Flomec F116-P Owner's Manual

Page 1
FL_F116P_v1702_01_EN
Signal input flowmeter: pulse, Namur and coil
Output: (0)4-20mA / 0-10V ref. flow rate, pulse ref. total and negative
flow
F-Series - Field mounted indicators for safe and hazardous areas.
Page 2
FL_F116P_v1702_01_EN
Page 3
FL_F116P_v1702_01_EN
SAFETY INSTRUCTIONS
Any responsibility is lapsed if the instructions and procedures as described in this manual
are not followed.
LIFE SUPPORT APPLICATIONS: The F116-P is not designed for use in life support
appliances, devices, or systems where malfunction of the product can reasonably be expected to result in a personal injury. Customers using or selling these products for use in such applications do so at their own risk and agree to fully indemnify the manufacturer and supplier for any damages resulting from such improper use or sale.
Electro static discharge does inflict irreparable damage to electronics! Before installing or
opening the F116-P, the installer has to discharge himself by touching a well-grounded object.
The F116-P must be installed in accordance with the EMC guidelines (Electro Magnetic
Compatibility).
Do connect a proper grounding to the metal enclosure as indicated if the F116-P has an
incoming power line which carries a 115-230V AC. The Protective Earth (PE) wire may never be disconnected or removed.
Intrinsically safe applications: follow the instructions as mentioned in Chapter 0 and
consult “Fluidwell F1..-..-XI - Documentation for Intrinsic safety
DISPOSAL OF ELECTRONIC WASTE
The WEEE Directive requires the recycling of disposed electrical and electronic
equipment in the European Union. When the WEEE Directive does not apply to your region, we support its policy and ask you to be aware on how to dispose of this product.
The crossed out wheelie bin symbol as illustrated and found on our products tells that this
product shall not be disposed of into the general waste system or into a landfill.
At the end of its life, equipment shall be disposed of according to the local regulations
regarding waste of the electrical and the electronic equipment.
Please contact your local dealer, national distributor or the manufacturer’s Technical
helpdesk for information on the product disposal.
SAFETY RULES AND PRECAUTIONARY MEASURES
The manufacturer accepts no responsibility whatsoever if the following safety rules and
precautions instructions and the procedures as described in this manual are not followed.
Modifications of the F116-P implemented without preceding written consent from the
manufacturer, will result in the immediate termination of product liability and warranty period.
Mounting, electrical installation, start-up and maintenance of this device may only be
carried out by trained persons authorized by the operator of the facility. Persons must read and understand this manual before carrying out its instructions.
This device may only be operated by persons who are authorized and trained by the
operator of the facility. All instructions in this manual are to be observed.
Check the mains voltage and information on the manufacturer's plate before installing the
unit.
Check all connections, settings and technical specifications of the various peripheral
devices with the F116-P supplied.
Open the enclosure only if all leads are free of potential. Never touch the electronic components (ESD sensitivity). Never expose the system to heavier conditions than allowed according the classification
of the enclosure (see manufacture's plate and chapter 4).
If the operator detects errors or dangers, or disagrees with the safety precautions taken,
then inform the owner or principal responsible.
The local labor and safety laws and regulations must be adhered to.
Page 4
FL_F116P_v1702_01_EN
ABOUT THE MANUAL
This manual is divided into two main sections:
The daily use of the F116-P is described in chapter 2 Operational. These instructions are
meant for users.
The following chapters and appendices are exclusively meant for electricians/technicians. These
provide a detailed description of all software settings and hardware installation guidance.
This manual describes the standard unit as well as the available options. For additional information, please contact your supplier.
A hazardous situation may occur if the F116-P is not used for the purpose it was designed for or is used incorrectly. Please carefully note the information in this manual indicated by the pictograms:
A "warning !" indicates actions or procedures which, if not performed correctly, may lead to personal injury, a safety hazard or damage of the F116-P or connected instruments.
A "caution !" indicates actions or procedures which, if not performed correctly, may lead to personal injury or incorrect functioning of the F116-P or connected instruments.
A "note !" indicates actions or procedures which, if not performed correctly, may indirectly affect operation or may lead to an instrument response which is not planned.
