FLOW Technology FT Series Installation, Operation And Maintenance Manual

4250 EAST BROADWAY ROAD PHOENIX, ARIZONA 85040 U.S.A.
TELEPHONE (602) 437-1315 FAX (602) 437-4459
FT SERIES
and Maintenance Manual
SERIAL NUMBER_________________________________
The specifications contained in this manual are subject to change without notice and any user of these specifications should verify from the manufacturer that the specifications are currently in effect. Otherwise, the manufacturer assumes no responsibility for the use of specifications that have been changed and are no longer in effect.
Installation, Operation
FT SERIES
TURBINE FLOWMETERS Installation, Operation and Maintenance Manual
TM-86675 REV. U
PUBLISHED BY FLOW TECHNOLOGY, INC. – April 2004
Thank you for selecting a FLOW TECHNOLOGY, INC. product for your flow measurement application.
Virtually every major commercial, government, and scientific organization is making use of our products, expertise and extensive technical support. This is a culmination of years of refinement in our flowmeter and calibrator designs, which has resulted in the technological leadership in the flow measurements field, which we enjoy.
We are proud of our quality products, our courteous service and welcome you, as a valued customer, to our growing family.
i
WARRANTY
Limited Warranty. Seller warrants that goods delivered hereunder will at delivery be free from defects in materials and workmanship and will conform to seller's operating specifications. Seller makes no other warranties, express or implied, and specifically makes NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
Limitation of Liability. under the warranty shall be limited to replacing or repairing at Seller's option, the defective goods within twelve (12) months from the date of shipment, or eighteen (18) months from the date of shipment for destination outside of the United States, provided that Buyer gives Seller proper notice of any defect or failure and satisfactory proof thereof. Defective goods must be returned to Seller's plant or to a designated Seller's service center for inspection. Buyer will prepay all freight charges to return any products to Seller's plant, or other facility designated by Seller. Seller will deliver replacements for defective goods to Buyer freight prepaid. The warranty on said replacements shall be limited to the unexpired portion of the original warranty. Goods returned to Seller for which Seller provides replacement under the above warranty shall become the property of the Seller.
The limited warranty does not apply to failures caused by mishandling or misapplication. Seller's warranty obligations shall not apply to any goods, which (a) are normally consumed in operation or (b) have a normal life inherently shorter than the warranty period stated herein.
Seller's obligation
In the event that goods are altered or repaired by the Buyer without prior written approval by the Seller, all warranties are void. Equipment and accessories not manufactured by Seller are warranted only to the extent of and by the original manufacturer's warranty. Repair or replacement goods furnished pursuant to the above warranty shall remain under warranty only for the unexpired portion of the original warranty period.
Should Seller fail to manufacture or deliver goods other than standard products appearing in Seller's catalog, Seller's exclusive liability and Buyer's exclusive remedy shall be release of the Buyer from the obligation to pay purchase price therefor.
THE FORGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES WHETHER ORAL, WRITTEN, EXPRESSED, IMPLIED OR STATUTORY. IMPLIED WARRANTIES OF FITNESS AND MERCHANTABILITY SHALL NOT APPLY SELLER'S WARRANTY OBLIGATIONS AND BUYER'S REMEDIES THEREUNDER (EXCEPT AS TO TITLE) ARE SOLELY AND EXCLUSIVELY AS STATED HEREIN. IN NO CASE WILL SELLER BE LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGE.
The total liability of Seller (including its subcontractors) on any claim whether in contract, tort (including negligence whether sole or concurrent) or otherwise, arising out of or connected with, or resulting from the manufacture, sales, delivery, resale, repair, replacement or use of any goods or the furnishing of any service hereunder shall not exceed the price allocable to the product or service or part thereof which gives rise to the claim.
ii
TM-86675
REVISIONS
DATE REVISION ECO NUMBER APPROVAL
A
B
C
D
E
12/22/92 F 10950 T. Roy
07/20/95 G 11278 M. Wusterbarth
06/13/96 H 12451 T. Roy
03/11/97 J 12724, 12744,12752 E. Knowles
08/20/97 K 13001 E. Knowles
02/09/97 L 13195 E. Knowles
06/13/98 M 13319 E. Knowles
05/03/99 N 14098 T Roy
3/11/02 P 15677 J Blasius
11/14/02 R 16185 J Blasius
5/8/03 S 16532 J Blasius
7/21/03 T 16700 J Blasius
4/5/04 U 17286 J Blasius
iii
TABLE OF CONTENTS
SECTION TITLE PAGE
1.0 INTRODUCTION 1
2.0 STANDARD LINE FLOWMETER 1
