Each Fluke product is warranted to be free from defects in material and workmanship under normal use and
service. The warranty period is two years and begins on the date of shipment. Parts, product repairs, and services
are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke
authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion,
has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation
or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications
for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software
will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but
have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if
product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased
in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or
replacement of a defective product which is returned to a Fluke authorized service center within the warranty
period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization
information, then send the product to that service center, with a description of the difficulty, postage and insurance
prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will
be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by
neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including
overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical
components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work.
Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the
repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSE
DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation
of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every
buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of
competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
S, INCLUDING LOSS OF
11/99
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
ООО «Флюк СИАЙЭС»
125167, г. Москва,
Ленинградский проспект дом 37,
корпус 9, подъезд 4, 1 этаж
The 700G Series Pressure Gauges (the Product) are high-accuracy digital pressure test gauges.
Accurate to 0.05 % Full Scale, the Products can be used as a calibration reference, or in any application
where high-accuracy pressure measurement is required.
The Product has user-configurable functions that include:
•Sample rate
•Auto off
•Ta re
•Damper
The Product includes password protect to lock and unlock the functions.
The Product ships with a Protective Cover, three AA Batteries (installed), and an NPT/metric Adapter.
•Units selection
•Max/Min
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
•Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
To register your product, visit http://register.fluke.com
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals
.
.
1
Page 6
700G Series
Users Manual
Safety Information
A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies
conditions and procedures that can cause damage to the Product or the equipment under test.
W Warning
To prevent possible fire, explosion, or personal injury:
•Read all safety information before you use the Product.
•Carefully read all instructions.
•Do not alter the Product and use only as specified, or the protection supplied by the
Product can be compromised.
•Do not use the Product if it operates incorrectly.
•Do not use the Product if it is altered or damaged.
•Disable the Product if it is damaged.
•Avoid any action that can generate an electrostatic charge. Electrostatic discharge is an
explosion hazard.
•Clean the Product only with a damp cloth.
•Do not rub the non-metallic parts of the enclosure (display overlay or holster) with a dry
cloth. This action can create an electrostatic discharge.
•Only assemble and operate high-pressure systems if you know the correct safety
procedures. High-pressure liquids and gases are hazardous and the energy from them
can be released without warning.
•Do not use the RS-232 interface in hazardous areas. The RS-232 interface port must be
sealed when used in a hazardous area.
•Substitution of components can impair suitability for hazardous locations.
•If the Product is exposed to overpressure or sudden physical shock (such as a drop),
examine the Product for damage that can cause a safety concern. If necessary, return
the Product to Fluke for evaluation.
•The battery door must be closed and locked before you operate the Product.
•Replace the batteries when the low battery indicator shows to prevent incorrect
measurements.
•Repair the Product before use if the battery leaks.
•Be sure that the battery polarity is correct to prevent battery leakage.
2
Page 7
Pressure Gauges
•Batteries must only be changed in an area known to be non-hazardous. Explosion
hazard.
•Remove the batteries if the Product is not used for an extended period of time, or if
stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can
damage the Product.
•Have an approved technician repair the Product.
W Caution
To avoid possible damage to Product or to equipment under test:
•If the display reads “OL” the range limit is exceeded and the pressure source must
immediately be reduced.
•Always apply thread seal tape to the threads of the gauge.
•Do not exceed the maximum torque allowed. Maximum torque allowed is
13.5 Nm = 10 lbft.
Table 1 lists the symbols that can be used on the Product or in this manual.
Table 1. Symbols
SymbolDescription
WWARNING - RISK OF DANGER. Consult user documentation.
Battery status
Battery
PConforms to European Union directives.
)Certified by CSA Group to North American safety standards.
Conforms to relevant Australian Safety and EMC standards.
Conforms to relevant South Korean EMC Standards.
(Conforms to the European Explosive Atmospheres (ATEX) directive.
This product complies with the WEEE Directive marking requirements. The affixed
label indicates that you must not discard this electrical/electronic product in
~
domestic household waste. Product Category: With reference to the equipment
types in the WEEE Directive Annex I, this product is classed as category 9
"Monitoring and Control Instrumentation" product. Do not dispose of this product as
unsorted municipal waste.
3
Page 8
700G Series
10K PSI
700 BAR
700G31 10K PSIG
PRESSURE
GAUGE
6
3
1
4
2
5
Users Manual
Ex-Hazardous Areas
An Ex-hazardous area as used in this manual refers to an area made hazardous by the potential
presence of flammable or explosive vapors. These areas are also referred to as hazardous locations,
see NFPA 70 Article 500.
