Flomag 3000 Installation And Operation Manual

FLOMAG s.r.o.
V Aleji 180/20a CZ-620 00 Brno Czech Republic tel: +420 541212539 fax: +420 549240356 e-mail: info@flomag.com
www.flomag.com
Magnetic Flowmeter
FLOMAG
®
3000
Installation and Operation
FLOMAG 3000 -
Installation and Operation Manual
2
FLOMAG 3000 -
Installation and Operation Manual
3
An magnetic flowmeter is used for volume flow measurement of electrically conductive liquids. Measurement principle is based on Faraday law on electromag­netic induction. A sensor con­sists of a non-magnetic tube with non-conductive lining, measuring electrodes and two coils generating electromag­netic field. Flowing liquid forms a conductor. Magnetic field in­duces voltage U in this conduc­tor that is proportional to mag­netic induction B, distance be­tween electrodes d and flow velocity v.
U = B x d x v
As magnetic induction and dis­tance between electrodes are constant, induced voltage is proportional to velocity of liquid flow in the tube. Volume flow rate is product of flow velocity and tube cross section.
Q = v x S
Fig.1 - Principle of measurement
Principle of measurement
Technical solution
The magnetic flowmeter itself consists of two basic parts – a flow sensor and a converter. The converter can be either an integral part of the sensor (compact version) or separated, connected with the sensor us­ing a cable (remote version). The sensor consists of a non­magnetic tube with non­conductive lining, measuring electrodes, excitation coils and cables. There are various sen­sor versions available enabling connection to adjacent tubes with flanges (type P) and fittings (gas fitting type G or food indus­try fitting type V) or wafer which are installed between flanges using clamps (type B). Non­conductive lining can be made of technical rubber (types TG, MG or NG) or Teflon (type T). The converter is used for gener­ating excitation current in coils, processing of signal from meas­uring electrodes, displaying of measured data and generating output signals. Current in exci­tation coils has constant value 250 mA or 125 mA and is pulse
generated with alternating po­larity to avoid permanent mag­netization of the sensor. Excita­tion pulse frequency can be chosen from six values – 25 Hz, 12,5 Hz, 8,33 Hz, 6,25 Hz, 3,125 Hz and 1,56 Hz. Excita­tion current of 250 mA with ex­citation frequency 3,125 Hz is suitable for all standard applica­tions. Other settings can be used for specific applications. Excitation current and fre­quency are factory set before sensor calibration and their later modifications are not allowed.
Voltage induced in measuring electrodes is measured always on the end of excitation pulse when magnetic field is steady. Each excitation pulse is fol­lowed by refreshing period. Sig­nal processing and parameter setting are performed digitally and the converter contains no setting controls or other moving parts what ensures its high reli­ability and long-term stability.
T
t
I
+I
-I
m
m
1/8T 3/8T
1/8T
0
Fig. 2 - Excitation pulse form
FLOMAG 3000 -
Installation and Operation Manual
4
min 5 x DN min 3 x DN
The flowmeter will give the best results when flow of liquid is steady. Therefore a few basic recommendations should be observed for its locating in a pipeline. There should be no transitions between the sensor and the adjacent pipeline that could be a source of turbulence. Correct axial alignment should be observed during installation. A gasket should not exceed internal edges of tubes.
If more interfering elements are present near the sensor (e.g. bends, fittings), required steady length should be multiplied by number of these interfering ele­ments. Reductions with slopes up to 8° can be included in steady lengths.
x
x
Minimum straight steady lengths of pipeline are required on both sides of the flow sen­sor. Their lengths have to be proportional to pipeline internal diameter.
Fig. 3 - Overlaps
Fig.4 - Steady lengths
8
°
m
a
x
m
a
x
4
5
°
min 2 x DN
Fig.5 - Reduction
If water in the pipeline is pumped by a water pump, the sensor should be always lo­cated behind the pump to avoid low pressure that can damage the sensor. Steady length of at
least 25DN is required between the pump and the sensor.
