensor
User in
ser
ts
frozen
gal
li
um
cell and p
owers
on u
nit
T
set=25°C
S
R=OFF
frz
mlt=0
(MELT/M
A
INT
)
melthtr=
1
(OFF
)
Unit st
ays
a
t 25°C until U
ser
sele
cts AU
TO or MAN
using
UP/D
Nan
d
presses SE
T
Un
it goes
into manu
al mode
if
MAN is se
lected
User s
elec
t
s
A
UTO
T
set=29.2
7°C
SR
=0.2°C
/min
fr
zmlt=0
(MELT/M
A
IN
T)
melthtr
=1
(OF
F)
Unit re
ma
in
sat
29.27°
Cfo
r
30 min
to allow
ga
lliu
m
cell to
stab
ilize
30 min
minimum
Unit
begins
m
elt
Tset=30.7
7°C
SR=0.2°C
/min
frzmlt=0
(ME
LT/MA
INT)
me
lthtr=1
(OFF)
4min
PREP
2
DISPL
AY
AU
TO
DISPL
AY
WA
IT
Display
PREP
Display
PRE
P/
HTR ON
Di
splay
PRE
P
8 beep
s
melthtr=1
(OFF)
remove
m
elt heate
r
DISPL
AY
MA
INT
Un
it
be
gins
freeze
H
Tse
t=35.0
0°C
SR
=OFF
frzm
lt=0
(MELT/MA
INT)
melthtr=1
(OFF)
Un
it
be
gins
freeze
C
Ts
et=0.00
°C
SR=
0.5°C
/min
frzm
lt=1
(FREEZ
E
)
melthtr=1
(OFF
)
Unit
returns
to
25°C
Tset=
25°C
SR=O
FF
frzmlt=0
(M
ELT/MA
INT)
m
elthtr=1
(OFF)
DISPL
AY
FREEZEH
(OPTIONA
L)
DISPL
AY
FREE
ZC
Unit rem
a
ins
at 25°C
unti
l User
selects
AUTO
or M
AN
(be
gin aga
in)
begin
again
DISPL
AY
AUTO
If in
MAN
mode, User
can re
turn
to AU
TO m
ode by adva
ncin
g
t
he progra
m(SET+
DN)
25°C
29
.27°C
30.77°C
29.86°C
Time - as indicated above
User c
an
ab
ort the
Auto
Prog
ram and
return t
oS
ta
ndby
Mode
(25°
C
)during
the "WAI
T
"tim
eby
adva
ncing
the
program (
SET+
DN)
Defaul
t-U
s
er initiates F
reez
e
by
advancing
prog
ram (SET+D
N)
Option - U
ser pr
e-sets MAIN
T
Duration
T
ime
after which
program in
itiates
Freeze
Default -
G
od
irectly
to FreezC
(no Fr
eezH)
O
ption - U
ser pre-
sets
FreezH Te
mper
ature
a
nd Durat
ion Time
to melt ga
llium ce
ll
be
fore Fre
eze
up to
36°C
Auto Program
Figure 1 Auto Program Diagram
13
Page 18
9230 Maintenance System
User’s Guide
calibrations of thermometer sensors. The duration of the melt plateau varies de
pending on usage, but lengths of 5 days or longer are achievable.
4.4.3Refreezing the Gallium Cell
At the end of the melt plateau (or when the user wishes to end the plateau), the
Auto Program is advanced by pressing the SET and DOWN buttons (press the
SET button and keep it pressed and then press the DOWN button and release
both) on the front panel (program advance). The dislay will flash “AdU”and
display “nnAint.” Use the UP and DOWN buttons to select the program state
“FrEE2C” then press the ENTER button. See Section 6.21.2, Maintain
Time-out, for optional user defined auto time-out. Gallium, like water, expands
in volume when freezing. Therefore, the cell must be frozen from the bottom
up or there is a potential for damage to the cell. The Auto Program process en
sures that the cell is properly refrozen.
The Auto Program initiates the freeze and the display shows free2E at this
point. The Peltier devices are reconfigured and controlled such that a large tem
perature gradient is generated between the top and bottom of the block. The instrument slowly scans to 0°C to maintain the gradient over several hours. See
Section 6.21.4, Freeze Hot Temperature, for optional user defined freeze hot
temperature.
4.4.4Return to Standby
Upon completion of the freeze, the Auto Program returns the unit to 25°C, the
front panel LED display shows auto and the unit remains there until the user
initiates the Auto Program again by pressing the SET button.
-
-
-
14
Page 19
5Parts and Controls
115V
Front
Foot
Fan
1
2
The user should become familiar with the Model 9230 parts. Successful use of
the instrument is dependent upon knowledge of important components and
theirproperuse.
5.1Bottom Panel
On the bottom panel are the fan and feet. See Figure 2.
1. The cooling fan inlet is at the bottom of the unit. The cooling air circulating
through the instrument keeps the electronics and the chassis cool. Keep the area
immediately around the instrument clear to allow adequate ventilation.
2. Three feet support the chassis permitting air space for the fan.
5 Parts and Controls
Bottom Panel
Figure 2 Bottom Panel
15
Page 20
9230 Maintenance System
SET
ADV
°C/°F
DOWN
UPEXIT
0.020 C
1
2
9230
GALLIUM MAINTENANCE APPARATUS 15°C to 30°C
User’s Guide
5.2Front Panel
The front panel contains the digital display and the controller keypad. See Fig
-
ure 3.
1. The digital display is an important part of the temperature controller because
it not only displays set and actual temperatures but also displays various instru
ment functions, settings, and constants. The display shows temperatures in
units according to the selected scale °C or °F. When in the Auto Program mode,
the display indicates the current status of the unit in the gallium melt process.
2. The four button controller keypad allows easy setting of the set-point tem
perature. The control buttons (SET, DOWN, UP, and EXIT) are used to set the
temperature set-point, access and set other operating parameters, and access
and set calibration parameters.
Setting the control temperature is done directly in degrees of the current scale
and can be set to 0.01 of a degree Celsius or Fahrenheit.
During the Manual mode of operation, the functions of the buttons are as
follows:
SET – Used to display the next parameter in the menu and to store parameters
to the displayed value.
DOWN – Used to decrement the displayed value of parameters.
UP –Usedtoincrementthedisplayedvalue.
EXIT – Used to exit a function and to skip to the next function. Any changes
made to the displayed value are ignored. Holding the EXIT button for about ½
second returns control to the main display.
-
16
Figure 3 Front Panel
Page 21
5.3Top Panel
Inner Melt Heater Storage Well
Cooling Air Vents
Preheat Wells
Constant Temperature
Block Assembly with
Gallium Cell
2
4
5
3
Top Insulation Plugs
1
SET
ADV
°C/°F
DOWN
UP EXIT
9230
GALLIUM MAINTENANCE APPARATUS 15°C to 30°C
Auto
5 Parts and Controls
Top Panel
The primary feature of the top of the unit is to provide access to the tempera
ture controlled block. The top panel consists of the constant temperature block
assembly; two preheat wells, top insulation plug, cooling air vents, the melt
heater and melt heater storage well. See Figure 4.
1. The constant temperature block assembly is where the gallium cell is in
serted. Peltier devices surround the cell in order to provide a uniform tempera
ture. These devices can be configured by the controller to provide a large
vertical gradient during the freezing of the gallium cell, or a very uniform tem
perature when maintaining a melting cell.
-
-
Figure 4 Top Panel
17
Page 22
9230 Maintenance System
User’s Guide
2. The inner melt heater is a low-power heater that creates an inner liquid layer
next to the reentrant tube of the gallium cell during the melting process. This
inner melt heater is controlled automatically by the microprocessor when using
the Auto Program mode, or may be used manually with the switch in back.
When not in use, the inner melt heater is stored in the well at the back of the
unit to prevent it from causing any damage.
3.The thermometer preheat wells are located at the back of the block access
well. Thermometers are preheated in these wells prior to insertion into the cell
in order to conserve its latent energy.
4. The cooling air vents in the top of the unit permit heated air to exit the unit.
5. The top insulation plugs fit into the well over the gallium cell. See Section
8.5, Inserting the Top Insulation Plugs.
5.4Rear Panel
The rear panel consists of the power connection, power switch, melt heater
switch, inner melt heater connection, and serial port. See Figure 5.
1. The power connection is where the power cord plugs into the back of the
instrument.
2. Set the power switch “ON” to operate the instrument and “OFF” to disconnect the power.
3. The melt heater plugs into the rear of the instrument into the inner melt
heater connector provided. Be sure it is plugged in during operation.
4. The melt heater switch can be set to MANUAL ON position or AUTO position. The instrument’s microprocessor has control when the switch is in the
AUTO position.
18
Warning: When the melt heater switch is in the MANUAL ON position,
the melt heater is continuously energized and can get hot!
5. The serial port is a DB-9 male connector for interfacing the instrument to a
computer or terminal RS-232 communications.
5.5Thermal Block Assembly
The thermal block assembly holds the gallium cell and heats or cools to main
tain the proper temperature. The thermal block assembly consists of the thermal
block, thermoelectric Peltier devices, the gallium cell, and the top insulation
plugs.
1. The thermal block is specifically designed to contain the gallium cell. The
thermal block also has 2 preheat wells for the test thermometers.
-
Page 23
5 Parts and Controls
RS-232
115V
RS-232
Power Switch
Powe r
Connection
RS-232
Interface
MANUAL ON
AUTO
INNER MELT
HEATER
Inner
Melt
Heater
Connector
Melt
Heater
Switch
Thermal Block Assembly
Figure 5 Rear Panel
19
Page 24
9230 Maintenance System
User’s Guide
2. The block assembly is heated or cooled as necessary to maintain the proper
temperature with thermoelectric Peltier devices.
3. The gallium cell is placed into the block. Use the removal tool to carefully
insert or remove the cell.
WARNING: DO NOT use the removal tool to carry the cell.
4. The top insulation plugs are placed over the gallium cell to help maintain the
proper temperature. See Section 8.5, Inserting the Top Insulation Plugs.
20
Page 25
6Controller Operation
6 Controller Operation
Standby
This section discusses in detail how to operate the instrument temperature con
troller using the front control panel. By using the front panel key-switches and
LED display, the user may initiate the Auto Program mode; a microproces
sor-controlled program that determines set temperatures, scan rates, duration
times, Peltier module configurations, and melt heater controls. By controlling
all of these parameters in the proper order, a temperature profile can be main
tained in the block assembly that assures a uniform melt, maintains the melt
over long periods of time, properly refreezes the cell and keeps the cell ready
for the next melt cycle. After selecting the Manual mode of operation, the user
may monitor the well temperature, adjust the set-point temperature in degrees
C or F, monitor the Peltier device output power, adjust the controller propor
tional band, and program the operating parameters, program parameters, serial
interface configuration, and the controller calibration parameters. Operation of
the functions and parameters is shown in the flowchart in Figure 6 on page 22.
