QUICK START VP40_VP42-QS-en-US_AA
4. Hold the sensor tip (1) close to a voltage
source. The meter’s detection distance is 1
cm (0.4”) max.
5. If voltage is present, the device vibrates and
the alarm indicator (3) flashes red (low
sensitivity mode) or amber (high sensitivity
mode)
6. Long press Flashlight button (7) to switch
the light ON/OFF (Flashlight and detector
will not function at the same time)
7. The device has an APO (auto power off) of 3
minutes. The Flashlight APO is 30 minutes
8. Long press the power button to switch OFF
9. Refer to detailed user manual, available at
http://support.flir.com for more
information
Cleaning
With the meter OFF, wipe the housing with a damp
cloth as needed, do not use harsh detergents,
abrasives, or solvents. Use compressed air to clear
dust particles from the Flashlight lens. Clean the lens
with a commercial lens cleaner when necessary.
Ensure that the device is dry before performing tests.
Battery Replacement
With the meter OFF,
unscrew the battery
compartment cap
(8) and remove the
batteries.
Insert two new ‘AAA’ batteries in the proper
orientation.
Secure the compartment cover ensuring that the
red plastic ring in the
battery cap aligns with
the battery, as shown.
Check that the device
powers up correctly
before use.