MC608A/B
Electromagnetic Transmitter
Installation, Operation and Maintenance Manual
TM-100467, Rev P
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 2
Thank you for selecting a Flow Technology, Inc. product for your flow measurement
application.
Virtually every major commercial, government, and scientific organization is making use of our
products, expertise and extensive technical support. This is a culmination of years of
refinement in our flow meter and calibrator designs, which has resulted in the technological
leadership in the flow measurements field that we enjoy.
We are proud of our quality products, our courteous service and welcome you, as a valued
customer, to our growing family.
The specifications contained in this manual are subject to change
without notice and any user of these specifications should verify from
the manufacturer that the specifications are currently in effect.
Otherwise, the manufacturer assumes no responsibility for the use of
specifications, which have been changed, and are no longer in effect.
TM-100467 REVISIONS
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 4
PUBLISHED BY FLOW TECHNOLOGY, INC. – March, 2016
WARRANTY
Limited Warranty
Seller warrants that goods delivered hereunder will at delivery be free from defects in materials and workmanship
and will conform to seller's operating specifications. Seller makes no other warranties, express or implied, and
specifically makes NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
Limitation of Liability
Seller's obligation under the warranty shall be limited to replacing or repairing at Seller's option, the defective
goods within twelve (12) months from the date of shipment, or eighteen (18) months from the date of shipment
for destination outside of the United States, provided that Buyer gives Seller proper notice of any defect or failure
and satisfactory proof thereof. Defective goods must be returned to Seller's plant or to a designated Seller's
service center for inspection. Buyer will prepay all freight charges to return any products to Seller's plant, or other
facility designated by Seller. Seller will deliver replacements for defective goods to Buyer freight prepaid. The
warranty on said replacements shall be limited to the unexpired portion of the original warranty. Goods returned
to Seller for which Seller provides replacement under the above warranty shall become the property of the Seller.
The limited warranty does not apply to failures caused by mishandling or misapplication. Seller's warranty
obligations shall not apply to any goods that (a) are normally consumed in operation or (b) have a normal life
inherently shorter than the warranty period stated herein. In the event that goods are altered or repaired by the
Buyer without prior written approval by the Seller, all warranties are void. Equipment and accessories not
manufactured by Seller are warranted only to the extent of and by the original manufacturer's warranty. Repair or
replacement goods furnished pursuant to the above warranty shall remain under warranty only for the unexpired
portion of the original warranty period.
Should Seller fail to manufacture or deliver goods other than standard products appearing in Seller's catalog,
Seller's exclusive liability and Buyer's exclusive remedy shall be release of the Buyer from the obligation to pay
purchase price therefore.
THE FORGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES WHETHER ORAL, WRITTEN, EXPRESSED,
IMPLIED OR STATUTORY. IMPLIED WARRANTIES OF FITNESS AND MERCHANTABILITY SHALL NOT APPLY SELLER'S
WARRANTY OBLIGATIONS AND BUYER'S REMEDIES THEREUNDER (EXCEPT AS TO TITLE) ARE SOLELY AND
EXCLUSIVELY AS STATED HEREIN. IN NO CASE WILL SELLER BE LIABLE FOR SPECIAL, INCIDENTAL OR
CONSEQUENTIAL DAMAGE.
The total liability of Seller (including its subcontractors) on any claim whether in contract, tort (including
negligence whether sole or concurrent) or otherwise, arising out of or connected with, or resulting from the
manufacture, sales, delivery, resale, repair, replacement or use of any goods or the furnishing of any service
hereunder shall not exceed the price allocable to the product or service or part thereof which gives rise to the
claim.
This manual provides information and guidance for personnel responsible for the installation,
operation, and maintenance of the MC608A/B Electromagnetic Flow Meter Transmitter
supplied by Flow Technology, Inc.
1.2 Purpose
The contents of this manual are for general information and to describe the operational
characteristics of the MC608A/B transmitter. This manual does not include instructions for
special application or factory repair.
1.3 Description
An Electromagnetic Flow Meter works off the principal of Faraday’s Law. In this
application of the principal the flow meter creates an electromagnetic field perpendicular
to the fluid flowing through the meter housing. An electromotive force (voltage) is
generated in a conductive medium when it passed through a magnetic field. The voltage
created is directly proportional to the density of the magnetic field and the velocity of the
conductive medium. Since the conductive medium is the liquid whose flow is to be
measured, and the length between the electrodes and the electromagnetic field strength
is known, therefore the signal generated is proportional to the velocity of the liquid in the
pipe.
In order for this technology to operate correctly the fluid to be measured does need to be
conductive. The minimum fluid conductivity required is 5 µS/cm for most fluids and 20
µS/cm for DI water.
