•It is designed specifically for use on FLOMEC® Industrial Grade
Turbine housings. It is also designed to work with several other
FLOMEC products.
Power Source Requirements
•This computer requires DC power. For this application, the power is
provided by (2) AAA size Alkaline batteries installed in the computer
at the factory.
•Upon receipt, examine your equipment for visible damage. The
computer is a precision measuring instrument and should be
handled as such. If any items appear damaged or missing, contact
your distributor.
•See General Safety Instruction, and all Cautions, Warnings, and
Dangers as shown.
TROUBLESHOOTINGOPERATION
TROUBLESHOOTINGOPERATION
REPAIR
REPAIR
4
MAINTENANCE /
MAINTENANCE /
GENERAL SAFETY INSTRUCTIONS
General Safety
Instructions
GETTINGSTARTED
GETTING STARTED
IMPORTANT: It is your responsibility to:
•Ensure that all equipment operators have access to adequate instructions
concerning safe operating and maintenance procedures.
The apparatus enclosure may contain aluminum and is
considered to constitute a potential risk of ignition by
impact or friction. Care must be taken into account during installation and use
to prevent impact or friction.
Part of the enclosure is constructed from plastic. To
prevent the risk of electrostatic charging the plastic
surface should only be cleaned with a damp cloth.
This product should never be electrically connected to any
other device in the hazardous location. Consult
manufacturer for exceptions.
Compatibility of this product’s material and the process
fluid and/or environment should be considered prior to
putting into service.
Product should never be operated outside its published
specifications for temperature.
(Battery) - Avoid mechanical or electrical abuse. Batteries
may explode or cause burns, if disassembled, crushed, or
exposed to fire or temperatures in excess of 212° F (100° C). Do not short
circuit or install with incorrect polarity. DO NOT INCINERATE.
When measuring flammable liquids, observe precautions
against fire or explosion.
When handling hazardous liquids, always follow the liquid
manufacturer’s safety precautions.
This device will operate using off-the-shelf AAA size
alkaline or lithium batteries. Do not mix brands or
technologies. Do not mix old and new batteries. Do not use rechargeable
batteries.
•When working in hazardous environments, always exercise appropriate
safety precautions.
•Open battery cells should be disposed of in accordance with local
regulations. Lithium batteries are best disposed of as a non-hazardous
waste when fully or mostly discharged. EPA does not list or exempt lithium
as a hazardous waste. If waste lithium batteries are still fully charged or
only partially discharged, they can be considered a reactive hazardous
waste because of unconsumed lithium remaining in the battery. Such
batteries may qualify as "Universal Waste" in many jurisdictions within the
U.S. and thus can be shipped for disposal or recycling in accordance with
Universal Waste requirements.
Minimum Display: 0.001
Maximum Display: 999,999 (x100)
Field
Calibration
Internal Power Supply: (2) Alkaline AAA batteries @ 1.5-volts each
Optional External Power Module: 5-26-volts DC
STANDARD FEATURES INCLUDE
• (2) Totalizing Registers
• (1) Field Calibration Curve
• Rate of Flow Feature
minutes, or Seconds
COMPUTER ELECTRONICS TERMINAL CONNECTIONS
Figure 1
GETTING STARTED
Operating
Temperature
SAFETY /
SAFETY /
SPECIFICATIONS
SPECIFICATIONS
Pulse
K-Factor
Transparent Amorphous Nylon
If wider operating temperature ranges are desired, reference
-40°F to +158°F (-40°C to +70°C)
Frequency Range: 0.25 Hz – 3kHz
Electrical / Mechanical
Specifications
Calibration
ASSEMBLY /
ASSEMBLY /
INSTALLATION
INSTALLATION
Power
• (1) Factory Calibration Curve
TROUBLESHOOTINGOPERATION
TROUBLESHOOTINGOPERATION
REPAIR
REPAIR
6
MAINTENANCE /
MAINTENANCE /
FEMALE TERMINAL HOUSING
(PICKUP COIL CONNECTION)
Minimum: -99.999%
Maximum: +99.999%
Yes
Battery Life: 2+ years
• Flowrate Time Period in Day, Hour,
SPECIFICATIONS (CONTINUED)
DIMENSIOINS
Length
Height
Height
“C”
Width (Widest Point)
3.40 in.
(8.6 cm)
0.85 in.
(2.1 cm)
0.72 in.
(1.8 cm)
2.14 in.
(5.4 cm)
COMPUTER DISPLAY FEATURES
FACTORY CALIBRATION
INDICATOR
FLOATING DECIMAL
POINT (3 PLACE)
RATE
BUTTON
TOTAL
BUTTON
FLOW INDICATOR
INFORMATION ROW
BATCH TOTAL,
ACCUMULATIVE TOTAL,
OR FLOWRATE
TOTAL SPILL-OVER
INDICATOR
(X10 OR X100)
TOTAL INDICATOR
(BATCH & ACCUMULATIVE)
BATCH INDICATOR
GETTINGSTARTED
GETTING STARTED
“A”
Figure 2
“B”
A
(Mounted)
“D”
SPECIFICATIONS
SPECIFICATIONS
SAFETY /
SAFETY /
D
INSTALLATION
INSTALLATION
ASSEMBLY /
ASSEMBLY /
C
B
OPERATIONTROUBLESHOOTING
OPERATIONTROUBLESHOOTING
Familiarize yourself with the computer features before installation and use.
Figure 3
MAINTENANCE /
MAINTENANCE /
REPAIR
REPAIR
7
SPECIFICATIONS (CONTINUED)
AGENCY APPROVALS
Approval Ratings
GETTING STARTEDOPERATION
SAFETY /
ASSEMBLY /
Q9 Computer Electronics:
SPECIFICATIONS
INSTALLATION
TROUBLESHOOTING
REPAIR
8
MAINTENANCE /
INSTALLATION
PRODUCT DESCRIPTION
This computer electronics is designed specifically for use on FLOMEC® Industrial
Grade
modules.
