Fluke 12E User Manual

Page 1
12E
Digital Multimeter
Users Manual
®
PN 3382138 October 2008
© 2008 Fluke Corporation. All rights reserved. Printed in China. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This war­ranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke's behalf. To obtain service during the warranty period, return the unit to point of purchase with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Service Center Address: Fluke Corporation Beijing Service Center Room 2301 Scite Tower 22 Jianguomenwai Dajie, Chao Yang District Beijing 100004, PRC Tel: 010-65123435
Shanghai Shilu Instrument Co., Ltd. #139, 2638 Nong, Hongmei Nanlu Shanghai 201108 Biao Zhun Hao: Q/SXAV 1-2002
Page 3
Table of Contents
Title Page
Introduction .................................................................................................................... 1
Safety Information .......................................................................................................... 1
Safe Working Practices .................................................................................................. 1
Instrument Overview ...................................................................................................... 4
Terminals ................................................................................................................... 4
Display ....................................................................................................................... 4
Battery Saver ................................................................................................................. 5
Making Measurements ................................................................................................... 5
Manual Ranging and Auto Ranging ........................................................................... 5
Data Hold .................................................................................................................. 5
Measuring AC and DC Voltage .................................................................................. 5
Measuring AC or DC Current ..................................................................................... 6
Measuring Resistance ............................................................................................... 7
Testing for Continuity ................................................................................................. 7
Testing Diodes ........................................................................................................... 8
Measuring Capacitance ............................................................................................. 8
Maintenance ................................................................................................................... 8
General Maintenance ................................................................................................ 9
i
Page 4
12E
Users Manual
Testing the Fuses...................................................................................................... 9
Replacing the Batteries and Fuses ........................................................................... 10
Service and Parts .......................................................................................................... 10
General Specifications ................................................................................................... 11
Accuracy Specifications ................................................................................................. 12
ii
Page 5
12E
Multimeter
Introduction
Warning
To avoid electric shock or personal injury, read "Safety Information” and “Warning and
Precautions” before using the Meter.
The Fluke 12E Multimeter (hereafter referred to as “the Meter”) is a 4,000 count instrument.
The Meter is battery powered with a digital display.
Safety Information
The Meter complies with IEC 1010-1 CAT I 1000 V, CAT II 600 V, and CAT III 300 V overvoltage standards. See Specifications.
Use the Meter only as specified in this manual, otherwise the protection provided by the Meter may be impaired.
In this manual, a Warning identifies conditions and actions that pose hazards to the user.
A Caution identifies conditions and actions that may damage the Meter or the equipment under test.
International symbols used on the Meter and in this manual are explained in Table 1.
Safe Working Practices
Review the safety information and comply with the safe working practices on pages 2 and 3.
1
Page 6
12E
Users Manual
Warnings and Precautions
To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, comply with the following practices:
Before using the Meter, inspect the case. Do not use the Meter if it is damaged. Look for cracks or missing plastic. Pay
particular attention to the insulation around the connectors.
Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged
test leads before using the Meter.
Verify a Meter’s operation by measuring a known voltage. Do not use the Meter if it operates abnormally. Protection
may be impaired. When in doubt, have the Meter serviced.
Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and
earth ground.
Use caution when working with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. These voltages pose a shock
hazard.
Use the proper terminals, function, and range for your measurements.
Do not operate the Meter around explosive gas, vapor, or dust.
When using the probes, keep the fingers behind the finger guards.
When making connections, connect the common test lead before connecting the live test lead; when disconnecting,
disconnect the live test lead before disconnecting the common test lead.
Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or
capacitance.
Do not use the Meter in a manner not specified by this manual or the safety features of the Meter may be impaired.
For all dc functions, including manual or auto-ranging, to avoid the risk of shock due to possible improper reading, verify the presence of any ac voltages by first using the ac function. Then select a dc voltage range equal to or greater than the ac range.
2
Page 7
Multimeter
Safe Working Practices
 Warnings and Precautions (Continued)
Before measuring current, check the Meter’s fuses (see “Testing the Fuses”) and turn OFF power to the circuit before
connecting the Meter to the circuit.