Hardware version : 03.01.xx Software version : 03.01.xx Manual : FL_F116-P_v1702_01_EN © Copyright 2017 : Fluidwell B.V. - The Netherlands.
Information in this manual is subject to change without prior notice. The manufacturer is not responsible for mistakes in this material or for incidental damage caused as a direct or indirect result of the delivery, performance or use of this material.
© All rights reserved. No parts of this publication may be reproduced or used in any form or by any means without written permission of your supplier.
Page 5
FL_F116P_v1702_01_EN
CONTENTS MANUAL
SAFETY INSTRUCTIONS .................................................................................................................... 3
DISPOSAL OF ELECTRONIC WASTE ............................................................................................... 3
SAFETY RULES AND PRECAUTIONARY MEASURES .................................................................... 3
ABOUT THE MANUAL ......................................................................................................................... 4
CONTENTS MANUAL .......................................................................................................................... 5
1 INTRODUCTION ...................................................................................................................... 6
1.1 System description ...................................................................................................................6
2 OPERATIONAL ....................................................................................................................... 7
2.1 Control panel ............................................................................................................................7
2.2 Operator information and functions ..........................................................................................8
3 CONFIGURATION ................................................................................................................... 9
3.1 How to program the F116-P .....................................................................................................9
3.1.1 Setup menu – Settings .......................................................................................................... 11
3.1.2 Explanation of SETUP-menu 1 - Total-A............................................................................... 12
3.1.3 Explanation of SETUP-menu 2 - Flow rate-A ........................................................................ 13
3.1.4 Explanation of SETUP-menu 3 - Total-B............................................................................... 14
3.1.5 Explanation of SETUP-menu 4 - Flow rate-B ........................................................................ 14
3.1.6 Explanation of SETUP-menu 5 - Display .............................................................................. 14
3.1.7 Explanation of SETUP-menu 6 - Power management .......................................................... 15
3.1.8 Explanation of SETUP-menu 7 - Flowmeter ......................................................................... 16
3.1.9 Explanation of SETUP-menu 8 - Analog output .................................................................... 16
3.1.10 Explanation of SETUP-menu 9 - Pulse ................................................................................. 18
3.1.11 Explanation of SETUP-menu A - Communication (option) ................................................... 18
3.1.12 Explanation of SETUP-menu B - Others ............................................................................... 18
4 INSTALLATION ..................................................................................................................... 19
4.1 General directions ................................................................................................................. 19
4.2 Installation / surrounding conditions ...................................................................................... 19
4.3 Dimensions- Enclosure.......................................................................................................... 20
4.4 Installing the hardware .......................................................................................................... 22
4.4.1. General installation guidelines .............................................................................................. 22
4.4.2. Aluminum enclosure - Field mounted .................................................................................... 23
4.4.3. Aluminum enclosure - Panel mounted .................................................................................. 23
4.4.4. Plastic (GRP) enclosure ........................................................................................................ 24
4.4.5. Terminal connectors .............................................................................................................. 25
5 INTRINSICALLY SAFE APPLICATIONS.............................................................................. 32
5.1 General information and safety instructions .......................................................................... 32
5.2 Terminal connectors Intrinsically safe applications ............................................................... 34
5.3 Configuration examples Intrinsically safe applications .......................................................... 36
6 MAINTENANCE ..................................................................................................................... 38
6.1 General directions ................................................................................................................. 38
6.2 Instructions for repair ............................................................................................................. 38
6.3 Battery replacement .............................................................................................................. 39
6.3.1 Safety instructions ................................................................................................................. 39
6.3.2 Replace the battery (hazardous area) ................................................................................... 39
6.3.3 Disposal of batteries .............................................................................................................. 39
TECHNICAL SPECIFICATION ................................................................................. 40
PROBLEM SOLVING ............................................................................................... 43
COMMUNICATION ................................................................................................... 44
DECLARATION OF CONFORMITY ......................................................................... 48
INDEX OF THIS MANUAL.................................................................................................................. 49
LIST OF FIGURES IN THIS MANUAL ............................................................................................... 49
Page 6
FL_F116P_v1702_01_EN
1 INTRODUCTION
1.1 SYSTEM DESCRIPTION Functions and features
The flow rate / totalizer model F116-P is a microprocessor driven instrument designed to show the flow rate, the total and the accumulated total as well as for the calculation of differential flow measurement and to sum two separate flows. This product has been designed with a focus on:
two multi-purpose pulse inputs; ultra-low power consumption to allow long-life battery powered applications (type PB/PC), Intrinsic safety for use in hazardous applications (type XI); several mounting possibilities with aluminum or GRP enclosures for harsh industrial
surroundings;
ability to process all types of flowmeter signals; transmitting possibilities with analog / pulse and communication outputs.