The CMOS, microprocessor
r
clearly displayed on
totals from .001 to 999,999 (x1),
operations are easily accessed with the two
In a FLOMEC Industrial Grade turbine met
causing an internal rotor to spin. As the rotor spins, an electrical signal is generated
in the
units shown
INSTALLATION
Before installation, ensure your computer model meets your specific needs. Refer to
the
computer is
computer housing on the floor of the battery holder.
If you ordered your computer electronics with a turbine body,
installed at the
If
coil connector and
the corners of the faceplate. Make sure the
screws.
If you ordered the computer with a turbine body and an acce
review and thoroughly un
All FLOMEC turbine meters are designed to measure flow in only one direction. The
d
upside down
and
Figure 4
Installation
Turbine Housings. It is also designed to work with several accessory output
equirements and data retention capabilities in both RAM and ROM. Information is
-based electronics have extremely low power
a large 6-digit LCD readout with three-point floating decimal for
9,999,990 (x10), or 99,999,900 (x100). All
buttons on the front panel.
er, liquid flows through the turbine housing
pickup coil. This pulse data from the turbine is translated into calibrated flow
on the computer's display readout.
Specifications Section to confirm required features. The model number of your
displayed on the outside wall of the computer housing and also inside the
the electronics are
factory.
you ordered your computer separately as a replacement, simply plug in the pickup
mount the computer on your turbine body with the four screws at
seal is fully seated before tightening the
ssory module, please
derstand all installation instructions before proceeding.
irection is indicated by the arrow on the turbine outlet port. If the computer display is
in your installation, remove the four screws, turn the display 180 degrees
reinstall the screws. (See Figure 4)
NOTE: When rotating the computer
display, it is not necessary to
disconnect the pickup coil connector,
however, care should be taken to
avoid inadvertent strain on the
connector wires.
GETTING STARTEDOPERATIONTROUBLESHOOTING
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
MAINTENANCE /
REPAIR
9
OPERATION
COMPUTER DISPLAY
ACTIVATE THE COMPUTER
Operation
SAFETY /
ASSEMBLY /
All operations are revealed on the LCD using the large 6-characters in the top row and
smaller characters and symbols in the second row. These characters and symbols
GETTING STARTED
indicate information regarding totals, flow, calibration, units of measure and
operational messages.
Push button operation varies dependent upon the various modes of operation, i.e.
Normal Operation mode, Field Calibration mode, and User Configuration mode. Their
operation will be described in their respective sections.
When batteries are installed, the computer is on continuously and always ready to
perform.
SPECIFICATIONS
The computer is powered by field replaceable commercially available batteries.
Reference the Maintenance Section for battery replacement details.
INSTALLATION
Figure 5
Figure 6
OPERATIONOPERATION
General
When batteries are initially installed or
replaced, the initialization routine will start
the LCD display blank, then display
“HELLO” on the top row and “Q9Disp” on
the information row for one second. (See
Figure 5)
The LCD will then display “HELLO” on the
top row and “FW Vxx” on the information
row for one second. The Vxx will be the
version of the software installed on the
display. Example: “FW V03” indicates
firmware version 3 installed on the display.
(See Figure 6)
The computer maintains two totals; Batch total and accumulative total. The batch
total can be reset to measure flow during a single use. The accumulative total
provides continuous measurement and can only be reset by removing the batteries,
holding down the Total button, and replacing the batteries.
The button usage map on the next page, and the user configuration process map later
in the manual, is useful for understanding where the various menus are located within
the software programming and the route to get to a specific menu.
Each of these maps are followed by user instructions explaining each menu.
TROUBLESHOOTINGOPERATION
REPAIR
10
MAINTENANCE /
OPERATION
NORMAL OPERATION MODE
Normal Operation
Mode
Button Usage Map – Normal Operation Mode
BUTTON
Total
Total
Rate
See InDesign for Graphic
Rate
Rate
Button Operation (Normal Operation Mode)
Toggles Between
When Batch Total
is Displayed
Change Display from
Total to Rate
When Rate is
Displayed
When Rate is
Displayed
Toggles Between
Hold for 3 Seconds
to Enter Diagnostics
Batch Total
Accumulative Total
Hold for 3 Seconds
to Reset Batch Total
Flowrate
Velocity
Input Frequency Hz
Output Frequency Hz
Battery Status
Flow Status
To Exit: Press Rate button,
or use “Timed-Out” exit.
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
OPERATIONOPERATION
Total Button: Toggle between batch total and accumulative total.
When a total is displayed, momentarily pressing Total button toggles the top row of
large display digits between batch total and accumulative total. The information row
will change to the proper units and the corresponding icons will also be shown, i.e.
TOTAL or BATCH TOTAL. (See Figure 7)
(Batch Total-Litres) (Accumulative Total-Gallons)
Figure 7
MAINTENANCE /
REPAIR
11
OPERATION (continued)
NORMAL OPERATION MODE (continued)
Total Button: Batch total reset.
GETTING STARTED
When a batch total (see Figure 8-1) is displayed, press and continue to hold Total
button for 3 seconds; the software will display a three second count down, then reset
the batch total to zero. (See Figures 8-2 thru 8-4)
NOTE: If Total button is released prior to count
down completion, the software returns to batch
SAFETY /
Figure 8-1
SPECIFICATIONS
Figure 8-2
ASSEMBLY /
INSTALLATION
Figure 8-3
OPERATION
OPERATION
Figure 8-4
NOTE: When the accumulative total is displayed in Normal Operation Mode, it cannot
8-5).
Figure 8-5
TROUBLESHOOTINGOPERATION
total screen. (See Figure 8-1)
NOTE: After count down completes, display will
show 0.000 until user releases Total button.