Do not operate the Meter with the case (or part of the case) removed.
Use only two AA batteries, properly installed in the Meter case, to power the Meter.
Replace the battery as soon as the battery indicator () appears. With a low battery, the Meter might produce false
readings that can lead to electric shock and personal injury.
Do not measure voltages above 600 V in Category II, or 300 V in Category III installations.
Remove test leads from the Meter before opening the Meter case or battery door.
When servicing the Meter, use only specified replacement parts.
Table 1. International Electrical Symbols
AC (Alternating Current) Earth Ground  DC (Direct Current) Fuse
AC or DC Double Insulated
Safety Information Shock Hazard
Battery  Complies with EU directives
3
Page 8
12E
Users Manual
Instrument Overview
Terminals
C
CAT I 1000 V
CAT II 600 V
3
V
A
10A
FUSED
mA A
1
2
400mA FUSED
COM
Item Description
1 Input terminal for ac and dc current
measurement to 10 A.
2 Input terminal for ac and dc microamp and
milliamp measurement to 400 mA.
3 Common (return) terminal for all
measurements.
4 Input terminal for voltage, resistance,
continuity, diode, capacitance, measurements.
4
apg1f.eps
Display
2
1
11
Item Description
1 Relative mode is active 2 Continuity selected 3 Display Hold is enabled 4 Diode test is selected 5 F – Farads for capacitance 6 A, V – amps or volts 7 DC, AC – dc or ac voltage or current 8 Ω - Ohms is selected
9 m, M, k – decimal prefix 10 Auto range selected
11 Battery is low and should be changed
3
10
9
4
5
6
7
8
apg2f.eps
4
Page 9
Multimeter
Battery Saver
Battery Saver
The Meter enters the “Sleep mode” and blanks the display if the Meter is not used and the input is inactive for 30 minutes. Press any button or turn the rotary switch to wake the Meter up. To disable the Sleep mode, hold down the YELLOW button while turning the Meter on.
Making Measurements
Manual Ranging and Auto Ranging
The Meter has both manual and auto range options. In the auto range mode, the Meter selects the best range for the input detected. This allows you to switch test points without having to reset the range. You can override auto ranging by selecting the range manually.
The Meter defaults to the auto range mode in measurement functions that have more than one range. When the Meter is in the auto range mode, Auto Range is displayed.
To enter and exit the manual range mode:
1. Press
2. To exit the manual range mode, press and hold
.
Each press of highest range is reached, the Meter wraps to the lowest range.
for two seconds.
increments the range. When the
 Warning
Dangerous voltages may be present at the input terminals and may not be displayed.
Data Hold
To hold the present reading, press . Press again to resume normal operation.
Measuring AC and DC Voltage
To minimize improper reading of an unknown voltage containing either ac or ac + dc voltage components, first select the ac voltage function on the meter making particular note of the ac range required for a good measurement. Next manually select the dc function with the dc range that either matches or is higher than the previously noted ac range. Using this procedure minimizes the effects of ac transients while making accurate dc measurements.
1. Choose ac or dc by turning the rotary switch to , or 
2. Connect the red test lead to the black test lead to the COM terminal
3. Measure the voltage by touching the probes to the desired test points of the circuit
4. Read the measured voltage on the display.
Note
The only way to access the 400 manual ranging.
terminal and the
range is by
5
Page 10
12E
Users Manual
Volts AC Volts DC
V
+
V
_ _
+
mV
Millivolts DC
_
apg3f.eps
+
Figure 1. Measuring AC and DC Voltage
Measuring AC or DC Current
1. Turn the rotary switch to , or .
2. Toggle between ac or dc current measurement by pressing the YELLOW button.
3. Connect the red test lead to either the A, or mA terminal based on the current to be measured and connect the black test lead to the COM terminal.
4. Break the circuit path to be measured. Then connect the test leads across the break and apply power.
AC
apg4f.eps
Figure 2. Measuring AC and DC Current
5. Read the measured current on the display.
6
Page 11
Multimeter
Making Measurements
Measuring Resistance
 Warning
To avoid electrical shock or damage to the Meter when measuring resistance or continuity in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged.
