FLOMEC TM300-N User manual

Page 1
Save theSe inStructionS
Tm Series Electronic Water meters
User Manual
TM Meter with Computer Display TM Digital Pulse Meter
TABLE OF CONTENTS
English ........................................................... 1
Español .......................................................... 9
Deutsch .......................................................16
Italiano ......................................................... 23
Français ....................................................... 30
ENGLISH
ImpOrTANT NOTICE
Use TM Series meters with water and other chemicals compatible with wetted components (see Specifications Section). Do not use to meter fuel or incompatible chemicals. TM Series meters are available with either a computer for local electronic display, or a conditioned signal output module to provide a digital signal to customer interfacing equipment. TM Series meters with computer display measure in gallons or litres. Refer to the Calibration Section for details.
These meters are not legal for trade applications. TM Series meters are very sensitive to electric
noise if operated within 1 to 2 inches of some electric motors or other sources of electronic noise.
04/13
www.GlobalTestSupply.com
INSTALLATION
Connections
Install your meter in-line either horizontally or vertically or at the end of the hose adjacent to the nozzle. Installation to metal connections is not recommended. Install as follows:
1. Plan to install turbine with a minimum straight pipe length as follows:
• Upstreamfromtheturbine,allowamini-
mum straight pipe length of 10 times the internal diameter of the turbine.
• Downstream from the turbine, allow a
minimum straight pipe length of 5 times the internal diameter of the turbine.
2. For Spigot (Pipe) End use only primer and solvents approved for PVC gluing.
Cut to Length The meter housing can be
shortened by the customer. Each meter has a “dotted” line feature molded on the top surface of the housing tube. The housing can be cut up to this line without harming any internals.
Most glue on fittings will fit without interfering
with the computer display area. However, the customer should check all parts before attempting cut.
Rev. A 920861-01
Page 2
For NPT Fittings wrap all connections with
+9 to 35 volt DC
Open Collector
Signal
3
2
1
4
Red
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
The terminal block is identified as follows:
Pin #1 6 volt square wave (not used) Pin #2 +9 to 35 volt DC Input Pin #3 Common Ground Pin #4 Open Collector signal Output
Resistor
3 to 4 wraps of thread tape. Make sure the tape does not intrude into the flow path.
3. Attach meter with arrow pointed in the direc­tion of fluid flow.
4. For NPT Fittings - Hand tighten the meter at the housing ends. Do not use a wrench or similar tool to tighten. This can damage the housing.
5. For ANSI or DIN Flange Fittings - Customer to provide:
• Ring Gaskets or Full-Face Gaskets ap-
proved for use with type flange installed (ANSI or DIN) and the fluid being monitored (2 required).
• 5/8”boltsandnutsforANSIanges.Four
per side for 3-inch meters; eight per side for 4-inch meters.
• 16 mm bolts and nuts for DIN anges.
Eight per side for 3-inch and 4-inch meters.
• Torqueboltsusingastarpatternto25ft-lbs.
Supplied flanges are two-piece Van Stone style and allow the meter to be oriented regardless of the mating flanges position.
For best results, always verify accuracy before use.
WARNING
Compatibility of this product’s material and the process fluid and/or environment should be considered prior to putting into service.
WARNING
Product should never be operated outside its published specifications for temperature or pressure. See specifications for your model.
WARNING
Make sure flow and pressure have been eliminated from process pipe prior to installing or removing product.
WARNING
Always use appropriate thread sealant or flange gaskets when connecting product to
process piping.
OPEN COLLECTOR SIGNAL OUTPUT
2
www.GlobalTestSupply.com
Wiring Diagram 1
Page 3
Wiring Diagram 2
3
2
1
4
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
6 volt Square Wave Signal
The terminal block is identified as follows:
Pin #1 6 volt square wave Pin #2 +9 to 35 volt DC Input (not used) Pin #3 Common Ground Pin #4 Open Collector signal Output (not used)
SQUARE WAVE OUTPUT
CAUTION
To protect against leakage, seal all pipe threads with an appropriate sealing com­pound. Make sure the sealing compound does not intrude into the flow path.
