Flomag 3000 Installation And Operation Manual

FLOMAG s.r.o.
V Aleji 180/20a CZ-620 00 Brno Czech Republic tel: +420 541212539 fax: +420 549240356 e-mail: info@flomag.com
www.flomag.com
Magnetic Flowmeter
FLOMAG
®
3000
Installation and Operation
FLOMAG 3000 -
Installation and Operation Manual
2
FLOMAG 3000 -
Installation and Operation Manual
3
An magnetic flowmeter is used for volume flow measurement of electrically conductive liquids. Measurement principle is based on Faraday law on electromag­netic induction. A sensor con­sists of a non-magnetic tube with non-conductive lining, measuring electrodes and two coils generating electromag­netic field. Flowing liquid forms a conductor. Magnetic field in­duces voltage U in this conduc­tor that is proportional to mag­netic induction B, distance be­tween electrodes d and flow velocity v.
U = B x d x v
As magnetic induction and dis­tance between electrodes are constant, induced voltage is proportional to velocity of liquid flow in the tube. Volume flow rate is product of flow velocity and tube cross section.
Q = v x S
Fig.1 - Principle of measurement
Principle of measurement
Technical solution
The magnetic flowmeter itself consists of two basic parts – a flow sensor and a converter. The converter can be either an integral part of the sensor (compact version) or separated, connected with the sensor us­ing a cable (remote version). The sensor consists of a non­magnetic tube with non­conductive lining, measuring electrodes, excitation coils and cables. There are various sen­sor versions available enabling connection to adjacent tubes with flanges (type P) and fittings (gas fitting type G or food indus­try fitting type V) or wafer which are installed between flanges using clamps (type B). Non­conductive lining can be made of technical rubber (types TG, MG or NG) or Teflon (type T). The converter is used for gener­ating excitation current in coils, processing of signal from meas­uring electrodes, displaying of measured data and generating output signals. Current in exci­tation coils has constant value 250 mA or 125 mA and is pulse
generated with alternating po­larity to avoid permanent mag­netization of the sensor. Excita­tion pulse frequency can be chosen from six values – 25 Hz, 12,5 Hz, 8,33 Hz, 6,25 Hz, 3,125 Hz and 1,56 Hz. Excita­tion current of 250 mA with ex­citation frequency 3,125 Hz is suitable for all standard applica­tions. Other settings can be used for specific applications. Excitation current and fre­quency are factory set before sensor calibration and their later modifications are not allowed.
Voltage induced in measuring electrodes is measured always on the end of excitation pulse when magnetic field is steady. Each excitation pulse is fol­lowed by refreshing period. Sig­nal processing and parameter setting are performed digitally and the converter contains no setting controls or other moving parts what ensures its high reli­ability and long-term stability.
T
t
I
+I
-I
m
m
1/8T 3/8T
1/8T
0
Fig. 2 - Excitation pulse form
FLOMAG 3000 -
Installation and Operation Manual
4
min 5 x DN min 3 x DN
The flowmeter will give the best results when flow of liquid is steady. Therefore a few basic recommendations should be observed for its locating in a pipeline. There should be no transitions between the sensor and the adjacent pipeline that could be a source of turbulence. Correct axial alignment should be observed during installation. A gasket should not exceed internal edges of tubes.
If more interfering elements are present near the sensor (e.g. bends, fittings), required steady length should be multiplied by number of these interfering ele­ments. Reductions with slopes up to 8° can be included in steady lengths.
x
x
Minimum straight steady lengths of pipeline are required on both sides of the flow sen­sor. Their lengths have to be proportional to pipeline internal diameter.
Fig. 3 - Overlaps
Fig.4 - Steady lengths
8
°
m
a
x
m
a
x
4
5
°
min 2 x DN
Fig.5 - Reduction
If water in the pipeline is pumped by a water pump, the sensor should be always lo­cated behind the pump to avoid low pressure that can damage the sensor. Steady length of at
least 25DN is required between the pump and the sensor.
Fig. 6 - A water pump
For the same reason, never locate stop valves behind the sensor.
Fig. 7 - Stop valves
The sensor can work both in horizontal and vertical positions; only axis of measuring elec­trodes inside the sensor must always remain in horizontal po­sition and tapping of the sensor should be directed upwards at horizontal installations.
Fig. 8 - Electrode axis
For vertical installations, liquid should flow upwards.
Fig. 9 - Vertical installation
To ensure correct measurement and to avoid air lock, whole sensor cross section should be flooded. Therefore never locate the sensor in upper parts of the pipeline or in vertical positions with liquid flowing downwards.
Fig.10 - Danger of air lock
If permanent flooding of whole pipeline cross section cannot be ensured, it is possible to locate the sensor in a low water trap so that it can be always com­pletely flooded. Free water dis­charge should be located 2DN higher than the sensor.
Fig.11 - Permanent flooding
To avoid vibrations that could damage the sensor, ensure that the adjacent pipeline is always supported as near to the sensor as possible.
Fig.12 - Danger of vibrations
Where continuous flow of fluid is required and removal of the sensor is impossible, a bypass should be installed. The same applies for locations where sen­sor removal would require draining of too long part of the pipeline.
Fig. 13 - A bypass
Installation instructions
FLOMAG 3000 -
Installation and Operation Manual
5
Correct function of the magnetic flowmeter requires perfect elec­trical connection between the sensor and the adjacent pipe­line, grounding potential and the power supply protective wire. For the flanged sensor with the adjacent conducting pipeline, flanges should be electrically connected and the pipeline grounded.
If the adjacent pipeline is non-conductive, grounding rings should be inserted in it or equivalent method should be
Fig.16: Grounding rings
Fig.17. The wafer sensor
Fig.18: Cathodic protection
Fig.15: Grounding of flanges
connection of sensor clamping flanges with grounding point of the sensor.
If electric current flows through the pipeline, e.g. for pipeline cathodic protection against cor­rosion, the sensor should be electrically isolated from the adjacent pipeline. The sensor should be bridged over using a wire and galvanic isolation of the flowmeter power supply should be provided so that the flowmeter can be isolated from all other devices.
Sensor grounding
Selection of suitable sensor lining and electrode material
Linings
Sensors have a non-conductive lining from various materials. Choice of material depends on measured fluid characteristics.
· Technical rubber
Technical rubber is suitable for low aggressive fluids with op­erational temperatures from 0.1 °C to 70 °C. It fits for most wa­ter management and sewage treatment applications. It is manufactured in two variants
“TG” – with hard structure and “MG” – with soft structure. Soft
structure is used for fluids with higher content of abrasive parti­cles (e.g. sand). It is not suit­able for drinking water.
· Resistant rubber
Type “NG” is suitable for me­dium aggressive fluids with op­erational temperatures from 0.1 °C to 90 °C. It can be used for measurement of hot service water, condensate etc., as well as for drinking water. If tem-
perature 100 °C can be ex­ceeded, Teflon (PTFE) lining is recommended.
· Teflone or Hallar
Type “T” is the most universal lining for aggressive fluids with operational temperatures from ­20 °C to 150 °C. It is suitable for chemical and food industry ap­plications.
Electrodes
Choice of material of measuring electrodes also depends on measured fluid characteristics.
· Stainless steel – “Ss”
Standard electrodes are made of stainless steel AISI 316Ti. They are suitable for all usual water based fluids and for lower concentrations of acids and caustics.
· Hastelloy C-22 – “Ha”
For some special applications, material of higher quality should be used. Hastelloy C-276 elec-
trodes are characterized by in­creased resistance against ac­ids and caustics and usually are suitable for most of industrial applications.
· Titanium - „Ti“
Suitable for some acids, lyes, chlorine, urea and sew­age.
· Platinum – “Pt”
For particularly aggressive flu­ids like concentrated acids and caustics, chemically extremely resistant material should be chosen – platinum. However, high cost of this material is its essential drawback.
* Note – We can recommend suitable lining and electrode materials for your particular ap­plication.
used to connect measured fluid electrical potential with ground.