(See Figure 8-3)
NOTE: After Total button is released, the
display will return to Batch Total screen and will
increment the total if flow is detected. (See Figure
8-4)
be reset using the Total button. Accumulative total
can only be reset by removing the batteries, holding
down the Total button, and replacing the batteries. If
the user attempts to reset it using the Total button,
the software will display a “Reset Denied” message
on the LCD until the button is released. (See Figure
REPAIR
12
MAINTENANCE /
OPERATION (continued)
NORMAL OPERATION MODE (CONTINUED)
Diagnostic Mode
GETTING STARTED
Rate Button: Change display from total to rate.
When a total is displayed, momentarily pressing the Rate button changes the top row
of large display digits from a total to a rate display. The information row will change to
the proper units and the corresponding icons will also be shown, i.e. GPM, L, etc.
(See Figure 9)
Figure 9
Rate Button: Change display from rate to velocity.
When a rate (see Figure 9) is displayed, momentarily pressing the Rate button
toggles the display between flow rate and velocity. The information row will change to
the proper units and the corresponding icons will also be shown. (See Figure 10)
(Flowrate) (Velocity)
Figure 10
Rate Button: Display diagnostic menu.
When a rate is being displayed, press and hold the Rate Button for 3 seconds; the
software will display a three second diagnostic count down (see Figure 11-1), then
display the first of four diagnostic screens (see Figures 11-2 thru 11-6).
NOTE: To go back to a total when a rate is
displayed, press Total and the display
changes to the last shown total, either
accumulative or batch. The information
row will change to the proper unit and the
corresponding icons will also be shown, i.e.
TOTAL or BATCH TOTAL.
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
OPERATIONOPERATION
Figure 11-1
NOTE: The diagnostic screens are reference information screens only. They cannot
be edited.
Within the diagnostic menu: Momentarily pressing the Total button will advance to
the next screen; pressing the Rate button at any time will exit the diagnostic menu.
Upon exit, the display will return to the rate display, and upon releasing the button will
continue on the last shown total, either accumulative or batch.
There is a 90 second “no action”clock running when in diagnostic mode. After 90
seconds with no button push to reset the clock, the computer will automatically exit
and return to the rate display.
MAINTENANCE /
REPAIR
13
OPERATION (continued)
NORMAL OPERATION MODE (CONTINUED)
Screen 1 – Input Frequency
After button is released, the software will display the input frequency in Hz.
GETTING STARTED
Figure 11-2
SAFETY /
Screen 2 – Output Frequency
SPECIFICATIONS
Advancing to next screen will display the output frequency in Hz.
Figure 11-3
ASSEMBLY /
INSTALLATION
Screen 3 – Battery Status
Advancing to the next screen will display the battery status.
Figure 11-4
OPERATION
Figure 11-5
Screen 4 – Flow Status
Advancing to the next screen will display the flow status.
TROUBLESHOOTINGOPERATION
Figure 11-6
REPAIR
14
MAINTENANCE /
The top row is used to indicate frequency.
The information row will display messages.
(See Figure 11-2)
The top row is used to indicate frequency.
The information row will display messages.
(See Figure 11-3)
The top row is used to indicate battery life
with each dash indicating a percentage of
battery life remaining. The information row
will display messages. (See Figure 11-4)
NOTE: If the battery life is sufficiently low,
“LOWBAT” will be displayed in the
information row. This low battery message
will be displayed automatically without
running a diagnostic battery check.
The top row is used to indicate flow rate in
the volume units programmed into the
computer. The information row will display
messages, i.e.: No Flo, Lo Flo, Hi Flo.
OPERATION (continued)
USER CONFIGURATION MODE
Configuration
Mode
General Notes for User Configuration Mode Menus
1. The term “focus” as used in this manual indicates an active editable item.
Focused items will constantly flash ON/OFF.
2. Advance focus: Toggle (momentarily press) the Total button. Advancing focus
will move focus to the next editable item to the right. If already on right most item
the focus will go back to the start position in a round robin fashion. If only (1) item
is editable, advancing focus will move focus to next menu.
3. Increment the value of editable item: Toggle the Rate button. If editing a
number, the number will increment in a round robin fashion. If editing a text option,
focus will advance thru a predefined list of text depending on the active menu and
item. Advancing to the next item or menu with a unit in focus will select that unit.
4. Advance focus to next menu: Press and hold the Total button and toggle the
Rate button.
5. Retrogress focus back to previous editable item or previous menu: Press
and hold the Rate button and toggle the Total button.
6. Exit any user configuration mode menu: Press and hold Total and Rate
buttons simultaneously for 3 seconds.
NOTE: There is a 90 second “no action”clock running when in user configuration
mode. After 90 seconds with no button push to reset the clock, the computer will
automatically exit and return to normal operating mode. There is no automatic “save”, all entries made up to that point will be lost.
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
Entering User Configuration Mode
While in normal operation mode, the user can enter the display user configuration
mode by pressing and holding the Total and Rate buttons simultaneously for 3
seconds from any screen; the software will display a three second count down
sequence (see Figure 12-1), then reset the display to configuration mode. (See Figure
12-2)
Figure 12-1
Figure 12-2
NOTE: If the user releases both or either button before the countdown completes, the
display will return to normal operation mode.
NOTE: After three seconds the LCD will display
“ConFig” on the top row and “Mode” on the bottom
row until the user releases both of the buttons; the
display will then immediately show the first menu
(PIN) of the configuration process menus.
15
OPERATIONTROUBLESHOOTING
OPERATION
MAINTENANCE /
REPAIR
USER CONFIGURATION MODE
General Notes for User Configuration Mode Menus
Once in configuration mode the user can then proceed thru the configuration process
GETTING STARTED
menus as shown on the “configuration process map” on the following page.
The software will retain your entries as you progress through the user configuration
menus, however, there is no automatic “save” when exiting. Instead, you will be
shown a “save” screen to enter “yes or no” to save your entries. (See Figure 13)
SAFETY /
SPECIFICATIONS
Figure 13
Configuration Process Map – User Configuration Mode
Use the Rate button to toggle between
“yes” or “no”. Use the Total button to
advance to the next display. Advancing to
the next display will perform the “save”
selection you made and return the display
to normal operation mode. The last shown
screen in that mode will be displayed.