Flowmeter and temperature input
This manual describes the unit with a pulse input from the flowmeter. Other versions are available to process (0)4-20mA signals. Two flowmeters with a passive or active pulse, Namur or sine wave (coil) signal output can be connected to the F116-P. To power the sensor, several options are available.
Standard output
Pulse output to transmit a pulse that represents a totalized quantity as programmed.  Linear (0)4-20mA or 0-10V analog output to represent the actual calculated differential flow rate
as programmed. The (0)4-20mA or 0-10V signal limits can be tuned.
Fig. 1: Typical application
Page 7
FL_F116P_v1702_01_EN
Configuration
The F116-P is designed for use in many types of applications. For that reason, a setup menu is available to program the F116-P according to your specific requirements.
The setup includes several important features, such as K-Factors, engineering units, signal selection, power management (to extend battery life-time), etc. All settings are stored in a non­volatile memory and therefore kept in the event of a power failure or an exhausted battery.
Display information
The unit has a LCD with (optional) backlight to show the process information, status and alarm messages. The display refresh rate is programmed in the setup menu.
At a key press, the display refresh rate will switch to FAST for 30 seconds. When 'OFF' is selected, the display goes off after 30 seconds after the last key press. The display temporarily comes on after a key press.
A backup of the total and accumulated total in EEPROM memory is made every minute.
Backlight
A backlight is available as an option. The brightness can be tuned as desired (requires power supply type PD/PF/PM). For battery and loop powered applications the backlight will not function.
Options
The following options are available: isolated or active (0)4-20mA / 0-10V analog output, full Modbus communication RS232/485/TTL (also battery powered), Intrinsic safety, mechanical relay or active output, power- and sensor-supply options, panel-mount, wall-mount and weather-proof enclosures, flame proof enclosure and LED backlight.
2 OPERATIONAL
This device may only be operated by persons who are authorized and trained by the
operator of the facility. All instructions in this manual are to be observed.
Take careful notice of the "Safety rules, instructions and precautionary measures" in the
front of this manual.
This chapter describes the daily use of the F116-P. This instruction is meant for users / operators.
2.1 CONTROL PANEL
The control panel has three keys. The available keys are:
Fig. 2: Control panel
Functions of the keys
This key is used to program and save new values or settings. The PROG/ENTER key is also used to gain access to the setup menu (read
chapter 3).
This key is used to select the differential or sum accumulated total, flow rate A and B and accumulated total A and B
The SELECT/ key is also used to increase a value after the PROG/ENTER key has been pressed (read chapter 3)..
This key is used to reset the total. The CLEAR/ key is also used to select a digit or an option after the PROG/ENTER
key has been pressed (read chapter 3).
Page 8
FL_F116P_v1702_01_EN
2.2 OPERATOR INFORMATION AND FUNCTIONS
In general, the F116-P operates in the operator mode. The shown information depends on the settings which are made in the setup menu.
The signal from the connected sensor is processed by the F116-P in the background, independent from the selected display refresh rate.
Fig. 3: Process information (typical)
For the Operator, the following functions are available:
Display (differential or summed) total/rate
Total/rate is the main display of the F116-P. After the selection of any other information, it will always return to this main display automatically. Total is shown on the upper line of the display and flow rate on the bottom line. When selected in the setup menu, the display shows the flow rate only or all. When the SELECT/ key is pressed, the other information shows momentarily.
When "-------" is shown, then the flow rate value is too high to be shown. The arrows indicate the increase/decrease of the flow rate trend. If the consumption is very low, it might be that a stable low flow rate and total is shown; this is due to the settings of the F116-P.