1. Turn the rotary switch to . Make sure power is
disconnected from the circuit to be measured.
<50
2. Connect the red test lead to the black test lead to the COM terminal.
3. Measure the resistance by touching the probes to the desired test points of the circuit.
4. Read the measured resistance on the display.
terminal and the
Figure 3. Measuring Resistance/Continuity
apg5f.eps
Testing for Continuity
With the resistance mode selected, press the YELLOW button twice to activate the continuity beeper. If the resistance is under 50 Ω, the beeper will sound continuously, designating a short circuit. If the meter reads , the circuit is open.
7
Page 12
12E
Users Manual
Testing Diodes
 Warning
To avoid electrical shock or damage to the Meter when testing diodes in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged.
1. Turn the rotary switch to
2. Press the YELLOW function button once to activate Diode Test.
3. Connect the red test lead to the black test lead to the COM terminal.
4. Connect the red probe to the anode side and the black test lead to the cathode side of the diode being tested.
5. Read the forward bias voltage value on the display.
6. If the polarity of the test leads is reversed with diode polarity, the display reading shows . This can be used to distinguish the anode and cathode sides of a diode.
.
terminal and the
Measuring Capacitance
Caution
To avoid damage to the Meter, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance.
1. Turn the rotary switch to
2. Connect the red test lead to the terminal and the
black test lead to the COM terminal.
3. Touch the probes to the capacitor leads.
4. After allowing the reading to stabilize (up to 15 seconds), read the capacitance value on the display.
.
Maintenance
Beyond replacing batteries and fuses, do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service instructions. The recommended calibration cycle is 12 months.
8
Page 13
Multimeter
Maintenance
General Maintenance
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals
1. Turn the Meter OFF and remove the test leads.
2. Shake out any dirt that may be in the terminals.
3. Soak a new swab with isopropyl alcohol and work around the inside of each input terminal.
4. Use a new swab to apply a light coat of fine machine oil to the inside of each terminal.
Testing the Fuses
 Warning
To avoid electric shock or injury, remove the test leads and any input signals before replacing the fuses.
1. Turn the rotary switch to
2. Plug a test lead into the terminal and touch the
probe to the A or mA terminal.
A good A terminal fuse is indicated by a reading between 000.0 Ω and 000.1 Ω. A good mA terminal fuse is indicated by a reading between 0.990 kΩ and 1.010 kΩ.
If the display reads , replace the fuse and test again.
If the display shows any other value, have the Meter serviced. See “Service and Parts” later in this manual.
.
9
Page 14
12E
Users Manual
Replacing the Batteries and Fuses
 Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the battery indicator () appears.
To prevent damage or injury, install ONLY replacement fuses with the specified amperage, voltage, and interrupt ratings.
Disconnect test leads before opening the case or the battery door.
B1
F1
F2
F1 Fuse, 500 mA, 1000V Fluke PN 1989732
F2 Fuse, 10A, 1000 V Fluke PN 1989726
B1 Battery, 2 X AA Alkaline
Fluke PN 376756
NEDA 15A, IEC LR6 Battery Door Fluke PN 1884065
Service and Parts
If the Meter fails, first check the batteries and fuses, then review this manual to make sure that you are operating the Meter correctly.