3.0 INSPECTION UPON RECEIPT 2
4.0 MECHANICAL CONNECTIONS 2
4.1 FLOW CONDITIONING 2
4.2 FLOW PULSATIONS 3
4.3 PURGING 3
4.4 INSTALLATION RECOMMENDATION 3
4.5 ORIENTATION AND CALIBRATION 3
4.6 FILTRATION 4
4.7 TORQUE REQUIREMENTS 4
5.0 PICKOFFS 4
5.1 INSTALLATION 4
5.2 EXPLOSION PROOF HOUSINGS 5
5.2.1 PICKOFF INSTALLATION SOCKETS 5
5.3 MAGNETIC PICKOFF 5
5.4 RF PICKOFF 6
6.0 ELECTRICAL CONNECTIONS 6
6.1 CONNECTIONS 6
6.2 CONNECTION CABLE 6
6.3 GROUNDING CONSIDERATIONS 6
6.4 SIGNAL PROCESSING 7
7.0 BIDIRECTIONAL FLOWMETERS 7
8.0 OPERATION 8
8.1 OVER RANGE 8
8.2 UNDER RANGE 8
8.3 TURBINE FLOWMETER LIQUID CHARACTERISTICS 8
8.3.1 INTRODUCTION 8
8.3.2 STANDARD CALIBRATION 8
8.3.3 SINGLE VISCOSITY CALIBRATIONS 9
8.3.4 MULTIPLE VISCOSITY CALIBRATIONS 9
8.4 TURBINE FLOWMETER GAS CHARACTERISTICS 9
8.4.1 INTRODUCTION 9
8.4.2 AIR CALIBRATION 10
8.4.3 SINGLE PRESSURE CALIBRATION 10
8.4.4 MULTIPLE PRESSURE CALIBRATIONS 11
9.0 SPECIFICATIONS AND OPTIONS 11
9.1 END FITTINGS 13
iv
TABLE OF CONTENTS (Continued)
SECTION TITLE PAGE
9.2 CALIBRATION 13
9.3 CONSTRUCTION MATERIALS 14
9.4 BEARINGS 14
9.5 PICKOFFS 15
10.0 PERIODIC MAINTENANCE 21
10.1 INSPECTION CLEANING AND STORAGE 21
10.2 REMOVING INTERNALS 22
10.3 GENERAL HANDLING TECHNIQUES 23
10.4 BALL BEARING REPLACEMENT 24
10.4.1 LIQUID FT4-8, AND GAS FT-10, FT-12 25
10.4.2 LIQUID FT6-8, FT8-8, FT-08, FT-10, FT-12, FT-16, FT-20, FT-24 GAS FT2-8, FT4-8, FT6-8, FT8-8, FT-08, FT-10, FT-12, FT-16, FT-20, FT-24
10.4.3 LIQUID AND GAS FT-32 26
10.4.4 LIQUID AND GAS FT-40, FT-48, FT-64 27
10.4.5 LIQUID AND GAS FT-96 28
10.4.6 LIQUID AND GAS FT128 29
10.4.7 LIQUID AND GAS FT192 30
10.5 JOURNAL BEARING REPLACEMENT 31
10.5.0.1 REPLACEMENT KITS 31
10.5.0.2 SELF-LUBRICATING BEARINGS 31
10.5.0.3 CARBIDE AND CERAMIC 31
10.5.1 FT-24 AND SMALLER GRAPHITE 32
10.5.2 FT-24 AND SMALLER CERAMIC AND TUNSTEN CARBIDE 33
10.5.3 FT-32 GRAPHITE 34
10.5.4 FT-32 CERAMIC AND TUNGSTEN CARBIDE 35
10.5.5 FT-40, FT48, FT64 CERAMIC AND TUNGSTEN CARBIDE 36
10.5.6 FT-96, FT128 TUNGSTEN CARBIDE 37
10.5.7 FT-192 TUNGSTEN CARBIDE 38
11.0 TROUBLESHOOTING GUIDE 39
12.0 PARTS LIST 44
25
v
TABLE OF CONTENTS (TABLES GUIDE) TABLE # TABLE TITLE PAGE TABLE 1 TORQUE REQUIREMENTS 4 TABLE 2 FT SERIES FLOWMETER MODEL NUMBERING SYSTEM 12 TABLE 3 BEARING APPLICATION GUIDE 15 TABLE 4 LIQUID SERVICE BALL BEARING 16 TABLE 5 LIQUID SERVICE JOURNAL BEARING 17 TABLE 6 GAS SERVICE BALL BEARING H CODE 18 TABLE 7 GAS SERVICE BALL BEARING A CODE 19 TABLE 8 METER READS HIGH 39 TABLE 9 METER READS LOW 40 TABLE 10 ZERO OUTPUT 41 TABLE 11 INTERMITTENT OPERATION 42 TABLE 12 NON-REPEAT METER OUTPUT 43 TABLE 13 CONSTANT NON-ZERO OUTPUT 44 TABLE 14 LIQUID SERVICE FLOWMETER PARTS LIST FT4-8, FT6-8 45 TABLE 15 LIQUID SERVICE FLOWMETER PARTS LIST FT8-8, FT-08 46 TABLE 16 LIQUID SERVICE FLOWMETER PARTS LIST FT-10, FT-12 47 TABLE 17 LIQUID SERVICE FLOWMETER PARTS LIST FT-16, FT-20 48 TABLE 18 LIQUID SERVICE FLOWMETER PARTS LIST FT-24, FT-32 49 TABLE 19 LIQUID SERVICE FLOWMETER PARTS LIST FT-40, FT-48 50 TABLE 20 LIQUID SERVICE FLOWMETER PARTS LIST FT-64, FT-96 51 TABLE 21 LIQUID SERVICE FLOWMETER PARTS LIST FT-128, FT-192 52 TABLE 22 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT2-8, FT4-8 53 TABLE 23 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT6-8, FT8-8 54 TABLE 24 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT-08 55 TABLE 25 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT-10, FT-12 56 TABLE 26 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT-16, FT-20 57 TABLE 27 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT-24, FT-32 58 TABLE 28 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT-40, FT-48 59 TABLE 29 GAS SERVICE FLOWMETER PARTS LIST (CODE H) FT-64 60 TABLE 30 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT2-8, FT4-8 61 TABLE 31 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT6-8, FT8-8 62 TABLE 32 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT-08 63 TABLE 33 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT-10, FT-12 64 TABLE 34 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT-16, FT-20 65 TABLE 35 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT-24, FT-32 66 TABLE 36 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT-40, FT-48 67 TABLE 37 GAS SERVICE FLOWMETER PARTS LIST (CODE A) FT-64, FT-96 68
vi
TABLE OF CONTENTS (FIGURES GUIDE) FIGURE # TITLE PAGE FIGURE 1 FLOWMETER BASIC PARTS 1 FIGURE 2 3-VALVE BYPASS MANIFOLD PIPE SCHEMATIC 2 FIGURE 3 PICKOFF INSTALLATION SOCKETS 5 FIGURE 4 DIMENSIONS 20 FIGURE 5 BEARING SEATING 23 FIGURE 6 ILLUSTRATED PARTS BREAKDOWN 69 FIGURE 7 LIQUID AND GAS BEARING (CODE A) FT2-8, FT4-8 70 FIGURE 8 LIQUID AND GAS BEARING (CODE A) FT6-8, FT8-8, FT-08 70 FIGURE 9 GAS BEARING (CODE H) FT2-8, FT8-8, FT6-8, FT8-8 FT-08 71 FIGURE 10 LIQUID AND GAS BEARING (CODE A) FT-10, THRU FT-24 71 FIGURE 11 GAS BEARING (CODE H) FT-10, FT-12 72 FIGURE 12 GAS BEARING (CODE H) FT-16, FT-20, FT-24 72 FIGURE 13 LIQUID AND GAS BEARING (CODE A) AND GAS (CODE H)
