Fluke 5622 Application Note

How to calibrate an RTD
or Platinum Resistance
Thermometer (PRT)
Topics Covered
Calibration by characterization ITS-90 Callendar-Van Dusen Polynomial
Tolerance testing ASTM 1137 IEC 60751
Application Note
Introduction
There are two types of calibrations applicable to PRTs—characterization and tolerance testing. The type of calibration to perform is determined by the way in which the UUT is to be used and the accuracy required by the user. Characterization is the type of calibra­tion in which the unit under test (UUT) resistance is determined at several temperature points and the data are fitted to a mathematical expression. Tolerance testing on the other hand is a calibration in which the UUT resistance is compared to defined values at specific temperatures. No data fitting is performed. In the laboratory, we are required to perform both types of calibration depending upon our customer’s needs.
Calibration Procedures
Characterization
Characterization is the method that is most often used for medium to high accuracy PRT calibration. With this method, a new resistance vs. temperature relationship is deter­mined anew with each calibration. Generally, with this type of calibration, new cali­bration coefficients and a calibration table are provided as a product of the calibration. There are five basic steps to perform as listed below:
1. Place the reference probe and the UUTs in the temperature source in close proxim­ity to one another.
2. Connect the leads to the readout(s) ensuring proper 2-, 3-, or 4-wire connection.
3. Measure the reference probe and determine the temperature.
4. Measure and record the resistance of the UUT(s).
5. Fit the data.
Some readouts simplify the technique by combining or eliminating some of the steps. In the following discussion, we will consider an application involving PRT characteriza­tion by comparison to an SPRT.
Step 1: Probe Placement
All temperature sources have instabilities and gradients. These translate into calibra­tion errors and/or uncertainties. To minimize the effects, the probes should be placed as close together as practical. In baths the probes to be calibrated should be placed in a radial pattern with the reference probe in the center (focus) of the circle. This ensures an equal distance from the reference probe to each of the UUTs. In dry-well temperature sources, the reference probe and probes to be calibrated should all be placed the same distance from the center for best results, but the reference may be placed in the center if needed.
Also, the sensing elements should be on the same horizontal plane. Even though sensing elements are different lengths, having the bottoms of the probes at the same level is sufficient. Sufficient immersion must be achieved so that stem losses do not occur. Generally, sufficient immersion is achieved when the probes are immersed to a depth equal to 20 times the probe diameter plus the length of the sensing ele­ment. For example, consider a 3/16 inch diameter probe with a 1 inch long sensing element. Using the rule of thumb, 20 x 3/16 in + 1 in = 3 3/4 in + 1 in = 4 3/4 in. In this example, minimum immersion is achieved at 4 3/4 inches. This rule of thumb is generally correct with thin wall probe construction and in situations of good heat transfer. If the probe has thick wall construction and/or poor heat transfer is present (such as in the case of a dry-well with incorrectly sized holes), more immersion is required.
F r o m t h e F l u k e D i g i t a l L i b r a r y @ w w w . f l u k e . c o m / l i b r a r y
Step 2: Connection to Readout
This step is straightforward. Connections must be tight and in proper 2-, 3-, or 4-wire configuration. If using 4-wire configuration, ensure that the current and voltage connections are correct. See Figure 1.
temperature data is available in real time. Some modern readouts also display the data in graphical format, allowing the operator to determine stabil­ity at a glance. Both of these features speed up the process and eliminate possible operator error due to incorrect table interpolation.
The second method is used when the readout does not provide for proper temperature calculation. (Some readouts, particularly DMMs, have some of the more popular temperature conversions built in. These typically do not allow use of unique calibra­tion coefficients and cannot be used for accurate temperature calibration.) In this case, the resistance is measured and the temperature is determined from either a calibration table or from a computer or calculator program.
Since the temperature must be calculated after the resistance is measured, the process is slower and does not provide immediate, real time tempera­ture data. See Tables 1 and 2 below.
Table 1. Interpolation from an RTD calibration table (resistance vs. temperature).
t( °C) R(t) (W) dR/dt(t) W/°C
400 249.8820 0.3514
401 250.2335 0.3513
402 250.5848 0.3512
403 250.9360 0.3511
450 267.3108 0.3456
451 267.6564 0.3455
452 268.0019 0.3454
453 268.3472 0.3452
Figure 1. Thermometer readout connection schematics
Step 3: Measurement of Reference Probe and Temperature Determination
There are two ways to measure the reference probe and determine the temperature. Both techniques have the same potential accuracy. That is, if done correctly, neither technique is inherently more accurate than the other.
