Flomec QS200, QS200-30, QS200-10, QS200-10PW, QS200-40 Product Owners Manual

...
Page 1
QS200
Insertion Ultrasonic
Flowmeter
STANDARD METERS:
QS200-10, QS200-15, QS200-20, QS200-30, QS200-40
U.S. Design Patent No. D845,804 - Ultrasonic Insert
U.S. Design Patent No. D845,805 - Tee Housing for Ultrasonic Insert
QS200-10PW, QS200-15PW, QS200-20PW, QS200-30PW, QS200-40PW
STANDARD RETROFIT INSERT:
QS200
Product Owner’s Manual
EN
920901-01 Rev C04/29/2019
Page 2
Please save these instructions for future reference. Read carefully before attempting
to assemble, install, operate or maintain the product described.
Protect yourself and others by observing all safety information. Failure to comply
with instructions could result in personal injury and/or property damage.
Please refer to back cover for information regarding this product’s warranty and
other important information.
SAVE FOR YOUR RECORDS
Model #: ___________________
Serial #: ___________________
Purch. Date: _______________
DO NOT RETURN THIS PRODUCT TO THE STORE!
Please contact Great Plains Industries, Inc.® before returning any product. If you are
missing parts, or experience problems with your installation, contact our Customer
Support Department. We will be happy to assist you.
Call: 888-996-3837 or 316-686-7361
Email: meters@gplains.com
Page 3
3
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
GETTING STARTED
BEFORE YOU BEGIN
Usage Requirements
This meter is for use with water only.
This meter is not legal for trade applications.
This meter has a permanent factory setting for measuring water only.
Power Source Requirements
This meter requires DC power from a customer-provided controller in order to provide flow information back to the controller.
Tools and Materials Needed
Wire strippers, wire cutters, screwdriver, tape measure
PVC pipe fittings (as needed), PVC pipe primer, PVC pipe cement
Direct burial wire splices, valve box
#18 AWG wire cable (Direct Burial) (Controller manufacturer may recommend thicker guage wire for longer distances)
UNPACKING
Contents: QS200-10, -15, -20, -30, -40 or QS200-10PW, -15PW, -20PW, -30PW, -40PW Meters
Note: The “PW” suffix denotes an NSF certified meter
(1) QS200 Ultrasonic insert assembly (2) K-factor Decals
(1) PVC Pipe Tee (for 1 in., 1-1/2 in., 2 in., 3 in., or 4 in. diameter pipe)
(1) Owners Manual
(1) Quick Release Pin
Contents: QS200 Retrofit Insert (Not available as NSF certified)
(1) QS200 Ultrasonic insert assembly (1) Quick Release Pin
(1) Set of two O-rings (round profile) (1) Owners Manual
(2) K-factor Decals
Inspect
After unpacking the unit, inspect carefully for any damage that may have occurred during transit. Check for loose, missing or damaged parts. Shipping damage claims must be filed with carrier.
See General Safety Instructions, and all Cautions, Warnings, and Dangers as shown.
tools.eps
Heavy_Hazard.eps
tools.eps
box.eps
Heavy_Hazard.eps
Hot_surface.eps
tools.eps
box.eps
magnify_2.eps
Heavy_Hazard.eps
Hot_surface.eps
Recycle.eps
tools.eps
box.eps
magnify_2.eps
warning symbol.eps
Heavy_Hazard.eps
Hot_surface.eps
Recycle.eps
Resp_protection1.eps
Page 4
MAINTENANCE /
REPAIR
TROUBLESHOOTING OPERATION
ASSEMBLY /
INSTALLATION
GETTING STARTED
4
SAFETY /
SPECIFICATIONS
GENERAL SAFETY INSTRUCTIONS
IMPORTANT: It is your responsibility to:
Ensure that all equipment operators have access to adequate instructions concerning safe operating and maintenance procedures.
This product is not approved for use with
petroleum products (diesel fuel, unleaded gasoline, jet fuel, kerosene, etc.), aromatic hydrocarbons or other incompatible chemicals
This product is not approved for use in hazardous
locations.