FT-32
FIGURE 14 LIQUID BEARING (CODE A) AND GAS (CODE A & H) FT-32,
FT-40, FT-48, FT-64 FIGURE 15 LIQUID AND GAS BEARING (CODE A) FT-96 74 FIGURE 16 LIQUID AND GAS BEARING (CODE A) FT-128 74 FIGURE 17 LIQUID AND GAS BEARING (CODE A) FT-192 75 FIGURE 18 LIQUID BEARING (CODE G & D) FT4-8 THRU FT-08 75 FIGURE 19 LIQUID BEARING (CODE G & D) FT-32 76 FIGURE 20 LIQUID BEARING (CODE E) FT4-8 THRU FT-08 76 FIGURE 21 LIQUID BEARING (CODE E) FT-16, FT-20, FT-24 77 FIGURE 22 LIQUID BEARING (CODE G & D) FT-32 77 FIGURE 23 LIQUID BEARING (CODE E) FT-32 78 FIGURE 24 LIQUID BEARING (CODE G & D) FT-40, FT-48, FT-64 78 FIGURE 25 LIQUID BEARING (CODE G & D) FT-96 79 FIGURE 26 LIQUID BEARING (CODE G & D) FT-128 79 FIGURE 27 LIQUID BEARING (CODE G & D) FT-192 80
73
73
vii
1.0 INTRODUCTION
This manual provides information and guidance for the installation, operation and maintenance of the Standard Line Turbine Flowmeters, manufactured by Flow Technology, Inc., Phoenix, Arizona.
2.0 STANDARD LINE FLOWMETER
The Flow Technology, Inc. Standard Line Turbine Flowmeter is a volumetric flow measuring instrument. The flow sensing element is a freely suspended, bladed rotor positioned axially in the flow stream with the flowing fluid pushing against the blades. The rotational speed of the rotor is proportional to the velocity of the fluid. Since the flow passage is fixed, the turbine rotors rotational speed is also a true representation of the volume of fluid flowing through the flowmeter. The rotation of the turbine rotor generates electrical pulses in the pickoff that is attached to the flowmeter housing in close proximity to the turning rotor. Each one of these pulses represents a discrete volume of fluid. The frequency or pulse repetition rate represents the volumetric flow rate and the accumulated pulse total represents the total volume measured. Meters provided for liquid applications are not interchangeable with meters provided for gas applications. All requests for information concerning a specific meter should contain the flowmeter model number and the flowmeter serial number.
The Standard Line Flowmeter consists of 3 basic assemblies. (See Figure 1)
Figure 1. Flowmeter Basic Parts
TM-86675 1
3.0 INSPECTION UPON RECEIPT
The flowmeter should be unpacked carefully and inspected to verify that no damage occurred during shipment. Make certain that the internal parts are clean and free from packing materials or debris.
C A U T I O N
The flowmeter is a precision instrument and
may be damaged if pressurized air
is used for cleaning the flowmeter or for
checking the rotation of the rotor.
4.0 MECHANICAL CONNECTIONS
4.1 Flow Conditioning
The turbine flowmeter is sensitive to velocity profile disturbances in the flow stream. For optimizing the velocity profile it is recommended that a straight run of constant diameter piping with length of at least 10 diameters upstream of the meter and at least 5 diameters downstream be provided. (See Figure 2) The upstream section should have straightening vanes or other flow conditioners. The presence of major flow disturbance generators such as pumps, valves or elbows may require longer straight sections. If swirl is present in the line ahead of the flowmeter installation, a longer straight section or additional flow conditioning may be required. Flow Technology, Inc. provides an array of innovative state of the art flow conditioners. Figure 2. 3-Valve Bypass Manifold Pipe Schematic
TM-86675 2
4.2 Flow Pulsations
Piping and system components should be arranged to minimize pulsations entering the turbine meter. Pulsations may cause the meter to read high, and excessive pulsations may cause permanent bearing damage. Pulsations should be kept below 10% of the current flow rate at the meter location.
C A U T I O N
Pressure should be built up gradually at start-up to avoid possible damage by
over-speeding the rotor. Any severe water hammering from improper start-up or
flow surges during operation must be avoided to prevent over-speeding, shaft or
rotor blade breakage.
NOTE
Water hammering is a term used during start-up (introducing fluid into the
piping) to describe a high velocity flow impact on the turbine rotor. This must be
avoided to prevent damage to the mechanical parts.
4.3 Purging
All flow lines in the meter system should be purged prior to installation of the meter. This will remove pipe dope, metal shavings, slag and debris that may damage the turbine meter. Control valves should be located downstream from the turbine meter. (See Figure 2) System start-ups with upstream control valves in an unfilled system can result in a hydraulic shock on the meter, causing damage and a change in calibration in liquid systems, or can cause over speed conditions in gas meter systems.