Display and Buttons
Table 2 lists the features of the Product.
Table 2. Product Features
ItemDescriptionItemDescription
Pressure value
On/Off button
Pressure unit
D
E
F
Bar graph
Control panel
1/4 in NPT Connector
4
Page 9
Pressure Gauges
Control Panel
The control panel is used to change settings or select functions and options. Table 3 lists the functions of
the buttons on the Control Panel.
Table 3. Control Panel
ItemDefault FunctionSetup Mode Function
Zeros the display.
For absolute versions of the gauge,
push to set the barometric reference
pressure.
The Product records minimum and
maximum pressure values and saves them
in memory.
Push 1X to show the maximum (MAX)
value on the display.
Push 2X to show the minimum (MIN) value
on the display.
After 2 seconds, the display shows the
live operation value again.
To clear the MIN/MAX memory values,
push and hold for 2 seconds until CLr
shows on the display.
Note
Note
Within a function, push to move forward
through options.
Within a function, push to move backward
through options.
Push to cycle through the functions.
Within a function, push to set the option
and go back to the Setup menu.
Within a function, push to set the option
and go back to the live view.
Opens the Setup menu to configure the
Product.
Toggles the backlight on/off.
Basic Operation
Push to turn on/off the Product. To maximize the life of the batteries, use the Auto Off feature. See
Table 4.
The analog bar graph at the bottom of the display shows the applied-pressure level relative to the full
range of the gauge.
Note
If you record a Tare value, the pressure value that shows on the display is not the actual pressure
applied to the gauge.
5
Page 10
700G Series
Users Manual
Setup the Product
Use the Setup menu to view and change settings.
To change settings:
1. Push
2. Push
3. Push
4. Push
Table 4 lists the functions that can be accessed, changed, locked, or unlocked in the Setup menu.
UNITS<options>
AUTO-OFF<options>Sets the time before the Product automatically turns off.
Battery Charge--
Temperature
to open the Setup menu.
again to select a function.
/ to select an option.
to set the option and go back to the previous function, or push to set the option and
go back to live view.
Table 4. Functions
FunctionOptionDescription
Selects a pressure unit from a list of predefined common
units and a custom unit/scale (CUST). See Set Custom Unit/Scale.
See Specifications for a list of available units.
Shows the battery voltage and a percent-of-life bar graph of
the battery charge.
F
C
Sets the temperature units to Celsius or Fahrenheit.
The Product is temperature compensated and shows the
temperature on the display that is measured by an internal
sensor.
On
DAMP
Off
RATE<options>
TARE<options>
FUnC LOCK<options>
6
Turns on/off the damper.
The damper prevents sudden changes in pressure values
when the source pressure pulsates.
Sets the rate, in samples per second, at which the Product
takes a pressure measurement and updates the value on the
display.
Sets a constant offset value to subtract from a pressure
measurement. See Set Tare.
Shows if functions are locked.
In Supervisory mode, use to lock/unlock functions to enable/
disable changes to the functions. See Supervisory Mode.
Page 11
Pressure Gauges
Set Tare
Use to set a constant offset value to subtract from a pressure measurement.
Note
If you record a Tare value, the pressure value that shows on the display is not the actual pressure
applied to the gauge.
For example, if a tare is set at 30 psi and the measured pressure is 37 psi, 7 psi shows on the display. If
a tare is set at 30 psi and the measured pressure is 27 psi, -3 psi shows on the display.
The tare value is based on the pressure units and resolution selected. The tare value can be set to the
maximum range of the gauge.
For safety, the bar graph always shows the actual pressure based on the full range of the gauge
regardless of the tare position. Even when the pressure value shows “0”, the bar graph shows the actual
applied pressure.
Supervisory Mode
If needed, use Supervisory mode to lock/unlock functions to enable/disable changes to the functions. If
FUnC LOCK shows on the display, at least one function is locked. The password for Supervisory mode
is “101”. The password is set at the factory and cannot be changed.
To open Supervisory mode:
1. Push
2. Push
3. Push
To lock/unlock a function:
1. Open Supervisory mode.
2. Push
3. Push
.
0 PWRD shows on the display.
to put in the password.
To move through the numbers in increments of 10, push and hold
numbers in increments of 1, push
/.
.
/ to select UnLOC or LOC.
to set the option.
/. To move through the
Set Custom Unit/Scale
Use to set a factor to multiply a pressure measurement by to make a custom unit/scale.
Example: 40 psi is the equivalent of 1000 lbs of product in a tank. To show the product weight with a
100 psi gauge, set the factor to 25. 40 psi pressure shows as 1000 (40 x 25) on the display. The units
show as CUST.