Fig. 6 - A water pump
For the same reason, never locate stop valves behind the sensor.
Fig. 7 - Stop valves
The sensor can work both in horizontal and vertical positions; only axis of measuring elec­trodes inside the sensor must always remain in horizontal po­sition and tapping of the sensor should be directed upwards at horizontal installations.
Fig. 8 - Electrode axis
For vertical installations, liquid should flow upwards.
Fig. 9 - Vertical installation
To ensure correct measurement and to avoid air lock, whole sensor cross section should be flooded. Therefore never locate the sensor in upper parts of the pipeline or in vertical positions with liquid flowing downwards.
Fig.10 - Danger of air lock
If permanent flooding of whole pipeline cross section cannot be ensured, it is possible to locate the sensor in a low water trap so that it can be always com­pletely flooded. Free water dis­charge should be located 2DN higher than the sensor.
Fig.11 - Permanent flooding
To avoid vibrations that could damage the sensor, ensure that the adjacent pipeline is always supported as near to the sensor as possible.
Fig.12 - Danger of vibrations
Where continuous flow of fluid is required and removal of the sensor is impossible, a bypass should be installed. The same applies for locations where sen­sor removal would require draining of too long part of the pipeline.
Fig. 13 - A bypass
Installation instructions
FLOMAG 3000 -
Installation and Operation Manual
5
Correct function of the magnetic flowmeter requires perfect elec­trical connection between the sensor and the adjacent pipe­line, grounding potential and the power supply protective wire. For the flanged sensor with the adjacent conducting pipeline, flanges should be electrically connected and the pipeline grounded.
If the adjacent pipeline is non-conductive, grounding rings should be inserted in it or equivalent method should be
Fig.16: Grounding rings
Fig.17. The wafer sensor
Fig.18: Cathodic protection
Fig.15: Grounding of flanges
connection of sensor clamping flanges with grounding point of the sensor.
If electric current flows through the pipeline, e.g. for pipeline cathodic protection against cor­rosion, the sensor should be electrically isolated from the adjacent pipeline. The sensor should be bridged over using a wire and galvanic isolation of the flowmeter power supply should be provided so that the flowmeter can be isolated from all other devices.
Sensor grounding
Selection of suitable sensor lining and electrode material
Linings
Sensors have a non-conductive lining from various materials. Choice of material depends on measured fluid characteristics.
· Technical rubber
Technical rubber is suitable for low aggressive fluids with op­erational temperatures from 0.1 °C to 70 °C. It fits for most wa­ter management and sewage treatment applications. It is manufactured in two variants
“TG” – with hard structure and “MG” – with soft structure. Soft
structure is used for fluids with higher content of abrasive parti­cles (e.g. sand). It is not suit­able for drinking water.
· Resistant rubber
Type “NG” is suitable for me­dium aggressive fluids with op­erational temperatures from 0.1 °C to 90 °C. It can be used for measurement of hot service water, condensate etc., as well as for drinking water. If tem-
perature 100 °C can be ex­ceeded, Teflon (PTFE) lining is recommended.
· Teflone or Hallar
Type “T” is the most universal lining for aggressive fluids with operational temperatures from ­20 °C to 150 °C. It is suitable for chemical and food industry ap­plications.
Electrodes
Choice of material of measuring electrodes also depends on measured fluid characteristics.
· Stainless steel – “Ss”
Standard electrodes are made of stainless steel AISI 316Ti. They are suitable for all usual water based fluids and for lower concentrations of acids and caustics.
· Hastelloy C-22 – “Ha”
For some special applications, material of higher quality should be used. Hastelloy C-276 elec-
trodes are characterized by in­creased resistance against ac­ids and caustics and usually are suitable for most of industrial applications.
· Titanium - „Ti“
Suitable for some acids, lyes, chlorine, urea and sew­age.