This chart may be copied for reference.
In the following discussion a button with the word SET, UP, DOWN, or EXIT
inside indicates the panel button while the dotted box indicates the display
reading. Explanation of the button or display reading are to the right of each
button or display value.
When two buttons are shown to initiate an action, for example,
S+D
6.1Standby
After the instrument is powered ON or at the end of a gallium cell refreeze pro
cess, the unit automatically is set to 25°C and the digital LED display reads:
-
-
-
-
Press and hold the first button (e.g., SET) and then press the
second button (e.g., DOWN) and release both
-
AutoReady to Start Auto Program
The unit is in a Standby condition and remains at 25°C indefinitely until the
user initiates a change by pressing a button sequence.
6.2Starting the Auto Program
The user begins the Auto Program by pressing the SET button.
S
Start the Auto Program
uuAltWAIT - Auto Program started
21
Page 26
9230 Maintenance System
Powe r ON
Display
“Auto”
T =25°C
SET
Select “nnAn”
Select Set-Point
UP/DN
SET
SET
Adjust Set-Point
Adjust Units °C/°F
Adjust Scan ON/OFF
Display
Temperature
Adjust Scan Rate
UP/DN
UP/DN
UP/DN
UP/DN
Toggles °C/°F
Display Set-Point
Resistance
SET
Adjust Proportional
Band
Configuration
Menu
Display Power
Display
“uuAIt”
SET
Display “PREP”
Display “nnaint”
Display “FrEE2H”
Program Advance
SET + DN
PRE-SET
TIMEOUT
Display “frEE2C”
Optional
SET+EXIT
SET + UP
SET
SET
EXIT
Adjust
Ready Temperature
SET
Program 1
Parameters
Menu
UP
DN
Adjust Prep Melt
Temperature
Adjust Prep
Scan Rate
Adjust
Beeper
Adjust
Prep 1Time
Adjust
Prep 2 Time
Adjust
Prep 3 Time
Adjust Maintain
Temperature
SET
Program 2
Parameters
Menu
UP
DN
Adjust Maintain
Time-Out
Adjust Maintain
Time-Out Time
Adjust Freeze
Hot Temperature
Adjust Freeze
Hot Time
Adjust Freeze
Cold Temperature
Adjust Freeze
Cold Scan Rate
Adjust Freeze
Melt State
SET
Program 3
Parameters
Menu
UP
DN
Adjust
Baud Rate
SET
Serial
Interface
Menu
UP
DN
Adjust
Sample Period
Adjust
Duplex Mode
Adjust
Linefeed
Adjust
R0
SET
Cal
Menu
Adjust Freeze
Cold Time
X5
EXITEXITEXITEXITEXIT
Menu Legend:
While in MANUAL Mode:
Press the SET button to step through the menu and to access the parameter. Use the UP and
DOWN buttons to adjust the parameter value. Press the SET button again to accept and store
the new value, and to advance to the next parameter.
Press the EXIT button briefly to skip a parameter without storing the parameter value.
Hold the EXIT button to exit the menu and to display the temperature.
Press the SET and DOWN buttons at the same time to advance the program to AUTO mode.
While in AUTO Mode:
Press the EXIT and UP buttons to momentarily display the block temperature.
The AUTO Mode parameters may be viewed, but no changes are allowed except for the
beeper and serial interface parameters.
•
•
•
•
•
•
Manual Mode
SET + DOWN
SET + DOWN
Auto Mode
UP
SET
User’s Guide
(see Figure
(see Figure
Figure 6 Controller Flowchart
22
Page 27
Plateau
Auto Program – Initiating and Maintaining the Gallium Melt
6 Controller Operation
At this point the Auto Program takes control of the instrument and initiates a
melt in the gallium cell in accordance with Figure 1 on page 13 and Figure 6 on
page 22.
At any time the user may press the program advance button sequence to access
the program advance mode.
S+D
Access program advance
toAutoExit from auto mode
From here, the set button can be pressed to go back to the Auto state or press
the up or down buttons to choose a different program state. Once the desired
program state is chosen press SET to accept the chosen state.
S
The chosen program state will now start.
Accept the program state
6.3Auto Program – Initiating and Maintaining the
Gallium Melt Plateau
Approximately 55-65 minutes after starting the Auto Program, the gallium cell
reaches 29.27°C and stabilizes enough to start the melt plateau. The display
shows:
prepPrEP – gallium cell melt initiated
The Auto Program takes the block assembly to 30.77°C, which begins melting
the outer surface of the gallium inside the cell.
After approximately 8 minutes, the Auto Program energizes the inner melt
heater inside the gallium cell entrant well. The instrument sounds an audible
beep 4 times and the display message alternates as indicated:
PrEPAlternating message between PrEP and Htr on
Htr on
The inner melt heater is energized for approximately four minutes after which
the Auto Program turns it OFF and sounds an audible beep eight times. At this
point the display indicates:
PrEPInner melt heater OFF
The user should remove the inner melt heater at this time. Note: No damage
will occur to the gallium cell if the heater is not removed, however, a shorten
-
23
Page 28
9230 Maintenance System
User’s Guide
ing of the melt plateau may result. The inner melt heater may be stored in the
melt heater well at the back of the unit.
After the inner melt heater has been powered OFF, the instrument remains at
30.77°C for a few more minutes to complete the outer melt of the gallium cell.
The unit then sounds an audible beep 16 times, the display changes from PrEP
to nnAint and the block temperature is reset to 29.86°C to maintain the melt
ing gallium cell.
The user may then insert a preheated monitor thermometer to monitor the tem
perature of the gallium cell. Due to the heat applied to the cell by the inner melt
heater, the temperature of the cell’s entrant well is higher than the gallium melt
ing point (29.7646°C) for a short period of time. Generally, after 30 to 60 min
utes, the temperature of the cell’s entrant well is sufficiently stable to begin
calibrations of thermometer sensors. The duration of the melt plateau varies de
pending on usage, but lengths of 5 days or longer are achievable.
6.4Auto Program – Ending the Gallium Melt Plateau
and Refreezing the Cell
At the end of the melt plateau (or when the user wishes to end the plateau), the
Auto Program is advanced by pressing the program advance button sequence.
S+D
Display will flash advance for 1 second then show the current program state.
Gallium, like water, expands in volume when freezing. Therefore, the cell must
be frozen from the bottom up or there is a potential for damage to the cell.
The Peltier devices are reconfigured and controlled such that a large tempera
ture gradient is generated between the top and bottom of the block. The instru
ment slowly scans from 29.86°C to 0°C to maintain the gradient over several
hours thus ensuring the cell freezes properly, beginning at the bottom and con
tinuing up to the top of the cell.
-
-
-
Page 29
See Section 6.21.2, Maintain Time-out, for optional user defined auto time-out
(an optional method of ending the melt plateau and initiating the refreeze pro
cess). See Section 6.21.4, Freeze Hot Temperature, for optional user defined
freeze hot temperature (a method to totally melt the cell before freezing it).
6.5Return to Standby
Upon completion of the freeze, the Auto Program returns the unit to 25°C, the
digital LED display shows:
AutoInstrument returning to 25°C
6 Controller Operation
Return to Standby
-
After reaching 25°C, the unit remains there until the user initiates the Auto Pro
gram again by pressing the SET button or goes into the Manual mode of
operation.
6.6Program Advance
From the manual mode, the program advance function allows the user to return
to the standby condition.
S+D
AdU“” flashes and then displays program mode
AutoProgram mode
Press the UP or DOWN button to toggle between the Auto Program and Man
ual mode.
S
At any time during the auto program the user may press the program advance
button sequence to access the program advance function.
S+D
The current program state will be displayed.
Access program advance
Accepts the new operation
Access program advance
-
-
toAutoReturn to main menu
PrEPInitiate cell melt with melt heater
nnAintMaintain melt plateau
25
Page 30
9230 Maintenance System
User’s Guide
FrEE2HInitiate total melt
or
FrEE2CStart freezing the cell
From the current program state, the up or down buttons can be pressed to
choose a different program state.
UorD
Once the desired program state is chosen, press “SET” to accept and start that
program.
S
To get to the wait state the user must select the Auto state then press “set” to
start the program in the wait state.
Choose the desired program state
Accept new program state and start the program
6.7Manual Mode
6.8Starting the Manual Mode
After the instrument is powered ON or at the end of a gallium cell refreeze process, the unit automatically is set to 25°C and the digital LED display reads:
AutoReady to start Auto Program
The unit is in a Standby condition and remains at 25°C indefinitely until the
user initiates a change by pressing a button sequence. To begin the Manual
mode of operation, the user presses the UP or DOWN buttons to change the
display reading as follows:
DorU
Toggle display between Auto and Manual
26
nnAnReady to go into Manual mode
By pressing the SET button while this is displayed, the instrument goes into the
Manual mode and begins to display the well temperature:
S
25.00CWell temperature in degrees Celsius
Accept Manual mode
Page 31
Once in the Manual mode, the temperature display function may be accessed
from any other function by pressing and holding the EXIT button a few
seconds.
6.9Temperature Set-point
6 Controller Operation
Temperature Set-point
The temperature set-point can be set to any value within the range and resolu
tion as given in the specifications. Be careful not to exceed the safe temperature
limit of any device inserted into the well.
Setting the temperature involves selecting one of the eight (8) set-points in
memory and then adjusting the set-point value.
6.9.1Programmable Set-points
The controller stores eight (8) set-point temperatures in memory. The set-points
can be quickly recalled to conveniently set the instrument to a previously pro
grammed temperature set-point.
To set the temperature, first select the set-point memory. This function is accessed from the temperature display function by pressing the SET button. The
number of the set-point memory currently being used is shown at the left on the
display followed by the current set-point value.
25.00CWell temperature in degrees Celsius
S
Access set-point memory
1 25.Set-point memory 1, 25.0°C currently used
To change the set-point memory to another preset value press the UP or DOWN
button.
-
-
4 30.CNew set-point memory 4, 30.0°C
Press the SET button to display the new selection and access the set-point
value.