ASSEMBLY /
OPERATION
TROUBLESHOOTINGOPERATION
INSTALLATION
See InDesign for Graphic
Save?
Yes or No
Exit
Message
Press Total + Rate for 3 Seconds to Enter
User Configuration Mode
Enter PIN
Incorrect
Involuntary
Flowrate / Time Period
Advanced Options
Correct
Exit
Reset PIN?
Volume Unit
Display Update
Rate
Low Frequency
Cutoff
Calibration
Method
Pipe
Diameter
Daughter Board
Options
Rate+Total
Buttons
To Exit
YesNo
Field Calibration Method Entry If
Dispense/Display, Percent Adjust or K-Factor Entry
Field Calibration Methods
REPAIR
16
MAINTENANCE /
Normal Operation Mode
Custom Volume Unit
Calibration
Method
Field Calibration
Method Entry
Must Pass Thru
Were selected
Save?
Yes or No
Exit
Message
OPERATION (continued)
USER CONFIGURATION MODE (continued)
PIN
PIN Menu
The first menu in the configuration mode is for the user to enter a Personal
Identification Number (PIN) to secure the configuration of the computer and continue
with making changes.
The default user PIN is 0000. Using this default PIN will unlock the meter unless the
PIN has been reset by the user.
Figure 14-1
Figure 14-2
NOTE: Using the user default PIN, or the user specific PIN (if previously reset), and
advancing focus to the next menu, the software will compare the entered PIN with the
PIN stored in the computer. If a match is found, the next menu will be shown allowing
the user to reset the PIN if desired.
Reset PIN Menu
If the PIN was entered correctly, the next menu will be displayed and allow reset of
the PIN if desired. (See Figure 15-1)
Figure 15-1 Figure 15-2
The “N” will be in focus (flashing). To decline resetting the PIN, advance focus to next
menu.
To reset the PIN, increment the “N” to “Y” (see figure 15-2) and advance focus to next
screen. (See Figure 16)
The initial user PIN entry screen will show
0000. The left most digit will be flashing
ON/OFF to indicate focus if the user wants
to change the PIN. (See Figure 14-1)
If the entered PIN was incorrect, the screen
advances to the “Config Exit” menu (See
Figure 14-2) momentarily, then returns
automatically to normal operation display.
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
OPERATION
MAINTENANCE /
REPAIR
17
OPERATION (continued)
USER CONFIGURATION MODE (continued)
GETTING STARTED
Figure 16
SAFETY /
ASSEMBLY /
Volume Unit Menu
SPECIFICATIONS
On this menu the user will be allowed to enter their volume unit of measure from a list
of options. (See Figure 16).
INSTALLATION
Figure 17
OPERATION
Figure 18
Switching between different volume units will not corrupt the Total's contents. For
example, in GAL (gallon) mode, the computer totalizes 10.000 gallons, if the user
switches to L (litre) mode, the display will read 37.854 litres (the same volume,
different unit).
Custom Volume Unit Label Menu
Volume
Unit
If “Y” was selected, the previous PIN will be
shown. The left most digit of PIN is in
focus. Increment the value of each digit as
desired as you advance focus through the
digits.
After reset of the PIN, advance focus to
next menu.
On the volume unit menu (see Figure 18)
the bottom row of characters will be in
focus to indicate they are editable.
Advancing to the next menu with a unit in
focus will select that unit of measure.
TROUBLESHOOTINGOPERATION
Figure 19
REPAIR
18
MAINTENANCE /
The software also allows the user to create
their own “Custom” volume unit (see Figure
19). Advancing to the next menu with the
“Custom” volume unit in focus will shunt
the user to a new menu (see below) that
will allow the user to create a new label, or
reuse one already stored in computer
memory.
OPERATION (continued)
USER CONFIGURATION MODE (continued)
If the computer is new or has no previous custom volume unit label stored in memory,
the screen will appear with three dashes. (See Figure 20)
Figure 20
Figure 21
NOTE: Once the custom volume unit label is created, the user will be required to
enter a calibration method for the custom volume unit created. Therefore, upon
advancing focus to next menu, the user will be shunted to the calibration method
menu to select a method of calibrating their custom volume unit.
NOTE: The calibration method section that defines the methods of calibration can be
found later in this manual.
After selecting the calibration method, advance focus to the next menu that will be
flow rate/ time period menu.
NOTE: The computer will remember your calibration method selection. If you
selected field calibration methods dispense display, percent adjust, or K-factor
method, you will be given the opportunity to configure that method at the point you
decide to exit the user configuration mode. At that point, the software will
automatically shunt you to the field calibration method entry menu.
The field calibration method entry section that defines the field calibration procedure
of the selected method can be found later in this manual.
NOTE: The factory calibration stored in the unit will never be lost
The left most of the dashes will be in focus
and the user can select a letter from the list
of upper case letters. An empty space will
be shown on the menu screen as an
underscore “_”during creation, but the
underscore will not show when displaying
the unit label in operation.
If the user has previously entered a custom
volume unit label, it will be read from
computer memory and the unit label will be
displayed (see Figure 21) with the left most
character in focus. The label can then be
changed if desired.
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
OPERATION
MAINTENANCE /
REPAIR
19
OPERATION (continued)
USER CONFIGURATION MODE (continued)
Flow Rate / Time Period Menu
GETTING STARTED
This menu is used to choose the period of time over which the flow rate is calculated.
SAFETY /
ASSEMBLY /
The options for this menu are: per day, hour, minute, or second. (See Figure 22-1)
Time Base
Day
Hourh
Minute
SPECIFICATIONS
When the screen is displayed, the software will show the current selected volume unit
and time period (see Figure 22-2) and will display both in the bottom row with a backslash between them. The time period unit will be in focus and can be edited by the
user.
