Specifications
Length: 8" to 10' (20 cm to 3m) Switch points: 1 t 4 (field adjustable) Orientation: ± 20˚ vertical Process temp.: F: -40˚ to 1 7 6 ˚ C: -40˚ to 80˚ Pressure: Atmospheric Wette d material: PP (20% glass fill) Process mount: 2" NPT (2" G) Enclosure rating: NEMA 4X (IP65) Installed height: 5.7" (14.4 cm) above tank process mount Encl. material: PP , UL94VO Conduit entrance: Single, 1/2" NPT Termination: 12 poles
ULTRASONIC SENSOR
Supply voltage: 12-36 VDC Consumption: 25 mA maximum Contact type: (1) SPST relay Contact rating: GP: 120 VAC/VDC @ 60 VA IS: 32 VDC @ 0.5A Cable type: 4-conductor / sensor, #22 AWG (shielded) Contact output: Selectable NO/NC Classification: Intrinsically safe CE compliance: EN 50082-2 immunity EN 55011 emission EN 61010-1 safety
VIBRATION SENSOR
Step One
Supply voltage: 12-30 VDC Consumption: 25 mA maximum Contact type: (1) SPST relay Contact rating: 120 VAC/VDC @ 60 VA Cable type: 5-conductor / sensor, #24 AWG (shielded) Contact output: Selectable NO/NC Classification: General purpose CE compliance: EN 50082-2 immunity EN 55011 emission EN 61010-1 safety
BUOYANCY SENSOR
Contact type: (1) SPDT reed Contact rating: 120 VAC/VDC @ 15 VA Cable type: 3-conductor / sensor, #22 AWG (shielded) Contact output: Selectable NO/NC Classification: General purpose CE compliance: EN50082-2 immunity EN 55011 emission EN61010-1 safety
| 2
MN301420 rev A
Components
V
V
A
A
A
V
V
A
A
A
Step Two
ONE SENSOR CONFIGURATION:
(AU18-434_, AV16-434_ or AZ18-434_)
Ultrasonic AU18-4343
1 x LU10-1305 1 x LM10-1_01 1 x LM30-1001 1 x LC05-1001
AU18-4347
1 x LU10-1325 1 x LM10-1_61 1 x LM30-1051 1 x LC05-1051
Buoyancy AV16-4343
1 x LV10-1301 1 x LM10-1_01 1 x LM30-1001 1 x LC05-1001
AV16-4347
1 x LV10-1351 1 x LM10-1_61 1 x LM30-1051 1 x LC05-1051
ibration
AZ18-4343
1 x LZ12-1405 1 x LM10-1_01 1 x LM30-1001 1 x LC05-1001
AZ18-4347
1 x LZ12-1405 1 x LM10-1_61 1 x LM30-1001 1 x LC05-1051
THREE SENSORS CONFIGURATION:
(AU18-434_, AV16-434_ or AZ18-434_)
TWO SENSORS CONFIGURATION:
(AU28-434_, AV26-434_ or AZ28-434_)
Ultrasonic AU28-4343
2 x LU10-1305 1 x LM10-1_01 2 x LM30-1001 1 x LC05-1001
U28-4347
2 x LU10-1325 1 x LM10-1_61 2 x LM30-1051 1 x LC05-1051
FOUR SENSORS CONFIGURATION:
(AU28-434_, AV26-434_ or AZ28-434_)
Buoyancy AV26-4343
2 x LV10-1301 1 x LM10-1_01 2 x LM30-1001 1 x LC05-1001
V26-4347
2 x LV10-1351 1 x LM10-1_61 2 x LM30-1051 1 x LC05-1051
ibration
AZ28-4343
2 x LZ12-1405 1 x LM10-1_01 2 x LM30-1001 1 x LC05-1001
Z28-4347
2 x LZ12-1405 1 x LM10-1_61 2 x LM30-1001 1 x LC05-1051
Ultrasonic AU38-4343
3 x LU10-1305 1 x LM10-1_01 3 x LM30-1001 1 x LC05-1001
AU38-4347
3 x LU10-1325 1 x LM10-1_61 3 x LM30-1051 1 x LC05-1051
COMPONENT LIST:
Buoyancy AV36-4343
3 x LV10-1301 1 x LM10-1_01 3 x LM30-1001 1 x LC05-1001
AV36-4347
3 x LV10-1351 1 x LM10-1_61 3 x LM30-1051 1 x LC05-1051
Smart Trak Fitting
P/N: LM10-1_01 or LM10-1_61
ibration
AZ38-4343
3 x LZ12-1405 1 x LM10-1_01 3 x LM30-1001 1 x LC05-1001
AZ38-4347
3 x LZ12-1405 1 x LM10-1_61 3 x LM30-1001 1 x LC05-1051
Compact Junction Box
P/N: LC05-1001 or LC05-1051
Ultrasonic AU48-4343
4 x LU10-1305 1 x LM10-1_01 4 x LM30-1001 1 x LC05-1001
U48-4347
4 x LU10-1325 1 x LM10-1_61 4 x LM30-1051 1 x LC05-1051
Switch Car Kit
P/N: LM30-1001 or LM30-1051
Buoyancy AV46-4343
4 x LV10-1301 1 x LM10-1_01 4 x LM30-1001 1 x LC05-1001
V46-4347
4 x LV10-1351 1 x LM10-1_61 4 x LM30-1051 1 x LC05-1051
ibration
AV48-4343
4 x LZ12-1405 1 x LM10-1_01 4 x LM30-1001 1 x LC05-1001
Z48-4347
4 x LZ12-1405 1 x LM10-1_61 4 x LM30-1001 1 x LC05-1051
Switch-Tek Level Switch
P/N: LU10-1305. LU10-1325,
LV10-1301, LV10-1351
or LZ12-1405
MN301420 rev A 3 |
Safety Precautions Step Three
About this Manual: PLEASE READ THE ENTIRE MANUAL PRIOR TO INSTALLING OR USING THIS
PRODUCT. This manual includes information on the Smart Trak™ with Compact Junction Box: AU_5­434_, AZ_8-434_ and AV_6-343_. The units are identical except for the number of switch points and the sensors technology.