· Platinum – “Pt”
For particularly aggressive flu­ids like concentrated acids and caustics, chemically extremely resistant material should be chosen – platinum. However, high cost of this material is its essential drawback.
* Note – We can recommend suitable lining and electrode materials for your particular ap­plication.
used to connect measured fluid electrical potential with ground.
For the wafer sensor, grounding can be provided by electrical
FLOMAG 3000 -
Installation and Operation Manual
6
Converter is capable to detect flow rates as low as 0.1 m/s. Upper limit is determined by capability of liquid to maintain continuous flow at higher veloci­ties. This is usually true for flow rates up to 12 m/s.
Measurement error rapidly increases for too low flow rates, as can be seen in the diagram. It shows limits of maximum rela­tive measurement error as func­tion of liquid flow rate.
On the other side, too high flow rate causes discontinuity of flow and results in chaotic tur­bulence and vacuum traps. This results in instable measurement and too high drift of flow rate values.
Ideal operational range of the sensor is in range from 0.5 to 5 m/s. This range is highlighted in the diagram for correct size se­lection.
Flow rate ranges for individual sizes are chosen to meet EN
14154 standard and they are shown in table. Preferred ranges are highlighted in bold. For non-specified working me­ters, other range can be also specified on request.
If range is not specified in a purchase order, the sensor will be calibrated in preferred range in accordance with the table above.
Correct sensor size selection
10
1
0.1
Flow velocity [ m/s ]
Volume flow rate
10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10
-2
-1
0
2
3
DN 10
D
N 20
DN 25
D
N 32DN
4
0
D
N 50DN 65DN 80
D
N
100
DN 125
D
N
1
50
D
N 15
D
N
20
0
D
N
25
0
DN 300
D
N 350DN
400
DN 500
D
N
600
D
N
700
D
N 800DN
900
D
N 1000
DN 1200
2 3 4 5 10
10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10
-1
3
4
2 34 5
3
2
Tab. Sensor ranges in m3/h according to their sizes
Fig. Limit of maximum relative error of measurement
Diagram for correct sensor size selection.
DN
S10 A25 B25 C25 C50 D25 D50 D100
Range Q3/Q1
R10 R25 R25 R25 R50 R25 R50 R100
10
1 0.63
1
1.6 1.6 2.5 2.5 2.5
15
2.5 1.6
2.5
4 4 6.3 6.3 6.3
20
4 2.5
4
6.3 6.3 10 10 10
25
6.3 4
6.3
10 10 16 16 16
32
10 6.3
10
16 16 25 25 25
40
16 10
16
25 25 40 40 40
50
25 16
25
40 40 63 63 63
65
40 25
40
63 63 100 100 100
80
63 40
63
100 100 160 160 160
100
100 63
100
160 160 250 250 250
125
160 100
160
250 250 400 400 400
150
250 160
250
400 400 630 630 630
200
400 250
400
630 630 1000 1000 1000
250
630 400
630
1000 1000 1600 1600 1600
300
1000 630
1000
1600 1600 2500 2500 2500
350
1000 630
1000
1600 1600 2500 2500 2500
400
1600 1000
1600
2500 2500 4000 4000 4000
450
1600 1000
1600
2500 2500 4000 4000 4000
500
2500 1600
2500
4000 4000 6300 6300 6300
600
2500
4000
6300 6300 10000 10000 10000
700
2500
4000
6300 6300 10000 10000 10000
800
4000
6300
10000 10000
900
4000
6300
10000 10000
1000 6300
10000
1200 6300
10000
Range marking
0.25 0.5
0.5
1
1.5
2
2.5
3
0.75 1
±F
10 11 12
[ % ]
v [ m/s ]
9
DN400..1200 DN10..25, 250..350 DN32..200
FLOMAG 3000 -
Installation and Operation Manual
7
Volba výstelky snímače Block diagram of the flowmeter
Main advantage of the mag­netic flowmeter FLOMAG3000 is its significant variability. Flow­meter converter in basic version consists only of power supply, microcomputer and sensor in­put module (module 1). Display, outputs and other optional fea­tures are available as plug-in modules. Thus, customer pays only for features that he really uses. Plug-in modules contain memories where all configura­tion data is stored. In this way, optional features can be added or modified as required anytime during the service life of the flowmeter.