NOTE: If connecting to new male pipe threads,
burrs and curls can adversely affect accu­racy. Correct the problem prior to turbine installation.
NOTE: Do not over tighten the flange bolts. This
may cause the gasket to be compressed into the flow stream and may decrease the accuracy of the meter.
CAUTION
Installation near high electromagnetic fields and high current fields is not recommended and may result in inaccurate readings.
www.GlobalTestSupply.com
Conditioned Signal Output Module Wiring
This conditioned signal output module can be wired to provide an open collector signal output or 6-volt square wave output.
Open Collector Signal Output
To achieve an open collector signal output, reference Wiring Diagram 1. The terminal block is located on the back side of the module. The module is factory assembled for open collec­tor signal output. Please provide the (820 ohm minimum) resistor.
Ten feet (3 m) of cable is provided with the module. Trim it to desired length or extend it as necessary. Distances up to 5,000 feet (1,524 m) can be achieved for open collector signal output.
Square Wave Output
To achieve square wave output, reference Wiring Diagram 2 and use an Electronic Digital Meter Battery Kit (sold separately) for battery power. The terminal block and battery location are located on the back side of the module. Access as follows:
3
Page 4
1. Remove the four Phillips-head screws from the front of the module and lift the module from the turbine.
2. To change terminal block connections, loosen the appropriate screws. Reconnect the wires in the proper positions and tighten the screws.
3. Install the batteries. Make sure the positive post is in the correct position.
4. Position the module on the turbine housing. To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws on the front of the module.
Ten feet (3 m) of cable is provided with the mod­ule. Trim the cable to desired length or extend it as necessary.
Verify Meter Accuracy
Before using, check the meter’s accuracy and verify calibration.
1. Make sure there is no air in the system by starting the flow until it runs steadily. Then, stop or divert the flow using a valve or nozzle.
2. Meter an exact known volume into an ac­curate container. For best results, meter with one continuous full stream.
3. Check the volume against the display or recording equipment. If the amount metered is accurate, further calibration is not neces­sary. If not, refer to the Calibration Section for further instructions.
OpErATION
Computer Display – Batch and Cumulative Totals
The computer maintains two totals. The Cumula­tive Total provides continuous measurement and cannot be manually reset. The Batch Total can be reset to measure flow during a single use. The Cumulative Total is labeled with TOTAL 1, Batch Total is labeled TOTAL 2 BATCH.
When the Cumulative Total reaches a display reading of 999,999 the computer will highlight an X10 icon. This indicates to the operator that a zero must be added to the 6 digits shown. When the next rollover occurs, the computer will highlight an X100 icon. This indicates to the operator that two zeroes must be added to the 6 digits shown.
Press the DISPLAY button briefly to switch between the TOTAL 1, TOTAL 2 BATCH and FLOWRATE. Press DISPLAY briefly to display the TOTAL 2 BATCH. Hold the DISPLAY button for 3 seconds to reset the Batch Total to zero.
When fluid is flowing through the meter, a small propeller icon is highlighted.
NOTE: Totalization counts total units without
differentiating between gallons, litres or field calibrated units.
Flowrate Feature
To use this feature, press and release DISPLAY until FLOWRATE icon appears. The factory set time base will be highlighted to the right of FLOWRATE (M = minutes, H = hours, D = days). When FLOWRATE is invoked, the display will be indicating rate of flow.
Activate the Meter
Computer is on continuously and always ready to perform. The computer is powered by field replaceable batteries. When display becomes dim, faded or the low battery message appears (see below), the batteries need to be replaced. Reference the Maintenance Section for details.
Factory and Field Calibration
All calibration information is visible to the user as icons on the top line of the display, above the numeric digits.
All units are configured with a “factory” calibra-
tion.Bothgallonsandlitresareavailable(“GL”
or “LT” will be displayed). While holding the CALIBRATE button, briefly press DISPLAY to toggle between gallons and litres. This factory calibration (indicated with FAC) is permanently programmed into the computer and is not user adjustable.
