Industrial Oval Gear Flowmeters
with pulse output or electronic display
Operation Manual
Your flow-meter and any associated electronics are precision instruments,
CAUTION
•
General Information
This manual provides the necessary information for installation of your Oval Gear
flowmeter; for information on any integral electronics or accessories fitted to your
flowmeter please consult the relevant electronics or accessory manual. Your Oval Gear
flowmeter should only be installed by persons familiar with local regulations, particularly
those for workplace Health and Safety, or Hazardous Area regulations where relevant.
For best results, please make yourself familiar with the contents of all relevant product
manuals prior to installation and commissioning. If further assistance is required please
consult the distributor from whom you purchased your flowmeter.
to avoid unnecessary damage please treat them with care.
DISPOSAL WITHIN THE EUROPEAN UNION - WEEE
The WEEE Directive requires that this product be recycled
when disposed of within the European Union
•The crossed out wheelie bin symbol shown in this manual
signifies that this product should not be disposed of in general
waste or landfill.
•Please contact the local dealer or national distributor from
whom this product was purchased for information on recycling
electronic equipment within your region.
Users installing product in Hazardous Areas must read this entire manual
CAUTION
The ‘quick start’ instructions shown below are intended for users who are experienced in
the use of flowmeters and who want to quickly set up their new meter with limited
functionality, and start using their product right away. The ‘quick start’ instructions will
allow the user to set up their meter without the risk of damage, allowing the use of the
product while the complete instructions are read in detail at a later date.
before installing their product.
Damage caused to meters by users who have only read the ‘Quick Start Guide’
will not be accepted as a justification for a warranty claim; if you are unsure,
read the whole manual before installing your meter.
4
1. Introduction
The Oval gear meter is a precise positive displacement flowmeter incorporating a pair of
oval geared rotors. These meters are capable of measuring the flow of a broad range of
clean liquids.
Stainless Steel flowmeters are suited to most liquid products and chemicals; including
many water based liquids, acids, bases and salt solutions, and Aluminium meters are
suitable for fuels, oils & most non-aggressive lubricating liquids.
Oval Gear flowmeters are available as blind meters with a pulse signal output capable of
interfacing to most monitoring and control instrumentation, or the meter can be fitted with
instruments such as totalisers, rate totalisers or batch controllers. These instruments also
have monitoring and control output options including 4-20mA, scaled pulse, flow-rate
alarms and batch control logic (preset metering).
1.1 Operating Principle
Oval Gear flowmeters are positive displacement devices where the passage of liquid
causes two oval geared rotors to rotate within a precision measuring chamber. Each
rotation of the Oval rotors will transmit 4 identical volumes of liquid from the meter inlet to
outlet (as shown in the diagram below); providing electronic pulses via magnetic sensors to
a digital instrument.
5
1.2Specifications
Model Prefix
004
006
008
Nominal Size
1/8” (4mm)
1/4” (6mm)
3/8” (8mm) 1
Flow range2 (USG/hr)
0.26 ~ 9.5
0.5 ~ 27
4 ~ 145
Flow range2 (L/hr)
1.0 ~ 36
2 ~ 100
15 ~ 550
Accuracy ( liquids ≥ 3cP)
± 1% of reading (± 0.2% with optional RT14)3
Repeatability (liquids ≥ 3cP)
Typically ± 0.03%
Temperature Range
-4oF ~ +250oF (-20oC ~ +120oC)4
Pressure Ratings – PSI (Bar) – threaded meters only
Reed Only Option (Intrinsically Safe Simple Apparatus)
Voltage
24V (dc) Maximum
1
Current
16mA Maximum
2
Power
0.4W Maximum3
1. OM008H meter have a nominal port size of ¼” (6mm)
2. Maximum flow rate must be reduced with increased viscosity, maintain maximum pressure drop across the
meter at below 14.5psi (1 Bar)
3.0.2% accuracy achievable using RT14 with non-linearity correction and multi-point calibration.