After advancing to the next configuration menu, the time period will be saved to
computer memory and used in the flow rate calculation.
INSTALLATION
Seconds
Figure 22-1 Figure 22-2
Abbreviati on
d
min
Time
OPERATION
TROUBLESHOOTINGOPERATION
REPAIR
20
MAINTENANCE /
OPERATION (continued)
USER CONFIGURATION MODE (Advanced Options)
Advanced Options
Refresh Rate / Low
Frequency Cutoff
Advanced Options Menu
This menu is used to access the Advanced Options. The Rate button is used to
toggle between NO and YES selections. The Total button is used to select either NO
or YES for the Advanced Options menu.
Select NO to return to Normal User
Configuration menu (See Figure 23-1).
Figure 23-1
Figure 23-2
Display Update Rate Menu
Select YES to enter into the Advanced
Options menu (See Figure 23-2).
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
This menu allows a validated user to change the update rate of the display for flow
rate or advance focus to the next menu.
The range of this menu is from 1 to 60
seconds. The factory default rate is 5
seconds. Although the computer is
maintaining constant calculations, the rate
that the new information is displayed is
controlled by this menu. (See Figure 24)
Figure 24
Low Frequency Cutoff Menu
This menu allows a validated user to enter a minimum allowed pulse input frequency
to filter out any low frequency input pulses which can be generated when the meter is
subject to mechanical vibrations. These vibrations could cause the totalizer to add
“phantom” volume to the display totalization values. (See Figure 25)
The software will read the minimum input frequency from the computer and display it
on the top row with the left most digit in focus. Increment range is from 0 to 255 Hz.
If 0 is set, the software will disable the input
frequency filtering algorithm.
When advancing out of this menu, the
software will save the selected value to the
computer.
21
OPERATIONTROUBLESHOOTING
OPERATION
MAINTENANCE /
REPAIR
OPERATION (continued)
USER CONFIGURATION MODE (Advanced Options - continued)
ASSEMBLY /
Calibration Method Selection Menu
GETTING STARTED
This menu is used to define which calibration method will be used in normal operation.
The calibration method the user previously selected (if any) will be shown on the
display first. The bottom row of characters will be in focus indicating that the user can
change the calibration method. Use the Rate button to move through the various
calibration method options.
Factory Calibration: All units are configured with a "factory" calibration. This factory
calibration (FAC icon displayed below 6-digit display) is permanently programmed into
the computer and is not user adjustable.
SAFETY /
NOTE: The factory calibration stored in the unit will never be lost
SPECIFICATIONS
INSTALLATION
Figure 26-1
Field Calibration: If one of the following calibration methods is selected (dispense
display, percent adjust, or K-factor entry), the user will be routed through the field
calibration method entry menu, before saving configuration options, to make specific
entries for the selection.
OPERATION
Figure 26-2
TROUBLESHOOTING
Figure 26-3
Factory calibration is the typical use
method (see Figure 26-1). Prior to
shipping from the factory, the display will
be configured and calibrated to the meter
body on which it is installed (or going to be
installed).
When the user selects this option the FAC
icon will be “ON” to indicate that the display
is using its factory calibration table. This
method does not support field calibration.
Dispense/Display is a field calibration
method by which the user will dispense a
known volume of fluid and the software will
keep track of the pulse count during the
dispense operation. The user will then
enter the known volume into the display
and the software will calculate a K-factor
for the volume dispensed.
(See Figure 26-2)
Percent Adjust method is a field
calibration method by which the user will
enter a percentage amount to adjust Kfactor values in the factory programmed Kfactor table. Once the percentage is input
by the user, the software will apply this
percentage adjustment to all calculated
and displayed values. (See Figure 26-3)
This method is good for users who simply
want to adjust their meter by a percentage
amount.
REPAIR
22
2222
MAINTENANCE /
OPERATION (continued)
USER CONFIGURATION MODE (Advanced Options - continued)
Inner Pipe
Diameter
GETTING STARTED
Figure 26-4
Pipe Diameter Menu
This menu is to allow the user to enter a pipe inner diameter (ID) which will allow the
display to calculate and display the fluid velocity thru the meter. The pipe dimension
unit can be selected as either inches or millimetres. (See Figures 27-1 & 27-2)
Figure 27-1
Pressing Rate will toggle between inches (IN) and millimetres (mm). Pressing Total
will save the selection and allow the user to edit the pipe inner diameter value.
K-Factor Entry method (see Figure 26-4)
is a field calibration method by which the
user can input either a single point or
multipoint calibration table.
From 1 to 5 unique K-factor and frequency
values can be entered and will be stored in
the custom K-factor table in the computer.
On this screen, the software will read the
pipe ID measurement unit label from the
computer and display it on the bottom row
as either English (IN) or metric (mm) units.
The unit label (IN or mm) will have the
focus to indicate that it can be changed.
The software will also read the pipe ID
numeric value including decimal point from
the computer. The decimal point is in
focus to indicate it can be changed.
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
Figure 27-2
Press the Rate button to move the focus right to the next decimal and eventually turns
off all three. An additional button press will cycle the focus back to the left most
decimal point. Pressing the Total button will lock in the selection of the decimal point
and allow the user to edit the diameter value.
Pipe ID values can be from 0.001 to 999999. The pipe diameter entries will be used
in the fluid velocity rate calculation during normal operation.
Pressing Rate will increment the digit in focus. Pressing Total will move focus to the
next digit. Holding Total and pressing Rate will exit this menu.
Upon exit from this menu, the software will temporarily store the pipe ID entries in the
computer, and then display the “Save” screen (see Figure 13) and return to normal
operation mode.
NOTE: Failure to save your entries on the “save” screen will cause the entries to be
discarded.
