Fluke 1507, 1503 Specifications

1507/1503
Insulation Testers
Technical Data
The Fluke 1507 and 1503 Insulation Testers are compact, rugged, reliable, and easy to use. With their multiple test voltages, they are ideal for many troubleshooting, commissioning, and preventative maintenance applications. Additional features, like the remote probe on these tools save both time and money when performing tests.
Features and benefits:
Insulation test range:
7: 0.01 M
50
1
1503: 0.1 MΩ to 2000 MΩ
Insulation test voltages:
1507: 50 V, 100 V, 250 V,
500 V, 1000 V
503: 500 V, 1
1
Save both time and money with
Automatic calculation of Polar­ization Index and Dielectric Absorption Ratio
Make repetitive tests simple and
ith the 1
easy w (Pass/Fail) function (only avail­able on the 1507)
0 G
to 1
000 V
7’s Compare
50
Repetitive or hard-to-reach
testing is easy with the remote test probe
Live circuit detection prevents
insulation test if voltage > 30 V is detected for added user protection
Auto-discharge of capac
voltage for added user protection
/DC voltage: 0.1 V to 600 V
C
A
200 mA Continuity
Resistance: 0.01 to 20.00 K
ve batter
Sa
power off
y power w
itive
ith auto
Read measurements easily w
large, backlit display CAT IV 600 V overvoltage
category rating for added user protection
Remote probe, test leads, probes
and alligator clips included w each tester
Accepts optional Fluke TPAK
netic hang
mag your hands for other work
Four AA alkaline batteries (NEDA
15 A or IEC LR6) for at least 1000 insulation tests
One-year warranty
ing system to free
ith
ith
1507/1503 Specifications
AC/DC voltage measurement
Accuracy
Range Resolution
600.0 V 0.1 V ± (2 % + 3)
Input impedance: 3 M(nominal), < 100 pF Common mode rejection ratio (1 kunbalanced): > 60 dB at dc, 50 or 60 Hz Overload protection: 600 V rms or dc
Earth bond resistance measurement
Range Resolution
20.00 0.01
200.0 0.1
2000 1
20.00 k 0.01 k
1
curacies apply from 0 to 100 % of range.
Ac
Overload protection: 2 V rms or dc Open circuit test voltage: > 4.0 V, < 8 V Short circuit current: > 200.0 mA
50 Hz to 400 Hz
± (% of Rdg + Digits)
Accuracy
+ (% of Rdg + Digits)
± (1.5 % + 3)
1
Insulation specifications
Measurement range: 0.01 Mto 10 Gmodel 1507, 0.01 Mto 2000 Mmodel 1503 Test voltages: 50 V, 100 V, 250 V, 500 V, 1000 V Test voltage accuracy: + 20 %, - 0 % Short-circuit test current: 1 mA nominal Auto discharge: Discharge time < 0.5 second for C = 1 µF or less Live circuit detection: Inhibit test if terminal voltage > 30 V prior to initialization of test Maximum capacitive load: Operable with up to 1 µF load
Accuracy (Model 1507)
Output Voltage Display Range Resolution Test Current
50 V
(0 % to + 20 %)
100 V
(0 % to + 20 %)
250 V
(0 % to + 20 %)
500 V
(0 % to + 20 %)
1000 V
(0 % to + 20 %)
0.01 Mto 20.00 M 0.01 M
20.0 M
to 50.0 M 0.1 M
1 M
0.0
to 20.00 M 0.0
20.0 MΩ to 100.0 M 0.1 M
0.01 Mto 20.00 M 0.01 M
20.0 Mto 200.0 M 0.1 M
to 20.00 M 0.0
1 M
0.0
to 200.0 M 0.1 M 1 mA @ 500 k ± (1.5 % + 5)
20.0 M
200 Mto 500 M 1 M
1 M
0.
to 200.0 M 0.
to 2000 M 1 M 1 mA @ 1 M
200 M
2.0 Gto 1
0.0 G
0.
1 M
1 M
1 M
1 G
±
1 mA @ 50 k ±
1 mA @ 100 k ± (3 % + 5)
1 mA @ 250 k ± (1.5 % + 5)
(3 % + 5)
1.5 % + 5)
(
0 % + 3)
1
(
y
Accurac
(% of Rdg + Digits)
±
±
2 Fluke Corporation 1507/1503 Insulation Testers