Clear total
The value for total can be reset. To do so, press the CLEAR/ key twice. When the key is pressed once, the text "PUSH CLEAR" is shown. To avoid a reset at this stage, press another key other than the CLEAR/ key or wait for 20 seconds. A reset of the total does not influence the accumulated total.
Display accumulated total
When the SELECT/ key is pressed, total and accumulated total are shown. The accumulated total cannot be reset. The value will count up to 99,999,999,999. The unit and number of decimals are shown according to the settings for the total.
Display Flow rate-A, Flow rate-B, (accumulated) total-A, (accumulated) total-B
The setting All shows the (differential or summed) flow rate. When the SELECT/ key is pressed again, the total/accumulated total, the flow rate A, the accumulated total A, the flow rate B and the accumulated total B are shown temporarily.
Low-battery alarm
Only use original batteries. Original batteries can be ordered at the manufacturer. The use of unapproved batteries will void the warranty.
At the end of the battery’s life-time, the voltage starts to drop. When the voltage becomes too low, the battery indicator comes on. When the battery indicator is on, install a
new and fresh battery as soon as possible.
Fig. 4: Low-battery alarm (typical)
Alarm
When the alarm indicator is shown, refer to Appendix B: Problem Solving.
RUN
Page 9
FL_F116P_v1702_01_EN
3 CONFIGURATION
This and the following chapters are exclusively meant for electricians and non-operators. In these, an extensive description of all software settings and hardware connections are provided.
Mounting, electrical installation, start-up and maintenance of this device may only be
carried out by trained persons authorized by the operator of the facility. Persons must read and understand this manual before carrying out its instructions.
This device may only be operated by persons who are authorized and trained by the
operator of the facility. All instructions in this manual are to be observed.
Make sure, the measuring system is correctly wired up according to the wiring diagrams.
Protection against accidental contact is no longer assured when the housing cover is removed or the panel cabinet has been opened (danger from electrical shock). The housing may only be opened by trained persons authorized by the operator of the facility.
Take careful notice of the "Safety rules, instructions and precautionary measures" in the
front of this manual.
The SETUP menu is used to program the F116-P The SETUP menu is accessible at all times while the F116-P remains fully operational. Be aware
that in this case any change to the settings may have an influence on the operation.
It is possible to prevent access to the SETUP menu with a password. A password may be required to enter the SETUP menu. Without this password, access to SETUP is denied.
3.1 HOW TO PROGRAM THE F116-P
The setup menu has different submenus. Each submenu has an unique number which is shown in front of the menu name.
Each setting has an unique two­digit number which is shown in front of the setting. The first digit refers to the submenu and the second digit refers to the setting.
Note that sometimes the name of the setting is shown on the upper line of the display.
How to enter the setup menu
When the setup menu is protected by a password, the F116-P asks for a password to access the setup menu. When in the operator mode, press and hold the PROG/ENTER key for 7 seconds to access the setup menu.
How to navigate in the setup menu
The setup menu has different submenus to program the F116-P. For navigation, the submenus and the settings are identified with numbers (for the submenu: e.g. 1; for the setting: e.g. 12.).
The CLEAR/► key and the PROG/ENTER key are used for navigation. The explanation
assumes that you are in the submenu TOTAL.
Action
Result
Remark
1
Press the CLEAR/► key to
select the next submenu.
The submenu FLOW
RATE shows
-
2
Press again to go to the next submenu.
The submenu DISPLAY
shows.
-
3
Momentarily, press the PROG/ENTER key to select the previous submenu.
The submenu FLOW
RATE shows
The PROG/ENTER key is used as a key. 4
Press again to go to the previous submenu.
The submenu TOTAL
shows
The PROG/ENTER key is used as a key.
Page 10
FL_F116P_v1702_01_EN
The SELECT/▲ key and the CLEAR/► key are used for navigation. The explanation assumes that you are in the submenu TOTAL. When you are: in the first setting and you navigate to the previous setting, the F116-P goes back to the related
main menu.
in the last setting and you navigate to the next setting, the F116-P goes to the related main
menu.