MAINTENANCE /
REPAIR
232323
OPERATION (continued)
USER CONFIGURATION MODE (Advanced Options - continued)
Daughter
Board
Output K-Factor Menu
(Requires Pulse Access or 4-20mA Daughter Board Module)
GETTING STARTEDOPERATION
The scaled pulse output K-factor menu allows the user to enter a K-factor for the
pulse output of the meter or advance focus to the next menu.
The software will read the pulse output K-factor from the computer and display it on
the top row. The left most digit will be in focus to indicate that it can be changed. The
decimal can also be moved (see Figure 28).
SAFETY /
SPECIFICATIONS
Figure 28
4-20mA Menu (Requires 4-20mA Daughter Board Module)
ASSEMBLY /
INSTALLATION
Figure 29-1
Figure 29-2
Figure 29-3
TROUBLESHOOTING
When advancing focus to the next menu
the software will save the Scaled Pulse Kfactor to the computer.
Zero Setting This figure will flash for 1
second before allowing the user to change
the 4-20mA zero setting (see Figure 29-1).
This setting will already be configured for the
minimum flowrate of the meter this Q9 was
attached to when shipped from the factory.
The button presses required to change this
menu item are the same as in the Pipe
Diameter Menu above (see Figure 29-2)
Span Setting This figure will flash for 1
second before allowing the user to change the
4-20mA span setting (see Figure 29-3). This
setting will already be configured for the
maximum flowrate of the meter this Q9 was
attached to when shipped from the factory.
REPAIR
24
MAINTENANCE /
USER CONFIGURATION MODE (Advanced Options - continued)
Calibration Mode
The button presses required to change
this menu item are the same as in the
Pipe Diameter Menu above
(see Figure 29-4).
Figure 29-4
OPERATION (continued)
Field Calibration Method
General
The field calibration method may be set by the user, and can be changed or modified
at any time using a field calibration method described in this section. Totals or
flowrate derived from the field calibration are being invoked when the (FAC) icon is no
longer visible below the 6-digit display.
Factory calibration settings are programmed into each computer during
manufacturing, using Stoddard test solvent at 70° F (21° C) for meters up to 1 inch.
Meters 1-1/2 inch and larger are factory calibrated using water at 70° F (21° C).
Settings are correct for light liquids such as water, gasoline or diesel. Readings using
the factory calibration (FAC) may not be accurate in some situations, for example,
"heavy" liquids such as motor oil, under extreme temperature conditions, nonstandard plumbing configurations or with fluids other than mentioned above.
For improved accuracy under such conditions, the computer allows for field
calibration, that is, user entry of custom calibration parameters. A "single point" field
calibration curve may yield acceptable accuracy when used in a non-standard
application, however, the computer is capable of programming a “five point” field
calibration curve.
NOTE: If the calibration method is changed when in user configuration mode, i.e., not
using the default factory calibration (FAC), the programming will allow the user to
adjust the user programmable calibration table for the calibration method selected.
This is done by shunting the user to the field calibration method entry menu specific to
that calibration method upon exit from the user configuration mode.
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
Verify Accuracy before Beginning Field Calibration
For the most accurate results, dispense at a flowrate which best simulates your actual
operating conditions. Avoid "dribbling" more fluid or repeatedly starting and stopping
the flow. This can result in less accurate calibrations. Make sure you meet the
meter's minimum flowrate requirements.
The use of a uniformly dependable, accurate calibration container is recommended for
the most accurate results. A five gallon calibration container is available in the parts
section of this manual. For best results, the meter should be installed and purged of
air before field calibration.
Due to high flowrates on meters 2 inch and larger, it is strongly recommended that
field calibration be completed with a combination of volume and weight determined
with fine resolution scales.
MAINTENANCE /
REPAIR
25
OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode)
Field Calibration Method Entry
GETTING STARTED
At the beginning of the calibration method entry menu, the software will allow the user
to start the calibration process or to exit back to normal operation. (See Figures 30-1
& 30-2)
The bottom row of characters will be in focus to indicate that the user can select
between either “Start” or “Exit” by advancing focus.
SAFETY /
SPECIFICATIONS
Figure 30-1
ASSEMBLY /
INSTALLATION
Figure 30-2
Display Dispense Method
Dispense/Display is a field calibration method by which the user will dispense a known
volume of fluid and the software will keep track of the pulse count during the dispense
operation. The user will then enter the known volume into the display and the
software will calculate a K-factor for the volume dispensed. (See Figure 31-1)
OPERATION
Figure 31-1
TROUBLESHOOTING
Figure 31-2
Dispense
Display
If “Start” is in focus when advancing to the
next menu, the software will automatically
advance to the calibration menu for the
calibration method the user previously
selected.
If “Exit” is in focus when advancing to the
next menu, the software will display a
“save” screen (see Figure 13) to enter “yes
or no” to save your entries.
The software will allow the user to
dispense a known volume in order to
create a custom K-factor. This will not over
write the existing factory calibration table.
The first screen in this calibration method
will show “run 1” indicating that computer is
waiting for flow to start. (See Figure 31-2)
NOTE: Pressing any single button while
on this screen will not have any effect.
Pressing both buttons and holding will
allow the user to exit calibration mode.
REPAIR
26
MAINTENANCE /
OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode - continued)
Percent AdjustK-Factor Entry
Figure 31-3
When the user is finished with the run, press and hold the Total button until the left
most digit begins to flash indicating it is in focus, then release the Total button.
Pressing the Rate button will increment the digit in focus. Pressing the Total button
will advance focus to the next digit.
Holding Total and pressing Rate will lock in the entered volume.
The software will then calculate the K-factor for the volume entered based on the
pulse count for the run and the volume entered. The average frequency and the Kfactor will be entered into the custom user K-factor table for run 1.
Exit Dispense Display Method
Figure 32
NOTE: Failure to save your entries on the “save” screen will cause the entries to be
discarded.