For the wafer sensor, grounding can be provided by electrical
FLOMAG 3000 -
Installation and Operation Manual
6
Converter is capable to detect flow rates as low as 0.1 m/s. Upper limit is determined by capability of liquid to maintain continuous flow at higher veloci­ties. This is usually true for flow rates up to 12 m/s.
Measurement error rapidly increases for too low flow rates, as can be seen in the diagram. It shows limits of maximum rela­tive measurement error as func­tion of liquid flow rate.
On the other side, too high flow rate causes discontinuity of flow and results in chaotic tur­bulence and vacuum traps. This results in instable measurement and too high drift of flow rate values.
Ideal operational range of the sensor is in range from 0.5 to 5 m/s. This range is highlighted in the diagram for correct size se­lection.
Flow rate ranges for individual sizes are chosen to meet EN
14154 standard and they are shown in table. Preferred ranges are highlighted in bold. For non-specified working me­ters, other range can be also specified on request.
If range is not specified in a purchase order, the sensor will be calibrated in preferred range in accordance with the table above.
Correct sensor size selection
10
1
0.1
Flow velocity [ m/s ]
Volume flow rate
10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10
-2
-1
0
2
3
DN 10
D
N 20
DN 25
D
N 32DN
4
0
D
N 50DN 65DN 80
D
N
100
DN 125
D
N
1
50
D
N 15
D
N
20
0
D
N
25
0
DN 300
D
N 350DN
400
DN 500
D
N
600
D
N
700
D
N 800DN
900
D
N 1000
DN 1200
2 3 4 5 10
10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10 2 3 4 5 10
-1
3
4
2 34 5
3
2
Tab. Sensor ranges in m3/h according to their sizes
Fig. Limit of maximum relative error of measurement
Diagram for correct sensor size selection.
DN
S10 A25 B25 C25 C50 D25 D50 D100
Range Q3/Q1
R10 R25 R25 R25 R50 R25 R50 R100
10
1 0.63
1
1.6 1.6 2.5 2.5 2.5
15
2.5 1.6
2.5
4 4 6.3 6.3 6.3
20
4 2.5
4
6.3 6.3 10 10 10
25
6.3 4
6.3
10 10 16 16 16
32
10 6.3
10
16 16 25 25 25
40
16 10
16
25 25 40 40 40
50
25 16
25
40 40 63 63 63
65
40 25
40
63 63 100 100 100
80
63 40
63
100 100 160 160 160
100
100 63
100
160 160 250 250 250
125
160 100
160
250 250 400 400 400
150
250 160
250
400 400 630 630 630
200
400 250
400
630 630 1000 1000 1000
250
630 400
630
1000 1000 1600 1600 1600
300
1000 630
1000
1600 1600 2500 2500 2500
350
1000 630
1000
1600 1600 2500 2500 2500
400
1600 1000
1600
2500 2500 4000 4000 4000
450
1600 1000
1600
2500 2500 4000 4000 4000
500
2500 1600
2500
4000 4000 6300 6300 6300
600
2500
4000
6300 6300 10000 10000 10000
700
2500
4000
6300 6300 10000 10000 10000
800
4000
6300
10000 10000
900
4000
6300
10000 10000
1000 6300
10000
1200 6300
10000
Range marking
0.25 0.5
0.5
1
1.5
2
2.5
3
0.75 1
±F
10 11 12
[ % ]
v [ m/s ]
9
DN400..1200 DN10..25, 250..350 DN32..200
FLOMAG 3000 -
Installation and Operation Manual
7
Volba výstelky snímače Block diagram of the flowmeter
Main advantage of the mag­netic flowmeter FLOMAG3000 is its significant variability. Flow­meter converter in basic version consists only of power supply, microcomputer and sensor in­put module (module 1). Display, outputs and other optional fea­tures are available as plug-in modules. Thus, customer pays only for features that he really uses. Plug-in modules contain memories where all configura­tion data is stored. In this way, optional features can be added or modified as required anytime during the service life of the flowmeter.
There are 4 free positions available (module 4, 5, 6 and 7) for binary and analog output modules. Their signals are usu­ally processed by connected technological devices. All output modules have galvanic isola­tion. At the same time, up to 4 binary output modules can be fitted. These can operate either as pulse or frequency outputs for flow rate indication. Alterna-
Power supply
module 4 B1 Binary output
module V1 Display
module 1 S1 Sensor input
module 2 F1 Electrode cleaning
module 5 B1 Binary output
module 6 C1 RS 232 interface
module 7 A3 Analog. output
Microcontroller
module 3 M1 Data logger
Sensor
85 - 265 VAC (24V, 12V, AC/DC)
Fig. 19: Block diagram of the flowmeter
tively, they can serve for indica­tion of flowmeter limit condi­tions. Galvanic isolation is en­sured by an optoelement or a relay. One position (module 7) is dedicated for the active ana­log output module. Modules with various accuracy and ranges are available. One posi­tion (module 6) is designed for
the serial communication mod­ule. RS 232, RS 485 or M­Bus interface can be plugged in.
Position (module 2) is for the electrochemical electrodes cleaning module.
A
Module 1
B
Sensor connection
C
Connected
D
internally
E
for compact version
1
module 2 F2 - F3
2 3
Not connected
4 5 6
module 4
7
A4, B1-B5, E1
8
module 5
9
A4, B1-B5, E1
10
module 6
11
A4, B1-B5, C1, D1, D2, E1
12
module 7
13
A1 - A5, B1-B5, E1
14 15
Not connected
16 17
L
18
N
Power supply
19
PE
Terminal connections
T0,5A 250V
ABCDE
12345678910111213141516PELN
B1B5C1A1
+
-
S1 V1
--
Magnetic flowmeter www.flomag.com
FLOMAG3000
Display
Sensor connection
Relay ou
tpu
t
250VAC/
1
A
RS 232
0(4)..20mAAnalog o
utput
85 - 2
40
V
50 - 6
0 H
z
Power
T0,5A / 250V
Tab. Converter terminals Fig. 20 Converter – location of terminals
The converter is integrated in a rugged aluminium box. After opening the box you will gain access to terminals. Terminals 17, 18 and 19 are for power
supply. Terminals A, B, C, D and E are used for the sensor. For compact version, the sen­sor is connected internally and terminals remain free. Termi-
nals 1 to 16 are used for con­nection of inputs and outputs of optional modules (binary out­puts, current output, RS232, RS485 etc.)
FLOMAG 3000 -
Installation and Operation Manual
8
For the remote sensor, there is a terminal box in its tapping (see figure 21). The sensor should be connected to the con­verter using a double shielded
cable. You can use our special sensor cable PAAR-LiYCY-CY
[1X(2X0,25 LiYCY)+1X(2X0,75 LiYCY)+1X0,75]CY (length up
to 200 m) or standard double
Maximum length of the cable between evaluation unit and the sensor is significantly limited by conductivity of measured fluid, as shown in Figure 23. Remote version should be used when measured fluid is too hot to avoid heat transfer to con­verter. See Figure 24 for as­sessment of remote version utilization. Parallel running of power and
0
25
50
100
200
0255075
100125150175200
Maximum length of cable
[ m ]
Fluid conductivity
[ S/cm ]
m
shielded cable Lapp UNI­TRONIC Cy PiDy 2x2x0.25 or Alpha 1243/2C (length up to 50
m).
Fig. 21: Sensor terminal box Fig. 22: Remote sensor connection
signal wires is highly inappropri­ate; especially in case of the cable that connects the sensor with the remote converter. If the instrument is used in environ­ment with strong electromag­netic interference, cables should be rather as short as possible. For connection of electronic converter input and output ter­minals, shielded cables are suit-
able. For connection of mains volt­age, a standard three-core ca­ble, e.g. CYKY 3x1,5 (wire) or VM03VQ-F 3x1 (wire strand) is recommended. The instrument has no switch so it should be fused and switched using other device.