4.4 Installation Recommendation
For liquid flowmeters, it is recommended that the flowmeter be installed so that it remains full of fluid when the flow ceases. When the flowmeter is left installed in a line that is temporarily out of service and has been partially or fully drained, severe bearing corrosion may occur. The type and corrosiveness of the fluid being metered, the type of bearing used in the flowmeter and the length of time the line will be out of service are factors which may affect the life and operation of the flowmeters. If it is economically feasible and conditions permit, the flowmeter should be removed, cleaned and stored when there is any doubt about the fluid level in the line during these out-of-service periods. See section 10.1 inspection cleaning and storage.
4.5 Orientation and Calibration
The orientation of the turbine flowmeter will influence the nature of the load on the rotor bearings, and thus, the performance of the meter at low flow rates. For optimum accuracy a turbine meter should be installed in the same orientation in which it was calibrated. Standard calibration orientation is with the meter axis horizontal.
TM-86675 3
4.6 Filtration
A filter should be installed upstream of the flowmeter. (See Figure 2) For 1/2 inch flowmeters, a 10 micron nominal filter should be used; for 3/4 inch and 1 inch flowmeters, a 20 micron nominal filter should be installed; for flowmeters 1-1/2 inch in diameter or larger, a 50 micron filter is recommended.
4.7 Torque Requirements
The following table provides the recommended torques in pound-feet for tightening MS-33656 flared­tube end fittings:
TABLE 1
TORQUE REQUIREMENTS
POUND-FEET
SIZE ALUMINUM TUBING STEEL TUBING
MIN. MAX. MIN. MAX.
½” 19 21 37 40
5/8” 27 30 54 58
¾” 35 40 75 85
1” 41 58 100 116 1-1/4” 66 75 126 140 1-1/2” 66 75 158 175
2” 150 166 221 245
5.0 PICKOFFS
5.1 Installation
Pickoffs should bottom in the well of the flowmeter housing but should only be finger tightened to approximately 4 lb-in (4500 gm-cm max) to prevent distortion of the coil housing. The pickoff is secured in position by tightening the lock nut to approximately 25 lb-in (30000 gm-cm). The pickoff is removed by loosening the hex lock nut and unscrewing the pickoff from the housing.
C A U T I O N Meter pressure ratings are established with a pickoff installed. Do not operate a flowmeter under pressure
without a pickoff installed.
5.2 Explosion Proof Housings
Flowmeters with explosion proof housings may have the pickoff installed inside a short section of conduit pipe (spud) that is welded to the housing. Since the spud is longer than the pickoff, the pickoff cannot be finger tightened directly and an alternate method must be used to install the pickoff. A modified 11/16-inch deep socket is required to screw the pickoff into the housing and a modified 13/16­inch deep socket is used to tighten the lock nut.
TM-86675 4
5.2.1 Pickoff Installation Sockets
Modify a standard 11/16-inch deep socket with an elongated 3/8-inch hole as shown in Figure 3. This socket will fit the pickoff and allow easy feed through of the lead terminals without damaging the leads. Modify a standard 13/16-inch deep socket with an elongated hole. (See Figure 3) Cut the outside diameter of the socket to 1.015 inches as shown to permit the socket to tighten the lock nut inside the conduit without damaging the pickoff leads.
Figure 3. Pickoff Installation Sockets
5.3 Magnetic Pickoff
The magnetic pickoff output is a low level signal that ranges from 10 mV to several volts peak-to-peak. A pulse amplifier may be needed to convert the pickoff low level signal to a 10 V peak-to-peak pulse signal suitable for process instrumentation. Typical resistance of magnetic pickoffs are 2275 ± 20%.
5.4 RF Pickoff
The modulated carrier (RF) pickoff must be installed with an appropriate amplifier (consult factory). The amplifier is needed to convert the modulated carrier signal to a 10 V peak-to-peak pulse signal suitable for process instrumentation. Typical resistance of modulated carrier pickoffs is 10 ± 10%.
TM-86675 5
6.0 ELECTRICAL CONNECTIONS
6.1 Connections
Standard pickoffs are available with a two-contact type MS3102A-10SL-4P connector or with threaded body and pigtail connectors.
6.2 Connection Cable
The connecting cable between the flowmeter and the electronic instrumentation should be a two conductor, 22 AWG, shielded and twisted cable with a vinyl jacket (Belden 8761 or equivalent). The cable should not be installed in a conduit or tray containing power lines, or close to strong electromagnetic sources such as electric lines, electric motors, transformers, welding machines, or high voltage lines. These sources may induce transient electrical noise in the coil and cause false pulse signals. Connections from standard pickoffs are not polarized and may be connected in either position. For non-standard pickoffs please refer to manufacturer's specifications.
6.3 Grounding Considerations
The shield of the cable is to be grounded at only one point in accordance with the instruction of the display instrument. Flow Technology, Inc. display instruments specify where the shield is to be grounded.
TM-86675 6
6.4 Signal Processing
An electronic signal conditioning circuit is required to either convert the frequency output of the flowmeter into a visual presentation on a display or to provide process control signals. Flow Technology, Inc. manufactures a complete line of electronic packages, which include rate and total displays, rate converters and microprocessor based units for linearization and temperature / pressure compensation.