Action
Result
Remark
1
Press the SELECT/▲ key
to select the first setting.
The setting UNIT shows.
-
2
Press the SELECT/▲ key
again to go to the next setting.
The setting DECIMALS
shows.
­3
Press the CLEAR/► key to
select the previous setting.
The setting UNIT shows.
-
4
Press the CLEAR/►key
again to go to the previous setting.
The submenu TOTAL
shows
This is normal behavior because the setting UNIT is the first setting of the submenu TOTAL.
How to make a setting
Changes are only saved when you press the PROG/ENTER key.
The explanation assumes that you are in the submenu TOTAL and the setting UNIT. When you do not want to save the change, wait for approximately 20 seconds or press and hold the PROG/ENTER key for approximately 3 seconds.
Action
Result
Remark
1
Momentarily, press the PROG/ENTER key.
The PROG indicator blinks. The engineering unit L shows.
To access the setting.
2
Press the SELECT/▲ key to
select the next engineering unit.
The PROG indicator blinks. The engineering unit m3 shows.
If you wait too long, the program mode goes off and changes are not saved, this is normal behavior.
3
Press the SELECT/▲ key to
select the next engineering unit.
The PROG indicator blinks. The engineering unit US GAL
shows.
-
4
Press the CLEAR/► key to
select the previous engineering unit
The PROG indicator blinks. The engineering unit m3 shows.
-
5
To confirm the changes: Momentarily, press the PROG/ENTER key.
The PROG indicator goes off. The change is saved. The engineering unit m3 shows.
If you do not press the PROG/ENTER key to confirm, your selection is not saved.
To discard the changes: Press and hold the PROG/ENTER key for approximately 3 seconds.
The PROG indicator goes off. The change is discarded. The engineering unit L shows.
-
Page 11
FL_F116P_v1702_01_EN
3.1.1 SETUP MENU – SETTINGS
1
TOTAL-A
11
unit
L; m3; kg; lb; GAL; USGAL; bbl; no unit
12
decimals
0000000; 111111.1; 22222.22; 3333.333
13
K-factor
0.000010 - 9999999
14
decimals K-factor
0 - 6
2
FLOW RATE-A
21
unit
mL; L; m3; mg; g; kg; ton; gal; bbl; lb; cf; rev; - - - - (no unit); scf; nm3; nL; p
22
time
/sec; /min; /hour; /day
23
decimals
0000000; 111111.1; 22222.22; 3333.333
24
K-factor
0.000010 - 9999999
25
decimals K-factor
0 - 6 26
filter
0 - 99
27
period
0.1 - 99.9 seconds
3
TOTAL-B
31
K-factor:
0.000010 - 9999999
32
decimals K-factor
0 - 6
4
FLOW RATE-B
41
K-factor
0.000010 - 9999999
42
decimals K-factor
0 - 6
5
DISPLAY
51
function
total; rate; all
52
light
0% (off); 20%; 40%; 60%;- 80%; 100% (full brightness)
53
calculate
add; differ
54
measurement
bi-direct; not negative; threshold; stationary
55
stationary flow rate
0000.000 - 9999999
56
stationary total
0000.000 - 9999.999
6
POWER MANAGEMENT
61
LCD new
fast; 1 sec; 3 sec; 15 sec; 30 sec; off
62
battery mode
operational; shelf
7
FLOWMETER
71
signal A
npn; npn-lp; reed; reed-lp; pnp; pnp-lp; namur; coil-hi; coil-lo; 8-1 DC; 12 DC; 24 DC
72
signal B
npn; npn-lp; reed; reed-lp; pnp; pnp-lp; namur; coil-hi; coil-lo; 8-1 DC; 12 DC; 24 DC
8
ANALOG
81
output
disable; enable
82
rate-min
000.000 - 999999
83
rate-max
000.000 - 999999
84
cut-off
0.0 - 9.9%
85
tune-min
0 - 9999
86
tune-max
0 - 9999
87
filter
01 - 99
9
PULSE
91
mode
signed; not negative; separated
92
width
0.001 – 9
93
decimals
0000000; 111111.1; 22222.22; 3333.333
94
amount
0.001 – 9999999
Page 12
FL_F116P_v1702_01_EN
A
COMMUNICATION
C1
speed
1200; 2400; 4800; 9600
C2
address
1 - 247
C3
mode
bus-rtu; bus-asc; off
D
OTHERS
D1
model
F116-P
D2
software version
nn:nn:nn
D3
serial no.
nnnnnnn
D4
password
0000 - 9999
D5
tag-nr
0000000 - 9999999
3.1.2 EXPLANATION OF SETUP-MENU 1 - TOTAL-A
1
TOTAL-A
11
unit
This setting is used to select the engineering unit for the indication of the total (A and B), the accumulated total (A and B), and the pulse output.