7
Page 12
700G Series
Users Manual
To set a factor:
1. Go to Supervisory mode.
2. Push
3. Push
4. Push
until SET FACTR shows on the display.
/ to select a factor.
to set the option and go back to the previous function, or push to set the option and
go back to live view.
Maintenance
Clean the Product
Clean the Product with a soft cloth dampened with water or water and weak soap.
WCaution
To prevent possible damage to the Product, do not use solvents or abrasive cleansers.
WCaution
For safe operation and maintenance of the product:
•Repair the Product before use if the battery leaks.
•Be sure that the battery polarity is correct to prevent battery leakage.
•Remove the batteries if the Product is not used for an extended period of time, or if
stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can
damage the Product.
•Replace the batteries when the low battery indicator shows to prevent incorrect
measurements.
•Have an approved technician repair the Product.
Batteries
When the battery voltage is low, shows on the top left of the display.
WWarning
To prevent possible fire, explosion, or personal injury:
•Batteries must only be changed in an area known to be non-hazardous. Explosion
hazard.
•Use replacement batteries of the same type. The replacement batteries must be the same
design, chemistry, age, and manufacturer.
8
Page 13
Pressure Gauges
To replace the batteries (See Figure 1.):
1. Use a Phillips screwdriver to loosen the captive screw on the battery door.
2. Remove the battery door and seal.
3. Replace the batteries. For more information, see Specifications.
4. Replace the battery door and seal.
Note
Align the battery door and seal properly with the Product to protect the Product from moisture and
dust.
5. Tighten the captive screw.
Figure 1. Battery Replacement
9
Page 14
700G Series
Users Manual
RS-232 Interface
The Product includes an RS-232 interface. You can use serial communication to configure and calibrate
the Product and move measurement data from the Product to a PC. An RS-232/USB cable is sold
separately and includes 700G/TRACK Software. For specifications on the interface, see Specifications.
WWarning
To prevent possible fire, explosion, or personal injury, do not use the RS-232 interface in
hazardous areas. The RS-232 interface port must be sealed when used in a hazardous area.
Specifications
Available Input Ranges................................See Ranges and Resolution tables for available ranges in psi plus equivalent
ranges and resolution for all pressure units.
Accuracy 700G Ranges
Positive Pressure (700G01, 700G02).........±0.1 % Full Scale
Positive Pressure (all other ranges)............±0.05 % Full Scale
Vacuum.......................................................±0.1 % Full Scale
Temperature Compensation .......................15 °C to 35 °C (59 °F to 95 °F) to rated accuracy
Note
For temperatures from -10 °C to +15 °C and +35 °C to +55 °C, add 0.003 % FS / °C.
Accuracy 700RG Ranges
Positive Pressure........................................±0.04 % rdg ±0.01 % Full Scale
Vacuum (700RG05)....................................±0.05 % Full Scale
Vacuum (all other ranges) ..........................±0.1 % Full Scale
Temperature Compensation .......................0 °C to +50 °C (+32 °F to +122 °F) to rated accuracy
Note
For temperatures from -10 °C to 0 °C and +50 °C to +55 °C, add 0.005 % FS / °C.
Media Compatibility
700G01, 700G02, 700G04,
700G05, 700RG05......................................any clean, dry, non-flammable, non-corrosive gas
All other ranges
100 psi to 1000 psi ..................................any non-flammable liquids or gases compatible with 316 stainless steel
Above 1000 psi .......................................any non-flammable, non-toxic, non-oxidizing liquid or gas compatible with 316
Mechanical Specifications
Dimensions .................................................11.4 cm x 12.7 cm x 3.7 cm (4.5 in x 5.0 in x 1.5 in)
Pressure
Connection ..............................................¼ in NPT Male
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is
necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-
voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in
ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to
radio reception in such environments.
Korea (KCC) ...............................................Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take
notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) .................................................47 CFR 15 subpart B. This product is considered an exempt device per clause
15.103.
RS-232 Interface ...........................................The Serial Interface (J4) is a 3.5 mm Stereo Jack connector. Use the RS-232/
USB cable sold with the 700G/TRACK Software. USB to RS-232 converter with
5 V RS-232 signals. Connection to RS-232 Interface in non-hazardous area only:
Ui=18V, Pi=0.5W.
Compliance Marks
) ............................................................Class I, Division 2, Groups A-D
P.................................................................................( II 3 G Ex ic IIB T6 Gc
SIRA 17ATEX4160X
Input Entity Parameters:
Ta = -10 °C...+55 °C
Ui = 18 V, Pi = 0.5 W (NON HAZARDOUS AREA)