Pickoff Connector Specifications
Standard 2 Pin MS Non-polarized Pins 27-31386 3 Pin MS Pins: A = Coil B = Coil C = N/C
Inductance = 0.350 mh ± 10 % Resistance = 3.5 ohms ± 10 % Amplified Pickoffs 27-94057
Pick-off w/RTD RF 27-62730 Mag 27-62731
3 Pin MS
Wire Leads
4 Pin MS
Wire Leads
Pins: A = Power B = Ground C = Pulse
Red = Power Black = Ground White = Pulse
Input Power = 8 to 32 VDC @ 10 ma
Output = 0 to 5 VDC Pulse
Output Impedance = 2.2 K ohms
Mag Amp: Frequency Range = 10 Hz to 10 KHz
Input Sensitivity = 20 MV p-p
RF Amp: Frequency Range = 10 to 3200 Hz
Oscillator Carrier Frequency = 45 KHz
Pins: A=Coil B=Coil Non-polarized C= RTD high D= RTD low & RTD compensation Leads: White= Coil Non-polarized Red= RTD high Black= RTD low & RTD compensation
RTD= 100 Ohm Platinum
7.0 BIDIRECTIONAL FLOWMETERS
Turbine flowmeters can be configured to measure flow in both directions and provide direction-sensing capability. This is accomplished by adding a second pickoff located with respect to the first pickoff in such a way as to achieve a 90-phase shift. The location of the pickoffs are determined by the number of blades on the rotor.
TM-86675 7
8.0 OPERATION
8.1 Over Range
In general, turbine flowmeters remain quite linear when they are over ranged, and may not provide any indication that the instrument is being misused. However, the pressure drop will become excessive and over speeding of the bearings could cause permanent damage. Bearings may also be damaged by excessive downstream thrust load. The probability of an over speed condition for a liquid meter usually occurs during system start up when there is still air in the lines. Air should be bled carefully from the lines before high flow range is established. The flow rate or output frequency should be monitored to insure maximum capability is not exceeded. Flow Technologies’s specifications should be consulted for specific maximum operating flow rates. See tables 4,5,6 and 7. Under extreme conditions, the maximum operating flow rates can be exceeded for brief periods of time without meter damage. Following are the maximum allowable over range capabilities:
Bearing Type Liquid Gas Ball 50% 10% Pivot 10% 10% Journal 50% N/A
8.2 Under Range
When used below the minimum specified range, turbine meters may become very non-linear. The repeatability of the meter may also be reduced due to bearing and magnetic pickoff drag.
8.3 Liquid Turbine Flowmeter Characteristics
8.3.1 Introduction
Optimum performance of a turbine meter system depends upon a valid calibration as well as the correct selection of supporting equipment. The rotational speed of a turbine rotor depends upon fluid properties as well as the fluid velocity. The most significant fluid property for a liquid meter is the kinematic viscosity. As liquid viscosity increases, the slip of the turbine rotor due to viscous drag is increased, and the rotational speed and hence pick-off frequency is decreased. Due to these effects, the kinematic viscosity of the calibration fluid should approximate the service conditions as closely as possible.
8.3.2 Standard Calibration
Standard liquid calibrations at FTI are done with MIL-C-7024 Type II solvent or water at room temperature. The viscosity of these fluids is approximately 1.25 and 1 centistokes respectively. The standard calibration consists of 10 data points distributed over the normal 10:1 range of the flowmeter. If viscosities or flow ranges other than these are required, they must be specified.
TM-86675 8
8.3.3 Single Viscosity Calibration
If the flowmeter is to be used at a viscosity other than the standard calibration viscosities, an oil blend calibration should be done on the meter to simulate the operating conditions. The calibration curve produced will represent the flowmeter’s output characteristics for that specific viscosity. If the flowmeter is used with liquids having viscosities greater than 3 centistokes, the linearity of the K-factor will be reduced.
8.3.4 Multiple Viscosity Calibrations
If the viscosity is changing due to varying temperature in the system, the performance characteristics over a range the viscosities can be established. This is done by performing multiple calibrations at different viscosities to cover the range of interest. The K-factor of the meter is then plotted as a function of Hz/v. The K-factor is the number of pulses generated by a flowmeter for every unit volume of fluid passing through it. Hz is the output frequency of the meter and v is the kinematic viscosity of the fluid in centistokes. The plot of K vs. Hz/v is commonly referred to as a universal viscosity curve. The data is plotted in this manner because all points fall on a single smooth curve. To obtain a useful curve, calibration points for calibrations at several viscosities are required. By observing the output frequency of the flowmeter and obtaining the viscosity or temperature of the fluid, the value of Hz/v can be calculated. Using the universal viscosity curve, the value of K corresponding to the known value of Hz/v can be determined. With the K-factor known, the flow through the meter can be determined from the expression:
Hz (60)
GPM = K - factor
8.4 Gas Turbine Flowmeter Characteristics
8.4.1 Introduction
Accurate performance of gas turbine flowmeters depend on a valid calibration that simulates the conditions the meter will operate in. Changes in the pressure and temperature of a gas directly affect the density and kinematic viscosity of the fluid. These changing fluid properties affect the performance of gas flowmeters in much the same way as liquid flowmeters.
TM-86675 9
Due to the nature of gasses to be compressed, the volume of gas measured is dependent on the pressure and temperature as established by Boyle's Law and Charles' Law. Using these relationships, the actual volume of gas measured can be related to a standard set of conditions that provide useful technical data. The standard conditions for pressure and temperature used at Flow Technology, Inc. are 14.7 pounds per square inch absolute (PSIA) and 520 degrees Rankine (60 degrees Fahrenheit) respectively. For proper conversions, the absolute measurement for pressure and temperature must be used.
The following equation is used to convert the actual volumetric flow rate (QA) in Actual Cubic Feet per Minute (ACFM) to the equivalent standard flow rate (QS) in Standard Cubic Feet per Minute (SCFM).
P Q
= Q A ( ---- ) ( ---- )
S
P
T S
A
T A
S
Where: QS = Standard flow rate in SCFM QA = Actual measured flow rate in ACFM PA = Actual measured pressure at pressure tap on meter (PSIA) PS = Standard Pressure (14.7 PSIA) TA = Actual measured temperature downstream of meter ( R) TS = Standard temperature (520 R)
8.4.2 Air Calibration
Gas calibrations at FTI are done using air at ambient conditions. These conditions are typically 14.2 PSIA and 72 F. The standard calibration consists of 10 data points distributed over the normal 10:1 range of the flowmeter. If conditions or flow ranges other than these are required, they must be specified.