When you change the engineering unit, you must recalculate and reprogram the K-factor for the (accumulated) total. When you recalculate and reprogram the K-Factor, the history for (accumulated) total is not correct anymore, because the (accumulated) total is not recalculated. For future reference, best practice is to make a note of the accumulated total before you program the recalculated K-Factor.
12
decimals
This setting is used to set the amount of digits behind the decimal point for the (accumulated) total indication (A and B).
13
K-factor
This setting is used to set the K-Factor for the total (A). With the K-Factor, the flowmeter pulse signals are converted to a quantity. The K-Factor is based on the number of pulses generated by the flowmeter per selected engineering unit, for example per m3. A more accurate K-Factor (more decimals, as set in decimals K-Factor) allows for a more accurate operation of the system.
Example 1: Calculating the K-Factor.
The flowmeter generates 2.4813 pulses per liter and the selected unit is m3. A cubic meter consists of 1000 liter which gives 2.4813 pulses*1000 liter=2481.3 pulses per m3. So, the K-Factor is 2481.3. Enter for the Flowmeter K-Factor: 24813 and for the flowmeter K-Factor decimals: 1.
Example 2: Calculating the K-Factor.
The flowmeter generates 6.5231 pulses per gallon and the selected engineering unit is gallons. So, the K-Factor is 6.5231. Enter for the Flowmeter K-Factor: 65231 and for the Flowmeter K-Factor decimals: 4.
When you recalculate and reprogram a new K-Factor, the history for (accumulated) total is not correct anymore, because the (accumulated) total is not recalculated. For future reference, best practice is to make a note of the accumulated total before you program the recalculated K-Factor.
14
decimals K-factor
This setting is used to set the amount of digits behind the decimal point for the K-Factor (A).
Page 13
FL_F116P_v1702_01_EN
3.1.3 EXPLANATION OF SETUP-MENU 2 - FLOW RATE-A
The settings for total and flow rate are entirely separate. In this way, different engineering units can be used for each e.g. cubic meters for total and liters for flow rate.
2
FLOW RATE-A
21
unit
This setting is used to select the engineering unit for the indication of the flow rate (A and B).
Alteration of the engineering unit will have consequences for operator and setup values, they will not be automatically recalculated to the value of the new selected unit. The K-Factor has to be adapted as well; the calculation is not done automatically.
22
time
This setting is used to set the time unit for the flow rate calculation (A and B). Note that the flow rate is given in engineering unit/time unit, e.g. liters/minute (l/min).
When you change this setting, also recalculate and change the settings for the analog rate-min and analog rate-max.
23
decimals
This setting is used to set the amount of digits behind the decimal point for the flow rate indication (A and B).
24
K-factor
This setting is used to set the K-Factor for the flow rate (A). With the K-Factor, the flowmeter pulse signals are converted to a quantity. The K-Factor is based on the number of pulses generated by the flowmeter per selected engineering unit, for example per m3. A more accurate K-Factor (more decimals, as set in decimals K-Factor) allows for a more accurate operation of the system.
25
decimals K-factor
This setting is used to set the amount of digits behind the decimal point for the K-Factor (A).
26
filter
This setting is used to stabilize the output signal. With the help of this digital filter a more stable but less actual representation of the flow rate can be obtained.
The filter principal is based on three input values: the filter level (01-99), the last calculated flow rate and the last average value. The higher the filter level, the longer the response time on a value change will be.
27
period
This setting is used to calculate the flow rate by counting the number of pulses within a certain time, for example 1 second. The longer the time the more accurate the flow rate will be.