Fig. 23: Maximum length of the cable and conduc-
tivity
Fig. 24: Selection of version according to temperature
A B C D E
Yellow/Green
Yellow
Red
White
Brown
Green
AB
CDE12345678910111213141516PEL
N
B1B1C1A3
+
-
S1 V1
Magnetic flowmwter www.flomag.com
FLOMAG3000
Display
Sensor input
RS 232
4..20mA
Analog. output
85 - 265 V
48 - 63 Hz
T0,5A / 250
V
AABBCCDDEE
PE
N L
M1F
1
17181
9
AABBCCDDEE
Sensor
Sensor
ABCDEABCDE
Cable
PAAR-LiYCY-CY [1X(2X0,25 LiYCY) +1X(2X0,75 LiYCY) +1X0,75]CY
Cable
Lapp UNITRONIC Cy PiDy 2x2x0.25 or Alpha 1243/2C
-20
-10
0
10
20
30
40
50
60
-60-40-20
0204080
100
120
140
160
Fluid temperature [ °C ]
Ambient temperature [ °C
]
TG, MG
NG
T
Remote version
Heating
Remote or compact version
FLOMAG 3000 -
Installation and Operation Manual
9
Displayed data
Fig. 25: Displayed data
The instrument is equipped with a high quality backlit two-line alphanumerical display with character height 9.6 mm (2x16 characters) providing good readability even from longer distances. Backlight function works in energy saving mode. Backlight time is limited to 254 seconds after last pressing of any key. If backlight is off, pressing of any key will switch it on again. Backlight time can be set in menu from 20 seconds to 254 seconds. Setting to 0 switches backlight permanently off; setting to 255 switches it permanently on.
Up to 8 basic readings can be read from the converter display. You can alternate them using
1 key. Additional information accessible via 2 key is avail-
able for some displayed data.
Flow rate
- Flow rate value treated by floating averaging. Number of averaging steps can be changed in range from 1 to 256. Flow rate units can be changed as required.
Number of displayed decimal places can be set in range from 0 to 4.
Total volume (+)
- Total volume of liquid flowed in direction of arrow on the sen­sor from start of measurement.
Total volume (-)
- Total volume of liquid flowed in opposite direction of arrow on the sensor from start of meas­urement.
Volume difference
- Difference between positive and negative volumes flowed from start of measurement.
Operation time
- Total time of operation from initial switching instrument on in hours and minutes.
Percent. flow rate
- Flow rate information indicated by horizontal bar (its width cor­responds to flow rate) and as numeric value in per cents of chosen maximum value.
Last error
- Abbreviated text of the last error message.
Current flow rate
- Flow rate value untreated by floating averaging.
Temporary volume +
- User resettable value of vol­ume flowed in direction of arrow on the sensor.
Temporary volume -
- User resettable value of vol­ume flowed in opposite direc­tion of arrow on the sensor.
Temporary difference
- User resettable value of differ­ence between volumes flowed
in direction and in opposite di­rection of arrow on the sensor.
Temporary time
- User resettable value of vol­ume flowed in direction of arrow on the sensor. Values of temporary counters
can be reset by holding 3 key and simultaneous pressing 4
key. This will reset all counters at the same time – both vol­umes and time.
Batching
- Shows information about the running batch. Detailed informa­tion is given in chapter Batch­ing.
FLOMAG 3000 -
Installation and Operation Manual
10
Archiving
Electromagnetic flowmeter FLOMAG3000 automatically saves in fixed time intervals , the value of the flowed volume. There are three archives.
Hour archive where it is possi­ble to find the flowed volumes about the last 192 hours (8 days).
Daily archive where it is pos­sible to find the flowed volumes about the last 192 days (more than half year).
Month archive where it is pos­sible to find the flowed volumes
about the last 12 months.
Upper line shows always time interval of the item in the ar­chive. Second line shows the flowed volume in the fixed time interval and the power off time in the fixed time interval (the flowmeter was without power supply).
Listing in archive
Pushing key 1, roll to item Total volume+. By repeated
pushing of key 2 you can find gradually Flow rate +, hour ar­chive, day archive and month archive.
By repeated pushing of key 3, in the hour archive you can
see gradually the saved sam-
ples from the previous hour, totally 192 hours back.
By repeated pushing of key
3, in the daily archive you can see gradually the saved sam-
ples from the previous day, to­tally 192 days back.
By repeated pushing of key
3, in the month archive you can see gradually the saved samples from the previous
month, totally 12 months back.
If you want to go back to the
last saved sample, please push and hold key 3 and together with it push key 4.
Fig. Hour archive
Fig. Daily archive
Fig. Month archive
Fig. Legend
Fig. Hour archive moving
Fig. Daily archive moving
Fig. Month archive moving
Ok - without power off
-#s - power off time in secunds
-#m - power off time in minutes
-#h - power off time in hours
-#d - power off time in days
Error messages
Fig. Error messages
In case of a fault, an error message with short de­scription of the fault is shown immediately at the LCD.
The error message begins with character E followed by error's number. If the error is connected with a defect of a module than it is followed also by character M and the number of the module. After pushing but-
ton 1 the flowmeter re­turns to value display mode
and at the same time the er­ror message is saved to last error register. During the indi­cation of the error message, the flowmeter is measuring. In case of error E-7, E-8 and E
-13, the flowmeter indicates 0 if the mentioned error mes­sage is not forbidden in the menu.
Error messages are listed below together with recom­mendation how to repair.
E0: No error
E1: EEPROM Checksum error
Checksum error saved in the module - re-check data in the module and save
again
E2: Stack overflow
For module „B“ in pulse mode - the time constants are too long, the flow rate is higher than it is possible to send pulses, stack overflow of un-sent pulses - change pulse length and space length or volume for 1 pulse
E3: Frequency limit ex-
ceeded
For module „B“ in fre­quency mode, it is required higher output frequency than the module is able to send, flow rate is higher than it was assumed - set higher flow rate value for 1kHz.
E4: Power fail
FLOMAG 3000 -
Installation and Operation Manual
11
Appears for short after
power fail
E5: Old software
For proper operation of the module is required newer firmware version than it is installed in the convertor – upgrade firmware.
E6: Can't use this mode
For module B placed in position 6 and 7 is not possi­ble to use frequence mode (it is possible only in position 4 and 5) – change the posi­tion of the module or change the mode to pulse.
E7: Sensor loop discon­nected
No current to the coils – for remote version check cables and terminals
E8: Empty pipe
For modules F2 and F3, indicates that the controlling electrode is not submerged
E9: Low medium conductiv-
ity
For modules F1 and F3 in electrodes cleaning mode, no current in the electrodes, sensor is not submerged, electrodes are furred or low medium conductivity - clean the sensor
E10: MBus conflict
Module D3 – exist two sta­tions M-Bus with same ad­dress – change the setting of module D3
E11: Current output overrange
For module A it is required higher output current than 20mA, flowrate is higher
than it was assumed – set higher flow rate value for Imax
E12: Serial line fail -
communication error
Communication module C1 or Dx sends data but does not receive confirma­tion for receipt of data – check cables, could be caused also by external in­terference, high capacity of the cables or too long ca­bles.
E13: Sensor signal over-
range
Signal from the sensor is overrange of the convertor ­electrodes are not sub­merged or there is short cir­cuit of the cables – check the sensor and the cables
User outputs – plug-in modules
T0,5A 250V
position 8
position 1
position 2
position 3
position 4
position 5
position 6
position 7
Flowmeter converter in basic configuration contains the power supply and boards re­quired for measurement func­tions. All other inputs, outputs and display units can be added as plug-in modules. Customer pays only for features that he really uses. At the same time,
this concept allows using of various types of inputs and out­puts tailored to customer needs.
Following table and figure indicate positions and functions of individual modules.
Fig. Module positions
Tab. Module positions
Po­sitio n
Modules
Terminals
1
S1 sensor input
module, always plugged in
A, B, C, D, E
2
F1-F3 module for
sensor full pipe checking and elec­trode cleaning
1, 2, 3, 4, 5,
3
M1 extended memory
module of measured data
-
4
A4,A7 passive cur-
rent output 4 - 20 mA B1-B5 binary outputs incl. frequency up to 12 kHz E1 binary input
6, 7
5
A4, A7 passive cur-
rent output 4 - 20 mA B1-B5 binary outputs incl. frequency up to 1,2 kHz E1 binary input
8, 9
6
A4, A7 passive c.o. B1-B5 except for
frequencies,
C1, D1, D2, D3, G1, H1 data communica-
tion E1 binary input
10, 11
7 A1-A3, A5, A6 active
current output
A4, A7 passive c.o. B1-B5 except for
Frequencies E1 binary input
12, 13
8
V1 display and key-
pad
10 pin connec­tor
FLOMAG 3000 -
Installation and Operation Manual
12
Parameter setting
The magnetic flowmeter con­verter can be configured in two ways, as required: either using a PC connected via serial inter­face, or using keys.