The first and best method is used with sophis­ticated readouts designed for temperature work. The resistance is measured and the temperature calculated from calibration coefficients which were entered into the readout previously. Once these calibration coefficients have been entered, the tem­perature calculations are accomplished internally and the readout displays in temperature units. The
1. Measure the reference probe resistance
2. Locate where it falls on the table
3. Subtract lower table value from measured value
4. Divide by dR/dT(t) (slope of curve)
5. Add fractional temperature to table value
249.9071 W
between 249.8820 W and
250.2335 W
249.9071 W – 249.8820 W =
0.0251 W
0.0251 / 0.3514 = 0.0714 °C
0.0714 ’C + 400 =
400.0714 °C
Tech Tip
Manual calculation is more prone to human error and is more time consum­ing than using a readout designed for temperature work.
2 Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT)
Table 2. Interpolation from an RTD calibration (resistance ratio (W) table).
t( °C) W(t) dt/dW(t)
300 2.1429223 275.2199
301 2.1465557 275.3075
302 2.1501880 275.3951
303 2.1538192 275.4827
350 2.3231801 279.6655
351 2.3267558 279.7559
352 2.3303304 279.8464
353 2.3339037 279.9369
1. Measure reference probe resistance
2. Calculate W (Rt/R = 25.54964)
3. Locate where it falls on the table
4. Subtract lower table value from measured value
5. Multiply by dt/dW(t) (inverse slope of curve)
6. 6) Add fractional temperature to table value
) (R
tpw
54.75258 W
54.75258 W / 25.54964 W =
tpw
2.1429883
between 2.1429223 and
2.1465557
2.1429883 – 2.1429223 =
0.000066
0.000066 • 275.2199 =
0.0182 °C
0.01821 °C + 300 °C =
300.0182 °C
Number of readouts - will the reference probe
•
and UUTs be measured with the same readout or different readouts?
Type of readout - a readout designed for temper-
•
ature calibration often has features which allow flexibility in the measurement scheme.
UUT characteristics - self-heating time, source
•
current requirements, stability, and overall qual­ity influence the measurement process.
It is not possible for us to anticipate all of the variables and discuss the optimum solutions here. However, in the following examples, we will con­sider some typical calibration scenarios and sug­gested measurement schemes.
Example 1: 2 DMM readouts, 1 reference probe and 5 UUTs
Step 4: Measurement of Units Under Test (UUTs)
Since the UUTs are resistance thermometers simi­lar to the reference probe, they are measured in a similar manner. If several UUTs are undergoing calibration, ensure that when they are connected or switched in, sufficient time is allowed for self­heating to occur before the data is recorded. Also, ensure that the readout is set to the correct range to provide the proper source current and to prevent range changes between the measurements at dif­ferent temperatures. Typically, the measurements are conducted starting at the highest temperature of calibration and working down. Additionally, it increases the precision of the calibration to use a mean (average) value calculated from multiple measurements at the same temperature. Often, the readout is designed with statistical features to facili­tate this practice. It is also a good practice to close the process with an additional measurement of the reference probe. The sequence in which the probes (reference and UUT) are measured is referred to as a measurement scheme. There are many variables to consider when designing a measurement scheme. Some points to consider are:
Accuracy - the higher the accuracy desired, the
•
more all of the following must be considered. Temperature source stability - the more stable
•
the source, the more time exists to conduct the measurements before temperature changes cause unwanted error.
Number of UUTs - the higher the number, the
•
longer it takes to cycle through all UUTs.
The reference probe is connected to one readout and the first UUT is connected to the second read­out. This places the probes to be measured under current at all times, thus, eliminating self-heating errors caused by changing current conditions. The UUTs will be connected and measured individually.
The scheme is as follows:
REF(1)-UUT (1) - REF(2)-UUT (2) - REF(3)-UUT (3) -
REF(4)-UUT (4) - REF(5)-UUT (5)
This provides 5 readings each of the reference and the UUT. Take the average of the readings and use it for the data fit. If the reference probe read­ings are in resistance, the temperature will have to be computed. After completion, repeat the process for the additional UUTs.
Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT) 3
Example 2: 1 DMM readouts, 1 reference probe and 5 UUTs
This example is similar to the first except that the reference probe and UUT must be measured by the same readout. The same scheme can be followed but more time must be allowed between readings to allow for self-heating. Since more time is involved, it might be beneficial to reduce the number of read­ings from five to three unless the heat source is extremely stable. Each probe will be connected and measured individually
The scheme is as follows:
wait-REF(1)-wait-UUT (1) - wait-REF(2)-wait-
UUT(2) - wait-REF(1)-wait-UUT(3)-done
This provides 3 readings each of the reference and the UUT. Take the average of the readings and use it for the data fit. Again, the reference probe readings are in resistance so the temperature will have to be computed. After completion, repeat the process for the additional UUTs.
Example 3: 1 multi-channel thermometer readout, 1 reference probe and 5 UUTs
In this example, all of the probes are connected directly to the thermometer readout, a Hart Scien-
tific 1560 Black Stack. The readout controls the measurement and scans through all probes perform­ing statistics in real time. Current may or may not be supplied at all times depending on the type of thermometer readout. If current is supplied at all times, there will be no self-heating errors. If current is not supplied at all times, ensure that the switch­ing is done rapidly enough to reduce self-heating errors to a negligible level.
The scheme is as follows
REF - UUT 1 - UUT 2 - UUT 3 - UUT 4 - UUT 5 -
repeat 10 or more times
This provides many readings each of the refer­ence and all of the UUTs. The average can be cal­culated and displayed directly by the readout. Also, the reference probe readings are in temperature so no further computation is required - the data is ready to fit.
Step 5: Thermometer calibration curve fitting
Data fitting is simple in concept but can be compli­cated in practice. Essentially it is a process of solv­ing a set of simultaneous equations which contain the calibration data to arrive at a set of coefficients unique to the PRT and calibration. There are several commercial software programs available specifically written to accomplish this task. Some are limited in function and do no more than solve the basic temperature functions. Others are more flexible and allow options regarding the number and location of calibration points and provide analysis regarding the precision of the resultant fit. The latter type of program is preferred. For metrologists who wish to tackle the algorithms themselves, a good mathemat­ics application software like Mathcad or Mathemat­ica or even a spreadsheet like Excel is extremely helpful. Hart Scientific offers two programs: TableWare for calculating calibration coefficients and MET/TEMP II for automating calibration tasks and calculating calibration coefficients. Of course, programs can be written in any of the modern com­puter languages (with double precision or better floating point capability) to perform the calculations with equal accuracy.
There are several equations which are used for PRT characterization. Among the most common are the International Temperature Scale of 1990 (ITS-90) series, the Callendar-Van Dusen, and third through fifth order polynomials. Obviously, with more than one model available to describe the behavior of a physical system, we must choose which one is best for our situation. The following discussion covers the features and purpose of each of these models and describes the form of the equa­tions. The steps necessary to actually fit the data will be discussed in the section on mathematics later in this manual.
ITS-90: The ITS-90 series of functions were developed through a concerted effort from the international metrology community’s leading tem­perature experts. These functions are intended to describe how the behavior of the SPRT relates, with a very high degree of precision, to the fixed points
4 Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT)
on which the scale is based. It does this extremely well SPRTs and with high quality PRTs. The ITS­90 uses a reference function - deviation function structure that has many advantages over traditional polynomials and is the preferred model for high accuracy applications. In the equations below, capi­tal T refers to ITS-90 temperatures expressed in Kelvin units.
Equation 1. ITS-90 resistance ratio
R T
( )
W T
( )
90
90
=
R
TPW
Where: W(T90) = resistance ratio at temperature T
R(T90) = measured resistance at tempera­ture T RTPW = measured resistance at the triple point of water
Equation 2. Deviation from ITS-90 reference func­tion.
W T W T W T
( ) ( ) ( )
=
90 90 90
r
Where: W(T90) = deviation of calculated W from
reference function at temperature T W(T90) = calculated resistance ratio at tem­perature T (from equation (1)) Wr(T90) = reference function value at tem­perature T
Equation 3. Reference function for the range
13.8033 K to 273.16 K (–259.3467 °C to 0.01 °C):
i
.
1 5
+
  
 
12
ln W T A A
( ( ))
= +
r i
90 0
i
=
1
T
ln
 
 
90
.
273 16
.