S
Display selected set-point memory
29.86New set-point value 29.86°C
6.9.2Set-point Value
If the set-point value is correct, hold the EXIT button to accept the new value
and resume displaying the well temperature. Otherwise, press the UP or
DOWN button to adjust the set-point value.
27
Page 32
9230 Maintenance System
User’s Guide
22.00New set-point value
When the desired set-point value is reached press the SET button to accept the
new value and to access the temperature scale units. If the EXIT button is
pressed, any changes made to the set-point are ignored.
S
Accept new set-point value
The default programmable set-points (factory set) are given in Table 2.
Tabl e 2 Default Programmable Set-points (Factory
Set)
Set-point Memory
Number
125.00
229.27
330.77
429.86
535.00
600.00
710.00
820.00
NOTE: Setpoint 1 will always be set back to 25.00°C af-
ter the power has been cycled off then on.
Default Set-point Temperature (°C)
6.10Temperature Scale Units
The temperature scale units of the controller can be set by the user to degrees
Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well tem
perature, set-point, scan rate, and proportional band.
Press the SET button after adjusting the set-point value to change display units.
-
28
Un= CScale units currently selected
Press the UP or DOWN button to change the units.
Un = FNew units selected
The units can also be changed quickly from the main display by pressing the
SET and UP buttons.
Page 33
6.11Scan
The scan rate can be set and enabled so that when the set-point is changed the
instrument heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. With the scan disabled the instrument heats or cools at the
maximum possible rate.
6.11.1Scan Control
The scan is controlled with the scan ON/OFF function that appears in the main
menu after the set-point and unit’s functions.
ScAn=OFFScan function OFF
Press the UP or DOWN button to toggle the scan ON or OFF.
ScAn=OnScan function ON
Press the SET button to accept the present setting and continue.
6 Controller Operation
Scan
S
Accept scan setting
6.12Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set
from 0.1 to 5.0°C/minute (0.2 to 9.0°F/min). The maximum scan rate, however,
is actually limited by the natural heating or cooling rate of the instrument. This
is typically less than 2°C/minute.
The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees per minute, degrees C or F depending on the
selected units.
Sr= 0.2Scan rate in °C/min
Press the UP or DOWN button to change the scan rate.
Sr=2.0New scan rate
Press the SET button to accept the new scan rate and continue.
S
The display returns to the current well temperature.
Accept scan rate
29
Page 34
9230 Maintenance System
User’s Guide
6.13Temperature Scale Units
To toggle between °C and °F, press and hold the SET button and then press the
UP button when the temperature is displayed.
6.14Secondary Menu
Functions used less often are accessed within the secondary menu. Press and
hold the SET and EXIT buttons to access the secondary menu. The first func
tion in the secondary menu is the power display. (See Figure 6 on page 22.)
6.15Peltier Device Power
The temperature controller controls the temperature of the well by pulsing the
Peltier devices ON and OFF. The total power being applied to the Peltier de
vices is determined by the duty cycle or the ratio of device on time to the pulse
cycle time. By knowing the amount of heating the user can tell if the instrument
is heating up to the set-point, cooling down, or controlling at a constant temperature. Monitoring the percent power lets the user know the stability of the well
temperature. With good control stability the percent heating power should not
fluctuate more than ±1% within one minute.
The device power display is accessed in the secondary menu. Press and hold
the SET and EXIT buttons. The Peltier device power is displayed as a percentage of full power.
-
-
30
25.00Well temperature
S+E
Access Peltier device power in secondary menu
SECFlashes and then percent power flashes
Pct PoFlashes prior to displaying Peltier device power
6.0 PPeltier device power in percent
To exit out of the secondary menu press the EXIT button and hold for a brief
moment. To continue on to the proportional band setting function, press the
EXIT button momentarily or the SET button.
6.16Set-point Resistance
The set-point resistance is the resistance associated with the set-point tempera
ture. Allow the temperature to stabilize at the desired set-point before using the
-
Page 35
6 Controller Operation
Proportional Band
displayed resistance value. The set-point resistance function is available for di
agnostic purposes.
Press the SET and EXIT buttons to enter the secondary menu and show the
heater power. Then press the SET button twice to access the set-point
resistance.
S+E
Access Peltier device power in secondary menu
SECFlashes and then percent power flashes
Pct PoFlashes prior to displaying Peltier device power
6.0Peltier device power in percent
S
Access set-point resistance
SrFlashes “” (Set-point Resistance) and then displays the
setting
109.715Resistance in ohms
6.17Proportional Band
-
In a proportional controller such as this, the Peltier device output power is proportional to the well temperature over a limited range of temperatures around
the set-point. This range of temperature is called the proportional band. At the
bottom of the proportional band, the Peltier device output is ±100%. At the top
of the proportional band, the Peltier device output is 0. Thus as the temperature
rises the Peltier device power is reduced, which consequently tends to lower the
temperature back down. In this way the temperature is maintained at a fairly
constant value.
The temperature stability of the well and the response time depend on the width
of the proportional band. If the band is too wide, the well temperature deviates
excessively from the set-point due to varying external conditions. This devia
tion is because the power output changes very little with temperature and the
controller does not respond well to changing conditions or noise in the system.
If the proportional band is too narrow the temperature may swing back and
forth because the controller overreacts to temperature variations. For best con
trol stability the proportional band must be set for the optimum width.
The proportional band width is set at the factory to about 8.0°C. The propor
tional band width may be altered by the user to optimize the control character
istics for a particular application.
-
-
-
-
31
Page 36
9230 Maintenance System
User’s Guide
The proportional band width is easily adjusted from the front panel. The width
may be set to discrete values in degrees C or F depending on the selected units.
The proportional band adjustment can be accessed within the secondary menu.
Press the SET and EXIT buttons to enter the secondary menu and show the de
vice power. Then press the SET button twice to access the proportional band.
-
S+E
SECFlashes “” and then percent power flashes
Pct PoFlashes prior to displaying Peltier device power
6.0Peltier device power in percent
S
SrFlashes “” (Set-point Resistance) and then displays the
109.715Resistance in ohms
S
PropFlashes “” and then displays the setting
8.0Proportional band setting
To change the proportional band press the UP or DOWN button.
Access Peltier device power in secondary menu
Access set-point resistance
setting
Access proportional band
32
9.0New proportional band setting
To store the new setting press the SET button. Press the EXIT button to con
tinue without storing the new value.
S
Accept the new proportional band setting
6.18Controller Configuration, Operating, and
Calibration Parameters
The controller has a number of configuration, operating, and calibration param
eters that are programmable via the front panel. The first set of parameters is
associated with the Auto Program and is located under the configuration
submenus called prog 1, prog 2,andprog 3. The next set of parameters is
-
-
Page 37
Auto Program Configuration Parameters
6 Controller Operation
associated with the serial RS-232 interface and is located under the configura
tion submenu called serIAL. The last set of parameters is associated with the
calibration of the instrument and is located under the configuration submenu
called CAL.
6.19Auto Program Configuration Parameters
CAUTION: The Auto Program configuration parameters should not be
adjusted under normal use. These parameters directly affect the melt pla
teau duration time and proper freezing of the gallium cell.
The first set of parameters is associated with the Auto Program (see Figure 1,
page 13). The Auto Program is a microprocessor-controlled program that deter
mines set temperatures, scan rates, duration times, Peltier module configura
tions and melt heater controls. By controlling all of these parameters in the
proper order, a temperature profile can be maintained in the block assembly
that assures a uniform melt, maintains the melt over long periods of time, properly refreezes the cell and keeps the cell ready for the next melt cycle. The de-
-
-
-
-
33
Page 38
9230 Maintenance System
User’s Guide
fault programmable set-points (factory set) are given in Table 3, along with
minimum and maximum set values.
Tabl e 3 Auto Program Parameters (Prog 1, Prog 2, Prog 3)
Parameter NameParameter
Ready Temperature (WAIT)
Prep Melt Temperature (PREP)
Prep Scan Rate (PREP)
Beeper (PREP)
Prep 1 Time (PREP)
Prep 2 Time (PREP)
Prep 3 Time (PREP)
Maintain Temperature (MAINT)
Maintain Time-out (MAINT)
Maintain Time-out Time (MAINT)
Freeze Hot Temperature (FREEZH)
Freeze Hot Time (FREEZH)
Freeze Cold Temperature (FREEZC)
Freeze Cold Scan Rate (FREEZC)
Freeze Cold Time (FREEZC)
Freeze Melt State (Manual Mode Only)
*Factory Set
rdu t
PrEPt
prEPSr
bEEPEr
dur 1
dur 2
dur 3
nnAint
ti out
dur 4
FrEE2H
dur 5
FrEE2C
FrECSr
dur 6
frnELt
Low
Value
29.00029.27029.300°C1xx.xxx
30.00030.77035.000°C1xx.xxx
0.10.20.5°C/min1x.x
-ON--1ON/OFF
360480600sec1xxx
120240360sec1xxx
240360480sec1xxx
29.79029.86035.000°C2xx.xxx
-OFF--2ON/OFF
1720043,200min2xx,xxx
29.86029.86036.000°C2xx.xxx
00360min2xxx
-1.0000.00010.000°C2xx.xxx
0.40.50.6°C/min2x.x
120150180min2xxx
-MELT--3MELT/
Default
Value*
High
ValueUnitsProgSet Resolution
FREEZE
34
6.20Prog 1
The parameters within the Prog 1 submenu are accessed from the secondary
menu after the proportional band function by pressing the SET button. ConfIg
flashes and then the name of the first Auto Program parameter menu ProG 1
is displayed:
8.0Proportional band setting
S
ConFIGFlashes before entering configuration menus
Select the configuration menus
Page 39
ProG 1Prog 1 submenu
Pressing the SET button accesses the parameters within the Prog 1 submenu.
Pressing the UP button accesses Prog 2. The well temperature is displayed by
pressing the EXIT button.
6.20.1Ready Temperature
The Ready Temperature parameter is the temperature at which the instrument
stabilizes before initiating a melt in the gallium cell. The default value is
29.27°C and can be seen on Figure 1 as the temperature that the well goes to
when the display shows uuAit. The instrument holds the well at this tempera
ture for 30 minutes minimum to assure that the gallium cell has stabilized, be
fore allowing the melt to proceed.
When Prog 1 is displayed, press the SET button to access the Ready Tempera
ture parameter. The display flashes rdu t and then displays the current Ready
Temperature value:
ProG 1Prog 1 submenu
6 Controller Operation
Prog 1
-
-
-
S
Access the ready temperature parameter
rdu tFlashes before entering ready temperature
29.270Current ready temperature
Use the UP or DOWN buttons to adjust the value.