When applying power, adhere to specifications
listed in appropriate electronics manual.
Disconnect external power before attaching or
detaching input or output wires.
NOTE: Be sure O-rings and seals are kept in good repair.
Compatibility of this product’s material and the
process fluid and/or environment should be considered prior to putting into service.
Product should never be operated outside its
published specifications for temperature or pressure. See specifications for your model.
Make sure flow and pressure have been eliminated
from process pipe prior to installing or removing product.
Installation near high electromagnetic fields and
high current fields is not recommended and may result in inaccurate readings.
Do not allow water to freeze in meter. Ice
expansion may burst the plastic housing.
Do not allow this meter to be used with steam.
Page 5
5
GETTING STARTED
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
SAFETY /
SPECIFICATIONS
INSERTS - ALL METER SIZES
Transducer Excitation
OFF State Current 200µA (typical)
OFF State V-High
Supply Voltage - (OFF State Current * Supply impedance)
ON State Current
(Supply Voltage / (Supply
impedance + 50Ω))
ON State V-Low ON State Current * 50Ω
Output Frequency 0 to 100 Hz
Output Pulse Width 4mSec (Approx.)
SPECIFICATIONS
QS200-10 QS200-15 QS200-20 QS200-30 QS200-40
QS200-10PW QS200-15PW QS200-20PW QS200-30PW QS200-40PW
Tee Housing Material
SCH 80 PVC (Polyvinyl chloride)
Insert Housing Material
PPS (Polyphenylene sulde) / ULTEM® (Polyetherimide)
Type Ultrasonic Flowmeter
Powered by
DC power provided by customer controller
7.5V (dc) min to 36V (dc) max
OFF State Current: 200μA (typical)
Unit of
Measure
Controller Dependant
Flow Rate
0.22-33 GPM
0.83-124.92 L/min
0.1-15 ft/sec
0.55-82 GPM
2.08-310.41 L/min
0.1-15 ft/sec
0.92-138 GPM
3.48-522.39 L/min
0.1-15 ft/sec
2.06-309 GPM
7.80-1169.70 L/min
0.1-15 ft/sec
3.58-537 GPM
13.55-2032.78 L/min
0.1-15 ft/sec
Accuracy +/- 2% of Reading
Uncertainty
0.04 GPM
0.018 ft/sec
0.10 GPM
0.018 ft/sec
0.17 GPM
0.018 ft/sec
0.37 GPM
0.018 ft/sec
0.65 GPM
0.018 ft/sec
Max.
Working Pressure
150 PSI @ 70°F (10.3 bar @ 60°C)
Operating
Temperature
+32°F to +140°F (0°C to +60°C)
Storage
Temperature
+32°F to +140°F (0°C to +60°C)
Field Calibration
No
Inlet / Outlet
Connections
1 in. Female
Socket
1 1/2 in.
Female
Socket
2 in. Female
Socket
3 in. Female
Socket
4 in. Female
Socket
Weight
0.95 lbs. (0.43kg)
1.12 lbs. (0.51kg)
1.39 lbs. (0.63kg)
2.52 lbs. (1.14kg)
3.21 lbs. (1.46kg)
Page 6
MAINTENANCE /
REPAIR
TROUBLESHOOTING OPERATION
ASSEMBLY /
INSTALLATION
GETTING STARTED
6
SPECIFICATIONS (CONTINUED)
QS200-10,
QS200-10PW
1 INCH
QS200-15,
QS200-15PW
1-1/2 INCH
QS200-20,
QS200-20PW
2 INCH
QS200-30,
QS200-30PW
3 INCH
QS200-40,
QS200-40PW
4 INCH
Dimensions
A. Length
4.25 in.
(108mm)
4.90 in.
(124mm)
5.56 in.
(141mm)
6.63 in.
(168mm)
7.38 in.
(187mm)
B. Height
5.38 in.
(137mm)
5.63 in.
(143mm)
6.12 in.
(156mm)
7.20 in.
(183mm)
8.41 in.
(213mm)
C. Width (at widest point)
2.50 in.
(64mm)
2.50 in.