Press 4 to switch the display to programming mode. Pro­gramming mode is password protected against unauthorized access. Correct password (4­digit number) must be entered to obtain access to main menu. Password of a new instrument is always set to 0000.
Fig. Enter password
This is also initial value dis­played as default. Simply con­firm it to enter in menu.
Password can be changed as required before you leave the programming mode.
Warning! You can switch the instrument to data display mode
anytime by pressing 1 and check current parameter set­tings. However, the instrument is not password protected against unauthorized access until you enter EXIT command. Programming runs in back­ground and with only a few ex­ceptions has no influence to measurement.
Fig. Cursor movement 2 key moves cursor to the
right. When the utmost right position is reached, the cursor returns to the left.
Fig. Character changing
system returns to the first avail­able character. Character set is always se­lected with regard to possibility of character occurrence in text: [0..9] for integers, [0..9,- , .] for decimals and complete alpha­bet for text variables (including Czech characters).
Confirm your selection by 4
key to finish editing.
A status message will be dis­played. If your password is not accepted, program returns to editing mode. If correct pass­word was entered, you will get to main menu.
Use 3 to move in menu. This
key moves the lower line item to upper line. In all menus, the upper line with blinking first character is always the active line.
Press 4 to enter in submenu
or to edit item. Pressing 2 in submenu brings you always back to previous menu (“Escape” function). If you are in main menu, pressing of this key will offer exit from programming mode.
Menu legend
Some menu items can be used only for viewing and do not allow change of values.
Fig. Movement in menu
Fig. Status message
+ Noise filter
+ Dynamic filter
» Enter password
125 mA
250 mA
¤ Production date
3 key changes character at cursor position. When the last available character is reached,
Fig. Read only
Fig. Enter value
Press 4 to return to previous
menu.
Other menu items can be
used to enter value directly.
When you enter the value and
press 4, a status message will be displayed.
If the value entered is ac­cepted, press any key to return to previous menu or to edit next item.
If the value entered is out of range, an error message will be displayed; press any key to edit the value.
In some cases, one of listed values has to be selected.
Fig. Selection of one value
Use 3 to select required item.
When the required value is in the upper line, press 4 to con-
firm your selection. A status message will be displayed to confirm that your selection has been accepted. Press any key to return to previous menu or to edit next item.
In some cases, more of listed
values can be selected.
Fig. Selection of more items
There is a sign “+” (indicating that the item is selected) or “­“ (indicating that the item is not selected) before each of items.
Fig. Selection of more items
Press 2 to change selection for the item displayed in the
upper line. Press 4 to finish your selection. A status mes­sage will be displayed to con­firm that your selection has been accepted. Press any key to return to previous menu.
FLOMAG 3000 -
Installation and Operation Manual
13
Flowmeter menu
» Enter password
0.Production data Production date ¤ Production date Serial number ¤ Serial number Software ¤ Software Meter's type ¤ Meter's type Modules used ¤ Embeded modules Date setting » Date Time setting » Time Permit upgrade » Upgrade PIN Delete history » Delete PIN Reset volumes » Reset PIN
1.Sensor Sensor constants Constatnt 1 » Sensor constant1
Constant 2 » Sensor constant2
Excitation freq. →○ 2.775 Hz
3.125 Hz 5.55 Hz 6.66 Hz 12.5 Hz 25 Hz
Excitation curr. →○ 125 mA
250 mA
Supressed flow » Do not meas. Q< Samples » Samples Filters →√ Noise filter
Dynamic filter
Zero setting » Autozero PIN
2.Module 2 Specific settings for individual modules
7.Module 7 Specific settings for individual modules
8.Display Language →● [CZ] Česky
[D] Deutsch [GB] English [PL] Polski [I] Italiano [NL] Nederlands [S] Svenska [E] Espanol [F] Francais
100 per cent » 100 per cent Flowrate units →● l/s
m3/h
User's » Flowrate multipl
» Unit's name
Decimal places » Decimal places Backlight » Backlight time Display select →√ Flowrate
Total volume + Total volume + Total difference Operation time Percent flowrate Last error Real time
Error messages →√ E1
E13
9.Exit Exit menu
Exit menu
New password » Access password
FLOMAG 3000 -
Installation and Operation Manual
14
0. Production data
This submenu relates to the flowmeter converter.
·Production date – of con- verter
·Serial number – of con- verter
·Software – current soft- ware version
·Type of meter – type num- ber of flowmeter converter
·Modules used – types of currently used modules
(Above listed items are only for information and user cannot change them)
·Date setting – setting of current date
·Time setting – setting of current time
·Upgrade enabled – a new firmware version can be uploaded after entering PIN
·Delete history – archives will be deleted after entering PIN
·Reset volumes – All total- izers will be reset after en­tering PIN
1. Sensor
This submenu relates to the sensor.
·Sensor constants – sen- sor calibration constants
·Excitation frequency – of sensor coils
·Excitation current – of sensor coils
·Suppressed flow rate – when flow rate is lower than this value, it is considered for zero. This setting is used to suppress creeping flows.
·Number of samples – for floating averaging that filters measured flow rate value. Higher number of samples
provides more stable flow rate value, however it in­creases time constant and
causes delayed reaction to flow rate changes.
·Filters ¨Noise filter partially re-
duces jump changes but mainly removes lower peri­odical interference. Tran­sient edges are rounded as can be seen in figure of re­sponse to unit jump. The filter is applied already on input and thus influences immediate flow rate value and cumulated volume cal­culated from it. The noise filter introduces only negligi-
ble delay (about 0.3 sec) and can be used almost at all circumstances. ¨Dynamic filter reduces rapid jump changes of flow rate. It protects very effec­tively against high short peaks caused by interfer­ence. Unlike averaging, dy­namic filter cuts input signal and interference is not in­cluded in cumulated vol­ume. It can however cause delay of flow rate jump change indication. This fact should be considered if the flowmeter is used for dosing applications. Response to unit jump can be seen in figure.
8. Display
This submenu relates to data shown on the display.
·Language – language of displayed data. You can select from 9 languages.
·100 per cent – 100 % flow rate for bar diagram. It is
Fig. Dynamic filter
Fig. Noise filter
used only for percentage bar diagram display; it is not meant as range of the me­ter.
·Flow rate units – You can select from 12 preset units or add your own user de­fined unit. In such case you have to enter multiple of
flow rate in l/s and a unit name.
·Decimal places – Number of decimal places of dis­played flow rate. You can enter 0-4 decimal places. If 5 places are entered, num­ber of decimal places will be dynamically changed to 4 valid decimal places.
·Time of backlight – Time of display backlight in sec­onds. When you press any key, backlight of display goes on. When time period set in seconds expires since you pressed the last key, the backlight goes off. You can set time period from 1 to 254 seconds. If you set 0, backlight will never be on. If you set 255, the display will be permanently backlit.
·Displayed values – De- fines what items will be dis­played. You can select any of available items. These will be alternately displayed on the flowmeter display.
Press 1 to alternate dis­played values.
·Error messages – Enable or disable displaying of indi­vidual error messages.
9. Exit
·Exit menu – When you finish editing, you have to exit menu because only af­ter that settings are perma­nently stored in module memories. If you will not exit the menu and power failure occurs, previously entered settings will be loaded. Also access to menu is password protected only when you exit menu.
·New password – You can change the access pass­word before you finish your editing.
Time
NtN
Sampling moments
Current flow rate
Flow rate
Average flow rate
Fig. Averaging
FLOMAG 3000 -
Installation and Operation Manual
15
Modules A – analog current output – are used for flow data transmission. There are 4 differ­ent types available with various ranges, accuracy and functions.
