Fluke 50 Series II, 52 II, 53 II, 54 II Specifications

Page 1
C25 80PK-26 80PK-25 FVF-SC1 TPAK
50 Series II Thermometers
Laboratory accuracy. Wherever you go.
Additional features for the 53 and 54 Series II:
•
Data Logging up to 500 points of data with user adjustable recording interval
•
Real time clock captures the exact time of day when events occur
•
Recall function allows logged data to be easily reviewed on the meter display
•
IR communication port allows data to be exported to optional FlukeView
®
Temperature PC software
Features
Specifications
Recommended Accessories
Battery life: 1000 hours typical, AA Size (HxWxD): 17 3 x 86 x 38 mm
Weight: 0.4 kg Three Year Warranty
11080-eng Rev. 01
Included Accessories
Impact absorbing holster Two bead probe thermocouples 80PK-1 (54+52) One bead probe thermocouple 80PK-1 (51+53)
Ordering Information
Fluke 51 II Thermometer Fluke 52 II Thermometer Fluke 53 II Thermometer Fluke 54 II Thermometer FVF-SC1 FlukeView Forms
Software including interface cable
●● ● ●
●● ● ●
●● ● ●
●●
●●
●●
●●
51 I I 52 II 53 II 54 II
Thermocouple types J,K,T,E J,K,T,E J,K,T,E,N,R,S J,K,T,E,N,R,S Number of inputs Single Dual Single Dual Time stamp Relative Time Relative Time Time of Day Time of Day Splash/Dust resistant Dual display with backlight Min/Max/Avg recording (T1-T2) True differential Data logging up to 500 points IR data port for interface
to PC Compatible with optional
FlukeView Software
Temperature range:
J-type Thermocouples -210°C to 1200°C (-346°F to 2192°F) K-type Thermocouples -200°C to 1372°C (-328°F to 2501°F) T-type Thermocouples -250°C to 400°C (-418°F to 752°F) E-type Thermocouples -150°C to 1000°C (-238°F to 1832°F) N-type** Thermocouples -200°C to 1300°C (-328°F to 2372°F) R** and S-type** Thermocouples 0°C to 1767°C (32°F to 3212°F)
Temperature accuracy
Above –100°C (-148°F):
J, K, T, E, and N-type** ± [ 0.05% + 0.3°C (0.5°F) ] R** and S-type** ± [ 0.05% + 0.4°C (0.7°F) ]
Below –100°C (-148°F):
J, K, E, and N-types ± [ 0.20% + 0.3°C (0.5°F) ] T-type ± [ 0.50% + 0.3°C (0.5°F) ]
**Only the Fluke Models 53 and 54 Series II thermometers are capable of measuring N, R, and S-type thermocouples.
The Fluke 50 Series II contact thermometers offer fast response and laboratory accuracy (0.05% + 0.3°C) in a rugged handheld test tool.
•
Large backlit dual display shows any combination of T1, T2(52 and 54 only), T1-T2(52 and 54 only) plus MIN, MAX, or AVG
•
Relative time clock on MIN, MAX, and AVG provides a time reference for major events
•
Electronic Offset function allows compensation of thermocouple errors to maximize overall accuracy
•
Readout in °C, °F, or Kelvin (K)
•
Sleep mode increases battery life
•
Battery door allows easy battery replacement without breaking the calibration seal
Fluke 51 II Fluke 52 II Fluke 53 II
Fluke 54 II