(64mm)
2.88 in.
(73mm)
4.18 in.
(106mm)
5.23 in.
(133mm)
D. Depth
4.47 in.
(114mm)
4.47 in.
(114mm)
4.94 in.
(125mm)
4.91 in.
(124mm)
6.17 in.
(156mm)
SAFETY /
SPECIFICATIONS
Figure 1
SECT
SECT
B
C
A
SECT VIEW (TEE ONLY)
D
For 3 in. & 4 in. Meters
A
SECT VIEW (TEE ONLY)
D
B
C
SECT
SECT
For 1 in., 1-1/2 in., & 2 in. Meters
Page 7
7
GETTING STARTED
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
SPECIFICATIONS (CONTINUED)
K-Factor Information
NOTE: The meter size is molded on the vertical arm of the Tee
FLOMEC ultrasonic meters use K-factor plus offset numbers for greater accuracy during calibration. These values are derived by calibrating the meters using NIST traceable instrumentation. Using both sets of values to calibrate the meters provides greater accuracy than using only a K-factor value. The K-factor and offset values for each meter are listed below.
IMPORTANT: The K-factors provided are for reference. Accuracy can be affected by plumbing configuration, fluid condition, adjoining pipe schedule, type of meter tee (if using QS200 insert in non-FLOMEC tee), and entrapped air. Customers should always validate accuracy and adjust K-factor as needed.
Meter
Model
Size
FLOMEC TEE
K-Factor
(frequency)
Offset
NON-FLOMEC TEE
K-Factor
(frequency)
QS200-10 1 in. 0.5386 0 N/A
QS200-15 1-1/2 in. 0.7926 0 0.7947
QS200-20 2 in. 1.3765 0 1.3583
QS200-30 3 in. 3.8444 0 4.2505
QS200-40 4 in. 7.1676 0 7.2229
SAFETY /
SPECIFICATIONS
NSF CERTIFICATION INFORMATION
NOTE: QS200 meter are available as a standard meter, or as an NSF certified meter. Only QS200 meters ordered as a complete unit (ultrasonic insert installed in FLOMEC tee) qualify to be NSF certified.
Standard meter: The identification plate on the top of the ultrasonic insert is marked “QS200”.
NSF certified meter: The identification plate on the top of the ultrasonic insert is marked “QS200PW”, and the tee body is marked with an NSF certification decal for additional identification.
NOTE: NSF certified meters use the same part number structure as non-NSF models, except for the addition of the “PW” suffix on the end. The “PW” suffix indicates an NSF certified meter. e.g., QS200-15PW.
NSF certification is valid only when product is marked per above information.
Page 8
8
MAINTENANCE /
REPAIR
TROUBLESHOOTING
SAFETY /
SPECIFICATIONS
GETTING STARTEDOPERATION
ASSEMBLY /
INSTALLATION
INSTALLATION
Below is a typical QS200 meter with labeled components. Familiarize yourself with the meter before installation.
Figure 2
TEE
QUICK RELEASE PIN
INSERT ASSY.
BLACK WIRE
RED WIRE
GREEN LED
AMBER LED
SIZE INFO
I.D. PLATE WITH
SERIAL NUMBER
AND FLOW ARROW
TYPICAL QS200 METER
Page 9
9
GETTING STARTED OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
INSTALLATION (CONTINUED)
Provide a straight pipe run of at least 10Xs the pipe’s diameter upstream of the meter, and at least 5Xs the pipe’s diameter downstream of the meter. The arrow embossed on the insert body denotes the flow direction.
FLOW DIRECTION
10Xs
5Xs
DOWNSTREAM
UPSTREAM
INSTALL AT ANGLE, IF POSSIBLE
INSERT
TOP VIEW
FLOW DIRECTION ARROW
Figure 3
Install Meter Onto Pipe
NOTE: There is no need to remove the insert to install the meter. The meter must be installed with the arrow on the insert pointing in the flow direction.
If the meter is accidentally installed backwards, simply remove the insert, rotate 180 degrees so the arrow points in the flow direction, and reinstall the insert.