8.4.3 Single Pressure Calibrations
If the flowmeter is to be used at conditions that vary significantly from the standard calibration, a calibration at equivalent conditions should be performed to simulate the actual operating conditions. The calibration curve produced will represent the flowmeter's output characteristics for the specified gas at the actual operating conditions.
TM-86675 10
8.4.4 Multiple Pressure Calibrations
Performance characteristics over a range of pressures can be established for operating conditions where the temperature and pressure are changing. This is done by performing multiple calibrations at different pressures to cover the range of interest. The information obtained is then plotted on a curve of K-factor verses Hz/v where the K-factor is the pulses per unit volume generated by the flowmeter and the Hz/v is the frequency output of the flowmeter divided by the kinematic viscosity of the fluid. The procedure is similar to liquid meters using the universal viscosity curve. By reading the output frequency of the flowmeter during operation and dividing it by the kinematic viscosity of the fluid being used, the volumetric flow rate can be determined by reading the K-factor from the universal viscosity curve and calculating the flow rate:
Hz (60) CFM = K - factor
Electronic instrumentation is available that can be programmed with the universal viscosity curve and setup with pressure and temperature transducers that will automatically make these calculations and read out the corrected flow rate.
9.0 SPECIFICATIONS AND OPTIONS
Table 2 shows the complete model numbering system for the FT series flowmeters. The sections that follow describe the contents of Table 2 in detail.
Tables 3 thru 7 describe the different bearings and their applications.
TM-86675 11
TABLE 2 FT MODEL NUMBERING SYSTEM FT _ _ _ _ _ _ _ _ _ _ __ _ _ _ _ _
BASE MODEL 2-8 Through 192 ( 1/2" Through 12" )
END FITTINGS Refer to Section 9.1
CALIBRATION Refer to Section 9.2
NON STANDARD UNITS U = To signify required units of measure other than GPM or ACFM R = To signify special calibration flow range other than normal 10:1 or extended range B = to signify both changes in units and special flow range
SERVICE L = Liquid G = Gas P = Gas with pressure tap
CONSTRUCTION MATERIALS Refer to Section 9.3
BEARINGS A = Ball Bearing D = Carbide Journal E = Graphite Journal G = Ceramic Journal H = Ball Bearing Self Lubricating (FT64 & smaller gas meters) Refer to Section 9.4
PICKOFFS Refer to Section 9.5
SPECIAL CODE Special meter configurations will not be addressed in this manual please contact the factory for a description of the special code.
TM-86675 12
9.1 End Fittings
AE = AN (or MS) external straight BE = British Standard external pipe threads 1/2" to 4" NE = NPT external threads, 1/2" to 6" HB = Hose Barb WF = Wafer type serrated surface C1 = 150# Raised Face Flange C2 = 300# Raised Face Flange C3 = 600# Raised Face Flange C4 = 900# Raised Face Flange C5 = 1500# Raised Face Flange C6 = 2500# Raised Face Flange J1 = 150# Ring Joint Flange J2 = 300# Ring Joint Flange J3 = 600# Ring Joint Flange J4 = 900# Ring Joint Flange J5 = 1500# Ring Joint Flange J6 = 2500# Ring Joint Flange D1 = DIN Flange PN16 D2 = DIN Flange PN40 D3 = DIN Flange PN100 D4 = DIN Flange PN160 D5 = DIN Flange PN250 D6 = DIN Flange PN400
9.2 Calibration
KA = 3 Point, K-factor average, in Air KW = 3 Point, K-factor average, in Water KS = 3 Point, K-factor average, in Solvent KB = 3 Point, K-factor average, in Oil Blend NA = 10 Point, normal 10:1 range, in Air NW = 10 Point, normal 10:1 range, in Water NS = 10 Point, normal 10:1 range, in Solvent NB = 10 Point, normal 10:1 range, in Oil blend XA = 10 Point, extended range, in Air XW = 10 Point, extended range, in Water XS = 10 Point, extended range, in Solvent XB = 10 Point, extended range, in Oil blend TA = 20 Point, normal 10:1 range, in Air TW = 20 Point, normal 10:1 range, in Water TS = 20 Point, normal 10:1 range, in Solvent TB = 20 Point, normal 10:1 range, in Oil blend YA = 20 Point, extended range, in Air YW = 20 Point, extended range, in Water
TM-86675 13
YS = 20 Point, extended range, in Solvent YB = 20 Point, extended range, in Oil blend GA = 30 Point, extended range, in Air GW = 30 Point, extended range, in Water GS = 30 Point, extended range, in Solvent GB = 30 Point, extended range, in Oil blend U2 = Universal Viscosity Curve, 2 viscosities U3 = Universal Viscosity Curve, 3 viscosities R1 = Reynolds # Calibration, 10 Point 1 Pressure R2 = Reynolds # Calibration, 10 Point 2 Pressure R3 = Reynolds # Calibration, 10 Point 3 Pressure E1 = Reynolds # Calibration, 20 Point 1 Pressure E2 = Reynolds # Calibration, 20 Point 2 Pressure E3 = Reynolds # Calibration, 20 Point 3 Pressure LW = 10 Point 10:1 Range, in Water, Premium linearity LS = 10 Point 10:1 Range, in Solvent, Premium linearity BA = Bidirectional, 1 Pickoff, 10 points each direction, Air BW = Bidirectional, 1 Pickoff, 10 points each direction, Water BS = Bidirectional, 1 Pickoff, 10 points each direction, Solvent BB = Bidirectional, 1 Pickoff, 10 points each direction, Oil blend CA = Bidirectional, 2 Pickoff, 10 points each direction, Air CW = Bidirectional, 2 Pickoff, 10 points each direction, Water CS = Bidirectional, 2 Pickoff, 10 points each direction, Solvent CB = Bidirectional, 2 Pickoff, 10 points each direction, Oil blend
9.3 Construction Materials
C = 304 Housing, 430F Rotor D = 304 Housing, 17-4 Rotor E = 316 Housing, 430F Rotor G = 316 Housing, 316 Rotor H = 316 Housing, 17-4 Rotor N = Hast C Housing, Hast C Rotor Q = PVC Housing, PVC Rotor R = Monel 400 Housing, Monel 400 Rotor T = Carp 20 Housing, Carp 20 Rotor
9.4 Bearing Code
A = 440C Ball Bearings. D = Carbide Journal - Carbide Shaft and Bearing. E = Graphite Journal - 316 SST Shaft and Graphite Bearing G = Ceramic Journal - Ceramic Shaft and Bearings H = 440C Ball Bearings. (Polymer retainer )
TM-86675 14
Table 3
BEARING APPLICATION GUIDE
CODE BEARING TYPE SERVICE BEARING
TEMPERATURE
RATING
A BALL LIQUID OR GAS -450° F TO 300° F 440C SST
MATERIAL
D CARBIDE
JOURNAL
E GRAPHITE
JOURNAL
G CERAMIC
JOURNAL
H BALL LIQUID OR GAS -450° F TO 300° F 440C SST POLYMER
LIQUID UP TO 1200° F C-2 CARBIDE
LIQUID UP TO 500° F 316 SHAFT GRAPH. BRG.
