Fluke 5622 Application Note

How to calibrate an RTD
or Platinum Resistance
Thermometer (PRT)
Topics Covered
Calibration by characterization ITS-90 Callendar-Van Dusen Polynomial
Tolerance testing ASTM 1137 IEC 60751
Application Note
Introduction
There are two types of calibrations applicable to PRTs—characterization and tolerance testing. The type of calibration to perform is determined by the way in which the UUT is to be used and the accuracy required by the user. Characterization is the type of calibra­tion in which the unit under test (UUT) resistance is determined at several temperature points and the data are fitted to a mathematical expression. Tolerance testing on the other hand is a calibration in which the UUT resistance is compared to defined values at specific temperatures. No data fitting is performed. In the laboratory, we are required to perform both types of calibration depending upon our customer’s needs.
Calibration Procedures
Characterization
Characterization is the method that is most often used for medium to high accuracy PRT calibration. With this method, a new resistance vs. temperature relationship is deter­mined anew with each calibration. Generally, with this type of calibration, new cali­bration coefficients and a calibration table are provided as a product of the calibration. There are five basic steps to perform as listed below:
1. Place the reference probe and the UUTs in the temperature source in close proxim­ity to one another.
2. Connect the leads to the readout(s) ensuring proper 2-, 3-, or 4-wire connection.
3. Measure the reference probe and determine the temperature.
4. Measure and record the resistance of the UUT(s).
5. Fit the data.
Some readouts simplify the technique by combining or eliminating some of the steps. In the following discussion, we will consider an application involving PRT characteriza­tion by comparison to an SPRT.
Step 1: Probe Placement
All temperature sources have instabilities and gradients. These translate into calibra­tion errors and/or uncertainties. To minimize the effects, the probes should be placed as close together as practical. In baths the probes to be calibrated should be placed in a radial pattern with the reference probe in the center (focus) of the circle. This ensures an equal distance from the reference probe to each of the UUTs. In dry-well temperature sources, the reference probe and probes to be calibrated should all be placed the same distance from the center for best results, but the reference may be placed in the center if needed.
Also, the sensing elements should be on the same horizontal plane. Even though sensing elements are different lengths, having the bottoms of the probes at the same level is sufficient. Sufficient immersion must be achieved so that stem losses do not occur. Generally, sufficient immersion is achieved when the probes are immersed to a depth equal to 20 times the probe diameter plus the length of the sensing ele­ment. For example, consider a 3/16 inch diameter probe with a 1 inch long sensing element. Using the rule of thumb, 20 x 3/16 in + 1 in = 3 3/4 in + 1 in = 4 3/4 in. In this example, minimum immersion is achieved at 4 3/4 inches. This rule of thumb is generally correct with thin wall probe construction and in situations of good heat transfer. If the probe has thick wall construction and/or poor heat transfer is present (such as in the case of a dry-well with incorrectly sized holes), more immersion is required.
F r o m t h e F l u k e D i g i t a l L i b r a r y @ w w w . f l u k e . c o m / l i b r a r y
Step 2: Connection to Readout
This step is straightforward. Connections must be tight and in proper 2-, 3-, or 4-wire configuration. If using 4-wire configuration, ensure that the current and voltage connections are correct. See Figure 1.
temperature data is available in real time. Some modern readouts also display the data in graphical format, allowing the operator to determine stabil­ity at a glance. Both of these features speed up the process and eliminate possible operator error due to incorrect table interpolation.
The second method is used when the readout does not provide for proper temperature calculation. (Some readouts, particularly DMMs, have some of the more popular temperature conversions built in. These typically do not allow use of unique calibra­tion coefficients and cannot be used for accurate temperature calibration.) In this case, the resistance is measured and the temperature is determined from either a calibration table or from a computer or calculator program.
Since the temperature must be calculated after the resistance is measured, the process is slower and does not provide immediate, real time tempera­ture data. See Tables 1 and 2 below.
Table 1. Interpolation from an RTD calibration table (resistance vs. temperature).
t( °C) R(t) (W) dR/dt(t) W/°C
400 249.8820 0.3514
401 250.2335 0.3513
402 250.5848 0.3512
403 250.9360 0.3511
450 267.3108 0.3456
451 267.6564 0.3455
452 268.0019 0.3454
453 268.3472 0.3452
Figure 1. Thermometer readout connection schematics
Step 3: Measurement of Reference Probe and Temperature Determination
There are two ways to measure the reference probe and determine the temperature. Both techniques have the same potential accuracy. That is, if done correctly, neither technique is inherently more accurate than the other.