The Tee is bidirectional, the insert is not. The insert operates correctly as long as the insert arrow is pointed in the direction of flow. The insert can be rotated 180 degrees, so its arrow can always be pointed in the direction of flow regardless of Tee installation.
If space allows, install Tee / insert at an angle rather than pointing up (see Figure 3).
NOTE: For 1 inch pipe installations, this angle is especially important and should be set at 45 degrees for most accurate meter operation.
Page 10
MAINTENANCE /
REPAIR
TROUBLESHOOTING OPERATION
SAFETY /
SPECIFICATIONS
GETTING STARTED
10
ASSEMBLY /
INSTALLATION
INSTALLATION (CONTINUED)
1. Remove all burrs from the pipe ends I.D. and O.D. edges and the Tee sockets I.D. edges (see Figure 4).
2. Clean and apply primer to the pipe ends and Tee sockets (see Figure 5).
3. Apply PVC cement to pipe ends and Tee sockets and quickly assemble the parts while the cement is fluid. Follow the cement manufacturer’s instructions (see Figure 6).
4. Hold the cemented parts together for at least 30 seconds.
Figure 4 Figure 5
APPLY PRIMER
APPLY CEMENT
REMOVE BURRS
Figure 6
For Below Ground Installation
1. Install a valve box around the insert. Valve box extensions may be
needed depending on depth (see Figure 7).
NOTE: A minimum of 10 in. thick layer of gravel should be installed immediately below the meter and valve box.
Figure 7
Page 11
11
GETTING STARTED
SAFETY /
SPECIFICATIONS
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
ASSEMBLY /
INSTALLATION
INSTALLATION (CONTINUED)
Wiring Connections
NOTE: Wiring diagram is shown in next section.
1. When using 18 AWG cable, cut off the unused wires so that they are
even with the sheath of the cable.
NOTE: For 18 AWG connections, remember the color of the twisted pair of wires you use so that you can make an identical connection with the same wires later.
2. 3M DBR/Y-6 Splice Kit Instructions:
Figure 8a
Figure 8c
Figure 8b
Figure 8d & 8e
Page 12
12
MAINTENANCE /
REPAIR
TROUBLESHOOTING
SAFETY /
SPECIFICATIONS
GETTING STARTEDOPERATION
ASSEMBLY /
INSTALLATION
INSTALLATION (CONTINUED)
a. Strip insulation ¾ in. (19 mm).
b. With wire ends even, insert wires into the connector and tighten until
secure.
c. Insert the connector all the way into the tube until the connector rests
on the bottom of the tube.
NOTE: If having difficulty getting the twist-on connector down into the tube when using small gauge wires, use a thin, non-conductive object to push the connector to the bottom of the tube. Upon removal of the object, ensure that no voids or water paths remain in the grease.
d. Fold the wires into the channels.
e. Close the cap.
Wiring Diagram
This diagram shows connection to a RAINBIRD® Flow Smart Module. Connections to other flow controllers may vary.
1212
RED
BLACK
DBR/Y-6 SPLICE KIT
FLOW INSERT
TO FLOW " + "
ON FLOW SMART MODULE
TO FLOW " - " ON FLOW SMART MODULE
18 AWG CABLE to CONTROLLER
Figure 9
Page 13
NOTES
13
GETTING STARTED
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
13
GETTING STARTED OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
1313
Page 14
14
MAINTENANCE /
REPAIR
TROUBLESHOOTING
SAFETY /
SPECIFICATIONS
GETTING STARTED
OPERATION
Each QS200 meter is shipped with two K-factor decals (placed loose inside the owner’s manual packet). One is to affix on or near the installed meter for on-site use if desired. The other decal is to affix to a convenient spot on or near the controller for immediate reference if desired.
LEDs Functionality
There are two LEDs on the QS200 product. The GREEN LED is used to indicate basic power and functionality of the meter. The AMBER LED is used to indicate that there is a flow of water through the QS200 meter insert. The LEDs will behave in the following manner to indicate different modes of operation:
MODE LED BEHAVIOR
Power Disconnected
or Meter Failure:
Both GREEN and AMBER LEDs are OFF.
