FlowLine LC30-1001, LC30-1051, LC30-1052, LC30-1002 Owner's Manual

Warranty, Service & Repair
WARRANTY
Flowline warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service for a period which is equal to the shorter of one year from the date of purchase of such products or two years from the date of manufacture of such products.
This warranty covers only those components of the products which are non-moving and not subject to normal wear. Moreover, products which are modified or altered, and electrical cables which are cut to length during installation are not covered by this warranty.
Flowline’s obligation under this warranty is solely and exclusively limited to the repair or replacement, at Flowline’s option, of the prod­ucts (or components thereof) which Flowline’s examination proves to its satisfaction to be defective. FLOWLINE SHALL HAVE NO OBLIGATION FOR CONSEQUENTIAL DAMAGES TO PERSON­AL OR REAL PROPERTY, OR FOR INJURY TO ANY PERSON.
not apply to products which are damaged during ship­ment back to Flowline’s factory or designated service center or are returned without the original casing on the products. Moreover, this warranty becomes immediately null and void if anyone other than ser­vice personnel authorized by Flowline attempts to repair the defective products.
a. There is evidence of a potentially hazardous material present
with product.
b. The product has remained unclaimed at Flowline for longer than
30 days after dutifully requesting disposition of the product.
THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE OF THIS WARRANTY. This warranty and the obligations and liabilities of Flowline under it are exclusive and instead of, and the original purchaser hereby waives, all other remedies, warranties, guarantees or liabilities, express or implied.
EXCLUDED FROM THIS WARRANTY IS THE IMPLIED WARRANTY OF FITNESS OF THE PRODUCTS FOR A PARTIC­ULAR PURPOSE OR USE AND THE IMPLIED WARRANTY OF MERCHANT ABILITY OF THE PRODUCTS.
This warranty may not be extended, altered or varied except by a writ­ten instrument signed by a duly-authorized officer of Flowline, Inc.
To register your product with the manufacturer, go to the Flowline website for on-line registration. The website address is as follows:
www.flowline.com
On-line Warranty Registration can be found under Contact Flowline on the Navigation Bar along the side of the home page.
If for some reason your product must be returned for factory service, go to the Flowline website listed above. Online Factory Service can be found under Contact Flowline on the Navigation Bar along the side of the home page. Click on Return Authorization to begin the registration process. You will need
the following information at the time of registration:
1. Part Number and full Serial Number from product
2. Name and telephone number of someone who can answer technical questions related to the product and its application.
3. Return Shipping Address
4. Brief Description of the Symptom
5. Brief Description of the Application
Once you have received a Material Return Authorization number, ship the product prepaid in its original packing to:
Flowline Factory Service MRA _____ 10500 Humbolt Street Los Alamitos, CA 90720
To avoid delays in processing your repair, write the MRA on the shipping label. Please include the information about the malfunc­tion with your product. This information enables our service tech­nicians to process your repair order as quickly as possible.
®
Version 3.2A © 2007 FLOWLINE Inc. All rights reserved. Manual # LC900008 09/07
Compact Relay Controller LC30 Series Owner’s Manual
Spec Tech Industrial 203 Vest Ave. Valley Park, MO 63088 Phone: 888 SPECTECH
E-mail: sales@spectechind.com www.spectechind.com
Step One
SPECIFICATIONS
Supply voltage: 120 / 240 VAC, 50 - 60 Hz. Consumption: 5 watts Strobe type: 10_1: N/A
10_2: Xenon tube
Strobe flash: 10_1: N/A
10_2: 1 per second Sensor input: (1) four-wire flow switch Sensor supply: 13.5 VDC @ 100 mA Contact type: (1) SPDT Contact rating: 250 VAC, 10A, 1/4 Hp Contact delay: 0 to 60 seconds LED indication: Sensor, relay & power status Electronics temp.: F: -40° to 158°
C: -40° to 70° Enclosure rating: NEMA 4X (IP65) Enclosure material: 10_1: PP (U.L. 94 VO)
10_2: PP (U.L. 94 VO) & polycarbonate Encl. mount: 100_: 3/4” NPT
105_: 3/4” R Enclosure rotation: 300° swivel base Conduit entrance: Single, 1/2" NPT Classification: general purpose Certificate: LR 79326-3 CE compliance: EN 61326 EMC
EN 61010-1 Safety
Part Number Information:
Part # Thread Description
LC30-1001
(1) (2)
3/4” NPT Flow or No-Flow Relay Controller
LC30-1051
(1) (2)
3/4” BSP Flow or No-Flow Controller
LC30-1002
(2)
3/4” NPT Flow or No-Flow Controller with Strobe Alert
LC30-1052
(2)
3/4” BSP Flow or No-Flow Controller with Strobe Alert
(1)
Adding -CE to end of part number signify CE configuration of controller
(2)
Adding -E to end of part number signify 240 VAC configuration of controller
1.0"
(25mm)
4.7" (119mm)
2.8"
(71mm)
3.2" (81mm)
1.0"
(25mm)
2.8"
(71mm)
3.2" (81mm)
6.4" (162mm)
Latch/Invert Logic
Time Delay
Sensor
Input
1
+ - S
NO NC
COM
AC AC
GND
RELAY
POWER SUPPLY
Dimension: LC30-10_1
Dimension: LC30-10_2
Internal Controller Logic
LC11-1001 Faceplate
AC
PP
RR
115 VAC 220 VAC
AC GND NC C NO
DELAY
INVERINVERT +/- +/-
Input 1A
(S)
(-)
(+)
Loading...