Outputs of A1 to A3, A5 and A6 modules are active (forced current) and are galvanically isolated from other flowmeter parts. Outputs can be loaded up to 1000 Ω. They can be plugged only in position 7.
A4 and A7 modules have pas­sive current output (it has to be powered externally) and are also galvanically isolated. Unlike A1-A3, A5, A6 modules, it can be plugged in positions 4, 5 and 6.
0..+Q output
0..-Q output IQI output
-Q..+Q output
Fixed current 0..20
Depending on flow rate (see diagrams), output can work in 5 modes:
The first four modes generate output current dependent on flow rate, the fifth mode enables direct entering of current. Fol­lowing 4 ranges can be se­lected for all modes (except for the fixed current mode):
A3, A4, A6 and A7 modules can work only with range 4..20 mA.
0..20mA output
4..20mA output
0..10mA output
0..5mA output
Range
4..20mA 0..10mA 0..5mA
Flow rate / current
-Q
max
0 Q
max -Qmax
0 Q
max -Qmax
0 Q
max -Qmax
0 Q
max
0..+Q output 0 0 20 4 4 20 0 0 10 0 0 5
0..-Q output 20 0 0 20 4 4 10 0 0 5 0 0
0..IQI output 20 0 20 20 4 20 10 0 10 5 0 5
-Q.+Q output 0 10 20 4 12 20 0 5 10 0 2,5 5
0..20mA
A1
Range 0(4)..20 mA Resolution 12 bit Accuracy ± 0.2%, ± 0.2mA Active - replaced by A5
A2
Range 0(4)..20 mA Resolution 16 bit Accuracy ± 0.1%, ± 0.1mA Active - replaced by A5
A3
Range 4..20 mA Resolution 16 bit Accuracy ± 0.1%, ± 0.1mA Active - replaced by A6
A4
Range 4..20 mA Resolution 16 bit Accuracy ± 0.1%, ± 0.1mA Passive - replaced by A7
A5
Range 0(4)..20 mA Resolution 16 bit Accuracy ± 0.1%, ± 0.1mA
Active
A6
Range 4..20 mA res. 16 bit Accuracy ± 0.1%, ± 0.1mA Active - compatible with module H1 ( HART )
A7
Range 4..20 mA res. 16 bit Accuracy ± 0.1%, ± 0.1mA Passive - compatible with module H1 ( HART )
Current output modules A1 – A4
Fig. 0..+Q output
Fig. 0..+Q output
Fig. IQI output
Fig. -Q..+Q output
Fig. Menu structure of A modules
7.Current output →● 0..+Q output 0..20mA output»Imax flowrate
0..-Q output ● 4..20mA output 0..|Q| output 0..10mA output -Q..+Q output 0..5 mA output
Fixed current»Fixed current [mA]
1213
A5
0..20 mA
Current output
+
-
A
+
-
1011
A7
4..20 mA
Current output
+
-
A
+
+
-
Modules A6 and A7 enables with module H1 (Bell 202 mo­dem) to communicate through current loop with a protocol compatable with HART (only Universal Commands)
FLOMAG 3000 -
Installation and Operation Manual
16
B1
Passive (max 4 kHz) Max. voltage 350 V
p-p
Max. perm. current 120 mA Max. pulse current 300 mA Resistance 27 Ω
B2
Passive (max 12 kHz) Max. voltage 60 V
p-p
Max. perm. current 300 mA Max. pulse current 500 mA Resistance 5 Ω
B3
Active Voltage 5 V Max. current 10 mA Max. frequency 12 kHz
B4
Active Voltage 24 V Max. current 40 mA Max. frequency 12 kHz
B5
Relay contacts Max. voltage 250 VAC Max. current 1 A
Binary output modules B1 - B5
Flowmeter converter can con­trol up to 4 multifunctional bi­nary outputs in positions 4 – 7. Following table indicates differ­ences between individual mod­ules:
Outputs can work as pulse, frequency or status outputs. Individual functions are ex­plained in detail in following section.
Binary module func­tions
· Normally closed/open
These modes are used for ser­vice purposes.
· Pulse outputs (not)
In this mode, a pulse is gener­ated immediately after preset volume has flowed. Pulse gen­eration is determined by three factors: pulse length “tu”, mini-
Tab. Binary output modules
mal delay between two pulses “tD” and volume per pulse “V”.
Flow rate values are time inte­grated. Immediately after preset volume per pulse has flowed, the pulse of length tu is gener­ated. After the pulse, there is a delay of length at least tD. If the delay expires and the preset volume has not flowed again, the output remains inactive; otherwise, another pulse and delay are generated immedi­ately. If the preset volume flows through sooner than the previ­ous pulse is finished, the unsent pulse will be stored in a buffer with maximum capacity of 255 pulses. If buffer overflow oc­curs, an error message will be generated. It follows from above mentioned that parameters of pulse output should be set such that expected pulse frequency cannot exceed limit frequency determined by pulse length and delay.
It applies: Maximum pulse frequency [s-1] = 1 / (tu + tD)
Volume per pulse can be set in range from 1 to 109 ml with 1 ml step, i.e. from 1 ml to 1000 m3. Delay and pulse lengths can be set in range form 10 ms to 2550 ms with 10 ms step. It follows from above mentioned that the maximum pulse frequency is 50 s-1.
Pulses can be generated in three modes depending on flow rate, and pulse polarity can be set (output is closed during pulse or open in not modes).
Fig. Pulse generation
Fig. IQI pulses
Fig. Q+ pulses
Fig. Q- pulses
· Frequency outputs
In this mode of operation, fre­quency is generated on output modules. Pulse to delay ratio is always 1:1. Warning! Only two frequency generators in posi­tions 4 and 5 are available for the converter. This function is blocked in positions 6 and 7. Maximum frequency in position 4 is 12 kHz and in position 5 only 1.2 kHz. If these limit fre­quencies are exceeded, output frequency will be limited and an error message will be gener­ated.
Frequency outputs can work in three modes of frequency de­pendence on flow rate.
Setting is done by selecting of flow rate corresponding to 1 kHz output frequency.
Fig. Q+ frequency
Fig. Q- frequency
Fig. IQI frequency
FLOMAG 3000 -
Installation and Operation Manual
17
Fixed frequency mode is used for service purposes. Required frequency is set directly in Hz in range 1-120000 Hz in position 4 and 1-1200 Hz in position 5.
· Negative/non-negative
flow rate
This mode is used for flow di­rection differentiation. Output is closed/open for negative flow rate.
· Failure occurred/not oc-
curred
If failure with mode set to active (see error messages) occurs, the output closes/opens for at least 5 seconds. If failure per­sists, output is closed/open for whole period of failure duration.
· Limit flow rate exceeded/
undergone (not)
If flow rate is higher/lower than limit value set, the output closes (opens). When flow rate returns to limits, the output opens (closes) again taking in account preset hysteresis. This function
Fig. Q > Qlimit
Fig. Q < Qlimit
Fig. IQI > Qlimit
Fig. IQI < Qlimit
Fig. Menu structure of B modules
Fig. Terminal connection
6
7
B1
Bi nar y o utput
Opt oMOS 35 0V
8
9
B2
Bi nar y out pu t
ac tiv e 5V/ 100mA
+
-
8
9
B5
Bi nar y out pu t
Re lay
4.Binary output ─►● Permanently open
Perm. Closed IQI pulses » Pulse length [10ms] IQI not pulses » Pause length [10ms] Q+ pulses » Vol./ pulse [ml] Q+ not pulses Q- pulses Q- not pulses Q+ frequency » Flowrate at 1 kHz Q- frequency IQI frequency Fixed frequency » Fixed frequency [ Hz ] Negative flow Non-neg. flow 'Error occured E1 'No error occured √ …
E13
Q>Qlim. » Flow limit Q>Qlim. not » Hysteresis Q<Qlim. Q<Qlim. not IQI>Qlim. IQI>Qlim. not IQI<Qlim. IQI<Qlim. not Cleaning Not cleanig Batch Opened » Volume Advance [ml] Batch Opened not » Time Advance [ms]
Batch Stop puls » Pulse length [10ms] Batch Stop /puls » Volume Advance [ml]
» Time Advance [ms]
works in four modes of depend­ence on flow rate with output polarity differentiation.
