The Fluke 572-2 Infrared Thermometer is the one product
you can use in high-temperature industrial environments all
around the world. Whether you work in power utility, metal
refining and smelting, glass, cement or petrochemical environments, the new 572-2 allows you to carry the most
trusted name in test tools anywhere you need accurate, hightemperature and high distance-to-spot measurements.
With a straight-forward user interface and soft-key menus,
the Fluke 572-2 makes even complex measurements easy.
Quickly navigate and adjust emissivity, start data logging, or
turn on and off alarms, with just a few pushes of a button.
Product highlights
With a rugged, easy-to-use,
ergonomic design, the Fluke
575-2 can stand up to tough
industrial, electrical, and
mechanical environments.
Measure -30 °C to 900 °C
•
(-22 °F to 1652 °F)
60:1 distance to spot ratio with
•
dual laser sighting for fast,
accurate targeting
Multi-language interface
•
(user select)
Current Temperature plus MAX,
•
MIN, DIF, AVG temperature
displays
Compatible with standard
•
mini-connector K-type thermocouples, including ones
you already own and have
installed.
Adjustable emissivity and pre-
•
defined emissivity table
Infrared and thermocouple
•
temperature on backlit display
Last reading Hold (20 seconds)
•
High and low temperature
•
alarm
Data storage and review (99
•
data sets)
Tripod mount
•
12 or 24 hour clock
•
USB 2.0 computer interface
•
cable
FlukeView® Forms Document-
•
ing Software
Two-year warranty
•
Page 2
572-2 specifications
Infrared measurements
Infrared temperature range-30 °C to 900 °C (-22 °F to 1652 °F)
IR accuracy (Calibration geometry with ambient
temperature 23 °C ± 2 °C)
IR Repeatability± 0.5 % of reading or ± 0.5 °C (± 1 °F), whichever is greater
Display Resolution0.1 °C / 0.1 °F
Distance:Spot60:1 (calculated at 90 % energy)
Minimum spot size19 mm
Laser sightingOffset dual laser, output <1 mW
Spectral Response8 μm to 14 μm
Response Time (95 %)<500 ms
EmissivityDigitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials
Measurement options
Hi/Low alarmsAudible and two-color visual
Min/Max/Avg/DifYes
Switchable celsius and fahrenheitYes
BacklightTwo levels, normal and extra bright for darker environments
Probe inputK-type thermocouple
Trigger lockYes
Data storage99 points
DisplayDot matrix 98 x 96 pixels with function menus
CommunicationUSB 2.0
K-type thermocouple specifications
K-type thermocouple input temperature range-270 °C to 1372 °C (-454 °F to 2501 °F)
K-type thermocouple input accuracy (with ambient
temperature 23 °C ± 2 °C)
K-type thermocouple resolution0.1 °C/0.1 °F
K-type thermocouple repeatability± 0.5 % of reading or ± 0.5 °C (± 1 °F), whichever is greater
Measurement range
(K-type thermocouple bead probe)
Accuracy± 1.1 °C (± 2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F). Typically within 1.1 °C (2.0 °F) from
Cable length1 m (40 in) K-type thermocouple cable with standard miniature thermocouple connector
General specifications
Operating temperature0 °C to 50 °C (32 °F to 122 °F)
Storage temperature-20 °C to 60 °C (-4 °F to 140 °F)
Relative humidity10 % to 90 % RH non-condensing up to 30 °C (86 °F)
Operating altitude2000 meters above mean sea level
Weight0.322 kg (0.7099 lb)
Power2 AA Batteries
Battery life8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100 % duty
Safety and complianceIEC 60825-1
≥ 0 °C: ± 1°C or ± 1 % of the reading, whichever is greater
≥-10 °C to <0 °C: ± 2 °C
<-10 °C: ± 3 °C
≥32 °F: ± 2 °F or ± 1 % of the reading, whichever is greater
≥14 °F to <32 °F: ± 4 °F
<14 °F: ± 6 °F
Simultaneous display of probe and IR temperature
< -40 °C: ± (1 °C + 0.2 °/1 °C)
≥ -40 °C: ± 1 % or 1 °C, whichever is greater
< -40 °F: ± (2 °F + 0.2 °/1 °F)
≥ -40 °F: ± 1 % or 2 °F, whichever is greater
-40 °C to 260 °C (-40 °F to 500 °F)
-40 °C to 0 °C (-40 °F to 32 °F)
and bead termination
cycle (thermometer continuously on)
FDA Laser Class II
EMC 61326-1
CE Compliance
CMC 沪制01120009
2 Fluke Corporation 572-2 Infrared Thermometer
Page 3
Recommended temperature probes
ProbeUsage
80PK-1 The general purpose bead probe is for quick, accurate surface temperatures and air temperatures within ducts,
80PK-8Pipe clamp probes (2) are essential for tracking continuously changing temperature differentials on hydronic
80PK-9The insulation-piercing probe provides a sharp tip to pierce pipe insulation and flat probe tip for good surface
80PK-11Flexible cuff thermocouple temperature probe is a convenient way to attach a thermocouple to a pipe while
80PK-25The piercing probe is the most versatile option. Good for checking air temperature in ducts, surface temperature
80PK-26The tapered probe is a good general-purpose gas and surface probe, with a good length and low mass tip casing
Ordering information
572-2 Infrared Thermometer
Includes
Infrared thermometer with contact thermometer
capabilities, K-type thermocouple bead probe, USB
2.0 computer interface cable, FlukeView
Documenting Software, hard carrying case, getting
started guide (print) and user’s manual (CD).
vent temperatures.
tubing and pipe loops, and good for quick, accurate refrigerant temperatures.
thermal contact, air temperatures within ducts, and vent temperatures.
keeping hands free.
under carpets/pads, liquids, thermometer wells, vent temperatures, and for penetrating pipe insulation.
for faster reaction to surface and air temperatures.
®
Forms
3 Fluke Corporation 572-2 Infrared Thermometer
Fluke. The Most Trusted Tools
in the World.
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com