Begin dispensing into a container of known
accurate volume. As soon as pulses are
detected by the software, the screen will
switch to display the volume being
dispensed on the top row and the volume
unit on the bottom row. (See Figure 31-3)
NOTE: “FldCAL” will then be displayed on
the top row and “Exit” on the bottom row
(see Figure 32). The software will
temporarily store the sorted table of
frequencies and K-factors, then display the
“Save” screen. (See Figure 13)
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
MAINTENANCE /
REPAIR
27
OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode - continued)
Percent Adjust Method (Correction Factor)
Percent adjust method is a field calibration method by which the user will enter a
GETTING STARTED
percentage amount to adjust K-factor values in the factory programmed K-factor table.
Once the percentage is input by the user, the software will apply this percentage
adjustment to all calculated and displayed values. (See Figure 33-1)
SAFETY /
SPECIFICATIONS
Figure 33-1
ASSEMBLY /
INSTALLATION
Figure 33-2
Momentarily press Rate to toggle between (dash “-“) and (underscore “_”). A dash
indicates a negative value. An underscore indicates a positive value. Press Total to
edit the adjustment value.
When all digits are entered correctly, exit by holding the Total and pressing the Rate
button. The software will store the adjusted percentage correction, then display the
“Save” screen. (See Figure 13)
NOTE: Failure to save your entries on the “save” screen will cause the entries to be
discarded.
This will not over write the existing factory
calibration table, instead, it applies a
calculation to the existing values for use by
the computer.
This method is good for users who simply
want to adjust their meter by a percentage
amount.
The top row left most character will be in
focus to indicate the user can change the
value. The left most character is the sign of
the value. The decimal point is not
adjustable. The range for the values are: -
99.999% to +99.999%. The software will
display the user applied percent
adjustment on the top row.
K-factor Entry Method
K-factor entry method (see Figure 34-1) is a field calibration method by which the user
can input either a single point or multipoint calibration table.
From 1 to 5 unique K-factor and frequency
values can be entered and will be stored in
the custom K-factor table in the computer.
This will not over write the existing factory
TROUBLESHOOTINGOPERATION
Figure 34-1
calibration table.
REPAIR
28
MAINTENANCE /
OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode - continued)
GETTING STARTED
Figure 34-2
Momentarily pressing the Rate button moves the decimal focus right to the next
decimal until reaching the end, then cycles back to the start position in a round robin
fashion. When the decimal point is moved to the correct position, pressing the Total
button will lock in its positional location and allow the user to edit the frequency value.
Pressing the Rate button will increment the digit. Momentarily pressing Total will
move to the next digit. Pressing and holding Total then pressing Rate will advance
to the next screen (see Figure 34-3).
NOTE: User input K-factor entry values will disable the (FAC) icon. The K-factor
range is 0.001 to 999999.
Figure 34-3
NOTE: To advance character value focus or move decimal point position for the K-
factor screen, the buttons operate the same as on the frequency screen above.
After completing the first table entry of frequency and K-factor, the display will ask the
user if there are additional points to enter. (See Figures 34-4 & 34-5)
Figure 34-4 Figure 34-5
If the user chooses YES, the previous screens will repeat for Frequencies 2 – 5 and
K-factors 2 – 5.
NOTE: If the table stops short of 5 points, then only the points entered will be used to
define the calibration curve.
The software will read the existing K-factor
table from the computer and will display the
first frequency on the top row (see Figure
34-2) with the left most character in focus
to indicate the user can change the value.
Minimum frequency value allowed is 0.25
and the maximum allowable frequency
value is 3000.00 Hz. The decimal point is
adjustable.
The software will read the existing K-factor
table from the computer and will display the
first K-factor on the top row. The bottom
row displays the K-factor table entry
position (1 thru 5) for the volume unit
shown. The decimal point will be in focus
to indicate the user can change the
position. The value is also adjustable (See
Figure 34-3)
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
MAINTENANCE /
REPAIR
29
OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode - continued)
If the user chooses NO, the software will show the FLdCAL Exit screen (see Figure
32) for one second, and will then exit field calibration mode. The software will
GETTING STARTED
temporarily store the K-factor table values in the computer, and then display the
“Save” screen (see Figure 13) and return to normal operation mode.
NOTE: Failure to save your entries on the “save” screen will cause the entries to be
discarded.
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
TROUBLESHOOTINGOPERATION
REPAIR
30
MAINTENANCE /
7. Meter installed too close to
TROUBLESHOOTING
Troubleshooting
GETTING STARTED
Symptom Probable Cause Corrective Action
A. DISPLAY IS
INACCURATE
1. Field Calibration not
performed properly.
2. Factory Calibration not
suitable for liquid being
measured.
3. Meter operated below
minimum flowrate.
4. Meter partially clogged
with dried liquid.
5. Turbine bearings partially
clogged with dried liquid.
6. Sealant material wrapped
around meter rotor.
fittings.
8. Meter installed to close to
motors or electrically
“noisy” environment.
B. DISPLAY IS
FADE D OR
BLANK.
C. NORMA L
FLOW RATE
BUT DISPLAY
DOES NOT
SHOW CO UNT
1. Batteries weak, dead, or
not connected.
2. Computer defective.
1. Field Calibration not
performed correctly.
2. Rotor stuck or damaged.
D. REDUCED
FLOW RATE
AND DISPLAY
DOES NOT
SHOW CO UNT
E. CANNOT GET
COMPUTER
INTO FIELD
CALIBRATION
3. Sealant material wrapped
around rotor.
4. Computer defective.
1. Meter clogged with dried
liqu id.
2. Below minimum flowrate.
1. Wrong button sequence.
2. Computer circuit board
defective.
3. Button defective.
Field calibrate again or select
Factory Calibration.
Perform a Field Calibration
according to Field Calibration
Section.
Increase flowrate.
Remove meter. Clean carefully.
Make sure rotor spins freely.
Remove meter. Clean carefully.
Make sure rotor spins freely.