· Electrode cleaning/no cleaning
The output is closed/open dur­ing cleaning.
· Batch Opened
· Batch /Opened
Output is switched on / switched off during the batch running. It is possible to set the advance of the output before batch ending. Advance could be set by time or volume. Detailed information is given in chapter Batching.
· Batch Stop pulse
· Batch Stop /pulse
Output generates pulse for ending of the batch. It is possi­ble to set pulse length (10ms ­2,5s) and pulse advance before batch ending. Advance could be set by time or volume. Detailed information is given in chapter Batching.
FLOMAG 3000 -
Installation and Operation Manual
18
C1, D1, D2 and D3 modules
Modules C1, D1, D2 and D3 are used for data communica­tion. All these modules are gal­vanically isolated from flow­meter circuits. They can be in­stalled only in position 6 and their signals are output to termi­nals 10 and 11. Interface RS232 is the only exception, as it needs 3 wires.
C1 – RS232 inter­face
It is used primarily for service purposes, be­cause maximum cable length from the converter to a computer is 15 me­ters and only one con­verter can be connected to a single link.
Connection to the com­puter equipped with RS232 in­terface is done using an in­cluded cable. Thread one end of the cable through a cable bushing and connect it to 3-pin connector behind terminals 10-
11. The other end of the cable is equipped with CANON 9M connector. This ensures leading of RS232 interface signals out of converter box while protec­tion level IP66 is maintained.
The interface is connected to the computer using a crosslink cable Laplink 9F-9F.
D1 – RS485 interface
It is used for permanent con­nection of multiple converters to the computer. It enables con-
C1
RS232
D1
RS485 (MODBUS)
D2
0/20mA data current loop
D3
M-BUS
G1
GSM modem
H1
HART modem (with A6,7)
12345
6789
CANON 9F
12345
6789
CANON 9M
1011
C1
RS 232
Fig.RS232 cable connection
necting of up to 31 stations in a communication network using a twisted pair link cable with total length of 1200 meters. Number of stations and length of cables can be increased by using re-
peaters. Stations are connected to the link in parallel. The most distant ends of line have to be equipped with terminating resis­tors 120 Ω.
When setting stations, be sure to set the same baud rate for all of them and to set a unique address for each of them. MASTER address is al­ways set to “0” and addresses of individual stations can be set in range 1-254.
D2 – Data current loop interface 0/20 mA
It is used for permanent con­nection of multiple converters to the computer. It enables con­necting of multiple stations on long distances. Data transfer via current coding 0/20 mA is highly resistant against interfer­ence and is suitable for indus-
Fig. RS485 network connection
Fig.Data current loop 0/20 mA
network connection
trial environment. Individual sta­tions are connected in series. Disadvantage of this configura­tion is that if one station fails, then whole network breaks down.
D3 – M-BUS interface
Standard M-Bus (Meter-Bus) is designed for applications of data acquisition from various media consumption meters. It enables connecting of many devices (hundreds of them) on distance of several kilometers.
Fig. M-Bus network connection
Fig. Communication module menu
M-BUS
DEVICE 1
DEVICE 2
DEVICE
3
M-BUS
M-BUS
M-BUS
MASTER
10
1110111011
6.RS232 ─► Baudrate ─►○ 300 Bd
600 Bd 1200 Bd 2400 Bd 4800 Bd
9600 Bd 19200 Bd 38400 Bd 57600 Bd 76800 Bd 115200 Bd
Adress ─► »Own adress Protocol ─►● FLOMAG
M-BUS MODBUS-RTU MODBUS-ASCII HART
Parity ─►● No parity
Odd parity Even parity
Word order ─►● Low word first
High word first
FLOMAG 3000 -
Installation and Operation Manual
19
Modul E1
Module E1 is galvanically separated active binary input for nonvolatile contacts or open collector. Input has SW tran­sient immunity. It is necessary min pulse length of 60 ms for switching .
Functions of the binary input
· Switch Off
· Batch Start
Starts the batch (see section
Batching).
· Batch Hold
Stops and holds the batch. Counted batch can be contin­ued after restarting it (see sec­tion Batching).
· Batch Stop
Asynchronously finishes the batch and sets again the preset
batch volume (see section Batching).
· Reset Volumes
Enables reset of the chosen
totalizers and working time.
· Cleaning Start
Starts electrochemical elec­trodes cleaning in modules F1 or F3 (requires plug-in modules
Fig Menu of module E1 Fig. Terminal connection
2. Electrode cleaning
─►
OFF Only once During Power ON Periodiccally
─►
»Clean every x [h]
»Time of start [h:m]
F1, F2 and F3 modules
F1 – F3 modules are used for full pipe check and for electro­chemical cleaning of electrodes.
F1-Electrochemical electrode cleaning mod-
ule
During operation of the flow­meter, a non-conductive layer can be formed on sensor elec­trodes. This layer increases contact resistance between electrodes and measured fluid and results in lower measure­ment accuracy.
F1 module enables measur­ing electrode cleaning without need of sensor deinstallation. The method is based on elec­trochemical effect. Alternating voltage is connected to elec­trodes and the deposited layer dissolves in liquid. This cleaning should be performed periodi-
F1
Electrochemical elec­trode cleaning module
F2
Empty pipe detection module
F3
Module with combined functions F1+F2
cally.
The cleaning cycle takes 1 minute. During cleaning, no real measurement is performed. Flow rate measured before start of cleaning is simulated. Clean­ing cycle duration can be indi­cated using binary outputs. Running cleaning process is indicated on display by moving full character on upper line.
The instrument offers several possibilities of cleaning cycle start:
If ONCE is selected, a single cleaning cycle is immediately performed and then the instru­ment returns to Off mode.
If AFTER SWITCH-ON is se­lected, the cleaning cycle will be always started if power supply
is switched on. Option PERI­ODIC starts cleaning in regular intervals that can be set by user in range from 1 to 255 hours. Timer starts counting after time value is entered. Starting time of cleaning can be set to match to real time. It makes sense only if cleaning period is set in multiples of 24 hours. Then cleaning will be performed al­ways in preset time. F1 module has no wires connected to the terminal block.
F2 - Full pipe check
module
Correct measurement of flow rate by the magnetic flowmeter is conditioned by full flooding of the whole sensor cross section by measured liquid. If flooding is only partial, the magnetic flowmeter will indicate flow rate higher than real. However, if both electrodes are not im­mersed in liquid, interfering volt­age can be generated on elec-
OFF Only once During power ON
Periodically
Fig. F1 module menu
4. Binary input ─►● Off
Batch Start Batch Hold Batch Stop Reset Counters
─►
Total volume +
Cleaning Start
Total volume - Operational time Temp. volume + Temp. volume - Temporary time
FLOMAG 3000 -
Installation and Operation Manual
20
trode wires and the flowmeter can indicate totally random val­ues. To avoid this situation, the sensor can be equipped with checking electrode and con­verter with full pipe check mod­ule. The module continuously checks if the checking electrode
is immersed in liquid. If it is not, an error message will be dis­played and zero flow rate will be
indicated. The checking elec­trode is connected to terminal 1. This electrode cannot be retro-
fitted and should be con­sidered on initial order.
Fig. F2 module menu
F3 - Electrode cleaning and full pipe check module
This module combines func­tions of modules F1 and F2. Besides different menu, all
above mentioned information applies for this module.
Fig. F3 module menu
2. Full pipe check
─►
ON
2. Clean/Full pipe
─►
Electrode cleaning
─►
Off Once During Power ON Periodiccally
─►
»Clean every x [h]
»Time of start [h:m]
Full pipe check
─►
ON
Batching
Batching mode is used for control of external equipment ( valves, pumps) in order to measure repeatedly the preset volume (batch). It is not recom­mended for batching of very small volumes. The period of one batch has to be min 30 s.