4. Temperature range stated for standard pulse output meters; higher and lower temperature rating options are
available. Meters fitted with integral instruments will have a reduced maximum temperature. OM008 meters
fitted with PPS rotors are limited to +176
5.Flanged meter pressure rating is in accordance with applicable flange standard, or with threaded meter
rating, whichever is lower.
6. Filtration requirements are for soft particles only; hard particles of any size are not acceptable.
o
F (+80oC).
1.2.4 Electrical Specifications
Hall Effect Output
Reed Switch Output
Reed Switch Output
(per switch)
1. 30V(dc)maximum peak voltage allowed for non-hazardous (safe area) installations
2. Up to 200mA possible with internal current limit bypassed (not recommended, safe area only)
3.Up to 6W possible with internal current limit bypassed (not recommended, safe area only)
Maximums for
7
5V (dc) @ 10mA
2. Installation
Please note that all flow-meters are calibrated with either Castrol ISO4113
available from the manufacturer or via an internet search.
All flowmeters are inspected and calibrated prior to shipment, and are sent out in perfect
condition. Should damage be present on receipt of the product please inspect the delivery
packaging for visible mishandling and contact the parcel service / freight forwarder.
Maintain any protective plugs/caps until installation.
or Exxsol D130 immediately prior to shipment, residual oil may be present;
please take the appropriate precautions for health and safety. An MSDS is
2.1 Mechanical Installation
Before installing your flowmeter, it is recommended that you confirm the meter is
suitable for your application conditions such as; fluid compatibility with meter materials,
flow rate, pressure, and temperature. Fluid entering the meter must remain a liquid at all
times; avoid solidification or gelling of the metered medium. If hydraulic shock or pressure
surges of any kind are possible, the system upstream of the meter must be fitted with a
surge suppressor or pressure relief valve to protect the meter from damage.
2.1.1 Installation Orientation
The flowmeter MUST be mounted with the rotor shafts in a horizontal plane. Failure to
mount your Oval Gear flowmeter in the correct orientation (as shown in the diagrams below) will cause the weight of the rotors to bear down on the thrust bearings. The short
term effects of incorrect mounting orientation will be a loss in accuracy, with long term
effects ranging from reduced lifespan to fairly rapid catastrophic damage.
Liquid can flow in a horizontal direction, or a vertical direction, but in each case the rotor
shafts must be in a horizontal plane. This is achieved by mounting the meter so that the
terminal cover, or integral instrument display, is facing in a horizontal direction. For pulse
output flowmeters the direction of flow is not important, as the output is non-directional.
CAUTION
8
2.1.2 Piping Construction
Strainer
Flowmeter
Isolation valve
Isolation valve
By-pass valve
It is preferred to install the flowmeter upstream of a flow control or shut-off valve, as the
back pressure provided by the valve will be beneficial to system accuracy; do not operate a
flowmeter directly discharging to atmosphere. Piping should be designed so that the
flowmeter is full of liquid at all times; this is achieved by designing the inlet and outlet
piping for the flowmeter to be lower than all surrounding piping.
For vertical installations the liquid should travel from bottom to top, i.e. it should rise
vertically through the flowmeter. This will ensure that the flowmeter remains full of liquid
and will stop air entrapment in the meter.
All piping surrounding your flowmeter should be well supported on secure footings,
preferably at the point where the piping joins to the flowmeter; unsupported piping will
cause severe pipe stress on the flowmeter.
The best piping designs provide a bypass line, as shown in the following diagram, which
allows isolation of the flowmeter and strainer from the main process line. The benefits of
installing a bypass line are that your system can be purged before start-up, and any
maintenance on the flowmeter or strainer can be carried out quickly and economically
without interrupting critical processes.
2.1.3 Mechanical Support
All flowmeters of nominal size 3” and 4” must have adequate mechanical support. Failure to
adequately support the flowmeter and the connected piping may result in structural damage
to the flowmeter housing. It is recommended that a solid structural support be provided at the
first flange connection, on both sides of the flowmeter. If possible, the housing of the
flowmeter can be directly supported, however piping should still be supported as close as
possible to the flowmeter. It is never acceptable to support the flowmeter and use the housing
of the flowmeter to support the weight of your piping.
