Flowline DS14 Specifications

EchoPod DS14 Ultrasonic Level Switch & Controller
®
Application
The general purpose ultrasonic level switch provides
continuous level detection up to 49.2” (1.25m) with 4
programmable relays for level switch or level control functions, and is congured via our free WebCal software. The embedded level controller can lower cost by replacing external control hardware. This non-contact liquid level sensor is ideally suited for corrosive, sticky or dirty liquids, and is broadly selected for small day tank, skid, IBC, sump and
process tank level applications.
®
Features
Level detection, switch and control functions up to
49.2” (1.25m)
Conguration is fast and easy via WebCal software
and USB adapter
Narrow 2” beam width and short 2” dead band
optimized for small tanks
Four (4) programmable relays rated @ 1A / 60VA for switch, pump or valve control and fail-safety
• 1 pump or valve with 3 alarms
• 2 pumps (lead-lag) with 2 alarms
• 2 pumps (duplexing) with 2 alarms
• 4 independent switch point alarms
PVDF transducer and 6P polycarbonate enclosure for corrosive liquids
Automatic temperature compensation for accurate measurement
Success
Metal parts are often
painted in dip tanks
to achieve a clean,
uniform coating. In
this process, sus-
pended parts are
briey submerged,
and then removed
to rack dry in temperature and humidity controlled
rooms. Dip tanks are generally 2’ to 4’ deep with level
and mixer automation. Here an EchoPod multi-func-
tion level sensor controls the ll valve on each tank
with a high and low level alarm. Additionally, the two-
wire, 4-20 mA level measurement signal is monitored
by the PLC. An optic level switch is also deployed as
a redundant high level alarm.
Compatible Products
PodView®
Level Indicator
LI40
© 2018 Flowline Inc. • 10500 Humbolt Street Los Alamitos, CA 90720 • 562.598.3015 • www.owline.com • DS300220 REV C
EchoPod DS14 Ultrasonic Level Switch & Controller
(23 mm)
3.20" (81 mm)
48"
Red
- Power
EchoPod
Ordering Notes
1) EchoPod can not be configured without the Fob USB interface tool (LI99-1001) and Web Cal. One Fob will configure all EchoPods.
2) WebCal is a free download from our website at www.flowline.com (Windows® XP Compatible, USB 2.0).
®
®
®
Specications
Range: 49.2” (1.25m)
Accuracy: 0.125” (3mm)
Resolution: 0.019” (0.5mm)
Dead band: 2” (5cm)
Beam width: 2” (5cm)
Conguration: WebCal® PC
Windows® USB 2.0
Memory: Non-volatile
Supply voltage: 12-24 VAC / VDC
Consumption: 0.5W
Contact type: (4) SPST relays, 1A
Contact fail-safe: Power loss: Hold last
Power on: Open, close
or hold last
Hysteresis: Selectable
Process temp.: F: 20° to 140°
C: -7° to 60°
Temp. comp.: Automatic
Ambient temp.: F: -31° to 140°
C: -35° to 60°
Pressure: MWP = 30 PSI (2 bar)
Enclosure rating: Type 6P, encapsulated,
corrosion resistant &
submersible
Encl. material: Polycarbonate
Strain relief mat: Santoprene
Trans. material: PVDF
Cable jacket mat: Polyurethane
Cable type: 9-conductor, shielded
Cable length: 48” (1.2m)
Process mount: 1” NPT (1” G)
Mount gasket: Viton®
Classication: General purpose
Compliance: CE, RoHS
Approvals: cFMus
Dimensions
(1.2 m)
2.00"
(51 mm)
Blk Wht Gry Blu Org Yel Pur Brn
- Return
- TX (Out)
- RX (In)
- Rly 1
- Rly 2
- Rly 3
- Rly 4
- Rly Com
Conguration
LI99-1001
The level sensor is congurable via our free WebCal PC software and Fob USB adapter. The sensors are offered with and without Fobs. Fobs are universal and can be used to congure any WebCal compatible
product.
2.70" (69 mm)
.89"
1" NPT
(1" G)
Download your free copy of WebCal in English or Chinese at www.owline.com/webcal.php.
Fittings
For optimum performance, install the level sensor using the below recommended or direct equivalent ttings.
P/N Description
LM52-1400 2” NPT x 1” NPT (Sch. 40) LM52-1800 2” NPT x 1” NPT (Sch. 80)
LM52-1410 2” Socket x 1” NPT (Sch. 40)
LM52-1810 2” Socket x 1” NPT (Sch. 80)
LM52-1850 1” NPT Flange - 150# (Sch. 80)
LM50-1001-1 1” NPT Bracket, Polypropylene
Ordering
Process mount (1) 0 NPT (US) 1 G (Metric)
Fob USB interface (2)
0 Without Fob 1 With Fob
© 2018 Flowline Inc. • 10500 Humbolt Street Los Alamitos, CA 90720 • 562.598.3015 • www.owline.com • DS300220 REV C
DS14-
Notes
1) Install the level sensor using Flowline installation ttings or equivalents.
2) The level sensor is congured via our WebCal software and one LI99-1001 Fob USB adapter.
The level sensor is offered with and without
a Fob. Fobs are universal and can be used to congure any WebCal compatible product. WebCal is a free download from our website at www.owline.com/webcal.php.
Loading...