Fluke 12B/18 Instruction Sheet

Page 1
®
12B/18
MultiMeter
Instruction Sheet
To ensure that the meter is used safely, follow these instructions:
Do not use the meter if the meter or test leads appear damaged, or if
Disconnect the live test lead before disconnecting the common test
When using the probes, keep your fingers behind the finger guards on
Do not use the Model 12B’s Automatic Selection mode to measure
Turn off power to the circuit under test before cutting, desoldering, or
Do not apply more than 600V rms between a meter terminal and earth
Read First: Safety Information
you suspect that the meter is not operating properly.
lead.
the probes.
voltages in circuits that could be damaged by this mode’s low input impedance (2 kΩ).
breaking the circuit. Small amounts of current can be dangerous.
ground.
Use caution when working with voltages above 60V dc or 30V ac rms. Such voltages pose a shock hazard.
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator (U) appears.
PN 602000 June 1996 Rev.3, 1/99
© 1996,1998,1999 Fluke Corporation, All rights reserved. Printed in U.S.A.
WWarning
Page 2
Symbols
riešenia na presné meranie
http://www.elso.sk
RANGE
VDC VAC
Press button.
Press button to switch between modes.
Double insulation.
Not applicable to identified model.
Manual ranging mode.
ag14i.eps
Automatic Range Selection
The meter defaults to autoranging when you turn it on.
Manual Range Selection
M
RANGE
RANGE
To return to autoranging, press for 2 seconds, or change the measurement
CAT III
COM
600V
+
mode.
RANGE
ag10i.eps
Standby Mode
If the meter is ON but inactive and not connected to voltage for more than 45 minutes, the display goes blank to preserve battery life. To resume operation, press any button or turn the rotary switch.
Page 3
DC and AC Voltage
riešenia na presné meranie
http://www.elso.sk
For Model 12B, refer also to Automatic Selection.
Volts dc Volts ac
Input Impedance 10 M Input Impedance 5 M
VDC VAC
DC AC
V
600V
COM
+
Continuity and Resistance
For Model 12B, refer also to Automatic Selection.
Continuity Resistance
VDC VAC
COM
V
600V
+
ag01i.eps
Mk
LOW IMPEDANCE
AUTOMATIC SELECTION
CAT III
600V
COM
+
<25
Detects shorts and opens 250 µS.
12B Only
LOW IMPEDANCE LOW IMPEDANCE
AUTOMATIC SELECTION
CAT III
600V
COM
LOW IMPEDANCE
AUTOMATIC SELECTION
COM
AUTOMATIC SELECTION
CAT III
600V
++
ag02i.eps
Page 4
Diode Test
riešenia na presné meranie
http://www.elso.sk
Good Diode
VDC VAC
LOW IMPEDANCE
AUTOMATIC SELECTION
CAT III
600V
COM
Typical voltage for silicon diode
12B Only
LOW IMPEDANCE
AUTOMATIC SELECTION
LOW IMPEDANCE
AUTOMATIC SELECTION
CAT III
+
COM
600V
+
Single beep
ag03i.eps
Bad Diode
Forward Bias Reverse Bias
12B Only
LOW IMPEDANCE
AUTOMATIC SELECTION
LOW IMPEDANCE
COM
AUTOMATIC SELECTION
CAT III
600V
+
COM
AUTOMATIC SELECTION
CAT III
600V
+
or
Shorted Open
ag04i.eps
Page 5
LIMITED WARRANTY & LIMITATION O F LIABILITY
riešenia na presné meranie
http://www.elso.sk
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is two years for Model 12B and three years for Model 18 and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Fluke’s opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that soft war e will oper ate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that softwar e will be er r o r f ree or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the product will be ret urned to Buyer, transportation prepaid (FO B Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormal condition of operation or handling, Fluke will provide an estimate of repair cost s and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the r epair and r et ur n t r anspor tation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, I NCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FI TNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, I NCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not aff ect t he validity or enforceability of any other provision.
Fluke Corporation Fluke Europe B.V. P.O. Box 9090 P.O. Box 1186 Everett WA 5602 B.D. Eindhoven 98206-9090 The Netherlands
5/94
Page 6
Continuity Capture
riešenia na presné meranie
http://www.elso.sk
(Open/Short Capture on Model 18)
To set up the meter to capture intermittent shorts and opens, turn the switch to J, connect the leads to the circuit; then press
M
M
M.