This setting does influence the update time for the analog output directly. If the
output response is too slow, decrease the number of pulses.
The shorter the update time, the higher the power consumption of the unit will
be (important for battery powered applications).
Page 14
FL_F116P_v1702_01_EN
3.1.4 EXPLANATION OF SETUP-MENU 3 - TOTAL-B
The engineering units are the same as used in SETUP-menu 1 - Total-A.
3
TOTAL-B
31
K-factor
This setting is used to set the K-Factor for the total (B). With the K-Factor, the flowmeter pulse signals are converted to a quantity. The K-Factor is based on the number of pulses generated by the flowmeter per selected engineering unit, for example per m3. A more accurate K-Factor (more decimals, as set in decimals K-Factor) allows for a more accurate operation of the system.
32
decimals K-factor
This setting is used to set the amount of digits behind the decimal point for the (accumulated) total indication (B).
3.1.5 EXPLANATION OF SETUP-MENU 4 - FLOW RATE-B
The engineering units are the same as used in SETUP-menu 2 - Flow rate-A.
4
FLOW RATE-B
41
K-factor
This setting is used to set the K-Factor for the flow rate (B). With the K-Factor, the flowmeter pulse signals are converted to a quantity. The K-Factor is based on the number of pulses generated by the flowmeter per selected engineering unit, for example per m3. A more accurate K-Factor (more decimals, as set in decimals K-Factor) allows for a more accurate operation of the system.
42
decimals K-factor
This setting is used to set the amount of digits behind the decimal point for the K-Factor (B).
3.1.6 EXPLANATION OF SETUP-MENU 5 - DISPLAY
5
DISPLAY
51
function
This setting can be set to display total or rate.
When 'total' is selected, simultaneously, (differential or summed) total and
(differential or summed) flow rate is shown. When SELECT is pressed, the (differential or summed) accumulated total is shown temporarily.
When 'rate' is selected, only (differential or summed) flow rate is shown
together with its engineering unit. When SELECT is pressed, the (differential or summed) total and the (differential or summed) accumulated total are shown temporarily.
When ‘All’ is selected, only (differential or summed) flow rate is shown together
with its engineering unit. When SELECT is pressed again, the (differential or summed) total/accumulated total, the flow rate A, the accumulated total A, the flow rate B and the accumulated total B are shown temporarily.
52
light
The backlight brightness can be adjusted from 0% (off) to 100% (full brightness) in steps of 20%. When the F116-P is only loop powered, the backlight is disabled. An external power supply is required to supply the backlight.
53
measurement
To solve undesired display readings during low or even negative consumption situations, four different measurement methods have been implemented. Note that the selection does influence the analog output value (ref. flow rate) as well.
bi-directional
Shown flow rate: positive and negative. Shown total: increases or decreases.
not negative
Shown flow rate: only positive or zero. Shown total: increases or decreases.
Page 15
FL_F116P_v1702_01_EN
threshold
Shown flow rate: as soon as the flow rate is lower than SETUP 54 or negative, zero flow rate is shown. Shown total: as soon as the flow rate is lower than SETUP 54 or negative, totalization will stop.
stationary
Shown flow rate: as soon as the flow rate is lower than SETUP 54 or negative, the stationary flow rate (SETUP 54) is shown. Shown total: as soon as the flow rate is lower as SETUP 54 or negative, stationary totalization (SETUP 55) will be activated. However, if the value of setting 54 is zero, totalization increases or decreases.
54
stationary flow rate
Enter here the flow rate according SETUP 53: threshold or stationary. The time and measuring units are according to flow rate SETUP 21 and 22.
threshold
Flow rate zero is shown as soon as the flow rate will be lower as this setting.
stationary
As soon as the flow rate is lower as this setting, this flow rate is shown.
If the flowmeters do not generate pulses, the flow rate shows zero.
55
stationary total
Enter here a flow rate per hour according to SETUP 53 – ‘stationary’. The measuring unit is according to TOTAL (A and B) - SETUP 11. This flow rate is converted to a total which will be used as long as the flow rate is lower as SETUP 54.
If the flowmeters do not generate pulses, the totalization will stop. This function is disabled if value zero has been entered.