If applicable, remove rotor. Make
sure rotor spins freely.
Install correctly.
Install correctly.
Remove computer. Check and
replace batteries.
Contact the factory.
Field calibrate again or select
Factory Calibration.
Remove meter. Make sure rotor
spins freely.
Remove meter. Make sure rotor
spins freely.
Contact the factory.
Remove meter. Clean carefully.
Make sure rotor spins freely.
Increase flow.
Proceed with calibration using Field
Calibration Section instructions.
Replace computer. Contact factory.
Replace computer. Contact factory.
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
MAINTENANCE /
REPAIR
31
MAINTENANCE
Maintenance
Batteries
The computer electronics are powered by alkaline batteries. Removing the batteries
GETTING STARTED
before storing the meter will extend battery life since the computer is always on (either
standby or active) when the batteries are installed.
If the meter's readout should become dim, blank or the low battery message appears
(see below), the batteries should be replaced.
Note: If the battery life is sufficiently low,
“LOWBAT” will be displayed in the
message area on the bottom row of
SAFETY /
SPECIFICATIONS
Battery Replacement Information
The computer is shipped with (2) alkaline AAA size batteries (1.5-volts each) installed.
Battery replacements are readily available as an off-the-shelf item. To maintain the
FLOMEC® warranty, the batteries listed below are approved for use.
Alkaline (AAA size, 1.5-volts each)
Energizer, Alkaline, E92 or En92
ASSEMBLY /
Duracell, Alkaline, Mn2400
INSTALLATION
Lithium (AAA size, 1.5-volts each)
Energizer, Lithium, L92
characters. This low battery message will
be displayed automatically without running
a diagnostic battery check.
Do not mix brands or technologies. Do not mix old and new batteries. Do not
use rechargeable batteries.
When batteries are disconnected or fail, the computer memory will retain the batch
total, accumulative total, factory calibration curve, and field calibration curve
indefinitely.
It is strongly recommended that battery checks and terminal cleaning be a part of a
routine maintenance schedule. Battery terminals should be cleaned annually.
Batteries can be replaced without removing the meter from the piping system.
Replace Batteries
1. Remove the (4) corner screws attaching the computer electronics to the meter and
lift the computer electronics from the meter.
2. Remove the batteries.
3. Check the battery terminals and remove any corrosion.
4. Install the new batteries and make sure the positive posts are positioned correctly.
TROUBLESHOOTINGOPERATION
When the batteries are installed correctly, the computer powers on automatically
and the display will show information.
5. Make sure the seal is fully seated before placing the computer electronics back on
the turbine. Tighten the (4) screws previously removed.
6. Do not clean exterior of computer assembly with Isopropyl Alcohol.
REPAIR
REPAIR
32
MAINTENANCE /
MAINTENANCE /
GETTING STARTED
PARTS
SERVICE
The factory, when provided with model number and serial number, can replace your
entire computer electronics assembly. Order replacement kits, parts, and accessories
with the part numbers shown below.
For warranty consideration, parts, or other service information, please contact your
local distributor. If you need further assistance, contact the GPI Customer Service
Department in Wichita, Kansas, during normal business hours.
A toll free number is provided for your convenience.
1-888-996-3837
To obtain prompt, efficient service, always be prepared with the following information:
• The model number of your computer.
(Printed on label located on bottom of one of the battery holders.)
• The serial number or manufacturing date code of your computer.
(Printed on label located on bottom of one of the battery holders.)
• Part descriptions and numbers.
For warranty work, always be prepared with your original sales slip or other evidence
of purchase date.
IMPORTANT: Please contact GPI before returning any parts. It may be possible to
diagnose the trouble and identify needed parts in a telephone call.
Description
No.
Reqd.
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
MAINTENANCE /
REPAIR
33
GETTING STARTED
SAFETY /
SPECIFICATIONS
ASSEMBLY /
INSTALLATION
NOTES
TROUBLESHOOTINGOPERATION
MAINTENANCE /
REPAIR
34
NOTES
GETTING STARTED
SPECIFICATIONS
SAFETY /
INSTALLATION
ASSEMBLY /
OPERATIONTROUBLESHOOTING
MAINTENANCE /
REPAIR
35
Warranty
FLOMEC® TWO-YEAR LIMITED WARRANTY
Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205,
hereby provides a limited warranty against defects in material and workmanship on all
products manufactured by Great Plains Industries, Inc. This product includes a 2 year
warranty. Manufacturer’s sole obligation under the foregoing warranties will be limited
to either, at Manufacturer’s option, replacing or repairing defective Goods (subject
to limitations hereinafter provided) or refunding the purchase price for such Goods
theretofore paid by the Buyer, and Buyer’s exclusive remedy for breach of any such
warranties will be enforcement of such obligations of Manufacturer. The warranty shall
extend to the purchaser of this product and to any person to whom such product is
transferred during the warranty period.
The warranty period shall begin on the date of manufacture or on the date of
purchase with an original sales receipt. This warranty shall not apply if:
A. the product has been altered or modified outside the warrantor’s duly appointed
representative;
B. the product has been subjected to neglect, misuse, abuse or damage or has been
installed or operated other than in accordance with the manufacturer’s operating
instructions.
To make a claim against this warranty, contact the GPI Customer Service Department at
316-686-7361 or 888-996-3837.
Or by mail at:
Great Plains Industries, Inc.
5252 E. 36th St. North
Wichita, KS, USA 67220-3205
The company will step you through a product troubleshooting process to determine
appropriate corrective actions.
GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY
FOR DIRECT, INDIRECT, INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED
IN THE USE OR LOSS OF USE OF THE PRODUCT WARRANTED HEREUNDER.
The company herewith expressly disclaims any warranty of merchantability or fitness for
any particular purpose other than for which it was designed.
This warranty gives you specific rights and you may also have
other rights which vary from U.S. state to U.S. state.
Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT –
Part 702 (governs the resale availability of the warranty terms).