User's Responsibility for Safety: Flowline manufactures a wide range of liquid level sensors, controllers,
and mounting systems. It is the user's responsibility to select components that are appropriate for the application, install them properly, perform tests of the installed system, and maintain all components. The failure to do so could result in property damage or serious injury.
Proper Installation and Handling: Use a proper sealant with all installations. Never over tighten the
components. Always check for leaks prior to system start-up.
Material Compatibility:
o Glass filled Polypropylene (PP, a polyolefin): Track, end cap, wire retainer clips, bayonet
adapter, level switch and sensor car for all Smart Trak Assemblies.
o Polychlorotrifluoroethylene (PCTFE, a fluoroplastic): Sensor car locking bolt and screw. o Polypropylene (PP, a polyolefin): Sensor, top compression fitting, thrust plate, locking pin and 2"
NPT fitting.
o Viton (a fluorocarbon): O-ring. o Neoprene (w/silicon gel for lubrication): Wire gasket. o Santoprene (w/silicon gel for lubrication): Seal plug.
Make sure that the application liquids are compatible with the materials that will be wetted. To determine the chemical compatibility between the components and its application liquids, refer to the Compass Corrosion Guide, available from Compass Publications.
Temperature and Pressure: Smart Trak™ is designed for use in application temperatures up to 80° C
(176° F). It is not designed for pressurized applications due to the wiring that must travel through a gasket at the head.
Wiring and Electrical: Electrical wiring of any liquid level control system should be performed in
accordance with all applicable national, state, and local codes. Take care not to cut or break the outer insulation jacket of wiring that may be immersed while routing cables in the Smart Trak™ system. Such breaks of the liquid seal of the sensor system may lead to component failure.
Flammable, Explosive and Hazardous Applications: Smart Trak™ may be used within flammable or
explosive applications only if the associated components are rated intrinsically safe for such use. In hazardous applications, use redundant measurement and control points, each having a different sensing technology.
Make a Fail-Safe System: Design a fail-safe system that accommodates the possibility of transmitter or
power failure. In critical applications, Flowline recommends the use of redundant backup systems and alarms in addition to the primary system.
| 4 MN301420 rev A
Assembly of Smart Trak™
Step Four
About Smart Trak™: Flowline’s Smart Trak™ with Compact Junction Box Assembly is an adjustable
mounting system for installing multiple level sensors vertically within a tank. Mounted through a single point at the top of the tank, up to 4 different sensors can be located at any depth on Smart Trak™. The compact junction box features termination for the various wires from each level switch as well as a 1/2” conduit connection. Smart Trak™ mounts vertically through a standard 2" NPT tank adapter, or on a side mount bracket (such as the LM50-1001). Unlike prefabricated “trees” or pipes, Smart Trak™ allows you to experiment with sensor position to account for variations in the point of actuation of each sensor during process testing.
Track: The track itself is approximately 1" square, and is from 8” to 10' long depending on the A-Dimension. The track may be cut to length if desired. Four separate grooves run the length of the track, one on each side of the square. These grooves hold the sensor cars that attach to Flowline sensors, and also serve to contain the switch cable. The bottom of the track is capped with an end cap.
Level Switches: Smart Trak™ will include from 1 to 4 level switches used to identify it’s own unique wet / dry
condition. The technologies used to indicate level are either Ultrasonic, Buoyancy or Vibration. Each technology features a unique wiring/power configuration (Ultrasonic and Vibration technologies require 12 to 36 VDC power for operation, see below). All of the switches are terminated in the Compact Junction Box. The Compact Junction Box provides a 1/2” Conduit connection and 12 poles for wire termination (for models AU45-434_ and AZ48-434_, common terminals such as Positive (+) and Negative (-) power must be shared).
Compact Junction Box
(inside shown)
Vibration (LZ12-1405) Wire Configuration
Ultrasonic (LU10-1305 or LU10-1325) Wire Configuration
Buoyancy (LV10-1301 or LV10-1351) Wire Configuration
MN301420 rev A 5 |