There are 4 free positions available (module 4, 5, 6 and 7) for binary and analog output modules. Their signals are usu­ally processed by connected technological devices. All output modules have galvanic isola­tion. At the same time, up to 4 binary output modules can be fitted. These can operate either as pulse or frequency outputs for flow rate indication. Alterna-
Power supply
module 4 B1 Binary output
module V1 Display
module 1 S1 Sensor input
module 2 F1 Electrode cleaning
module 5 B1 Binary output
module 6 C1 RS 232 interface
module 7 A3 Analog. output
Microcontroller
module 3 M1 Data logger
Sensor
85 - 265 VAC (24V, 12V, AC/DC)
Fig. 19: Block diagram of the flowmeter
tively, they can serve for indica­tion of flowmeter limit condi­tions. Galvanic isolation is en­sured by an optoelement or a relay. One position (module 7) is dedicated for the active ana­log output module. Modules with various accuracy and ranges are available. One posi­tion (module 6) is designed for
the serial communication mod­ule. RS 232, RS 485 or M­Bus interface can be plugged in.
Position (module 2) is for the electrochemical electrodes cleaning module.
A
Module 1
B
Sensor connection
C
Connected
D
internally
E
for compact version
1
module 2 F2 - F3
2 3
Not connected
4 5 6
module 4
7
A4, B1-B5, E1
8
module 5
9
A4, B1-B5, E1
10
module 6
11
A4, B1-B5, C1, D1, D2, E1
12
module 7
13
A1 - A5, B1-B5, E1
14 15
Not connected
16 17
L
18
N
Power supply
19
PE
Terminal connections
T0,5A 250V
ABCDE
12345678910111213141516PELN
B1B5C1A1
+
-
S1 V1
--
Magnetic flowmeter www.flomag.com
FLOMAG3000
Display
Sensor connection
Relay ou
tpu
t
250VAC/
1
A
RS 232
0(4)..20mAAnalog o
utput
85 - 2
40
V
50 - 6
0 H
z
Power
T0,5A / 250V
Tab. Converter terminals Fig. 20 Converter – location of terminals
The converter is integrated in a rugged aluminium box. After opening the box you will gain access to terminals. Terminals 17, 18 and 19 are for power
supply. Terminals A, B, C, D and E are used for the sensor. For compact version, the sen­sor is connected internally and terminals remain free. Termi-
nals 1 to 16 are used for con­nection of inputs and outputs of optional modules (binary out­puts, current output, RS232, RS485 etc.)
FLOMAG 3000 -
Installation and Operation Manual
8
For the remote sensor, there is a terminal box in its tapping (see figure 21). The sensor should be connected to the con­verter using a double shielded
cable. You can use our special sensor cable PAAR-LiYCY-CY
[1X(2X0,25 LiYCY)+1X(2X0,75 LiYCY)+1X0,75]CY (length up
to 200 m) or standard double
Maximum length of the cable between evaluation unit and the sensor is significantly limited by conductivity of measured fluid, as shown in Figure 23. Remote version should be used when measured fluid is too hot to avoid heat transfer to con­verter. See Figure 24 for as­sessment of remote version utilization. Parallel running of power and
0
25
50
100
200
0255075
100125150175200
Maximum length of cable
[ m ]
Fluid conductivity
[ S/cm ]
m
shielded cable Lapp UNI­TRONIC Cy PiDy 2x2x0.25 or Alpha 1243/2C (length up to 50
m).