29.300New ready temperature
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new ready temperature value and continue to Prep Melt Tem
perature
6.20.2Prep Melt Temperature
The Prep Melt Temperature parameter is the temperature at which the instru
ment begins to melt the outer surfaces of the gallium within the gallium cell.
The default value is 30.77°C and can be seen on Figure 1 as the temperature
that the well goes to when the display shows Prep. The instrument holds the
well at this temperature for approximately 18 minutes (the sum of Prep 1, Prep
2 and Prep 3 duration times) to begin melting the outer surfaces of the gallium.
-
-
-
35
Page 40
9230 Maintenance System
User’s Guide
After accessing the Ready Temperature parameter, the display flashes PrEP t
and displays the current value of the Prep Melt Temperature.
PrEP tFlashes before entering prep melt temperature parameter
30.770Current prep melt temperature
Use UP or DOWN buttons to adjust the value.
35.000New prep melt temperature
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new Prep Melt Temperature value and continue to the Prep
Scan Rate
6.20.3Prep Scan Rate
The Prep Scan Rate parameter is the rate in °C/minute or °F/minute at which
the instrument scans from the Ready Temperature to the Prep Melt Temperature. The default value is 0.2°C/minute and can be seen on Figure 1, where the
unit begins the melt.
After accessing the Prep Melt Temperature parameter, the display flashes
PrEP sr and displays the current value of the Prep Scan Rate.
PrEP SrFlashes before entering prep scan rate parameter
Sr = 0.2Current prep scan rate
Use the UP or DOWN buttons to adjust the value.
Sr = 0.5New prep scan rate
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
-
-
36
S
6.20.4Beeper
The Beeper parameter is used to enable or disable the audible warnings that
may be generated during the initiation of the melt. The default value is ON.
Accept new prep scan rate value and continue to the inner melt
heater Beeper
Page 41
6 Controller Operation
As seen on Figure 1, the Auto Program sounds an audible beep 4 times when it
energizes the inner melt heater and the display shows an alternating message
from PrEP to Htr on. After approximately 4 minutes, the Auto Program
sounds an audible beep 8 times and turns the inner melt heater OFF (the display
returns PrEP). After approximately 6 more minutes, the Auto Program sounds
an audible beep 16 times and displays nnaint, an indication that the gallium
melt has begun.
Changing the Beeper parameter to OFF disables the beeper during this phase of
the Auto Program.
After accessing the Prep Scan Rate parameter, the display flashes beeper and
displays the current value of the Beeper parameter.
Prog 1
bEEPErFlashes before entering beeper parameter
OnCurrent beeper setting (enabled)
Use the UP or DOWN buttons to adjust the value.
OFFNew beeper setting
If no changes are desired, press the EXIT button to continue to the next parameter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new beeper setting and continue to the prep 1 time
6.20.5Prep 1 Time
After beginning the gallium melt, see Figure 1, the Auto Program waits a short
time before energizing the inner melt heater. The Prep 1 Time parameter speci
fies this time in seconds. The default value is 480 seconds (8 minutes). At the
end of this time, the Auto Program energizes the inner melt heater and sounds 4
audible beeps.
After accessing the Beeper parameter, the display flashes dur 1 and displays
the current value of the Prep 1 Time parameter.
dur 1Flashes before entering the prep 1 time parameter
SEC480Current prep 1 time value (8 minutes)
Use the UP or DOWN buttons to adjust the value.
SEC600New prep 1 time value (10 minutes)
-
37
Page 42
9230 Maintenance System
User’s Guide
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new prep 1 time value and continue to prep 2 time
6.20.6Prep 2 Time
After energizing the inner melt heater during the initialization of the gallium
melt, see Figure 1, the Auto Program waits a short time before turning OFF the
inner melt heater. The Prep 2 Time parameter specifies this time in seconds.
The default value is 240 seconds (4 minutes). At the end of this time, the Auto
Program turns the inner melt heater OFF and sounds 8 audible beeps.
After accessing the Prep 1 Time parameter, the display flashes dur 2 and dis
plays the current value of the Prep 2 Time parameter.
dur 2Flashes before entering the prep 2 time parameter
SEC240Current prep 2 time value (4 minutes)
Use the UP or DOWN buttons to adjust the value.
SEC360New prep 2 time value (6 minutes)
If no changes are desired, press the EXIT button to continue to the next parameter. Press the SET button to accept the new value and continue on to the next
parameter.
-
-
38
S
Accept new prep 2 time value and continue to prep 3 time
6.20.7Prep 3 Time
After turning OFF the inner melt heater during the initialization of the gallium
melt, see Figure 1, the Auto Program waits a short time before beginning the
maintenance process of the gallium melt plateau. During this time the instru
ment remains at 30.77°C to complete the melting of the outer surfaces of the
gallium in the cell. The Prep 3 Time parameter specifies this time in seconds.
The default value is 360 seconds (6 minutes). At the end of this time, the Auto
Program begins the maintenance process and sounds 16 audible beeps.
After accessing the Prep 2 Time parameter, the display flashes dur 3 and dis
plays the current value of the Prep 3 Time parameter.
dur 3Flashes before entering the prep 3 time parameter
SEC360Current prep 3 time value (6 minutes)
-
-
Page 43
Use the UP or DOWN buttons to adjust the value.
SEC480New prep 3 time value (8 minutes)
6 Controller Operation
Prog 2
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and to return to the Prog 1
submenu.
S
ProG 1Prog 1 submenu
Press the UP button to access Prog 2. The well temperature is displayed by
pressing the EXIT button. Press the SET button to access the parameters within
the Prog 1 submenu.
6.21Prog 2
Prog 2 is displayed by pressing the UP button while the display shows ProG
1 . Pressing the SET button then accesses the parameters within the Prog 2
submenu.
ProG 1Prog 1 submenu
U
ProG 2Prog 2 submenu
Pressing the DOWN button may access Prog 1. Pressing the UP button may ac
cess Prog 3. The well temperature is displayed by pressing the EXIT button.
Pressing the SET button accesses the parameters within the Prog 2 submenu.
-
Accept new prep 3 time value and return to the prog 1 submenu
Select the Prog 2 submenu
-
6.21.1Maintain Temperature
The Maintain Temperature parameter is the temperature at which the instru
ment stabilizes and remains during the melt plateau in the gallium cell. The de
fault value is 29.86°C and can be seen on Figure 1 as the temperature that the
well goes to when the display shows nnAint. The instrument holds the well at
this temperature until the melt plateau is ended.
ProG 2Prog 2 submenu
S
Access the maintain temperature parameter
nnaintFlashes before entering the maintain temperature
parameter
-
-
39
Page 44
9230 Maintenance System
User’s Guide
29.860Current maintain temperature value
Use the UP or DOWN buttons to adjust the value.
29.790New maintain temperature value
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new maintain temperature value and continue to the main
tain time-out
6.21.2Maintain Time-out
The Maintain Time-out setting is used to enable a preset delay time after which
the gallium melt plateau is terminated. The default setting is OFF and at this
setting the instrument remains at the Maintain Temperature until the user ad
vances the Auto Program by depressing the SET and DOWN buttons and then
selecting the FrEE2C program state. A refreeze of the gallium cell is then
started.
If the Maintain Time-out setting is ON, the Auto Program automatically ends
the gallium melt plateau after the preset delay time has been reached and
refreezes the gallium cell. It can be seen on Figure 1 under the Maintain portion
of the Auto Program.
After accessing the Maintain Temperature parameter, the display flashes ti
out and displays
ti outFlashes before entering the maintain time-out parameter
OFFCurrent maintain time-out setting
-
-
-
40
Use UP or DOWN buttons to adjust the value.
OnNew maintain time-out setting
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new maintain time-out setting and continue to the maintain
time-out time
6.21.3Maintain Time-out Time
The Maintain Time-out Time parameter is the time that the instrument main
tains the gallium cell at the Maintain Temperature before terminating the gal
-
-
-
Page 45
6 Controller Operation
lium melt plateau. The default value is 7200 minutes (5 days). It is only used if
the Maintain Time-out parameter is set to ON. If the Time-out parameter is set
to ON, the instrument ends the Maintain portion of the Auto Program after the
Maintain Time-out Time has been reached and refreezes the gallium cell. It can
be seen on Figure 1 under the Maintain portion of the Auto Program.
After accessing the Maintain Time-out parameter, the display flashes dur 4
and displays the current value of the Maintain Time-out Time.
Prog 2
dur 4Flashes before maintain time-out time parameter
7200Current maintain time-out time value (7200 minutes)
Use the UP or DOWN buttons to adjust the value.
1440New maintain time-out time value (1440 minutes)
If no changes are desired, press the EXIT button to continue to the next parameter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new maintain time-out time value and continue to the freeze
hot temperature
6.21.4Freeze Hot Temperature
After the melt plateau is finished or the user has ended the melt, the instrument
normally proceeds to refreeze the cell. If desired, the Freeze Hot Temperature
and Freeze Hot Time parameters may be used to preset the instrument to totally
melt the cell before refreezing.
The Freeze Hot Temperature parameter is the temperature at which the instru
ment totally melts the gallium in the cell. The value can be set up to 36.00°C
and can be seen on Figure 1 as the temperature that the well goes to when the
display shows FrEE2H. The instrument holds the well at this temperature for
up to 6.0 hours to ensure that all of the gallium has been melted before
refreezing the cell.
After accessing the Maintain Time-out Time parameter, the display flashes
FrEE2H and displays the current value of the Freeze Hot Temperature.
FrEE2HFlashes before entering the freeze hot temperature
parameter
29.86Current freeze hot temperature value
Use the UP or DOWN buttons to adjust the value.
-
41
Page 46
9230 Maintenance System
User’s Guide
36.00New freeze hot temperature value
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new freeze hot temperature value and continue to the freeze
hot time
6.21.5Freeze Hot Time
The Freeze Hot Time parameter is the time that the instrument maintains the
gallium cell at the Freeze Hot Temperature to totally melt the gallium in the
cell. The value can be set up to 360 minutes (6 hours) and can be seen on Fig
ure 1 as the time duration when the display shows FrEEZH. The instrument re
mains at the Freeze Hot Temperature for this time to ensure that all of the
gallium has been melted before refreezing the cell.
Note: The default program does not use the Freeze Hot option and thus the
Freeze Hot time is set to 0 seconds.