To contact Fluke call:
+86-10-65123435 ext 15 in China Visit Fluke’s Web site at www.fluke.com
10
apg6f.eps
Page 15
Multimeter
g
g
g
r
General Specifications
General Specifications
Maximum Voltage between any Terminal and Earth Ground:
Fuse Protection for mA μA inputs: 500 mA, 1000 V FAST, Min. Interrupt Rating 20,000 A
Fuse Protection for A input: 10 A, 1000 V FAST, Min. Interrupt Rating 20,000 A Display: Di Temperature: Operatin Operating Altitude: 0 to 2000 meters Temperature Coefficient: 0.1 X (specified accuracy)/ °C (<18 °C or >28 °C) Electroma Rated Transient Overvoltage: 4 kV (1.2 x 50 μs) Peak for measu
Relative Humidity: Noncondensing < 10 °C
Relative Humidity, 40 MΩ Range: 80% from 10 °C to 30 °C; 70% from 30 °C to 40 °C
Battery Type: 2 X AA, NEDA 15A, IEC LR6
Battery Life: Alkaline: 500 Hours
Size (H x W x L): 180 mm x 89 mm x 51.5 mm (with holster)
Weight: 425 grams
Certifications: CMC,
Safety Compliance: IEC 61010-1, 2000 CAT I 1000 V, CAT II 600 V, and CAT III 300 V overvoltage standards
Overvoltage installation categories per IEC 61010-1, 2000: The Meter is designed to protect against transients in these catagories: CAT I CAT II CAT III
netic Compatibility: Complies with FCC Part 15, Class B, IEC 61326, 3 V/m, performance criterion B
From high-voltage low-energy sources, e.g., electronic circuits or a copy machine. From equipment supplied from the fixed installation, e.g., TVs, PCs, portable tools and household appliances. From equipment in fixed equipment installations, e.g., installation panels, feeders and short branch circuits, and lighting
systems in large buildings.
1000V
ital: 4000 count updates 3/sec
: 0 °C to 40 °C, Storage: -30 °C to 60 °C indefinitely (to -40 °C for 100 hours)
ement Categories I, II, and III.
90% from 10 °C to 30 °C; 75% from 30 °C to 40 °C
11
Page 16
12E
Users Manual
Accuracy Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0% to 75%. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits])
Function Range Resolution Accuracy
400.0 mV 1 0.1 mV 3.0% + 3
AC Volts (40 to 500 Hz)
DC Millivolts
4.000 V
40.00 V
400.0 V 1000 V
400.0 mV 0.1 mV 1.0% + 10
0.001 V
0.01 V
0.1 V 1 V
1.0% + 3
DC Volts
Diode Test 2
1. Manual Range only
2. Diode test open circuit test voltage is 1.1 V to 1.6 V and short circuit current is < 0.6 mA (typical).
4.000 V
40.00 V
400.0 V 1000 V
1.000 V 0.001 V 10 %
0.001 V
0.01 V
0.1 V 1 V
0.5% + 3
12
Page 17
Multimeter
Accuracy Specifications
Function Range Resolution Accuracy
400.0 Ω
Resistance (Ohms)
Capacitance 3
AC Current (40 to 200 Hz)

AC Current (40 to 200 Hz)

AC Current (40 to 200 Hz)
DC Current

3. Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 50 nF range).
4. When in the 4A range, display will show 4000 counts, please ignore the last digit.
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
50.00 nF
500.0 nF
5.000 μF
50.00 μF
100.0 μF
400.0 μA 4000 μA
40.00 mA
400.0 mA
4
4.000 A
10.00 A
400.0 μA 4000 μA
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
0.1 μA 1 μA
0.01 mA
0.1 mA
0.01 A
0.01 A
0.1 μA 1 μA
0.5% + 3
0.5% + 2
0.5% + 2
0.5% + 2
0.5% + 2
1.5% + 3
2% + 5 2% + 5 5% + 5 5% + 5 5% + 5
1.5% + 3
1.5% + 3
1.5% + 3
1.5% + 3
13
Page 18
12E
Users Manual
Function Range Resolution Accuracy
DC Current

DC Current
4. When in the 4A range, display will show 4000 counts, please ignore the last digit.
Function
40.00 mA
400.0 mA
4
4.000 A
10.00 A
Overload
Protection
AC Volts 1000 V >10 MΩ <100 pF
DC Volts 1000 V >10 MΩ <100 pF
106 V Hz Max
0.01 mA
0.1 mA
0.01 A
0.01 A
Input Impedance
(Nominal)
1.5% + 3
1.5% + 3
Common Mode Rejection Ratio
>60 dB at dc,
50 or 60 Hz
>100 dB at dc,
50 or 60 Hz
14