The first and best method is used with sophis­ticated readouts designed for temperature work. The resistance is measured and the temperature calculated from calibration coefficients which were entered into the readout previously. Once these calibration coefficients have been entered, the tem­perature calculations are accomplished internally and the readout displays in temperature units. The
1. Measure the reference probe resistance
2. Locate where it falls on the table
3. Subtract lower table value from measured value
4. Divide by dR/dT(t) (slope of curve)
5. Add fractional temperature to table value
249.9071 W
between 249.8820 W and
250.2335 W
249.9071 W – 249.8820 W =
0.0251 W
0.0251 / 0.3514 = 0.0714 °C
0.0714 ’C + 400 =
400.0714 °C
Tech Tip
Manual calculation is more prone to human error and is more time consum­ing than using a readout designed for temperature work.
2 Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT)
Table 2. Interpolation from an RTD calibration (resistance ratio (W) table).
t( °C) W(t) dt/dW(t)
300 2.1429223 275.2199
301 2.1465557 275.3075
302 2.1501880 275.3951
303 2.1538192 275.4827
350 2.3231801 279.6655
351 2.3267558 279.7559
352 2.3303304 279.8464
353 2.3339037 279.9369
1. Measure reference probe resistance
2. Calculate W (Rt/R = 25.54964)
3. Locate where it falls on the table
4. Subtract lower table value from measured value
5. Multiply by dt/dW(t) (inverse slope of curve)
6. 6) Add fractional temperature to table value
) (R
tpw
54.75258 W
54.75258 W / 25.54964 W =
tpw
2.1429883
between 2.1429223 and
2.1465557
2.1429883 – 2.1429223 =
0.000066
0.000066 • 275.2199 =
0.0182 °C
0.01821 °C + 300 °C =
300.0182 °C
Number of readouts - will the reference probe
•
and UUTs be measured with the same readout or different readouts?
Type of readout - a readout designed for temper-
•
ature calibration often has features which allow flexibility in the measurement scheme.
UUT characteristics - self-heating time, source
•
current requirements, stability, and overall qual­ity influence the measurement process.
It is not possible for us to anticipate all of the variables and discuss the optimum solutions here. However, in the following examples, we will con­sider some typical calibration scenarios and sug­gested measurement schemes.
Example 1: 2 DMM readouts, 1 reference probe and 5 UUTs
Step 4: Measurement of Units Under Test (UUTs)
Since the UUTs are resistance thermometers simi­lar to the reference probe, they are measured in a similar manner. If several UUTs are undergoing calibration, ensure that when they are connected or switched in, sufficient time is allowed for self­heating to occur before the data is recorded. Also, ensure that the readout is set to the correct range to provide the proper source current and to prevent range changes between the measurements at dif­ferent temperatures. Typically, the measurements are conducted starting at the highest temperature of calibration and working down. Additionally, it increases the precision of the calibration to use a mean (average) value calculated from multiple measurements at the same temperature. Often, the readout is designed with statistical features to facili­tate this practice. It is also a good practice to close the process with an additional measurement of the reference probe. The sequence in which the probes (reference and UUT) are measured is referred to as a measurement scheme. There are many variables to consider when designing a measurement scheme. Some points to consider are:
Accuracy - the higher the accuracy desired, the
•
more all of the following must be considered. Temperature source stability - the more stable
•
the source, the more time exists to conduct the measurements before temperature changes cause unwanted error.
Number of UUTs - the higher the number, the
•
longer it takes to cycle through all UUTs.
The reference probe is connected to one readout and the first UUT is connected to the second read­out. This places the probes to be measured under current at all times, thus, eliminating self-heating errors caused by changing current conditions. The UUTs will be connected and measured individually.
The scheme is as follows:
REF(1)-UUT (1) - REF(2)-UUT (2) - REF(3)-UUT (3) -
REF(4)-UUT (4) - REF(5)-UUT (5)
This provides 5 readings each of the reference and the UUT. Take the average of the readings and use it for the data fit. If the reference probe read­ings are in resistance, the temperature will have to be computed. After completion, repeat the process for the additional UUTs.
Fluke Corporation, Hart Scientific Division How to calibrate an RTD or Platinum Resistance Thermometer (PRT) 3
Loading...
+ 5 hidden pages