After accessing the Freeze Hot Temperature parameter, the display flashes dur
5 and displays the current value of the Freeze Hot Time.
dur 5Flashes before entering the freeze hot time parameter
0Current freeze hot time value (0 minutes)
Use the UP or DOWN buttons to adjust the value.
-
-
-
42
360New freeze hot time value (360 minutes)
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new freeze hot time value and continue to the freeze cold
temperature
6.21.6Freeze Cold Temperature
The Freeze Cold Temperature parameter is the temperature to which the instru
ment scans during the refreezing of the gallium cell. The default value is
0.00°C and can be seen on Figure 1 as the temperature the unit scans to while
the display shows FrEE2C.
After accessing the Freeze Hot Time parameter, the display flashes Free2C
and displays the current value of the Freeze Cold Temperature.
-
-
Page 47
6 Controller Operation
FrEE2CFlashes before entering the freeze cold temperature
parameter
0.000Current freeze cold temperature value
Use the UP or DOWN buttons to adjust the value.
5.000New freeze cold temperature value
Prog 2
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
S
Accept new freeze cold temperature value and continue to the freeze
cold scan rate
6.21.7Freeze Cold Scan Rate
The Freeze Cold Scan Rate parameter is the rate in °C/minute or °F/minute at
which the instrument scans from the Freeze Hot Temperature to the Freeze
Cold Temperature. The default value is 0.5°C/minute and can be seen on Figure
1 where the unit begins to refreeze the gallium cell. This parameter is very important for the proper gradient to be maintained during the refreeze process.
After accessing the Freeze Cold Temperature parameter, the display flashes
frECsr and displays the current value of the Freeze Cold Scan Rate.
FrECSrFlashes before entering the freeze cold scan rate parameter
Sr = 0.5Current freeze cold scan rate
Use the UP or DOWN buttons to adjust the value.
Sr = 0.6New freeze cold scan rate
If no changes are desired, press the EXIT button to continue to the next param
eter. Press the SET button to accept the new value and continue on to the next
parameter.
-
-
S
Accept new freeze cold scan rate value and continue to the Freeze
Cold Time
6.21.8Freeze Cold Time
The Freeze Cold Time parameter is the time that the instrument scans to and
maintains the gallium cell at the Freeze Cold Temperature to totally refreeze the
gallium in the cell. The default value is 150 minutes (2.5 hours). At the end of
43
Page 48
9230 Maintenance System
User’s Guide
this time, the Auto Program returns the well to 25°C to wait for the next melt
process.
After accessing the Freeze Cold Scan Rate parameter, the display flashes dur
6 and displays the current value of the Freeze Cold Time parameter.
dur 6Flashes before entering the freeze cold time parameter
150Current freeze cold time value (150 minutes)
Use the UP or DOWN buttons to adjust the value.
180New freeze cold time value (180 minutes)
If no changes are desired, press the EXIT button to Prog 1 submenu. Press the
SET button to accept the new value and continue on to the next parameter.
S
Prog 1 may be accessed by pressing the DOWN button. Prog 3 may be accessed by pressing the UP button. The well temperature is displayed by pressing the EXIT button. Pressing the SET button accesses the parameters within
the Prog 2 submenu.
Accept new freeze cold time value and continue to the Prog 2
submenu
6.22Prog 3
Prog 3 is displayed by pressing the UP button while the display shows ProG
2. Pressing the SET button then accesses the parameters within the Prog 3
submenu.
ProG 2Prog 2 submenu
U
ProG 3Prog 3 submenu
Prog 2 may be accessed by pressing the DOWN button. The serial interface
menu may be accessed by pressing the UP button. The well temperature is dis
played by pressing the EXIT button. Pressing the SET button accesses the pa
rameters within the Prog 3 submenu.
6.22.1Freeze Melt State
Select the Prog 3 submenu
-
-
44
The Freeze Melt State parameter is used to control the configuration of the
Peltier devices and is only usable in the manual mode of operation. During nor
mal operation of the instrument this parameter should not be altered.
-
Page 49
6 Controller Operation
Serial Interface Parameters
During the melt and maintain processes, the Peltier devices are configured to
apply a uniform heat source to the well. This minimizes vertical gradients in
the well and gallium cell to maximize the length of the melt plateau. During a
Freeze process, the Peltier devices are configured to generate a large vertical
gradient in the well to ensure the proper refreezing of the gallium cell.
The default setting for this parameter is MELT.
The Freeze Melt State parameter is accessed from the Prog 3 submenu.
ProG 3Prog 3 submenu
S
Access the Freeze Melt State parameter
FrnELTFlashes before entering the freeze melt state parameter
nnELTCurrent freeze melt state setting
Use the UP or DOWN buttons to adjust the setting.
FrEE2ENew freeze melt state setting
If no changes are desired, press the EXIT button to exit the Prog 3 submenu.
Press the SET button to accept the new value and return to the Prog 3 submenu.
S
The display then shows ProG 3 and returns to the Prog 3 submenu.
Accept new Freeze Melt State setting
ProG 3Prog 3 submenu
Prog 2 may be accessed by pressing the DOWN button. The serial interface
menu may be accessed by pressing the UP button. The well temperature is dis
played by pressing the EXIT button. Pressing the SET button accesses the pa
rameters within the Prog 3 submenu.
-
-
6.23Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SEriALSerial RS-232 interface parameters menu
The serial interface parameters menu contains parameters, which determine the
operation of the serial interface. The parameters in the menu are – BAUD rate,
sample period, duplex mode, and linefeed.
45
Page 50
9230 Maintenance System
User’s Guide
6.23.1BAUD Rate
The BAUD rate is the first parameter in the menu. The BAUD rate setting de
termines the serial communications transmission rate. The BAUD rate of the
serial communications may be programmed to 300, 600, 1200, 2400, 4800, or
9600. The default setting is 2400 baud.
bAUdFlashes before displaying the serial baud rate setting
2400 bCurrent BAUD rate
Use the UP or DOWN buttons to adjust the setting.
1200 bNew BAUD rate
Press the SET button to store the new setting as the baud rate or press the EXIT
button briefly to skip to the next parameter without storing the new setting.
6.23.2Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measurements transmitted from the serial interface. If the sample rate is set to 5, the instrument transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0.
SPerFlashes before displaying serial sample period setting
-
46
SP=1Current sample period (seconds)
Use the UP or DOWN buttons to adjust the setting.
SP=60New sample period
Press the SET button to store the new setting as the sample period or press the
EXIT button briefly to skip to the next parameter without storing the new
setting.
6.23.3Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du
plex or half duplex. With full duplex any commands received by the instrument
via the serial interface are immediately echoed or transmitted back to the device
of origin. With half duplex the commands are executed but not echoed.
dUPLFlashes before displaying serial duplex mode setting
-
Page 51
d=FULLCurrent duplex mode
Use the UP or DOWN buttons to adjust the setting.
d=HALFNew duplex mode setting
Press the SET button to store the new setting as the duplex mode or press the
EXIT button briefly to skip to the next parameter without storing the new
setting.
6.23.4Linefeed
6 Controller Operation
Calibration Parameters
The final parameter in the serial interface menu is the linefeed mode. This pa
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return.
LFFlashes before displaying serial linefeed setting
LF=OnCurrent linefeed setting
Use the UP or DOWN buttons to adjust the setting.
LF=OFFNew linefeed setting
Press the SET button to store the new setting as the linefeed.
Prog 3 may be accessed by pressing the DOWN button. The calibration menu
may be accessed by pressing the UP button. The well temperature is displayed
by pressing the EXIT button. Pressing the SET button accesses the parameters
within the serial interface menu.
6.24Calibration Parameters
The operator of the instrument has access to one of the calibration constants,
R0. This value is set at the factory and must not be altered. The correct value is
important to the accuracy and proper and safe operation of the instrument. This
parameter should not be adjusted except in the event the controller’s memory
fails. The user may then restore this value to the factory setting. This constant
is supplied to the user on the Report of Calibration with the manual.
-
CAUTION: DO NOT change the value of the instrument calibration con
stant from the factory set value. The correct setting of this parameter is
important to the safety and proper operation of the instrument.
CALCalibration parameters menu
-
47
Page 52
9230 Maintenance System
User’s Guide
Press the SET button five times to enter the menu.
6.24.1R0
The calibration parameter R0 characterizes the resistance temperature relation
ship of the platinum control sensor. This parameter may be adjusted by an ex
perienced user to improve the accuracy of the instrument.
To properly maintain the gallium cell during the melt plateau and during the
refreezing process, the accuracy of the Model 9230 near the gallium melting
point (29.7646°C) must be very precise, ±0.05°C or better. The value of R0 for
this instrument refers to the resistance of the control probe at 0°C. For calibra
tion purposes, the value of R0 is adjusted to compensate for the error at
29.27°C. The value of this parameter is set at the factory for best instrument ac
curacy. This parameter may be adjusted by an experienced user to improve the
accuracy of the instrument (see Section 10, Calibration Procedure).
From the CAL submenu, pressing the SET button 5 times accesses the R0
parameter:
-
CALCalibration parameters menu
Press the SET button five times to enter the menu.
r0Flashes before displaying the R0 setting
99.788Current R0 value (ohms)
Use the UP or DOWN button to adjust the setting.
99.999New R0 value (ohms)
-
-
-
48
If no changes are desired, press the EXIT button to return to the CAL submenu.
Press the SET button to accept the new value.
The display then shows CAL and returns to the CAL submenu. The serial inter
face menu may be accessed by pressing the DOWN button. The well tempera
ture is displayed by pressing the EXIT button. Pressing the SET button five
times accesses the R0 parameter within the CAL submenu.
-
-
Page 53
7 Digital Communication Interface
7Digital Communication Interface
The Model 9230 is capable of communicating with and being controlled by
other equipment through the digital interface.
With a digital interface the instrument may be connected to a computer or other
equipment. This allows the user to set the set-point temperature, monitor the
temperature, and access any of the other controller functions, all using remote
communications equipment. Communications commands are summarized in
Table 4 on page 52.
7.1Serial Communications
The instrument is installed with an RS-232 serial interface that allows serial
digital communications over fairly long distances. With the serial interface the
user may access any of the functions, parameters and settings discussed in Sec
tion 7, Controller Operation with the exception of the BAUD rate setting. Addi
tionally, when the instrument is in Standby mode (well at 25°C and display
shows Auto), the user is required to use the front panel buttons, to enter Manual mode or to initiate the Auto Program.
The serial interface operates with eight data bits, one stop bit, and no parity.