1 5
 
Where: Wr(T90) = reference function value at tem-
perature T Ai = reference function coefficients from definition
Equation 4. Deviation function for the subrange
83.8058 K to 273.16 K (–189.3442 °C to 0.01 °C):
W T a W T b W T W T
4 90 4 90 4 90 90
1 1( ) ( ( ) ) ( ( ) ) ln( ( ))= + − ⋅
Where: W(T90) = calculated deviation value at
temperature T (from equation (2)) W(T90) = calculated resistance ratio at temperature T (from equation (1)) a4, b4 = resulting calibration coefficients
Where: Wr(T90) = reference function value at tem-
perature T Ci = reference function coefficients from definition
Equation 6. Deviation function for the subrange
273.15 K to 692.677 K (0.00 °C to 419.527 °C):
W T a W T b W T
8 90 8 90 8 90
1 1( ) ( ( ) ( ( ) )= − +
2
Where: W(T90) = calculated deviation value at tem-
perature T (from equation (2)) W(T90) = calculated resistance ratio at tem­perature T (from equation (1)) a8, b8 = resulting calibration coefficients
The designations 4 and 8 in the deviation func­tions, equations (4) and (6) were inserted by NIST for identification of specific subranges. The values for the coefficients Ai and Ci in the reference func­tions, equations (3) and (5) are given in Table 3.
Table 3. ITS-90 Reference Function Coefficients
Coefficient Value
A
0
A
1
A
2
A
3
A
4
A
5
A
6
A
7
A
8
A
9
A
10
A
11
A
12
C
0
C
1
C
2
C
3
C
4
C
5
C
6
C
7
C
8
C
9
–2.135 347 29
3.183 247 20
–1.801 435 97
0.717 272 04
0.503 440 27
–0.618 993 95
–0.053 323 22
0.280 213 62
0.107 152 24
–0.293 028 65
0.044 598 72
0.118 686 32
–0.052 481 34
2.781 572 54
1.646 509 16
–0.137 143 90
–0.006 497 67
–0.002 344 44
0.005 118 68
0.001 879 82
–0.002 044 72
–0.000 461 22
0.000 457 24
Equation 5. Reference function for the range
273.15 K to 1234.93 K (0.00 °C to 961.78 °C):
i
.
  
Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT) 5
9
W T C C
( )
= +
r i
90 0
i
=
1
T
90
754 15
K
481
 
Callendar-Van Dusen: The Callendar-Van Dusen (CVD) equation has a long history. It was the main equation for SPRT and PRT interpolation for many years. It formed the basis for the temperature scales of 1927, 1948, and 1968. This equation is far simpler than the ITS-90 equations but has seri­ous limitations in the precision of fit. As a result, it is not suitable for high accuracy applications but is perfectly suited to modest accuracy applications. Partly due to its history and simplicity, but mostly due to its continued suitability, it continues to be the preferred model for industrial platinum resis­tance thermometers today. In the equations below, lower case t refers to ITS-90 temperature in Celsius units.
Equation 7. Callendar-Van Dusen resistance ratio
R t
( )
( )
=
R
0
W t
Where: W(t) = resistance ratio at temperature t
R(t) = measured resistance at tempera­ture t R0 = measured resistance 0 °C
And...
Equation 8. Callendar-Van Dusen equation
W t At Bt Ct t( ) ( )= + + + ⋅ −1 100
2 3
Where: W(t) = resistance ratio at temperature t (ref-
erence 0 °C) A,B,C = calibration coefficients (C is = 0 for temperatures above 0 °C)
NOTE: All temperatures are expressed in °C and
the resistance ratio (W) is referenced to 0 °C rather than the triple point of water (0.010 °C) as with the ITS-90.
Polynomials: Polynomials are frequently used to model physical phenomena from all fields of sci­ence. They have limited use with PRTs because of the high order required to achieve a suitable fit. (Recall that the reference functions for the ITS-90 are 9th and 12th order polynomials for the ranges above 0 °C and below 0 °C.) Additionally, the previ­ous models use resistance ratio as the variable to fit. Most polynomials in use fit the resistance directly. Since resistance is not as stable as the resistance ratio, these models have serious limitations. That having been said, polynomials can be very useful over limited ranges and in applications where accu­racy requirements are very modest.
Equation 9. Typical RTD polynomial equation
t a bR cR dR eR= + + + +
2 3 4
Tolerance Testing Method
PRT calibrations involving tolerance testing are reserved for low accuracy applications. With this type of calibration the UUT resistance is compared to defined values at specific temperatures. The values are defined by one of the common models such as the ASTM 1137 or IEC 60751 curve. PRTs calibrated in this way are generally used in indus­trial style applications where the readout is unable to accept unique coefficients but is pre-programmed with a common PRT curve. The probe must be tested to ascertain its compliance to the curve of interest. There are accuracy classes defined that probes are intended to fit.