M
M
ag12i.eps
Captures transitions longer than 250 µS. Transitions after the first transition cause the meter to beep, but the
display does not change. To reset the display to the current condition, press To exit, press
M for 2 seconds, or change the measurement mode.
M.
Disabling the Beeper
To disable the beeper for all modes, hold down r for 2 seconds while
turning the meter on.
Page 7
MIN MAX
riešenia na presné meranie
http://www.elso.sk
µF
(Records the lowest and highest measurements)
Automatic Selection (12B only), autoranging, and standby are disabled. Put the meter in the proper range before entering MIN MAX.
When the reading changes more than about 50 digits, the meter gives a short beep. When a new minimum or maximum is recorded, the meter gives a long beep.
Enter
MIN MAX
MIN
MAX
Present Reading with
Input Change Alert
M
MIN
MAX
TM
MIN
MAX
MIN
MAX
MIN
MAX
Maximum Reading
M
MAX
M
MIN
Minimum Reading
2 sec
To Exit
ag06i.eps
Page 8
MIN MAX with Elapsed Time
riešenia na presné meranie
http://www.elso.sk
µF
Records the hours and minutes between when MIN MAX was entered and the last high and low was recorded. OL is displayed for times longer than 99:59.
To enable the MIN MAX timer, hold down
M while turning the rotary
switch from OFF to either measurement mode.
Enter MIN MAX. View
Maximum Reading Maximum Reading
Present Reading
M
MIN
MAX
MAX
M
MIN
MAX
Elapsed Time
M
MAX
MIN
MAX
MIN MIN
MIN
MAX
Minimum Reading
Elapsed Time
MM
MIN
MAX
MIN
MAX
2 sec
Minimum Reading Exit MIN MAX
Page 9
Capacitance
riešenia na presné meranie
http://www.elso.sk
Turn off circuit power; then disconnect and discharge the capacitor before measuring capacitance.
F
µ
If the capacitor requires more discharging, diSC is displayed while the capacitor discharges.
12B Only
LOW IMPEDANCE
AUTOMATIC SELECTION
CAT III
600V
COM
+
+
Note correct probe polarity for polarized capacitors.
LOW IMPEDANCE
AUTOMATIC SELECTION
ag05i.eps
Page 10
Automatic Selection
riešenia na presné meranie
http://www.elso.sk
18
M
µF
(12B only)
If a dc or ac voltage greater than about 4.5V is present across the inputs
when you are attempting to measure continuity or resistance, the meter switches to dc or ac voltage mode.
W Warning
Repetitive transients on a dc bus will cause
k
CHEK
to select ac volts, even though a hazardous dc voltage may be present. To avoid a misleading display and possible electric shock, manually select the proper volts function for measurements on these circuits.
When dc or ac voltage is automatically selected, the meter has low input impedance (2 k). This low impedance, which places a moderate load on the circuit under test, is appropriate only for measuring power supply voltages under load. Do not use Automatic Selection to measure voltage in circuits that could be damaged by a 2 k load.
If AC detected If DC detected
Z
Lo
Locked-Mode Functions
RANGE
Disables Automatic Selection and locks
Meter puts 2 k
meter in selected mode.
load on circuit.
LOW IMPEDANCE
AUTOMATIC SELECTION
CAT III
600V
COM
+
or
RANGE
or
VDC VAC
or
: :
RANGE
RANGE
0.000 00.00 000.0 0000 2 sec
Reenables Automatic Selection
ag07i.eps
Page 11
Maintenance
riešenia na presné meranie
http://www.elso.sk
Clean the case with a damp cloth and detergent. Do not use abrasives or solvents.
Battery Replacement
Remove the test leads before disassembling the case.
COM
600V
+
+
ag13i.eps
Page 12
Test Lead Replacement
riešenia na presné meranie
http://www.elso.sk
Y
Replace the test leads with Fluke TL-75 PN 855705 double-insulated leads.
Service and Parts
This meter should be serviced only by a qualified service technician. To contact Fluke, call one of the following telephone numbers:
USA and Canada: 1-888-99-FLUKE (1-888-993-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-356-5500
Or, visit Fluke’s Web site at www.fluke.com.