The electromagnetic flow meters are composed of:
The sensor that must be mounted between two portions of pipe by flanges, threaded
joints, or triclamp fittings.
The converter which drives the sensor and displays or transmits data.
The MC 608A/B converter is provided in a cylindrical aluminum instrument enclosure and
can be interfaced to all Flow Technology magnetic flow meters. It can be mounted directly
on top of the flow meter or remote from the flow meter connected through a pair of
cables, whose maximum length depends on the conductivity of the liquid.
The flow meter Ka and Kb coefficients are established during the factory flow meter
calibration. These factors are normally programmed into the MC 608A/B converter at the
factory prior to shipment, however the key pad of the converter allows the user to enter
or modify these coefficients. This key pad also allows the user to set the full scale flow
rate, define the scaled pulse output, adjust alarm set points, and adjust dosage volume.
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 9
The MC608 is available in two main versions:
MC608A – Mains powered 90…264Vac or 12/24 Vac/dc
MC608B – Battery powered or 12/24Vac/dc
2 INSTALLATION
2.1 Inspection
Verify that all parts listed on the packing list are included with your shipment. Note that
there may be more than one package for your shipment. Please check the factory set up
sheet and verify the meter size, full scale flow rate, and analog/pulse output are set up as
required.
2.2 Mechanical Installation of the transmitter to the flow meter
2.2.1 Compact (mounted on flow meter)
The MC 608A/B converter is supplied mounted directly to the flow meter junction box.
In this configuration all interconnection wiring between the flow meter sensor and the
converter have been completed at the factory. Upon flow meter installation, external
wiring for power input and signal outputs will need to be completed. For optimum
installation flexibility the converter display is capable of rotating up to 180˚.
Figure 1 – MC608A – Mains Powered Unit
Figure 2 – MC608B – Battery Powered Unit
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 10
2.2.2 Separate (remote mounted from flow meter)
The MC 608A and MC 608B converters can be mounted up to a maximum of 328 feet
(100 meters) and 98 feet (30 meters) respectively from the flow meter, dependent on
fluid conductivity. When supplied for remote mounting, the flow meter can optionally
be purchased with various cable lengths for interconnection to the converter. This
interconnection cable can be supplied attached to and resin sealed in the flow meter
junction box or the cables can be supplied separately and attached and sealed in the
field. For optimum installation flexibility the converter display is capable of rotating up
to 180˚.
Figure 3 – Remote mounting
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 11
2.2.3 Maximum cable length
When the converter is to be mounted remote from the flow meter, the maximum cable
distance is limited by the conductivity of the fluid to be measured. The maximum
distance is 328 feet (100 meters). Below is a chart to determine maximum cable length
based on the installation operating conditions.
Figure 4 – Cable length vs. conductivity
2.2.4 Grounding
The sensor and converter must be properly grounded to earth potential for the system
to operate correctly. Earth potential is the reference parameter for the flow meter and
without proper grounding issues with zero offset, or flow indication at zero flow will
occur. The sensors and converters all have earth grounding points that can be used to
insure a direct contact with earth ground.
The Flow Technology electromagnetic flow meters are equipment with a third ground
electrode that eliminates the need for grounding rings when installing the flow meter in
insulating piping.
2.3 Electrical Installation
This section provides the professional installer with information for connecting the
MC608A/B to the user's system.
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 12
Verify that the power is off before connecting or servicing!
Electrical connection of the device must be carried out by properly trained
2.3.1 MC608A/B terminal locations
Figure 5- High voltage: 90 – 264 Vac (PCBA up to version 6-2)
WARNING:
personnel!
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 13
Figure 6 – Low voltage: 12 – 24 VDC/Vac (PCBA up to version 6-2)
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 14
Figure 7- Low and high voltage (PCBA version 6-3)
SF-100834 Rev A
TM-100467, Rev P
MC608A/B/P Transmitter Manual
Page 15
Figure 8- Low and high voltage (PCBA from version 6-4)
Note: Line (L) and Neutral (N) terminals are not polarity sensitive when used with DC
power.
2.3.2 Remote mounted converter, without empty pipe detector
The remote mounted version of the MC 608A/B converter is supplied with an auxiliary
junction box where cables, C012 and C013 from the flow meter junction box will need to
be connected. The converter is normally supplied scaled for a specific flow meter. Please
check the program sheet supplied with the equipment to ensure the converter is
connected to the proper flow meter. Cables C012 and C013 are supplied with metal
ends and identifying letters. Refer to the diagram below for correct terminal location.
SF-100834 Rev A