Serial Communications
-
-
7.1.1Wiring
The serial communications cable attaches to the instrument
through the DB-9 connector at
the back of the instrument. Figure 7 shows the pin-out of this
connector and suggested cable
wiring. To eliminate noise, the
serial cable should be shielded
with low resistance between the
connector (DB-9) and the
shield. If the instrument is used
in a heavy industrial setting, the
serial cable must be limited to
one meter in length.
7.1.2Setup
Before operation, the serial in
terface must first be set up by
programming the BAUD rate
and other configuration parame
ters. These parameters are pro
-
-
-
Figure 7 Serial Cable Wiring
49
Page 54
9230 Maintenance System
User’s Guide
grammed within the serial interface menu.
To enter the serial interface menu from the standby mode do the following:
AutoInstrument in standby mode
Use the UP or DOWN buttons to select the manual mode.
nnAnInstrument in manual mode
S
25.00 CWell temperature displayed
S+E
Pct PoFlashes prior to displaying Peltier device power
18.0 PPeltier device power in percent
S
ConFIGFlashes configuration menu
ProG 1Program 1 submenu
U
SErIALSerial interface menu
Enter the manual mode
Enter the secondary menu
SECFlashes and then
Continue pressing the SET button until the display shows
Press the UP button three times untilis displayed
50
S
7.1.2.1Baud Rate
The baud rate is the first parameter in the menu. The display prompts with the
baud rate parameter by showing “bAUd”. Press the SET button to choose to set
the baud rate. The current baud rate value is displayed. The baud rate may be
programmed to 300, 600, 1200, or 2400 baud. The baud rate defaults to 2400
baud.
Access the serial interface parameters
Page 55
Use the UP or DOWN buttons to change the baud rate value. Press the SET
button to set the baud to the new value or the EXIT button to abort the opera
tion and skip to the next parameter in the menu.
7.1.2.2Sample Period
The sample period is the next parameter in the menu and prompted with
“SPEr”. The sample period is the time period, in seconds, between tempera
ture measurements transmitted from the serial interface. If the sample rate is set
to 5, the instrument transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. Press the SET button to choose to set the sample period.
Adjust the period with the UP or DOWN button and then use the SET to set the
sample rate to the displayed value.
7.1.2.3Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex
mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full
duplex any commands received by the instrument via the serial interface are
immediately echoed or transmitted back to the device of origin. With half duplex the commands are executed but not echoed. The default setting is full duplex. The mode may be changed using the UP or DOWN buttons and pressing
the SET button.
7 Digital Communication Interface
Interface Commands
-
-
7.1.2.4Linefeed
The final parameter in the serial interface menu is the linefeed mode. This parameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac-
ter (LF, ASCII 10) after transmission of any carriage-return. The default setting
is with linefeed on. The mode may be changed using the UP or DOWN buttons
and pressing the SET button.
7.1.3Serial Operation
Once the cable has been attached and the interface set up properly the control
ler immediately begins transmitting temperature readings at the programmed
rate. The set-point and other commands may be sent to the instrument via the
serial interface to set the temperature set-point and to view or program the vari
ous parameters. The interface commands are discussed in Section 8.3, Interface
Commands.
The serial interface operates with eight data bits, one stop bit, and no parity.
7.2Interface Commands
The various commands for accessing the instrument functions via the digital in
terface are listed in this section. These commands are used with the RS-232 se
rial interface. The commands are terminated with a carriage-return character.
The interface makes no distinction between upper and lower case letters, hence
-
-
-
-
51
Page 56
9230 Maintenance System
User’s Guide
either may be used. Commands may be abbreviated to the minimum number of
letters, which determines a unique command. A command may be used to ei
ther set a parameter or display a parameter depending on whether or not a value
is sent with the command following an “=” character. For example, an “s”<cr>
returns the current set-point and an “s=28.00”<cr> sets the set-point to 28.00
degrees.
In the following list of commands, characters or data within brackets, “[“ and
“]”, are optional. A slash, “/”, denotes alternate characters or data. Numeric
data, denoted by “n”, may be entered in decimal or exponential notation. Char
acters are shown in lower case although upper case may be used. Spaces may
be added within command strings and are simply ignored. Backspace (BS,
ASCII 8) may be used to erase the previous character. A terminating CR is im
plied with all commands.
Tabl e 4 Serial Interface Commands
-
-
-
Command DescriptionCommand Format
Display Temperature
Read current set-points[etpoint]sset: 9999.99{C or F}set: 25.00 C
Set current set-point to ns[etpoint]=ns=28-5.00C to
Read unitsuuu:{C or F}u: C
Read temperaturettt: 9999.99{C or F}t: 29.86 C
Set temperature units:u[nits]=c/fCorF
Set temperature units to Celsius u[nits]=cu=c
Set temperature units to
Fahrenheit
Read scan functionsc[an]scscan:{ON or OFF}scan:ON
Set scan function:sc[an]=on/of[f]ON or OFF
Turn scan function onsc[an]=onsc=on
Turn scan function offsc[an]=of[f]sc=of
Read scan ratesr[ate]srsrat: 999.99{C or F}/min srat: 0.5C/min0.1C/min to
Setscanrateto
minute
Read program stateadvadvadv:{OFF, WAIT, PREP,
Advance program stateadv={adv/auto}adv=advADV, AUTO
Secondary Menu
Read heater power
(duty cycle)
n
degrees per
u[nits]=fu=f
sr[ate]=n]sr=5
po[wer]popo: 999.9po: 12.3
Command
ExampleReturned
MAINTAIN, FREEZHOT,
FREEZCOLD}
Returned
Example
WAIT
Acceptable
Values
40.00C
5.0C/min
52
Page 57
7 Digital Communication Interface
Interface Commands
Command DescriptionCommand Format
Read proportional bandpr[op-band]prpb: 999.9pb: 8
Set proportional bandpr[op-band]=npr=8.830.1C to 100C
Set serial duplex mode:du[plex]=f[ull]/h[alf]FULL or HALF
Set serial duplex mode to fulldu[plex]=f[ull]du=f
Set serial duplex mode to halfdu[plex]=h[alf]du=h
Set serial linefeed mode:lf[eed]=on/of[f]ON or OFF
Set serial linefeed mode to onlf[eed]=onlf=on
Set serial linefeed mode to offlf[eed]=of[f]lf=of
Calibration Menu
Read R0 calibration parameterr[0]rr0: 999.999
Set R0 calibration parameterr[0]=
These commands are only used for factory testing
Miscellaneous (not on menus)
Read firmware version number*ver[sion]*verver: 999,9.99ver.9230,v1.00
Read structure of all commandsh[elp]hlist of commands
Legend:[] Optional Command data
NOTE:
n
sa[mple]=
n
n
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Command
ExampleReturned
F)/min
sa=00 to 10,000
r=100.71398.0 to 102.0
Returned
Example
freezCsrate:
0.2(C or F)/min
MELT Mode
Acceptable
Values
FREEZE
ohms
54
Page 59
8 Melting Point of Gallium Realization
8Melting Point of Gallium Realization
This section explains the procedure for realizing the melting point of gallium
using a Model 5943 Stainless Steel Gallium Cell and the Model 9230 Mainte
nance System.
8.1General
The gallium melting point (29.7646°C) is a critical temperature. Thermometers
used in life science, environmental monitoring, at room or body temperatures,
and many other applications depend on it for accurate calibrations. Laboratory
standards rely on it as an ITS-90 check standard and as a means of measuring
drift between calibrations. Use of the Model 5943 Stainless Steel Gallium Cell
and the Model 9230 Maintenance System provides an easy way to generate and
maintain the gallium melting point. Plateaus lasting up to 5 days are typical.
The Model 5943 Stainless Steel Gallium Cell holds a gallium sample that is
99.99999+% pure. The gallium is sealed in a Teflon envelope in a high purity
argon atmosphere, which is then sealed inside a stainless steel housing. This
double-sealing method reduces leaching of the gallium sample and ensures a
life of ten years or longer for the cell. Gallium expands on freezing by 3.1%
and, to compensate for this, the inner construction of the cell incorporates a
flexible Teflon wall. Although the cell is more rugged than other gallium cells,
care must be taken to prevent damage to the cell, particularly during the
refreeze process. Because gallium expands on freezing (like water), the cell
must be frozen from the bottom up or there is potential for damage to the cell.
The Auto Program process of the Model 9230 ensures that the cell is frozen
properly. The cell’s entrant well dimensions are 8.7” (length) x 0.324” (diameter) with an immersion depth of 6.58” in gallium.
The Model 9230 Gallium Maintenance Apparatus is specifically designed to
maintain the Model 5943 Gallium Cell. Once properly installed in the Model
9230, the Auto Program may be started and subsequently initiates the melt pla
teau, maintains the plateau for up to 5 days typical, and properly refreezes the
cell for the next melt. When not actively melting or refreezing a cell, the Model
9230 automatically maintains the cell at 25.00°C (Standby), well below the
melting point. To initiate a melt, the inner melt heater is inserted into the cell
and the Auto Program is started through the Model 9230’s front-panel buttons.
The Model 9230 increases its temperature to just above the gallium point and
powers the 8-watt inner melt heater. Approximately 30 minutes later, the gal
lium cell is ready to be used (see Figure 1).
The Model 9230 uses a Proportional Integral Derivative PID controller with
specialized analog circuitry to control several Peltier devices. The controlled
Peltier devices provide the appropriate amount of heating or cooling to the
block assembly into which the gallium cell is inserted. The controller also con
trols the inner melt heater, cooling fan and beeper alarm. Two preheat wells en
sure that thermometers are properly warmed when inserted into the gallium
cell. The preheat wells are oversized so thermometers fit loosely and heat loss
General
-
-
-
-
-
55
Page 60
9230 Maintenance System
User’s Guide
from the gallium cell is neither too sudden nor too severe when cold thermome
ters are initially placed into the wells.
At the end of the melting plateau (or when the user wishes to end the plateau),
the Auto Program is advanced. When totally melted, the Peltier devices are re
configured and controlled such that a large temperature gradient is generated
between the top and bottom of the block. The instrument slowly scans to 0°C to
maintain the gradient over several hours. This ensures that the gallium cell is
frozen from the bottom up.
In summary, the following steps are involved in the realization of the melting
point of gallium:
Prepare the Model 9230 to receive the gallium cell.
•
Prepare the frozen Model 5943 gallium cell to be installed in the Model
•
9230.
Install the frozen Model 5943 Stainless Steel Gallium Cell into the Model
•
9230.