NOTE: Your computer may have other units of
measure programmed into it. If so, holding the CALIBRATE button and momentarily pressing the DISPLAY button will toggle through all factory set units. Other possible units are:
IGL(imperial gallon), QT (quart),CF (cubic
feet), CM (cubic meter), BL (42 gal. barrel), CC (cubic centimeter) or OZ (ounce).
4
www.GlobalTestSupply.com
Page 5
Switching between different units will not corrupt
theTotal’scontents.Forexample,inGLmode,
the computer totalizes 10.00 gallons, if the user switches to LT mode, the display will read 37.85 litres (the same volume, different unit).
The “field” calibration may be set by the user, and can be changed or modified at any time using the calibration procedure described below in the Calibration Section. Totals or flowrate derived from the field calibration are invoked when the FAC icon is no longer visible on the top line of the display.
CALIBrATION
Verify Accuracy Before Beginning Field Calibration
For the most accurate results, dispense at a flowrate which best simulates your actual oper­ating conditions. Avoid “dribbling” more fluid or repeatedly starting and stopping the flow. This can result in less accurate calibrations.
Make sure you meet the meter’s minimum flowrate requirements:
TM Series Meters
3inchmeter 30GPM(113.6LPM)  4inchmeter 40GPM(151.4LPM)
The use of a uniformly dependable, accurate calibration container is recommended for the most accurate results. Due to high flowrate, it is strongly recommended that calibration be completed with a combination of volume and weight using fine resolution scales.
For best results, the meter should be installed and purged of air before field calibration.
Field Calibration with Computer Display
Field Calibration and Factory Calibration are de­fined in the Operation Section. Factory calibration settings are programmed into each computer during manufacturing, using water at 70° F (21° C). Readings using the Factory Calibration (FAC) may not be accurate in some situations, for ex­ample, under extreme temperature conditions, non-standard plumbing configurations or with fluids other than water.
Field Calibration Procedures (Correction Factor Method)
1. To calibrate, press and hold the CALIBRATE and DISPLAY buttons for about 3 seconds until you see FLdCAL. Release both buttons and you will see CF - 00.0. You are now in the field calibration mode and values from
-99.9% to +99.9% can be entered.
+
2. The
3. The DISPLAY button can then be pushed to
4. All enabled units-of-measure remain visible
5. To return to factory calibration (FAC), press
/– position appears either as an “under­score” character for plus, or as a “hyphen” character for minus. The DISPLAY button selects the position and the CALIBRATE button toggles this character.
select the numeric positions. Press the CALI­BRATE button to scroll from 0 to 9. Enter the percentage of change you want the display to correct. When satisfied with the value, press both CALIBRATE and DISPLAY buttons simultaneously. CALEnd will be displayed and unit will go back to normal operation, less the FAC (factory calibration) icon.
and selectable – the entered correction will be applied to all enabled units.
and hold both CALIBRATE and DISPLAY buttons for about 3 seconds until FAcCAL is displayed. Then release buttons. Unit should return to normal operation and FAC icon is visible.
Calibration with Conditioned Signal Output Module
The K-factor of your meter appears on the calibration report as the number of pulses per gallon. The factor is determined during produc­tion using water at 70° F (21° C). This K-factor may be used for “single point” calibration and provide acceptable accuracy. However, readings may not be accurate when using this calibration method in some situations. For example, extreme temperature conditions, non-standard plumbing configurations or with fluids other than water.
www.GlobalTestSupply.com
5
Page 6
mAINTENANCE
Proper handling and care will extend the life and service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. How­ever, it is important the rotor moves freely. Keep the meter clean and free of contaminants.
The rotor can be removed for cleaning and inspec­tion. Begin by unscrewing the nose cone from the outlet end of the meter. A 1/4" square socket extension can be used. Remove the lock and flat thrust washers. Rotor can then be removed from shaft. Note orientation because rotor is not bi-directional. Remove debris or deposits using soft brush or small probe. Reassemble in reverse order.
CAUTION
Blowing compressed air through the turbine assembly could damage the rotor.
CAUTION
Do not allow liquids to dry inside the turbine.
CAUTION
Batteries should ONLY be replaced with P/N 113520-1 Kit (Includes two each P/N 902004-2 Batteries). Do not mix old with new. Do not use other brands or technologies.