No Flow Low
Power Mode:
GREEN LED flashes ON/OFF at a rate
of approximately 2 blinks per second.
No Flow or Reverse
Flow Active Mode:
GREEN LED flashes ON/OFF at a rate
of approximately 8 blinks per second.
Low Flow
Active Mode:
GREEN LED flashes ON/OFF at a rate of
approximately 8 blinks per second and the
AMBER LED flashes ON/OFF at a rate that is
proportional to the rate of water flow thru the
meter.
**High Flow
Active Mode:
GREEN LED flashes ON/OFF at a rate of
approximately 8 blinks per second and the
AMBER LED will appear to be constantly ON, but
will be dim.
**NOTE: To save power it is normal for the LEDs to be dim when flashing at higher flow rates. At very high flow rates, the AMBER LED will appear to be constantly ON but dim.
1414
ASSEMBLY /
INSTALLATION
OPERATION
Page 15
15
GETTING STARTED TROUBLESHOOTING
SAFETY /
SPECIFICATIONS
1515
ASSEMBLY /
INSTALLATION
OPERATION
MAINTENANCE /
REPAIR
TROUBLESHOOTING
1. The LEDs are the primary indicators of meter performance. Refer to the Operations Section for LED indicating functionality.
2. Entrained air is air bubbles suspended in the water flow. Entrained air creates errors in accuracy of ultrasonic technology meters. Recommend a maximum of 10% entrained air in the water flow.
3. The faces of the transducers need to be clean and free of oily substances for accurate operation. Do not touch transducers with fingers, oily rags, etc. DO NOT use wire brushes or abrasives to clean the faces of the transducers (see Figure 10).
4. Ensure the flow direction arrow is pointing in the direction of flow for correct LED functionality.
5. Ensure the quick release pin is installed in the tee to maintain pressure and alignment of the insert in the tee.
Figure 10
TRANSDUCER FACES
Page 16
REPAIR
QS200 Insert Replacement and QS200 Retrofit Insert Installation Instructions
NOTE: These instructions are intended to assist in replacing the insert in an existing FLOMEC® QS200 ultrasonic Meter.
These instructions also apply when a FLOMEC QS200 insert is used as a replacement insert in other brands of Tee type meters, that have an obsolete or unreliable insert.
The FLOMEC QS200 is a direct replacement for a paddle wheel type insert in most meters that have a Tee type housing.
Keep the new insert free of dirt and debris during installation.
Check Meter Size
If your flow controller requires a meter K-factor, use the K-factor for your size meter (see SPECIFICATIONS section). If the meter size is unknown, it is easily determined using one of the following methods (for FLOMEC and other brands):
1. Dipstick Method (For 2 in., 3 in. and 4 in. meters only):The interior depth of the meter body indicates its size. To check, remove the old insert from the meter body and insert the end of a stiff tape measure down into the insert bore until it touches the bottom of the meter body bore. Read the depth shown on the tape measure at the top of the Tee housing (see Figure 11) and compare with dimension “D” in the specifications section (see Figure 1).
2. Line Pipe Size Method: Normally, the line pipe size is the same as the meter size. Figure 12 shows line pipe sizes.
NOTE: FLOMEC Tees are marked with hallmark of size, part number, SCH80 and pressure.
MAINTENANCE /
REPAIR
TROUBLESHOOTING OPERATION
ASSEMBLY /
INSTALLATION
SAFETY /
SPECIFICATIONS
GETTING STARTED
16
Figure 11
Page 17
REPAIR (CONTINUED)
GETTING STARTED
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
1.315
[1 5/16]
Ø
1.900
[1 29/32]
Ø
2.375
[2 3/8]
Ø
1" PIPE
1 1/2" PIPE
2" PIPE
3" PIPE
4" PIPE
3.500
[3 1/2]
Ø
4.500
[4 1/2]
Ø
Figure 12
17
Page 18
TROUBLESHOOTING OPERATION
ASSEMBLY /
INSTALLATION
SAFETY /
SPECIFICATIONS
GETTING STARTED
MAINTENANCE /
REPAIR
REPAIR (CONTINUED)
Remove Old Insert From Tee
1. Clean all dirt and debris away from the immediate area of the old insert and the top of the meter, then pull out the quick release pin from the meter (see Figure 13).