• Insert the top insulation plugs into the well on top of the cell.
• Insert the inner melt heater into the entrant well of the gallium cell.
• Start the Auto Program.
• Generate the melt plateau and remove the inner melt heater when
prompted by the Model 9230 beeper alarm.
• Calibrate thermometers during the gallium melt plateau.
• Advance the Auto Program to refreeze the cell.
The details of operating the Model 9230 to realize the melting point of gallium
are explained for each step in the following sections.
-
-
56
8.2Preparing the Model 9230 to Receive the
Gallium Cell
The Model 9230 is designed to be used in a controlled laboratory environment
(see Section 3.2, Environmental Conditions). Control the ambient temperature
to a fixed temperature within the specified operating range (18°C to 28°C).
Rapid changes in ambient temperature of more than ±1.0°C reduces the stabil
ity of the instrument. Avoid drafty areas.
Place the Model 9230 on a flat surface with at least 6 inches of free space
around and 18 inches above the instrument.
Make sure that the well of the instrument is clean and clear of any foreign ob
jects, debris, dirt or grit.
Connect the inner melt heater to the back of the instrument making sure that it
is securely connected. Place the inner melt heater cartridge into the melt heater
storage well at the top back of the instrument.
-
-
Page 61
Installation in the Model 9230
Preparing the Frozen Model 5943 Gallium Cell for
8 Melting Point of Gallium Realization
Verify that the POWER switch is in the OFF position (“0”). Verify that the In
ner Melt Heater switch is in the AUTO position. Install the power cord into the
power entry module on the back of the instrument and plug the cord into a
mains outlet of the proper voltage, frequency, and current capability. Turn on
the power to the instrument by toggling the switch on the power entry module
located on the back of the instrument. The fan should begin blowing air
through the instrument, the beeper sounds one beep, and the display illuminates
after approximately 3 seconds. After a brief self-test the controller should begin
normal operation, the instrument will begin to approach 25°C, and the display
will show Auto.
8.3Preparing the Frozen Model 5943 Gallium Cell
for Installation in the Model 9230
Maintain the Model 5943 gallium cell at temperatures below the melting point
(29.7646°C). The cell must be frozen when installed in the Model 9230.
Make sure that the outer surfaces and entrant well of the frozen Model 5943
gallium cell are clean and that the entrant well of the cell is clear of any foreign
objects, debris, dirt or grit. Use pure or reagent grade alcohol and clean, soft
cotton cloths to clean all surfaces if necessary. Thoroughly dry the cell before
continuing.
Add a small amount of silicone oil 200.05, or equivalent transfer fluid, to the
entrant well of the cell. Do not overfill the entrant well. The oil level in the
well should be approximately the same level (52 mm below the top of the cell)
as that of the gallium in the cell when the thermometer is in place in the well.
-
8.4Installing the Frozen Model 5943 Gallium Cell
into the Model 9230
The instrument comes with a cell removal tool to aid in installing and removing
the gallium cell from the Model 9230 well. The tool has a spring-loaded handle
with a compressible rubber tip. When the handle is squeezed, thereby com
pressing the spring, the rubber tip is extended and the tip may be inserted into
the entrant well of the cell. Releasing the handle causes the rubber tip to com
press and become firmly positioned in the entrant well. Once lodged in the en
trant well, the cell can be gently lowered into the Model 9230 well. Wet the
rubber tip of the tool with transfer fluid (e.g. silicone oil 200.05) for ease in in
serting into the cell’s entrant well.
CAUTION: Do not use the cell removal tool to carry the cell. The tool is
to be used only to aid in installing and removing the cell from the Model
9230 well. Do not drop the cell into the well. Damage to the cell and
Model 9230 may occur if the cell is dropped.
-
-
-
-
57
Page 62
9230 Maintenance System
User’s Guide
With one hand holding the outside of the cell, insert the cell removal tool tip
into the cell’s entrant well as described above and verify that the rubber tip is
firmly lodged after releasing the tool handle. Hold onto the removal tool with
the other hand and gently lower the cell with the tool the rest of the way while
carefully releasing the grip on the outside of the cell. When the cell is securely
lowered into the Model 9230 well, compress the removal tool handle to loosen
the rubber tip and gently remove the tool from the cell’s entrant well. The gal
lium cell is now installed in the Model 9230.
8.5Inserting the Top Insulation Plugs
The instrument comes with 4 rubber insulation plugs to be inserted into the
Model 9230 well on top of the cell (2 spares). The plugs have a hole in the cen
ter through which thermometers can be inserted. The plugs serve two purposes.
First, they help ensure an even temperature profile in the block and cell allow
ing a longer lasting and more accurate melting-point realization. Second, they
protect against excessive moisture condensation during the cell refreezing pro
cess. Condensation can lead to corrosion over the long term and freezing water
can damage the block.
Firmly insert two insulation plugs into the Model 9230 well on top of the cell.
Make sure that they are positioned such that the center holes are aligned with
the entrant well of the gallium cell.
-
-
-
-
58
8.6Inserting the Inner Melt Heater
The inner melt heater’s function is to melt a small volume of gallium around
the entrant well of the cell when initiating the melt. The recommended practice
for realizing the gallium melt by national standards laboratories requires two
solid/liquid interfaces. An inner liquid interface exists at the reentrant well with
an outer liquid interface completely surrounding the inner interface at the con
tainer wall. For a detailed description of the liquid/solid/liquid interfaces in the
realization of the gallium melt see NIST Note 1265.
Verify that the inner melt heater is securely connected to the back of the instru
ment. Remove the inner melt heater from the storage well and carefully insert it
through the holes in the top insulation plugs and into the entrant well of the gal
lium cell. Make sure that the heater cartridge is completely inserted to the bot
tom of the cell’s entrant well.
8.7Auto Program
The Model 9230 is designed to automate the gallium cell melt plateau. This is
accomplished through the use of the Auto Program, a microprocessor-con
trolled program that determines set temperatures, scan rates, duration times,
Peltier module configurations and melt heater controls. By controlling all of
these parameters in the proper order, a temperature profile can be maintained in
the block assembly that assures a uniform melt, maintains the melt over long
-
-
-
-
-
Page 63
8 Melting Point of Gallium Realization
Auto Program
periods of time, properly refreezes the cell and keeps the cell ready for the next
melt cycle. The Auto Program is illustrated in Figure 1.
8.7.1Starting the Auto Program
After the frozen gallium cell has been properly installed in the instrument and
the unit has been powered ON, the front panel LED display shows Auto.The
instrument goes to 25°C and remains there until the SET button is pressed on
the front panel starting the Auto Program. After pressing the SET button, the
display changes to uuait and the Auto Program begins. The instrument auto
matically resets to 29.27°C at a scan rate of 0.2°C/minute in accordance with
the Auto Program (see Figure 1).
8.7.2Generating the Melt Plateau and Removing the Inner
Melt Heater
Once the unit has stabilized at 29.27°C, it waits for a minimum of 30 minutes
to ensure that the gallium cell has equalized in temperature. At the end of the
equalization time, the Auto Program begins to initiate the melt. The display
changes to PrEP as the set temperature of the block is changed to 30.77°C.
This initiates a melt on the outer surfaces of the gallium within the cell. A few
minutes into the program, the unit beeps four times and the display changes to
a message that flashes between PrEP and Htr on to indicate that the inner
melt heater has been turned ON. The inner melt heater remains ON for a few
minutes, then shuts off as the display changes back to PrEP , and the unit
beeps 8 times. The user should remove the inner melt heater at this time (Note:
Damage does not occur to the gallium cell if the heater is not removed, however, a shortening of the melt duration may result).
After the inner melt heater has been powered OFF, the instrument remains at
30.77°C for a few more minutes to complete the outer melt of the gallium cell.
The unit then beeps 16 times, the display changes from PrEP to nnAinT and
the block temperature resets to 29.86°C to maintain the melting gallium cell.
-
8.7.3Calibrating Thermometers
After the melt has been initiated, a preheated monitor thermometer may be in
serted to monitor the temperature of the gallium cell. Due to the heat applied to
the cell by the inner melt heater, the temperature of the cell’s entrant well is
higher than the gallium melting point (29.7646°C) for a short period of time.
Generally, after 30 to 60 minutes, the temperature of the cell’s entrant well is
sufficiently stable to begin calibrations of thermometer sensors. The duration of
the melt plateau varies depending on usage, but lengths of 5 days or longer are
achievable.
The two preheat wells in the block may be used to heat up the thermometers
before inserting them into the cell. The preheat wells are slightly oversized so
thermometers fit loosely and heat loss from the gallium cell is neither too sud
den nor too severe when cold thermometers are initially placed in the preheat
wells.
-
-
59
Page 64
9230 Maintenance System
User’s Guide
8.7.4Advancing the Auto Program and Refreezing the Gallium
Cell
At the end of the melt plateau (or when the user wishes to end the plateau), the
Auto Program is advanced by pressing the SET and DOWN buttons and then
using the UP and DOWN buttons to select the next program state. Press SET to
start the program. See Section 6.21.2, Maintain Time-out, for optional user de
fined Auto Time-Out. Gallium, like water, expands in volume when freezing.
Therefore, the cell must be frozen from the bottom up or there is a potential for
damage to the cell. The Auto Program process ensures that the cell is properly
refrozen.
The Auto Program initiates the freeze and the display shows FrEE2C at this
point. The Peltier devices are reconfigured and controlled such that a large tem
perature gradient is generated between the top and bottom of the block. The in
strument slowly scans to 0°C to maintain the gradient over several hours.
8.7.5Return to Standby
Upon completion of the freeze, the Auto Program returns the unit to 25°C, the
front panel LED display shows Auto , and the unit remains there until the user
initiates the Auto Program again by pressing the SET key.
-
-
-
60
Page 65
9Calibration Procedure
9 Calibration Procedure
Calibration Equipment
Note:This procedure is to be considered a general guideline. Each labora
tory should write their own procedure based on their equipment and their
quality program. Each procedure should be accompanied by an uncer
tainty analysis also based on the laboratory’s equipment and environment.
To properly maintain the gallium cell during the melting plateau and during the
refreezing process, the accuracy of the Model 9230 near the gallium melting
point (29.7646°C) must be very precise, ±0.05°C or better. To achieve this, the
instrument must be calibrated near this temperature. The value of R0 for this
instrument refers to the resistance of the control probe at 0°C. For calibration
purposes, the value of R0 is adjusted to compensate for the error at 29.27°C.