Open battery cells should be disposed of in accordance with local regulations. Lithium batteries are best disposed of as a non-hazardous waste when fully or mostly discharged. EPA does not list or exempt Lithium as a hazardous waste. If waste lithium batteries are still fully charged or only partially discharged, they can be considered a reactive hazardous waste because of unconsumed lithium remaining in the battery. Such batteries may qualify as “Universal Waste” in many jurisdictions within the U.S. and thus can be shipped for disposal or recycling in accordance with Universal Waste requirements.
If the display becomes dim, blank or the low battery message appears (see below), replace the batteries as follows:
CAUTION
Handle the rotor carefully. Small scratches or nicks can affect accuracy.
NOTE: Make sure the arrow on the turbine outlet
is pointed in the direction of fluid flow.
Battery Replacement
The computer display is powered by two 3-volt lithium batteries which may be replaced while the meter is installed. When batteries are removed or lose power, the batch and cumula tive totals and the field and factory calibrations are re tained.
WARNING
(Battery) – Avoid mechanical or electrical abuse. Batteries may explode or cause burns, if disassembled, crushed or exposed to fire or temperatures in excess of 212°F (100°C). Do not short circuit or install with incorrect polarity. DO NOT INCINERATE.
6
www.GlobalTestSupply.com
1. Remove the four Phillips-head screws from the face of the meter and lift the faceplate from the turbine.
2. Remove the old batteries and clean any cor­rosion from the terminals.
3. Install new batteries. Make sure the positive post is in the correct position.
4. When the batteries are replaced, the faceplate will power ON. Check the display to ensure normal functions have resumed before as­sembling again.
5. Reseat batteries, if necessary, and position the faceplate on the turbine housing. To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws on the faceplate.
Page 7
SpECIFICATIONS
Inlet and Outlet:
Spigot (Pipe) End Models: TM300/TM300-P 3 inch Schd. 80, Spigot (Pipe)
TM400/TM400-P 4 inch Schd. 80, Spigot (Pipe)
NPT Models: TM300-N/TM300-N-P 3 inch NPT TM400-N/TM400-N-P 4 inch NPT
ANSI Flange Models: TM300-F/TM300-F-P 3 inch 150# ANSI Flange
TM400-F/TM400-F-P 4 inch 150# ANSI Flange DIN Flange Models: TM300-D/TM300-D-P 3 inch PN10/PN16 DIN Flange TM400-D/TM400-D-P 4 inch PN10/PN16 DIN Flange
Design Type: Turbine Wetted Components:
Housing: PVC Journal Bearings: PEEK Shaft: Stainless Steel Rotor and Nose Cone: Acetal Washers: Stainless Steel
Max. Working Pressure:
TM300(AllModels):225PSIG@73°F  TM400(AllModels):225PSIG@73°F  (NonCE),135PSIG@73°F
(CE applications)
U.S. Measurement
Unit of Measure: Gallon Flow Range:
3inch: 40-400GPM
(Linearity: ±3.0% of Reading)
 30-600GPM
(Accuracy: ±2.0% Full Scale)
4inch: 60-600GPM
(Linearity: ±3.0 of Reading)
 40-800GPM
(Accuracy ±2.0% Full Scale)
Accuracy with Computer:
± 3.0% of reading (Accuracy can be improved with field calibration)
Operating Temperature: +32° to +140° F
(Do not allow fluid to freeze inside meter.)
Battery Life: 5 years Storage Temperature: –40° to +158° F Product Weight - lbs.:
Spigot ANSI DIN (Pipe) NPT Flange Flange
3 inch: 2.4 3.9 5.8 6.5 4 inch: 3.7 6.1 9.2 8.8
Dimensions - Inches (L x H x W):
Spigot (Pipe): 3 inch: 11.5 x 5.34 x 3.5
4 inch: 13.5 x 6.34 x 4.5
*
**
NPT: 3 inch: 14.7 x 5.78 x 4.37
4 inch: 17.0 x 6.7 x 5.87
ANSI Flange:
3 inch: 12.0 x 7.5 x 7.5
4 inch: 14.0 x 9.0 x 9.0
DIN Flange:
3 inch: 12.0 x 8.0 x 8.0
4 inch: 14.0 x 8.5 x 8.5
* Add .3 lbs. for Conditioned Signal Module. ** Add 1.1 inches to height for Conditioned Signal
module.
