1507/1503
Insulation Testers
Technical Data
The Fluke 1507 and 1503 Insulation
Testers are compact, rugged, reliable,
and easy to use. With their multiple
test voltages, they are ideal for many
troubleshooting, commissioning, and
preventative maintenance applications.
Additional features, like the remote
probe on these tools save both time
and money when performing tests.
Features and benefits:
Insulation test range:
•
7: 0.01 M
50
1
••
1503: 0.1 MΩ to 2000 MΩ
••
Insulation test voltages:
•
1507: 50 V, 100 V, 250 V,
••
500 V, 1000 V
503: 500 V, 1
1
••
Save both time and money with
•
Automatic calculation of Polarization Index and Dielectric
Absorption Ratio
Make repetitive tests simple and
•
ith the 1
easy w
(Pass/Fail) function (only available on the 1507)
0 G
Ω to 1
000 V
7’s Compare
50
Ω
Repetitive or hard-to-reach
•
testing is easy with the remote
test probe
Live circuit detection prevents
•
insulation test if voltage
> 30 V is detected for added
user protection
Auto-discharge of capac
•
voltage for added user protection
/DC voltage: 0.1 V to 600 V
C
A
•
200 mA Continuity
•
Resistance: 0.01 Ω to 20.00 KΩ
•
ve batter
Sa
•
power off
y power w
itive
ith auto
Read measurements easily w
•
large, backlit display
CAT IV 600 V overvoltage
•
category rating for added user
protection
Remote probe, test leads, probes
•
and alligator clips included w
each tester
Accepts optional Fluke TPAK
•
netic hang
mag
your hands for other work
Four AA alkaline batteries (NEDA
•
15 A or IEC LR6) for at least
1000 insulation tests
One-year warranty
•
ing system to free
ith
ith
1507/1503 Specifications
AC/DC voltage measurement
Accuracy
Range Resolution
600.0 V 0.1 V ± (2 % + 3)
Input impedance: 3 MΩ (nominal), < 100 pF
Common mode rejection ratio (1 kΩ unbalanced): > 60 dB at dc, 50 or 60 Hz
Overload protection: 600 V rms or dc
Earth bond resistance measurement
Range Resolution
20.00 Ω 0.01 Ω
200.0 Ω 0.1 Ω
2000 Ω 1 Ω
20.00 kΩ 0.01 kΩ
1
curacies apply from 0 to 100 % of range.
Ac
Overload protection: 2 V rms or dc
Open circuit test voltage: > 4.0 V, < 8 V
Short circuit current: > 200.0 mA
50 Hz to 400 Hz
± (% of Rdg + Digits)
Accuracy
+ (% of Rdg + Digits)
± (1.5 % + 3)
1
Insulation specifications
Measurement range: 0.01 MΩ to 10 GΩ model 1507, 0.01 MΩ to 2000 MΩ model 1503
Test voltages: 50 V, 100 V, 250 V, 500 V, 1000 V
Test voltage accuracy: + 20 %, - 0 %
Short-circuit test current: 1 mA nominal
Auto discharge: Discharge time < 0.5 second for C = 1 µF or less
Live circuit detection: Inhibit test if terminal voltage > 30 V prior to initialization of test
Maximum capacitive load: Operable with up to 1 µF load
Accuracy (Model 1507)
Output Voltage Display Range Resolution Test Current
50 V
(0 % to + 20 %)
100 V
(0 % to + 20 %)
250 V
(0 % to + 20 %)
500 V
(0 % to + 20 %)
1000 V
(0 % to + 20 %)
0.01 MΩ to 20.00 MΩ 0.01 MΩ
20.0 M
Ω to 50.0 MΩ 0.1 MΩ
1 M
0.0
Ω to 20.00 MΩ 0.0
20.0 MΩ to 100.0 MΩ 0.1 MΩ
0.01 MΩ to 20.00 MΩ 0.01 MΩ
20.0 MΩ to 200.0 MΩ 0.1 MΩ
Ω to 20.00 MΩ 0.0
1 M
0.0
Ω to 200.0 MΩ 0.1 MΩ 1 mA @ 500 kΩ ± (1.5 % + 5)
20.0 M
200 MΩ to 500 MΩ 1 MΩ
1 M
0.
Ω to 200.0 MΩ 0.
Ω to 2000 MΩ 1 MΩ 1 mA @ 1 MΩ
200 M
2.0 GΩ to 1
0.0 G
Ω 0.
1 M
Ω
1 M
Ω
1 M
Ω
1 G
Ω ±
1 mA @ 50 kΩ ±
1 mA @ 100 kΩ ± (3 % + 5)
1 mA @ 250 kΩ ± (1.5 % + 5)
(3 % + 5)
1.5 % + 5)
(
0 % + 3)
1
(
y
Accurac
(% of Rdg + Digits)
±
±
2 Fluke Corporation 1507/1503 Insulation Testers