The value of this parameter is set at the factory for best instrument accuracy
and is located on the Report of Calibration shipped with the instrument. This
parameter may be adjusted by an experienced user to improve the accuracy of
the instrument. A calibration should be done at regular intervals, such as once a
year, to ensure that it is operating properly.
9.1Calibration Equipment
Calibration of the Model 9230 requires a standard thermometer and readout device that are adequately accurate. The combined accuracy of the thermometer
and readout should be ±0.0125°C or better. Recommended equipment includes
a laboratory grade PRT or thermistor probe with a length of 230 to 300 mm (9
to 12 inches). The calibration may be performed in a frozen Model 5943 gallium cell or in an aluminum test block that simulates the gallium cell. A small
amount of silicone oil 200.05 (or equivalent) should be used in the entrant well
for good thermal transfer from the cell to the thermometer.
-
-
9.2Calibration Procedure
The accuracy of the Model 9230 at the gallium melting point (29.7646°C) is
determined by the value of the calibration parameter R0. The calibration proce
dure involves measuring the error between the Model 9230 and the reference
thermometer at a temperature near the gallium melting point and adjusting the
calibration parameter R0 as necessary to reduce the errors to within acceptable
limits. The stated accuracy of the Model 9230 is given in Section 3.1, Specifi
cations. For calibration purposes, the temperature is measured at 29.27°C and
R0 is adjusted as applicable. The following steps may be followed to calibrate
the Model 9230:
1. Enter the Manual mode of operation and set the temperature of the Model
9230 to 29.270°C with a frozen gallium cell installed. Use a small amount of
silicone oil 200.05 (or equivalent) in the entrant well for good thermal transfer
from the cell to the thermometer. Be sure to install the 3 top insulation plugs
-
-
61
Page 66
9230 Maintenance System
User’s Guide
on top of the cell. Insert the reference thermometer and monitor the tempera
ture of the cell with an appropriate readout.
2. Allow the system to stabilize at 29.270°C for approximately 1 to 2 hours.
When the stability of the reading is less than ±0.02 over a 20 to 30 minute pe
riod record the average reading of the reference thermometer. This reading
gives the error of the Model 9230 and may be used for “as found” data if re
quired. No further action is required if the unit is found to be within acceptable
limits.
3. If step 2 above shows that the Model 9230 is out of specification or accept
able limits, use the following equation to calculate a new R0 value:
R
= R1– (T2– T1) * (0.3850)
2
Where,
R
= the new calculated value for R0
2
R
= the old value for R0 currently programmed in the Model 9230 mem
1
ory
T
= the temperature measured in step 2 above
2
T
= the set temperature of the 9230 from step 1 above (29.270°C)
1
The value of R
9230. Use the following process with the front panel buttons to obtain the currently programmed value of R0:
(old R0) should be obtained from the memory of the Model
1
-
-
-
-
25.00 CInstrument in manual mode; well temperature displayed
-
62
S+E
Press the SET button and then the EXIT button to enter the
secondary menu
SECFlashes secondary menu
Pct PoFlashes percent power
18.0PPeltier device power in percent
S
Press the SET button untilis displayed
ConFIGFlashes configuration menu
ProG 1Prog 1 submenu
U
Press the UP button untilis displayed
Page 67
CALCalibration parameter menu
9 Calibration Procedure
Calibration Procedure
S
Press the SET button 5 times to access the R0 value
R0Flashes before entering the R0 setting
99.983Current R0 value (ohms)
4. Program the new value for R0 (R
memory (see Section 6.24.1, R0).
5. Repeat steps 1 and 2 to verify that the newly calibrated Model 9230 is now
within acceptable limits. If not, perform steps 3, 4, and 5 again as necessary.
Note: If it is known that the Model 9230 is more than 0.5°C out of specifica
tion, such that, if the unit was calibrated at 29.270°C the cell would begin melt
ing, then the above procedure should be performed below the 29.270°C set
point as applicable (e.g., 28.5°C). After adjusting the unit at the lower temperature (e.g., 28.5°C), and verifying that the error had been reduced to less than
0.5°C, then the above procedure may be safely used.
) from step 3 above into the Model 9230
2
-
-
63
Page 68
10Maintenance
The calibration instrument has been designed with the utmost care. Ease
•
of operation and simplicity of maintenance have been a central theme in
the product development. Therefore, with proper care the instrument
should require very little maintenance. Avoid operating the instrument in
an oily, wet, dirty, or dusty environment.
If the outside of the instrument becomes soiled, it may be wiped clean
•
with a damp cloth and mild detergent. Do not use harsh chemicals on the
surface, which may damage the paint.
It is important to keep the well of the instrument clean and clear of any
•
foreign matter.
The instrument should be handled with care. Avoid knocking or dropping
•
the instrument.
Examine the gallium cell for deformities before inserting it in the well. If
•
there is any chance of jamming the cell in the well, file or grind off the
protuberance.
• Do not drop the probes into the well. This type of action can cause damage to the entrant well of the gallium cell.
• If a hazardous material is spilled on or inside the equipment, the user is
responsible for taking the appropriate decontamination steps as outlined
by the national safety council with respect to the material.
• If the mains supply cord becomes damaged, replace it with a cord with
the appropriate gauge wire for the current of the instrument. If there are
any questions, call Hart Scientific Customer Service for more information.
•
Before using any cleaning or decontamination method except those rec
ommended by Hart, users should check with Hart Scientific Customer
Service to be sure that the proposed method does not damage the equip
ment.
•
If the instrument is used in a manner not in accordance with the equip
ment design, the operation of the instrument may be impaired or safety
hazards may arise.
•
Do not drop the gallium cell into the instrument well. This type of action
can cause a shock to the control sensor, thermoelectric modules, or dam
age the cell.
10 Maintenance
-
-
-
-
65
Page 69
11 Troubleshooting
11Troubleshooting
This section contains information on troubleshooting, CE Comments, and a
wiring diagram.
11.1Troubleshooting Problems, Possible Causes,
and Solutions
In the event that the instrument appears to function abnormally, this section
may help to find and solve the problem. Several possible problem conditions
are described along with likely causes and solutions. If a problem arises, please
read this section carefully and attempt to understand and solve the problem. If
the problem cannot otherwise be solved, contact Hart Scientific Customer Ser
vice for assistance (1-801-763-1600). Be sure to have the model number and
serial number of your instrument available.
ProblemPossible Causes and Solutions
Incorrect temperature
reading
Incorrect R0 parameter. Find the value for R0 on the Report of Calibration
that was shipped with the instrument (or from subsequent calibrations of the instrument). Reprogram the parameter into the Model 9230 memory (see Section
6.24.1, R0). Allow the instrument to stabilize and verify the accuracy of the
temperature reading.
-
Blank display after mains
power applied
The Instrument heats or
cools too quickly or too
slowly
Controller locked up. The controller may have locked up due to a power surge
or other aberration. Initialize the system by performing the Factory Reset
Sequence.
Factory Reset Sequence. Hold the SET and EXIT buttons down at the same
time while powering on the instrument. After the instrument displays
release the buttons. The display shows
plays the firmware version, and then displays
ter reset sequence, all of the configuration parameters are reset to their default
values. Reprogram the R0 parameter into the Model 9230 memory (see Sec
tion 6.24.1, R0) and any other applicable configuration parameters. Allow the
instrument to stabilize and verify the accuracy of the temperature reading.
Blown fuse. A fuse may have blown due to a power surge or failure of a com
ponent. Replace the fuse once. If the fuse blows a second time, it is likely
caused by the failure of a component. Always replace the fuse with one of the
same rating, voltage, and type. Never replace the fuse with one of a higher cur
rent rating.
Incorrect scan and scan rate settings. The scan and scan rate settings may
be set to unwanted values. Check the Scan and Scan Rate settings. The scan
may be off (if the unit seems to be responding too quickly). The scan may be
on with the Scan Rate set low (if unit seems to be responding too slowly).
-init-
, then displays
Auto
. After performing the mas
9230
-init-
, then dis
,
-
67
-
-
-
-
Page 70
9230 Maintenance System
User’s Guide
ProblemPossible Causes and Solutions
The display shows any of
the following:
err 2 , err 3 , err
4,orerr5
The display shows
6
The display shows
8
After starting the Auto
Program the display
shows
uuAit
n’t go into
err 1 ,
PrEP
err
err
and does
.
Controller problem. The error messages signify the following problems with
the controller.
Err 1
- a RAM error
Err 2
- a NVRAM error
Err 3
- a Structure error
Err 4
- an ADC setup error
Err 5
- an ADC ready error
Initialize the system by performing the Factory Reset Sequence describe
above.
Defective control sensor. The control sensor may be shorted, open or other
wise damaged. Disconnect the power cord from the instrument. Looking at the
front of the instrument (display), remove the front panel and right side panel.
Locate the analog control board (on the block side of the electronics panel).
The Molex connector at the top of the board connects the control probe to the
controller (see wiring diagram, connector J4). Remove the connector and mea
sure the resistance of the control sensor. It should be approximately 110 ohms
at room temperature. If it measures close to this, make sure that it is securely
re-connected to the controller, replace the panels and reapply power to the in
strument. If the error message returns, there is a problem with the controller.
Improper freezing. The Auto Program of the instrument is designed to monitor
power levels and scan rates during the freezing cycle. If an unacceptable level
or rate is detected, the unit will show
To prevent any damage from occurring to the cell, remove it immediately and
freeze it manually. To freeze manually, place the lower 1/3 of the cell in an ice
bath with the top 2/3 exposed to the ambient air for approximately one hour until the cell freezes. The instrument may be reset by pressing any of the front
panel buttons, but if the error message continues, the unacceptable freeze conditions are still present. Power the unit OFF and wait a few minutes. Then
power the unit ON and reset the instrument by pressing any of the front panel
buttons to see if the problem has been eliminated.
Ambient temperature may be changing too rapidly. Rapid changes in the
ambient temperature of more than ±1.0°C reduce the stability of the instrument.
-
The Auto Program monitors the stability of the block at 29.27°C and if it is too
unstable the melt does not proceed. Check the ambient temperature around
the instrument for rapid changes. Avoid drafty areas.
err 8
and the beeper alarm will sound.
-
-
-
68
11.2Comments
11.2.1EMC Directive
Hart Scientifics’ equipment has been tested to meet the European Electromag
netic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration
of Conformity for your instrument lists the specific standards to which the unit
was tested.
11.2.2Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (73/23/EEC),
Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN
61010-1) and the IEC 1010-2-010 (EN 61010-2-010) standards.
-
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.