Metric Measurement
Unit of Measure: Litre Flow Range:
3 inch: 151-1514 LPM
113-2271 LPM
4 inch: 227-2271 LPM
151-3028 LPM
Accuracy with Computer: ± 3.0% of reading (Accuracy can be
Operating Temperature: 0° to +60° C
(Do not allow fluid to freeze inside meter.)
Battery Life: 5 years Storage Temperature: –40° to +70° C Product Weight - kgs:
Spigot ANSI DIN (Pipe) NPT Flange Flange
3 inch: 1.09 1.77 2.63 2.95 4 inch: 1.68 2.77 4.17 3.99
(Linearity: ±3.0% of Reading)
(Accuracy: ±2.0% Full Scale)
(Linearity: ±3.0 of Reading)
(Accuracy ±2.0% Full Scale)
improved with field calibration)
*
www.GlobalTestSupply.com
7
Page 8
Dimensions - cm (L x H x W):**
Spigot (Pipe): 3 inch: 29.2 x 13.6 x 8.9 4 inch: 34.3 x 16.1 x 11.4 NPT: 3 inch: 37.3 x 14.7 x 11.1
4 inch: 43.2 x 17.0 x 14.9 ANSI Flange:
3 inch: 30.5 x 20.3 x 20.3 4 inch: 35.6 x 22.9 x 22.9
DIN Flange: 3 inch: 30.5 x 20.3 x 20.3 4 inch: 35.6 x 21.6 x 21.6
* Add .14 kg to weight for Conditioned Signal
Module.
** Add 2.8 cm to height for Conditioned Signal
module.
SErVICE
For warranty consideration, contact your local distributor. If you need further assistance, contact
theGPICustomerServiceDepartmentat:
1-888-996-3837
You will need to:
• Provideinformation fromthe decal on your
meter.
• ReceiveaReturnAuthorizationnumber.
• Flushanyuidfromthemeterbeforeshipping
to the factory.
If possible leave customer installed fittings or ample length of bare pipe for reinstallation.
pArTS
The following replacement parts and accessories are available for the TM Series meters:
Part No. Description
113435-1 Conditioned Signal Output Module 113520-1 Battery Replacement Kit
12517601 Gasket
12519001 Computer Mount Assembly 901002-52 Seal 125518-01 3 in. (Spigot) Turbine Assy Kit 125518-02 3 in. (NPT) Turbine Assy Kit 125518-03 3 in. (Flange) Turbine Assy Kit 125520-01 4 in. (Spigot) Turbine Assy Kit 125520-02 4 in. (NPT) Turbine Assy Kit 125520-03 4 in. (Flange) Turbine Assy Kit 125519-01 3 in. Computer Assy Kit 125521-01 4 in. Computer Assy Kit 125518-04 3 in. (DIN Flange, PVC) Turbine Assy Kit 125520-05 4 in. (DIN Flange, PVC) Turbine Assy Kit
CAUTION
Do not return the meter without specific authority from the GPI Customer Service Department. Due to strict regulations governing transportation, handling, and disposal of hazardous or flammable liquids, GPI will not accept meters for rework unless they are completely free of liquid residue.
WEEE DIrECTIVE
The Waste Electrical and Electronic Equipment (WEEE) directive (2002/96/EC) was approved by the European Parliament and the Council of the European Union in 2003. This symbol indicates that this product contains electrical and electronic equipment that may include bat-
teries, printed circuit boards, liquid crystal displays or other components that may be subject to local disposal regulations at your location. Please understand those regulations and dispose of this product in a responsible manner.
RoHS Compliant (2011/65/EU)
This product is in compliance with the RoHS Directive of the European Parliament and of the Council on the Restriction of the Use of Certain Hazardous Substances in Electrical and Elec­tronic Equipment.
Environmental Rating: IP65
8
www.GlobalTestSupply.com
Loading...