The two common accuracy classes are class A
and class B::
IEC 60751 ASTM 1137
Class A
Class B
± [0.15 + (0.002 · t)] °C ± [0.13 + (0.0017 · t)] °C
± [0.30 + (0.005 · t)] °C ± [0.25 + (0.0042 · t)] °C
These include errors arising from deviations in R0 and from errors in slope. Frequently, we will see probes rated at a fraction of Class A. For example,
0.1 ASTM Class A. Fractional accuracy is achievable in sensors alone, but are very difficult to achieve in probes. The calculations are straightforward. See below:
Example 4: Calculate the accuracy of a 0.1 ASTM Class A probe at 100 °C
1. = (0.13 + (0.0017 · t)) · 0.1
2. = (0.13 + (0.0017 · 100)) · 0.1
3. = (0.13 + 0.17) · 0.1 = 0.03
PRTs that conform to a standard specification such as ASTM 1137 or IEC 60751 are expected to be within tolerances of defined resistance values for any given temperature. The resistance values are defined by a form of the Callendar-Van Dusen (CVD) equation and specified values for coefficients A,B and C (see table 4). These values may be deter­mined using a published table or calculated by solv­ing the equations.
Measurements for tolerance testing are carried out in the same manner as measurements for char­acterization. ITS-90 temperature is determined by the reference thermometer. The resistance of the UUT is then compared to the defined resistance val­ues, and pass or fail status is determined based on the specified tolerances (i.e. Class A or Class B).
Where: t = temperature
R = resistance a,b,c,d,e = calibration coefficients
6 Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT)
Table 4. Equations for ASTM 1137 and IEC 60751
° ≤ < °200 0 C Ct
0 650 ° ≤ ≤ °C Ct
Range Callendar-Van Dusen Equation
R R At Bt C t t
= + + +
1 100[ ( ) ]
t
0
R R At Bt
= + +
1[ ]
t
0
ASTM 1137 and IEC 60751 coefficient values
A = 3.9083 X 10–3 B = –5.775 X 10–7 C = –4.183 X 10–12
2 3
2
Example 5: Calculate the tolerance of a 0.1 ASTM Class A probe at 100 °C
Measure the reference probe temperature
Measure the indicated UUT temperature using ASTM 1137 equation and coefficients
Calculate the error 0.05 °C
Calculate the tolerance at
100.00 °C
Determine tolerance status Fail (0.05 °C > 0.03 °C)
100.00 °C
100.05 °C (Given)
0.03 °C (See example 4)
To be certain of the tolerance status of a cali­brated instrument it is necessary to have calibra­tion uncertainties that are significantly better than the tolerance of the instrument being calibrated. Typically a ratio of 4:1 or four times better than the tolerance of the instrument being calibrated is required. When this is not the case the risk may be unacceptably high that out of tolerance instruments will be falsely accepted or that in tolerance instru­ments will be falsely rejected. As the magnitude of the detected error approaches the tolerance of the calibrated instrument, the risk of incorrectly assign­ing a tolerance status increases. Guard bands may be helpful in these circumstances. For example if a guard band is 80% of the tolerance then instru­ments found within 80% of their tolerance will pass, instruments outside of the tolerance will fail and instruments that are in between will be inde­terminate. The better the calibration uncertainties the tighter the guard band can be.
Conclusion
Platinum Resistance Thermometer calibration pro­cedures are similar whether the method selected is characterization or tolerance testing. For best accuracy with modern equipment choose character­ization. For equipment that does not allow charac­terization, tolerance testing may be your only choice and this is a common situation in industry today. When conducting tolerance testing it is important to use equipment and procedures with sufficient accu­racy to determine the tolerance status confidently.
Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT) 7
8 Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT)
Fluke. Keeping your world up and running.
Fluke Corporation Hart Scientific Division
799 E Utah Valley Drive American Fork, UT 84003
Tel: 801.763.1600 Fax: 801.763.1010 E-Mail: info@hartscientific.com www.hartscientific.com
Fluke Eurpoe B. V., Hart Scientific Division
PO Box 1186, 5602 BD Eindhoven The Netherlands Tel: +31 (0)40 2675 403 Fax: +31 (0)40 2675 404 E-mail: Hart.Logistics@Fluke.NL
All other countries: Tel: +1 801.763.1600 Fax: +1 801.763.1010
©2009 Fluke Corporation. All rights reserved. Specifications subject to change without notice. 6/2009 3498460 A-EN-Rev A
®
Loading...