Page 13
Specifications
riešenia na presné meranie
http://www.elso.sk
Maximum Voltage Between any Terminal and Earth Ground
Display 3 3/4-digits, 4000 counts, updates 4/sec Operating Temperature -10°C to 50°C (14°F to 122°F) Storage Temperature -30°C to 60°C (-22°F to 140°F)
Temperature Coefficient 0.1 x (specified accuracy)/°C
Relative Humidity 0% to 90% (-10°C to 35°C; 14°F to 95°F)
Battery Type 9V, NEDA 1604 or IEC 6F22 Battey Life 650 continuous hours with alkaline
600V rms
indefinitely (to -40°C (-40°F) for 100 hrs)
(<18°C or >28°C; <64°F or > 82°F)
0% to 70% (35°C to 50°C; 95°F to 122°F)
450 continuous hours with carbon-zinc
Shock, Vibration 1 meter shock. Per MIL-T-28800D for a
Class 3 instrument
Size (HxWxL) 3.46 cm x 7.05 cm x 14.23 cm
(1.35 in x 2.75 in x 5.55 in) Weight 286g (10 oz) Safety Designed to Protection Class II
requirement of UL3111, ANSI/ISA-S82,
CSA C22.2 No 231, and VDE 0411, and
IEC 1010 overvoltage Category III (CAT
III). EMI Regulations Complies with FCC Part 15,Class B, and
VDE 0871B.
s
Trademark of TÜV Product Services.
Complies with EN 61010-1: 1993.
Page 14
Accuracy is specified for a period of one year after calibration, at 18°C to
riešenia na presné meranie
http://www.elso.sk
28°C (64°F to 82°F) with relative humidity to 90%. AC conversions are ac­coupled, average responding, and calibrated to the rms value of a sine wave input. Accuracy specifications are given as follows:
±([% of reading] + [number of least significant digits])
Function Range Resolutio
1
1
1 mV
0.001V
00.01V
000.1V 0001V
1 mV
0.001V
00.01V
000.1V 0001V
0.1
0.001 k
0.01 k
f
E
J
4000 mV
4.000V
40.00V
400.00V 0600V
4000 mV
4.000V
40.00V
400.00V 0600V
400.0
4.000 k
40.00 k
Accuracy (50 to 400 hz)
n
±(1.9% + 3) ±(1.9% + 3) ±(1.9% + 3) ±(1.9% + 3) ±(1.9% + 3)
±(0.9% + 2) ±(0.9% + 2) ±(0.9% + 1) ±(0.9% + 1) ±(0.9% + 1)
±(0.9% + 2) ±(0.9% + 1) ±(0.9% + 1)
|
400.0 k
4.000 M
40.00 M
1.000 µF
10.00 µF
100.0 µF
10000 µF
0.1 k
0.001 M
0.01 M
0.001 µF
0.01 µF
0.1 µF 1 µF
±(0.9% + 1) ±(0.9% + 1) ±(1.5% + 3)
±(1.9% + 2) ±(1.9% + 2) ±(1.9% + 2)
1000 µF ±(1.9% + 2)
>1000 µF ±(10% + 90)
typical
2
ML
2.000V
0.001V
±(0.9% + 2)
1. The 4000 mV range can be entered only in manual range mode. Use the 4000 mV range with accessories.
2. The beeper is guaranteed to come on at <25 and turn off at >250Ω. The meter detects opens or shorts 250 µs.
Page 15
Function Overload
riešenia na presné meranie
http://www.elso.sk
Protection
Input Impedance (Nominal)
1
f
E
f
E
600V rms
600V rms
600V rms
600V rms
>5 M <100 pF Automatic Selection and LoZ = >2 k
<200 pF (ac coupled)
2
>10M <100 pF Automatic Selection and LoZ = >2 k
<200 pF
Common Mode Rejection Ratio
2
Normal Mode
Rejection
(1 k
Unbalanced)
>60 dB at dc 50 or 60 Hz
>100 dB at dc, 50 or 60 Hz
Open Circuit
>50 dB at 50 Hz or 60 Hz
Full Scale Voltage
J
L
600V rms
600V rms
Test Voltage
<1.5V dc <450 mVdc<1.5V dc
2.4-3.0V dc 2.400V dc
To 4.0 M 40 M
Short Circuit Current
J
L
600V rms <500 µA 600V rms 0.95 mA (typical)
1. 3 x 106 V Hz maximum
2. 2 k input impedance up to 50V. Impedance increases with input voltage to >300 k at 600V.
Page 16
MIN MAX Recording Accuracy and Response Time
riešenia na presné meranie
http://www.elso.sk
Specified accuracy of the measurement function ±12 digits for changes >200 ms in duration (±40 digits in ac). Typical 100 ms response to 80%.
MIN MAX Recording with Elapsed Time
Elapsed Time Resolution Accuracy
0 to 100 hours
(99:59)
1 minute 0.3% typical