Electromagnetic flowmeter FLOMAG3000 has very sophis­ticated batching functions. It enables simple batching with manual starting but also fully automatic batching by remote control.
Batching process is divided into 4 phases. Single phases are distinguished on LCD by a symbol in square brackets. Up­per line shows remaining vol­ume in ml to the end of the
batch. Bottom line shows help for manual batch control.
Phase 0 - Stop
During phase 0 outputs are inactive, no batching runs, wait­ing for start of the batching process. Manual start is real-
ized by holding button 3 and simultaneous pushing of button
4 or from outside with the help of module E in Batch Start mode. During that phase the batch volume is set. To enter edit mode it is necessary to
hold button 2 and simultane­ously push button 4.
Phase 1 - Hold
Phase 1 temporary batch in-
terruption. Outputs give signal for interruption of the batch (close valve, switch off pump). Counted batch could ve contin­ued again by manual start choosing Restart or from out­side with the help of module E in Batch Start mode. It is possi­ble also to abort the batch by choosing Reset or from outside with the help of module E in Batch Stop mode.
Phase 2 - Run
During that phase runs the measuring of the batch (valve is open, pump is running). This mode can be interrupted using Phase 1 (Hold) activated manu­ally or with the help of module E in Batch Stop mode. It is possi­ble also to abort it choosing Re-
Fig. Batching phase 0
Fig. Batching phase 1
Fig. Batching phase 2
Module G1
Module H1
Module G1 is a GSM modem enabling sending by SMS mes­sages on preset numbers, in preselect intervals (or on re-
quest) inforamtion about the status of the flowmeter and the measured values. Detailed de­scription of the module func-
tions is given in a separate document.
Module H1 is Bell 202 mo­dem, extensioning the function of modules A6 and A7. It en­ables data communication through current loop with a pro-
tocol compatible with HART (only Universal Command). As­signing the variables of the flowmeter is following:
PV - Flow rate [ l/s ] SV - Current flow rate [ l/s ] TV - Total volume + [ m3/h ] QV - Total volume - [ m3/h ]
FLOMAG 3000 -
Installation and Operation Manual
21
Fig. Batching phase 3
Fig. Batch running
Fig. Minimum batching option
Fig. More complex batching option
event which first takes place. The advantage is that you can set different advance for each output. This way it is possible to switch off a pump and later close the valve.
Binary outputs have 4 options
for setting in batch mode. They
set or from outside with the help of module E in Batch Stop mode. During this phase there is no reaction to external signal Batch Start. If there is no any preset outputs' advance, after flowing of the set batch volume comes switching to phase 0 (Stop). Outputs will send signal for batch stop (switch off pump, close valve). Because of the late reaction to the output signal a partial overflow of the set batch volume will take place and display will show negative volume. That's why it is suitable to send a signal for batch stop in advance as it is described below.
Phase 3 - Finish
In practice we need to send a signal for batch stop in ad­vance. It is mainly because of the inertia of the technical equipment (valves, pumps). Phase 3 represents time when one of the outputs have sent in advance signal for batch stop. According to the ability to fore­cast exactly the advance, the batch can slightly overflow which means to go to phase 0 or the batch will not finish and will go to phase 3. In this case it is possible to go to phase 0 by choosing Reset or from outside with the help of module E in Batch Stop mode. In the same time takes place presetting of the volume batch accroding to the preset value. Second possi­bility is to start new batch choosing restart or from outside with the help of module E in Batch Start mode.
The binary outputs in batch mode enable setting of the ad­vance by volume when the out­put is activated, if the remaining batch volume is smaller than the preset volume advance. Second possibility is to set time advance when the flowmeter's logic calculates what time re­mains to the end of the batch according to the actual flow. Both possibilites can be com­bined. Output reacts to the
time
volume
output 1
output 2
output 2 (puls)
output 1 (puls)
start
hold
volume - advance 2
volume - advance 1
volume of batch
Flowmeter
Electrically controled valve
module B
FLOMAG3000
Pump
Flowmeter
Electrically controled valve
module B module B module E module E
FLOMAG3000
Start Hold
can indicate condition when runs the batch (phase 2) or to sent a pulse for batch finish/ interuption. Pulse length can be set from 10 ms up to 2550 ms, step 10 ms. Both output modes can work in both polarities.
FLOMAG 3000 -
Installation and Operation Manual
22
L
A
DN
o D
o d
l
160
168
80
114
58
65
70
75
100
136
* Standard construction length meets ISO 13359, different construction lengths should be indicated, e.g. “l=215” Construction length tolerance for DN£200: +0/-2 mm DN³200: +0/-3 mm ** Weight of PDIN sensor without converter and terminal box
- weight of terminal box has to be added for remote version – 0.25 kg
- weight of converter has to be added for compact version – 0.9 kg
Flowmeter dimensions – flanged versions “P”, “PDIN” and “PANSI”
DN PN D d A l Weight**
ISO 13359
EN 14154
Optional
[mm] [inches] [mm] [mm] [ kg ]
15 1/2
16 95 62 164 200 138 66 3.5
20 3/4
16 105 62 170 200 138 66 3.5
25 1
16 115 72 180 200 215 96 3.5
32 1 1/4
16 135 82 199 200 215 96 6
40 1 1/2
16 145 92 209 200 215 96 7
50 2
16 160 107 223 200 215 96 8
65 2 1/2
16 180 127 244 200 215 96 10
80 3
16 195 142 260 200 215 96 12
100 4
16 215 162 280 250 215 96 16
125 5
16 245 192 310 250 305 126 21
150 6
16 280 218 340 300 305 126 28
200 8
16 335 274 398 350 380 211 35
250 10
10 405 370 480 450 380 211 42.5
300 12
10 440 420 535 500 515 320 55
350 14
10 500 480 584 550 515 320 65
400 16
10 565 530 642 600 515 320 94
450 18
10 565 530 642 600 515 320 94
500 20
10 670 640 752 600 515 320 122
600 24
10 780 760 870 600 615 320 158
700 28
10 895 880 990 700 715 420 230
800 32
6 1010 980 1100 800 815 420 325
900 36
6 1115 1040 1185 900 815 520 420
1000 40
6 1220 1140 1290 1000 1015 520 510
1200 48
6 1455 1340 1510 1200 1015 520 680
L*
[Bar] [mm] [mm] [mm]
[mm]
Type acc. to Flance DN PN Flange dimensions meet
PDIN
15..1200 2.5, 6, 10, 16, 25, 40, 63 EN 1092-1 BS 4504
PANSI
1/2“..40“ 150lb, 300lb ASA / ANSI B 16.5
Protection: compact version IP66, remote version IP67 (optionally IP68) Electrodes: Ss – stainless steel AISI316Ti, Ha - Hastelloy C22, Ti - Titanium, Pt – platinum Lining: TG – hard rubber, MG – soft rubber, NG – resistant rubber, PTFE – Teflon Accessories: optional for a surcharge – grounding rings or grounding electrodes for nonconductive tube
FLOMAG 3000 -
Installation and Operation Manual
23
L
A
DN
o d
160
168
80
114
58
65
70
75
100
136
DN D A L* Weight**
lining
TG, MG
lining
NG
lining PTFE
[mm] [inches] [mm [mm] [mm] [mm] [mm] [ kg ]
10 3/8
62 145 - - 62 0.8
15 1/2
62 145 74 72 70 0.9
20 3/4
62 145 74 72 70 1.1
25 1
72 158 104 102 100 1.5
32 1 1/4
82 168 104 102 100 1.8
40 1 1/2
92 179 104 102 100 2.2
50 2
107 192 104 102 100 2.8
65 2 1/2
127 212 104 102 100 3.2
80 3
142 227 104 102 100 3.5
100 4
162 247 104 102 100 4
125 5
192 277 134 132 130 6
150 6
218 303 134 132 130 8
Flowmeter dimensions – wafer version “B”
* Standard construction length meets ISO 13359, different construction lengths should be indicated, e.g. “l=215” Construction length tolerance: +0/-2 mm ** Weight of the sensor without converter and terminal box
- weight of terminal box has to be added for remote version – 0.25 kg
- weight of converter has to be added for compact version – 0.9 kg
The sensor is designed for installation between flanges and for fastening by clamps (not included in delivery). For sensor sizes DN20..DN120, flanges with corresponding dimensions are used. For sensor
sizes DN10..D15 the flange DN20 has to be used, because the sensor body is bigger than space between openings of corresponding flanges.