2. Remove the old insert. Grasp the insert flange with your hand and pull straight up and out of the Tee, making sure no dirt or other particles fall into the insert bore of the meter (see Figure 14).
NOTE: The FLOMEC Multi-Tool (p/n 146055-501) can be used to make removal of old insert easier (see Figure 17).
NOTE: When replacing another manufacturer’s insert with the QS200 Retrofit Insert, the old insert could have a sediment or mineral buildup and need to be leveraged out of the tee.
18
Figure 13 Figure 14
Page 19
GETTING STARTED
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
REPAIR (CONTINUED)
Install New or Replacement Insert Into Tee
1. The Tee insert bore must be clean before installing the insert and the (2) black O-rings on the insert should be fully lubricated.
2. Orient the insert over the insert bore so that the arrow on the insert is pointing in the direction of flow.
3. Insert the insert straight down into the Tee.
4. Push down on the insert flange and twist slightly to install, and to align the retaining pin holes in the insert with the holes in the Tee (see Figure 15).
NOTE: The FLOMEC Multi-Tool (p/n 146055-501) can be used to make installing new insert easier (see Figure 18).
NOTE: The O-rings on the outside of the QS200 Retrofit Insert are square profile O-rings. When installing in another brand of Tee, if the insert fits too tight into the Tee, replace the square profile O-rings with the round profile O-rings included with the QS200 Retrofit Insert. Their installation could make installation easier in tight fitting Tees. The O-rings must be fully lubricated before insert installation.
5. Next, replace the quick release pin (see Figure 16).
6. Using the splicing and wiring information (see WIRING CONNECTIONS and WIRING DIAGRAM in the INSTALLATION section), connect the wiring from the new insert to the wiring cable of the controller.
7. Energize the insert and verify that it is operating properly (see Troubleshooting section if required).
19
Figure 15 Figure 16
Page 20
TROUBLESHOOTING OPERATION
ASSEMBLY /
INSTALLATION
SAFETY /
SPECIFICATIONS
GETTING STARTED
MAINTENANCE /
REPAIR
REPAIR (CONTINUED)
Using the FLOMEC Multi-Tool
The multi-tool was designed exclusively for tee type meters that use an insertion type insert. It facilitates easier removal of the insert, whether from a QS200 meter or other brands of tee type insertion meters.
It is especially useful when attempting to remove an old, inoperative, or inaccurate insert from another brand of tee type insertion meter for replacement with a QS200 insert.
The multi-tool is versatile, compact and fits easily into the valve box for close quarters use.
The pointed end is designed for use as:
1. A versatile pry bar and driving wedge. On other brands of tee type insertion meters, when the insert lip may be too close to the top of the tee to use the U-formed end of the multi-tool, use the pointed end (as pry bar or driving wedge) between the insert lip and top of the tee to break the insert loose and gain room between the insert lip and tee to use the U-formed end to lever the insert up and out.
2. An alignment tool to align the pin holes of the new insert with the pin holes in the tee.
3. A cleaning tool for clearing debris from the valve box lid groove or other crevices.
The U-formed end is designed for use as:
4. A handle for leverage when using the pointed end.
5. A levering tool by slipping the U-formed end around the insert between the insert lip and the top of the tee. This positions the tool for use as a lever for levering inserts up and out of the tee for replacement. It can be levered up or down, and can be rotated to any quadrant around the insert in order to give the best position for leverage.
The multi-tool (P/N 146055-501) is a separate item available to the customer, and is listed in the repair parts list near the back of the manual.