Fig. 21: Sensor terminal box Fig. 22: Remote sensor connection
signal wires is highly inappropri­ate; especially in case of the cable that connects the sensor with the remote converter. If the instrument is used in environ­ment with strong electromag­netic interference, cables should be rather as short as possible. For connection of electronic converter input and output ter­minals, shielded cables are suit-
able. For connection of mains volt­age, a standard three-core ca­ble, e.g. CYKY 3x1,5 (wire) or VM03VQ-F 3x1 (wire strand) is recommended. The instrument has no switch so it should be fused and switched using other device.
Fig. 23: Maximum length of the cable and conduc-
tivity
Fig. 24: Selection of version according to temperature
A B C D E
Yellow/Green
Yellow
Red
White
Brown
Green
AB
CDE12345678910111213141516PEL
N
B1B1C1A3
+
-
S1 V1
Magnetic flowmwter www.flomag.com
FLOMAG3000
Display
Sensor input
RS 232
4..20mA
Analog. output
85 - 265 V
48 - 63 Hz
T0,5A / 250
V
AABBCCDDEE
PE
N L
M1F
1
17181
9
AABBCCDDEE
Sensor
Sensor
ABCDEABCDE
Cable
PAAR-LiYCY-CY [1X(2X0,25 LiYCY) +1X(2X0,75 LiYCY) +1X0,75]CY
Cable
Lapp UNITRONIC Cy PiDy 2x2x0.25 or Alpha 1243/2C
-20
-10
0
10
20
30
40
50
60
-60-40-20
0204080
100
120
140
160
Fluid temperature [ °C ]
Ambient temperature [ °C
]
TG, MG
NG
T
Remote version
Heating
Remote or compact version
FLOMAG 3000 -
Installation and Operation Manual
9
Displayed data
Fig. 25: Displayed data
The instrument is equipped with a high quality backlit two-line alphanumerical display with character height 9.6 mm (2x16 characters) providing good readability even from longer distances. Backlight function works in energy saving mode. Backlight time is limited to 254 seconds after last pressing of any key. If backlight is off, pressing of any key will switch it on again. Backlight time can be set in menu from 20 seconds to 254 seconds. Setting to 0 switches backlight permanently off; setting to 255 switches it permanently on.
Up to 8 basic readings can be read from the converter display. You can alternate them using
1 key. Additional information accessible via 2 key is avail-
able for some displayed data.
Flow rate
- Flow rate value treated by floating averaging. Number of averaging steps can be changed in range from 1 to 256. Flow rate units can be changed as required.
Number of displayed decimal places can be set in range from 0 to 4.
Total volume (+)
- Total volume of liquid flowed in direction of arrow on the sen­sor from start of measurement.
Total volume (-)
- Total volume of liquid flowed in opposite direction of arrow on the sensor from start of meas­urement.
Volume difference
- Difference between positive and negative volumes flowed from start of measurement.
Operation time
- Total time of operation from initial switching instrument on in hours and minutes.
Percent. flow rate
- Flow rate information indicated by horizontal bar (its width cor­responds to flow rate) and as numeric value in per cents of chosen maximum value.
Last error
- Abbreviated text of the last error message.
Current flow rate
- Flow rate value untreated by floating averaging.
Temporary volume +
- User resettable value of vol­ume flowed in direction of arrow on the sensor.
Temporary volume -
- User resettable value of vol­ume flowed in opposite direc­tion of arrow on the sensor.
Temporary difference
- User resettable value of differ­ence between volumes flowed
in direction and in opposite di­rection of arrow on the sensor.
Temporary time
- User resettable value of vol­ume flowed in direction of arrow on the sensor. Values of temporary counters
can be reset by holding 3 key and simultaneous pressing 4
key. This will reset all counters at the same time – both vol­umes and time.
Batching
- Shows information about the running batch. Detailed informa­tion is given in chapter Batch­ing.
Loading...
+ 19 hidden pages