3.1.7 EXPLANATION OF SETUP-MENU 6 - POWER MANAGEMENT
When used with the internal battery option (type PB/PC), the user can expect reliable measurement over a long period of time. The F116-P has several smart power management functions to extend the battery life time significantly. Two of these functions can be set.
6
POWER MANAGEMENT
61
lcd new
The calculation of the display-information influences the power consumption significantly. When the application does not require a fast display refresh rate, it is strongly advised to select a slow refresh rate. Please understand that NO information will be lost; every pulse will be counted and the output signals will be generated in the normal way. At a key press, the display refresh rate will switch to FAST for 30 seconds. When 'OFF' is selected, the display goes off after 30 seconds after the last key press. The display temporarily comes on after a key press.
Example battery life-time with a coil pick-up:
1kHz pulse and FAST update: about 2 years; 1kHz pulse and 1 sec update: about 5 years.
62
battery mode
The F116-P has two modes: operational or shelf. After "shelf" has been selected, the F116-P can be stored for several years; it will not process the sensor signal; the display is switched off but all settings and totals are stored. In this mode, power consumption is extremely low. To wake up the F116-P again, press the SELECT/ key two times.
Page 16
FL_F116P_v1702_01_EN
3.1.8 EXPLANATION OF SETUP-MENU 7 - FLOWMETER
The F116-P is able to handle several types of input signal. The pickup / signal is selected with:
SETUP 71 (Input A), Read also chapter 4 SETUP 72 (Input B), Read also chapter 4.
The selections "active pulse" offer a detection level of 50% of the supply voltage.
7
FLOWMETER
71/72
SIGNAL
TYPE OF SIGNAL
EXPLANATION
RESISTANCE
FREQ. / mV
REMARK
NPN
NPN input
100 kΩ pull-up
max. 6 kHz.
(open collector)
NPN-LP
NPN with low pass filter
100 kΩ pull-up
max. 1.2 kHz.
(open collector)
less sensitive
REED
Reed-switch input
1 MΩ pull-up
max. 600 Hz.
REED-LP
Reed-with low pass filter
1 MΩ pull-up
max. 120 Hz.
Less sensitive
PNP
PNP input
100 kΩ pull-down
max. 6 kHz.
PNP-LP
PNP with low pass filter
100 kΩ pull-down
max. 1.2 kHz.
Less sensitive
NAMUR
NAMUR input
820 Ω pull-down
max. 4 kHz.
External power
required
COIL-HI
High sensitive coil input
-
min. 20 mVpp
Sensitive for interference!
COIL-HI (option ZF)
min. 10 mVpp
COIL-HI (option ZG)
min. 5 mVpp
COIL-LO
Low sensitive coil input
-
min. 80 mVpp
Normal sensitivity
8-1 DC
Active pulse input
detection level 8.2V DC
3.9
max. 10 kHz.
External power
required
12 DC
Active pulse input
detection level 12V DC
4
max. 10 kHz.
External power
required
24 DC
Active pulse input
detection level 24V DC
3
max. 10 kHz.
External power
required
3.1.9 EXPLANATION OF SETUP-MENU 8 - ANALOG OUTPUT
A linear 4-20mA signal (option AB: 0-20mA or option AU: 0-10V) output signal is generated that represents the flow rate. The settings for the flow rate influence the analog output directly. The relationship between the flow rate and the analog output is set with the following settings.
8
ANALOG OUTPUT
81
output
If the analog output is not used, select disable to minimize the power consumption (e.g. save battery life-time).
Option AP: When a power supply is available but the output is disabled, a 3.5mA signal will be generated.
82
rate-min
Enter here the flow rate at which the output should generate the minimum signal (0)4mA or 0V - in most applications at zero flow. The number of decimals shown depend upon setup 23. The engineering units/time (e.g. L/min) are dependent upon setup 21 and 22.
83
rate-max
Enter here the flow rate at which the output should generate the maximum signal (20mA or 10V) - in most applications at maximum flow. The number of decimals shown depend upon setup 23. The engineering units/time (e.g. L/min) are dependent upon setup 21 and 22.
Loading...
+ 36 hidden pages