20
Page 21
Figure 17
Figure 18
GETTING STARTED
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
REPAIR (CONTINUED)
Remove Old Insert From Tee with FLOMEC Multi-Tool
21
Install New QS200 Insert FLOMEC Multi-Tool
PRY UP - To break a stuck insert loose
LEVERAGE - To Lift insert up and out
PIN HOLE ALIGNMENT ­For easy pin install
CLEAN LID GROOVE ­Remove dirt, roots, debris
DRIVING WEDGE - To lift insert more
Page 22
REPAIR PARTS ILLUSTRATION FOR QS200
TROUBLESHOOTING OPERATION
ASSEMBLY /
INSTALLATION
SAFETY /
SPECIFICATIONS
GETTING STARTED
MAINTENANCE /
REPAIR
Do not return this product without prior approval
from the GPI Customer Service Department. Due to strict government regulations, GPI cannot accept parts unless they have been drained and cleaned.
22
3
2
1
Page 23
PARTS & SERVICE
For warranty consideration, parts, or other service information, please contact your local distributor. If you need further assistance, contact the GPI Customer Service Department in Wichita, Kansas, during normal business hours.
A toll free number is provided for your convenience.
1-888-996-3837
To obtain prompt, efficient service, always be prepared with the following information:
The model number of your meter.
The serial number or manufacturing date code of your meter.
Part descriptions and numbers.
For warranty work, always be prepared with your original sales slip or other evidence of purchase date.
23
GETTING STARTED
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
OPERATION TROUBLESHOOTING
MAINTENANCE /
REPAIR
REPAIR PARTS LIST FOR QS200
Ref. No.
Description
Part Number
1 O-Ring Seal Kit 146500-01
Includes (2) square profile O-rings & (2) round profile O-rings
2 Hardware Kit 146500-2
Includes (1) quick release pin
3 Ultrasonic Insert QS200
Includes (1) insert assembly w/square profile O-rings, (2) round profile O-rings, (1) quick release pin, (2) K-factor decals & (1 owner’s manual)
3 Ultrasonic Insert, Potable Water QS200PW
Includes (1) insert assembly w/square profile O-rings, (2) round profile O-rings, (1) quick release pin, (2) K-factor decals & (1 owner’s manual)
- Housing, 1 inch tee, PVC, Spare 146010-511
- Housing, 1-1/2 inch tee, PVC, Spare 146015-511
- Housing, 2 inch tee, PVC, Spare 146020-511
- Housing Assembly, 3 inch tee, PVC, Spare 146030-512
- Housing Assembly, 4 inch tee, PVC, Spare 146040-512
- Multi-Tool, Spare 146055-501
IMPORTANT: Please contact GPI before returning any parts. It may be possible to diagnose the trouble and identify needed parts in a telephone call.
Page 24
© 2019 Great Plains Industries, Inc., All Rights Reserved.
Great Plains Industries, Inc. / 888-996-3837 / FLOMEC.net
FLOMEC® TWO-YEAR LIMITED WARRANTY
Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205, hereby provides a limited warranty against defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This product includes a 2 year warranty. Manufacturer’s sole obligation under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or repairing defective Goods (subject to limitations hereinafter provided) or refunding the purchase price for such Goods theretofore paid by the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of Manufacturer. The warranty shall extend to the purchaser of this product and to any person to whom such product is transferred during the warranty period.
The warranty period shall begin on the date of manufacture or on the date of purchase with an original sales receipt. This warranty shall not apply if:
A. the product has been altered or modified outside the warrantor’s duly appointed representative;
B. the product has been subjected to neglect, misuse, abuse or damage or has been installed or operated other than in accordance with the manufacturer’s operating instructions.
To make a claim against this warranty, contact the GPI Customer Service Department at
316-686-7361 or 888-996-3837.
Or by mail at:
Great Plains Industries, Inc.
5252 E. 36th St. North
Wichita, KS, USA 67220-3205
The company will step you through a product troubleshooting process to determine appropriate corrective actions.
GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT, INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED IN THE USE OR LOSS OF USE OF THE PRODUCT WARRANTED HEREUNDER.
The company herewith expressly disclaims any warranty of merchantability or fitness for any particular purpose other than for which it was designed.
This warranty gives you specific rights and you may also have other rights which vary from U.S. state to U.S. state.
Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT – Part 702 (governs the resale availability of the warranty terms).
920901-01 Rev C04/29/2019