LIQUID UP TO 1200° F ALUMINUM BASED
CERAMIC
RETAINER
Meter temperature rating may be limited by the pickoff temperature rating
9.5 Pickoffs
-1 = RF MS Connector 400 ° F, 27-31199-101
-2 = MAG MS connector 400° F, 27-30880-101
-3 = MAG Explosion Proof 400° F, 27-30880-102 leads / 27-30931-102 EP
-5 = RF Explosion Proof 400° F, 27-31199-102 leads / 27-31949-101 EP
-6 = MAG MS connector 750°F, 27-80666-104
-7 = MAG Explosion Proof 750° F, 27-80666-104 MS / 27-82333-102 EP
-8 = RF MS Cox Equivalent 400° F, 27-84097-102
-9 = RF MS 5/8 Thd. 400° F, 27-84097-101 S8 = RF F & P 400° F, 27-31386-101
-L = RF MS connector 750° F, 27-88628-102
-M = RF Explosion Proof 750° F, 27-88826-103
-Y = RF Explosion Proof (CSA) 400° F, 27-13869-101
-Z = MAG Explosion Proof (CSA) 400° F, 27-13868-101
-U = MAG MS (Factory Mutual) 400° F, 27-32400-101
-X = RF MS (Factory Mutual) 300° F, 27-32404-101 PP = MAG Leads (Factory Mutual) 400
° F, 27-32400-103
SS = RF Leads (Factory Mutual 300° F, 27-32404-103 TT = MAG Explosion Proof (Factory Mutual) 400° F, 27-32400-102 XX = RF Explosion Proof (Factory Mutual) 300 A1 = RF MS Amplified 230
° F, 27-61313-104
° F, 27-32404-102
A2 = MAG MS Amplified 230° F, 27-61313-101 A3 = MAG Explosion Proof 230
° F, 27-61313-102
A4 = MAG Leads Amplified 230° F, 27-61313-103 A5 = RF Explosion Proof Amplified 230 A6 = RF Leads Amplified 230
° F, 27-61313-106
° F, 27-61313-105
TM-86675 15
TABLE 4
LIQUID SERVICE - BALL BEARING
MODEL
FT4-8 .25-2.5 .03-3 .1-3 48000 2000
FT6-8 .5-5 .05-5 .12-5 25000 2100
FT8-8 .75-7.5 .08-8 .16-8 16000 2000
FT-08 1-10 .1-10 .2-10 12000 2000
FT-10 1.25 - 12.5 .15-15 .3-15 9600 2000
FT-12 2-20 .25-25 .5-25 5400 1800
FT-16 5-50 .6-60 1-60 2400 2000
FT-20 9-90 1-100 1-100 1300 1950
STD RANGE 10:1 (GPM) RF & MAG PICKOFF
EXTENDED RANGE (GPM) RF PICKOFF
EXTENDED RANGE (GPM) MAG PICKOFF
NOMINAL K FACTOR PULSE/ GALLON
MAX FREQ. (HZ)
FT-24 15-150 1.6-160 2.5-160 600 1500
FT-32 22-225 2.5-250 3.5-250 350 1300
FT-40 40-400 4.5-450 5.0-450 180 1200
FT-48 65-650 N/A 7.5-750 75 812
FT-64 125-1250 N/A 15-1500 30 625
FT-96 300-3000 N/A 50-3500 28 1400
FT128 550-5500 N/A 60-6000 14 1300
FT160 850-8500 N/A 100-10000 8.5 1200
FT192 1200-12000 N/A 1500-15000 5.0 1000
Notes:
1. Repeatability
2. Linearity = ± .5% of reading except as noted. Values are valid for viscosities of 3 centistokes or less based on standard 10:1 range.
