Pressure module specifications (all specifications in % of full span. Specifications reflect a confidence interval of 95%.)
Range/ Range (approx)/ uncertainty Stability Temperature Total
Reference High2Low
1
side side Fitting pressure
Model Resolution Resolution (23 ± 3ºC) (1 year) (0 to 50ºC) uncertainty media media material (x nominal)
Differential
FLUKE-700P00 1 in. H2O/0.001 0.25 kPa/0.0002 0.300 0.025 0.025 0.350 Dry Dry 316 SS 30x
FLUKE-700P01 10 in. H
FLUKE-700P02 1 psi/0.0001 6900 Pa/0.7 0.150 0.070 0.080 0.300 Dry Dry 316 SS 3x
O/0.01 2.5 kPa/0.002 0.200 0.050 0.050 0.300 Dry Dry 316 SS 3x
2
FLUKE-700P22 1 psi/0.0001 6900 Pa/0.7 0.100 0.020 0.030 0.150 316 SS Dry 316 SS 3x
FLUKE-700P03 5 psi/0.0001 34 kPa/0.001 0.050 0.020 0.030 0.100 Dry Dry 316 SS 3x
FLUKE-700P23 5 psi/0.0001 34 kPa/0.001 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3x
FLUKE-700P04 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 Dry Dry 316 SS 3x
FLUKE-700P24 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3x
Gage
FLUKE-700P05 30 psi/0.001 207 kPa/0.01 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P06 100 psi/0.01 690 kPa/0.07 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P27 300 psi / 0.01 2070 kPa / 0.1 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P07 500 psi/0.01 3400 kPa/0.1 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P08 1000 psi/0.1 6900 kPa/0.7 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P09 1500 psi/0.1 10 MPa/0.001 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 2x
Absolute (not compatible with Fluke 701 or 702)
FLUKE-700PA3 5 psi/0.0001 34 kPa/0.001 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x
FLUKE-700PA4 15 psi/0.001 103 kPa/0.01 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x
FLUKE-700PA5 30 psi/0.001 207 kPa/0.01 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x
FLUKE-700PA6 100 psi/0.01 690 kPa/0.07 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x
Vacuum (not compatible with Fluke 701 or 702)
FLUKE-700PV3 -5 psi/0.0001 -34 kPa/0.001 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3x
FLUKE-700PV4 -15 psi/0.001 -103 kPa/0.01 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3x
Dual
FLUKE-700PD2 ± 1 psi/0.0001 ±6900 Pa/0.7 0.150 0.025 0.025 0.200 316 SS Dry 316 SS 3x
FLUKE-700PD3 ± 5 psi/0.0001 ± 34 kPa/0.001 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3x
FLUKE-700PD4 ± 15 psi/0.001 ± 103 kPa/0.01 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3x
FLUKE-700PD5 -15/30 psi/0.001 -100/207 kPa/0.01 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700PD6 -15/100 psi/0.01 -100/690 kPa/0.07 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700PD7 -15/200 psi/0.01 -100/1380 kPa/0.1 0.040 0.015 0.015 0.070 316 SS N/A 316 SS 3x
High
FLUKE-700P29 3000 psi/0.1 20.7 M Pa/0.001 0.050 0.010 0.020 0.080 C276 N/A C276 2x
FLUKE-700P30 5000 psi/0.1 34 M Pa/0.001 0.050 0.010 0.020 0.080 C276 N/A C276 2x
FLUKE-700P31 10000 psi/1 69 M Pa/0.007 0.050 0.010 0.020 0.080 C276 N/A C276 1.5x
2
Max over-
1
Total uncertainty, one
year for temperature
range 0°C to +50°C. Total
uncertainty, 1.0% of full
span for temperature
range -10°C to 0°C.
For P00 module only,
compensated
temperature range is
15 ° to 35° C.
2
“Dry” indicates dry air
or non-corrosive gas as
compatible media.
“316 SS” indicates media
compatible with Type
316 Stainless Steel.
“C276” indicates media
compatible with
Hastelloy C276.
Use of pressure zero is
required prior to
measurement or source.
Maximum overpressure
specification includes
common mode pressure.
Modules are rated.
Metric adapter(s):
1/4" NPT female to male
BSP/ISO 1/4-19, tapered
thread, included with all
modules except P29,
P30, and P31. Effective
October 1996, all
modules include a NIST
traceable certificate and
test data.