Protection: compact version IP66, remote version IP67 (optionally IP68) Electrodes: Ss – stainless steel AISI316Ti, Ha - Hastelloy C22, Ti - Titanium, Pt – platinum Lining: TG – hard rubber, MG – soft rubber, NG – resistant rubber, PTFE – Teflon Pressure: PN16, PN25, PN40, PN63 Accessories: optional for a surcharge – grounding rings or grounding electrodes for nonconductive
tube
FLOMAG 3000 -
Installation and Operation Manual
24
L
A
DN
o d
l
160
168
80
114
58
65
70
75
100
136
DN d l A* Weight**
ISO 13359
EN 14154
Optional
[mm] [inches] [mm] [mm] [mm] [mm] [mm] [ kg ]
15 1/2
62 66 145 200 134 0.9
20 3/4
62 66 145 200 150 1.1
25 1
72 96 158 200 213 1.5
32 1 1/4
82 96 168 200 213 1.8
40 1 1/2
92 96 179 200 213 2.2
50 2
107 96 192 200 213 2.8
65 2 1/2
127 96 212 200 213 3.2
80 3
142 96 227 200 213 3.5
100 4
162 96 247 250 213 4
125 5
192 126 277 250 301 6
150 6
218 126 303 300 301 8
L*
Flowmeter dimensions – version with aseptic screwed fitting “B” (DIN 11851)
* Standard construction length meets ISO 13359, different construction lengths should be indicated, e.g. “l=213” Construction length tolerance: +0/-2 mm ** Weight of the sensor without converter and terminal box
- weight of terminal box has to be added for remote version – 0.25 kg
- weight of converter has to be added for compact version – 0.9 kg
The sensor is connected to the pipeline using an aseptic screwed fitting that meets DIN 11 851 stan­dard. Part of the fitting with cap nut is firmly fixed to the sensor. Welded counterpart with male thread and sealing are part of delivery. This sensor version is suitable for foodstuff flow. Non-conducting lining of the sensor extends over its edges to avoid both leakage and depositing of measured fluid on edges. Due to cap nuts, the sensor can be easily deinstalled and comfortably cleaned.
Screwed fitting: DIN 11 851 Protection: compact version IP66, remote version IP67 (optionally IP68) Electrodes: Ss – stainless steel AISI316Ti, Ha - Hastelloy C22, Ti - Titanium, Pt – platinum Lining: NG – resistant rubber (for drinking water), PTFE – Teflon (for foodstuff) Pressure: PN16, PN25, PN40, PN63 Accessories: optional for a surcharge – grounding electrodes for nonconductive tube
FLOMAG 3000 -
Installation and Operation Manual
25
L
A
DN
o d
l
160
168
80
114
58
65
70
75
100
136
G
DN G d l A L* Weight**
thread
ISO 13359
EN 14154
Optional
[mm] [inches] [inches] [mm] [mm] [mm] [mm] [mm] [ kg ]
15 1/2
1“ 62 66 145 200 134 0.9
20 3/4
1 1/4“ 62 66 145 200 150 1.1
25 1
1 1/2“ 72 96 158 200 213 1.5
32 1 1/4
2“ 82 96 168 200 213 1.8
40 1 1/2
2 1/2“ 92 96 179 200 213 2.2
50 2
3“ 107 96 192 200 213 2.8
65 2 1/2
3 1/2“ 127 96 212 200 213 3.2
80 3
4“ 142 96 227 200 213 3.5
Flowmeter dimensions – version with tube thread “G” (DIN ISO 228 )
The sensor is connected to the pipeline using a fitting with a cap nut and a gasket. Both ends of meas­uring tube are equipped with male tube thread. Adjacent tubes have to be equipped with fitting with cap nut (not included in delivery) and sealed using gasket (not included in delivery).
Protection: compact version IP66, remote version IP67 (optionally IP68) Electrodes: Ss – stainless steel AISI316Ti, Ha - Hastelloy C22, Ti - Titanium, Pt – platinum Lining: TG – hard rubber, MG – soft rubber, NG – resistant rubber, PTFE – Teflon Pressure: PN16, PN25, PN40, PN63 Accessories: optional for a surcharge – grounding electrodes for nonconductive tube
* Standard construction length meets ISO 13359, different construction lengths should be indicated, e.g. “l=213” Construction length tolerance: +0/-2 mm ** Weight of the sensor without converter and terminal box
- weight of terminal box has to be added for remote version – 0.25 kg
- weight of converter has to be added for compact version – 0.9 kg
FLOMAG 3000 -
Installation and Operation Manual
26
PDIN 50 16 TG Ss Ge Fe Cv
Cv
compact version
Rvx
remote version (x = cable length in m)
_
without full pipe check electrode
Fe
optional full pipe check electrode
_
without grounding electrode
Ge
optional grounding electrode
Ss
stainless steel electrodes
Ha
Hastelloy electrodes
Pt
platinum electrodes
TG
hard rubber lining
MG
soft rubber lining
NG
resistant rubber lining
PTFE
Teflon lining
6, 10, 16, 25, 40
nominal pressure [bars]
150lb, 300lb
nominal pressure [lb]
10..1200
nominal bore diameter [mm]
3/8"..50"
nominal bore diameter [inches]
PDIN
flanged version – flanges according to DIN
PANSI
flanged version – flanges according to ANSI
B
wafer version
V
version with aseptic fittings for food industry
G
threaded version
Ti
Titanium electrodes
Sensor – marking and label
Tab. Sensor marking
Fig. Sensor label
FLOMAG 3000 -
Installation and Operation Manual
27
FLOMAG 3 0 0 0 S1 F1 -- B1 B1 C1 A1 V1
--
module not fitted
V1
display and keypad
--
module not fitted
A1
active output module 0(4)..20 mA (12 bit) - replaced by A5
A2
active output module 0(4)..20 mA (16 bit) - replaced by A5
A3
active output module 4..20 mA (16 bit) - replaced by A6
A4
passive output module 4..20 mA (16 bit) - replaced by A7
A5
active output module 0(4)..20 mA (16 bit)
A6
active output module 4..20 mA (16 bit) (HART w. mod.H1)
A7
passive output module 4..20 mA (16 bit) (HART w. mod.H1)
B1..B5, E1
--
module not fitted
A7
pasive output module 4..20 mA (16 bit)
C1
RS232 module
D1
RS485 module
D2
communication loop module 0/20 mA
D3
M-Bus
G1
GSM
H1
Bell 202 modem HART compatible (only with module A6 or A7)
--
module not fitted
A7, B1..B5, E1
--
module not fitted
A7
pasive output module 4..20 mA (16 bit)
B1
binary output – passive optoMOS 250V(AD,DC) max.120mA max.4kHz
B2
binary output – passive optoMOS 60V(AD,DC) max.300mA max.10kHz
B3
binary output – active 5 VDC max.10mA max.12kHz
B4
binary output – active 24 VDC max.40mA max.12kHz
B5
binary output – relay 250VAC/1A
--
module not fitted
M1
memory module for data archiving
--
module not fitted
F1
electrochemical electrode cleaning module
F2
Full pipe check module
F3
module with F1+F2 functionality
S1
sensor signal input amplifier module (allways included)
0
Power supply 85-265 VAC
1
Power supply 24 V (18-36 VDC, 18-26 VAC)
2
Power supply 12 V (9-18 VDC, 9-14 VAC)
0
Compact version
1
Remote version
0
Unspecified (working) meter
1
Specified meter
B1..B5, E1
E1
binary input - active (for nonvolatile contacts or open collector)
Converter – marking and label
Tab. Converter marking
Fig. Converter label
Converter in remote version should be always completed with sensor with the same serial number!
FLOMAG 3000 -
Installation and Operation Manual
28
Loading...