3. RF Pickoff not applicable for meter sizes FT-48 and larger.
4. Linearity = ± .75% of reading for FT4-8 and FT6-8 when high temp mag pickoff is used.
5. Premium linearity available over select ranges consult factory.
= ± .05%
TM-86675 16
TABLE 5
LIQUID SERVICE - JOURNAL BEARING
MODEL
FT4-8 .25-2.5 .1-3.0 .12-3.0 48000 2000
FT6-8 .5-5.0 .15-5.0 .2-5.0 25000 2000
FT8-8 .75-7.5 .2-8.0 .25-8.0 16000 2000
FT-08 1.0-10.0 .25-10 .3-10 12000 2000
FT-10 1.25 - 12.5 .3-15 .4-15 9600 2000
FT-12 2-20 .5-25 .5-25 6000 2000
FT-16 5-50 1.0-60 1.0-60 2400 2000
FT-20 9-90 1.0-100 1.5-100 1300 1950
STD RANGE 10:1 (GPM) RF & MAG PICKOFF
EXTENDED RANGE (GPM) RF PICKOFF
EXTENDED RANGE (GPM) MAG PICKOFF
NOMINAL K FACTOR PULSE/ GALLON
MAX FREQ. (HZ)
FT-24 15-150 1.6-160 2.5-160 600 1500
FT-32 22-220 2.5-250 3.5-250 350 1300
FT-40 40-400 4.5-450 5.0-450 180 1200
FT-48 65-650 N/A 7.5-750 75 812
FT-64 125-1250 N/A 15-1500 30 625
FT-96 300-3000 N/A 50-3500 28 1400
FT128 550-5500 N/A 60-6000 14 1300
FT160 850-8500 N/A 100-10000 8.5 1200
FT192 1200-12000 N/A 150-15000 5.0 1000
Notes:
1. Repeatability = +/- .1% for FT-12 and smaller, +/- .05% for FT-16 and larger.
2. Linearity = +/- .5% of reading except as noted. Values are valid for viscosities of 3 centistokes or less based upon standard 10:1 range.
3. Linearity is +/- 2% of reading for FT4-8.
4. Linearity is +/- .75% of reading for FT6-8 when high temp mag pickoff is used
5. Premium linearity available over select ranges consult factory.
6. RF pickoff not applicable for meter sizes FT-48 and larger.
TM-86675 17
TABLE 6
GAS SERVICE - BALL BEARING “H” CODE
MODEL
FT2-8 .1-1.0 .09-1.25 N/A 114000 1900
FT4-8 .25-2.5 .2-3.0 N/A 40800 1700
FT6-8 .5-5.0 .25-5.0 .4-5.0 24000 2000
FT8-8 .75-7.5 .4-8.0 .5-8.0 16000 2000
FT-08 1.0-10.0 .5-10 .75-10 12000 2000
FT-10 1.25 - 12.5 .6-15 1-15 9600 2000
FT-12 2-20 1-25 1.5-25 6000 2000
FT-16 5-50 1.5-60 2.5-60 2400 2000
STD RANGE 10:1 (ACFM) RF & MAG PICKOFF
EXTENDED RANGE (ACFM) RF PICKOFF
EXTENDED RANGE (ACFM) MAG PICKOFF
NOMINAL K FACTOR PULSE / ACF
MAX FREQ. (HZ)
FT-20 9-90 2.5-100 5-100 1300 1950
FT-24 15-150 4-160 6-160 600 1500
FT-32 22-220 5-250 8-250 350 1300
FT-40 40-400 9-450 10-450 180 1200
FT-48 65-650 N/A 15-750 75 812
FT-64 125-1250 N/A 30-1500 30 625
Notes:
1. Gas service - Air @ 14.7 PSIA and 60 F.
2. Repeatability = +/- .1%.
3. Linearity = +/- 1.0% of full scale based on standard 10:1 range.
4. MAG Pickoff not applicable for FT2-8.
5. MAG Pickoff range is .35 - 2.5 ACFM for FT4-8.
6. Linearity for FT2-8 is 4% of full scale, repeatability is +/- .5%.
7. Linearity for FT4-8 is 3% of full scale, repeatability is +/- .3%.
8. Linearity for FT6-8 is 1.5% of full scale, repeatability is +/- .15%.
9. RF Pickoff not applicable for meter sizes FT-48 and larger.
TM-86675 18
TABLE 7
GAS SERVICE - BALL BEARING “A” CODE
MODEL
FT2-8 .1-1.0 .09-1.25 N/A 93000 1550
FT4-8 .25-2.5 .2-3.0 N/A 45000 2000
FT6-8 .5-5.0 .25-5.0 .4-5.0 24000 2000
FT8-8 .75-7.5 .4-8.0 .5-8.0 16000 2000
FT-08 1.0-10.0 .5-10 .75-10 12000 2000
FT-10 1.25 - 12.5 .6-15 1-15 9600 2000
FT-12 2-20 1-25 1.5-25 6000 2000
FT-16 5-50 1.5-60 2.5-60 2400 2000
STD RANGE 10:1 (ACFM) RF & MAG PICKOFF
EXTENDED RANGE (ACFM) RF PICKOFF
EXTENDED RANGE (ACFM) MAG PICKOFF
NOMINAL K FACTOR PULSE / ACF
MAX FREQ. (HZ)
FT-20 9-90 2.5-100 5-100 1300 1950
FT-24 15-150 4-160 6-160 600 1500
FT-32 22-220 5-250 8-250 350 1300
FT-40 40-400 9-450 10-450 180 1200
FT-48 65-650 N/A 15-750 75 812
FT-64 125-1250 N/A 30-1500 30 625
FT-96 300-3000 N/A 70-3500 9.0 467
FT128 550-5500 N/A 120-6000 4.0 325
FT160 850-8500 N/A 200-10000 2.0 240
FT192 1200-12000 N/A 300-15000 1.0 167
Notes:
1. Gas service - Air @ 14.7 PSIA and 60 F.
2. Repeatability = +/- .1%.
3. Linearity = +/- 1.0% of full scale based on standard 10:1 range.
4. MAG Pickoff not applicable for FT2-8.
5. MAG Pickoff range is .35 - 2.5 ACFM for FT4-8.
6. Linearity for FT2-8 is 4% of full scale.
7. RF Pickoff not applicable for meter sizes FT-48 and larger.
TM-86675 19
Loading...
+ 61 hidden pages