9
2.1.4 Filtration / Straining
Flange bolts in Aluminium flanges should never be tightened to torque
CAUTION
It is recommended to fit the inlet side of your flowmeter directly to a strainer of
appropriate pipe size and mesh size. The minimum mesh size as shown in the specifications
section of this manual should be adhered to where ever possible. When metering medium
or high viscosity fluids it may be necessary to use a strainer one pipe size larger than the
flowmeter nominal size, in order to limit the pressure drop across the strainer basket and
maintain strainer efficiency (i.e. a 1.5” strainer on a 1” meter).
In systems where there is potential for hard particles of any size, the filtration levels of the
system should be improved so as to eliminate the flow of hard particles through the meter.
While soft particles may pass through the rotating meter components without damage (if
they are small enough) hard particles are abrasive and will always cause rotor and bearing
damage regardless of size.
2.1.5 Pipe Connections
When installing a flowmeter, it is important that no upstream pipe join between the
flowmeter and the strainer are sealed using PTFE sealing tape. Lose pieces of PTFE tape
are common causes of failure in new meter installations due to the tendency to wrap
around the rotating components inside the meter. It is recommended that the sealing of
pipe joins should be done with a sealing paste or liquid (such as Loctite® 565 thread
sealant paste).
Flanged connections should be made using gaskets and bolting suitable for the metered
liquid and the system pressure; flowmeters fitted with stainless steel flanges may be
installed using metallic gaskets and high strength bolting if required. For meters fitted
with Aluminium flanges it is only suitable to make pipe connections in accordance with
the relevant flange standard requirements for cast iron flanges; gaskets according to
ASME B16.5 Annex C gasket group 1a are recommended, and only low strength or
intermediate strength bolting may be used.
For all flanged pipe connections to flowmeters it is essential that flange faces are well
aligned and closely fitting.
values greater than 110ft.lb (150Nm)
10
2.2 Electrical Installation
Two types of output are available from an Oval Gear flowmeter; NPN Hall Effect and Reed
Switch (contact closure). Some meter configurations will have one of these outputs, some
will have both; a flowmeter may be installed using any of the available outputs.
2.2.1 Wiring
All wiring of electrical outputs should use high quality instrument cable; twisted pair low
capacitance shielded instrument cable (20AWG [0.5mm
only high temperature cable where process temperatures exceed 185
drain or screen should be terminated on a DC common or a specifically assigned shield
terminal at the readout instrument end only; in order to protect the signal from mutual
inductive interference. The cable shield at the meter end of the cable must be isolated
with tape or similar, do not connect the cable shield to ground at the meter.
The cable should not be run in a common conduit, or parallel with, power cables or high
inductive load carrying cables; as interference will affect the transmitted pulse signal. Run
all instrument cables in their own separate conduit. Where instrument cables must cross
high power cables be sure that the cables intersect at 90 degrees in order to limit induced
interference.
Do not combine any inductive loads on the same voltage supply as your flowmeter wiring,
as these components are commonly sources of high frequency interference that may affect
the quality of the output signals. Inductive loads on a common voltage source also have the
potential for voltage spikes well in excess of the 24V (dc) limit of the flowmeter electronics.
The maximum wire cross section that can be connected to the terminals of an Oval Gear
pulse meter is 16AWG (1.5mm
2
).
2
] minimum) is recommended. Use
o
F (85oC). The cable
2.2.2 Hall Effect Outputs
The Hall Effect is a solid state 3 wire device which provides an open collector, NPN signal.
The output of the Hall Effect must be fitted with a pull-up resistor between the signal
output ( ) and the voltage supply. The Hall Effect output provides a square wave pulse
signal, which alternates between ground potential and the DC voltage available at the pullup resistor.
The NPN Hall Effect output is a reliable output type, producing a consistent output
irrespective of supply voltage variations, temperature variations, or mechanical shock. The
service life of the Hall Effect output is theoretically infinite, so long as it is protected from
high energy voltage spikes. Hall Effect outputs are protected against reverse polarity, and
against low energy voltage spikes; however, they are not protected against constant overvoltage above the maximum limit of 24V (dc) (±5%).
11
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.